4-0002428K - PINPOINT - System - Operators - Stryker Laparos
4-0002428K - PINPOINT - System - Operators - Stryker Laparos
Document: 4-0002428
Edition: English – INTL
Revision: K, 2021-01
Printed in Canada
Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied Permission is hereby granted, free of charge, to any person obtaining a copy of the fonts
for the following trademarks or service marks: NOVADAQ, PINPOINT, SPY-PHI, STRYKER. All accompanying this license ("Fonts") and associated documentation files (the "Font Software"), to
other trademarks are trademarks of their respective owners or holders. The absence of a product, reproduce and distribute the Font Software, including without limitation the rights to use, copy,
feature, or service name, or logo from this list does not constitute a waiver of Stryker’s trademark merge, publish, distribute, and/or sell copies of the Font Software, and to permit persons to whom
or other intellectual property rights concerning that name or logo. the Font Software is furnished to do so, subject to the following conditions:
The above copyright and trademark notices and this permission notice shall be included in all
Copyright © 2006 by Tavmjong Bah. All Rights Reserved. copies of one or more of the Font Software typefaces.
The Font Software may be modified, altered, or added to, and in particular the designs of glyphs
Permission is hereby granted, free of charge, to any person obtaining a copy of the fonts or characters in the Fonts may be modified and additional glyphs or characters may be added to
accompanying this license ("Fonts") and associated documentation files (the "Font Software"), to the Fonts, only if the fonts are renamed to names not containing either the words "Bitstream" or
reproduce and distribute the modifications to the Bitstream Vera Font Software, including without the word "Vera".
limitation the rights to use, copy, merge, publish, distribute, and/or sell copies of the Font This License becomes null and void to the extent applicable to Fonts or Font Software that has
Software, and to permit persons to whom the Font Software is furnished to do so, subject to the been modified and is distributed under the "Bitstream Vera" names.
following conditions: The Font Software may be sold as part of a larger software package but no copy of one or more
The above copyright and trademark notices and this permission notice shall be included in all of the Font Software typefaces may be sold by itself.
copies of one or more of the Font Software typefaces. THE FONT SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND,
The Font Software may be modified, altered, or added to, and in particular the designs of glyphs EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY WARRANTIES OF
or characters in the Fonts may be modified and additional glyphs or characters may be added to MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT
the Fonts, only if the fonts are renamed to names not containing either the words "Tavmjong Bah" OF COPYRIGHT, PATENT, TRADEMARK, OR OTHER RIGHT. IN NO EVENT SHALL
or the word "Arev". BITSTREAM OR THE GNOME FOUNDATION BE LIABLE FOR ANY CLAIM, DAMAGES OR
This License becomes null and void to the extent applicable to Fonts or Font Software that has OTHER LIABILITY, INCLUDING ANY GENERAL, SPECIAL, INDIRECT, INCIDENTAL, OR
been modified and is distributed under the "Tavmjong Bah Arev" names. CONSEQUENTIAL DAMAGES, WHETHER IN AN ACTION OF CONTRACT, TORT OR
The Font Software may be sold as part of a larger software package but no copy of one or more OTHERWISE, ARISING FROM, OUT OF THE USE OR INABILITY TO USE THE FONT
of the Font Software typefaces may be sold by itself. SOFTWARE OR FROM OTHER DEALINGS IN THE FONT SOFTWARE.
THE FONT SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, Except as contained in this notice, the names of Gnome, the Gnome Foundation, and Bitstream
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY WARRANTIES OF Inc., shall not be used in advertising or otherwise to promote the sale, use or other dealings in
MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT this Font Software without prior written authorization from the Gnome Foundation or Bitstream
OF COPYRIGHT, PATENT, TRADEMARK, OR OTHER RIGHT. IN NO EVENT SHALL Inc., respectively. For further information, contact: [email protected]
TAVMJONG BAH BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY,
INCLUDING ANY GENERAL, SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES, WHETHER IN AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING
FROM, OUT OF THE USE OR INABILITY TO USE THE FONT SOFTWARE OR FROM OTHER
DEALINGS IN THE FONT SOFTWARE.
Except as contained in this notice, the name of Tavmjong Bah shall not be used in advertising or
otherwise to promote the sale, use or other dealings in this Font Software without prior written
authorization from Tavmjong Bah. For further information, contact: [email protected].
Copyright © 2003 by Bitstream, Inc. All Rights Reserved. Bitstream Vera is a trademark of
Bitstream, Inc.
4-0002428 Rev. K
Chapter 1 Health and Safety Compliance............................................................................ 1
Important Information ................................................................................................................................................. 1
Intended Use ............................................................................................................................................................. 2
Overview of the PINPOINT System ........................................................................................................................... 2
Components .......................................................................................................................................................... 2
User Qualifications ..................................................................................................................................................... 4
General System Safety .............................................................................................................................................. 4
Electrical Safety – General .................................................................................................................................... 4
Electrical Safety – Power....................................................................................................................................... 5
Light Safety- Illumination ....................................................................................................................................... 5
Warning Labels on the PINPOINT System ............................................................................................................ 7
Electromagnetic Compatibility (EMC) and Environmental Safety .......................................................................... 7
Advice When Used with High Frequency Surgical Equipment .............................................................................. 8
Damaged or Malfunctioning Equipment ................................................................................................................ 8
Cleaning and Sterilizing ......................................................................................................................................... 9
Repair and Modification ....................................................................................................................................... 10
Disposal of PINPOINT System Components ........................................................................................................... 10
Symbols Used on the PINPOINT System ................................................................................................................ 11
Chapter 2 Unpacking and Setting Up the PINPOINT System .......................................... 17
System Parts ........................................................................................................................................................... 17
Setting Up the PINPOINT Imaging System.............................................................................................................. 19
Selecting an Appropriate System Installation Location ........................................................................................ 19
Connecting the Video Processor / Illuminator (VPI) ............................................................................................ 20
Connecting to a Medical-grade Video Monitor..................................................................................................... 21
Connecting to a Medical Recorder ...................................................................................................................... 22
Connecting to Ancillary Video Devices (Optional) ............................................................................................... 22
Selecting the Appropriate Video Output Format .................................................................................................. 22
Connecting the PINPOINT Camera to the VPI .................................................................................................... 24
Connecting the Light Guide Cable to the Laparoscope ....................................................................................... 25
Connecting and Disconnecting the Laparoscope to the PINPOINT Camera ....................................................... 27
Storing and Transporting Laparoscopes ............................................................................................................. 28
Chapter 3 Handling, Preparation, and Administration of ICG ......................................... 29
About Indocyanine Green (ICG) Imaging Agent ...................................................................................................... 29
ICG Safety ............................................................................................................................................................... 29
Clinical Pharmacology ......................................................................................................................................... 29
ICG Contraindications ......................................................................................................................................... 29
ICG Warnings ...................................................................................................................................................... 30
ICG Adverse Reactions ....................................................................................................................................... 31
General ICG Preparation Instructions ...................................................................................................................... 31
ICG Administration via Central or Peripheral Venous Line ...................................................................................... 32
Supplies Required for each Imaging Sequence .................................................................................................. 32
Dosage ................................................................................................................................................................ 32
Preparation for ICG Administration ...................................................................................................................... 32
ii 4-0002428 Rev. K
Limitations on Processing.................................................................................................................................... 83
Reprocessing Laparoscopes and Light Guide Cables ............................................................................................. 84
Using Polishing Paste for Laparoscopes ............................................................................................................. 84
Step 1: Preparing Laparoscopes and Light Guide Cables ................................................................................... 85
Step 2: Cleaning Laparoscopes and Light Guide Cables .................................................................................... 87
Step 3: Inspecting Laparoscopes and Light Guide Cables .................................................................................. 89
Step 4: Sterilizing Laparoscopes and Light Guide Cables ................................................................................... 91
Sterilization .......................................................................................................................................................... 95
Expected Service Life of Laparoscopes and Light Guide Cables ............................................................................ 97
References .............................................................................................................................................................. 98
Reprocessing the PINPOINT Camera (PC9002) ..................................................................................................... 99
Step 1: Preparing the PINPOINT Camera ........................................................................................................... 99
Step 2: Cleaning the PINPOINT Camera .......................................................................................................... 101
Step 3: Inspecting the PINPOINT Camera ........................................................................................................ 103
Step 4: Sterilizing the PINPOINT Camera ......................................................................................................... 104
Appendix A: Troubleshooting ............................................................................................... 109
Clearing PINPOINT Camera, Illumination, and Recorder Errors ........................................................................... 109
PINPOINT Camera Communication Error ......................................................................................................... 109
Illumination Error ............................................................................................................................................... 110
Video Recording and Screenshot Errors ........................................................................................................... 111
Setting up the Sony 1000MD Recorder for Control from the VPI ...................................................................... 112
Setting up the Sony 3300MT Recorder for Control from the VPI ....................................................................... 115
General Troubleshooting ....................................................................................................................................... 119
Fuse Replacement Procedure ............................................................................................................................... 122
Appendix B: VPI Menu Options and Settings....................................................................... 123
Image Quality Menu Settings ................................................................................................................................. 123
Sharpening (0-10) ............................................................................................................................................. 123
Brightness (0-255) ............................................................................................................................................. 123
Red Saturation (0-1024) .................................................................................................................................... 123
Blue Saturation (0-1024) ................................................................................................................................... 123
Peak/Mean ........................................................................................................................................................ 123
Default Profile Menu .............................................................................................................................................. 124
Service Menu ......................................................................................................................................................... 125
Display Options Menu ............................................................................................................................................ 125
SPY Image ........................................................................................................................................................ 125
Overlay Image ................................................................................................................................................... 125
SPY Color Segmented Fluorescence (CSF) Image .......................................................................................... 125
Appendix C: Specifications and Standards ......................................................................... 127
Appendix D: Software License Agreement .......................................................................... 133
iv 4-0002428 Rev. K
WARNING: Read before use.
