Repair of Blood Pressure Measurement Devices
Repair of Blood Pressure Measurement Devices
Measurement Devices
Blood pressure measurement is a routine examination that determines the
pressure in the arteries. The measurement provides information about the
pumping capacity of the heart and the condition of the blood vessel walls.
Blood pressure measurement is a harmless and easy-to-perform diagnostic
measure.
Blood pressure
Blood pressure (BP) is the pressure of the circulating blood in the large
arteries and is therefore the result of the pumping action of the heart. Blood
pressure is one of the vital signs that indicate the status of the body's life-
sustaining functions.
Two values are used to express arterial blood pressure: systolic pressure,
the maximum pressure during one heartbeat, and diastolic pressure, the
minimum pressure between two heartbeats. During each heartbeat, the
blood pressure varies between the systolic and diastolic pressures. The
pressure is measured in millimetres of mercury (mmHg).
In addition to systolic and diastolic pressure, mean arterial pressure
(MAP) is sometimes used. This is the average arterial pressure during a
cardiac cycle.
Blood pressure is not static, but varies throughout the day. Normal resting
blood pressure for an adult is about 120 mmHg systolic and 80 mmHg
diastolic. This result is expressed as '120 over 80' and written as BP 120/80.
A patient has hypotension, or low blood pressure, when the blood pressure
is lower than 90/60. High blood pressure is called hypertension and is
defined when the blood pressure is consistently higher than 140/90. Both
conditions are indicators of disease.
Principles of operation
There are two main methods of measuring blood pressure:
the auscultatory method and the oscillometric method.
Auscultatory method
Systolic and diastolic pressures are measured by listening to the flow of
blood in the brachial artery, the main blood vessel in the arm, with a
stethoscope. To do this, the cuff of a sphygmomanometer is wrapped around
the patient's upper arm and a stethoscope is placed over the brachial
artery. When the cuff is pumped up and the pressure becomes higher than
the patient's systolic blood pressure, the blood flow through the artery stops.
No sound can be heard at this point.
The pressure is then slowly released. When the pressure in the cuff is lower
than the pressure produced by the heart, blood begins to flow through the
artery and a flow noise can be heard. This noise, which is the result of
turbulence in the blood, is also called the Korotkoff sound. The pressure at
which this sound first appears is the systolic blood pressure. If the pressure
in the cuff drops but is still high enough to narrow the artery, the turbulence
can be heard. When the sound finally disappears, the pressure in the cuff has
reached the diastolic blood pressure.
Flow noises are nothing unusual. It can also be heard in everyday life when
water pipes are partially blocked or water valves are not fully opened.
The beginning of the flow sound marks the systolic pressure and the last
audible sound marks the diastolic pressure.
Oscillometric method
Automated blood pressure monitors work a little differently. Instead of
listening to the sound of the blood flowing, a sensor detects oscillations in
the blood flow. These oscillations can be detected by a pressure sensor as
small changes in pressure. This method also uses a cuff to control the flow of
blood through the artery. Again, the beginning of the oscillation marks the
systolic pressure and the last detectable oscillation marks the diastolic
pressure.
Mercury sphygmomanometer
In a mercury sphygmomanometer, the weight of the mercury column in the
riser tube acts against the cuff pressure and thus against the blood pressure.
The height of the mercury column also acts as an indicator. This direct
measurement means that no other mechanics are required.
As the working principle is based on the specific weight of mercury, which
does not change, mercury sphygmomanometers do not require calibration.
They are considered the gold standard. Readings are always correct as long
as the riser is vertical and the mercury level is '0'.
Mercury sphygmomanometers are no longer used in Europe and the USA
because of the toxicity and environmental impact of mercury. They are being
replaced by aneroid instruments. However, mercury sphygmomanometers
are still common in developing countries.
