Data Processing Manual
Data Processing Manual
DATA PROCESSING
USER'S MANUAL
DATA PROCESSING
USER'S MANUAL
・ The information described in this manual, and the specifications and contents of the software described in this manual are
subject to change without prior notice due to the ongoing improvements made in the instrument.
・ In no event shall JEOL RESONANCE be liable for any direct, indirect, special, incidental or consequential damages, or any
other damages of any kind, including but not limited to loss of use, loss of profits, or loss of data arising out of or in any way
connected with the use of the information contained in this manual or the software described in this manual. Some countries do
not allow the exclusion or limitation of incidental or consequential damages, so the above may not apply to you.
・ This manual and the software described in this manual are copyrighted, all rights reserved by JEOL RESONANCE and/or
third-party licensors. Except as stated herein, none of the materials may be copied, reproduced, distributed, republished,
displayed, posted or transmitted in any form or by any means, including, but not limited to, electronic, mechanical,
photocopying, recording, or otherwise, without the prior written permission of JEOL RESONANCE or the respective copyright
owner.
・ When this manual or the software described in this manual is furnished under a license agreement, it may only be used or
copied in accordance with the terms of such license agreement.
TRADEMARK
・ Windows is a trademark of Microsoft Corporation.
・ All other company and product names are trademarks or registered trademarks of their respective companies.
CONTENTS
1 FUNDAMENTALS OF DELTA
1.1 STARTING AND EXITING DELTA................................................. 1-1
1.2 COMPONENTS OF WINDOWS ...................................................... 1-3
1.2.1 1D Processor Window .................................................................... 1-3
1.2.2 Data Slate Window......................................................................... 1-4
1.2.3 nD Processor Window .................................................................... 1-4
1.2.4 2D Viewer Window........................................................................ 1-5
1.2.5 3D Viewer Window........................................................................ 1-5
1.2.6 3D Slicer Window .......................................................................... 1-6
1.2.7 Curve Analysis Window ................................................................ 1-6
1.3 POINTER BAR .................................................................................. 1-7
1.3.1 Changing Mode by Selecting Button ............................................. 1-8
1.3.2 Button to Display Key-Commands ................................................ 1-9
1.3.3 Buttons to Select Mode ................................................................ 1-11
1.4 CHANGING THE DISPLAY OF A SPECTRUM .......................... 1-17
1.4.1 Expanding a Spectrum (Zoom) .................................................... 1-17
1.4.2 Moving a Spectrum ...................................................................... 1-19
1.4.3 Expanding or Reducing a Spectrum (Zoom + Modifier 1) .......... 1-21
1.4.4 Changing the Y-Axis Gain ........................................................... 1-24
1.5 PROCESS LIST ............................................................................... 1-25
1.5.1 Process List .................................................................................. 1-25
1.5.2 Process List Composition ............................................................. 1-27
1.5.2a Process list.............................................................................. 1-27
1.5.2b Process control buttons .......................................................... 1-28
1.5.3 Creating and Editing a Process List ............................................. 1-29
1.5.3a Appending a process to the process list ................................. 1-29
1.5.3b Inserting a process in the process list ..................................... 1-29
1.5.3c Replacing a process in the process list ................................... 1-30
1.5.3d Moving the process ................................................................ 1-30
1.5.3e Deleting the process you select .............................................. 1-30
1.5.3f Changing the process parameters ........................................... 1-31
1.5.4 Deleting All the Processes............................................................ 1-33
1.5.5 Loading the Process List File ....................................................... 1-34
1.5.5a Loading the process list file from the hard disk ..................... 1-34
1.5.5b Loading the process list file from another window on
the screen ............................................................................... 1-35
1.5.6 Saving the Process List................................................................. 1-36
1.5.7 Executing the Process List ........................................................... 1-37
1.6 LOADING AND SAVING DATA ................................................... 1-39
1.6.1 Loading a Data File ...................................................................... 1-39
1.6.1a Loading a data file to the window
which automatically selected ............................................... 1-39
1.6.1b Loading a data file to the specified window .......................... 1-41
1.6.1c Searching for data .................................................................. 1-42
1.6.2 Saving NMR Data ........................................................................ 1-46
1.6.2a Saving measurement data....................................................... 1-46
1.6.2b Saving processed data ............................................................ 1-47
NMECZ-USP-1 C-1
1.7 CONVERSION OF NMR DATA ..................................................... 1-50
1.7.1 Data Formats Supported by Delta................................................. 1-50
1.7.2 Importing Data.............................................................................. 1-51
1.7.3 Exporting Data.............................................................................. 1-52
2 1D NMR DATA PROCESSING
2.1 1D PROCESSOR WINDOW ............................................................. 2-1
2.2 1D DATA PROCESSING................................................................... 2-5
2.2.1 Canceling the DC Offset................................................................. 2-5
2.2.2 Zero-Filling..................................................................................... 2-6
2.2.3 Window Function ........................................................................... 2-7
2.2.3a Setting a window function........................................................ 2-7
2.2.3b Changing the parameters for the window function .................. 2-8
2.2.3c Changing the window function ................................................ 2-9
2.2.3d Setting multiple window functions......................................... 2-10
2.2.4 Fast-Fourier Transformation ......................................................... 2-11
2.2.5 Phase Correction ........................................................................... 2-12
2.2.5a Automatic phase correction .................................................... 2-12
2.2.5b Manual phase correction ........................................................ 2-12
2.2.6 Baseline Correction ...................................................................... 2-15
2.2.6a Automatic baseline correction ................................................ 2-15
2.2.6b Manual baseline correction .................................................... 2-15
2.2.7 Unit for Axis ................................................................................. 2-18
2.2.8 Reference Setting .......................................................................... 2-18
2.2.8a Setting the reference using the pointer bar ............................. 2-18
2.2.8b Setting the reference using the Reference command.............. 2-19
2.2.9 Peak Detection .............................................................................. 2-20
2.2.9a Automatic peak detection ....................................................... 2-20
2.2.9b Manual peak detection ........................................................... 2-23
2.2.10 Integration..................................................................................... 2-26
2.2.10a Automatic integration ............................................................. 2-26
2.2.10b Manual integration ................................................................. 2-26
2.2.11 Transfer Integral Tool Window ..................................................... 2-30
2.2.12 Comment ...................................................................................... 2-31
2.2.13 PiP (Picture in Picture) ............................................................... 2-33
2.2.14 Setting Display Range of Spectrum.............................................. 2-34
2.2.15 Plotting ......................................................................................... 2-36
2.3 DECONVOLUTION ........................................................................ 2-37
3 2D NMR DATA PROCESSING
3.1 nD PROCESSOR WINDOW ............................................................. 3-1
3.1.1 nD Processor Window .................................................................... 3-1
3.1.2 Domain ........................................................................................... 3-3
3.1.3 Process List ..................................................................................... 3-3
3.1.4 Process Levels ................................................................................ 3-4
3.2 2D DATA PROCESSING................................................................... 3-5
3.2.1 Zero-Filling..................................................................................... 3-5
3.2.2 Window Function ........................................................................... 3-6
3.2.2a Setting a window function........................................................ 3-6
3.2.2b Changing the parameters for the window function .................. 3-7
3.2.2c Changing the window function ................................................ 3-8
3.2.2d Setting multiple window functions........................................... 3-9
3.2.2e Checking the window function using 1D slice data ............... 3-10
3.2.3 Symmetrization............................................................................. 3-12
3.2.4 Adjusting Levels of the Contour Lines (1) ................................... 3-13
3.2.5 Adjusting Levels of the Contour Lines (2) ................................... 3-16
C-2 NMECZ-USP-1
3.2.6 Projection ..................................................................................... 3-17
3.2.6a Whole range projection .......................................................... 3-17
3.2.6b Partial projection .................................................................... 3-17
3.2.6c Projection modes .................................................................... 3-17
3.2.7 Loading and Processing 1D Slice Data ........................................ 3-18
3.2.7a Displaying slice data using the process list ............................ 3-18
3.2.7b Displaying slice data in the 2D Viewer window .................... 3-19
3.2.8 Pasting a High-Resolution 1D NMR Spectrum ............................ 3-20
3.2.8a Pasting data from a disk ......................................................... 3-20
3.2.8b Pasting the displayed data ...................................................... 3-21
3.2.8c Pasting data displayed in another window ............................. 3-22
3.2.9 Reference Setting ......................................................................... 3-23
3.2.10 Peak Pick ...................................................................................... 3-24
3.2.10a Setting the J value .................................................................. 3-25
3.2.10b Adjusting the threshold and base levels ................................. 3-25
3.2.10c Displaying the results of peak pick in the Spread
Sheet window ......................................................................... 3-26
3.2.11 Data Processing of Phase Sensitive 2D Data ............................... 3-26
3.2.11a Phase-correction processing ................................................... 3-26
3.2.11b Positive and negative peak display control ............................ 3-31
3.2.12 Eliminating the T1 Noise ............................................................. 3-33
3.2.13 Other Processing .......................................................................... 3-34
4 ARRAY DATA PROCESSING
4.1 OUTLINE OF ARRAY MEASUREMENT
DATA PROCESSING ........................................................................ 4-1
4.2 PROCESSING METHOD FOR
ARRAY MEASUREMENT DATA .................................................... 4-2
4.2.1 Processing All Data as a Single Data File ...................................... 4-2
4.2.1a When process conditions are saved in the process list ............. 4-2
4.2.1b When the process list is not provided ...................................... 4-3
4.2.1c Displaying arbitrary data in the Data Slate window ................ 4-7
4.2.2 Processing arbitrary Data taken from the Measurement Data ........ 4-8
5 ADDING AND SUBTRACTING THE DATA
5.1 DISPLAYING THE OPERATE ON FILES WINDOW .................... 5-1
5.2 ADDING AND SUBTRACTING
THE MEASUREMENT DATA (FID) ............................................... 5-2
5.3 ADDING OR SUBTRACTING THE PROCESSED DATA ............. 5-5
6 BATCH PROCESSING
6.1 DISPLAYING THE BATCH PROCESS FILES WINDOW ............. 6-1
6.2 BATCH PROCESSING OF MEASUREMENT DATA (FID)........... 6-2
7 COMBINING THE 1D NMR DATA SETS
7.1 COMBINE FILES WINDOW ........................................................... 7-1
7.2 COMBINING THE 1D NMR DATA SETS....................................... 7-1
8 DATA SLATE WINDOW
8.1 DATA SLATE WINDOW .................................................................. 8-1
8.2 OPENING THE DATA SLATE WINDOW ....................................... 8-2
8.3 DISPLAYING ADDITIONAL PROCESSED DATA ........................ 8-3
8.3.1 Displaying Additional Data in Another Geometry......................... 8-4
8.3.2 Overlaying Additional Data in a Single Geometry ........................ 8-5
8.4 CONTROLLING THE DISPLAY ..................................................... 8-6
8.4.1 Display of Multiple Geometries ..................................................... 8-6
8.4.2 Controlling Display of Multiple Data in a Geometry..................... 8-9
8.5 PEAK PICKING AND INTEGRATION ......................................... 8-12
NMECZ-USP-1 C-3
9 SPREAD SHEET
9.1 SPREAD SHEET WINDOW ............................................................. 9-1
9.2 DISPLAYING RESULTS OF PEAK PICKING
AND INTEGRATION ........................................................................ 9-2
9.3 PLOTTING THE SPREADSHEET ................................................... 9-4
9.4 SAVING THE SPREADSHEET AS A CSV FILE ............................ 9-4
9.5 USING CALCULATOR PANEL ....................................................... 9-5
10 3D NMR DATA PROCESSING
10.1 DISPLAY OF 3D DATA .................................................................. 10-1
10.1.1 3D Viewer Window ...................................................................... 10-1
10.1.2 Data Slate Window ....................................................................... 10-2
10.2 SLICE OPERATION OF 3D DATA ................................................. 10-3
10.2.1 How to Use 3D Slicer ................................................................... 10-3
11 HOW TO USE PROCESSING FUNCTIONS
11.1 PRE TRANSFORM.......................................................................... 11-1
11.2 WINDOW......................................................................................... 11-3
11.3 TRANSFORM .................................................................................. 11-4
11.4 POST TRANSFORM ....................................................................... 11-5
11.5 DISPLAY.......................................................................................... 11-7
12 DISPLAY FUNCTIONS OF PROCESSING DATA
12.1 POP UP MENU IN 1D DATA SCREEN ......................................... 12-1
12.1.1 Pop-Up Menu ............................................................................... 12-1
12.1.2 Setting Each Item ......................................................................... 12-4
12.2 POP-UP MENU IN 2D DATA SCREEN ......................................... 12-8
12.2.1 Pop-Up Menu ............................................................................... 12-8
12.2.2 Setting Each Item ....................................................................... 12-11
12.3 POP-UP MENU IN LEVEL TOOL ............................................... 12-16
12.3.1 Pop-Up Menu ............................................................................. 12-16
12.3.2 Setting Each Item ....................................................................... 12-17
INDEX
C-4 NMECZ-USP-1
FUNDAMENTALS OF
1
DELTA
1.1 STARTING AND EXITING DELTA............................................ 1-1
1.2 COMPONENTS OF WINDOWS.................................................. 1-3
1.2.1 1D Processor Window .................................................................... 1-3
1.2.2 Data Slate Window......................................................................... 1-4
1.2.3 nD Processor Window .................................................................... 1-4
1.2.4 2D Viewer Window........................................................................ 1-5
1.2.5 3D Viewer Window........................................................................ 1-5
1.2.6 3D Slicer Window .......................................................................... 1-6
1.2.7 Curve Analysis Window ................................................................ 1-6
1.3 POINTER BAR ............................................................................... 1-7
1.3.1 Changing Mode by Selecting Button ............................................. 1-8
1.3.2 Button to Display Key-Commands ................................................ 1-9
1.3.3 Buttons to Select Mode ................................................................ 1-11
1.4 CHANGING THE DISPLAY OF A SPECTRUM..................... 1-17
1.4.1 Expanding a Spectrum (Zoom) .................................................... 1-17
1.4.2 Moving a Spectrum ...................................................................... 1-19
1.4.3 Expanding or Reducing a Spectrum (Zoom + Modifier 1) .......... 1-21
1.4.4 Changing the Y-Axis Gain ........................................................... 1-24
1.5 PROCESS LIST ............................................................................ 1-25
1.5.1 Process List .................................................................................. 1-25
1.5.2 Process List Composition ............................................................. 1-27
1.5.2a Process list .............................................................................. 1-27
1.5.2b Process control buttons ........................................................... 1-28
1.5.3 Creating and Editing a Process List ............................................. 1-29
1.5.3a Appending a process to the process list .................................. 1-29
1.5.3b Inserting a process in the process list ...................................... 1-29
1.5.3c Replacing a process in the process list .................................... 1-30
1.5.3d Moving the process ................................................................. 1-30
1.5.3e Deleting the process you select ............................................... 1-30
1.5.3f Changing the process parameters............................................ 1-31
1.5.4 Deleting All the Processes............................................................ 1-33
1 FUNDAMENTALS OF DELTA
1.b NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
NMECZ-USP-1 1-1
1 FUNDAMENTALS OF DELTA
1-2 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Output data
NMECZ-USP-1 1-3
1 FUNDAMENTALS OF DELTA
1-4 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
NMECZ-USP-1 1-5
1 FUNDAMENTALS OF DELTA
1-6 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Pointer bar
Fig. 1.3 Pointer bar in the display area for processing data
Clicking each of the buttons at the left side of the pointer bar selects its mode. At the right
side of the pointer bar are the indicators of the modifier functions and the button to display
key-commands. The indicators of modifier state change depending on their functions.