This Operator’s Manual contains essential information about using the PINPOINT
Endoscopic Fluorescence Imaging System (PINPOINT System) safely and
effectively. Before use, thoroughly review this manual and use the system as
instructed.
Keep this manual in a safe, accessible location. Questions or comments about any
information in this manual should be sent to Stryker Customer Support.
The words WARNING, CAUTION, and Note carry special meaning and the
associated clauses should be carefully reviewed:
2 4-0002428 Rev. K
Figure 1: Overview of the PINPOINT Endoscopic Fluorescence Imaging System
▪ Medical cart
▪ Laparoscopic insufflator
▪ Medical-grade HD video recorder
▪ Printer
▪ Sterilization trays for PINPOINT camera, light guide cable, and laparoscope
WARNING: This manual does not explain or discuss clinical surgical procedures.
Therefore, the healthcare professional using PINPOINT must be a licensed
physician or medical personnel under the supervision of a licensed
physician and must have received sufficient training in clinical procedures
and the use of the PINPOINT System.
WARNING: To avoid the risk of electric shock, this equipment must only be connected to
supply mains with protective earth.
CAUTION: Do not use or store liquids around the VPI. If liquid enters the VPI, immediately turn
the system off and unplug it from the power outlet.
Do not insert objects into the ventilation holes of the VPI enclosure.
Do not connect or disconnect the camera cable while the VPI is powered on.
4 4-0002428 Rev. K
WARNING: Grounding reliability can only be achieved when equipment is connected to
an equivalent receptacle marked 'hospital only' or 'hospital grade’.
Do not use the imaging system if the power cord or plug is damaged or
modified in any way.
CAUTION: Unplug power cords by grasping the plug. Do not unplug power cords by pulling on
the cable.
WARNING: Avoid looking at light emitted directly from the laparoscope or the light guide
cable tip. Table 16 in Appendix C contains specifications for NIR radiation
emitted in fluorescence mode.
Always use standby mode by turning illumination off (see Turning Off
System Illumination, in Chapter 5) when connecting or disconnecting the
light guide cable from the laparoscope, or prior to removing the laparoscope
from the patient.
There is high-energy light at the distal end of the laparoscope. This can
cause the temperature of the body tissue to rise to 41°C. Avoid direct contact
of the distal end with body tissue or flammable materials as it can cause
burns.
6 4-0002428 Rev. K
The PINPOINT camera and VPI* are labeled with the following laser safety warning
labels in accordance with applicable standards:
Laser Aperture
indicator, see
Figure 2
* The Video Processor/Illuminator used with the PINPOINT camera is a component that is shared with
the SPY Portable Handheld Imaging System (SPY-PHI System).
The user is expected to follow all precautions and warnings outlined in this manual,
and adhere to the electromagnetic compatibility (EMC) limits set out in the
electromagnetic compatibility guidance document. Failure to do so may
compromise the system’s ability to maintain basic safety under abnormal EMC
conditions.
The imaging system has been certified for compliance with international standards
for EMC and is suitable for use in professional healthcare environments. The
imaging system generates radio frequency energy and should be installed and
used in accordance with these instructions to minimize the possibility of
interference with other electro-medical equipment. However, there is no guarantee
that interference will not occur in any particular installation.
If in turning the imaging system off and on, the PINPOINT System is shown to
cause harmful interference to other electro-medical equipment, the user is
encouraged to try to correct the interference by:
WARNING: Use of accessories, transducers and cables other than those specified or
provided by Stryker could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in
improper operation.
Note: The emission characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 Class A). It is not intended for use in a residential
environment (for which CISPR 11 Class B is required).
If the imaging system becomes unresponsive and does not resume normal function
after turning off and back on, stop using the imaging system, and contact Stryker
Customer Support.
WARNING: PINPOINT rigid laparoscopes are not designed to provide insulation against
high-frequency (HF) electrical currents. To prevent burns and unintended
thermal injury of surrounding tissue, activate electrodes only when visible
through the laparoscope and avoid contact between active electrodes and
the laparoscope.
WARNING: Do not use the PINPOINT camera if any part of the system is damaged or
does not function properly. Failure to follow this warning may lead to injury.
In the event of a drop or impact, inspect the PINPOINT camera and do not
use it:
If any of the above events occur, please contact Stryker Customer Support.
8 4-0002428 Rev. K
WARNING: Follow the instructions in Chapter 7 to clean and sterilize the PINPOINT
components and accessories.
Keep the light guide cable connectors clean at all times. Contaminants on the
light guide cable connectors may cause overheating.
WARNING: For storage, transport and processing, ensure that the laparoscope is not
subjected to mechanical strain to prevent damage to the sensitive lens
system.
Ensure that the processing, material and personnel are suitable for achieving
the results necessary.
➢ Observe all local regulations for operator safety, protection and training
in performance of manual cleaning and drying processes.
➢ Clean and sterilize the laparoscope prior to initial use as well as each
subsequent use of the laparoscope.
➢ Observe appropriate protective measures to prevent contaminating the
environment.
➢ Clean and sterilize the camera, light guide cable, and laparoscope
thoroughly prior to returning for repair.
➢ Ideally, return the part in its original packaging. If this is not possible,
package the part to secure it for transport.
➢ Stryker is not liable for damage resulting from improper shipping.
10 4-0002428 Rev. K
The following section describes symbols and indicators that appear on the
PINPOINT camera and VPI.
