Atlas991992Manual
Atlas991992Manual
F O R A T L A S M O D E L 9 9 1
A T L A S E C L I P S E ™ M O D E L 9 9 2
REV. C PN 48113-1
HUMPHREY SYSTEMS
A Division of Carl Zeiss, Inc.
5160 Hacienda Drive
Dublin, California 94568
P.O. Box 8111
Pleasanton, California 94588
(1) 877-486-7473
This book may not be reproduced in whole or in part by any means of infor-
mation storage, retrieval, or reproduction without permission. For information,
call or write to Humphrey Systems.
TABLE OF CONTENTS
1 INTRODUCTION
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
ATLAS Instrument and Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Ethernet Lan Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Network Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Twisted-Pair Installation Cables and Connectors . . . . . . . . . . . . . . . . . . . . . . . 1-7
Using Network Operating Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Network OS Drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Setting Up the ATLAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Using this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
2 PATIENT EXAMS
Full Exam Versus QuickVue Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Full Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
QuickVue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Aligning and Focusing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Position the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
How to Get the Best Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Align the Projection Head and Focus the Image . . . . . . . . . . . . . . . . . . . . . . . .2-3
Capture the Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Configuring Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Shutting Down the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Conducting the Full Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Building Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Saving the Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Displaying Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Conducting the QuickVue Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
4 ALTERNATE VIEWS
Open MasterVue Control and Select View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Corneal Power Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Diopter Scale Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Statistical Indices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Tangential Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Diopter Scale Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Numerical View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Diopter Scale Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Keratometry View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Photokeratoscope View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Profile View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
MultiVue Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
5 DATABASE MANAGEMENT
Database Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Opening the Utilities Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Backup the Hard Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Backup all Data to Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Backup New Data to Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Format Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Reload Hard Disk from Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Archiving and Restoring Examinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
FileVue™ Archiving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Archive Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Archive Oldest Exams without Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Select by Group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Select by Doctor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Select by Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Select Specific Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Archive Exams to Media/External Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Restore Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Exporting and Importing Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Export Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Import Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Maintaining the Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Edit Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Select Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Edit Patient Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Delete Patient Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Add New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Edit Exams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Select Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Select Exam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Edit Exam Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Save Exam Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Edit Patient Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Delete Exam Infomation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Edit Doctor Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Add Doctor Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Change Doctor Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Delete Doctor Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Edit Group Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Add Group Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Edit Group Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Delete Group Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Edit Operator Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
Add Operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
Edit Operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
6 OPERATING ENVIRONMENT
Change Preferences Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
General Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Selection Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Install New Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Set Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Set the Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Set the Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Changing Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
7 PRINTING
Printing to the Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Printing to a Slide File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Printing to an Image File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
15 SPECIFICATIONS
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Environmental Requirements (No Condensation) . . . . . . . . . . . . . . . . . . . . . . . 15-1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
User Input Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Operating System and Data Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Data Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-2
Data Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-2
Communication File Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-2
A GLOSSARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
I INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1
1 INTRODUCTION
SYSTEM OVERVIEW
The Humphrey ATLAS™ Model 991 and Atlas Eclipse Model 992
(referred to herein as ATLAS) Corneal Topography Systems are
diagnostic instruments that measure the curvature of the cornea of the
eye, and produce a topographical map. This map can be viewed on a
computer screen and printed in color. The information can be used to
help diagnose eye pathology and make surgical decisions.
The ATLAS projects a series of light rings onto the cornea of the eye.
The reflected images are detected and captured by a video camera. The
instrument determines corneal curvature values by measuring the
distance between the rings. These are converted by the software into a
topographical map of the cornea.
Keyboard
Zip Drive
Color Printer
A variety of printers can be used with the system, but generally any
Windows compatible printer can be connected to the ATLAS. Other
printer options may be available and can be discussed with Technical
Service. Please refer to the owner’s manual enclosed with your printer
for routine maintenance and set up procedures.
