The thermal flow measurement principle bases on the research of Louis Vessot King. He also developed
the first hot-wire anemometer which is something like the elder of modern thermal flowmeters.
Measuring principle
Two sensors are installed in the gas stream. One sensor monitors the actual process temperature whilst
the second sensor is heated to a defined differential temperature. Even with no flow, the gas extracts
some energy from the heated sensor due to convection. As the gas flow increases, more energy is
drawn from the heated sensor. To maintain the differential temperature, the power to the heated
sensor must be increased. The electrical power needed to maintain the differential temperature of the
two sensors is in relation to the mass flow in the pipe. The relation is not linear, therefore a thermal
mass flow meter must be calibrated to define the characteristics of the unit. The sensitivity of the
principle is best a low flow. At high flow rates less and less increase of heating power is needed to
compensate an increased mass flow. The resolution of the flow signal is therefore reduced.
1/7
Gas groups
This group of gases can reliably be Due to the properties of this group of These gases should not be
measured. Basic knowledge of the gases the application might be attempted without approval
application is needed to select and compromised. Changing gas of an experienced expert.
size the meter. composition of Biogas, condensing Sometime thermal mass
Butane, or the very high specific heat flowmeters are the only
Example: capacity of Helium can lead to technical solution to measure
• Air unacceptable results or even failure of these gases however. All
• Oxygen the measurement. Expert knowledge these gases bear some
• Nitrogen and detailed information of the additional risks that can lead
• Carbon Dioxide process conditions and the customer to corrosion, unacceptable
• Argon expectations are needed to select the errors, zeropoint shifts and so
• Methane best solution. on.
Example: Example:
• Biogas • Ammonia
• Natural Gas • Chlorine
• Hydrogen • Always check all other
• Helium gases…
• Butane
• Propane
2/7
Basic meter construction
Transmitter
Amplifies and filters signals, computes relevant data (totalizer, flowrate), generates interface signals
(Hart, Profibus etc). Is the human machine interface (HMI) and display. It can be mounted integrally or
remote, depending on the installation situation. Is available in aluminum (non-Ex and Ex versions).
Sensor
Cointains all parts and functions of the device mounted in the pipe and connected to the process.
Process connection
The obstacle prevents the green heater “decider” from being cooled as
much as the red heater if the flow goes reverse. At forward flow both
heaters will be cooled with the same rate.
If the flow goes reverse, the obstacle prevents the green heater from being cooled as much as the red
one. The job of this green heater “decider” is to detect if the flow goes in forward or backward direction.
3/7
Heater (in every sensor form)
This temperature sensors measures the actual temperature of the gas as a reference.
Advantages
4/7
Limitations
Process influences:
• Pulsating flow can result in an unstable measurement signal. In some cases, pulsating flow may
lead to reverse flow, which for some measurement technologies either remains undetected or is
regarded as positive (forwards) flow.
• Moisture/wet gas can create the impression of increased flow rate with thermal mass meters. Due
to its high thermal conductivity, water strips the heated sensor of additional heat energy, which
appears as increased flow in the device. This problem can be alleviated by installing the devices at a
radial angle of 135° to ensure that condensation flows to the base of the sensor and away from the
sensor tips.
• Dirt/coating on the sensors of a thermal mass meter can also have an influence on the
measurement. Dirt can function as an insulation to the sensors restricting heat flow from the heated
sensor to the measurement fluid. This can give the impression of lower flow.