Button 1 Camera
Button 2 Camera
Front panel
Power
Front panel
Camera socket
Rear panel
Refer to instruction manual/booklet
or
Accompanying
General Warning
Documentation
12 4-0002428 Rev. K
Laser Warning Front panel
Rear panel
Manufacturer
Rear panel
Date of manufacture
Accessory
Reference or Model name/number
Packaging
Accessory
Serial number
Packaging
Accessory
Lot number/Date code
Packaging
Accessory
Made in Germany
Packaging
Accessory
Made in Canada
Packaging
Accessory
Made in USA
Packaging
Accessory
Made in Taiwan
Packaging
Accessory
Do not reuse - single use only
Packaging
Accessory
Expiry (date/time)
Packaging
Rear panel,
CE Marking declaring EU Directive compliance Accessory
Packaging
Accessory
Authorized European (EC) Representative
Packaging
Accessory
Medical Device
Packaging
14 4-0002428 Rev. K
White balance failed See Chapter 6
Tools menu – Flip image (unavailable with PINPOINT System) See Chapter 6
Amber Off
Above “Power”
Power indicator
button
Green On
Above “Illumination”
Illumination indicator White Illumination on
button
Below “Illumination”
Green Down
button
(part of menu
Green Up
feature)
Below “White
Green Select
Balance” button
1
A laser-on indicator light is also located on the PINPOINT camera.
16 4-0002428 Rev. K
WARNING: If any items are missing or damaged, do not use the PINPOINT System and
contact a Stryker service representative.
Use Table 6 to identify and inspect all PINPOINT System parts before proceeding
with setting up the system.
Model PC9001*
PINPOINT camera
Model PC9002
Fluorescence Imaging
Laparoscopes (if available)
10 mm diameter
Fluorescence Imaging
Laparoscopes (if available)
5 mm diameter
* The Video Processor/Illuminator used with the PINPOINT camera is a component that is shared with
the SPY-PHI System.
Model PC9004
Power Cord
HD-SDI Cable
DVI Cable
PINPOINT Endoscopic
Fluorescence Imaging System
Operator's Manual
18 4-0002428 Rev. K
WARNING: The PINPOINT System has been certified for compliance with international
standards for electromagnetic compatibility (EMC).
WARNING: Read and follow the information in “Electrical Safety – Power” in Chapter 1
and only connect the PINPOINT VPI to a “Hospital Only” or “Hospital Grade”
power outlet.
CAUTION: Ensure that the selected location provides a minimum 5 cm (2 inches) gap around
the VPI cooling vents to allow for required ventilation.
Figure 3: PINPOINT System wiring diagram showing connection to video monitors and
recorder
20 4-0002428 Rev. K
Figure 4: Video Processor / Illuminator rear panel showing connectors and ports
CAUTION: The communication ports on the rear panel are intended for use by Stryker service
representatives only.
Note: Normal use of the system does not require the user to connect to the equipotential
post. The potential equalization conductor is only to be used by qualified personnel.
A medical-grade color video monitor can be connected via the SDI and DVI
connectors on the rear panel of the VPI (see Figure 4).
See Table 11 in Appendix C for specifications of the video output signals and
connectors to ensure compatibility of the selected video monitor. Consult the
operator instructions provided by the manufacturer of the video monitor for details
on operation and adjustment of the video monitor.
Note: For the onscreen recorder controls and status to be visible to the operator during
normal operation of the PINPOINT System, the VPI SDI 1 video output must be
connected to the SDI-IN connector on the recorder. The SDI-OUT connector on the
recorder should be connected to the medical-grade video monitor. For more
information, see Figure 3.
For more information about setting up the recorder, please refer to Chapter 6.
The VPI may be interconnected to other medical-grade video devices such as:
Use the video-format switch on the rear panel of the VPI to select one of the
following video output formats:
Note: Ensure that the video-format switch on the VPI is in the correct position for the
recorder that is being used.
22 4-0002428 Rev. K
The correct output formats are as follows:
WARNING: If the video output is routed through a video recorder, operators must always
be aware of whether they are viewing a live or recorded image. Wave your
hand in front of the camera to determine whether or not you are viewing a
live image.
1. Remove the cap from the end of the camera cable and orient the camera cable
connector with the raised arrow facing up (see Figure 5).
2. Insert the camera cable into the camera port on the VPI (see Figure 6).
24 4-0002428 Rev. K
WARNING: The following procedures must be performed using proper sterile technique
if they are being performed in preparation for surgery.
CAUTION: The camera and laparoscope are precision medical instruments and must be
handled with great care.
➢ Inspect the laparoscope for damage before and after use. If the laparoscope
is damaged, discontinue use and contact the manufacturer.
➢ Do not subject to impact. Put the laparoscope down carefully.
➢ Hold laparoscope only by the ocular funnel or body and not by the sheath.
Do not bend the sheath.
➢ Store laparoscopes safely in a tray or similar container.
2. Connect the light guide cable to the threaded connector on the laparoscope.
3. Fully insert the other end of the light guide cable into the light guide cable port
of the VPI (see Figure 7).
WARNING: Both ends of the light guide cable and the distal end of the laparoscope can
become hot and may cause patient or operator burns or thermal damage to
surgical equipment (for example, surgical drapes, plastic material, etc.).
Do not touch the glass tip of the light guide cable or the distal end of the
laparoscope. Do not allow the glass tip of the light guide cable or distal end
of the laparoscope to contact the patient, surgical drape, or any other
flammable material.
Avoid looking at light emitted directly from the ends of the light guide cable
and laparoscope.
26 4-0002428 Rev. K
To connect the laparoscope to the camera:
▪ Firmly insert the laparoscope eyepiece into the camera until the coupler
rotates to the locked position (see Figure 9).
1. Hold both the camera and laparoscope and rotate the coupler toward the
unlock symbol on the camera until it clicks (see Figure 10).
28 4-0002428 Rev. K
ICG imaging agent is a sterile, water soluble tricarbocyanine dye with a peak
spectral absorption at 800-810 nm in blood plasma or blood. ICG contains not
more than 5.0% sodium iodide. ICG is to be administered intravenously.
The Sterile Water for Injection, pH of 5.0 to 7.0, provided with the ICG is used to
reconstitute the ICG.
Instructions for preparation, handling and administration of ICG imaging agent are
provided in this chapter.
Note: Stryker does not distribute ICG – ICG imaging agent should be acquired through
normal hospital channels.
Regulation of ICG varies by country. In certain countries, ICG is not registered for
use as a pharmaceutical product. Contact your local distributor for more
information on the availability of ICG at your location.
ICG contains sodium iodide and should be used with caution in patients who have
a history of allergy to iodides or iodinated imaging agents due to a risk of
anaphylaxis.
WARNING: Anaphylactic deaths have been reported following ICG administration during
cardiac catheterization.
Each vial of ICG and accompanying Sterile Water for Injection are intended for use
with only 1 patient and within 6 hours of reconstitution. Discard any unused
reconstituted ICG after each surgery is completed or 6 hours have lapsed since
reconstitution. If a precipitate is present upon reconstitution, do not use and discard
the solution.
ICG powder may cling to the vial or lump together because it is freeze-dried in the
vials. This is not due to the presence of water - the moisture content is carefully
controlled. The ICG is suitable for use.
The outside packaging of needles, syringes, stopcock, ICG vials, and the Sterile
Water for Injection are NOT sterile. The contents of the ICG vial are sterile and
must be handled aseptically to maintain the sterile field during surgery.
Radioactive iodine uptake studies should not be performed for at least a week
following the use of ICG for injection.
Nursing Mothers: It is not known whether this drug is excreted in human milk.
Because many drugs are excreted in human milk, caution should be exercised
when ICG is administered to a nursing woman.
Only use ICG at indicated doses and concentrations as defined in the SPY
Portable Handheld Imaging System Operator’s Manual.
Do not use needles, syringes, stopcocks, ICG vials and Sterile Water for Injection
that appear to have packaging or seals that are compromised in any way.
30 4-0002428 Rev. K
Anaphylactic or urticarial reactions have been reported in patients with or without
history of allergy to iodides. If such reactions occur, immediate treatment with the
appropriate agents, for example, epinephrine, antihistamines, and corticosteroids
should be administered. Resuscitative measures may also be required.
WARNING: Do not use any ICG that has been reconstituted for more than 6 hours.