NMECZ-USP-1 1-7
1 FUNDAMENTALS OF DELTA
The two buttons in the tool-action group do not appear in the 1D Processor window.
1-8 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
You can see the state of the modifier function and accelerator function on the pointer bar.
While pressing the key according to the function, the function is extended.
● Modifier functions
Pressing the Alt or Shift key extends/limits the function of the mode that is being
selected on the pointer bar. The function of the modifier changes according to the selected
mode.
If your computer is a Macintosh, press the Option or Shift key.
If you press the two keys at one time, both functions take effect.
● Accelerator functions
During pressing the Ctrl key, the specific mode that was selected on the pointer bar
changes to the other mode.
If your computer is a Macintosh, press the Command key.
Also, during pressing the Space key, all modes change to the Zoom mode.
NMECZ-USP-1 1-9
1 FUNDAMENTALS OF DELTA
1-10 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
■ Zoom mode
■ Pan mode
Selected Overlay Among the spectra overlaid in the Data Slate window,
(Modifier 1) this button moves only the spectrum you specify.
Constrain Specifies the direction to expand in either the X- or
Direction Y-axis direction.
(Modifier 2)
For information on operating the Pan mode, refer to Sect. 1.4, “CHANGING THE
DISPLAY OF A SPECTRUM”.
■ Gain mode
The Gain mode increases or decreases the intensity of the spectrum peaks.
Button/indicator Mode Function
Gain Increases/decreases the intensity of the spectrum peaks.
Selected Overlay Among the spectra overlaid in the Data Slate window,
(Modifier 1) this button increases/decreases the gain of only the
spectrum you specify.
NMECZ-USP-1 1-11
1 FUNDAMENTALS OF DELTA
■ Select mode
Smart object selection Selects an object of the mode you used last time.
■ Phase mode
Constrain Aspect Ratio Reads and copies the square area you specify.
(Modifier 2)
Copy Amplitude Copies the intensity of the peak you read.
If the mode of the Copy Position button changes, the function of the modifier function
also changes.
This mode does not have a function for Modifier 1.
1-12 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
■ Threshold mode
The Reference mode sets up the reference of the chemical shift for the axis marker.
Button/indicator Mode Function
Paste reference Pastes the reference on the peak that is nearest to the
specified position.
Selected Overlay Among the spectra overlaid in the Data Slate window, this
(Modifier 1) button pastes the reference on the spectrum you specify.
No-Snap-to-Peak Pastes the reference on the position you specify, regardless
(Modifier 2) of the height of the threshold level.
You can set the reference value using the Options menu.
Reference”.
For information on operating the Reference mode, refer to Sect. 2.2.8, “Setting a
■ Peak mode
In the Peak mode, pressing the Ctrl key changes the mode to the Select mode.
(Accelerator function)
For information on operating the Peak mode, refer to Sect. 2.2.9, “Peak Detection”.
NMECZ-USP-1 1-13
1 FUNDAMENTALS OF DELTA
■ Integral mode
In the Integral mode, pressing the Ctrl key changes the mode to the Select mode.
(Accelerator function)
For information on operating the Integral mode, refer to Sect. 2.2.10, “Integration”.
■ Measure mode
In the Measure mode, pressing the Ctrl key changes the mode to the Select mode.
(Accelerator function)
■ Cursor mode
In the Cursor mode, pressing the Ctrl key changes the mode to the Select mode.
(Accelerator function)
1-14 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
■ Annotation mode
If the mode of the Annotation mode changes, the functions of the modifier keys also
change.
In the Annotation mode, pressing the Ctrl key changes the mode to the Select mode.
(Accelerator function)
The lines drawing by "Create Line" can be changed to arrows using right-click menu.
■ Molecule mode
In the Molecule mode, pressing the Ctrl key changes the mode to the Select mode
(Accelerator function).
NMECZ-USP-1 1-15
1 FUNDAMENTALS OF DELTA
■ PiP mode
The PiP mode selects a partial range of the spectrum and displays it in the same geometry.
Button/indicator Mode Function
Create PiP Creates a partial spectrum (PiP box) from the whole
spectrum range. Or, this button selects a PiP box that
already exists.
Force Create Creates a PiP box by specifying a range. This button does
(Modifier 1) not select a PiP box that already exists.
Constrain-Aspect-Ratio Creates a square PiP box.
(Modifier 2)
■ Region mode
■ Pick mode
The Pick mode performs the slice operation of two-dimensional data and reading of a peak
position.
Button/indicator Mode Function
Pick Position Specifies the peak position.
The vertical range you dragged will be expanded in the Y-axis direction.
NMECZ-USP-1 1-17
1 FUNDAMENTALS OF DELTA
The horizontal range you dragged will be expanded in the X-axis direction.
The area you dragged will be expanded in the X- and Y-axis directions.
1-18 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
NMECZ-USP-1 1-19
1 FUNDAMENTALS OF DELTA
1-20 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Expansion
Reduction
NMECZ-USP-1 1-21
1 FUNDAMENTALS OF DELTA
Reduced Expanded
1-22 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Expansion
Reduction
NMECZ-USP-1 1-23
1 FUNDAMENTALS OF DELTA
Gain increase
Gain decrease
1-24 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Process list
The process list, in which contents of standard data processing are described for typical
measurement data, is supplied. This process list is called the “standard process list”.
The standard process list is saved in the Global Process List directory.
NMECZ-USP-1 1-25
1 FUNDAMENTALS OF DELTA
Moreover, the standard process list suiting the measurement is set to a typical measurement
sequence. When measurement was complete, data will upload to a workstation, standard
processing set up will be performed automatically.
Thus, fundamental data processing for Delta will perform the process list set up. Moreover,
changing the data processing conditions described in the process list will perform a change
of the processing conditions.
The processing contents can be changed, or the newly created process list can also be saved
to a file as another process list. When performing data processing with the same conditions,
processing becomes possible by an easy operation that reads the purposed process list.
In the 1D Processor window, click the Processing button at the right side of the window to
show or hide the process list.
In the nD Processor window, click the button to display the process list.
For handling of the process list in the nD Processor window, refer to Sect. 3.1, “nD
PROCESSOR WINDOW”.
1-26 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Input
window
Output
window
Delete
Cut Restore
Paste Clear
NMECZ-USP-1 1-27
1 FUNDAMENTALS OF DELTA
You can use both Cut and Paste buttons only in the same process list window.
In the 1D Processor window, the display/phase process is always the last process in
the process list, and you cannot delete it.
1-28 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
1. In the 1D Processor window, click the button to enter the Append mode.
2. Select the process that you want to append using the tool bar or the menu bar.
The process you select is appended before display/phase.
1. Click the button in the nD Processor window to enter the Append mode.
2. Click the process list column of the axis to which you want to append the
process.
Click the process list column of the axis
to which you want to append.
3. Select the process you want to append using the tool bar or the menu bar.
The process you select is appended before transpose.
NMECZ-USP-1 1-29
1 FUNDAMENTALS OF DELTA
1-30 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
1. In the process list, click the button in the left of the process whose
parameter you want to change.
The parameter and its value used in the selected process are displayed on the
parameter display box or the check box under the process list display area.
Clicking the button shows all the processes.
Click the
button to
show its
process.
Parameter
box
Check
box
NMECZ-USP-1 1-31
1 FUNDAMENTALS OF DELTA
■ nD Processor window
1. In the process list, click the button in the left of the process whose
parameter you want to change.
The parameter and its value used in the selected process are displayed on the
parameter display box or the check box under the process list display area.
Clicking the button shows all the processes of the selected axis.
Click the
button to
show its
process.
Parameter
box
Check
box
1-32 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
1. Click the process list column of the axis from which you want to delete the
process.
Click the process list column of the axis
from which you want to delete.
NMECZ-USP-1 1-33
1 FUNDAMENTALS OF DELTA
1.5.5a Loading the process list file from the hard disk
1. Click the button.
1-34 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
1.5.5b Loading the process list file from another window on the screen
After right-clicking the process list file in another window, drag and drop it to the
process list display area where you want to paste it.
If you do not quickly drag and drop the process list file after you right-click it, the
context menu may appear so that you cannot copy the file.
NMECZ-USP-1 1-35
1 FUNDAMENTALS OF DELTA
Filename box
1-36 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
While the program executes the process list, the letter “Abort” appears on the
button.
The program processes the data according to the process list, and displays the resultant
spectrum in the data display area.
● Reference: Checking if the program has finished executing the process
list
If display/phase at the end of the process list is shown in green, all the processing in the
process list has finished.
If you click the button, the program shows the process that has just finished in
bright green, and shows its spectrum.
NMECZ-USP-1 1-37
1 FUNDAMENTALS OF DELTA
If all the processing in the process list has finished, this function does not work.
■ In the nD Processor window
1-38 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Version display
box
File list
Data information
box
NMECZ-USP-1 1-39
1 FUNDAMENTALS OF DELTA
Revision_time
If you want to specify the window to load the data, refer to Sect. 1.6.1b.
■ Version number
Single_pulse – 1 – 1 . jdf
Before processing the data set, the program automatically makes a data file of a version
different from the original one with the same filename, and then processes the data set. The
original data set are not processed, preventing corruption of the original FID data.
When you save the data with the same filename, the version number increments by one. If
you delete data files after processing them, the version numbers in the list becomes
discontinuous.
If you save data with the existing filename, the data are given a version number higher by
one. Copied data are sometimes deleted after they are processed. Therefore, some of the
version numbers in the file list are usually omitted.
If you do not specify the version number of the file in the File Browser window, the
program loads the data of the latest version (data with the largest version number).
In the File Browser window, the parts till the group number are displayed as a filename.
1-40 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
■ Directory
You can specify two directories in the File Browser window using the buttons. One is a
data directory. The other is a global directory.
Normally, the user uses the data directory. The global directory is read-only.
Clicking the button displays a list of the files in the data directory to the file list
box in the File Browser window.
The data directory is specified in the Data in the Directory tab of the window which
opens by selecting Options — Preferences in the Delta Console window.
The program opens the 1D Processor window, and loads the data.
(This button is unavailable if the data are multiple-dimensional.)
The program opens the nD Processor window, and loads the data.
(This button is not displayed if the data are one-dimensional.)
The program opens the Data Slate window, and loads the data.
The program opens the Data Viewer window, and loads the data.
(This button is not displayed if the data are one-dimensional.)
The program opens the Data Slicer window, and loads the data.
(This button is not displayed if the data is one-dimensional.)
The program opens the File Information window, and displays the parameters
for the data (NMR data is not displayed).
The program closes the File Browser window.
NMECZ-USP-1 1-41
1 FUNDAMENTALS OF DELTA
■ Searching for
1-42 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
If you want to stop searching for files at any time, click the button that
appears during the search for the files.
NMECZ-USP-1 1-43
1 FUNDAMENTALS OF DELTA
Button Function
Returns the File Search window to the state for specifying the file search
conditions.
Shows parameters for the file in the File Information window.
■ Search conditions
● Date
The program searches for files using the time stamp attached to the file. There are three
time stamps: for the operating system ( ), file creation date ( ), and file revision date
( ).
The above figure searches for the data files created before February 13, 2009.
1-44 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
● Filename
The program searches for files with the name you specify here. The wild card (*) is
available.
This example searches for the data files that contain “str” in their filenames.
● Data size
The program searches for files that have the number of points you specify here.
This example searches for the data files in which the number of points is greater than 5000
in the x-axis.
● Number of dimensions
The program searches for files that have the number of dimensions you specify here.
This example searches for the data files in which the number of dimensions is greater than 2.
● Processed or not processed
The program searches for the files that are processed, or the files that are not processed
(measurement data (FID)).
If the “Match Processed files only” is not selected, the program searches for only the data
files of the measurement data (FID).
● Parameter
The program searches for the files that have the parameter you specify here.
The parameter may be any of the parameters that are saved in the files of the Delta program,
such as number of accumulations, and kind of solvent.
NMECZ-USP-1 1-45
1 FUNDAMENTALS OF DELTA
Filename
1-46 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
The program saves the 1D processed data with the same filename and the version
number incremented by one.
