UTZ-PDF (7) (1)
UTZ-PDF (7) (1)
What is Ultrasound?
BASIC CONCEPTS OF
Ultrasound is the name given to high-frequency sound
ULTRASOUND waves over 20,000 cycles per second (20kHz). These waves,
inaudible to humans can be transmitted in beams and are
used to scan tissues of the body.
By:Associate Prof. Myna Marie Nerona RRT, MSRT Ultrasound’s frequency ranges from 2 to 10 MHz.
Dean The UTZ’s duration of the pulse is about 1 microsecond
University of Makati and the pulses are repeated about 1000 times per second.
1 2 3
CATEGORIES OF
SOUND
ULTRASOUND
1) ULTRASOUND High frequency sound waves AUDIBLE SOUND
2) AUDIBLE SOUND Above 20,000 cycles per second (20 kHz) Sound is a form of energy which causes a mechanical disturbance
in the form of vibration of molecules within a medium
3) INFRASOUND Inaudible to humans Sound is a mechanical wave that travels in a medium in a
Used to scan tissues of the body LONGITUDINAL WAVE
Ultrasound Pulse: 1-10 MHz Sound requires a medium containing molecules, and
Pulse Duration: 1 microsecond therefore cannot travel through a vacuum
Pulse Repetition: 1000 times/second sound requires a vibrating object
16 Hz to 20,000 Hz
4 5 6
INFRASOUND When a sound travels through a medium: TYPES OF SOUND WAVES
Below 16 Hz
The molecules of that medium are alternately 1. LONGITUDINAL OR COMPRESSION WAVES
compressed (squeezed) and rarefied (stretch).
7 8 9
1
9/9/23
10 11 12
FREQUENCY APPLICATION
FREQUENCY
MEDIUM SPEED IN
m/sec ACOUSTIC VARIABLES
1. Period
INFRA BELOW 16 Hz EARTHQUAKES
AUDIBLE BETWEEN AIR (at 32º F) 331
SOUND 2. Wavelength
SOUND 16 Hz TO 20,000 Hz
OR 3. Amplitude
16 Hz TO 20 kHz
4. Frequency
5. Velocity
WATER (SEA) (at 1531 6. Power
77º F)
7. Intensity
SOLID (STEEL) 5200
13 14 15
ACOUSTIC
VARIABLES
Period (T): the time taken for one complete cycle to 1. PERIOD (us) PERIOD
PERIOD FREQUENCY
& FREQUENCY
occur (s or µs)
Wavelength (λ): length of space over which one
cycle occurs (m or mm)
Amplitude (Depth): the maximum displacement PERIOD FREQUENCY
that occurs in an acoustic variable
Frequency: cycle per second (Hz) Period (T)
Velocity: frequency times wavelength The TIME taken for one complete cycle to occur
)
16 17 18
2
9/9/23
PERIOD & FREQUENCY 3. AMPLITUDE (Pa)
19 20 21
CHAPTER 1
3. AMPLITUDE (Pa) 3. AMPLITUDE (Pa) 3. AMPLITUDE (Pa)
22 23 24
4. FREQUENCY (Hz) 4. FREQUENCY (Hz) 4. FREQUENCY (Hz)
25 26 27
3
9/9/23
PRP PRF
Pulse repetition frequency (PRF)
Pulse Repetition Period (PRP) PRP decreases as PRF increases.
PRP PRF
Number of pulses occurring in 1 s. Time from the beginning of one pulse to
More pulses occur in a second, less
Usually expressed in kHz. the beginning of the next.
time from one to the next.
Pulse/ sec Usually expressed in microseconds (µs).
28 29 30
Pulse duration (ms)
PRP = 1 PRP (µs) = 1
PRF PRF (kHz)
TIME it takes for one pulse to occur
Period = 1 Period (s) = 1
Frequency Frequency (Hz)
.
PD= Period x # of cycles in the pulse.
.
31 32 33
34 35 36
4
9/9/23
M5. VELOCITY (m/sec) 5. VELOCITY (m/sec) 6. POWER (Watts)
E The faster the Utz Is the rate of energy transferred
D The velocity of a sound wave in a
I
propagation medium, is related to its wavelength and through the sound wave.
Density Speed
U frequency
M Hardness v Power is proportional to the
Compressibility amplitude squared of a sound
Stiffness
f λ 5 MHz ultrasound beam has a wave
wavelength of 0.308 mm in P = A²
soft tissue with a velocity of
1540 m/sec.
37 38 39
40 41 42
43 44 45
5
9/9/23
1915: PAUL LANGEVIN 1942: KARL DUSSIK 1948: GEORGE LUDWIG M.D.
After Titanic Sinking (1912) Invents first physician to use ultrasound for medical first described the use of ultrasound to
Hydrophone (1st Transducer) to detect diagnosis (brain tumors) diagnose gallstones
Icebergs and Submarines during WW1
46 47 48
1953: INGE EDLER AND HELLMUTH 1958: IAN DONALD 1950s:DOUGLASS HOWRY & JOSEPH
HERTZ HOLMES
Pioneer of OB-GYN Ultrasound
Performed the first successful echocardiogram
pioneer 2D B-mode Ultrasound
49 50 51
Early 1950s Late 1950sà the first contact compound 1966: DON BAKER, DENNIS
Because air does not transmitted B-scanner (using olive oil as a lubricant) WATKINS, AND JOHN REID
ultrasound waves efficiently, the air was developed. This equipment used an
interface between the transducer and the articulating arm to produce static Pulsed Doppler ultrasound technology
patient’s skin was initially a problem. images.
52 53 54
6
9/9/23
55 56 57
2000s - Present
The iPhone now has a
telesonography app. BIOLOGIC EFFECTS OF a. Thermal Index (TI)
ULTRASOUND???? Is a predictor of max. temperature increases under
clinically relevant conditions.
Heat Index
In general UTZ is considered to be safe image Absorption of sound waves may cause heating in
modality.. tissue
TI = Below 1.0 (1 degree celsius)
Ø TIB (bone)
a. Thermal Index (TI) Ø TIC (cranial)
b. Mechanical Index (MI) Ø TIS (soft tissue)
58 59 60
THE PIEZOELECTRIC
b. Mechanical Index (MI) EFFECT
The ISUOG and WFUMB, 2009..