The ICG imaging agent can be reconstituted and prepared for injection either at the
beginning of, or during the surgery, depending on the preference of the surgical
team. Prepare ICG for administration as follows:
1. Remove one (1) 25 mg vial of ICG and one (1) 10 ml vial of Sterile Water for
Injection from the kit.
3. Remove the flip off cap on the ICG vial (25 mg) and inject the 10 ml of Sterile
Water for Injection through the stopper into the ICG vial.
5. Mix the contents of the ICG vial thoroughly and inspect the reconstituted vial
for precipitation. If precipitation is noted, continue to gently shake until all ICG
is dissolved in solution.
WARNING: If precipitation persists, do not use the mixture. Discard the reconstituted vial
and prepare a new vial, as described above.
WARNING: The total dose of ICG injected should be kept below 2 mg/kg of patient body
weight.
Note: Dosing per procedure will vary based on the procedure and is determined at the
discretion of the imaging surgeon.
Increase in dose may be required in those patients exhibiting heavy deposits of fat
on the area of interest being imaged.
1. Prior to the NIR imaging procedure, withdraw the desired dosage of ICG
solution for each planned imaging sequence into separate single use 3 ml
syringes.
32 4-0002428 Rev. K
2. ICG administration is to be performed via a central or peripheral venous line.
Use a three-way stopcock attached to an injection port on the infusion line.
Attach one prepared 3 ml syringe of ICG solution and one prepared 10 ml
syringe of saline solution.
3. Inject the prepared 2.5 mg/ml ICG solution into the line as a tight bolus.
34 4-0002428 Rev. K
The PINPOINT camera is designed for single-handed operation with ergonomic
controls. The size and weight of the camera have been optimized to ensure user
comfort during imaging when connected to a laparoscope and light guide cable.
The buttons on the camera can be used to control all PINPOINT system imaging
functions, including:
The controls and indicators on the PINPOINT camera are shown in Figure 11.
Figure 12: Primary button functions on the PINPOINT camera (top view)
36 4-0002428 Rev. K
The VPI provides NIR and white light illumination, video processing, user interface,
and other control capabilities for the PINPOINT system.
The front panel includes buttons for controlling the primary functions of the system,
indicators for power and laser-on, and ports for the PINPOINT light guide and
camera cables (see Figure 13). The rear panel includes the device identification
label and power connection ports (see Figure 4).
6. Camera port
38 4-0002428 Rev. K
WARNING: Proper sterile technique must be used when preparing the camera and
laparoscope for surgical endoscopy. The following components are not
sterile and should not be handled by a sterile operator:
Note: High-Definition Imaging Laparoscopes are optimized for imaging with visible light
and will not perform ICG fluorescence imaging. To perform ICG fluorescence
imaging, use an ICG Fluorescence Imaging Laparoscope.
2. Plug the VPI power cord into an AC wall outlet or isolation transformer.
The power indicator on the front of the VPI above the VPI power button should
illuminate amber.
▪ If the power indicator does not illuminate, check to confirm that the main
power switch (on the rear panel of the VPI) is turned to the ‘on’ position.
▪ If, after confirming that the main power switch is in the ‘on’ position, the
indicator still does not illuminate, please consult Appendix A:
Troubleshooting.
3. Connect the PINPOINT camera to VPI by plugging the camera and light guide
cables into the appropriate ports on the VPI front panel (see Chapter 2 and
Figure 13).
CAUTION: For the PINPOINT camera to initialize correctly, it must be connected to the VPI,
prior to powering on the system.
The power indicator turns green and the system enters Standby mode. The
Standby mode icon displays in the top-right corner of the video image
(see Figure 14).
Note: In Standby mode, video display is enabled and ventilation fans are on. Illumination
is not active and the video image on the display may be dark.
As the VPI powers on, the PINPOINT System emits dim flashes of colored light.
40 4-0002428 Rev. K
Figure 14: Onscreen Standby mode icon
Standby mode
In White-light mode, the VPI provides visible-spectrum illumination through the light
guide cable and laparoscope for viewing a white-light image on the system display.
▪ Press the Illumination button on the PINPOINT camera (see Figure 15).
OR
▪ The White-light mode icon displays in the top-right corner of the display
image (see Figure 16).
Note: From Standby mode, pressing the illumination button activates White-light mode.
Fluorescence imaging cannot be activated directly from Standby mode.
42 4-0002428 Rev. K
Figure 16: White-light mode icon
White-light mode
The color fidelity of the color image may be affected by external sources of white
light, such as surgical lights, overhead lighting, or daylight. Performing a white
balance optimizes the image so that white objects appear white.
Always perform a white balance if the image color fidelity appears inaccurate.
A white balance can be performed using the controls on the front panel of the VPI,
or using the buttons on the PINPOINT camera.
2. Press the White Balance button on the front panel of the VPI.
OR
a. Press and hold the Display Mode button to open the Tools menu.
An icon will appear in the center of the video image for several seconds to indicate
that the white balance is in progress and the result of the adjustment (see Figure
18).
indicates that the system has been unable to complete the white balance.
Repeat the white balance procedure.
If the white balance procedure fails a second time, turn off the system power and
contact a qualified Stryker service representative.
44 4-0002428 Rev. K
Figure 18: Onscreen white balance icons
3. While viewing the image on the video monitor, press the focus buttons
on the camera (see Figure 19) to achieve a sharp and focused image.
If the focus-stop icon displays, the focus has reached the end of its
adjustment range, press the opposite focus button.
46 4-0002428 Rev. K
Figure 20: Onscreen focus adjustment icon
▪ The laser-on indicator LED on the camera and the laser-on indicator on the
front panel of VPI illuminate to indicate that the NIR laser is on (see Figure
22).
▪ The Overlay mode icon displays in the top-right corner of the video
image (see Figure 23).
48 4-0002428 Rev. K
Figure 22: Laser-on indicator in Fluorescence Imaging mode
Note: For fluorescence imaging using the Overlay display mode, ensure that a
fluorescence imaging laparoscope is used. Attempting to use Overlay mode with
any other laparoscope, including PINPOINT white-light imaging laparoscopes, will
produce an image in which the video screen is flooded with fluorescence signal,
obscuring the white light image.
50 4-0002428 Rev. K
Note: The fluorescence display modes which are available in the Display mode cycle can
be configured in the VPI Options Menu (see Appendix B, Display options).
2. Press the Display mode button on the PINPOINT camera (see Figure 27).
The current Display mode is shown in the main display area of the screen and
the Display mode icon is shown in the top right corner (see Figure 23).
52 4-0002428 Rev. K
The PINPOINT System enables the operator to toggle between Fluorescence
Imaging mode and White-light mode at any time during an imaging procedure. The
parallel display option is not available in White-light mode.
The White-light mode icon displays in the top right corner of the screen (see
Figure 28).
To turn off illumination at any time (from White-light mode or Fluorescence Imaging
mode) and return to Standby mode:
OR
1. Ensure that the system is in Standby mode (see Turning Off Illumination
(Entering Standby Mode)).
2. If using the Sony video recorder (PC9026), press the Stop button on the
recorder to close the video case.
3. Press the power button on the front panel of the VPI. The system shuts
down and the power indicator glows amber, indicating that the main power
switch on the rear panel is in the on position.
Note: After the PINPOINT System has been used for the first time, the VPI can be
powered off using the power button on the front panel. It is unnecessary to power
off the system using the main power switch on the back of the VPI.
54 4-0002428 Rev. K
If the PINPOINT system fails to produce continuous imaging, switch the main
power off (on the VPI rear panel) for five seconds and then switch the power on
again.
The PINPOINT camera, light guide cable, and camera cable should remain
connected to the VPI during this time.
56 4-0002428 Rev. K
The PINPOINT System offers several functions that are independent of the
imaging mode. These functions can be controlled from the PINPOINT Camera at
any time during a clinical imaging procedure:
Capture screenshot
1. Press and hold the Display Mode button at any time during imaging (see
Figure 30).
