I. InBody270 Installation
A. Product Components 4
B. Operating Environment 5
C. Installation Instructions 5
D. Initial Setup 8
E. Connecting Printer, Thermal Printer, Stadiometer, 12
Blood Pressure Monitor, Barcode Reader and SD400
F. Maintenance 15
V. Others
A. Exterior and Functions 24
B. Safety Information 27
C. Classification 29
D. Specifications 29
E. EMC declaration 31
❶ InBody270
❷ Adapter (DC 12V, 3.4A) 1 EA
Input Buttons
❸ Power cord 1 EA
❹ User’s Manual 1 EA
Upper part
Hand electrodes
Foot electrodes
Hinge
Footplate
❶
❶ ❹
❸
C. Installation Instructions
1. Open the packing box of the InBody270 and remove the packing pads. Then take the InBody270 out of
the box.
Polyethylene foam cover
Packing pads
Caution
참조 주의
• If you have any problems installing your InBody270, please contact InBody for assistance.
• Do not transport the equipment by holding the screen portion or the joints of the hand electrodes.
• Keep the packing materials provided for repacking the equipment in the future. Other wastes should be
disposed of according to relevant laws and regulations.
2. While holding the upper part lock button, raise the upper part of the InBody270 and remove the polyethylene
foam cover.
Hinge
Caution
참조 주의
• Using the InBody270 on carpet may cause static electricity, which could damage the equipment. If installing
the InBody270 on carpet is unavoidable, please use an antistatic mat.
• Install the InBody270 on a leveled, non-vibrating surface. Installing the equipment on an uneven surface
may cause the examinee to fall down. Test results may also be inaccurate.
• Never clean the hand and foot electrodes with liquid spray or detergent directly. The equipment may corrode
and/or malfunction if the liquid or detergent leaks inside. Use the InBody Tissue when cleaning the InBody270.
Level Indicator
Caution
참조 주의
• Avoid injuring your hands when rotating the leveling screws under the footplate.
4. Connect the adapter (❷) to the power input port, which is located on the rear panel (❶). Connect the adapter (❷)
to the power cord (❸). Plug the power cord (❸) into a grounded 3-socket outlet.
* The InBody270 can be used in connection with other equipment such as a stadiometer, a blood pressure monitor, data management
software called Lookin’Body, or a barcode reader. For more information, please refer to 'I. InBody270 Installation E.Connecting Printer,
Thermal Printer, Stadiometer, Blood Pressure Monitor, Barcode Reader and SD400' in User's Manual.
❶
Power input port
Rear panel InBody
Grounded outlet
❷ ❸
3-socket outlet
Warning
참조 주의
• Do not place the InBody270 in a location making it difficult to disconnect the power cord.
• Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock.
• Always use an outlet connected to the power (AC 100 - 240 V). Using other power rated outlets may result in
fire or malfunction.
• When using a power surge protector, make sure that the outlet or the extension cable has adequate power capacity.
• Do not disassemble or modify the equipment including internal parts without written consent from the
manufacturer. This may cause electric shock or injury, product malfunction, inaccurate results, and will
void the manufacturer’s warranty.
Caution
참조 주의
• If the InBody270 is not plugged into a grounded outlet, it may cause damage through electric surges or
product malfunction. This may affect the test results.
• The test results may be inaccurate if the InBody270 is under electrical interference. Do not install the
InBody270 near products that generate electrical interference such as fluorescent lights, large AC motor
equipment(treadmill, vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal
therapy equipments, or heating appliances. Do not share the power source of the InBody270 with other
electrical devices. This may affect the test results.
• When connecting the InBody270 with other test equipment, turn on the other equipment first. When
turning off other equipment, turn off the InBody270 first. This is necessary to minimize electrical surges
on the InBody270.
• Always use the specified adapter provided by InBody as it is a part of the InBody270. Using other adapters
may result in malfunction of the InBody270.
• Operation of the InBody270 2,000m above sea level may affect the weight measurement.
• Adapter must be arranged so that easily cut off the power when a problem occurs in the InBody270.
InBody
Power switch
2. Press the [Administrator Menu] button on the screen, which appears when no one is on the footplate.
Administrator Menu
3. Input the password (default password: 0000) to access the Administrator Menu.
4. The Administrator Menu will give you access to ‘Setup’ and ‘Troubleshooting’.
Setup Troubleshooting
8
01. Date/Time/Units/Country/Language/Password/Volume
: Change or modify the InBody’s basic settings.
03. N/A
04. N/A
17. Bluetooth
: Connect the InBody270 to data management software Lookin’Body via Bluetooth.
