BPR-5V628 968 User Manual
BPR-5V628 968 User Manual
Laboratory Refrigerator
BPR-5V628/968
User Manual
BIOBASE GROUP
Version 2022.05
Machine Translated by Google
Statement
Thank you for choosing and using the product of BIOBASE. For your safe and convenient use and
reasonable maintenance of the product, please read the operation instruction carefully and keep it
properly for reference.
For damages of any instrument due to the fact that the user does not use the product according to the
operating instrument environment declared in the handbook or injuries due to the fact that
individual does not operate the product according to safety instructions, BIOBASE has no
obligations and responsibilities to be in charge of them.
The user must accomplish the following three points when using the product:
1. Always use protective devices correctly (including clothes, gloves, goggles, etc.);
2. Always adopt good health habits and operate strictly according to the product instruction;
3. Everyone is obliged to be in charge of one's own safety.
Due to the fast product update of BIOBASE, there may be differences between function of the
product you bought and the function mentioned in the instruction, please in kind prevail.
Safety Instruction:
·Please read the handbook carefully when using the machine for the first time.
·The Laboratory refrigerator can only be operated by trained and authorized personnel.
·The maintenance of the equipment can only be accomplished by BIOBASE or agents authorized by
BIOBASE.
·if operator runs into conditions not mentioned in the instruction, please contact BIOBASE or
agents authorized by BIOBASE for inquiry correct processing method.
·please use accessories provided by BIOBASE. if user uses other accessories, BIOBASE will not
take charge of generated negative effects. However, the user can apply to BIOBASE to test and verify
that whether the accessories accord with the requirements of BIOBASE.
·It is necessary to inspect and maintain the Laboratory refrigerator at specified time intervals.
·The Laboratory refrigerator is strictly prohibited to store living things or other goods which have
strict temperature requirement and are unsuited for storage at constant temperature.
·The Laboratory refrigerator performs refrigeration through the heat dissipation of the back surface
(condenser). In order to ensure normal operation and ventilation and heat dissipation, the back and
The left and right sides of the refrigerator body should be at least 30cm away from the wall and not
Barriers are allowed to block the air inlet and the air outlet.
·In case of machine fault or power failure, the temperature in the Laboratory refrigerator will rise. If
the machine cannot be repaired in short period, please take out the stored goods and transfer them
to other place where it accords with the storing temperature to avoid damages.
1
Machine Translated by Google
Content
2
Machine Translated by Google
3
Machine Translated by Google
Notes: please note that do not let the power line be damaged by trundles when pushing the
Laboratory refrigerator.
Notes: be sure to take off the packaging pedestal on the bottom of the Laboratory
refrigerator.
Notes: Do not put goods into the Laboratory refrigerator which is just plugged in. Let the
empty body run for a while (about 12 hours) and then put the goods to be refrigerated into the
refrigerator.
Normal operating condition of the equipment:
to) Environmental temperature: 10ÿÿ32ÿÿ
b) Relative humidity: ÿ80ÿÿ
c) There is no strong sharp pounding and corrosive gas around;
d) There are no effects of direct radiation from sunshine and other cold and heat sources.
Working system of the Laboratory refrigerator: intermittent running
Safety precautions
·Supply voltage: the equipment needs 220V/50Hz alternating current power supply. If the service
voltage is lower than 187V or higher than 242V, it is necessary to add a proper automatic voltage
regulator to be used cooperatively;
·When the Laboratory refrigerator is used, the power supply is required to be equipped with a lower
voltage air circuit breaker and a leakage protection device; ·It is
necessary to use a dedicated independent socket which is grounded reliably. The length of the
power line cannot be lengthened at randomly. If it really needs to be lengthened, be sure to use a
cooper core conductor which is larger than 2.5mm2. And the cross area of the copper core
4
Machine Translated by Google
conductor which is in the wall and connected with the power socket must be above 4mm2ÿ
·Flammable and explosive dangerous goods and goods of acid and alkali etc. with strong
corrosiveness are not allowed to be put into the Laboratory refrigerator;
·Keys should be kept properly to avoid accidents that will happen if children get the key to open the
door;
·The zero line (N end) of the socket cannot be connected with the ground lead (B end). Otherwise, the
housing of the Laboratory refrigerator may be electrified and electric shock accidents may be
caused;
·The power line cannot be bundled up, pressed under weight and next to heat sources of compressor,
etc
3. Instructions
5
Machine Translated by Google
temperature
Notes: If there is indeed parameter change requirement, please contact the after-sale and distributors
of BIOBASE for setting.
Notes: When working normally, simultaneously press the down-regulation key and Defrost key for 5
seconds, enter the page of setting time parameter value, the default page displays t (min).The
corresponding parameters of H (hour), D (day), M (month), and Y (year) were displayed
respectively after press the up-regulation key.Select the parameters that needed to be adjusted, press
the set key and the digital tube flickers, adjust the parameters by pressing the up-regulation key or
down-regulation key, then press the set key to save the current parameter value. In the process, if
the down-regulation key and Defrost key are pressed at the same time, the temperature display
interface is restored.
3.2.3 Display error
When faults of the refrigerator temperature display probe including short circuit, open circuit, malfunction and incorrect
connection appear, “E0” is displayed. When faults of the refrigerator
6
Machine Translated by Google
temperature control probe including short circuit, open circuit, malfunction and incorrect connection
appear, “E1” is displayed. At the same time, the buzzer of the temperature controller will give out
alarm and the display screen displays corresponding alarm parameters (E0ÿE1).