● Saving the data with a different filename
NMECZ-USP-1 1-47
1 FUNDAMENTALS OF DELTA
Filename box
The program saves the 2D processed data with the same filename and the version
number incremented by one.
1-48 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
Filename box
NMECZ-USP-1 1-49
1 FUNDAMENTALS OF DELTA
Delta ― 1 1 2 1 4 4 4 3 2 1 4
GX/GSX/EX 1 ― 1 1 1 4 4 4 3 4 1 4
GX 2D
Phase
5 ― 5 5 ― 5 5 5 3 5 5 5
ALPHA 6 1 1 6 1 4 4 4 3 7 1 4
Lambda 1 1 1 1 1 4 4 4 3 4 1 4
Alice 1 1 ― 1 1 4 4 4 3 4 1 4
Generic 2 1 1 ― ― 4 4 4 3 2 1 4
CMX 6 1 1 1 4 4 4 4 3 4 4 4
Nuts 4 4 4 4 4 ― 4 4 3 4 4 4
VXR/Unity 7 4 4 4 4 4 4 4 3 7 4 4
AC 8 8 8 8 4 8 8 8 3 8 8 4
AMX/DMX 2 1 1 2 1 4 4 4 3 4 4 4
Galactic 4 4 4 4 4 4 ― 4 3 4 4 4
Felix 4 4 4 4 4 4 4 ― 3 4 4 4
Felix(Win) 3 3 3 3 3 3 3 3 ― 3 3 3
JCAMP 4 4 4 4 4 4 4 4 3 4 ― 4
SIMPSON 4 4 4 4 4 4 4 4 3 4 4 ―
The following table explains the conversion conditions for the numbers in the above table.
Number Conversion condition
1 1D, 2D (Absolute/Phase sensitive mode), Relaxation time
2 1D-4D
3 1D Only
4 1D, 2D (Absolute/Phase sensitive mode)
5 2D Phase sensitive mode
6 1D, 2D (Absolute/Phase sensitive mode), 3D (Absolute/Phase sensitive mode),
Relaxation time
7 1D, 2D (Absolute/Phase sensitive mode), 3D (Absolute/Phase sensitive mode)
8 1D, 2D (Absolute)
1-50 NMECZ-USP-1
1 FUNDAMENTALS OF DELTA
2. Select the directory that has the data file you want to import.
3. Select the file in the file list.
No information appears in the data information box.
4. Click the button.
The program automatically imports the data file, and then opens the 1D Processor
window.
Since there is no information on the process list, the process list is not set
automatically.
NMECZ-USP-1 1-51
1 FUNDAMENTALS OF DELTA
Filename box
1-52 NMECZ-USP-1
1D NMR DATA
2
PROCESSING
2.1 1D PROCESSOR WINDOW......................................................... 2-1
2.2 1D DATA PROCESSING............................................................... 2-5
2.2.1 Canceling the DC Offset ................................................................ 2-5
2.2.2 Zero-Filling .................................................................................... 2-6
2.2.3 Window Function ........................................................................... 2-7
2.2.3a Setting the window function ..................................................... 2-7
2.2.3b Changing the parameters to use for the window function......... 2-8
2.2.3c Changing the window function ................................................. 2-9
2.2.3d Setting multiple window functions ......................................... 2-10
2.2.4 Fast-Fourier Transformation ......................................................... 2-11
2.2.5 Phase Correction .......................................................................... 2-12
2.2.5a Automatic phase correction .................................................... 2-12
2.2.5b Manual phase correction ......................................................... 2-12
2.2.6 Baseline Correction ...................................................................... 2-15
2.2.6a Automatic baseline correction ................................................ 2-15
2.2.6b Manual baseline correction ..................................................... 2-15
2.2.7 Unit for Axis ................................................................................ 2-18
2.2.8 Setting a Reference....................................................................... 2-18
2.2.8a Setting the reference using the pointer bar.............................. 2-18
2.2.8b Setting the reference using the Reference command .............. 2-19
2.2.9 Peak Detection.............................................................................. 2-20
2.2.9a Automatic peak detection........................................................ 2-20
2.2.9b Manual peak detection ............................................................ 2-23
2.2.10 Integration .................................................................................... 2-26
2.2.10a Automatic integration ............................................................. 2-26
2.2.10b Manual integration .................................................................. 2-26
2.2.11 Transfer Integral Tool Window ..................................................... 2-30
2.2.12 Comment ...................................................................................... 2-31
2.2.13 PiP (Picture in Picture) ............................................................... 2-34
2.2.14 Setting Display Range of Spectrum ............................................. 2-35
2.2.15 Printing ......................................................................................... 2-37
2.3 DECONVOLUTION..................................................................... 2-38
2 1D NMR DATA PROCESSING
Macro buttons
Panel-display
Input switch buttons
FID display
area
Output
Process
list
Phasing Tools
panel
The name of the data file is on the title bar of the 1D Processor window.
This window has two display areas. The upper display area is the input domain for
displaying the measurement data (FID: Free Induction Decay) before Fourier transformation.
The lower display area is the output domain for displaying data after processing. This
window also displays a process list that relates the two data. At the bottom right of the
window, buttons and boxes for phase correction appears.
You can display or hide the Processing Tools panel and the Phasing Tools panel by
clicking the panel-display switch buttons at the right side of the window.
NMECZ-USP-1 2-1
2 1D NMR DATA PROCESSING
■ Tool bar
The 1D Processor window in the extended mode has the tool bar. Its buttons are for loading
data files, storing processed data files, printing the data, and automatic processing of the data.
Button Name Function
Load Data from Disk Loads a data file to the hard disk.
Finger Loads a data file that is in the other window on the screen.
The following example loads a data file from window B to
window A:
In window A, click the button, click the button to
change the pointer to a finger, move the pointer to the window
B and click the data to move it. This function is the reverse of
cut-and-paste.
Open File Loads a data into the 1D Processor window that is presently
open. If the process list is already set, it is automatically
applied to the loaded data.
Open Clear Loads a data into the 1D Processor window that is presently
open. The previous process list is deleted when the data is
loaded.
Put Processed Data Transfers a data in the 1D Processor window that is presently
into Data Slate open to the Data Slate window.
Save Processed Data Saves a processed data to the hard disk with the same file
to Disk name and the version number incremented by one.
Print Prints a processed data.
Generate Report Prints the spreadsheet that includes all peak information.
2-2 NMECZ-USP-1
2 1D NMR DATA PROCESSING
■ Macro buttons
Each of the macro buttons has a default function. You can modify the function, or you can
assign a different macro function to the button so that it suits your needs.
Button Name Function
DC Balance Eliminates zero-frequency noise produced by any shift in the
balance of the AD converter.
Window Function Applies the Hamming and trapezoid window functions to FID
data.
FFT Performs Fast-Fourier transformation (FFT). FFT transforms a
time-domain data to a frequency-domain data.
Auto Phase Performs phase correction. The Machine Phase function is
assigned to this button.
Base Correct Performs baseline correction to the frequency domain
spectrum.
Abs Calculates the absolute value of the data.
PPM Changes the unit of the data axis to ppm. You can replace the
function of this button to Hertz, or Auto Reference.
Peak Pick/Integrate Performs peak picking and integration.
1. In the 1D Processor window, set the process list so that it performs the data
processing that you want.
2. Using the right mouse button, drag the data process list to the macro button to
which you want to assign the function that the data process list contains.
The function of the macro button is replaced with the function that the data process list
contains. To verify the function of the macro button, right-click the button to display
Help.
NMECZ-USP-1 2-3
2 1D NMR DATA PROCESSING
You can set the macro button to the default function using the Macros tab of the Processing
Tools panel.
1. In the 1D Processor window, click the Macro tab of the Processing Tools panel.
The Macros tab opens.
Process-switch tabs
Restore default
macros button
2. Click button.
The program changes the macro function of the button to its default function. Clicking
the button at the bottom right changes functions of all buttons to their default
functions.
2-4 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-5
2 1D NMR DATA PROCESSING
2.2.2 Zero-Filling
The Zerofill function increases the number of data points by adding a specified number of
zeros after the original data. Since the data of every dimension must have data points of the
number to the nth power of 2, the number of zeros you add must be an nth power of 2, such
as 1, 2, 4, 8, and 16. If you enter a numerical value that is not an nth power of 2, the
numerical value of the power of 2 larger than the entered value will be applied.
If you set Power_2 to FALSE, it becomes possible to analyze the data having a
number of points that is not an nth power of 2.
2-6 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-7
2 1D NMR DATA PROCESSING
2. Enter numerical value to change the parameter for the displayed window
function.
3. Click the button.
The window function appears in the FID display area.
The program processes the spectrum using the new parameter of the window function,
and displays the processed spectrum in the processed data display area.
4. Verify the change of the processed spectrum.
5. Repeat steps 2 to 4, and find the optimum value for the parameter.
2-8 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-9
2 1D NMR DATA PROCESSING
2-10 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-11
2 1D NMR DATA PROCESSING
1. Verify that all the processes in the process list are complete.
Verify that display/phase at the end of the process list is green. If it is not green, click
the button to complete the all processes in the process list.
2. Set the phase reference position.
a. Click the button on the pointer bar to select Phase mode.
b. In the processed-data display area, drag and drop the pivot-point line (Fig. 2.1)
or click the X-axis ruler area to change the phase-reference position.
The pick-position marker appears at the position where you dragged the pivot point
line, or you clicked the X-axis ruler area.
The information on the position of the pick-position marker appears at the right of Φp
in the Phasing Tools panel.
2-12 NMECZ-USP-1
2 1D NMR DATA PROCESSING
Pivot-point line
NMECZ-USP-1 2-13
2 1D NMR DATA PROCESSING
When the process list has multiple phase processes, the program adds the P0 and P1 values
in the phase processes.
2-14 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-15
2 1D NMR DATA PROCESSING
3. Specify the approximation method for the baseline distortion in the Fit box.
The following four approximations are provided.
Approximation Description
Polynomial Approximates the baseline distortion using a polynomial equation.
Trigonometric Approximates the baseline distortion using a trigonometric function.
Piecewise Linear Approximates the baseline distortion using a line graph.
Akima Approximates the baseline distortion using the Akima method.
4. Enter values in the Order and Ave.Points boxes for the approximation method
in step 3.
5. Click the button on the pointer bar to select Cursor mode.
6. Move the mouse pointer to the processed data area in the iBase Correct
window, and click multiple points where there is no peak in the spectrum.
Baseline-point markers appear to show the points where you clicked.
2-16 NMECZ-USP-1
2 1D NMR DATA PROCESSING
Calculated
baseline
NMECZ-USP-1 2-17
2 1D NMR DATA PROCESSING
X Ref box
If the Options panel is not displayed, show it by clicking the Options button at
the right side of the 1D Processor window.
2-18 NMECZ-USP-1
2 1D NMR DATA PROCESSING
3. Expand the peak that you want it to be the reference in the X direction, a
reference peak such as a standard substance (e.g. TMS) and a solvent.
4. Click the top of the peak to set it as the reference peak.
The program adds the Reference process to the process list. The list also contains the
Position where you clicked, and the Reference value you set in step 2.
■ References
NMECZ-USP-1 2-19
2 1D NMR DATA PROCESSING
The markers of the peak-threshold level, noise-threshold level, and base plain level are
displayed in the processed-data display area.
2-20 NMECZ-USP-1
2 1D NMR DATA PROCESSING
Peak
threshold Noise
marker threshold
marker
Base plain
level marker
Here, the markers appear at the positions where the program automatically calculates.
2. Set the peak threshold level as shown below.
a. Click the button on the pointer bar to select Threshold mode.
b. Point to the peak threshold marker.
The pointer changes to the icon for changing the peak threshold.
c. Hold down the mouse button and drag it to set the position of the peak
threshold.
The peak threshold level marker moves to the position where you release the mouse
button.
3. Set the noise threshold level as shown below.
a. Point to the noise threshold marker.
The pointer changes to the icon for changing the noise threshold.
If the marker is near another marker, and you cannot catch the noise threshold
marker, point to one of the grab boxes at both ends.
b. Hold down the left mouse button and drag it to set the position of the noise
threshold.
The noise threshold level marker moves to the position where you release the mouse
button.
4. Set the base plain level.
a. Point to the base plain level marker.
The pointer changes to the icon for changing the base plain level.
If the marker is near another marker, and you cannot catch the baseline level
marker, point to one of the grab boxes at both ends.
b. Hold down the mouse button and drag it to set the position of the base plain
level.
The base plain level marker moves to the position where you release the mouse
button.
NMECZ-USP-1 2-21
2 1D NMR DATA PROCESSING
Peak markers
● Reference
If you click the button in the 1D Processor window, the program performs both
automatic peak detection and automatic integration at the same time, and displays the
results in the processed-data display area in the 1D Processor window.
2-22 NMECZ-USP-1
2 1D NMR DATA PROCESSING
NMECZ-USP-1 2-23
2 1D NMR DATA PROCESSING
Click the (button to display key-commands) button on the pointer bar and
select Select All [A] from the pull-down menu.
The program selects all peaks, and changes the color of the peak markers of the peaks.
1. Select a peak.
Select the peak using the procedure “■ Specifying a peak by clicking a peak marker”
that is explained in the previous section.
2. Click the button on the pointer bar to select Select mode.
3. Point to the peak marker of another peak to select, hold down Shift key, and
click it.
The program selects both peaks, and changes the color of the peak markers of the
peaks.
2-24 NMECZ-USP-1
2 1D NMR DATA PROCESSING
2. Click the (the button to display the key-command) button and select
J Coupling [J] from the pull-down menu.
The J value between the selected peaks appears.
The peak list appears under the spectrum in default. If you want to display the peak list
above the peaks, right-click the mouse to display a menu, and select Options — Peak —
Peak Above.
