RO180 Manual_rev100827
RO180 Manual_rev100827
rev100827
Subject to alternations without prior notice.
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
Table of contents
1 General ....................................................................................................................................1
1.1 About this user manual 1
1.2 Symbols used in this manual 1
1.3 Liability and warranty 2
1.4 Copyright 2
1.5 Spare parts 2
1.6 Disposal 3
2 Safety.......................................................................................................................................3
2.1 Intended use 3
2.2 Contents of this manual 3
2.3 Changes and modifications to the equipment 4
2.4 User responsibility 4
2.5 Personnel requirements 4
2.6 Health and safety at work 5
2.7 Personal protective equipment (PPE) 5
2.8 Potential hazards 5
3 System features......................................................................................................................6
3.1 Purification process 7
3.2 Purification diagram 8
3.3 Security advice 8
4 Installation and start up.........................................................................................................9
4.1 Hose connections 9
4.2 Level control connection and power supply 10
4.3 Start up 10
5 System control .....................................................................................................................11
5.1 Display and function keys 11
5.2 First start 11
5.3 Settings 12
5.4 Energy saving mode 15
5.5 Maintenance 17
5.6 Service 18
6 Warning messages...............................................................................................................18
7 Maintenance and servicing .................................................................................................20
7.1 Adjustment of the operation pressure 20
7.2 Filter replacement 21
®
7.3 El-Ion electricity adjustment 22
7.4 El-Ion® nuts tightening 23
8 Technical data ......................................................................................................................24
8.1 Feed water requirements 24
8.2 System data 25
9 Appendix ..................................................................................................................................I
9.1 P&ID I
9.2 E.-drawing II
9.3 Consumables and spare parts III
Table of contents
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
1 General
1.1 About this user manual
This user manual describes how to use the equipment safely and correctly. When operating this equipment,
it is essential that the user adhere to the safety instructions and operating instructions, the local accident
prevention regulations and general safety regulations.
Before using the equipment, the user must read the user manual, particularly the section on safety and the
relevant safety warnings. The user must fully understand the instructions before proceeding.
The user manual is a part of the equipment. It must be kept close to the equipment and be accessible at any
time for all users. Should the equipment be passed on to a third party, the manual must also be passed on to
the third party.
This symbol alerts the user to potential sources of hazards to health, injuries or fatal injuries. Do not operate
the equipment until you have completely understood the instructions and proceed accordingly.
This symbol alerts the user to the presence of uninsulated voltage that has the potential to cause serious or
even fatal injuries. Do not operate the equipment until you have completely understood the instructions and
proceed accordingly. All maintenance work must be performed by qualified service personnel.
Important!
This symbol alerts the user to potential sources of malfunctions or damage to the equipment if operated
improperly.
Note!
This symbol alerts the user to advice on efficient und trouble-free operation.
General Page 1 of 25
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
The user manual must be kept close to the equipment and available at all times for anyone intending to use
the equipment.
The user manual must be read before any work is carried out using the equipment! The manufacturer does
not accept liability for damage or faults that arise due to disregard for the instructions in the manual.
The text and diagrams depicted in this manual do not necessarily match the equipment you have received.
The diagrams and images are not on a scale of 1:1.
The equipment delivered to you may be slightly different to the equipment described and depicted here due
to specialized features, additional options or due to changes in technology. If you have any queries, please
contact the manufacturer.
We reserve the right to make technical changes to product and further develop it in order to improve its
functionality.
1.4 Copyright
This manual must be treated as a confidential document. It is only intended for use by those operating the
equipment.
All contents of this manual – text, diagrams, images etc. – are protected by copyright law and industrial
property rights. Anyone found violating these rights will be prosecuted.
The manual may not be passed on to third parties, copied – either in part or whole – or used for any other
purpose without the express and written consent of the manufacturers. Anyone found violating these
copyright laws must pay compensation. The manufacturer may also prosecute.
Important!
The wrong or faulty spare parts can lead to damage, malfunction or a complete breakdown of
the equipment.
All guarantee, service, damages and third party liability claims against the manufacturer, the manufacturer’s
representatives and dealers become null and void once unauthorized spare parts are used in the equipment.
General Page 2 of 25
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
1.6 Disposal
If no return or disposal contract has been agreed on, please dispose of the disassembled components as
follows:
Important!
