Topas 590 TDD Productsheet en

The TDD 590 thermodenuder is designed for precise temperature control up to 400°C to remove moisture and volatile particles from aerosol samples, ensuring accurate particle size measurements. It features high efficiency, long operating life of activated carbon, and is suitable for various applications including aerosol research and emissions analysis. The device operates with minimal loss and utilizes a microprocessor for precise temperature adjustments.

Uploaded by

Rahul Jain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
26 views2 pages

Topas 590 TDD Productsheet en

The TDD 590 thermodenuder is designed for precise temperature control up to 400°C to remove moisture and volatile particles from aerosol samples, ensuring accurate particle size measurements. It features high efficiency, long operating life of activated carbon, and is suitable for various applications including aerosol research and emissions analysis. The device operates with minimal loss and utilizes a microprocessor for precise temperature adjustments.

Uploaded by

Rahul Jain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Thermodenuder TDD 590

Special Advantages
 Exact temperature control up to 400°C
 High efficiency by high residence time and by
high temperature
 Long operating life of the activated carbon by
large storage capacity
 Manufactured with inert material
 Usable as flexible laboratory equipment and as
system element

Thermodenuder TDD 590 Applications


 Particle measurements in hot exhaust gases
The Model TDD 590 thermodenuder is designed to  Separation of vaporous phases off hot aerosols
remove moisture and volatile particles from an
 Removing volatile particles and precursors
aerosol sample for the subsequent measurement
from aerosol sample
of dry particle size.
 Aerosol research
There are a lot of industrial applications the  Emissions from direct heating boilers
particle size distribution of hot gases is interesting  Biomass burning
for.
The specification of many particle size measuring
instruments does not allow analysing gas with high
temperatures which required a cooling of the To Filling neck for From
sampling
sample. When cooling takes place condensation particle active carbon
sizer line
processes will occur and volatile particles can be
formed by nucleation. Both processes produce
particles or change particle size of solid
compounds in the gas by heterogeneous
condensation on particle surfaces. So results of Adsorption tube
Heater with
particle measurements include a systematic error. (volatile compounds temperature
diffuse to surface of controller
For avoiding such measuring errors active carbon) (up to 400°C)

thermodenuders are used. By regulating the Principle of Thermodenuder TDD 590


sample temperature such devices enable reducing
moisture as well as volatile particles in the sample
aerosol flow. Thus condensation processes during
cooling down are widely prevented.
Specifications

Details Technical Data


Active carbon surrounding the flowpath of the Flow rate 0.1 … 3 l/min
particle-gas mixture removes the volatile
compounds using the physical effect of adsorption. Temperature range 30 … 400°C
According to the adjusted temperature, specific Temperature WIKA MCS-100
components were extracted from the aerosol gas. controller
The size distribution of particles in the aerosol is
not influenced. Volume of activated 6l
The hot aerosol or exhaust gas containing particles carbon
is directed into the thermodenuder where it is Materials glass, stainless steel
heated to a pre-set temperature. A built-in
Sampling connector 6 mm Swagelok
microprocessor controller displaying actual and
set values is used to achieve precise adjustment of Power supply 230 V AC / 50-60 Hz
the temperature. The temperature can be varied
Dimensions 1200 x 250 x 200 mm
easily with three or four touch keys at the
(L x H x D)
temperature controller.
Weight 11.5 kg
Volatile particles completely evaporate in the
heating section. Hereby volatile particles are
defined as particles evaporating during the
residence time inside the thermodenuder at the
pre-set temperature. Adsorbents in the aerosol
diffuse to the surface of the active carbon due to
the different concentration of the adsorbents
between the volume with carbon and the gas.

As a result only solid particles are left in the


aerosol. The TDD 590 thermodenuder operates
with minimal loss as the particles never come in
direct contact with the active carbon.

Active carbon Gas permeable screen


For more information please
QMS certified to
visit our website at
DIN EN ISO 9001.
www.topas-gmbh.de

Aerosol Specifications are subject to


change without notice.
Diffusion of volatile
compounds 12 100 11908 TMS  Copyright 2019 Topas GmbH.

Topas GmbH Phone +49 (351) 21 66 43 - 0


Technologie-orientierte Fax +49 (351) 21 66 43 55
Partikel-, Analysen- und Sensortechnik E-mail [email protected]
Gasanstaltstraße 47 · DE - 01237 Dresden, GERMANY Internet www.topas-gmbh.de