Cuff
Traditionally a cuff consists of two parts, an inner bladder made from rubber
and the outer envelop made from strong fabric, often nylon, with a Velcro
fastener. Attached to the bladder are one or two rubber hoses. Cuffs for
mercury and aneroid sphygmomanometers have two tubes, one for the
pumping bulb and one for the pressure gauge. Electronic home care devices
only have one tube.
Cheap cuffs as they are found today in the developing world consist of one
piece only. They are made of plastic and are not as strong as those with a
separate bladder and fabric. They do not last very long and are almost
impossible to repair.
With good cuffs, it is easy to remove the bladder. The fabric sleeve can then
be washed or the bladder repaired.
Cuffs are available in various sizes. The size is based on the circumference of
the upper arm. The cuff should cover two-thirds of the length of the upper
arm. If the cuff is too small, the blood pressure reading will be too high. If the
cuff is too large, the result will be too low. The range of arm circumferences
is printed on each cuff.
Typical upper arm cuff sizes are:
The size is approximately the width of the cuff in cm. If in doubt, it is better
to take the next larger cuff than a cuff that is too small.
Stethoscope
The stethoscope is a simple acoustic examination instrument. It is used for
listening to the sounds of heart and lung and the blood flow during blood
pressure measurement.
The stethoscope consists of a sound-detecting device,
the chest piece, and two earpieces connected by a Y-shaped flexible plastic
tube.
The principle is simple: sound is transmitted through air-filled hollow tubes to
the user's ears. The sound is picked up by a resonator called the chest piece.
The chest piece is a metal housing that usually has two sides, one with a
larger diameter and one with a smaller diameter. The larger side usually has
a diaphragm, which is a thin plastic disc. The other, smaller side is called the
bell and is shaped like a small cup with a small hole. The larger, flat side with
the diaphragm is used to hear higher frequencies and the smaller bell is used
to hear lower frequencies.
For examination, the user must cover the opposite side with a fingertip if
there is no mechanical closure. The chest piece is placed on the patient,
body sounds move the diaphragm and the vibrations cause small changes in
air pressure in the tubes that travel to the earpieces.
NIBP monitor
In addition to manual blood pressure monitors, there are also automated
blood pressure monitors. These electronic blood pressure monitors are used
in operating theatres, intensive care units and other areas where patients'
vital signs need to be monitored around the clock. For this purpose, the
devices take readings regularly and independently. This type of automated
blood pressure monitor is also known as an NIBP (Non-Invasive Blood
Pressure) monitor. NIBP monitors have programmable alarms and also
measure mean arterial pressure and pulse rate.
Typical patient monitor measuring blood pressure. Only the cuff is connected
and only the blood pressure is shown on the display.
1. The tested person has to be relaxed, calm and should not speak during
the measurement.
The arm should be placed on a table with the palm facing upwards so
that the blood
pressure cuff is at the same height as the heart.
2. Fasten the cuff around the bare upper arm. The tube connectors should
point towards the
crook of the arm. There should be about 3 cm (1 in) of space between
the crook of the arm
and the cuff for the stethoscope.
3. Hold the inflating bulb with the pressure gauge with your right hand.
Close the release
valve on the bulb by turning the screw clockwise. Squeeze the bulb and
inflate the cuff.
The pressure should be about 180 to 200 mmHg.
4. Place the stethoscope in your ears and the stethoscope head between
the crook of the arm
and the cuff where the artery is located.
5. Open the release valve slightly and slowly deflate the cuff. Listen
carefully to the sound of
the pulses and watch the gauge carefully.
6. The systolic blood pressure is shown on the dial gauge when you first
hear a rhythmic
thumping sound. In a healthy person, the reading is around 110 to 120.
Memorise this
number.
7. Continue to slowly release the pressure and continue to listen to the
sound. When you no
longer hear the knocking sound, the diastolic pressure has been
reached. This reading is
normally around 60 to 80 mmHg.
8. The blood pressure measurement is complete. Now deflate the cuff
completely.