NMECZ-USP-1 2-25
2 1D NMR DATA PROCESSING
2.2.10 Integration
Automatic and manual spectrum integrations are provided.
2-26 NMECZ-USP-1
2 1D NMR DATA PROCESSING
● Reference
The difference of the level of the integration base point depending on the specified
integration range:
In default, the integration base line is traced between the both edge points of the integration
range. If you specify the integration range while holding down the Shift key, the levels of
the integration base point are set to that of the start and end points you specified using the
mouse.
1. Select the integration curve whose integration you want to set to the normal
value using the steps described in the above section, “■ Selecting the current
integration curve”.
2. Click the (the button to display the key-command) button, and select
Normalize [N] from the pull-down menu.
The program sets the integration value of the integration curve thtat you selected in
step 1 to the normal value, and changes the other integration values relatively
according to the normal value.
When you want to change the normal value, carry out the following step 3.
3. Enter the normal value in the Normal box of the Options panel.
The integration value changes.
Normal box
If the Options panel is not displayed, click the Options button at the right side
of the 1D Processor window to open the Options panel.
NMECZ-USP-1 2-27
2 1D NMR DATA PROCESSING
1. Click the (the button to display the key-command) button on the pointer
bar, and select Select All [A] from the pull-down menu.
All integration curves are selected, and appear in the color for selected state.
2. Click the button again, and select Delete [Backspace] from the
pull-down menu.
All integration curves are deleted.
4. Move the mouse pointer to change the size of an integration curve, and release
the mouse button.
Expansion will be performed if the mouse pointer is moved upward, and reduction
will be performed if the mouse pointer is moved downward.
2-28 NMECZ-USP-1
2 1D NMR DATA PROCESSING
1. Select the integration curve whose integration position you want to change by
using the steps described in the previous section, “■ Selecting the current
integration curve”.
The integration curve and its box appear.
2. Point to the grab box at the bottom center of the box, and hold down the mouse
button.
1. Select the integration curve whose integral baseline you want to correct using
the steps described in the previous section, “■ Selecting the current integration
curve”.
The integration curve and its box appear.
2. Point to the grab box at the bottom right or left of the box, and hold down the
mouse button.
NMECZ-USP-1 2-29
2 1D NMR DATA PROCESSING
1. Select the integration curve that you want to split using the steps described in
the previous section, “■ Selecting the current integration curve”.
The integration curve and its box appear.
2. Point to the grab box at the top left of the box.
The mouse pointer changes to scissors.
2-30 NMECZ-USP-1
2 1D NMR DATA PROCESSING
2.2.12 Comment
In the Delta program, you can write comments (characters) on the screen by selecting the
Annotation mode in the pointer bar.
When you want to display the comments, right-click the mouse and select Options —
Annotation.
Comments are saved together with the data. Therefore, after you write a comment, save the
data on the hard disk.
■ Creating a comment
NMECZ-USP-1 2-31
2 1D NMR DATA PROCESSING
2-32 NMECZ-USP-1
2 1D NMR DATA PROCESSING
■ Editing a comment
■ Moving a comment
1. Select the Select mode ( button) or the Annotation mode ( button) on the
pointer bar.
2. Click the comment you want to change.
The selected comment is surrounded with a blue box.
3. Right-click the mouse button.
NMECZ-USP-1 2-33
2 1D NMR DATA PROCESSING
2-34 NMECZ-USP-1
2 1D NMR DATA PROCESSING
Select Auto Arrange while clicking the button, or press the P key on the
keyboard.
NMECZ-USP-1 2-35
2 1D NMR DATA PROCESSING
The values displayed in the parameter boxes are the real range of the spectrum display.
They may differ from the numerical values you entered because the values are
rounded to fit with the digital resolution of the spectrum.
2-36 NMECZ-USP-1
2 1D NMR DATA PROCESSING
2.2.15 Plotting
You can simply perform the data plotting using the following procedures.
If you want to customize the printing conditions, use the “Page Layout Editor” window.
You can specify the data and parameters to print by selecting them in the Delta tab in
the Print window which appears by clicking the button.
● Reference
When you plot explanatory notes, such as peak comment, and sample names, measurement
conditions, in arbitrary places of the processing data display area, refer to Sect. 2.2.12,
“Comment”.
NMECZ-USP-1 2-37
2 1D NMR DATA PROCESSING
2.3 DECONVOLUTION
If you use deconvolution, you can separate the NMR peaks that overlap. By superposing
the waveform of Lorentzian, Gaussian or hybrid of them, the program approximates the
observed spectrum in which two or more peaks are overlapped, and optimizes its waveform
parameters. After deconvoluting the spectrum, the program automatically saves the
resultant integration values in the peak information of the spreadsheet.
Before deconvoluting the spectrum, you need to peak pick the data set; then expand the
range to deconvolute. The program deconvolutes all the peaks displayed on the display
window. If you deconvolute the whole range of the data set, it may take a long time.
1. Right-click the mouse, and select Options — Peaks — Deconvolve from the
pop-up menu to turn on the deconvolution display.
2. Click the button on the pointer bat to select Peak mode.
3. Click the button on the pointer bar, and select Fit Lorentzian, Fit
Gaussian, or Fit Mixed from the command menu to specify the peak shape.
When you select a peak shape, The deconvolution operation starts, and is performed
under iteration conditions. The χ2 value (Chi square) that appears in the Delta Console
window shows the state of the optimization. The program repeats the optimization
until χ2 becomes a constant.
The calculated (deconvoluted) peaks appear in yellow. The difference between the
sum of the calculated peaks and the measured spectrum appears in red.
4. If you want to display the sum of the calculated peaks, right-click the mouse,
and select Options — Peaks — Deconvolve Sum from the pop-up menu.
The sum of the calculated peaks appears.
2-38 NMECZ-USP-1
2 1D NMR DATA PROCESSING
Difference spectrum
(red)
Deconvoluted
spectra (yellow)
NMECZ-USP-1 2-39
2 1D NMR DATA PROCESSING
■ Adding a peak
2-40 NMECZ-USP-1
2D NMR DATA
3
PROCESSING
3.1 nD PROCESSOR WINDOW ........................................................ 3-1
3.1.1 nD Processor Window .................................................................... 3-1
3.1.2 Domain ........................................................................................... 3-3
3.1.3 Process List .................................................................................... 3-3
3.1.4 Process Levels ................................................................................ 3-4
3.2 2D DATA PROCESSING............................................................... 3-5
3.2.1 Zero-Filling .................................................................................... 3-5
3.2.2 Window Function ........................................................................... 3-6
3.2.2a Setting a window function ........................................................ 3-6
3.2.2b Changing the parameters for the window function ................... 3-7
3.2.2c Changing the window function ................................................. 3-8
3.2.2d Setting multiple window functions ........................................... 3-9
3.2.2e Checking the window function using 1D slice data ................ 3-10
3.2.3 Symmetrization ............................................................................ 3-12
3.2.4 Adjusting Levels of the Contour Lines (1)................................... 3-13
3.2.5 Adjusting Levels of the Contour Lines (2)................................... 3-16
3.2.6 Projection ..................................................................................... 3-17
3.2.6a Whole range projection ........................................................... 3-17
3.2.6b Partial projection ..................................................................... 3-17
3.2.6c Projection modes..................................................................... 3-17
3.2.7 Loading and Processing 1D Slice Data ........................................ 3-18
3.2.7a Displaying slice data using the process list............................. 3-18
3.2.7b Displaying slice data in the 2D Viewer window ..................... 3-19
3.2.8 Pasting a High-Resolution 1D NMR Spectrum ............................ 3-20
3.2.8a Pasting data from a disk .......................................................... 3-20
3.2.8b Pasting the displayed data ....................................................... 3-21
3.2.8c Pasting data displayed in another window .............................. 3-22
3.2.9 Reference Setting ......................................................................... 3-23
3.2.10 Peak Pick ...................................................................................... 3-24
3.2.10a Setting the J value ................................................................... 3-25
3.2.10b Adjusting the threshold and base levels .................................. 3-25
3.2.10c Displaying the results of peak pick in the Spread Sheet
window ................................................................................... 3-26
3 2D NMR DATA PROCESSING
3.b NMECZ-USP-1
3 2D NMR DATA PROCESSING
Tool bar/
tool buttons
Process-list
control buttons
Domain
buttons
Process lists
Fig. 3.1 nD Processor window
Toggle
button
In the nD Processor window, using the button, you can switch the display between the
process list before processing and the result that are processed according to the process list.
NMECZ-USP-1 3-1
3 2D NMR DATA PROCESSING
■ Tool bar
The nD Processor window has a tool bar that contains buttons for loading a data file,
saving a processed data file, printing, and other functions.
Button Name Function
Open Data From Disk Loads a data file from the hard disk.
Open Data By Fingering Open the data that you want to load before you click
Geometry this button.
For example, if you want to load a data file in window
B to window A, first click the button in window A,
click the button to change the pointer to a finger;
then click the data file in window B that you want to
load. This is the reverse of cut-and-paste.
Open Data File Loads new data file to the nD Processor window that is
presently open. If the nD Processor window has a
process list, the program processes the data according
to the process list.
Open Data File and Clear Loads new data file to the nD Processor window that is
Processing Lists presently open. At the same time, the present process
list in the nD Processor window is deleted.
Save Data File Saves the data file that is presently open to the hard
disk.
Remove File and Clear Deletes the data file and its process list that are
Processing Lists presently open.
Process File and Put in Processes the data according to the process list, and
Data Slate displays the processed data in the Data Slate window.
Process File and Put in Processes the data according to the process list, and
Data Viewer displays the processed data in the Data Viewer
window.
Toggle between processing Switches the display in the nD processor window
list view and geometry between the process-list display and processed-data
view display.
Abort current processing Aborts the running process in the process list.
3-2 NMECZ-USP-1
3 2D NMR DATA PROCESSING
3.1.2 Domain
Delta treats each axis of NMR data set as a “domain.”
The domain contains the nucleus, the observation range of a spectrum and the processing
parameters. The only one domain of an NMR data set can be processed at one time. This is
called the “major domain,” other domains are called “minor domains.”
In order to distinguish the order of evolution times of t1, t2, and t3 in NMR data, Delta
expresses the domain with notation such as x, y, z, a, b, c, d, and e. On the axis of the
x-domain, the NMR signals are always acquired by the receiver of a spectrometer, and
other domains follow the x-domain with evolution times in the order of y, z….
Processing is performed only in the major domain of a data set. Both a major domain and a
minor domain can be replaced using the transpose command. If the transpose command is
executed n times for n-dimensional data, a major domain will return to the first domain.
transpose
process
NMECZ-USP-1 3-3
3 2D NMR DATA PROCESSING
Process level
control button
3-4 NMECZ-USP-1
3 2D NMR DATA PROCESSING
3.2.1 Zero-Filling
The Zerofill function extends the data set by refilling zero data with the specified scale
factor to the original data. Since the data size of every dimension consists of the nth power
of 2, the specified number must be the nth power of 2 such as 1, 2, 4, 8, and 16.
If you enter the numerical value that is not the nth power of 2, the numerical value of the
power of 2 larger than the entered value will be applied.
1. In the nD Processor window, in the process list of the domain where you want
to add zerofill, click the process above which you want to add zerofill.
2. Click the button.
3. Select PreTransform — Zerofill in the nD Processor window.
NMECZ-USP-1 3-5
3 2D NMR DATA PROCESSING
3-6 NMECZ-USP-1
3 2D NMR DATA PROCESSING
Parameter
box
2. Type the numerical value to change the parameters for the displayed window
function.
3. Click the button.
The program reprocesses the data according to the process list and displays the
processed data.
4. Verify the change of the processed spectrum
5. Repeat steps 2 to 4, and find the optimum value for the parameter.
To find an optimum parameter, it will be helpful for you to slice the data using
the button, and evaluate the window function in the 1D NMR data.
NMECZ-USP-1 3-7
3 2D NMR DATA PROCESSING
3-8 NMECZ-USP-1
3 2D NMR DATA PROCESSING
NMECZ-USP-1 3-9
3 2D NMR DATA PROCESSING
1. In the nD Processor window, click the X or Y domain button to select the axis to
display.
2. Click the button in the Axes section to display the slice-position
parameters.
Axes
section
X domain
Slice
position
parameters
3-10 NMECZ-USP-1
3 2D NMR DATA PROCESSING
NMECZ-USP-1 3-11
3 2D NMR DATA PROCESSING
3.2.3 Symmetrization
The following two symmetrization methods are provided.
• Before performing FFT on the 2D data, add symmetrize to the process list in the
nD Processor window and carry out the process list.
• After performing FFT on the 2D data, symmetrize the data in the 2D Viewer
window or Data Slate window.
Symmetrization is useful to reduce asymmetric noise. But be careful because, if the noise
happens to have symmetric peaks, symmetrization may result in emphasizing the noise
peak that remains after the symmetrization.
■ Before performing FFT on the 2D data, add symmetrize to the process list
in the nD Processor window and carry out the process list.
3-12 NMECZ-USP-1
3 2D NMR DATA PROCESSING
■ After performing FFT on the 2D data, symmetrize the data in the 2D Viewer
window or Data Slate window.
1. In the 2D Viewer window, click the button on the pointer bar to select select
mode.
2. Move the mouse pointer to the 2D data display area.
3. Hold down the right mouse button.
The following pop-up menu appears.
NMECZ-USP-1 3-13
3 2D NMR DATA PROCESSING
4. While holding down the right mouse button, drag the pointer to Level Tool [∖]
in the pop-up menu, and release the mouse button.
The Level Tool window appears.
Contour-line
level button Top slider
Preset button
Bias slider
Threshold level
Bottom slider
Noise level
Apply
Base level
Fig. 3.5 Level Tool window
Button Function
Preset button Sets the number of contour-line levels. You can select from 2, 4, 6, 8,
12, 16 and 24.