Figure 30: Display Mode button on the PINPOINT Camera activates the Tools menu
The Tools menu displays on the left side of the screen (see Figure 31).
58 4-0002428 Rev. K
Figure 31: Tools menu
2. Navigate the menu by pressing the focus buttons, (up) and (down).
3. Press the Display Mode button to activate the selected icon in the menu.
4. To exit the Tools menu and return to a live image at any time, press the
Illumination button .
Video recording can be started and paused from the Tools menu or, if enabled,
using a button on the PINPOINT camera.
Note: MediCapture recorders cannot be controlled from the Tools menu or PINPOINT
camera.
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Record icon (see Figure 32).
3. Press the the Display Mode button on the camera to start and pause
recording.
A recording-active message displays in the top left corner of the screen during
recording (see Figure 33).
A recording-paused message displays in the top left corner of the screen when
recording is paused (see Figure 34).
60 4-0002428 Rev. K
Figure 34: Recording-paused message
Note: For more information about using the Stryker SDC3TM HD Information Management
System, see the device operating instructions.
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Record icon (see Figure 35).
A recording icon displays in the top right corner of the screen during
recording (see Figure 36).
Note: For more information about using Sony recorders, see the device operating
instructions.
62 4-0002428 Rev. K
To start and stop video recording:
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Record icon (see Figure 35).
3. Press the the Display Mode button to start and pause recording.
A recording icon displays in the top right corner of the screen during
recording (see Figure 36).
5. To pause video recording, press the VIDEO/PAUSE button on the front panel
6. To enable playback, press the CLOSE button on the front panel of the
recorder.
If the record icon displays briefly and then disappears, the TEAC recorder
may not be selected in the VPI Service menu. See Video Recording and
Screenshot Errors on page 111.
Note: For more information about using the TEAC UR-4MD recorder, see the device
operating instructions.
Note: This button function is disabled by default and must be enabled in the VPI Options
menu after the PINPOINT System is installed.
1. Press the rear focus button to start recording video (see Figure 38). A
recording icon displays on the monitor.
2. When using a Stryker SDC3 system or Sony recorder, press the rear focus
button again to pause recording. When using a TEAC recorder, press the rear
focus button again to stop recording.
Screenshots can be captured from the Tools menu or, if enabled, using a button on
the PINPOINT camera.
Note: When using a MediCapture recorder, control from the Tools menu or PINPOINT
camera is not available.
To capture a screenshot:
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Screenshot icon (see Figure 39).
64 4-0002428 Rev. K
The screenshot thumbnails momentarily display on the screen with the image
number in a picture-in-picture window bar (see Figure 40).
Note: For more information about using the Stryker SDC3TM HD Information Management
System, see the device operating instructions.
To capture a screenshot:
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Screenshot icon (see Figure 41).
The screenshot icon momentarily displays in the top right corner of the screen
and the captured screenshot appears briefly in a picture-in-picture window in
the lower-right corner of the screen, followed by the image number (see Figure
42).
Note: For more information about using Sony recorders, see the device operating
instructions.
66 4-0002428 Rev. K
To capture a screenshot:
1. Press and hold the Display Mode button to activate the Tools menu.
2. In the Tools menu, select the Screenshot icon (see Figure 41).
The screenshot icon momentarily displays in the top right corner of the
screen and the screenshot will be automatically saved to the system video
recorder.
Note: This button function is disabled by default and must be enabled in the VPI Options
menu after the PINPOINT System is installed.
▪ Image Quality
▪ Display Options (PINPOINT camera)
▪ Default Profile
▪ Load Profile (1-4)
▪ Save Profile
▪ Service
▪ Exit
▪ Language
For more information about the VPI Options menu, see Appendix B.
Figure 44: Options and settings that are available from the VPI Options menu
68 4-0002428 Rev. K
To access the VPI Options menu from the PINPOINT camera:
1. Press and hold the Display Mode button to open the Tools menu. Press
the bottom focus button to navigate to the wrench icon in the list
(see Figure 45).
Figure 45: Accessing the VPI Options menu in the Tools menu
2. Press Display Mode button again to open the VPI Options menu (see
Figure 46).
3. To exit the VPI Options Menu and return to a live image at any time, press the
Illumination button .
Figure 46: Display screen showing the VPI Options menu overlying the video image
1. With the VPI Options menu open, navigate by pressing the focus buttons,
2. Press Display Mode button to select the highlighted options in the menu.
The control indicators on the front panel of VPI also illuminate to reveal a down
arrow (Illumination button), up arrow (White Balance button) and right arrow
(Menu button). The buttons in the VPI front panel can also be used to navigate
the Options menu directly from the VPI.
3. Press the Illumination button on the camera to exit the VPI Options menu.
Note: If no menu selection is made within approximately 15 seconds, the VPI Options
menu will timeout and automatically exit.
To access and navigate the VPI Options menu from the VPI front panel:
70 4-0002428 Rev. K
Figure 47: Menu controls on the VPI front panel
Note: When it appears on the display, the VPI Options Menu will obstruct the video
image. It is therefore not intended to be used during a clinical procedure (see
Figure 46).
1. Press the Menu button on the front panel of the VPI or select the VPI Options
3. Press the Up or Down arrow button to highlight the desired image setting.
5. Press the Up or Down arrow button to adjust the setting. The setting takes
effect immediately and is visible in the image.
72 4-0002428 Rev. K
6. Press the Right arrow button or the button to deselect when done.
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Display Options and press the Right arrow button or the button
to select.
3. Press the Up or Down arrow button to choose the desired display options.
To restore the Image Quality settings and Display Modes selections to the default
settings:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Default Profile and press the Right arrow button or the button
to select.
Load User Profiles allows the user to retrieve a previously saved set of
configurable settings:
▪ Display Modes
▪ Image Quality
▪ On-screen Info
Up to 4 profiles are available.
To load a profile:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
Save Profile allows the user to save a set of configurable settings for future use.
These settings include:
▪ Display Modes
▪ Image Quality
▪ On-screen Info
Up to 4 profiles are available.
To save a profile:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Save Profile and press the Right arrow button or the button to
select.
3. Press the Up or Down arrow button to choose the profile number to save to.
4. Press the Right arrow button or the button to select the profile.
Selecting Service in the menu will display device specific information about the
device.
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Service and press the Right arrow button or the button to
select.
74 4-0002428 Rev. K
To select the recorder used with the system and enable recorder control from the
Tools menu:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Service and press the Right arrow button or the button to
select.
3. Highlight Recorder Setup and press the Right arrow button or the button
to select.
4. Press the Up or Down arrow button to choose a recorder from the list.
The focus buttons on the PINPOINT Camera have two optional function presets:
▪ Focus only
▪ Focus, start/pause record, capture screenshot
To set the focus button functions:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Service and press the Right arrow button or the button to
select.
3. Highlight Recorder Setup and press the Right arrow button or the button
to select.
4. Press the Up or Down arrow button to choose a recorder from the list.
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Service and press the Right arrow button or the button to
select.
3. Highlight Focus Button (PINPOINT Camera) and press the Right arrow button
4. Press the Up or Down arrow button to choose between Focus and Focus -
Record.
1. Press the Menu button on the front panel of the VPI or select the VPI Options
2. Highlight Service and press the Right arrow button or the button to
select.
3. Highlight Test Patterns and press the Right arrow button or the button to
select.
4. Press the Up or Down arrow button to choose the desired test pattern.
5. Press the Right arrow button or the button to display the test pattern.
6. Press the Right arrow button or the button to exit the test pattern when
done.
76 4-0002428 Rev. K
To change the language setting, complete the following steps:
1. Press the Menu button on the front panel of the VPI or select the VPI Options
3. Highlight the desired language and then press the Right arrow button or the
button to select.
78 4-0002428 Rev. K
The following chapter describes required processing (cleaning and sterilization) for
PINPOINT System components.
