ASTM E 2698 - 10 RT - Digital
ASTM E 2698 - 10 RT - Digital
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
--`,,`,,,````````,,`,,,`,``,,-`-`,,`,,`,`,,`---
1⁄2 of the measured unsharpness.
3.2.2 energy—a property of radiation that determines the
penetrating ability. In x-ray radiography, energy machine rating 3.2.15 ghosting—residual signal or image from a prior
is determined by kilo electron volts (keV), million electron exposure in a current image. Signal or image can be negative
volts (MeV). In gamma ray radiography, energy is a charac- or positive and may affect interpretation of the image.
teristic of the source used. 3.2.16 bad pixel—a pixel identified with a performance
3.2.3 like section—a separate section of material that is outside of the specification range for a pixel of a DDA as
similar in shape and cross section to the component or part defined in Practice E2597.
being radiologically inspected, and is made of the same or 3.2.17 relevant cluster—a cluster with a cluster kernel pixel
radiologically similar material. (CKP), where there are fewer than five good neighboring
3.2.4 material group—materials that have the same pre- pixels surrounding this pixel as defined in Practice E2597. A
dominant alloying elements and which can be examined using CKP is a pixel that does not have sufficient good neighboring
the same IQI. A listing of common material groups is given in pixels to perform interpolation, and is therefore not correctable.
Practice E1025.
3.2.5 NDT facility—the facility or entity performing the 4. Significance and Use
radiologic examination. 4.1 This practice establishes the basic parameters for the
3.2.6 digital detector array (DDA)—an electronic device application and control of the digital radiologic method. This
that converts ionizing or penetrating radiation into a discrete practice is written so it can be specified on the engineering
array of analog signals which are subsequently digitized and drawing, specification, or contract. It will require a detailed
transferred to a computer for display as a digital image procedure delineating the technique or procedure requirements
corresponding to the radiologic intensity pattern imparted upon and shall be approved by the CEO.
the input region of the device. The conversion of the ionizing
or penetrating radiation into an electronic signal may transpire 5. Basis of Application
by first converting the ionizing or penetrating radiation into
5.1 The following items are subject to contractual agree-
visible light through the use of a scintillating material. These
ment between the parties using or referencing this standard.
devices can range in speed from many minutes per image to
many images per second, up to and in excess of real-time 5.1.1 Personnel Qualification—If specified in the contrac-
radioscopy rates. tual agreement, personnel performing examinations to this
3.2.7 digital driving level (DDL)—for computer graphics standard shall be qualified in accordance with a nationally or
display boards, the digital value that corresponds to a particular internationally recognized NDT personnel qualification prac-
monochrome grayscale level. A particular DDL “drives out” a tice or standard and certified by the employer or certifying
particular visible shade of gray. For example, in an 8-bit agency, as applicable. The practice or standard to be used and
display, a DDL assumes 256 values from 0 to 255. its applicable revision shall be identified in the contractual
agreement between the using parties.
5.1.2 If specified in the contractual agreement, NDT agen-
5
Available from National Institute of Standards and Technology (NIST), 100 cies shall be qualified and evaluated as described in E543. The
Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov. applicable edition of E543 shall be specified in the contract.
--`,,`,,,````````,,`,,,`,``,,-`-`,,`,,`,`,,`---
image display test pattern, in accordance with the requirements
and limitations of each (see Practice E2597 and Guide E2736).
of SMPTE RP 133, shall be configured for the system display
The provider and the user of the examination system should be
resolution and aspect ratio. Alternate test patterns may be used
fully aware of the capabilities and limitations of each system
provided they include the features described in SMPTE RP 133
proposed.
required to perform the quality tests specified in this practice.