• Gives an estimation of cavitation
effect of UTZ Disapproved the use of UTZ for the sole
purpose of providing souvenir images of
MI = Below 1.0 (1 degree celsius) the fetus and also for entertainment
purposes..
61 62 63
7
9/9/23
B. PIEZOELECTRIC
MATERIALS
A. PIEZOELECTRIC EFFECT B. PIEZOELECTRIC MATERIALS/
ELECTROCERAMIC NATURAL SYNTHETIC
PIEZOà is a Greek word “PIEZIN” 1. Quartz 1. Lead Zirconate Titanate (PZT)
5. Lithium Sulphate
64 65 66
Normally these dipolar molecules If the materials are heated above
have a random arrangement
the Curie temperature in
•Piezoelectric materials are
crystalline materials composed of presence of an electric field, the
they are unable to align themselves
dipolar molecules without an applied electric field.
molecules align themselves with
that field.
•Positive at one end and negative
at the other.
67 68 69
70 71 72
8
9/9/23
ACOUSTIC
IMPEDANCE A. ACOUSTIC IMPEDANCE (Z) A substance which has densely
packed molecules will have a high
Acoustic impedance. Ex. Bone
The resistance offered by tissues to
the movement of particles caused by
the ultrasound waves. A substance which has loosely packed
particles will have a low acoustic
Ex: Bone tissue Vs. Soft Tissue
impedance. Ex. Air
73 74 75
76 77 78
CHAPTER III
B. ACOUSTIC IMPEDANCE C. ACOUSTIC IMPEDANCE C. ACOUSTIC IMPEDANCE MISMATCH
MISMATCH AND REFLECTION AND REFLECTION
MISMATCH
Substances with the same acoustic
impedanceà
The greater the impedance mismatch,
The difference in acoustic the greater the percentage of energy Have a total transmission of energy and
impedance between two that will be reflected at the interface or therefore no reflection.
substances is known as the boundary between one medium and ZR= 0% ; ZT=100%
acoustic impedance mismatch. another.
79 80 81
9
9/9/23
Substances with a small difference in Substances with a large difference If there is a large acoustic mismatch
acoustic impedanceà in acoustic impedanceà Ex: between (bone and muscle) ; (air and soft
tissue)
Large amount of reflected energy
Small amount of energy is reflected and a small amount of transmitted
but the majority is transmitted. energy.
large propor-
produces a do not have
bright image very little
tion of energy is enough energy
energy is
on the to produce
ZR= 99% ; ZT=1%
reflected.
display. transmit-ted im age.
ZR= 5% ; ZT =95%
82 83 84
MISMATCH REFLECTION TRANSMISSION Therefore, it is not practical to use
ultrasound to produce images of soft
Ultrasound Beam
tissue subjects containing GAS OR
BONE.
MISMATCH REFLECTION TRANSMISSION
Ultrasound imaging is very good at
discriminating between substances with
small differences in acoustic impedance
such as soft tissues.
85 86 87
88 89 90
10
9/9/23
CHAPTER IV
II. SIDE LOBES
These echoes will therefore be In addition, they cause a
incorrectly placed on the image.
degradation of lateral resolution
The extraneous beams of the
due to the effective widening of
ultrasound generated from the 15% à side lobes
the beam in the scan plane.
edges of individual transducer.
Grating lobes
91 92 93
III. BEAM WIDTH Ultrasound Artifacts
The beam width affects the
LATERAL RESOLUTION
Beam width refers to the
dimension of the beam in the scan
plane. the narrower the beam width, the
better the Lateral resolution.
94 95 96
1.
Different types of UTZ artifacts
Reverberation
REVERBERATION
2. Acoustic shadowing
3.
4.
Acoustic enhancement
Edge shadowing
This is the production of spurious (false) echoes
5. Beam width artifact due to repeated reflections between two interfaces
6. Slice thickness artifact with a high acoustic impedance mismatch.
An ultrasound artifact is a 7. Side lobe artifacts
Comet tail
structure in an image which 8. Mirror image
does not directly correlate with 9. Double image
10. Equipment-generated artifacts
actual tissue being scanned.
97 98 99
11
9/9/23
This artifact can be differentiated from the real echoes
Artifacts appearing as parallel, equally This artifact will often be seen: due to the lack of breathing movement occurring.
spaced lines is called reverberation
To help to eliminate this artifact it may be useful to:
Reverberation artifacts are a result of the 1. At the skin-transducer interface 1. Increase the amount of gel used
presence of two or more strong reflecting 2. Gas surface and the transducer 2. Used a stand-off pad (gel pad or a bag of saline)
3. Reduce the gain
surfaces 4. Move the position of the transducer.
Shotgun wound
ACOUSTIC
SHADOWING
This appears as an area of low amplitude echoes (hypoechoic or
If the shadowing contains reverberation
anechoic) behind an area of strongly attenuating tissue. Acoustic shadowing will occur at the (ringdown echoes), it is likely to be gas.
It is cause by severe attenuation of the beam at interference, resulting
in very little sound being transmitted beyond. interface with a large acoustic
The attenuation can be due to either absorption or reflection of the
sound waves or a combination of two.
mismatch such as:
If the shadowing is clear shadowing,
1. Soft tissue and gas absent of any ringdown, it is likely to be
2. Soft tissue and bone or calculus
a calculus
Calcified mass
ACOUSTIC
ENHANCEMENT
This artifact appears as a localized area of Can commonly be seen distal to fluid filled
increased echo amplitude behind an area of structures such as the urinary bladder, the Acoustic enhancement is
low attenuation.
gallbladder or a cyst. caused by weakly attenuating
structures.
This artifact can often be a useful diagnostic
aid, particularly when scanning a soft-tissue
mass or cyst containing low-level echoes. Fluid filled mass
12
9/9/23
BEAM WIDTH
EDGE SHADOWING
ARTIFACT
A combination of refraction and reflection
occurring at the edges of rounded structures (and All echoes returning to the transducer will
when the speed of sound is different from that in
surrounding tissue) will result in an edge
have arisen from across the full width of the
beam, which can vary by several millimetres.