Contour-line level Click the buttons to specify the level to display. The buttons you
button clicked are highlighted.
Top slider Determines the highest level (intensity) of the contour lines.
Bottom slider Determines the lowest level (intensity) of the contour lines.
Bias slider Determines the gradient of the levels (intensities) in the range between
the top slider (top right) and the bottom slider (bottom left). Moving the
bias slider changes the slope of the curve shown in the figure. Moving
it upward draws more contour lines in the low level range of signal
intensities.
Noise level The program automatically sets the noise level after processing the
data.
Threshold level The program automatically sets the threshold level after processing the
data. This is used for peak picking in 2D processing.
Base level Zero level of the signal intensity. Set it to the lowest level in the Level
Tool window.
Apply Click this button to draw the contour lines again according to the levels
you have set.
3-14 NMECZ-USP-1
3 2D NMR DATA PROCESSING
5. In the Level Tool window, adjust the levels of the contour lines.
In the Level Tool window, set the lowest level by taking the noise level into account.
Set the number of the contour lines to draw and the levels (intensities) to assign to the
lines between the minimum intensity and the maximum intensity.
Maximum intensity
Noise level
Usually, few or no contour lines are set to levels below the noise level. If you set
contour levels below the noise level, the program displays the contour lines of the
noise; this requires a large amount of system memory, and the processing speed may
become slow.
Adjusting the contour-line levels using the following procedure:
a. Among the contour-line level buttons, click the lowest one.
b. Move the bottom slider to specify the lowest level of the contour line.
c. Move the top slider to specify the highest level of the contour line.
d. Click one of the preset buttons to specify the number of contour lines to
display.
e. Click the button.
NMECZ-USP-1 3-15
3 2D NMR DATA PROCESSING
3. Specify the lower bound of the projection (slice) spectrum by dragging the
mouse.
4. Move the mouse pointer onto the 2D spectrum, right-click, and select Contour
Bounds — Set Lower Bound from the pop-up menu.
The bottom slider of the Level Tool is specified.
5. Select the projection (slice) spectrum again, and specify the upper bound of the
projection spectrum by dragging the mouse.
6. In the same manner as that for specifying the lower bound, move the mouse
pointer onto the 2D spectrum, right-click, and select Contour Bounds — Set
Upper Bound from the pop-up menu.
The top slider of the Level Tool is specified.
7. Using the bias slider of the Level Tool, adjust the level to the optimum position
and click Apply.
3-16 NMECZ-USP-1
3 2D NMR DATA PROCESSING
3.2.6 Projection
3.2.6a Whole range projection
In the 2D Viewer window, select Project — Full.
The program displays the projection of the data in the projection-data display area.
Projection-data
display area
NMECZ-USP-1 3-17
3 2D NMR DATA PROCESSING
Y-axis
counter
3-18 NMECZ-USP-1
3 2D NMR DATA PROCESSING
NMECZ-USP-1 3-19
3 2D NMR DATA PROCESSING
3-20 NMECZ-USP-1
3 2D NMR DATA PROCESSING
NMECZ-USP-1 3-21
3 2D NMR DATA PROCESSING
Mouse pointer
As soon as you see the mouse pointer change to the above shape while holding
down the right mouse button, right-drag the pointer quickly.
If you do not move the pointer quickly, a pop-up menu appears. If this happens,
release the mouse button, and hold down the right mouse button again.
6. While holding down the right mouse button, drag the pointer to the button in
the 2D Viewer window, and then release the mouse button.
The program displays the 1D spectrum in the area you selected in step 3.
3-22 NMECZ-USP-1
3 2D NMR DATA PROCESSING
3. Enter the reference values in the X-axis reference value input box and the
Y-axis reference value input box.
The units are the same as those used in the rulers.
4. Click the button on the pointer bar to select Reference mode.
5. Move the mouse pointer to the top of the peak that you want to specify as the
reference.
6. Click the left mouse button.
The position values of the peak that you clicked change to the reference values that
you set in step 3.
NMECZ-USP-1 3-23
3 2D NMR DATA PROCESSING
Fig. 3.7 Results of Peak Pick and Integrate with Cosy Peak J Width
Before you click the button, setting the J value is required. Set it according
to the following Sect. 3.2.10a, “Setting the J value”.
If the number of peaks is not appropriate for peak pick, adjust the threshold and
base levels according to the following Sect. 3.2.10b, “Adjusting the threshold and
base levels”.
The detailed results of the peak pick are displayed in the Spread Sheet window.
Refer to the following Sect. 3.2.10c”, Displaying the results of peak pick in the
Spread Sheet window”.
3-24 NMECZ-USP-1
3 2D NMR DATA PROCESSING
Threshold level
Base level
NMECZ-USP-1 3-25
3 2D NMR DATA PROCESSING
7. Move the mouse pointer to the threshold and base level setting area in the
Level Tool window, and hold down the middle mouse button.
8. While holding down the left mouse button, drag the pointer to the position
where you want to set the base level, and then release the mouse button.
The base level moves.
After this, perform peak pick again.
3.2.10c Displaying the results of peak pick in the Spread Sheet window
1. In the Delta Console window, select Analyze — Peak Spreadsheet.
The Spread Sheet window opens.
2. In the Spread Sheet window, click the button.
The mouse pointer changes to a finger.
3. Move the pointer to the 2D NMR data displayed on the screen, and click it.
The detailed results of the peak pick appear.
For information on the Spread Sheet window, refer to Chap. 9, “SPREAD SHEET”.
3.2.11 Data Processing of Phase Sensitive 2D Data
For processing the phase sensitive 2D data, phase-correction processing and display control
of positive-and-negative peak are necessary in addition to various kinds of data processing
of 2D data in the absolute-value display mode.
■ When you know the phase-correction values, or use the calculated values
When you know the correct phase-correction values in advance, or use the calculated
phase-correction values, add the phase-correction values into the process list in the nD
Processor window, and correct the phase according to the following operation.
3-26 NMECZ-USP-1
3 2D NMR DATA PROCESSING
4. Click the to display the Φ0, Φ1, ΦP parameter input boxes, and enter the
phase-correction values in each box.
NMECZ-USP-1 3-27
3 2D NMR DATA PROCESSING
3-28 NMECZ-USP-1
3 2D NMR DATA PROCESSING
When you do not know the suitable phase correction value, perform the 2D Fourier
transformation using the Phase 2D window, obtain the phase correction values for each
slice data using the functions of the Phase 2D window, and perform the phase correction.
1. In the nD Processor window that has the necessary process list, click the
button on the tool bar.
The Phase 2D window opens.
2. In the 2D data display area, click the button on the pointer bar to select
Cursor mode.
3. Select two peaks, whose phases are not completely correct yet, to use for
phase correction.
Select the two peaks that are positioned as far as possible in the diagonal direction.
NMECZ-USP-1 3-29
3 2D NMR DATA PROCESSING
3-30 NMECZ-USP-1
3 2D NMR DATA PROCESSING
NMECZ-USP-1 3-31
3 2D NMR DATA PROCESSING
+ button and
– button
3-32 NMECZ-USP-1
3 2D NMR DATA PROCESSING
T1 noise
1. In the Data Slate window, open the 2D spectrum whose T1 noise you want to
eliminate.
2. Select Actions — Box Action — Ridge Reduce X.
3. Click the button on the pointer bar, and select the .
NMECZ-USP-1 3-33
3 2D NMR DATA PROCESSING
3-34 NMECZ-USP-1
ARRAY DATA
4
PROCESSING
4.1 OUTLINE OF ARRAY MEASUREMENT DATA
PROCESSING ................................................................................ 4-1
4.2 PROCESSING METHOD FOR ARRAY MEASUREMENT
DATA ............................................................................................... 4-2
4.2.1 Processing All Data as a Single Data File ...................................... 4-2
4.2.1a When process conditions are saved in the process list .............. 4-2
4.2.1b When the process list is not provided ....................................... 4-3
4.2.1c Displaying arbitrary data in the Data Slate window ................. 4-7
4.2.2 Processing arbitrary Data taken from the Measurement Data ........ 4-8
4 ARRAY DATA PROCESSING
● Processing all the measurement data (first to nth data) at one time
under the same processing conditions (
Sect. 4.2.1).
FFT
2
● Taking out a data file from among the first to the n th measurement data,
and processing the data (
Sect. 4.2.2).
NMECZ-USP-1 4-1
4 ARRAY DATA PROCESSING
The processed data appears as one-dimensional data. These data are useful when
viewing the change in intensities while taking notice of the specific signal, such as a
90° pulse or decoupling map.
4-2 NMECZ-USP-1
4 ARRAY DATA PROCESSING
If you click the button instead of the button, the program processes the
first to the nth measurement data according to the process list, and then opens the
Data Slate window. In this case, the 1st to the nth processed data are arranged in
one window as the following figure. This presentation is useful when comparing
details of the spectra, such as homonuclear spin decoupling or differential NOE.
NMECZ-USP-1 4-3
4 ARRAY DATA PROCESSING
2. Click the button in the Axes section, and enter the slice data points (1 to n)
in the slice position parameter input box.
Select the data whose phase is easy to correct, and slice it.
Slice position
parameter
input box
4-4 NMECZ-USP-1
4 ARRAY DATA PROCESSING
The processed data appears as one-dimensional data. These data are useful when
viewing the change in intensities while taking notice of the specific signal, such as a
90° pulse or decoupling map.
NMECZ-USP-1 4-5
4 ARRAY DATA PROCESSING
If you click the button instead of the button, the program processes the
first to the nth measurement data according to the process list, and then opens the
Data Slate window. In this case, the 1st to the nth processed data are arranged in
one window as in the following figure. This presentation is useful when
comparing details of the spectra, such as homonuclear spin decoupling or
differential NOE.
4-6 NMECZ-USP-1
4 ARRAY DATA PROCESSING
Row-slice
display area
Column slice
display area
3. Move the pointer to the row-slice display area, and hold down the right mouse
button to show the pop-up menu.
NMECZ-USP-1 4-7
4 ARRAY DATA PROCESSING
2. Click the button in the Axes section, and enter the position (number of
points) of the slice data that you want to process into the Y-axis slice position
parameter input box.
Slice position
parameter
input box
4-8 NMECZ-USP-1
4 ARRAY DATA PROCESSING
4. Process the data in the same way as for the usual 1D data.
If you want to save the processed data, save the data with a different filename.
NMECZ-USP-1 4-9
ADDING AND
5
SUBTRACTING THE DATA
5.1 DISPLAYING THE OPERATE ON FILES WINDOW ............. 5-1
5.2 ADDING AND SUBTRACTING
THE MEASUREMENT DATA (FID)........................................... 5-2
5.3 ADDING OR SUBTRACTING THE PROCESSED DATA ....... 5-5
5 ADDING AND SUBTRACTING THE DATA
NMECZ-USP-1 5-1
5 ADDING AND SUBTRACTING THE DATA
3. On the upper left of the “Operate on Files” window, click the button the
same number of times as the number of files that you want to add.
As shown in the following figure, in the “Operate on Files” window, rows whose IDs
are f1, f2, f3, etc. Appear under “Input Data Files”.
5-2 NMECZ-USP-1
5 ADDING AND SUBTRACTING THE DATA
7. On the lower left of the Operate on Files window, click the button.
Under “Expressions & Processing Lists”, in the ID column, the row ID:e1 appears.
NMECZ-USP-1 5-3
5 ADDING AND SUBTRACTING THE DATA
8. Verify that Expression appears in the Type column of the e1 row, click the
space at the right of the button, and type the summation equation of the
spectra in the cell.
If the addition is a simple summation of the spectra, type “f1+f2+f3+others”.
5-4 NMECZ-USP-1
5 ADDING AND SUBTRACTING THE DATA
3. On the upper left of the “Operate on Files” window, click the button the
same number of times as the number of files that you want to add or subtract.
In the “Operate on Files” window, rows whose IDs are f1, f2, f3, etc. appear under
“Input Data Files”.
4. Click the button.
The pointer changes to a finger.
5. Move the pointer to one of the data display areas whose data you want to add
or subtract, and click it.
The file name of the data you clicked appears in the Files column under “Input Data
Files”.
6. As for each spectrum you want to add or subtract, repeat steps 4 and 5 to its ID.
7. At the lower left of the Operate on Files window, click the button.
Under “Expressions & Processing Lists”, in the ID column, the row ID:e1 appears.
8. Verify that Expression appears in the Type column of the e1 row, click the
space at the right of the button, and type an appropriate addition or
subtraction equation in the cell.
9. Click the button at the left of the equation you typed.
This displays two buttons to the right of the button: the button and the
button.
10. Click the button.
The calculated data appears in the Data Slate window.
The data to add or subtract must have the same measurement conditions such as
x_offset, x_sweep, and zerofill.
NMECZ-USP-1 5-5
6.1
BATCH PROCESSING
6
DISPLAYING THE BATCH PROCESS FILES WINDOW ...... 6-1
6.2 BATCH PROCESSING OF MEASUREMENT DATA (FID) ... 6-2
6 BATCH PROCESSING
NMECZ-USP-1 6-1
6 BATCH PROCESSING
If you want to batch process the files that have a series of group numbers, click the
button above the button, and then specify the smallest group number of the files
that you want to batch process.
A series of group numbers to batch process must have sequential numbers.
3. Click the button to load the process list of that measurement data.
If you want to process all the data using a single process list, click the button at
the bottom of the “Batch Process Files” window.
6-2 NMECZ-USP-1
COMBINING THE 1D NMR
7
DATA SETS
7.1 COMBINE FILES WINDOW ....................................................... 7-1
7.2 COMBINING THE 1D NMR DATA SETS ................................. 7-1
7 COMBINING THE 1D NMR DATA SETS
NMECZ-USP-1 7-1
7 COMBINING THE 1D NMR DATA SETS
3. Enter the file name in the Output File box at the lower part of the Combine Files
window.