21. N/A
25. Auto-Lock
: Set the password and wait time for auto-lock on the InBody270.
10
11
2) Plug the USB cable provided with the printer into the USB HOST port on the rear panel of the InBody270
and plug the other end of the USB cable into the printer.
2. Thermal Printer
Connect thermal printer to the InBody270 to print Thermal Results Sheet.
* For best result, the InBody270 should be turned off when connecting to thermal printer.
2) Connect the serial cable supplied with your thermal printer to the 9-pin serial port on the rear panel of the
InBody270. Then, connect the other end of the serial cable to the serial port on to thermal printer.
12
2) Plug the serial cable provided with the stadiometer to the 9-pin serial port on the rear panel of the
InBody270.
1) First turn off the InBody270 and then the blood pressure monitor.
* You may experience connection issues in connecting the blood pressure monitor to the InBody270 if the InBody is turned on.
2) Plug the serial cable provided with the blood pressure monitor to the 9-pin serial port on the rear panel of
the InBody270.
13
2) Plug the USB cable of the barcode reader into the USB HOST port on the rear panel of the InBody270.
3) Turn on the InBody270. If the barcode reader is connected to the InBody270, the barcode reader
icon( ) will appear on the top left corner of the screen when no one is on the footplate.
6. SD400
Connect the SD400 to the InBody270 in order to connect the stadiometer and the blood pressure monitor
simultaneously.
2) Connect the serial cable supplied with the SD400 to the 9-pin serial port on the rear panel of the
InBody270.
Then, connect the other end of the serial cable to the serial port of the SD400.
3) Connect your devices (such as stadiometer and blood pressure monitor) to the ports on the SD400 and turn
on the devices.
4) Turn on the InBody270. Go to Setup in Administrator Menu. Select ‘SD400’ under ‘26. Serial Connect’
and select the devices you have connected to the ports on the SD400. When your devices are connected,
the icons ( ) of your devices will appear in the top left of the test standby screen.
14
Caution
참조 주의
• Do not bend the handles of the hand electrodes or rotate them in the reverse direction beyond its limitation.
• Do not extend the handles of the hand electrodes beyond its limitation.
• Do not place any objects on the footplate.
• Do not apply excessive force on the equipment.
• Turn off the equipment if you are not using it for a day or longer.
• Do not allow any liquid substances to contact the equipment directly. Keep food and drinks away from the
equipment. Substances getting inside the equipment can cause critical damage to the electronic components.
• Use a lint-free cloth to gently wipe the external surface of the equipment about once every week. Be careful not
to scratch the LCD screen.
• InBody270 does not need regular maintenance. If some problems occur while operating the device, get in touch
with the store where you purchased it or A/S manager.
We do not take the responsibility about problems caused by any arbitrary repairs.
15
Warning
참조 주의
• Individuals with medical implant devices such as pacemakers, or essential support devices such as patient
monitoring systems, must not use this equipment. Safe, low-level currents will flow through the body during
the test, which may cause malfunctioning of the device or endanger lives.
• Bioelectrical Impedance Analysis (BIA) uses safe low level currents, which are not harmful to the body. However,
Caution
참조 주의
• Stand upright for about 5 minutes before testing. Taking the test immediately after lying in bed or sitting for
a long period of time might result in a slight change in the test results. This is because body water tends to
move to the lower body as soon as the person stands or gets up.
• Do not eat before testing. In cases where the examinee has already eaten, the test should be put off for at least
two hours after the meal. This is because food mass is included in the examinee’s weight and thus, may result
in measurement errors.
• Use the bathroom before testing. Waste is not included in the body’s compositional elements, but the volume
of urine and excrement is included in the weight measurement affecting accuracy of the test results.
• Do not exercise before testing. Strenuous exercise or sharp movements can cause temporary changes in body
composition. Even light exercise can change your body composition temporarily.
• Take the test in the morning, if possible. Body water tends to gravitate towards the lower body throughout
the day, affecting accuracy of the test results.
• Thoroughly wipe the palms and soles with the InBody Tissue before testing. Testing may be difficult if the
examinee’s palms and soles are too dry or if the examinee has too many calluses.
• Avoid contact with the examinee during testing. Contact may lead to interference affecting test results.
16
17
18
Place the heels on the rear sole Step on the footplate barefoot.
electrodes.
19
B. Repacking Instructions
Once the InBody270 is installed, avoid transporting the equipment. If it must be transported, repack it in
the following sequence.
Polyethylene
foam cover
Lock
button
4. Put the InBody into the packing box. Place the packing pad over the equipment and tape up the
packing box.