Correct positioning of the focal
shadowing artifact.
zone will help to reduce this
artifact.
SLICE THICKNESS
ARTIFACT
of the beam, and are
These occur due to the thickness
similar to beam width artifacts.
However, they occur at 90 degree to the scan plane These artifacts will typically be seen
with echoes from interfaces in front of and behind in transverse views of the urinary
the assumed plane of origin appearing in the
bladder This artifact is a result of inherent
display. characteristics of the transducer, and
apart from trying a different
Beam thickness artifact is primarily transducer, cannot be eliminated.
dependent upon beam angulation
13
9/9/23
DOUBLE IMAGE
ARTIFACT
An area close to a specular reflector will This artifact is caused by refraction of
be imaged twice, once by the original the beam and may occur in areas such
ultrasound beam and once by the beam as the rectus abdominis muscle on the In the transverse plane the edges of the
after it has reflected off the specular anterior abdominal wall. muscle act as a lens and cause the
reflector. ultrasound beam to be refracted and
this causes a single structure to be
fluid-air interface. interrogated by the two separate
Diaphragm refracted beams.
If too mach gain is applied then the
10. EQUIPMENT-GENERATED electronic noise, inherent in all systems,
In order to establish whether these will also be amplified.
echoes are genuine, it is necessary to ARTIFACTS
move the transducer slightly to one
side to avoid the junction of the rectus
If too little gain is applied, this can lead to
abdominis muscles. Incorrect use of the equipment controls loss of relevant information, and incorrect
can lead to artifacts appearing. diagnosis may occur.
ACOUSTIC
Velocity of sound
VARIABLES
Period (T): the time taken for one complete cycle to The velocity of sound in a medium is determined by Cycle is composed of one compression and one
occur (s or µs) the density and elastic properties of the medium. rarefaction, or a complete positive and negative
Wavelength (λ): length of space over which one It differs greatly among air, bone and soft tissue change in an acoustic variable
cycle occurs (m or mm) But varies only slight form one tissue to another
Amplitude (Depth): the maximum displacement
that occurs in an acoustic variable
Frequency: cycle per second (Hz)
Velocity: frequency times wavelength
14
9/9/23
Frequency is the number of cycles per second. Period is the time it takes for 1 cycle to occur; the Wavelength (λ) is the distance the wave must travel
It also describes how many times the acoustic variable inverse of frequency in 1 cycle.
changes in one second. Units of period: seconds (s)and microseconds (µs) It determined by both the source of the wave and the
Cycles of compressions and rarefactions in a sound medium in which it is propagating
waves Units of wavelength: meters (m) and millimeters (mm)
Period= 1/frequency
Units of frequency: hertz (Hz) or megahertz (MHz)
It is inversely proportional to the frequency.
Frequency =
Wavelength= propagation speed /frequency
Propagation Speed is the maximum speed with
which an acoustic wave can move through a
Propagation speed is greater than in solids>liquid>gases. Example:
medium, determined by the density and If the frequency of an UTZ wave traveling through
stiffness of the medium. Propagation speed (c) is equal to frequency (f) times
soft tissue is increased from 5 to 10 MHz, what
wavelength (λ) happens to the wavelength?
It increases proportionally with the stiffness Because the propagation speed is constant for a given
Density is the concentration of mass per unit medium, if the frequency increases, the wavelength will
volume, the propagation speed is inversely decrease. Conversely, if the frequency decreases, the
proportional to density wavelength will increase.
Units: meters/seconds (m/s),
millimeters/microseconds (mm/µs)
15
9/9/23
Pulse Duration
Duty Factor Spatial Pulse Length
Is the fraction of time that the transducer is generating a
Is the time it takes for a pulse to occur; the period of pulse. Is the distance over which a pulse occurs.
the ultrasound in the pulse multiplied by the Units: millimeters (mm)
number of cycles in the pulse.
Maximum value : 1.0 in continuous waves, the transducer SPL= wavelength x number of cycles in a pulse (n)
Units: seconds (s) milliseconds (ms) is always generating a pulse. A second transducer acts as
Pulse Duration= number of cycles (n) x period (p) the listening device
Minimum value 0.0. the transducer is not being excited. In
clinical imaging, using pulse echo system is the duty factor
from 0.001 to 0.01
Units: Unitless
DIFFERENT MODES
Amplitude OF ULTRASOUND
OPERATIONAL MODES
Is the maximum variation that occurs in an acoustic A-MODE 1. Static imaging modeà A-mode (useful for
variable. B-MODE measuring midline shifts of the brain) , B-mode
it indicates the strength of the sound waves REAL-TIME (used primarily for abdominal imaging)
To arrive at this variation, the undistributed value is M-MODE 2. Dynamic Imaging modesà M-mode (application in
subtracted from the maximum value, and the unit DOPPLER ULTRASOUND dynamic imaging of internal structures), Real-time
for the acoustic variable is applied. (allows for observation of structures in motion)
3. Ranging modeà Doppler mode (used for depth and
flow measurements and investigations of moving
surfaces.)
A mode scan
1.) A-MODE display (Amplitude Mode)
Modes of Display Echoes are shows as peaks
One-dimensional graphic display with vertical deflections of the baseline.
Y-axis (height of the deflection) represents the amplitude/ strength of the
echo
X-axis represents the distance in time is a function of where on the
horizontal baseline the deflection occurs.
Distance between various structures can be measured
Used to build two-dimensional B-mode image
16
9/9/23
B-Mode Scan
2.) B-MODE display (Brightness mode)
This type of image shows all the tissue traversed by
the ultrasound scan.
Displays the echoes as variations in the brightness of
üThe main purpose for employing the line spots on the image.
The position of the spot on the baseline is related to
A-mode is to measure the depth of the depth of the reflecting structure: the brightness is
proportional to the strength of the echo.
interferences and to detect their
separation accurately. Two-dimensional images in which the echo amplitude
is depicted as dots of different brightness.