4. On the top right side of the Combine Files window, in the Y box of the Actual
Points column in the “Output File Information” section, enter the number of the
data points in the Y-axis.
5. On the bottom right side of the Combine Files window, under Axis Info, set the
conditions of the Y-axis.
6. As necessary, click one of the following buttons: the , , , , or
button.
This specifies the window that displays the combined data.
If you want to change the order of the data, select the data to move, and then
click the or button to move the data to a desired position.
7-2 NMECZ-USP-1
DATA SLATE WINDOW
8
The data slate is the tool for displaying data sets and analyzing them.
You can use it for many purposes. The window can handle the 1D, 2D,
3D, and 4D NMR data files, and moreover, it can also handle multiple
data files in a window. This is contrasted with the Data Viewer window
and the Data Processor window in which you can only handle one data
set.
The data slate allows you to overlap and connect multiple data for
comparison, analysis, and plotting.
NMECZ-USP-1 8-1
8 DATA SLATE WINDOW
List box
Data information
box
8-2 NMECZ-USP-1
8 DATA SLATE WINDOW
● Reference
The geometry is an area to display the processed data.
This Data Slate window consists of two geometries.
NMECZ-USP-1 8-3
8 DATA SLATE WINDOW
8-4 NMECZ-USP-1
8 DATA SLATE WINDOW
1. Click the geometry where you want to display the additional data.
2. Click the button.
3. Click the button.
4. Click the button.
The File Browser window opens.
5. In the list box of the File Browser window, select the data file that you want to
additionally display, and click the button.
The data saved in the selected file are added to the geometry that you selected in step 1.
The additional data is overlaid with the primarily displayed data in a different
color. If necessary, you can change the display position (
Display that overlays multiple data in a geometry”).
Sect. 8.4.2, “■
1. Click the geometry where you want to display the additional data.
2. Click the button.
3. Click the button.
4. Click the button.
The mouse pointer changes to a finger.
5. Move the pointer to the area where the data file that you want to additionally
display appears, and click it.
The data you clicked are added to the geometry that you selected in step 1.
The additional data is overlaid with the primarily displayed data. If necessary,
you can change the display position (
multiple data in a geometry”).
Sect. 8.4.2, “■Display that overlays
NMECZ-USP-1 8-5
8 DATA SLATE WINDOW
■ To make the horizontal scale the same in all geometries displayed in one
Data Slate window
■ To make the horizontal and vertical scales the same in all geometries
displayed in one Data Slate window
8-6 NMECZ-USP-1
8 DATA SLATE WINDOW
The following operation is effective for data in two geometries displayed in the same Data
Slate window, and for data in two geometries displayed in different Data Slate windows.
1. In the Data Slate window, select Tools — Geometry Tools — Connection Tool.
The Connect Geometries window opens.
NMECZ-USP-1 8-7
8 DATA SLATE WINDOW
7. Click the button of the data axis of “Select Geometry 2” (selected in step 6) that
you want to connect with the selected data axis of “Select Geometry 1”
(selected in step 4).
For example, if you want to expand the horizontal and vertical axes of two 1D data to
the same ratio, set as fbelow:
a. Click the input box at the right of the X of Source Ruler.
b. Click X in the top row of the Destination Ruler.
c. Click the input box at the right of the Y of Source Ruler.
d. Click Y in the middle row of the Destination Ruler.
You can make a connection between a 1D data geometry and a 2D data geometry.
8. Click the button.
The two geometries are connected.
9. Expand the data of geometry 1 or geometry 2.
To the axes selected in step 7, the geometry 1 and geometry 2 data are expanded with
the same ratio.
10. Select File — Close.
The Connect Geometries window closes.
8-8 NMECZ-USP-1
8 DATA SLATE WINDOW
The first data is displayed in green, the data added next is displayed in brown, and the third
in blue.
Thus, when additional multiple data are displayed in a single geometry, each data is
displayed in a different color. Moreover, the selected-data display box is shown in the same
color as the presently selected data.
Selected-data
display box
For example, if you select the second data, the selected-data display box is shown in brown
for the real data and purple for the imaginary data.
Color for real data
Color for imaginary data
Also, the scales for the X-ruler and Y-ruler change corresponding to the data that is
presently selected.
NMECZ-USP-1 8-9
8 DATA SLATE WINDOW
Click the tab above the spectrum area in the Data Slate window.
The data for the clicked-tab is displayed.
Every time you click a tab, the colors of the selected-data display box changes. The colors
of the selected-data display box represent the display color of the currently selected data.
■ Changing the display position of the selected data in the Y-axis direction
1. Select the data whose position you want to change in the Y-axis direction.
For the operation procedure, refer to the previous section, “■Changing the
selected data”.
2. Click the button on the pointer bar.
The button is highlighted in blue.
3. Move the pointer to the processed-data display area. While pressing the Alt
and Shift key, hold down the left mouse button.
While the Alt key is pressed, the button is highlighted in blue. Also, while the
Shift key is pressed, the button is highlighted in blue.
When the button is highlighted in blue, you can only process the data of the
selected layer. When the button is highlighted in blue, the direction of process is
limited.
4. Drag the pointer in the Y-axis direction to the position where you want to
display the data, and release it.
The data selected in step 1 move in the Y-axis direction.
You cannot change the display position of the data that were displayed first in the
Data Slate window.
1. Select the data whose position you want to change in the X-axis direction.
For the operation procedure, refer to the previous section, “■Changing the
selected data”.
2. Click the button on the pointer bar.
The button is highlighted in blue.
3. Move the pointer to the processed-data display area. While pressing the Alt
and Shift key, hold down the left mouse button.
4. Drag the pointer in the X-axis direction to the position where you want to
display the data, and release it.
The data selected in step 1 move in the X-axis direction.
You cannot change the display position of the data that were displayed first in the
Data Slate window.
8-10 NMECZ-USP-1
8 DATA SLATE WINDOW
1. Select the data whose display position you want to change in the X- and Y-axis
directions.
For the operation procedure, refer to the previous section, “■Changing the
selected data”.
2. Click the button on the pointer bar.
The button is highlighted in blue.
3. Move the pointer to the processed-data display area. While pressing the Alt
key, hold down the left mouse button.
4. Drag the pointer in the X- and Y-axis direction to the position where you want to
display the data, and release it.
The data selected in step 1 move in the X- and Y-axis direction.
You cannot change the display position of the data that were displayed first in the
Data Slate window.
■ Moving the display positions of all data to the initial positions (where the
data were displayed first)
■ Moving the display position of the selected data to the initial position
(where the data were displayed first)
1. Select the data whose position you want to move to the initial position.
For the operation procedure, refer to the previous section, “■Changing the
selected data”.
2. Click one of the buttons on the pointer bar.
3. Move the mouse pointer to the button, and hold down the left mouse button.
A list of key commands appears.
4. Drag the pointer to Recenter [R], and release the mouse button.
This moves the display position of the data that you selected in step 1 to the initial
position.
NMECZ-USP-1 8-11
8 DATA SLATE WINDOW
■ Deconvolution
When you load data into the Data Slate window, the program also loads the information on
deconvolution included in the data. If the deconvolution procedure was carried out in the
1D Processor window, when you load the data to the Data Slate window, you can display
the deconvolution results by right-clicking the mouse button and selecting Options —
Peak — Deconvolve to turn on the display option.
■ Comments
Comments are also saved together with the data. You can display the comments by
right-clicking the mouse button and selecting Options — Annotation to turn on the
display option.
You can also type comments in the Data Slate window in the same manner as in the 1D
Processor window.
The spread sheet tool is used to display or print the information on the data, such as peaks,
integration and attributes.
You can use the spread sheet in the Data Slate window in the same manner as in the 1D
Processor window.
■ Printing data
You can use the button in the Data Slate window in the same manner as in the 1D
Processor window.
8-12 NMECZ-USP-1
SPREAD SHEET
9
The spread sheet tool is used to display or print information on data,
such as peaks, integration and attributes.
NMECZ-USP-1 9-1
9 SPREAD SHEET
If you carry out peak pick and integration, the results of peak pick and integration are
displayed in the order of ppm in the Spread Sheet window. The results of peak pick and
integration can be distinguished by the symbol in Class in the Spread Sheet window as
shown in the Table 9.1.
Table 9.1 Meaning of the symbol in Class
Symbol in Class Meaning
A Automatic peak picking
M Manual peak picking
AI Automatic integration
MI Manual integration
Presently selected peak picking (X : A
SX
or M) or integration (X : AI or MI)
D Deconvolution
9-2 NMECZ-USP-1
9 SPREAD SHEET
1. In the Spread Sheet window, click the blank cell under the button.
1. Move the mouse pointer onto the item you want to delete in the Spread Sheet
window.
NMECZ-USP-1 9-3
9 SPREAD SHEET
When the peak pick and integration are currently selected in the Data Processor window,
the letter “S” is displayed in Class and the row is highlighted in blue.
Highlighted
in blue
In the above example, two lines are displayed in blue, showing that the peak and
integration curves are selected at present.
Filename box
9-4 NMECZ-USP-1
9 SPREAD SHEET
NMECZ-USP-1 9-5
10 3D NMR DATA
PROCESSING
10.1 DISPLAY OF 3D DATA ............................................................... 10-1
10.1.1 3D Viewer Window...................................................................... 10-1
10.1.2 Data Slate Window....................................................................... 10-2
10.2 SLICE OPERATION OF 3D DATA ........................................... 10-3
10.2.1 How to Use 3D Slicer .................................................................. 10-3
10 3D NMR DATA PROCESSING
3D data
display area
XZ plane
display area
YZ plane
display area
XY plane
display area
1D data of X-axis
display area
1D data of Y-axis 1D data of Z-axis
display area display area
NMECZ-USP-1 10-1
10 3D NMR DATA PROCESSING
If you want to change the angle to view the spectrum, click the button and drag
the spectrum in the window.
If you select Options — Rotate, the spectrum automatically rotates. If you want to
change the rotation speed and the rotation axis, select Options — Rotation Speed,
and adjust them.
10-2 NMECZ-USP-1
10 3D NMR DATA PROCESSING
1. Select View — 3D Slicer from the menu bar in the Delta Console window.
2D data
display area
1D data
display area
3D data
display area
NMECZ-USP-1 10-3
10 3D NMR DATA PROCESSING
3. Select Layout — Show XZ 2D from the menu bar in the 3D Slicer window.
The projection data on the XZ plane is displayed in the 2D data display area.
4. Select Layout — Show 2D Slices from the menu bar in the 3D Slicer window.
The slice data of the XZ plane is displayed in the 2D data display area. At this time,
the data in the 1D data display area changes to the projection data on the Y-axis.
10-4 NMECZ-USP-1
10 3D NMR DATA PROCESSING
If you want to overlay an additional slice data at another point, select Layout —
Overlay 2D Slices, and select another 1D data point in the same manner as
above.
NMECZ-USP-1 10-5
HOW TO USE
11
PROCESSING FUNCTIONS
The Delta program has various processing functions for data processing.
Chapter 11 explains various processing functions in the order that they
displayed in the pull-down menu of 1D and nD Processor windows.
NMECZ-USP-1 11-1
11 HOW TO USE PROCESSING FUNCTIONS
Item Explanation
Flip Performs forward linear interpolation.
Order: The order of operation
FTS Creates a data file consisting of original data from which noise is
artificially eliminated.
No.Peaks: Specifies the number of peaks to extract.
Threshold: Specifies threshold.
Extend: Expansion ratio.
Phase: Whether or not to apply the automatic phase
correction.
FTS Automatic Performs FTS automatically.
Extend: Expansion ratio.
Phase: Whether or not to apply the automatic phase
correction.
Gauss Echo Shift Performs Gaussian window and its shift value for MQMAS.
Width: Spectrum width
High Pass Performs high-pass digital filtering.
K: Reciprocal of the filter width
Hypercomplex Changes the Hypercomplex Transpose saving style.
Transpose For example; (RR,RI,IR,II) –> (RR,IR,RI,II)
Dim: Axis to change the Hypercomplex Transpose style.
Inverse Chebyshev Performs Inverse Chebyshev IIR digital filter in the range specified by
Start and Stop.
Start: Start position
Stop: Stop position
Linear Prediction Performs linear prediction.
Order: The order of the operation.
Sample: Number of points to sample from the file.
Reconstruct: Number of points in the file to reconstruct.
Notch Filter Performs notch filtering at the point specified by Shift value.
K: K factor
PN to SHR Transforms P/N type 2D data to States-Haberkorn-Ruben type data.
Dim: Axis to transform.
SHR to PN Converts States-Haberkorn-Ruben type data to P/N type data.
Dim: Axis to convert.
Shift Shifts data by the number of data points specified by Shift.
Zero fill Performs zero filling by the number of data points specified by Times.
11-2 NMECZ-USP-1
11 HOW TO USE PROCESSING FUNCTIONS
11.2 WINDOW
Item Explanation
Blackman Applies the Blackman window function.
Blackman-Harris Applies the Blackman-Harris window function.
Cosine Applies the Cosine window function.
GAUSS Applies the Gaussian window function that is specified by Width and
Shift.
Hamming Applies the Hamming window function.
Kaiser Applies the Kaiser window function which is specified by πUnits.
Lorentzian-Gaussian Applies the Lorentzian-Gaussian window function.
Sine Applies the Sine window function.
Single Exponential Applies the Single Exponential window function with specified Width.
You can also specify the Shift value.
Single Exponential Performs the Single Exponential window function, setting the
Automatic parameters automatically by specifying decay rate with Factor.
Sinebell Automatic Applies the Sinebell window function by setting the parameters
automatically.
Sinebell Applies the Sinebell window function with the specified Shift and
Size.
Sinebell (order 2) Applies the Sinebell (order 2) window function with the specified Shift
and Size.
Sinebell (order 4) Applies the Sinebell (order 4) window function with the specified Shift
and Size.