Packing pads
Polyethylene foam cover
Caution
참조 주의
• Always use the protective packing materials provided by InBody when repacking.
20
Question Answer
• My InBody270 does • Insert the power plug completely into a grounded 3-socket outlet.
not turn on.
• When using a power surge protector, the equipment may not power on if
the power switch on the power surge protector is turned off. Check the
power surge protector which the power plug is connected to.
Power surge protector
power switch
• The problem may occur if the power cord is not completely plugged into
the adapter. Insert the power cord completely into the adapter.
• The problem may occur if you are using the power adapter that is not provide
by InBody. Always connect a power adapter (DC 12V, 3.4 A) provided
by InBody.
21
• My touchscreen • Calibrate
the touchscreen under Settings of the Administrator Menu
is inaccurate or not ‘23. Touchscreen Alignment’.
responsive. * Press firmly to optimize touchscreen response.
• If you cannot enter the Administrator Menu due to touchscreen problems,
please restart the InBody. The InBody stores the last touchscreen alignment
and will automatically recall the previous touchscreen settings. The InBody
can also recognize if its touchscreen alignment is off screen and will
automatically take the user to the calibration screen after restarting.
•Iwould like to connect • Please refer to ‘E. Connecting Printer, Thermal Printer, Stadiometer, Blood
other equipment to the Pressure Monitor, Barcode Reader and SD400’ in section ‘I. InBody270
InBody270. Installation’ in this User’s Manual.
22
Question Answer
• Must socks or stockings • Bare skin contact is essential in the analysis using the BIA method. Socks
be removed for the or stockings may cause a varying degree of distortion in the results. Socks
InBody Test? or stockings must be removed to obtain accurate data.
• Is it okay to wear accessories • Theideal condition for the analysis is simply standing with no clothes
(jewelry, watch, rings, etc) or and wearing no accessories. However, this may not always be possible.
metal objects while taking Therefore, we recommend that the examinee remove as many clothing items
the InBody Test? and accessories that may affect the weight as possible.
• Who cannot take the • Individualswith medical implant devices such as pacemakers, or essential
InBody Test or will have support devices such as patient monitoring systems, must not use this
difficulties taking the equipment. The currents will flow through the body during the
InBody Test? test, which may cause malfunctioning of the device or endanger lives.
• Can a person with metal • Theideal test methodology is where the examinee does not wear anything
implants in the body take metallic. Individuals with metallic implants may have skewed test results
the InBody Test? due to the conductivity of the metal affecting the results.
• Asthe weight of clothes and other wear affects the results of the body
composition analysis, it is strongly recommended to take off any heavy
clothing or metallic wear. Except for the weight, jewelry does not effect
the body composition analysis, as the contact point with the InBody270
are hands and feet.
• Isthe electric current • The physiological electric impedance method uses safe low level currents
harmful to the body? that is not harmful to the body. The safety of the InBody has been tested
and proven. The InBody products have been approved for medical use by
the CE and all over the world. Many medical institutions around the world
are actively using the InBody.
• How often should I take • Individualswho are undergoing any programs that may affect their body
the InBody Test? composition are strongly recommended to have the InBody Test done every two
to four weeks.
• Consistent testing will allow individuals to track and monitor their progress
over time.
• What are the precautionary • Please refer to ‘A. Precautionary Steps’ in section ‘II. InBody Test’ in this
steps to ensure accuracy of User’s Manual.
the InBody Test?
23
1. Upper Part
❶ Hand electrode pipe: The signal cables connecting to the electrode are in this pipe.
❷ Hand electrode: Examinee holds the hand electrode so that the 4 fingers wrap the surface of the
bottom hand electrode while the thumb is placed on the oval electrode.
❸ Body: Connects the upper part of the equipment to the lower part.
❹ LCD screen: Shows each stage of the test, instructions, test results, etc. You can touch the screen to
input the data required for the test, configure settings, or view test results.
❺ Number keypad: Used for inputting age, height, and other number-based data.
❻ Setup button : Used for entering ‘Settings’ under the Administrator Menu when no one is on the footplate.
❼ Function button: Used for entering ‘Troubleshooting’ under the Administrator Menu when no one is on
the footplate.
❽ Print button: Used for printing the test results.
❾ Decimal point button: Used for inputting the decimal point in ID, height, age, or weight.
⓫ Enter button: Used to finish inputting data or to save changes in Administrator Menu.
❶ ❺
❻
❼ ❿
❽
❷ ⓫
❾
⓬
24
❹
❺
❶
❷
25
❸ LAN port (10T Base): Used for connecting the InBody270 to Lookin’Body installed on a computer.