3. ) REAL-TIME
From the M-mode display the x-axis is a depth axis;
This mode display motion by showing the the y-axis is a time axis that can be used to
images of the part of the body under the
transducer as it is being scanned, the images
measure the time between successive heartbeat..
4.) M-MODE display (Motion-mode)
changes with each movement of the Shows movement as a function of time
transducer or id any part of the body is Used in cardiac scanning
moving. Another way of displaying motion
Examples: moving fetus or pulsating artery. Result is a wavy line.
The movement id shown on the monitor in ü If the A-mode image contains moving interfaces
real time as it occurs. In most real time units, it and is converted to a B-mode display so that the
is impossible to freeze the displayed image, vertical axis is time driven, M-mode display
results.
holding it stationary so that it can be studies
and measured if necessary.
M-Mode Scan
Sometimes called… 4.) Doppler Ultrasound
1. TM mode (time-motion mode); Demonstrates and measures blood flow
2. PM- mode (position-motion
mode) and
3. UCH (ultrasonic cardiography) Christian Johann Doppler noted in
1842 that the wavelength of light varied
since its principal application is to according to the relative motion of the source
monitor the heart.
of the light and the observer.
17
9/9/23
Stars moving away from the The Doppler shift frequency
earth produce light that appears is in the audible range, and
shifted to the red (longer therefore an audio amplifier and
wavelength). Stars moving speaker are all that is necessary
toward the earth appear more to listen to the fetal heart, called
blue. utz stethoscope.
2.) Pulsed Wave Doppler Unit 3.) Colour Doppler Unit 4.)Duplex Doppler System
Ultrasound is transmitted in pulses The distribution and direction of the flowing blood Combination of a B-mode and Doppler system
With good depth resolution are shown as a two dimensional image in which the A blood vessel is located by B-Mode ultrasound
Measures the speed of the blood in a particular vessel velocities are distinguished by different colours. imaging and then the blood flow is measured by the
Cannot measure high blood velocities in deep vessels Doppler Ultrasound.
High velocities may be wrongly displayed as low Allows the Doppler beam to be directed accurately at
velocities any particular blood vessels
18
9/9/23
WAVE
PROPAGATION WAVELENGTH
The transmission and spread of ultrasound waves to WAVELENGTH
different tissues The length of a single cycle of the ultrasound wave
Average Propagation for Soft Tissues: 1540 m/s Inversely proportional to the frequency
Average Propagation for Soft Tissues: 4620 m/s Determines the resolution of the scanner
Higher the frequency, the shorter the wavelength
The shorter the wavelength, the better resolution,
giving clearer image and more details on the screen
To improve resolution
Adjustment of the ultrasound beam
ultrasound in different ways.
Higher frequencies are more readily
Done by the time-gain-compensation (TGC) amplifier
May be electronic or by a lens attached to the absorbed and scattered that lower To compensate for ultrasound attenuation in any part of
transducer frequencies the body
To reach deeper tissues, it is necessary To improve the quality of the final image
to use lower frequencies because the
waves are less likely to be diverted as
they traverse intervening structures.
BOUNDARIES
The line at the periphery of two tissues which propagate
The greater the difference between the characteristic
ultrasound differently
acoustic impedances, the greater the fraction which
The zone of echoes at the interface is reflected.
The greater the ratio of the propagation speeds, the
greater the fraction which is refraction.
A fraction of the incident wave is reflected at an angle equal
to the angle of incidence.
Another fraction passes across the interface and is refracted,
continuing at an angle that is different from the angle of
incidence.
19
9/9/23
If reflecting boundaries is much wider than the wavelength, it
Part II Relevant
acts like a mirror and is called specular reflector.
Examples: Fetal skull, the diaphragm, walls of the vessels and
Terminologies PIEZOELECTRIC EFFECT
connective tissues
Ultrasound waves scatter when the width of the reflectors is
Piezein – “press or pressure”
Ability of a material to generate an electrical charge unresponse
smaller than the wavelength. Organs are called scattering to applied pressure
media
Examples: Liver and Kidney parenchyma
PIEZOELECTRIC
MATERIALS
Crystalline materials composed of dipolar molecules
PIEZOELECTRIC CRYSTALS
Generate ultrasound waves
Quartz – naturally occurring crystals Capable of changing electrical signals into mechanical Ceramic crystals that deform and vibrate when they
Lead zirconate titanate – man made ceramic (ultrasound) waves are electronically simulated generates the sound
Natural Materials: pulse used for diagnostic sonography.
PIEZOELECTRIC CRYSTAL AS TRANSMITTER OF SOUND
Quartz Converting electrical energy into mechanical energy (sound) Each pulse consists of a band of frequencies referred
Tourmaline
to as bandwidth.
Rochelle Salt PIEZOELECTRIC CRYSTAL AS TRANSMITTER OF SOUND
Synthetic Materials: Converting mechanical energy (sound) into electrical energy
Lead zirconate titanate (PZT)
Barium titanate NOTE:
Lead metaniobate Small crystal diameter
Ammonium dihydrogen phosphate Increased beam divergence
Lithium sulphate Larger crystal diameter
Decreased beam divergence
ACOUSTIC
IMPEDANCE Acoustic Impedance
PIEZOELECTRIC ELEMENTS ACOUSTIC IMPEDANCE Ultrasound wave is similar to a light beam in that it
Crystalline minerals that generate voltages when Property of a substance may be :
subjected to a mechanical force Describes how the particles of that substance behave Focused
when subjected to pressure wave
Piezein – “to press or squeeze” Refracted
High density substance – high acoustic impedance
Piezoelectric Effect – discovered by Jacques and Reflected
Low density substance – low acoustic impedance
Pierre Curie Formula: Z=pc Scattered
Thinner Piezoelectric Materials p = density of material (kg/m3) At interfaces between different media
c = speed of sound (m/s)
Higher resonant frequencies Z = acoustic impedance (rayls)
20
9/9/23
REFLECTION SCATTERING REFRACTION
Occurs when two large structure of significantly Occurs when an ultrasound wave strikes a boundary Occurs when the beam encounters an interface
different acoustic impedance form an interface or interface between two small structures between two different tissues at an oblique angle
Occurs when a sound wave strikes an object that is Occurs when a sound wave strikes an object that is The beam will be deviated as it travels through the
larger than the wavelength equal to or smaller than the wavelength tissue
Occurs due to difference in wave velocity across an
interface between two materials
DIVERGENCE A substance which has densely
Occurs when the beam travels through tissue and it
will diverge due to diffraction effects
packed will have a high Acoustic
impedance. Ex. Bone
ACOUSTIC IMPEDANCE AND
REFLECTION
Z= ρc The average velocity of
ultrasonic waves in soft tissue In substances with the same acoustic
impedanceà there is a total transmission
Density (ρ) kg/m3 is 1540 meters per second of energy and therefore no reflection.