Sinebell Shift Applies the N-shifted-Sinebell window function with the specified π/n
value.
Traf Applies the Traf window function with the specified width and Shift.
Traf Modified Applies the Traf window function with the specified width, Shift and
Pre-Slope.
Trapezoid Applies the trapezoidal window function with the specified P1, P2 and
P3.
Trapezoid 3 Applies the trapezoidal window function with the specified P2, P3 and
P4.
NMECZ-USP-1 11-3
11 HOW TO USE PROCESSING FUNCTIONS
11.3 TRANSFORM
Item Explanation
Covariance NMR The covariance NMR processing algorism.
Covariance: Performs the covariance process for the
homonuclear NMR measurement.
INADEQUATE Covariance:
Performs the covariance process for the
INADEQUATE data.
Indirect Covariance: Performs the covariance process for the
measured data of the heteronuclear correlation
spectrum.
DOSY The DOSY measurement data processing algorism.
DOSY CONTIN: Continuous type inverse Laplace
transformation. This is suitable for the data of a
system with a continuous molecular weight.
DOSY SPLMOD: Discrete type inverse Laplace transformation.
This is suitable for data in which different
component peaks are overlapped.
L-Marquardt: Nonlinear curve fitting. This is suitable for data
with few peak overlaps.
MCR: Multivariate Curve Resolve method.
This feature is that the calculation has few
parameters to set.
MCR-CONTIN: Combination of the MCR and CONTIN
methods.
MCR-LM: Combination of the MCR and L-Marquardt
methods.
MCR-SPLMOD: Combination of the MCR and SPLMOD
methods.
Decay Analysis Analyzes the decay curve.
CONTIN: Continuous type inverse Laplace
transformation. This is suitable for the data of a
system with a continuous molecular weight.
L-Marquardt: Nonlinear curve fitting. This is suitable for the
data with few overlaps of a peak.
MCR: Multivariate Curve Resolve method
This feature is that the calculation has few
parameters to set.
MCR-CONTIN: Combination of the MCR and CONTIN
methods.
MCR-LM: Combination of the MCR and L-Marquardt
methods.
SPLMOD: Discrete type inverse Laplace transformation.
This is suitable for data in which different
component peaks are overlapped.
FDM 1D Processing of the filter diagonal matrix (FDM).
FFT Fourier transforms.
Scale: Sets the multiplier for the first FID data point.
Clip: For data measured by using digital filtering, it is
usually set to TRUE. If TRUE, the number of
data points is reduced after Fourier
transformation.
Flatten Reduces the order of the multidimensional data.
11-4 NMECZ-USP-1
11 HOW TO USE PROCESSING FUNCTIONS
Item Explanation
Hilbert Hilbert transforms: Creates an imaginary data set from a real
frequency-domain data set.
IFFT Inverse Fourier transforms.
Maximum Entropy Creates Frequency domain data from time domain data using the
Maximum Entropy method. Specify Times, Normal values and Noise
values.
NMR Pipe Transforms NMR Pipe format data and prints out to a specified file.
Projection Processes the data measured by the projection reconstruction
Reconstruction measurement method.
Shear Performs SHEARING processing for 2D data measured by the
MQ-MAS method.
Sum Plane Sums data along the axis.
Transpose Transposes matrix data. This cannot be applied for 1D data.
Unfold Unwrap folded data connected at two sweep widths simultaneously.
NMECZ-USP-1 11-5
11 HOW TO USE PROCESSING FUNCTIONS
Item Explanation
Phase Apply Phase: Applies P0, P1 and PP, which are set to numerical
values, to actual data file.
Automatic Phase: Corrects phase by P0 and P1 automatically.
When Hold P1 is set to TRUE, the phase is
corrected by only P0.
Digital Filter Phase: Correct for phase distortion caused by digital
filter to the original phase.
Machine Phase: Corrects phase automatically taking pulse width,
filter type, filter-band width, dead time and delay
time into considering.
Phase Correct: Corrects phase on the basis of the phase
information written in the pulse sequence.
Phase Exclude: Eliminates the information on the data specified
by Center and Width, and corrects phase
automatically, taking pulse width, filter type,
filter-band width, dead time and delay time into
consideration.
Reference Performs Deconvolution with Lorentzian for the peak that is specified
Deconvolution by Center, Width and Line_Width values.
Ridge Detects a ridge and eliminates it in the range which is specified by
StartCoord and StopCoord.
Scale Adjusts the scale to the specified value.
Smooth Performs smoothing processing.
Sub Set Selects a part of the data and outputs it. Select dimension with Dim
and specify the range by Start and End.
Symmetrize Performs symmetrization. Select either Cosy, J-Resolve, Inadequate or
Filter for processing type. Moreover, you can specify Threshold.
Tilt Performs Tilt processing with the specified Angle value.
Zero Region Replaces the region specified by Start and Stop with the value of
ZERO or NOIZE.
11-6 NMECZ-USP-1
11 HOW TO USE PROCESSING FUNCTIONS
11.5 DISPLAY
Item Explanation
Copy Data Copies the data to the specified directory.
Display Signal / Noise Displays the SN value calculated by the specified method for the
specified range in the Delta Console window.
Display window Displays processed data in the Data Slate window.
Export File Converts the data to the file type specified by Format, and outputs it in
the foreign_data directory.
Hertz Sets unit to Hertz.
Integrals Apply Integral: Performs integration according to the integration
range specified by TemplateFile which is created
by the Transfer Integral Tool.
Integrate Region: Performs integration in the range that is specified
by Start and Stop.
Mask: Sets the signal intensity in the range where the
data is not integrated to zero.
Normalize Smallest Integral :
Sets the Normal value to the integration value of
the minimum peak and performs the same
processing as Auto Integrate.
Normalize Tallest Integral :
Sets the Normal value to the integration value of
the maximum peak and performs the same
processing as Auto Integrate.
Peak Deconvolve: Performs deconvolution.
Peak Chop: When the difference between the height of the
highest peak specified by Peaks and the height of
the next highest peak is larger than the value
specified by Max, the parts of peaks whose height
is smaller than that of Max are displayed.
Peak Limit: Restricts the number of peak-pickings to the
number specified by Limit.
Print Peak Width: Calculates the line width of an arbitrary height
peak. You can confirm the result by Report of File
Information.
Solvent peak limit: Omits solvent peaks or reference peaks.
Peak Pick / Integrate Performs peak picking automatically with the value specified by
J-coupling, detects the peaks having peak width specified by
IntegWidth, and integrates them.
Plot Plot: Prints.
Plot Expand Integrals:
Prints the integrated areas of data.
Presentation Plot: Prints with the format specified by Presentation
Manager.
Print Integral List: Prints integrations list.
Print Peak List: Prints peak lists.
PPM Set unit to ppm.
NMECZ-USP-1 11-7
11 HOW TO USE PROCESSING FUNCTIONS
Item Explanation
Reference Automatic Reference:
Sets reference automatically.
Copy Reference: Specifies source list and copies the reference
information.
Linear Reference: Sets the position specified by Center to the value
specified by Beta as a reference.
Moreover, expands/contracts the scale by the
value specified by Alpha.
Performs the reference process by linear
transformation.
x' = α∗(x-center) + β
Maximum Reference:
Detects the maximum peak in the range specified
by Ref and Width, and sets the detected peak to
the reference specified by Ref.
Reference: Sets the position specified by Position to the value
specified by Reference.
Unreference: When a reference has already been set, returns to
the original state.
Save Saves the processed data with the name specified by Filename.
Scale Axis Range Changes the scale of the axis.
Stop Stops data processing.
Threshold Expands threshold level with the value specified by Multi.
TSC Database Creates TSC data base file.
11-8 NMECZ-USP-1
DISPLAY FUNCTIONS OF
12
PROCESSING DATA
If you click the right button on a data processing screen, such as 1D
Processor, Data Slate and 2D Viewer, a pop-up menu appears.
Chapter 12 explains the contents of the pop-up menu. The displayed
items change according to the display window and the set-up mode.
Pop-up menu
NMECZ-USP-1 12-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Pop-up menu for every mode class set in the pointer bar
The pop-up menu changes with the mode of the pointer bar (Fig. 12.2).
At modes except :
Measure, Cursor, Annotation,Create rectangle
(Annotation), and Create arrow (Annotation) modes
Level Tool [¥]
Layers
Options
Display Modes
Draw Components
Logarithm Base
Display Units
Save Current View [V]
Restore Saved View [Shift-V]
Export To
Copy Geometry to Clipboard [^+C]
Copy Attributes top Paste Butter
At Annotation mode
Level Tool [¥]
Layers
Options
Display Modes
Draw Components
Logarithm Base
Display Units
Color
Font
Size
Attribute
Outline
Rotation
Alignment
Save Current View [V]
Restore Saved View [Shift-V]
Export To
Copy Geometry to Clipboard [^+C]
12-2 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Menu items
NMECZ-USP-1 12-3
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
Annotations [Alt-A] Displays the notes that are set in the Annotation mode.
Cursors [Alt-C] Displays the cursor that is set in the Cursor mode.
Measure [Sh-Alt-H] Displays the measuring range that is set in the Measure mode.
PiP Windows [Alt-W] Displays the area of the PIP box that is set in the PIP mode on the spectrum.
Regions [Alt-R] Displays the area that is set in the Region mode.
Integral Integrals [Alt-I]: Displays the integration curves that are set
in the Integral mode.
Integral Value [Alt-Q]: Displays the integration value.
Peak Integration [Alt-E]: Integrates the peaks that are picked up.
Peak Center of Mass [Sh-Alt-W]: Calculates the center of the peak from the
integration values and peak positions of two
peaks, and displays it.
Deconvolve [Alt-K]: Displays the result of deconvolution.
12-4 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
Data Data [Alt-D]: Displays data.
Data Points[Alt-N]: Displays the marker on each data points.
Statistics [Alt-T]: Displays base level, noise level and
threshold level of peak.
NMECZ-USP-1 12-5
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
Show Ruler Type Display ruler type on the x-axis.
[Sh-Alt-G]:
Show Ruler Unit: Display a unit of the scale.
Display Clock Time X: Displays with human time.
■ Logarithm Base
Item Explanation
2 Displays as base two logarithms.
e Displays as natural logarithms.
10 Displays as common logarithms.
■ Display Units
Item Explanation
Hertz Displays the x-axis in Hertz.
ppm Displays the x-axis in ppm.
Percent Displays the x-axis in percent.
Points Displays the x-axis in points.
Reset Returns to the initial settings.
Cycle [Sh+I] Returns to the settings of the previous step.
Peak Units Hertz: Displays the chemical shift of picked peak in Hertz.
ppm: Displays the chemical shift of picked peak in ppm.
Percent: Displays the chemical shift of picked peak in percent.
Points: Displays the chemical shift of picked peak in points.
Ruler Units: Displays the chemical shift of picked peak in the same unit
as that of the x-axis ruler.
Measure/Cursor Unit Hertz: Displays the unit of measure and cursor as hertz.
ppm: Displays the unit of measure and cursor as ppm.
Percent: Displays the unit of measure and cursor as percent.
Points: Displays the unit of measure and cursor as points.
Ruler Units: Displays the unit of measure and cursor in the same unit as
that of the x-axis ruler.
■ Measure/Cursor Labels
Item Explanation
Show Measure X Shows the X distance of measure and the X position of cursor when
Measure/Cursor is not selected.
Show Measure Amplitude Shows the peak intensity at the Measure/Cursor position when
Measure/Cursor is not selected.
Show Measure All Shows information about unselected Measure/Cursor.
Show Measure None Does not show information about an unselected Measure/Cursor.
12-6 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
6 to 38 Changes the font size in a range from 6 to 38.
Default Sets the font size to the default value.
Item Explanation
1 to 12 Changes decimal places in a range from 1 to 12 digits.
Default Sets decimal places to the default value.
■ Draw Components
Item Explanation
Draw Real [Sh-1] Displays real data.
Draw Imag [Sh-2] Displays imaginary data.
■ Export To
Item Explanation
Data Processor[Alt-Ctrl-1] Opens the data by 1D Data Processor.
Data slate [Alt-Ctrl-2] Opens the data by Data Slate.
Data Slicer [Alt-Ctrl-2] Opens the data by Data Slicer.
Phaser [Alt-Ctrl-2] Opens the data by 2D Phaser.
Viewer [Alt-Ctrl-2] Opens the data by 2D Viewer.
NMECZ-USP-1 12-7
12 DISPLAY FUNCTIONS OF PROCESSING DATA
12-8 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
The pop-up menu changes with the mode of pointer bar (Fig. 12.4).
At modes except : At Annotation mode
Measure, Cursor, Annotation, Create rectangle Level Tool [¥]
(Annotation), and Create arrow (Annotation) modes Layers
Options
Level Tool [¥] Display Modes
Layers Draw Components
Options Logarithm Base
Display Modes Display Units
Draw Components Color
Logarithm Base Font
Display Units Size
Peak Positions Attribute
Datadex 1D Outline
Datadex 2D Rotation
Save Current View [V] Alignment
Restore Saved View [Shift-V] Peak Positions
Contour Bounds Datadex 1D
Export To Datadex 2D
Copy Geometry to Clipboard [^+C] Save Current View [V]
Copy Attributes top Paste Butter Restore Saved View [Shift-V]
Contour Bounds
Export To
Copy Geometry to Clipboard [^+C]
At Measure mode
Level Tool [¥]
Layers At Create arrow
Options (Annotation) mode
Display Modes Level Tool [¥]
Draw Components Layers
Logarithm Base Options
Display Units Display Modes
Measure Labels Draw Components
Measure Font Size Logarithm Base
Measure Decimal Precision Display Units
Peak Positions Line Style
Datadex 1D Color
Datadex 2D Peak Positions
Save Current View [V] Datadex 1D
Restore Saved View [Shift-V] Datadex 2D
Contour Bounds Save Current View [V]
Export To Restore Saved View [Shift-V]
Copy Geometry to Clipboard [^+C] Contour Bounds
Export To
Copy Geometry to Clipboard [^+C]
At Cursor mode
At Create rectangle
Level Tool [¥]
Layers
(Annotation) mode
Options Level Tool [¥]
Display Modes Layers
Draw Components Options
Logarithm Base Display Modes
Display Units Draw Components
Cursor Labels Logarithm Base
Cursor Font Size Display Units
Cursor Decimal Precision Color
Peak Positions Peak Positions
Datadex 1D Datadex 1D
Datadex 2D Datadex 2D
Save Current View [V] Save Current View [V]
Restore Saved View [Shift-V] Restore Saved View [Shift-V]
Contour Bounds Contour Bounds
Export To Export To
Copy Geometry to Clipboard [^+C] Copy Geometry to Clipboard [^+C]
NMECZ-USP-1 12-9
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Menu items
12-10 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
Annotations[Alt-A] Displays notes when a mode other than Annotation is selected.
