Software-software upgrade
2
Preparation
Kernel upgrade:
If the system is abnormal or the operating system is significantly changed, Kernel upgrade(format the
hard disk) is required.
Simple upgrade:
If the version of application software is updated and the accompanying software is not significantly
changed, you can perform a simple upgrade, which is equivalent to patching the original software.
Attention :
1. Please back up the required patient data before upgrading.
2. After downloading the software, check the file size to avoid upgrade exceptions caused by
incomplete files
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
3
Preparation
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
4
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
5
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
6
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
7
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
8
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
9
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
10
TP upgrade
1. Ensure that the “TPUpdate” file is copied to the root directory of U disk
2. Insert U disk into the dedicated USB port on the right of the touch screen
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
12
TP upgrade
3. After the machine is powered on, hold down the key of soft keyboard on KBD to
enter the touch screen upgrade interface.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
13
TP upgrade
4. Click【 system update 】, then the upgrading will take several minutes. After the upgrade is
complete, the screen will display“Update Finished, Restart! ”,which means that the touch screen
upgrade is finished.
5. Unplug the U disk from touch panel and restart the machine .
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
14
Kernel upgrade
1. Insert the safe mode U disk and external keyboard into the machine's USB port
2. Pressing the power button, when the character interface appears, press the “delete” key of the external keyboard,
the system will prompt you to enter the password: "service", and then press "Enter" to enter the bios setting
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
15
Kernel upgrade
3. Set boot device as below steps:
Choose the Boot → Hard Disk Drives → 1st drive → USB drive
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
16
Kernel upgrade
4. Boot Device Priority → 1st boot device →USB drive.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
17
Kernel upgrade
5. Exit->Save Changes and Exit->OK
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
18
Kernel upgrade
6. In the system BIOS mode:
Press “up”, “down”, “left” and “right” key can choose the option;
Press “Enter” can enter the option;
Press “Freeze” key can exit the option.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
19
Kernel upgrade
7. After a few minutes, the following screen will appear. Use the USB keyboard to enter
Forcekill command and press Enter.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
20
Kernel upgrade
8. When the screen show us below message, press Enter key to shut down the system.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
21
Kernel upgrade
9. Unplug the U disk, boot up the system, and then software will be installed automatically.
When the system reboot, the screen will show a blue screen like left
image, this means software is installing, and when it is finished,
system will reboot by itself, during this time, we only need waiting.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
22
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
23
11. At the Exam screen, press Shift, E and Menu keys one by one. Enter Password(monthly password)
and press enter key. Press SET key to start the update.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
25
Dongle version
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
26
Simple upgrade
1. Copy the right Adata and Asecure files into the root directory of any empty USB drive.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
27
Simple upgrade
2. Boot up the system, and wait until the EXAM screen appears, then Insert the U disk.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
28
Simple upgrade
3. At the Exam screen, press Shift, E and Menu keys one by one. Enter Password(yearly password) and
press enter key. Press SET key to start the update.
4. Press Exam key and back to the Exam screen, remove the USB drive.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
29
Simple upgrade
5. Shut down the system and reboot it.
When the system reboot, the screen will show a blue screen
like left image, this means software is installing, and when
it is finished, system will reboot again by itself, during this
time, we only need waiting.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
30
Simple upgrade
6. After booting the machine, inserting the U disk. At the Exam screen, press Shift, E and Menu one by
one, then enter Password(yearly password) and press Enter. Press SET key to start the update.
(upgrade aconfig)
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
31
Simple upgrade
If the system have the u-scan option, we need upgrade the keypass , if not, this step can be cancelled.
7. At the Exam screen, pressing Shift, E and Menu keys one by one. Enter Password(monthly
password) and press enter key. Press SET key to start the update. (upgrade keypass)
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software-software upgrade
32
Simple upgrade
8. Enter the engineering mode with Monthly code. Press “i” directly, then the u-scan function will be
active. Reboot the machine.(Active the u-scan function)