Speed of sound (c) m/s
Acoustic impedance (Z) rayls =
kg/m2/s Z1= 0% ; Z2=100%
21
9/9/23
In substances with a large difference THE INTENSITY REFLECTION EQUATION
in acoustic impedanceà there is a
In substances with a small This equation allows you to calculate the intensity
large amount of reflected energy reflection coefficient (R) for an interface between two
difference in acoustic substances providing you know their acoustic impedance.
and a small amount of transmitted R= Ir or R = (Z 1 –Z 2)2
impedanceà a small amount of energy. Ii (Z 1 + Z 2)2
energy is reflected but the
majority is transmitted. Ir = reflected intensity
Z1 = 99% ; Z2 =1% Ii= transmitted intensity
Z1= 5% ; Z2=95%
Example: what is the intensity reflection ACOUSTIC IMPEDANCE AND
coefficient R for an interface between liver ULTRASOUND IMAGING If there is a large acoustic mismatchà Ex:
(Z1= 1.65Mrayls) and fat (Z2= 1.35 Mrayls) between bone and muscle or air and soft tissue,
then a large proportion of energy is reflected.
This results in a strong echo, which produces a
The difference in acoustic bright image on the display.
R = (Z1 –Z2)2 = (1.65-1.35)2 = (0.3)2
impedance between two However, very little energy is transmitted across
(Z1 + Z2)2 (1.65+1.35)2 = (3.0)2 the interface and any echoes produced beyond
substances is known as the the interface do not have enough energy to
= (0.1)2 = 0.01 = 1.0 % acoustic impedance mismatch. produce image.
ACOUSTIC
IMPEDANCE AND
REFLECTION
Substances with same acoustic impedance:
100% energy transmission
If the acoustic impedance mismatch is Therefore, it is not practical to use No reflection
smallà Ex: between two soft tissues, then ultrasound to produce images of soft Substances with a small difference in acoustic
impedance:
a small proportion of energy (1% or less) tissue subjects containing gas or bone.
95% energy transmission
is reflected. 5% reflection
The rest of the energy is transmitted Ultrasound imaging is very good at Substances with a large difference in acoustic
across the interface to produce further discriminating between substances with impedance:
echoes from other interfaces deeper 1% energy transmission
small differences in acoustic impedance 99% reflection
within the subject. such as soft tissues
22
9/9/23
TRANSDUCERS/SCANNING PROBES
The most expensive part of any ultrasound unit
Part III Transducer TRANSDUCER/PROBE
A device which converts one form of energy to
1.) Linear Array/ Linear sequenced Array Transducer
another
Converts electrical energy into ultrasound waves 2.) Sector/Curvilinear Array Transducer
and vice versa 3.) Convex Transducer
Contains piezoelectric crystals 4.) Phased Array Transducer
Transmit ultrasound beam
Receive reflected echoes
CHOOSING THE
APPOPRIATE TRANSDUCER
3.) Pediatric Ultrasound
4.) Phased Array Transducer 1.) Obstetric Ultrasound 5.0 MHz transducer: for children
Flat faced transducer Focused at 5-7 cm
Linear or convex transducer
Wide field of view
3.5 MHz: better in later pregnancy Sector transducer of 7 MHz:
useful in cardiac and cranial ultrasound
5.0 MHz: best during early pregnancy Neonatal brain scans
Focused at 7-9 cm For adult testis and neck
Focused at 4-5 cm
2.) General Purpose Ultrasound
Sector or convex transducer
3.5 MHz
Focused at 7-9 cm
23
9/9/23
BEAM SHAPE
Increasing the frequencyà will result II. INTENSITY OF THE BEAM III. SIDE LOBES
in a longer near field and less far
field divergence
CRYSTAL NEAR FIELD FAR FIELD
A narrow crystal diameterà will
The intensity of the beam is the Some of the energy from the
THIN
result DECREASE
in a narrower beam inINCREASE
the near power (measured in watts) flowing transducer radiates at various
field,
FREQUENCY and thereFIELD
is more divergence
THICK NEARINCREASE FARDECREASE
FIED in through a unit area. angles to the transducer face, and
the far field
INCREASE LONG LESS these are known as side lobes.
DECREASE SHORT MORE
24
9/9/23
25
9/9/23
REVERBERATION
This is the production of spurious (false) echoes
due to repeated reflections between two interfaces Artifacts appearing as parallel, equally This artifact will often be seen:
with a high acoustic impedance mismatch. spaced lines is called reverberation
Comet tail
Reverberation artifacts are a result of the 1. At the skin-transducer interface
presence of two or more strong reflecting 2. Gas surface and the transducer
surfaces
Shotgun wound
ACOUSTIC
SHADOWING
This artifact can be differentiated from the real echoes
This appears as an area of low amplitude echoes (hypoechoic or
due to the lack of breathing movement occurring. anechoic) behind an area of strongly attenuating tissue. Acoustic shadowing will occur at the
It is cause by severe attenuation of the beam at interference, resulting
in very little sound being transmitted beyond. interface with a large acoustic
The attenuation can be due to either absorption or reflection of the
To help to eliminate this artifact it may be useful to: sound waves or a combination of two. mismatch such as:
1. Increase the amount of gel used
2. Used a stand-off pad (gel pad or a bag of saline) 1. Soft tissue and gas
3. Reduce the gain
2. Soft tissue and bone or calculus
4. Move the position of the transducer.
ACOUSTIC
ENHANCEMENT
If the shadowing contains reverberation
(ringdown echoes), it is likely to be gas. This artifact appears as a localized area of Can commonly be seen distal to fluid filled
increased echo amplitude behind an area of
structures such as the urinary bladder, the
low attenuation.
gallbladder or a cyst.