Cursors[Alt-C] Displays a cursor when a mode other than Cursor is selected.
Measure[Sh-Alt-H] Displays the measuring range when a mode other than Measure is selected.
PiP Windows[Alt-W] Displays the area of the PIP box on the spectrum when a mode other than PiP
is selected.
Regions[Alt-R] Displays the area of the created region when a mode other than Region is
selected.
Integral Integrals [Alt-I]: Displays the integration curves that are set
in the Integral mode.
Integral Value [Alt-Q]: Displays the integration value.
Peak Center of Mass [Sh-Alt-W]: Calculates the center of the peak from the
integration values and peak positions of
two peaks, and display it.
Peak Pick [Alt-P]: Displays the peak information when in
any mode other than Peak mode.
Sequences: On the basis of the information which is
written in the spreadsheet, draws the line
between peaks and controls the display of
the peak information.
Sequence Marker [Alt-M]: Displays cursor on the selected peak after
peak-picking.
Molecules Atom Numbers [Sh-Alt-A]: Numbers the skeleton of molecular
structural formula that is displayed on the
geometry.
Atom Select Numbers Numbers the selected skeleton.
[Sh-Alt-S]:
Atom Sequence [Sh-Alt-Q: Displays the correction between the
selected skeleton and peaks.
Molecule Border [Sh-Alt-D]: Displays a frame on the molecule.
Molecule Title [Sh-Alt-T]: Displays the title of molecule.
Data Data [Alt-D]: Displays data.
Data Points [Alt-N]: Displays data points.
Line Hinding [Alt-L]: In the stack display, does not display the
behind line of the stacking data.
Statistics [Alt-T]: Displays base level, noise level and
threshold level of peak.
Center Contour Levels Corrects indication of contour plot based
Around Baseplane Portion: on the statistics that is set initially.
Extra Data Color Solid: Changes color when projection is
performed on the surface of the cube in
3D display.
Contour Symmetrically: Makes the positive and negative contour
lines symmetrical.
NMECZ-USP-1 12-11
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Item Explanation
Display Grid [Alt-G]: Displays grid lines.
Perspective [Alt-V]: Changes display in 3D display.
PiP Zoom Factor Displays the zoom ratio to the original
spectrum.
Plot Border [Sh-Alt-B]: Display the frame of the Plot box.
Layer Tabs [Sh-Alt-O]: In case of spectra are displayed as stacking
layer, displays tabs which select each
spectrum.
Scan to Threshold [Sh-Alt-R]: Turn this on if you want to search a slice
data set automatically using threshold
level.
Show Zoom Overview Displays the zoom navigator.
Window [Sh-tab]:
Wrap data [Sh-Alt-Space]: Wraps the data.
File Information All Lagers [Alt-O]: In overwriting, displays file names of all
spectra.
Clipped Indicator: Displays whether FID overflows (is
clipped) or not.
Comment [Sh-Alt-O]: Displays comment.
File Name [Alt-F]: Displays file name.
Selected Header Params Displays the selected parameters.
[Sh-Alt-G]:
Shadow Dimensions [Alt-B]: Displays the position of a slice data in a
unit, such as point, percent or second.
Ruler X Rule [Alt-X]: Displays the x-axis.
Y Ruler [Alt-Y]: Displays the y-axis.
Z Ruler [Alt-Z]: Displays the z-axis.
Logarithmic X Ruler Displays the x-axis with logarithmic
[Sh-Alt-X]: scaling.
Logarithmic V Ruler Displays the y-axis with logarithmic
[Sh-Alt-Y]: scaling.
Ruler Unit Abbreviated: Omits the ruler unit.
Show Axis Name: Displays the axis name
Show Number of Points Displays point numbers on the x-axis.
[Sh-Alt-E]:
Show Ruler Type [Sh-Alt-G]: Display the ruler type on the x-axis.
Show Ruler Unit: Displays the unit of the scale.
Display Clock Time X: Displays the x-axis with human time.
Display Clock Time Y: Displays the y-axis with human time.
■ Logarithm Base
Item Explanation
X Ruler Selects logarithm display from 2, e and 10.
Y Ruler Selects logarithm display from 2, e and 10.
12-12 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Display Units
Item Explanation
X Ruler Hertz Displays the x-axis in Hertz.
ppm Displays the x-axis in ppm.
Percent Displays the x-axis in percent.
Point Displays the x-axis in points.
Reset Returns to the initial settings.
Cycle [Shift-I] Switches the X-axis unit in the order of Hertz, ppm, Percent and Point.
Peak Units Hertz: Displays the chemical shift of the picked peak in hertz.
ppm: Displays the chemical shift of the picked peak in ppm.
Percent: Displays the chemical shift of the picked peak in
percent.
Point: Displays the chemical shift of the picked peak in point.
Ruler Displays the chemical shift of the picked peak in the
Units: same unit as that of the X-axis ruler.
Measure/Cursor Hertz: Displays the unit of measure and cursor as hertz.
Units
ppm: Displays the unit of measure and cursor as ppm.
Percent: Displays the unit of measure and cursor as percent.
Point: Displays the unit of measure and cursor as points.
Ruler Displays the unit of measure and cursor in the same unit
Units: as that of the x-axis ruler.
Y Ruler Hertz Displays the y-axis in hertz.
ppm Displays the y-axis in ppm.
Percent Displays the y-axis in percent.
Point Displays the y-axis in points.
Reset Returns to the initial settings.
Cycle [Shift-I] Switches the X-axis unit in the order of Hertz, ppm, Percent and Point.
Peak Units Hertz: Displays the chemical shift of the picked peak in hertz
ppm: Displays the chemical shift of the picked peak in ppm
Percent: Displays the chemical shift of the picked peak in percent
Points: Displays the chemical shift of the picked peak in point
Ruler Displays the chemical shift of the picked peak in the
Units: same unit as that of the y-axis ruler.
Measure/Cursor Hertz: Displays the unit of measure and cursor as hertz.
Units
ppm: Displays the unit of measure and cursor as ppm.
Percent: Displays the unit of measure and cursor as percent.
Points: Displays the unit of measure and cursor as points.
Ruler Displays the unit of measure and cursor in the same unit
Units: as that of the y-axis ruler.
NMECZ-USP-1 12-13
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Peak positions
Item Explanation
Show Peak Pos X Displays position on the x-axis for the selected peak.
Show Peak Pos Y Displays position on the y-axis for the selected peak.
Show Peak Pos All Displays position on the x and y axes for the selected peak.
Show Peak Pos None Does not display information on Peak Positions.
■ Measure/Cursor Labels
Item Explanation
Show Measure X Shows the x-length of the measure and the x-position of the cursor when
Measure/Cursor is not selected.
Show Measure Y Shows the y-length of the measure and the y-position of the cursor when
Measure/Cursor is not selected.
Show Measure Dist Shows the length of the measure and the signal intensity at the position
of the cursor when Measure/Cursor is selected.
Show Measure All Shows the information on the Measure/Cursor.
Show Measure None Does not show the information on the Measure/Cursor.
Item Explanation
6 to 38 Changes the font size in a range from 6 to 38.
Default Sets the font size to the default value.
Item Explanation
1 to 12 Changes decimal places in a range from 1 to 12 digits.
Default Sets decimal places to the default value.
■ Draw Components
Item Explanation
Draw Above Baseplane Displays the positive spectrum.
[Sh-1]
Draw Below Baseplane Displays the negative spectrum.
[Sh-2]
■ Datadex1D
Item Explanation
X Displays the slice data on the x-axis for the selected area in the Data
Slate.
Y Displays the slice data on the y-axis for the selected area in the Data
Slate.
12-14 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Datadex2D
Item Explanation
X,Y Returns to the 2D display.
■ Contour Bounds
Item Explanation
Set Upper Bound Sets the signal intensity on the cursor position to the upper limit for the
contour plot.
Set Lower Bound Sets the signal intensity on the cursor position to the lower limit for the
contour plot.
Set Threshold Sets a bias for the contour plot. Clicking the desired contour line by the
middle button of a mouse sets the height of the contour line to the value
of the threshold.
■ Export To
Item Explanation
Data Processor[Alt-Ctrl-1] Opens the data by nD Data Processor.
Data slate [Alt-Ctrl-2] Opens the data by Data Slate.
Data Slicer [Alt-Ctrl-2] Opens the data by Data Slicer.
Phaser [Alt-Ctrl-2] Opens the data by 2D Phaser.
Viewer [Alt-Ctrl-2] Opens the data by 2D Viewer.
NMECZ-USP-1 12-15
12 DISPLAY FUNCTIONS OF PROCESSING DATA
Pop-up menu
The pop-up menu changes with the setting of mode and the displayed data (Fig. 12.6).
At Level mode At Color mode
Preserve On/Off State Preserve On/Off State
Preserve Attributes Preserve Attributes
Statistics [Alt-T] Statistics [Alt-T]
1 Level Per Button [1] 1 Level Per Button [1]
3 Level Per Button [3] 3 Level Per Button [3]
7 Level Per Button [7] 7 Level Per Button [7]
Line Limit [L] Line Limit [L]
Drawing Options Drawing Options
Auto Apply [A] Auto Apply [A]
Auto Apply All [U] Auto Apply All [U]
Switch to Colors [C] Switch to Levels [C]
Reset Levels [R] Reset Color Button [R]
Put Away[Esc] Save Options[S]
Put Away[Esc]
12-16 NMECZ-USP-1
12 DISPLAY FUNCTIONS OF PROCESSING DATA
■ Menu items
Item Explanation
Smooth Performs antialiasing of the color display in the geometry.
Dither Performs dither processing for the geometry display.
Blend For the 3D geometry display, smoothes the border between the contour
line and the background.
Backface Culling Neglects the object behind the object displayed on the geometry of the
3D data.
Double Side Performs the double side process to the object displayed on the
geometry of the 3D data.
Never Transparent Does not Perform the transparence process to the object displayed on
the geometry of the 3D data.
Always Transparent Always performs the transparence process to the object displayed on the
geometry of the 3D data.
NMECZ-USP-1 12-17
INDEX
B F
Baseline correction..................................................... 2-15 Fast-Fourier transformation ....................................... 2-11
Automatic baseline correction ............................... 2-15 FFT ............................................................................ 2-11
Manual baseline correction ................................... 2-15 FID.................................................................... 1-40, 1-45
Other functions in the iBase Correct window........ 2-17 File Search window ................................................... 1-42
“Batch Process Files” window.................................... 6-1
G
C
Gain mode ................................................................. 1-11
Calculator panel ........................................................... 9-5 Geometry ..................................................................... 8-3
Combine Files window ................................................ 7-1 Global directory ......................................................... 1-41
Group number ............................................................ 1-40
NMECZ-USP-1 I-1
Inporting data ........................................................ 1-51
H Saving NMR data .................................................. 1-46
I-2 NMECZ-USP-1
Executing the process list ...................................... 1-37 Threshold mode ......................................................... 1-13
Loading the process list ......................................... 1-34 Tool bar ......................................................... 2-1, 2-2, 3-2
Saving the process list ........................................... 1-36 “Transfer Integral Tool” window ............................. 2-30
Standard process list .............................................. 1-26
Processed data
Adding and subtracting ........................................... 5-5 U
Displaying additional processed data ...................... 8-3
Loading a data file to the specified window .......... 1-41
Unit for axis ............................................................... 2-18
Saving processed data ........................................... 1-47
Searching for data ................................................. 1-42
Process-list control button ............................................ 3-1
Projection ................................................................... 3-17 V
Partial projection ................................................... 3-17
Projection mode .................................................... 3-17 Version number ......................................................... 1-40
View Control window................................................ 2-34
R
W
Reference mode ......................................................... 1-13
Reference setting ............................................... 2-18, 3-23 Window function .................................................. 2-7, 3-6
Region mode .............................................................. 1-16 Changing the parameters
for the window function .............................. 2-8, 3-7
Changing the window function ........................ 2-9, 3-8
S Checking the window function
using 1D slice data ........................................... 3-10
Setting a window function ............................... 2-7, 3-6
Select mode ................................................................ 1-12
Setting multiple window functions ................ 2-10, 3-9
Spectrum
Expanding a spectrum ........................................... 1-17
Expanding or reducing a spectrum ........................ 1-21
Moving a spectrum................................................ 1-19 Y
Setting display range of spectrum ................ 2-34, 3-34
Spread sheet Y-axis gain
Spread sheet tool ................................................... 8-12 Changing the Y-axis gain...................................... 1-24
Spread Sheet window .............................................. 9-1
Statistic ........................................................................ 2-2
Symmetrization .......................................................... 3-12 Z
Zero-filling ........................................................... 2-6, 3-5
T Zoom mode ................................................................ 1-11
Short cut key to the Zoom mode ........................... 1-17
T1 noise
Eliminating the T1 noise ....................................... 3-33
NMECZ-USP-1 I-3