WARNING: The PINPOINT components must be cleaned and sterilized prior to the first
use and after every subsequent use. Follow the instructions provided in this
manual.
Use only the sterilization cycles outlined in this guide. Using unspecified
sterilization cycles can damage the device or result in incomplete
sterilization.
Failure to power off the VPI before starting to clean may expose personnel to
unsafe conditions and result in damage to the system.
Inspect camera head cables and light guide cables for cuts and breaks
before soaking in any fluid. See the Inspection section for more information.
Do not pull, stretch, kink, puncture, or otherwise alter the light guide cables.
Doing so will cause irreversible damage to the glass optical fibers, which will
impair light transmission through the cable.
Brush carefully around light guide cable tips to avoid damaging the optical
fibers, or light transmission loss can occur.
Do not process the VPI. The device is not designed to withstand the
procedures described in this guide. To properly clean the VPI:
80 4-0002428 Rev. K
The following table uses the Spaulding classification to define the minimum level of
sterilization required for each PINPOINT component.
Light Guide 1
Nonsterile Reusable Non-Critical
Cable
Refer to Manufacturer’s
Recorder Nonsterile Reusable
Instructions
Refer to Manufacturer’s
Printer Nonsterile Reusable
Instructions
Sterilization 4
Nonsterile Reusable Non-Critical
Trays
All Phases
Cleaning/Thermal Disinfection
Utility Water Water as it comes from the tap. Used to rinse during pre-treatment
and to prepare cleaning solutions
Soft-bristle brush2 To clean exterior of device and hard-to-reach areas of the device
Sterilization
82 4-0002428 Rev. K
1Thefollowing detergents were validated for cleaning efficacy according to the detergent manufacturer’s instructions.
Choose one (1) detergent listed below or a substantially equivalent detergent. Do not exceed the concentration and
temperature recommended by the detergent manufacturer.
Minimum
Detergent Type Minimum Soak Time
Concentration
Prolystica® HP
1/4 oz/gallon
Enzymatic Enzymatic 1 minute
(2 ml/L)
(1C22/1C24) – pH: 8.0
Prolystica HP
1/4 oz/gallon
Neutral Detergent Neutral 1 minute
(2 ml/L)
(1C21/1C27) – pH: 8.0
Prolystica HP
1/4 oz/gallon
Alkaline Alkaline 1 minute
(2 ml/L)
(1C20/1C26) – pH: 11.0
Step 1 Prepare
Step 2 Clean
Step 3 Inspect
Step 4 Sterilize
WARNING: Following all four steps is critical to ensuring the component is ready for use.
Failure to properly prepare and clean the devices could result in
contaminated components after reprocessing.
CAUTION: Laparoscope optics may be damaged if the instruments are dropped. Always
handle with care.
Handle the light guide cable with care as damage to the protective outer cover
causes damage to fiber optic light guide cable. Protect the light guide cable from
bending tightly which may damage the inner fiber bundle. It is very important to
prevent impacts, especially to the fused light entry, as this can lead to the
destruction of the fusing.
Polishing paste is included with the laparoscope. If any of the three optical surfaces
(distal end, light guide cable connector, or proximal end) have deposits that cannot
be removed in normal cleaning practices, they can be removed with polishing paste
prior to cleaning, disinfection, and sterilization.
2. Gently rub the cotton swab on the optical surface, cleaning in a circular motion.
3. Use acetone or alcohol on a clean cotton swab to remove the paste residue.
CAUTION: Polishing optical surfaces should not be part of routine cleaning procedures.
Repeated application of polishing paste to the laparoscope lens can cause
damage. It should only be performed when the laparoscope image is cloudy due to
deposits that cannot be removed in normal cleaning practices.
84 4-0002428 Rev. K
1. Wipe excess soil from the device.
2. Prevent soil from drying by keeping the device moist until it is ready for further
processing.
1. Transport the device in a tray to avoid damage. Follow the facility’s internal
procedures for the transportation of contaminated surgical instruments and
devices.
WARNING: Compatibility with an ultrasonic treatment has not been evaluated and it can impact
functional performance.
CAUTION:
1. Rinse the device under cool, running utility water to remove gross soil from
device. Ensure all external surfaces and any accessible internal surfaces are
rinsed.
86 4-0002428 Rev. K
1. Soak
a. With the device immersed in the solution, thoroughly brush the exterior
with a soft-bristle brush. Focus brushing on dead-end holes and mated or
rough surfaces.
b. Actuate and brush any movable parts in all extreme positions.
c. With the device immersed in the solution, use a bottle brush to thoroughly
clean any dead-end holes or hard-to-reach areas of the device.
3. Rinse
a. Rinse the device with critical water until all detergent residue is removed.
b. Once all detergent residue is removed, continue to rinse for 10 seconds.
c. Drain excess water from the device and dry it using a clean, lint-free cloth
or pressurized air.
d. Visually inspect the device for cleanliness, paying close attention to hard-
to-reach areas. If visible soil remains, repeat steps 1-3.
1. Place the device in the automated washer (on an incline, if feasible, to facilitate
drainage).
The following minimum and maximum washer parameters have been validated for cleaning
efficacy and device functionality:
3. Drain excess water from the device and dry it using a clean, lint-free cloth or
pressurized air.
4. Visually inspect the device for cleanliness, paying close attention to hard-to-
reach areas. If visible soil remains, repeat steps 1-4.
Note: Thermal disinfection value A0 ≥3000 may be utilized for compatible devices.
Do not exceed the time and temperature specified in the table above.
88 4-0002428 Rev. K
Inspect the laparoscope before each use. If a problem relating to functionality or
appearance is observed or suspected, the device should be returned to Stryker for
repair.
2. Test the image quality of the laparoscope before each use. The image shall not
be dark, blurry, or cloudy, or have any obstruction, debris, or abnormalities in
the field of view. Also ensure that light is transmitting efficiently from the light
guide connector to the distal end. Fiber transmission can be evaluated by
orienting the light post toward an overhead light and looking at the fiber
illumination pattern at the distal end of the laparoscope. If a problem is
observed that cannot be resolved from cleaning or polishing the optical
surfaces, the device should be returned to Stryker for repair.
4. Inspect the optical surfaces of the laparoscope for clarity. If there are deposits
that need to be removed, follow the instructions in the “Using Polishing Paste”
section.
a. Point one end of the light guide cable toward a light source such as a
window or lamp.
b. Visually check the other end of the cable for dark spots. Dark spots
indicate broken optical fibers.
c. If the dark spots cover more than 20 to 25% of the light-output surface,
discard the light guide cable.
d. Inspect the light entry and exit surfaces of the light guide cable to ensure
they are smooth and clean. Debris deposits on these surfaces may result
in insufficient light output. Continued use of light guide cables with debris
may result in progressive damage to the laparoscope.
OR
▪ Corrosion
▪ Sharp edges
1. Corrosion or discoloration
90 4-0002428 Rev. K
WARNING: Laparoscopes must be sterilized in the sterilization tray (PC9017) supplied by
Stryker.
WARNING: Maintain sterilization trays regularly and per the manufacturer’s instructions
provided on Case Medical’s website.
CAUTION: Ensure that optical surfaces do not contact hot metal surfaces, as this may damage
the entire system.
2. Ensure that all devices are disassembled for sterilization; refer to Figure 48
and 49.
3. Insert four new filters, provided in the sterilization accessories kit (PP9076 or
PP9066), two each in both the sterilization tray (PC9017) bottom and lid as
shown in the following figures. The tray must be clean and dry prior to use.
WARNING: Using the incorrect filters will result in the product not being properly
sterilized and may lead to the failure of the sterile barrier.
All four filters must be replaced with new filters prior to sterilization.
92 4-0002428 Rev. K
Figure 51: Installing filters in the sterilization tray (PC9017) lid
4. Pack the cleaned, dried, and inspected laparoscopes in the sterilization tray
(PC9017) supplied by Stryker (see Figure 52).