7.2 The DDA cannot be operated without computing hard-
ware and software for image acquisition, image display and 7.5.7 The monitor shall be capable of discriminating the
image storage/retrieval. horizontal and vertical low contrast (1 %) modulation patterns
7.2.1 The software shall be capable of acquiring images at the display center and each of the four corner locations.
frame by frame from the DDA and integrating, or averaging the 7.6 Image Quality Indicators (IQI):
frames, or both. 7.6.1 IQIs shall be in accordance with a recognized standard
7.2.2 The software shall perform an image calibration to or approved by the Cognizant Engineering Organization. Hole
correct the inhomogenities of the detector and to determine and plate type indicators shall comply with Practice E1025 or
correct bad pixels (that is, bad pixel map) as defined in Practice Practice E1742, Annex 1. Wire type indicators shall be in
E2597. accordance with Practice E747 and correlated to the hole type
7.3 The software to display resulting imagery from a DDA penetrameters in accordance with Practice E747.
shall be capable to scale images in size (geometrical 7.6.2 The IQI shall be constructed from material in the same
magnification—zoom) and gray levels (window and level material group (see Practice E1025) as the material to be
adjustment—brightness and contrast, for example from 16 bit radiologically inspected. If an IQI material of the same
to 8 bit). Additional functions shall be required such as a line material group is not available, a material that is radiologically
profile measurement, histogram, and statistics window for less dense shall be used.
measuring an region of interest (ROI) (mean and standard
7.6.3 The use of alternative IQIs that are approved by the
deviation).
CEO shall be documented in a written procedure with the
7.4 The Digital Detector Array (DDA):
design, materials designation, and thickness identification or
7.4.1 Only DDAs shall be used in practice as established in
documented on a drawing and that drawing referenced in the
Guide E2736.
procedure.
7.4.2 Users shall comply with the manufacturers’ require-
ments of temperatures and humidity conditions for both 7.6.4 The IQIs shall be procured or fabricated to the
operation and shipping. requirements of Practice E1025 or Practice E747 with a
7.4.3 The DDA shall be calibrated using the manufacturers’ manufacturer’s certification of compliance with respect to alloy
recommendation both for frequency of calibration and the and dimensions. Users shall visually inspect IQIs for damage
method used. Other calibration methods are allowed as long as and cleanliness in accordance with Appendix X1.
approved by the CEO. 7.7 Radiation Sources:
7.4.4 The user shall ensure that all exposures are within the 7.7.1 X-Radiation Sources—Selection of appropriate X-ray
linear operating range of the DDA, using either information voltage and current levels is dependent upon variables regard-
obtained from the manufacturer or data obtained by the ing the specimen being examined (material type and thickness)
user/CEO. and exposure time. The suitability of these exposure param-
7.5 The image display shall meet the following require- eters shall be demonstrated by attainment of the required
ments as a minimum. Alternate image displays or requirements radiographic quality level and compliance with all other
may be used with CEO approval. requirements stipulated herein.
00 1-1T 1 1T 0.7
0 1-2T 1 2T 1.0
1 2-1T 2 1T 1.4
2 2-2T 2 2T 2.0
3 2-4T 2 4T 2.8
A
Expressed as a percentage of material thickness.
B
Expressed as multiple thickness of IQI.
C
Equivalent IQI sensitivity is that thickness of the IQI expressed as a percentage of the specimen thickness in which a 2T hole would be clearly visible under the same
radiologic conditions.
FIG. 1 Flowchart for Setting Geometry, Image Unsharpness and Image Quality (see 10.19)
ymin 5 1.6 * SRb/UIm (1) 10.19.2.3 Third, if the maximum unsharpness is less than
The required value of UIm is given in Table 2. required in Table 2, the inspection shall proceed with the
10.19.2 Determine and Test for Maximum Normalized Im- defined geometry and continue in 10.19. Otherwise, a change
age Unsharpness (UIm): shall be made to decrease the x-ray focal spot size until the
10.19.2.1 First, determine the maximum image unsharp- recomputed image unsharpness meets Table 2. If at the smallest
ness, UIm, that shall be met for the inspection as shown in Table focal spot size, the required total image unsharpness is not met,
2. it means that the DDA being tested is not suitable for the
10.19.2.2 Second, compute the expected image unsharpness inspection.
for the geometry and DDA in use as defined in Guide E1000, 10.19.3 Evaluate IQI Visibility Using Contrast to Noise
Eq. 9 (shown here in a modified form for DDAs): Ratio (CNR):
1 3 10.19.3.1 Once the image unsharpness requirement is met,
UIm 5 y · =U3g 1 ~1.6 · SRb!3 (2)
the image quality (IQI visibility) must be assessed. If the
Ug 5 ~ y – 1 ! * F (3) required hole is not clearly visible it shall be evaluated by
measuring CNR on a hole-type IQI or other method approved
where Ug is the geometrical unsharpness (see Eq 3, where F
and documented by the CEO.
is the focal spot size.