If you have done something wrong, perhaps you released the pressure too
quickly, do not inflate the cuff again right away. Wait at least a minute
before repeating the measurement.
There is no single reference value for blood pressure. However, the following
blood pressure values can be considered normal for adults:
Construction
Mercury and aneroid sphygmomanometers and digital blood pressure
monitors differ in the way they work and display blood pressure. The first two
are mechanical devices, whereas NIPB monitors are electronic devices where
not only is the display digital, but the measurement process is automatic.
Mercury sphygmomanometer
NIBP monitor
While the sphygmomanometer blood pressure method is based on listening
to Korotkoff sounds, electronic blood pressure monitors work differently.
They use the oscillometric method.
A pressure transducer (pressure sensor) in the device detects small pressure
changes in the cuff pressure caused by oscillations in blood flow. A filter and
amplifier then separates the oscillation signal, an A/D (Analogue/Digital)
converter converts the signal into digital data, and a microprocessor
interprets the signals and displays the result.
Additionally the inflation and deflation of the cuff work automatically.
Hospital NIPB monitors are also be timer controlled so that a long term
monitoring can be carried out.
Home blood pressure monitor
The following is the construction of an automated blood pressure monitor for
home use. These devices are also called digital blood pressure devices.
The word digital refers only to the display, the blood pressure measurement
itself is analogue. As the accuracy of the device depends on the conversion
of the measurement signal from analogue to digital, or on its calibration,
these devices are not fundamentally more accurate than an aneroid
sphygmomanometer. They are just easier to use.
The air pump consists of a small DC motor with attached diaphragm pump.
The pump pumps air through the connector into the cuff. At this moment the
solenoid valve is closed. When the device releases the pressure from the
cuff, the pump stops and the solenoid valve opens.
Also connected to the tubing is the pressure sensor on the electronic board.
As the pressure in the system is everywhere the same, the sensor must not
be close to the cuff. That is why the sensor can be mounted directly on the
PCB. A microcontroller on the board analyses the pressure in the cuff as well
as the oscillations during blood flow through the artery, drives the display
and controls pump motor and solenoid valve.
The inside of a typical automated blood pressure monitor for home use.
In principle, more expensive clinical blood pressure monitors do not work
much differently. However, there are two things that are different, apart from
additional display options. For more accurate measurement, these devices
usually have a second tube. Inflating and deflating the cuff is separate from
measuring the pressure. The pressure sensor has its own tube. Also the
pump is bigger and more robust, because the clinical monitors need to be
able to take measurements regularly and automatically.
Let's take a look at the tubing. The pump is at the top right. A one-way valve
(red) is connected to it. A pressure sensor is connected to the first T-piece.
This is followed by another T-piece to which a solenoid valve (yellow) is
connected. This gradually releases the pressure during the measurement.
From there, a tube leads to another pressure sensor and to the patient
connector.
Repair
Blood pressure monitors should always be delivered for repair with the cuff
and connecting tube, as this is often where the fault lies.
Note accessories such as cuff and power adapter on the job card to avoid
confusion later.
Put the cuff on yourself and carry out a test run in order to confirm the
fault. Make sure that
the problem is really a fault and not an operating error.
If you are not sure how to operate the unit yourself, consult the user
manual.
Leakage test
Leaks in the pressure cuff or connecting tubing can be detected with a
simple test set-up. Wrap the cuff around a glass bottle approximately the
diameter of an arm. Apply pressure to the cuff with the rubber bulb until
200 mmHg is reached. The pressure should remain stable and should not fall
by more than 2 mmHg in 10 seconds.
If you are using an automatic blood pressure monitor, disconnect the
connecting tube and connect it to a rubber bulb and follow the same
procedure.
Performance test
Calibration is performed by comparing the device under test with a reference
instrument. This reference instrument, for example a mercury
sphygmomanometer, is attached to the device under test by a Y-piece. The
cuff is wrapped around a glass bottle, such as an infusion bottle, and
calibration can be performed.