Electrical and electronic scrap and components, lubricants and other process materials must
be treated as hazardous waste and can only be disposed of by authorized waste disposal
facilities.
2 Safety
At the time this equipment was developed and produced, it complied with the current state-of-the-art
technology and is accepted as reliable and safe.
The equipment can represent a hazard if it is not operated by untrained personnel, if operated improperly or
not according to the manual.
The section on safety provides an overview of all the important issues relevant to personal safety and safe
and trouble-free operation.
The other sections of the manual also contain specific safety warnings and warning symbols that highlight
the possible hazards. Please also comply with the pictograms, signs, markings and inscriptions on the
equipment. These may not be removed and must always be visible.
Important!
Any form of use or operation other than the intended use is improper and prohibited.
Claims of any kind made against the manufacturer or the manufacturer’s representatives are null and void if
the equipment is not operated as intended.
The operating company is liable for all damages caused by improper use. The intended use is described in
the manual and includes compliance with the operating conditions. The equipment may only be operated
together with the parts supplied.
Reading and understanding the manual is the only way to protect against hazards, prevent mistakes and
operate the equipment safely and trouble-free.
We recommend that the operating company obtain written confirmation from each user that he/she has read
and understood the manual.
Safety Page 3 of 25
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
Ensure that all pictograms, signs, markings and inscriptions on the equipment are always visible and never
removed. Damaged, missing or illegible pictograms, signs, markings and inscriptions must be replaced
immediately.
The equipment may only be operated in a technically flawless and safe state. The equipment must be
checked for damage before each start-up.
The instructions in the manual must be followed completely and without restriction!
When operating this equipment, it is essential that the user adhere to the safety instructions and operating
instructions, the local accident prevention regulations, general safety regulations and the relevant
environmental protection legislation.
The operating company and the authorized personnel within that company are responsible for operating the
equipment trouble-free and must define who is responsible for installing, operating, servicing and cleaning
the equipment.
Authorized personnel are persons who sufficient training, knowledge, experience and understanding of the
relevant regulations to be able to assess the task given to him/her and identify potential hazards.
If a potential user does not have sufficient knowledge at his/her disposal, this person must be trained to be
qualified.
It must be clear who is responsible for equipment installation, operation, maintenance and servicing. These
competencies must be clearly defined and conformed to in order to prevent any uncertainties that may
compromise safety.
Only reliable personnel may work with and operate the equipment. Do not allow anyone to operate the
equipment in a way that compromises the safety of others or the environment or damages the equipment.
No one under the influence of drugs, alcohol or medication that impairs the ability to respond/react is allowed
to operate the equipment.
Please consider all other directives related to age and qualification when selecting potential users.
The user must ensure that other, unauthorized personnel keep a safe distance from the equipment.
Personnel must report to the operating company immediately if any changes become apparent which may
impair the safe operation of the equipment.
Safety Page 4 of 25
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
If these instructions are not observed, the equipment may become damaged or destroyed or may harm
persons.
The manufacturer is not liable for third party damages or compensation if the safety and operating
instructions in this manual or the local accident prevention regulations and general safety regulations are not
observed.
When working with this equipment, please wear goggles or another form of protective eyewear.
The equipment is safe for use, provided it is operated as intended. A minimal residual risk cannot be
eliminated completely.
Electrical energy in the equipment can lead to serious injuries. The equipment presents a
danger to life if the insulation or individual components become damaged.
- Switch off the main power switch before carrying out maintenance, cleaning or repair work and
ensure that it cannot be switched on again while carrying out this work.
- Eliminate all voltage from the equipment when working on the electrical equipment.
- Do not remove or deactivate any safety equipment.
The manufacturer does not take any responsibility for injuries to third parties
caused by improper use of the system.
Safety Page 5 of 25
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
3 System features
This water purification system is designed for the production of deionized Pure Water out of softened
drinking water from public supplies. Pure Water will be produced by a Reverse-Osmosis unit and the final
® ®
deionization by the El-Ion cell. The El-Ion electro-deionization removes last dissolved ions, particles and
®
bacteria from the water. The resin inside the El-Ion cell will be regenerated permanently by an electro-
chemical process and does not need to be exchanged. The system has to be connected to a level controlled
tank to regulate the production.
All components of the system are assembled in a plastic coated steel frame standing on wheels.