10.19.3.2 For hole-type IQIs, the appropriate IQI shall be
placed on the object and the mean of the pixels inside of the
TABLE 2 Maximum normalized image unsharpness (UIm) hole of interest and a neighbouring region on the IQI (but
Under 1 in. through 1 in. 0.010 in. outside of the hole) shall be measured. The difference in the
(# 25.4 mm) (0.254 mm) two means shall be divided by the standard deviation of the
Over 1 through 2 in. 0.020 in.
(>25.4 through 50.8 mm) (0.508 mm) same region outside of the hole and the resulting quantity
Over 2 through 4 in. 0.030 in. (contrast to noise ratio) shall be at least 2.5 unless otherwise
(>50.8 through 101.6 mm) (0.762 mm) specified by the CEO. If the CNR is sufficient, the DDA and
Over 4 in. 0.040 in.
(>101.6 mm) (1.016 mm) geometry are ready for inspection. If the contrast to noise ratio
is less than the required amount proceed to 10.19.4.
--`,,`,,,````````,,`,,,`,``,,-`-`,,`,,`,`,,`---
--`,,`,,,````````,,`,,,`,``,,-`-`,,`,,`,`,,`---
10.19.4.4 Average multiple frames of data into one image parts or assemblies.
using the manufacturer provided acquisition software. 12.2 Impression stamping, laser marking, or vibro engrav-
10.19.5 After increasing the exposure, if the IQI is still not ing shall be used only where permitted by the applicable
visible, the DDA system is not suitable for inspection. If the specifications or drawings. Unless otherwise specified, mark-
IQI is visible or meets the CNR criteria, the DDA and ing shall be located in areas adjacent to the part number.
geometry are acceptable and ready to proceed with the inspec- 12.3 When impression stamping, laser marking, or vibro
tion. engraving is prohibited, parts may be marked by etching.
10.20 Window Width and Window Level : Suitable echants and application methods shall be employed.
Etching methods other than fluid etching may be used.
10.20.1 The inspections shall be conducted using a defined
12.4 Where etching, impression stamping, laser marking, or
window width. The window width value shall be based on the
vibro-engraving are not appropriate, identification may be
feature of interest and approved by the CEO.
accomplished by dyeing or ink stamping.
10.20.2 Window level shall be adjusted during inspection
12.5 Each part that has successfully passed radiologic ex-
with a fixed window width. Window level adjustment maybe
amination shall be marked as follows:
necessary to view the entire part thickness range.
12.5.1 When impression stamping, laser marking, vibro-
10.21 Image Zoom:
engraving, etching, or ink stamping is applicable, symbols
10.21.1 The inspections shall be conducted with a fixed shall be used. The symbol shall contain an identification
image zoom. The image zoom shall be based on the feature of symbol of the facility.
interest and approved by the CEO. 12.5.2 Except for specialized applications, use the symbol
“X” enclosed in a circle to denote 100 % radiologic examina-
11. Quality Assurance Provisions tion.
11.1 The NDT facility is responsible for furnishing all 12.5.3 When using the sampling method, parts that are
supplies in conformance to contract or purchase order require- actually radiographed shall be marked as specified in 12.5.1.
ments and, unless otherwise specified in the contract or All items in the lot accepted on a sampling basis (part of the lot
purchase order, the performance of all examination require- but not actually radiographed) shall be marked using the
ments contained herein. The examination provisions contained symbol “X”: enclosed in an ellipse.
herein shall become a part of the NDT facility overall exami- 12.6 When dyeing is applicable, blue dye shall be used to
nation system or quality program. indicate 100 % radiologic examination. When sampling is
11.2 The results of all radiologic examinations shall be used, orange dye shall be used to indicate parts accepted on a
recorded and kept on file in accordance with the contract or sampling basis (part of the lot not actually examined), while
purchase order. Additionally, refer to 7.1.7 of Practice E2002. the parts of the lot actually examined shall receive blue dye.
APPENDIX
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
--`,,`,,,````````,,`,,,`,``,,-`-`,,`,,`,`,,`---