If the shadowing is clear shadowing,
absent of any ringdown, it is likely to be
a calculus This artifact can often be a useful diagnostic
aid, particularly when scanning a soft-tissue
mass or cyst containing low-level echoes.
Calcified mass
26
9/9/23
BEAM WIDTH
EDGE SHADOWING
ARTIFACT
A combination of refraction and reflection
occurring at the edges of rounded structures (and
when the speed of sound is different from that in All echoes returning to the transducer will
Acoustic enhancement is surrounding tissue) will result in an edge have arisen from across the full width of the
caused by weakly attenuating shadowing artifact. beam, which can vary by several millimetres.
structures.
SLICE THICKNESS
ARTIFACT
of the beam, and are
These occur due to the thickness
similar to beam width artifacts.
However, they occur at 90 degree to the scan plane These artifacts will typically be seen
Correct positioning of the focal with echoes from interfaces in front of and behind in transverse views of the urinary
the assumed plane of origin appearing in the
zone will help to reduce this display. bladder
artifact.
Beam thickness artifact is primarily
dependent upon beam angulation
SIDE LOBE
ARTIFACT
beam exists as
The energy within the ultrasound
several side lobes radiating at a number of angles
from a central lobe.
Echoes are generated by these side lobes in They may also arise within a cyst, where
addition to the main lobe, but all the returning adjacent structures are portrayed as if arising
This artifact is a result of inherent echoes are assumed by the transducer to have from within the cyst. These appearances can
arisen from the central axis of the main lobe.
characteristics of the transducer, and give rise to a false diagnosis of septations
apart from trying a different within a cyst.
transducer, cannot be eliminated.
27
9/9/23
fluid-air interface.
Diaphragm
In the transverse plane the edges of the In order to establish whether these 10. EQUIPMENT-GENERATED
muscle act as a lens and cause the echoes are genuine, it is necessary to ARTIFACTS
ultrasound beam to be refracted and move the transducer slightly to one
this causes a single structure to be side to avoid the junction of the rectus
interrogated by the two separate abdominis muscles. Incorrect use of the equipment controls
refracted beams. can lead to artifacts appearing.
COMPONENTS AND
CONSTRUCTION OF A TYPICAL
TRANSDUCER
FREQUENCY
If too mach gain is applied then the Affects the quality the ultrasound image 1.) PHYSICAL HOUSING
electronic noise, inherent in all systems, Higher Frequency Contains all individual components
will also be amplified. Shorter wavelength Provides the necessary structural support
Better Resolution Acts as an electrical and acoustic insulator
Lower Penetration
Higher Absorption
2.) ELECTRICAL CONNECTIONS
If too little gain is applied, this can lead to Lower Frequency
Formed in front and back of the crystal
loss of relevant information, and incorrect Longer wavelength
Made of thin film of gold or silver
diagnosis may occur. Poor Resolution
Higher Penetration
Lower Absorption
28
9/9/23
BANDWIDTH
BACKING/DAMPING MATERIALS
5.) ACOUSTIC LENS
Shortens the ultrasound pulse length Reduce the beam width of the transducer The range of frequencies contained within an
Improve image resolution
Width of the Beam: determines lateral resolution
ultrasound pulse
Eliminates the vibrations from the back face
Lateral Resolution: the ability to resolve structure across or perpendicular to the
beam axis
Wide Bandwidth:
Controls the length of vibrations from the front face Materials:
Aluminum
Shorter spatial pulse length
Improves axial resolution Perspex Wider range of frequency
Materials: Polystyrene
Narrow Bandwidth:
Plastic or epoxy resin 6.) IMPEDANCE MATCHING LAYER
Sandwich between the piezoelectric crystal and the patient Longer spatial pulse length
Cork Chosen to improved transmission into the body
BANDWIDTH
Narrower range of frequency
Rubber
Araldite loaded with tungsten powder
CHOOSING THE APPOPRIATE TRANSDUCER 3.) Pediatric Ultrasound
5.0 MHz transducer: for children TWO REGIONS OF ULTRASOUND BEAM
1.) Near Field/Fresnel zone
1.) Obstetric Ultrasound Focused at 5-7 cm 2.) Far Field/Fraunhofer zone
Linear or convex transducer Sector transducer of 7 MHz: Increasing Frequency
3.5 MHz: better in later pregnancy Neonatal brain scans Longer near field
5.0 MHz: best during early pregnancy For adult testis and neck Less far field divergence
Narrow Crystal Diameter
Focused at 7-9 cm Focused at 4-5 cm Narrower near field
More far field divergence
ULTRASOUND BEAM Thin Crystal
Decreased near field
2.) General Purpose Ultrasound Area through which the sound energy emitted from the ultrasound
transducer Increased far field
Sector or convex transducer Thick Crystal
3.5 MHz Three dimensional and symmetrical around its central axis Increased near field
Focused at 7-9 cm Decreased far field
BEAM INTENSITY SLICE THICKNESS SPATIAL RESOLUTION
The power (measured in watts) flowing through a unit area Three dimensional volume displayed as a two Detail Resolution
The ability to display two structures situated close together as
SIDE LOBES/GRATING LOBES dimensional image separate images
Lobes at various angles to the main beam Higher Frequency:
Approximately 15% of the energy in the beam RESOLUTION Better resolution
Cause a degradation of lateral resolution The ability of an imaging system to differentiate between Lower penetrability
structures Higher absorption
BEAM WIDTH Lower Frequency:
The dimension of the beam in the scan plane Spatial Resolution: resolution in space
Poor resolution
Affects the spatial resolution Contrast Resolution: resolution of gray shades
Higher penetrability
Narrow Beam Width Temporal Resolution: resolution in time
Lower absorption
Better spatial resolution
29
9/9/23
TWO COMPONENTS
OF SPATIAL
RESOLUTION
LATERAL RESOLUTION CONTRAST RESOLUTION
TWO COMPONENTS OF SPATIAL RESOLUTION Azimuthal, Transverse, Angular or Horizontal The ability of the imaging system to differentiate
between body tissue and display them as different