The side coverings can easily be opened for maintenance works. The system corresponds to the latest
technology and to recent safety regulations. It can be installed and put into operation without the help of
customer service engineers.
Pressure gauge
®
El-Ion electricity indicators
Fine Filter
Reverse-Osmosis module
®
El-Ion cell
®
Schematic diagram of the El-Ion cell. The conductivity will be decreased by an electro-chemical process
®
within the El-Ion cell. The ion exchange chamber provide for a salt removal from the water. Concentrate will
be drained. The conductivity value of the ultra pure water will be lower than 0.5 µS/cm.
Important!
The tank opening must be clearly marked with the enclosed fire warning labels.
No fire or fire causing devices are allowed near the unit and the tank.
Fire warning labels are enclosed this manual.
Before installing the system please check if the feed water meets the specification
demands. They are listed in chapter 8.1 - Feed water requirements
equirements.
Concentrate Outlet
Brine outlet
– Lead the CONCENTRATE and BRINE tubes to the waste water disposal (D=8mm)
Important!
Important!
The connector for the level control is located on the rear inner panel of the system.
Terminal block
Level sensor
connection
OPTION:
– Open the control cabinet door
– Stick the distribution cable / UV-Lamp cable through a free cable entrance rear side of the control
cabinet
– Tighten the wires on the terminal block in accordance to the circuit diagram
4.3 Start up
– Open feed water supply and check for impermeabilities of all connections
– Choose the First Start item within the Start Selection Screen
for the first water filling of the system after installation
Note!
Before Normal Operation it is highly recommended to study the chapter “System Control”
keenly.
5 System control
5.1 Display and function keys
Function keys
ON : System Start and Start Selection Screen
: Menu selection and number input
ENTER: Confirmation
: Menu selection and number input
Menu : Superior menu item
OFF : Switch off key (Attention: the system is still power connected)
Flush
The modules will be rinsed for 120 seconds.
After the flushing program the system will switch into normal opration
mode.
5.3 Settings
ON ► Start selection screen ► Main Menu
Confirmation starts the main menu or simply press the menu key during
normal operation.
Main Menu
– Settings: System configuration
– Maintenance: Operation counter and Service Operation
– Service: Technical service configuration
Note:
The timer P5.1 will only be displayed, if external pump 5.1 is switched on
5.5 Maintenance
For some changes in the maintenance menu a password is required.
The user password consists of the first four ciphers of the serial number.
5.6 Service
For some changes in the maintenance menu a password is required.
Note: Only change any settings after consultation with the customer
service!
In the service menu both I/O-boards, the flow transmitter and the tank can
be calibrated, settings concerning the tank can be adjusted.
The extended menu is requires a service password. It is allotted to service
technicians only.
6 Warning messages
Warning Meaning Measure
1: Q1 The measured feed water conductivity Check the feed water pretreatment.
above setpoint QIA+Q1 exceeded a set limit. Check the feed water conductivity.
2: Q2 The measured permeate conductivity If the measured conductivity of the RO
above setpoint QIA+Q2 exceeded a set limit. permeate is too high, the RO modules
must be replaced.
3: Q3 The measured electro-deionized If the measured conductivity of the
above setpoint permeate conductivity QIA+Q3 electro deionized permeate is too high,
®
exceeded a set limit. the functionality of the El-Ion cell must
be tested.
Inform the customer service.
5: concentrate RO The measured flow rate F3.2 of the Check the feed water supply.
below alarm value concentrate exceeded a set limit. Check hoses for any constrictions.
Check the functionality of the RO
pump P3.1
6: temperature The measured product water Check the settings of the temperature
above setpoint temperature TA+4.1 exceeded a set limit.
limit. Check the temperature of the ambient
temperature and of the product water.
8: feed pressure The feed water pressure measured by Raise the feed water pressure.
below alarm value PSA-3.1 is below the set limit. Change the filter cartridge F3.1.
9: system stop System is stopped due to an operating Reset the error by pressing the on key.
start with on key fault. This warning appears together
with other warning messages.
10: tank is empty The product water tank is empty. The Check the feed water supply.
level below setpoint circulation pump is stopped to protect it Check the functionality of the RO
against running dry. process.
11: cable fault A fault in the tank measurement cable Check the cable of the tank level
tank measurement has been detected. The tank level measurement and its connection to the
cannot be displayed correctly. system.
