INSTRUMENT PARAMETER (1)
INSTRUMENT PARAMETER (1)
Characteristics of
measurement systems
Solution:
%FSD = (4.35 psi/50 psi) 100 = 8.7
Solution:
Deviation at 50 lb = (50 2.5/100) lb = 1.25 lb
Deviation at 100 lb = (100 2.5/100) lb = 2.5 lb
Maximum deviation occurs at FSD, that is, 5 lb or 2.5%
FSD
Precision
• Precision is defined as the ability of instrument to
reproduce a certain set of readings within given
accuracy.
• Precision describes an instrument’s degree of
random variations in its output when measuring a
constant quantity.
• Precision depends upon repeatability.
The precision is obtained from the number of significant figures, in
which the reading is expressed and can be mathematically expressed
as
𝑋𝑛 − 𝑋𝑛
𝑃=1 −
𝑋𝑛
Where, P = precision Xn = Value of nth measurement 𝑋𝑛 = Average
value the set of measurement values
Precision is often confused with accuracy. High
precision does not imply anything about measurement
accuracy.
Low accuracy measurements from a high precision
instrument are normally caused by a bias in the
measurements, which is removable by recalibration
Accuracy Precision
• Accuracy represents degree • Precision represents degree
of correctness of the of repeatability of several
measured value with independent measurements
respect to true value. of desired input at the same
reference conditions
• Accuracy of instrument • Precision of instruments
depends on systematic depends on factors that
errors. cause random or accidental
errors.
Measuring a fixed target position
Low
precision,
low
accuracy
High High
precision, precision,
low High
accuracy accuracy
Repeatability
Repeatability is defined as ability of instrument to reproduce a group
of measurements of same measured quantity, made by same observer,
using same instrument, under same conditions.
Reproducibility
Describes the closeness of output readings for the same input when
there are changes in the method of measurement, observer,
measuring instrument, location, conditions of use and time of
measurement.
Both terms thus describe the spread of output readings for the same input. This
spread is referred to as repeatability if the measurement conditions are constant
and as reproducibility if the measurement conditions vary.
The degree of repeatability or reproducibility in measurements from an instrument
is an alternative way of expressing its precision.
Sensitivity of measurement
The sensitivity of measurement is a measure of the change in instrument
output that occurs when the quantity being measured changes by a given
amount. Thus, sensitivity is the ratio:
Static Sensitivity Change of Output Signal ∆𝑄𝑜
𝐾= =
Change in Input signal ∆𝑄𝑖
Thus, if the calibration curve is liner, as shown, the sensitivity of the instrument
is the slope of the calibration curve. If the calibration curve is not linear as
shown, then the sensitivity varies with the input. Inverse sensitivity or
deflection factor is defined as the reciprocal of sensitivity.
Example
The following resistance values of a platinum resistance thermometer
were measured at a range of temperatures. Determine the measurement
sensitivity of the instrument in ohms/C
solution
330
Sensitivity = slope
325
321 − 314
Resistance (Ω)
=
320
260 − 230
= 0.233 Ω/℃
315
310
305
0 50 100 150 200 250 300 350
T (°C)
Linearity
Linearity is usually reported as
It is normally desirable that the output non-linearity, which is the
reading of an instrument is linearly maximum of the deviation
proportional to the quantity being
between the calibration
measured.
Linearity is a measure of the curve and a straight line
proportionality between the actual value positioned so that the
of a variable being measured and the maximum deviation is
output of the instrument over its operating minimized.
range
Instrument Drift
• Drift is an undesired change in the output of a measured
variable over a period of time due to change in sensitivity of
the instrument.
• Drift may be caused by environmental factors, mechanical
vibrations, changes in temperatures, age, wear and tear, scale
formation, stray electric fields, stray magnetic fields, thermal
EMFs.
• Prime sources occur as chemical structural changes and
changing mechanical stresses.
Solution
It is then used in an environment
At 20C, deflection/load characteristic is a straight
at a temperature of 30°C and the
line. Sensitivity = 20 mm/kg.
following deflection/load
characteristic is measured.
At 30C, deflection/load characteristic is still a
straight line. Sensitivity = 22 mm/kg.
Stability:
It is the ability of an instrument to retain its performance
throughout is specified operating life.
Dynamic characteristics
Is the relationship between the system input and output
when the measured quantity (measurand) is varying
rapidly.
The dynamic characteristics of a measuring instrument
describe its behaviour between the time a measured
quantity changes value and the time when the instrument
output attains a steady value in response.