The ability to distinguish two objects perpendicular
1.) AXIAL RESOLUTION to the ultrasound beam shades of gray
Longitudinal, Linear, Depth or Range
Depends upon the beam diameter Optimized by using the correct overall gain
The ability to distinguish two objects parallel to the
ultrasound beam Smaller Beam Width: better lateral resolution
Depends upon the spatial pulse length and wavelength Larger Beam Width: poor lateral resolution
Short Spatial Pulse Length: good axial resolution
Longer Spatial Pulse Length: poor axial resolution
FIVE MAIN PROCESSES THAT CAUSE ATTENUATIONS
TEMPORAL RESOLUTION
Frame Rate ULTRASOUND INTERACTIONS AND 1.) ABSORPTION
Occurs when ultrasound energy is lost to tissues by its conversion to heat
The ability of the imaging system to display events ATTENUATIONS Main factor causing attenuation
which occurs at different times as separated images Higher Frequency:
Greater amount of absorption
Higher Frame Rate: better temporal resolution ATTENUATION Bone: higher absorption coefficient
Increasing protein content gives increasing absorption
Decrease in the intensity and amplitude of the Blood –> Fat –> Nerve –> Muscle –> Skin –> Tendon
ultrasound waves as they pass through tissues –> Cartilage –> Bone
Best Absorption: tendon, ligament, fascia, joint capsule & scar tissue
Unit: decibels per centimeter
REFLECTION SCATTERING REFRACTION
Occurs when two large structure of significantly Occurs when an ultrasound wave strikes a boundary Occurs when the beam encounters an interface
different acoustic impedance form an interface or interface between two small structures between two different tissues at an oblique angle
Occurs when a sound wave strikes an object that is Occurs when a sound wave strikes an object that is The beam will be deviated as it travels through the
larger than the wavelength equal to or smaller than the wavelength tissue
Occurs due to difference in wave velocity across an
interface between two materials
30
9/9/23
ULTRASOUND ARTIFACTS 1.) REVERBERATION
DIVERGENCE Comet tail
Occurs when the beam travels through tissue and it A structure in an image which does not directly The production of spurious or false echoes due to repeated reflections
between two interfaces with a high acoustic impedance mismatch
will diverge due to diffraction effects correlate with actual tissue being scanned The presence of two or more strong reflecting surfaces
Often occur at:
Skin-transducer interface
Gas surface and transducer
Prevention/Elimination:
Increase the amount of gel used
Used a stand-off gel pad
Reduce the gain
Move the position of the transducer
ACOUSTIC SHADOWING ACOUSTIC ENHANCEMENT EDGE SHADOWING
Caused by highly attenuating structure Caused by weakly attenuating structures Combination of refraction and reflection occurring at
Often occur at: the edges of rounded structures
Soft tissue and gas Often occur at:
Soft tissue and bone or calculus Distal to fluid-filled urinary bladder, gallbladder
Calcified mass or cyst
Fluid-filled mass
BEAM WIDTH ARTIFACT SLICE THICKNESS ARTIFACT SIDE LOBE ARTIFACT
Variations of all echoes returning to the transducer Occurs due to the thickness of the beam Echoes generated by side lobes assumed by the
Prevention/Elimination: Dependent upon beam angulation transducer to have arisen form the central axis of the
Often seen in: main lobe
Correct positioning of the focal zone
Transverse view of the urinary bladder Appearance can give rise to a false diagnosis
Inherent characteristic of the transducer Inherent characteristic of the transducer
31
9/9/23
MIRROR IMAGE ARTIFACT DOUBLE IMAGE ARTIFACT EQUIPMENT-GENERATED ARTIFACT
Caused by specular reflection of the beam at a large Caused by refraction of the beam Caused by incorrect use of the equipment control
smooth interface Often occur at:
Often seen in: Rectus abdominis muscle
Fluid-air interface Prevention/Elimination:
Diaphragm Move the transducer slightly to one side to avoid the
junction of rectus abdominis muscle
A-mode
CHAPTER IX
A-mode (A-scan, amplitude modulation) is a one-
Ultrasound techniques 1.) A-MODE display (Amplitude Mode) dimensional examination technique in which a transducer
with a single crystal is used.
ü Echoes are shown as peak and the distance The echoes are displayed on the screen along a time
(distance) axis as peaks proportional to the intensity
between the various structures can be (amplitude) of each signal. The method is rarely used today,
measured. as it conveys limited information, e.g. measurement of
distances.
ü Strength , depth , height of echo.
B-mode
CHAPTER IX CHAPTER IX
mode)
2.) B-MODE display (Brightness
1.) A-MODE display (Amplitude Mode) B-mode (brightness modulation) is a similar
2D mode; Static Mode technique, but the echoes are displayed as
The echoes are seen as bright dots which the points of different grey-scale brightness
ü A-Mode, or Amplitude Modulation, is the corresponding to the intensity (amplitude) of
display of amplitude spikes of different position of the reflecting structures on a two-
dimensional image each signal .a
heights.
32
9/9/23
CHAPTER IX CHAPTER IX
Doppler Ultrasound Doppler Ultrasound
Doppler flow
üChristian Johann Doppler noted in 1842 ü This mode makes use of the Doppler effect
in measuring and visualizing blood flow
33
9/9/23
CHAPTER IX
very short pulses. All echoes arriving at the
transducer between the pulses in a certain
time Pulsed wave the
interval (termed Doppler
gate) are registered
ü The Doppler shift frequency is in the crystals, one permanently frequencies is limited by the pulse repetition
emitting ultrasound and the frequency (PRF).
audible range, and therefore an audio other receiving all the echoes. When the Doppler frequency is higher than
amplifier and speaker are all that is No information is provided the pulse repetition frequency, high
about the distance of the velocities are displayed as low velocities in
necessary to listen to the fetal heart, called reflector(s), but high flow the opposite direction (spectral Doppler) or
utz stethoscope. velocities can be measured. in the wrong colour (colour Doppler).