12:attention The service operation mode has been Initiate necessary actions in case of a
service operation started, which means that the system tank overflow.
produces product water permanently.
Tank levels will be ignored.
13: Maintenance The system is maintained by the
by customer service customer service.
14: ext. fault 1 An external device signals a fault. Analyze the external fault signal.
start with on key
15: failure controller A problem of the power supply of the Inform the customer service.
®
El-Ion cells El-Ion cell has been detected.
16: replace The maximum operating hours of the Replace the UV lamp 5.1 of the tank.
ext. UV lamp 5.1 external UV lamp 5.1 are exceeded. Reset the expired operating hours of
the UV lamp.
®
17: concentrate flow The flow rate of the El-Ion cell Check the flow rate of the concentrate.
below setpoint concentrate is below the set limit. Check the limit settings for FTA-4.1.
Check hoses for any constrictions.
If the problem maintains, inform the
customer service.
18: product flow The flow rate of the product water is Check the flow rate of the product
below setpoint below the set limit. water.
Check the limit settings for FTA-3.1.
Check the function of the RO process.
Check the RO concentrate flow rate
FTA-3.2.
Check the functionality of the RO
pump P3.1.
Check the feed water supply.
Check hoses for any constrictions.
If the problem maintains, inform the
customer service.
20: con measurement All four conductivity measurements are Check if air is located the
out of range displayed in an accumulative alarm. measurement chamber and/or if there
On the display an arrow indicates the is a cable break.
direction in which the conductivity Deaerate the system.
exceeded the range (LF1 – LF4).
The replacement of the fine filter (F3.1) has to be chosen individually and depends on the contamination of
the feed water.
In case of feed water with potable water quality, a replacement of the filter is recommended all 3 months.
Filter attachment
Filter body
– Wait for the pressure reduction indicated on the pressure gauge (PI3.1)
– Unscrew the filter case anticlockwise with the supplied filter gripper
– Check for impermeability of all connections and close the left cabinet side again
– System is ready
The current is factory preset to 8 A Cell 1 and 4 A Cell 2. The voltage of the cells depends on the feed water
quality.
The adjustment potentiometer is located on the aluminum plate mounted on the ground of the electric
cabinet.
Control cabinet
®
EL-ION Potentiometer (blue)
®
El-Ion circuit board
Important!
®
The nuts of the El-Ion cell must be tightened every 12 months with a torque wrench.
®
EL-ION cell
Nuts and counter nuts
– Stop the system with the OFF function key and turn OFF the feed water supply
– Hold the counter nuts with a wrench, tighten all nuts with a torque wrench one by one for 2 or 3
times
– Plug in power supply cable, turn on water supply, start the system with the ON function key
8 Technical data
8.1 Feed water requirements
Before installation, please check if feed water fulfills following specification:
pH 3–9
°C 5 – 35
Temperature
°F 45 – 90
All given flow rates are based on feed water temperature of 15°C (59°F).
The RO permeate flow will change approx. 3% for each °C (decrease for lower values, increase for higher
values).
Important!
Weight kg 150
9 Appendix
9.1 P&ID
Appendix I
1 2 3 4 5 6 7 8
PSA- 13[bar]
input: 4.5 to 24 VDC; 12mA
output: 13-186Hz;
3.1 0-232[psi]; 0-16[bar]
art.00279-1
art.07377-5 cr.(delay 300s)
PI
distribution system
0,2[µS/cm]; (0-0,3/3µS/cm)
FTA-
cr.(delay 300s)
QIA+ QIA+
Attention!