5. Mount the cable basket onto the four posts and then pack the clean, dried, and
inspected light guide cable in the cable basket (see Figure 53).
7. Secure the lid to the sterilization tray (PC9017) and apply the tamper-evident
seals (see Figure 55).
Note: The color of the seal will vary depending on the sterilization process used.
8. Sterilize the sealed sterilization tray (PC9017) supplied by Stryker using one of
the methods below.
94 4-0002428 Rev. K
Perform one of the following sterilization cycles.
Sterrad NX Standard
1
Caution: The Prion Cycle is only validated for use with the laparoscope. Do not use this cycle for
the light guide cable or the camera.
2
Caution: This cycle, recommended for control or elimination of Transmissible Spongiform
Encephalopathies, may be used. However, a device exposed to a long cycle should be expected to
have reduced functional life.
3
Drying time depends on several variables; including altitude, humidity, preconditioning, size of
chamber, mass of load, and placement in the chamber. Users must verify that drying time set in their
autoclave yields dry surgical equipment.
WARNING: After sterilization, allow all devices to cool to room temperature and dry
completely before reassembly and reconnection to consoles. Any moisture
on the device can cause fog during use, resulting in potential image
occlusion.
96 4-0002428 Rev. K
The device’s service life is largely determined by wear, processing methods, and
any damage resulting from use. To extend the time between device servicing,
always follow the care and handling instructions in this user manual.
Before each use, test the device functionality and inspect it for any sign of damage
per the Inspection section. If the device does not function properly or appears to be
damaged, return it to Stryker for evaluation and possible repair or replacement.
Repairs through Stryker as the equipment manufacturer bring the device back to
manufacturer specifications. Clean and (when applicable) sterilize all potentially
contaminated devices before returning them to Stryker.
1. AAMI TIR12: Design, testing and labeling reusable medical devices for
reprocessing in healthcare facilities: a guide for medical device manufacturers
98 4-0002428 Rev. K
Reprocessing the PINPOINT camera requires a four-step process:
Step 1 Preparation
Step 2 Manual Cleaning
Step 3 Inspection
Step 4 Sterilization
WARNING: Following all four steps is critical to ensuring the component is ready for
use. Failure to properly prepare and clean the device could result in a
contaminated component after reprocessing.
When preparing and cleaning, ensure all foreign matter is removed from the
surface of the device. This allows the active ingredients of the sterilization method
to reach all the surfaces of the device.
CAUTION: Do not use metal or abrasive brushes, scrub pads, or rigid tools when cleaning the
PINPOINT camera. The optical elements can be scratched, permanently damaging
the equipment.
Note: The PINPOINT camera may be contaminated by contact with human tissue and
fluids during clinical use. To avoid drying of blood, protein, and other substances
on the camera, prepare it for cleaning within one hour after use.
2. Measure and dispense enzymatic detergent into the water based on the
detergent manufacturer instructions.2
5. With the device immersed in the solution, thoroughly brush the exterior with a
soft-bristled brush, focusing on any mated or rough surfaces.
7. Rinse each device with water1 until all detergent residue is removed.
10. Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas (see Figure 56).
1 Reverse Osmosis/Deionized water (RO/DI) at 30°C was used for validating cleaning efficacy.
CAUTION: Do not scratch contaminants off the camera using hard instruments as this can
damage the lens, cable, or buttons.
After preparation and following the General Practice guidelines, manually clean the
PINPOINT camera according to the following detailed manual cleaning instructions.
2. Wipe the entire surface of the device using a soft clean cloth dipped in the
detergent solution.
3. Immerse the device in the detergent solution, ensuring the solution contacts all
inner and outer surfaces.
5. With the device immersed in the solution, thoroughly brush the exterior with a
soft-bristled brush, focusing on any mated or rough surfaces.
7. Rinse each device with water1 until all detergent residue is removed.
9. Drain excess water from the device and dry it using a clean cloth or
pressurized air.
10. Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas.
1 Reverse Osmosis/Deionized water (RO/DI) at 30°C was used for validating cleaning efficacy.
After cleaning, inspect the camera for corrosion, damaged surfaces, chips, or
contamination. Soil may accumulate on the surfaces of mechanics and optics. If
the PINPOINT camera is found to be contaminated, it must be cleaned again
following the procedures described in Steps 1 and 2.
▪ damage
▪ sharp edges
▪ dents
▪ cracks
▪ signs of mechanical or thermal damage
2. Verify that the proximal and distal glass surfaces are clean and free of debris.
Remove any soil or debris deposits using cotton swabs soaked in alcohol.
To sterilize the PINPOINT camera using the Stryker Sterilization Tray (PC9018):
1. Insert three new filters provided in the sterilization accessories kit (PP9066):
two in the sterilization tray (PC9018) bottom and one in the lid (see Figure 57
and Figure 58).
The sterilization tray (PC9018) must be clean and dry prior to use.
WARNING: Using the incorrect filters will result in the product not being properly
sterilized and may lead to the failure of the sterile barrier.
All three filters must be replaced with new filters prior to sterilization.
2. Pack the cleaned, dried, and inspected PINPOINT camera in the sterilization
tray (PC9018) supplied by Stryker (see Figure 59 and Figure 60).
3. Secure the lid to the sterilization tray (PC9018) and apply the tamper-evident
seals (see Figure 61).
Note: The color of the seal will vary depending on the sterilization process used.
As the system enters Standby mode, it will initialize the connected camera. If a
2. Disconnect and reconnect the PINPOINT camera, and then power on the VPI
again to clear the error.
3. If the camera fails a second time, turn the VPI power off and contact a qualified
Stryker service representative.
Camera failed
1. Press the illumination button on the VPI twice to clear the error.
2. If the illumination fails a second time, turn off the system power off and contact
a qualified Stryker service representative.
Illumination-failed
If the VPI is configured for recording but not properly connected to the recorder, or
the recorder is unresponsive and Video Recording or Screen Capture is selected,
the screen may momentarily pause (<0.5s) before returning to normal without
providing confirmation of the recording occurring.
Figure 64: Recorder or image flip unavailable icons in the Tools menu
To select the recorder in the VPI Options menu, see Recorder Setup on page 72.
For more information about configuring Sony recorders, see Setting Up the Sony
1000MD Recorder for Control from the VPI on page 102 or Setting up the Sony
3300MT Recorder for Control from the VPI on page 105.
1. Press the MENU button on the front panel of the recorder. The main recorder
menu opens on the display.
2. Using the arrow buttons on the front panel, navigate to Settings and press the
ENTER button.
3. In the Settings menu, select System Admin Settings and press the ENTER
button.
5. In the System Admin Settings menu, select Device Settings and press the
ENTER button.
6. In the Device Settings menu, select the RS-232C 1 field and press the
ENTER button.
8. Next to the RS-232C field, select Advanced and press the ENTER button.
9. In the System Control Setting dialog box, in the Speed field, select 115200
bps from the list.
11. In the Device Settings menu, select Apply and press the ENTER button.
1. Press the MENU button on the front panel of the recorder. The main recorder
menu opens on the display.
2. Using the arrow buttons on the front panel, navigate to Settings and press the
ENTER button.
4. In the warning dialog box, select OK and press the ENTER button.
5. In the System Admin Settings menu, select Device Settings and press the
ENTER button.
8. Next to the RS-232C field, select Advanced and press the ENTER button.
12. In the Device Settings dialog, click Apply and then press the MENU button to
exit.
Table 9 Troubleshooting
Operating mode does not PINPOINT is in If PINPOINT is in standby (illumination off) mode:
change when illumination Standby
1. Press the illumination button on the front
button is pressed and (illumination off) panel of the VPI.
released. mode.
2. Press illumination button on the camera
to change operating mode.
Display option does not PINPOINT is not in If PINPOINT is in white light operating mode:
change when display Fluorescence 1. Press the illumination button on the front
button is pressed and Imaging Mode panel of the VPI.
released.