REAL-TIME IMAGING,
DOPPLER EFFECT, AND
DOPPLER
COLOR flOW
The Doppler effect refer to a change in frequency when the
Real-time imaging is the presentation of multiple
motion of laminar or turbulent flow is detected within a
image frames per second over selected areas of the vascular structure.
Clinical applications
body.
The transducer may be composed of several elements Color flow Doppler is a technique that assigns a color scale
that can be electronically focused and fired in rapid to the change in frequency or Doppler shift.
sequence to produce a real-time image. Generally red signifies a shift toward the transducer,
Thus structures are visible as they change position in whereas blue signifies a shift away from the transducer
time. With this dynamic imaging it is possible to see,
for example, pulsatile vascular and cardiac structures,
diaphragm motion, and peristalsis in the bowel and
stomach.
ABDOMEN
Once a mass is suspected, the sonographer
must evaluate its acoustic properties to
determine if it is heterogeneous,
AND
homogeneous, hypoechoic, hyperechoic,
isoechoic, or anechoic. RETROPERITONEUM
A hypoechoic lesion is characterized by The upper abdominal ultrasound examination generally
very low-level echoes with a good
posterior border. includes a survey of the abdominal cavity from the
Characteristics of the
sonographic image A hyperechoic mass may represent a diaphragm to the level of the umbilicus.
tumor, thrombus, or calcification; the lesion Specific protocols are followed to image the texture,
present with bright echo reflectors and
borders, anatomic relationships, and blood flow patterns
possibly shadowing beyond it.
An isoechoic mass shows nearly the same within the liver, biIiary system, pancreas, spleen, vascular
texture pattern as the surrounding structures, retroperitoneum, and kidneys.
parenchyma with no significant change in Patients may be examined in the supine, decubitus,
the through transmission
upright, or prone positions.
An anechoic mass shows no internal
echoes, has smooth walls, and displays
increased through transmission.
34
9/9/23
Pancreas
The pancreas is a retroperitoneal gland.
Its head lies in the curve of the duodenum, and its
body and tail lie posterior to the antrum of the
stomach.
The texture of the pancreas varies, depending on
the amount of fat that is interspersed throughout
its islets of Langerhans.
Biliary system
On ultrasound examination the gallbladder and
bile ducts are seen just inferior to the right lobe of
the liver and anterior to the right kidney.
The intrahepatic bile duct runs slightly anterior to
the portal vein as it drains bile from the liver into
the storage cistern (gallbladder) via the cystic
duct.
Bile is released into the common bile duct, where
it joins the pancreatic duct secretions to drain into
the duodenum at a small raised area called the
duodenal papilla.
The biliary system is evaluated with ultrasound
to assess size and wall thickness and to detect
sludge, stones, polyps, or other masses.
35
9/9/23
Kidneys
The kidneys rest on the posterior surface of the
abdominal cavity in an oblique plane lateral to
the psoas muscles
The large right lobe of the liver causes the right
kidney to be slightly lower than the left kidney.
The renal vascularity may be imaged with color
Doppler as the patient is rolled into a slight
decubitus position.
GYNECOLOGIC
APPLICATIONS
SUPERFICIAL STRUCTURES
Anatomic features of the pelvis
Superficial structures such as the thyroid, The pelvis is divided into the greater and lesser
breast, scrotum, and penis are imaged very
pelvic cavity, with the pelvic brim being the
well with ultrasound using high frequency
circumference of a plane divide ng the two
cavities.
transducer .
The greater pelvic cavity, or "false" pelvis, is
Routine ultrasound examination superior to the pelvic brim and is bounded on
demonstrate minute structures such a the each side by the ileum.
lactiferous ducts within the breast and the The lesser pelvic cavity, or "true" pelvis, is
spermatic cord within the testes. In caudal to the pelvic brim.
conjunction with mammography, The walls of the pelvic cavity are formed by
ultrasound is able to characterize the several muscles collectively called the pelvic
texture of a breast mass to determine if the diaphragm. These muscles include the levator
mass is fluid-filled, complex, or solid. ani, piriformis, and coccygeus muscles.
36
9/9/23
Sonography of the
female pelvis
A complete transabdominal examination of the female A high-frequency transducer is inserted into the vagina to
pelvis includes visualization of the distended urinary image the uterus, cervix, fallopian tubes, ovaries, and
bladder, uterus, cervix, endometrial canal, vagina, adnexal area in coronal and sagittal planes.
ovaries, and supporting pelvic musculature. Sonography of the pelvis is clinically useful for imaging
The full bladder helps to push the small bowel normal anatomy, identifying the size of ovarian follicles
superiorly out of the pelvic cavity, flattens the body of as part of an infertility workup, measuring endometrial
the uterus, and serves as a sonic window to image the thickness, evaluating the texture of the myometrium,
pelvic. determining if a pregnancy is intrauterine or extrauterine,
detecting tumors or abscess formations, and localizing an
The rectum and other bowel structures may also be seen intrauterine contraceptive device
and must be distinguished from the normal pelvic
structures.
OBSTETRIC
APPLICATIONS
The pregnant woman is the ideal candidate for an
ultrasound examination.
The amniotic fluid enhances the sound penetration to
differentiate fetal anatomy, umbilical cord, placenta, and
amniotic membranes within the uterine cavity.
Endovaginal ultrasound is the procedure of choice
during the first trimester of pregnancy to delineate the
gestational sac with the embryo, yolk sac, chorion, and
amniotic cavities . The gestational sac may be visualized
as early as 4 weeks from the date of conception with
endovaginal ultrasound
The embryo, heartbeat, and site of the placenta may be
seen at 5 weeks of gestation.
• The beginning of the second trimester
( 1 3 to 28 weeks of gestation) allows
to image the detailed anatomy of the
fetus.
Thank you!
• Structures such as the brain, face,
limbs,
neck, abdominal wall, liver, gallbladder,
kidneys, stomach, pancreas, bowel ,
heart,
lungs, and bladder may be seen with
high resolution transducers.
• Heart motion, fetal size and position,
and number
of fetuses may be easily assessed with
sonography.
331 332
37