20[µS/cm]
Q3
female thread
female thread
Q1
output 1/4"
input 1/4"
B art.02501 B
TA+
24VDC; 0,15A)
art.03352-6
QISA+
24VDC; 0,15A)
4.1
D=6mm;
1/4"male
F3.2 Q2
3/4"male
RO Pass 1 3/4"BSP
3/4"female 3/4"male
3/4" female DN12; L/- M 3/4"BSP
hose DN10; L/-
3/4"male
3/4"female 3/4"male - + - 3/4"male
DN10; L/-
softener 0,88[GPM];
4,01[GPM]; 3/4"; d32 0,792[GPM];
art.04990-1 200[l/h] art.04990-1
910[l/h] DN25
P3.1 V3.2
180[l/h]
art.08550
V3.1 3/4"BSP
hose
3/4"BSP
hose
art.03350 V3.5 3/4"female
DN10; L/-
820 mm
0-100[PSI];
V4.1
3/4"BSP 24VDC; 0,15A) 3/4"BSP
male thread 3/4" 0-6,8[bar] 3/4" male thread
art.04990-1
femal male
3/4"
JG;8mm
art.03242-2 DN10; L/- 3/4"male
1300 mm
output 1/8"male
E4.1
art.03175
{single filter housing}
input 1/8"male
C C
{art.01224-2 wrench}
V3.4 Two-Pass EDI
DN10; L/L art.16250
150 mm
2,42[GPM]; 3/8"female
200[l/h]
FTA-
0,079[GPM]; 18[l/h]
3/4"BSP
4.1
Recovery=90%
hose
1/4" flex tube art.06533
800 mm; black 0,9-60[l/h];
0,004-0,264[GPM]
art.03161-1 cr.(delay 120s)
0,062[GPM]; 14[l/h]
V3.7 n.cr.(delay 120s)
xx[GPM]; xxx[l/h] V3.6
If Recovery>93%
Recovery: 60%
0,6[GPM];
FTA-
input 1/4"
135 [l/h]
female thread
min.2,5cm; 1 inch
3/4"BSP
male thread
free entrance
Feed water specification
V3.1 solenoid valve {24VDC, NC; art. 04990-1} Feed water pressure 2,0-6,0 bar
V3.2 solenoid valve {24VDC, NC; art. 04990-1} 29-87 psi
V3.3 solenoid valve {24VDC, NC; art. 04990-1} 0,34 m³/h
V3.4 check valve {art. 03175} Feed water flow 1,5 GPM
V3.5 flow control valve {art. 02742}
V3.6 adjustable flow restriction Feed water conductivity < 1400 µS/cm
E V3.7 ball valve {only for service operation{art. 03161-1} Free Clorine < 0,1 mg/l
CO2 max. < 20 mg/l
Scope of supply
F3.1 active carbon filter {with 1µm screen}
{cartridge art. 08550} Silt Density Index (SDI) <3
{wrench art. 01224-2} Silica < 15 mg/l
F3.2 RO Pass 1 {1x modul TW 30 - 4040; art. 03205} Iron (FE) < 0,1 mg/l Designed for: DO NOT SCALE DRAWING REVISION 05
Manganese (Mn) < 0,05 mg/l
Siemens Water Technologies ID.No.: 3055
P3.1 RO-pump {brass; clamp type; RO pass1} pH-Range 3-9
standard-catalog-system
V4.1 adjustable flow restriction Langelier-Index <0 (negativ)
No. Changing Date NAME NAME DATE TITLE:
V4.2 tap valve 5 -35 °C
Temperature 01 from AutoCAD to Solidw. 09.12.2008 BC
45 - 90 °F DRAWN BC 09.12.2008
PSA- pressure switch {switch & alarm function; min. detection}
The Langelier-Index has to be negativ
02 Conversion to SolidWorks 26.02.2009 TK
CHK'D TK 20.08.2010 Protegra CS™ RO EDI 180
PI pressure gauge {indication}
QI conductivity meter {indication}
03 Layer added 16.11.2009 BC (flow sheet_P&ID)
Attention! all given flow rates are based on feed water
F QISA+ conductivity meter {switch & alarm function; max. detection} temperature of 59[F]; 15°[C]
04 new siemens logo 08.06.2010 BC Dry weight: 120 [kg]
TA+ temperature sensor {alarm function; max. detection} The RO-permeate flow will change approx. 3% for each
FTA- flow turbine {with alarm generation; min. detection} degree C (drop for lower values; increase for higher values
05 new product names 20.08.2010 BC Siemens AG
I IS WT IND HPS GER
DWG NO. A3
05 Flow-turbines
F3.2 geändert v.
20.08.2010 BC Fahrenberg 8
22885 Barsbüttel, Deutschland
3055_P&ID_Rev05 page
05 art.03205-11 in 03205 20.08.2010 BC
Tel.: +49 (40) 670868-6
1/1
1 2 3 4 Fax: +49 (40) 670868-44 Order-No.: Customer-PO-No.:
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
9.2 E.-drawing
Appendix II
TM
Protegra CS RO EDI 180
Reverse-Osmosis system
®
with 2-pass El-Ion Electro-Deionization
Catalogue No. 3055
Appendix III