* The InBody270 can be connected to Lookin’Body installed on a computer using one of the ports ❷, ❸, or ❹.
❹ USB SLAVE port: Used for connecting the InBody270 to Lookin’Body installed on a computer.
* The InBody270 can be connected to Lookin’Body installed on a computer using one of the ports ❷, ❸, or ❹.
❺ USB HOST port: Used for connecting to a printer, a barcode reader, or a USB Thumb Drive.
❻ Power Inlet: Used for connecting the power adapter.
* Always use the adapter supplied by InBody Co., Ltd..
❼ Power switch: Used to turn the device on/off.
CAUTION NOTE DANGER
Warning
참조 주의
• Do not touch signal input, signal output or other connectors, and the patient simultaneously.
• External equipment intended for connection to signal input, signal output or other connectors, shall comply
with relevant IEC Standard(e.g., IEC60950 for IT equipment and IEC60601-1 series for medical electrical
equipment). In addition, all such combination-system-shall comply with the standard IEC60601-1 and/or
IEC60601-1-1 harmonized national standard or the combination. If in doubt, contact qualified technician
or your local representative.
❻ ❷ ❺ ❸ ❹
❼ ❶
26
USB port
Safety Symbols
Warning, Caution
4
BF Type Equipment
4
Adapter
4
4 Power On
44
4
Power Off
4
Etc. Symbols
0120 SN
European Conformity SN Serial number
0120
0120 SN
Manufacturer Direct current
EC REP EC REP
Authorized
SN representative in the
EC 0120
REP
0120 SN Operating instructions
EUROPEAN COMMUNTY
EC REP
EC REP
27
WARNING
Electric shock hazard – do not dismantle.
Dismantling will void the warranty.
AVERTISSEMENT
Risque de choc électrique - ne pas démonter.
Le démontage annulera la garantie.
DANGER
Do not use this equipment with electrical medical device such as a pacemaker.
Ne pas utiliser cet équipement avec des appareils médicaux électriques comme un stimulateur cardiaque.
CAUTION
Do not spray any liquid substance directly onto the device.
Ne pulverisez aucune substances liquids directement sur l’appareil.
CAUTION
No excessive force on shoulder joint
Ne pas appliquer de force excessive sur les bars articulés.
28
D. Specifications
Bioelectrical 10 Impedance Measurements by Using 2 Different Frequencies
Bioelectrical
Impedance Analysis (BIA) (20kHz, 100kHz) at Each of 5 Segments (Right Arm, Left Arm,
Impedance (Z)
Measurement Items Trunk, Right Leg, and Left Leg)
Measurement Method Direct Segmental Multi-frequency Bioelectrical Impedance Analysis Method (DSM-BIA )
Simultaneous Multi-frequency Impedance Measurement (SMF-BIA)
Body Composition
No Empirical Estimation
Calculation Method
29
Logo Name, Address, and Contact Information can be shown on the InBody Results Sheet.
Types of Result Sheets InBody Test Results Sheet, InBody Test Results Sheet for Children,
InBody Test Thermal Results Sheet
Voice Guidance Provides audible indication for test in progress, test complete, and successfully saved settings changes.
Database Test results can be saved if the member ID is utilized. The InBody can save up to 100,000 results.
USB Thumb Drive Copy, backup, or restore the InBody test data (data can be viewed on Excel or Lookin’Body
data management software)
Barcode Reader The member ID will be automatically inputted when the barcode ID is scanned.
Backup data Backup data saved in the InBody by using a USB Thumb Drive, Restore results on the
InBody from a backup file.
Model BPM040S12F07
External Interface RS-232C 1EA, USB HOST 2EA, USB SLAVE 1EA, LAN (10T) 1EA,
Bluetooth 1EA, Wi-Fi 1EA
30
Electromagnetic emissions
Emissions test Compliance Electromagnetic environment
The InBody270 uses RF energy only for its internal function.
RF emissions CISPR 11 Group 1 Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class A
Harmonic emissions The InBody270 is suitable for use in all establishments,
Class A
IEC 61000-3-2 including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
Voltage fluctuations/ flicker
buildings used for domestic purposes.
emissions Complies
IEC 61000-3-3
31
NOTE UT is the a.c. mains voltage prior to application of the test level.
IEC 60601
Immunity test Compliance level Electromagnetic environment
test level
Portable and mobile RF communications
equipment should be used no closer to any part
of the InBody270, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
ᵃ Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the InBody270 is used exceeds the
applicable RF compliance level above, the InBody270 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the InBody270.
ᵇ Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
32