YG-0.36 360L Wanrooe Specificaciones
YG-0.36 360L Wanrooe Specificaciones
Pulsating vacuum sterilizer is strictly designed and manufactured in accord with GMP technical
Standard. It has passed ISO9001 quality management qualification standard and ISO13485:2003
This autoclave is suitable for pharmaceutical factory, hospital, scientific research institute and other
units to sterilize and dry asepsis clothes, instruments, culture medium, etc.
Overall Size Inner Size Steam Compressed City water Power N.W
Model
W*H*D (mm) W*H*D(mm) (kg/cycle) Air (m³/cycle) (kg/cycle) (Kw) (kg)
(1) Adopting the jacketed structure, a full dimpled jacket shall be provided wrapping around the
chamber in order to avoid steam condensation and facilitate, and proper temperature
distribution. which can be peripherally heated uniformly, preventing material from non-uniform
heating.
(2) Chamber: the chamber is made of SUS316L, According to mechanical and chemical
treatment, the chamber surface is bright, beautiful and corrosion resistant. Two stainless steel
rail guides in the chamber shall be provided as tracks for the loading cart.
(3) The insulating layer is made by aluminium silicate which is the best insulating material, and
The door seal is the inflatable type, pressurized with compressed air, and can withstand the
The doors shall have a safety mechanism to prevent opening of the door when the chamber
Sterilization cycle can start only after the door is fully closed and locked.
(5) Pipeline system is consisted of air filter, vacuum pump, pneumatic valve,etc.
Air filter for is using PALL company brand. The performance is excellent with no medium
Angle type of pneumatic valve: strong switch valve, without any mistake, remote
compressed gas control. with the inner chamber of interlinked valve for stainless steel clamp
type seat.
Vacuum pump:The autoclave include a liquid ring vacuum pump to obtain vacuum in the
chamber.
(6) Control system: PLC+ HMI control + Micro-printer. a timely data printing, should be able to
default one or more sterilization cycle. Electric apparatus system adopts domestic or imported
When automatic controller in the failure of the circumstances, the security device makes the
sterilization indoor pressure safety in the pressure of the atmosphere to the back of the state,
In order to maintenance, test and emergency needs, manual operation can be through the
The master controller system: can set the user name and password, also can set up
different user access. A manager, users, maintainers triple interface function. Managers can
be used according to different level, set on use of a password, ensure use safety, prevent
Touch screen: display work process parameters, control is simple, the operation is
Sterilization cart made of SUS 316L stainless steel is provided for loading, sterilization and
unloading the items. The cart have support guides at different levels in order to place the
Transfer trolley:
Transfer trolley made of SUS 304 stainless steel is provided for transporting and
loading/unloading the sterilization cart to and from the chamber. For proper operation of a
(8) The system have an E-Stop mechanism designed to stop all physical movement of the
equipment immediately.
(9) The GMP confirming connecter fixed can provide the GMP validation for the customer freely.
4.Process flow
Dynamic pre-vacuum:
Pulling vacuum in the chamber, and then while still pumping out air, allowing steam into the
chamber. This way, the air is removed from the chamber with no heat build-up on the material.
Pressure pulsing:
The aim is to provide rapid removal of non-condensable gases and condensates and to pre-heat the
Sterilization:
pressure and temperature are obtained in the chamber, sterilization time begins. If the temperature
Drying:
Pulling vacuum in the chamber, and then start drying time count while vacuum is still being made.
Throughout the drying process, the jacket is kept at working pressure with hot steam. In this way, the
high vacuum and heat radiation from the chamber walls helps drying the material.
Cooling:
Once sterilization is done, slow vacuumize made until pressure drops to 1.100 mbar. After that, while
the vacuum system is still working, vacuum break valve opens preventing drop of pressure and thus
accelerating the cooling process. The phase ends when cooling ending temperature is reached.
CONFIGURATION LIST
No. N am e M od el M an uf act ur er R em ar k
Door sealing
2 01-03 Runde China Medical used silicon rubber
ring
Ⅱ Door 02- 00
Door proximity
2 CLJ series CoRon China Sharp,easy to install
switch
Safety interlock
3 02-02 WANROOEMED High temperature resistance
device
Sterilizing
1 03-01 WANROOEMED
software
Pressure DANFOSS,
6 MBS-3000 High control accuracy, and reliability
transmitter Denmark
Pressure Reliable, performance, accurate
7 SNS series TOKYO Japan
controller control
Integration installation with manual
8 Solenoid valve 2V025-06 AMISCO Italy
operation, good performance
5 Pressure gauge YTF-98ZK Brighty, China Simple structure and good reliability
Loading
Ⅴ 05-00
system