However, before calibrating, perform a leakage test as described above.
Then calibrate and adjust the '0' position of the sphygmomanometer.
Aneroid sphygmomanometers are set to '0' as described above.
Aneroid sphygmomanometers can now be calibrated at different pressures.
Mercury sphygmomanometers do not require calibration as the specific
weight of mercury does not change. It is only necessary to ensure that the
level indicates "0" when no pressure is applied. Aneroid manometers are
different. The mechanics can wear out or get damaged over time. They need
a regular calibration.
1. Inflate the cuff to 200 mmHg and compare the readings. The result
should not differ by
more than ± 3 mmHg.
2. Repeat the procedure at 150 mmHg, 100 mmHg and 50 mmHg.
3. Inflate the cuff again to 200 mmHg and check that the pressure remains
stable for
10 seconds. The system should not lose more than 2 mmHg during this
time.
Common problems
Apart from dropped aneroid sphygmomanometers that need to be
readjusted, most problems are related to leaks. Leaks can be caused by
holes in the bladder, worn tube connections, porous tubes and rubber bulbs,
dirty release valves or damaged pressure gauges.
Leaks in the bladder of the cuff can usually be repaired. This involves taking
the bladder out of the fabric. The fabric must be opened on one side and
sewn back together later. A rubber bladder can then be repaired in the same
way as a bicycle inner tube. The first step is to locate the leak. Inflate the
bladder and watch and listen for air escaping. Major leaks will be easy to
hear or see. A quicker way is to put the inflated bladder in a bucket of water.
The escaping air will create bubbles.
Cheap cuffs without inner bladders can often not be repaired because rubber
glue does not work on synthetic or plastic materials. Instead of using rubber
glue and bicycle repair patches, try gluing a piece of similar synthetic
material, for example from an old cuff, with elastic contact adhesive or
silicone.
After the mercury is pressed through the gauze, the black oxide remains.
If you have a sphygmomanometer where the bulb and pressure gauge are
combined, this method will not work. In this case, the gauge housing must be
opened and the position of the pointer itself corrected. To do this, unscrew
the metal ring, remove the glass and carefully lift the pointer off the shaft.
Sometimes a tool such as a small, flat screwdriver is needed. Do not use
pliers or forceps to remove the pointer. It can be easily deformed. Once
removed, the pointer can be carefully placed back in the correct position.
To lift the pointer without damaging it, you can cut a strip of epoxy from
an old circuit board and drill and file a small slot.
Automated blood pressure device for home care use are usually cheaply
made and often hardly repairable. They are also not designed for later
calibration and adjustment. They are disposable products. Battery-powered
devices suffer from high battery consumption. This is normal because the
pump motor draws quite a lot of current. In developing countries, where
these devices are often used in hospitals, this is a problem, especially
because batteries are either of very poor quality or very expensive.
Replacing the batteries with rechargeable ones does not work either,
because the resulting operating voltage will be too low. Typically, these
devices run on 6 V, which is supplied by 6x 1.5 V AA batteries. But 6x 1.2 V
AA rechargeable batteries will only give you 4.8 V, which is often too low to
run the device, or not long enough.
The best solution is to use an external power supply. This should provide at
least 500 mA. Universal power supplies are available in any city. If your
blood pressure monitor works reliably on 5 V, you can also use a USB power
supply, which is cheaper and easier to get.
Clinical blood pressure monitors are robust devices that do not break
often. When a problem does occur, it is often related to the tubing. Leaks are
possible due to ageing of the tubes. It is also possible that dirt has got into
the system and is blocking a sensor or the solenoid valve. This should not
actually happen, as there should be a filter before the pump.
Further literature
On Wikipedia you can find further articles about these topics:
Blood pressure
Blood pressure measurement
Diastole
Korotkoff sounds
Pressure measurement
Pressure sensor
Pulse
Sphygmomanometer
Stethoscope
Systole