After the power is switched off, the control function of the controller fails and the back-up source is
used for displaying. If the back-up source is charged for more than 10 hours, the power failure
symbol “Pd”, the current inside temperature and the alarm display alternatively for one time and
then display again after 30 seconds' stop for cycle operation. After 8 hours, the nixie tube stops
displaying. The high temperature alarm mark light will twinkle continuously until the electric
quantity of the back-up source is used up. The inside temperature can be continuously displayed
during the power failure. Please pay attention to this and take relevant measures to avoid damages.
Suggest: If you judge that the temperature sensing probe goes wrong, you should cut off the power
and first check whether the connecting wire of the temperature sensing probe is loose. Please
contact professional after-sale personnel or distributor of BIOBASE for processing.
Tips: if internal code appears when user regulates the temperature controller, the user should
wait for 1 minute and then operate it until the inside temperature is displayed to avoid fault of
the Laboratory refrigerator caused by the change of the control parameter.
When the Laboratory refrigerator stalls temporarily due to power failure or other faults, the inside
Temperature of the Laboratory refrigerator will rise from 5ÿto 15ÿ within 1 hour. If power cannot
be supplied in short time, the user should consider taking out the stored goods and putting them into
other normal Laboratory refrigerators for storing to avoid losses caused by the damages of the
stored goods.
Before putting goods into the Laboratory refrigerator to be stored, the user should confirm in
advance that whether inside temperature of the Laboratory refrigerator is in accordance with the set
temperature for storing to avoid losses caused by goods damages due to the improper temperature.
Due to the refrigeration inertia, the Laboratory refrigerator cannot be kept at constant temperature.
And there is certain up and down difference between the inside temperature and the set temperature.
The inside temperature varies with the change of the service environment and the set temperature. Item
belongs to normal phenomenon.
7
Machine Translated by Google
4. Routine Maintenance
In routine maintenance, in order to prevent electric shock or personnel injuries, be sure to cut off the
power before repairing or maintaining the equipment and do not inhale drugs or particulate matters
surrounding the equipment when maintaining the equipment. It is necessary to dry wet gloves to
protect your hands. Otherwise, your hands may be cut by edges or corners of the refrigerator body.
·The Laboratory refrigerator should be cleaned once every month. Regular cleaning may keep the
appearance of the incubator new;
·Use dry cloth to wipe off a small amount of dust on the housing, the inner chamber and all
accessories of the Laboratory refrigerator. If the Laboratory refrigerator is very dirty, it is advised to
use neutral detergent for cleaning;
·After cleaning, use cloth which has been soaked in clean water to wipe off the detergent;
·Do not pour water on the housing or in the preserve room of the Laboratory refrigerator. Otherwise,
the electrical insulation may be damaged, thereby causing faults.
·During the rainy season, vapors may be easily condensed on the surface of the glass door of the
refrigerator. In serious condition, water may drip. Please use dry cloth to wipe it dry in due time.
Normal use will not be affected.
4.2 Defrost
The Laboratory refrigerator belongs to air cooling frostless refrigerator which needs no defrosting.
In summer with high environmental humidity, condensation may appear on the buccal frame of the
Laboratory refrigerator. This is normal. It is advised to wipe it dry with dry cloth.
·No heavy things are allowed to be placed on the door or the header of the Laboratory refrigerator to
avoid deformation of the door body or the header caused by the press.
·The Laboratory refrigerator should be cleaned and maintained every once in a while.
·Use warm wet soft cloth to wipe the inner and outer surfaces of the Laboratory refrigerator.
·When serious dirt appears, use neutral detergent for washing tableware to wipe and then use
cleaned soft cloth to wipe out the water spots.
·Once starting the Laboratory refrigerator, you'd better keep it running continuously.
Notes: be sure not to sprinkle water directly on the Laboratory refrigerator to avoid
performance reduction and metal rust of electrical components. Be sure not to clean the
Laboratory refrigerator with hot water and corrosive detergent or organic solvent. Also do
do not clean the Laboratory refrigerator with scrubbing brush and wire brush. Children are not
allowed to play games with the Laboratory refrigerator.
8
Machine Translated by Google
9
Machine Translated by Google
6.2 Wiring
Diagram
10
Machine Translated by Google
Packing list
User Instruction 1 1
Key 4 4
Shelf 10 10
Snap 40 40
Fuse (10A) 2 2
Power Cord 1 1
11
Machine Translated by Google
7. Warranty
ÿ If the instrument and equipment during the warranty period fails or damages due to improper
use by the user, our company shall not bear the warranty obligation.
ÿ After the warranty period, our company is also responsible for the maintenance, but the
corresponding maintenance fee will be charged.
ÿ For the production date, please refer to the product label. ÿ
We can provide the drawings and some necessary technical data for the maintenance units and
maintenance personnel trained and approved by our company.
1. From the date of purchase, the user can enjoy one-year free warranty for the whole machine and
three-year warranty for the compressor (man-made damage is not within the scope of warranty), and
the purchase time is subject to the valid purchase documents.
2. The following situations are not within the scope of warranty, only maintenance services are
provided, but corresponding cost are charged.
(1) All damages caused by human factors, including damages caused by using in abnormal working
environment, not following the user manual and improper operation, etc.
(2) If the user disassembles, repairs or refits the machine without permission, or the device is not
repaired by the authorized unit of our company, or there is no warranty card or valid document.
(3) Damage caused by improper transportation or other irresistible factors (such as flood, fire, lightning,
earthquake, etc.) after purchase.
12
Machine Translated by Google
ADDRESS: OLABO Intelligent Manufacturing Industrial Park, Ancheng Town, Pingyin County,
Tel: +86-531-81307661
Inquiry: [email protected]
Complaints: [email protected]