2. Press illumination button on the camera
to change operating mode.
If PINPOINT is in standby (illumination off) mode, see
previous troubleshooting item.
The video image is out of The focus Press the focusing buttons on the camera until a clear
focus. adjustment on the image is achieved.
camera has not
been set to the
correct position.
The video image appears The Sharpening 1. Check the Image Quality value for
grainy or noisy. value is too large. Sharpening using the option menu.
2. If the value is 5 or greater, reduce the
value.
The video image appears The Brightness 1. Check the Image Quality value for
too bright. value is too large Brightness using the option menu.
2. If the value is greater than 192, reduce
the value.
The video image appears The Brightness 1. Check the Image Quality value for
too dark. value is too small Brightness using the option menu.
2. If the value is less than 175, increase
the value.
Image quality is poor due Image quality Reset the Image Quality settings by selecting the
to poor color or brightness. settings have been Default Profile from the user options menu.
altered significantly
Other equipment in the Radio frequency 1. Check to see if any of the following
vicinity of PINPOINT interference may be resolve the problem:
seems to malfunction occurring between ▪ Reorient or relocate PINPOINT or
when PINPOINT is PINPOINT and the the equipment receiving the
powered on, but works malfunctioning interference.
normally when PINPOINT equipment.
is powered off. ▪ Increase the separation between
PINPOINT and the equipment
receiving the interference.
▪ Connect PINPOINT to a power
outlet on a different circuit from that
to which the other equipment is
connected.
2. If PINPOINT becomes unresponsive,
and does not resume normal function
after turning off and back on, stop using
PINPOINT and contact a qualified
Stryker service representative.
2. Carefully remove the fuse cover with the fuse located next to the three-prong
power connector on the rear panel of the VPI.
5. If the VPI fails to operate properly again, contact a qualified Stryker service
representative for repair.
Increasing the sharpening value increases the degree to which the displayed video
is sharpened. Video sharpening is an image processing feature and does not affect
the focus of the PINPOINT camera.
The Sharpening setting can be set to any value between 0 and 10.
The Brightness setting controls the overall image brightness displayed on the video
monitor. The system maintains the displayed video brightness at a constant value
regardless of the distance between the distal tip of the laparoscope and the tissue.
The Brightness setting can be set to any value between 0 and 255. Increasing this
value, increases the brightness of the video displayed on the monitor.
Increase the Red Saturation value to increase the overall redness of the video
displayed on the monitor.
The Red Saturation can be set to any value between 0 and 1024.
Increase the Blue Saturation value to increase the overall blueness of the video
displayed on the monitor.
The Blue Saturation can be set to any value between 0 and 1024.
The Peak/Mean menu enables control of the method that the imaging system uses
to set the scene brightness for viewing objects at different distances from the tip of
the laparoscope. There are three modes in the Peak/Mean menu: Peak, Mean, and
Balanced.
Select Peak to set the brightness for viewing small, foreground objects that are
closer to the tip of the laparoscope. Background objects farther from the
laparoscope may fall into darkness.
For example, use the Mean setting to set the brightness for viewing objects that are
farther from the front of the laparoscope. Foreground objects may appear too bright
to see detail.
Select Balanced to set the brightness by balancing the Peak and Mean settings.
For example, use the Balanced setting when viewing objects both near to and
farther away from the tip of the laparoscope. Balanced is the default setting and is
suitable for most scenarios.
Default Profile allows the user to return to the settings as originally installed on the
imaging system. The following are the default settings:
SPY
Sharpening: 4
Brightness: 192
Peak/Mean: Balanced
▪ DSP version
▪ Boot version
▪ FPGA version
▪ LCB version
▪ VPI serial number
▪ PINPOINT camera version
▪ PINPOINT camera serial number
During fluorescence imaging, the imaging system offers up to three different video
display options.
The SPY image display shows only the NIR fluorescence on the monitor in gray
scale. No white-light image is displayed.
The Overlay image display combines the white light image and the NIR fluorescent
image. In this mode, the NIR fluorescence appears green on top of a high-definition
white-light image.
In this imaging display, the white-light image is shown as a gray scale image. The
NIR fluorescent image is color-scaled, with red representing most fluorescence and
blue representing least fluorescence and displayed on top of the white-light image.
In addition, the color scale of fluorescence is shown such that as the distance
between the tip of the laparoscope and the tissue changes, the colors remain
approximately the same. This allows assessment of the fluorescent image at
different imaging distances.
White light/PINPOINT
Camera
mode
Cable length 3m
Resolution HD compatible
Physical Length 3m
Laser class Complies with 21CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice No. 50, dated June
24, 2007.
Apertures for NIR radiation emission Laparoscope tip, light guide cable tip
Wavelength 805 nm
THE PINPOINT SYSTEM INCLUDES PINPOINT SYSTEM SOFTWARE WHICH NOVADAQ IS PREPARED TO
LICENSE YOU TO USE AS PART OF YOUR USE OF THE PINPOINT SYSTEM BUT ONLY ON THE TERMS AND
CONDITIONS SET OUT IN THIS AGREEMENT. BY CLICKING “I Accept" ON THE SCREEN, OR BY USING THE
PINPOINT SYSTEM SOFTWARE IN ANY WAY, YOU CONFIRM YOUR AGREEMENT TO BE BOUND BY THE
TERMS AND CONDITIONS OF THIS AGREEMENT. IF YOU DO NOT AGREE, YOU MUST CEASE ALL USE OF
THE PINPOINT SYSTEM SOFTWARE AND RETURN IT TO THE DISTRIBUTOR OR VENDOR FOR REFUND OR
CREDIT WHERE APPLICABLE.
1. DEFINITIONS
1.1 “Documentation” means the operator guides, manuals and other instructions for use of the Program that
Novadaq makes generally available.
1.2 “Program” means Novadaq’s PINPOINT System Software for operating and using the PINPOINT System (in
object code form only) and any updates released by Novadaq, together with associated Documentation and
media (if any) provided to You.
1.3 “PINPOINT System” means the fluorescent imaging system designed and manufactured by Novadaq having
the following components: a surgical laparoscope, a camera head (“PINPOINT Camera”), a SPY-PHI
handheld camera head, a flexible light guide cable, an endoscopic video processor, PINPOINT cart, monitor,
recorder, printer and sterilization trays.
2. PROGRAM LICENSE
2.1 Limited License. Subject to Your compliance with the terms of this Agreement, Novadaq hereby grants to
You, during the term of this Agreement, a limited, non-exclusive, non-transferable license:
(a) to use a single copy of the Program solely for Your own internal business operations solely on the PINPOINT
System on which the Program was first installed, consistent with the Documentation;
(b) to use the Documentation provided with the Program in support of Your authorized use of the Program; and
(c) to copy the Program for archival or backup purposes.
2.2 Restrictions. You will not copy or use the Program except as expressly permitted by this Agreement. You will
not relicense, sublicense, rent or lease the Program or use the Program for third-party training, commercial
time-sharing or service bureau use. You will not, and will not permit any third party to, (a) copy, modify, translate,
reverse engineer, disassemble or decompile any Program, or otherwise determine or attempt to determine
source code of the Program, or (b) create any derivative work based upon the Program embedded in the
PINPOINT System, except to the extent expressly permitted by applicable law. You will not use the Program
on multiple processors without the prior written consent of Novadaq. Novadaq may, at its own expense and
upon reasonable advance notice to You, conduct an audit of Your use of the Program to verify Your compliance
with this Agreement.
2.3 Ownership. Novadaq will retain all right, title and interest in and to patent, copyright, trademark, trade secret
and any other intellectual property rights in the Program and any derivative works thereof, subject only to the
limited licenses set forth in this Agreement. You do not acquire any other rights, express or implied, in the
Program or the Documentation other than those rights expressly granted under this Agreement.
Copyright
Novadaq® is trademark of Novadaq Technologies ULC