Projection Head
Assembly
Joystick
The joystick moves the base of the unit from side to side and back and
forth for positioning. The base can also be locked in place when you
move the unit by turning the joystick counter-clockwise until it stops.
The joystick is also used to focus the image. The button on top of the
joystick activates the camera and captures the image.
The left chin rest is positioned at an angle for the right eye, and the
right chin rest is angled for the left eye.
The chin rest (from the patients perspective) is labeled with blue arrows
to indicate the head position for both right and left eyes. The patient
should be instructed to place their chin in the appropriate side
depending on which eye is being examined.
Canthus Marks
Chin Rest
Computer
The computer is an IBM-compatible unit with a hard disk drive. It also
has a 3 1/2” drive, which uses standard 3 1/2” diskettes. An external,
optional iomega Zip drive is used for archiving and backing up of data.
The Zip drive uses special 100 MB disks designed solely for use with the
iomega drive.
Please refer to the installation instructions included with the iomega Zip
drive or to the following for more information:
Keyboard Serial port Parallel port for
connector connection (not printer and/or
port operational) Zip drive
Floppy drive
Modem or Networking Model, patent, and serial VGA port (for Video In port
PCMCIA sockets numbers with certification high resolution (not operational)
(interchangeable) stickers and electrical external
requirements. monitor)
• Attach both the cable from the computer’s parallel port and the printer
cable to the printer icon connector on the Zip drive.
The iomega Zip drive uses 100 MB disk cartridges for easy and
convenient operation. There is a disk cartridge eject button for disk
cartridge removal. In an emergency (use only when power is
disconnected), a disk cartridge may be ejected by inserting a jumbo
paper clip into the rear hole next to the printer icon.
The Zip drive has a green power light indicator and an amber data light
indicator that flashes whenever data is transferred to or from the drive.
Network OS Drivers
The following Ethernet drivers are provided on the hard disk in the
ethernet directory. The table shows what driver to use with each
network operating system. For instance, if you are using Netware 4.1 on
a network server, you would use the SMC9000.LAN driver.
If the drivers intended for use are not listed here, Humphrey cannot
insure or guarantee that these drivers will work with the ATLAS.
To shut down the system, select “SHUT DOWN” from the main menu
which will allow Windows ‘95 to shut down properly. Failure to shut
down the system properly can result in fatal system errors.
Note: The Zip drive needs to be powered on and connected in order for
the computer to recognize it and assign a drive for it to use. This must
be done in the “boot up” process of the computer.
If you should need to replace any of the components, you must be sure
that the cables are connected properly.
The side panel for the ATLAS for Models 991 and 992 have modem
and ethernet PCMCIA sockets to allow the user to connect any style or
type of modem or networking card using standard "plug and play"
capabilities. Windows '95 may install the software for the modem or
networking card automatically, or ask the user to insert the installation
disk that was provided by the manufacturer of the PCMCIA card.
NOTE: Please see the user installation instructions for the type of
modem or networking card that is being installed. Humphrey Systems
does not assume any liability or responsibility for the installation,
troubleshooting, repair, or maintenance of third party hardware or
software. Only limited technical support may be provided by Humphrey
in the event of a problem. Please refer to the ATLAS User Manual,
pages 1-9 for a list of approved network drivers for the ATLAS system,
or contact Humphrey Technical Support for a list of approved
PCMCIA network/modem cards.
Floppy
Modem or Networking Model, patent, and serial VGA port (for Video In port
PCMCIA sockets numbers with certification high resolution (not operational)
(interchangeable) stickers and electrical external
requirements. monitor)
Caution:
This instrument is not anesthetic proof. DO NOT use in the presence
of flammable anesthetic since this creates a risk of explosion.
• The ATLAS has ventilation openings to allow for the release of heat
generated during operation. If these openings are blocked, built-up
heat can cause failures which may result in a fire hazard.
• DO NOT use accessories that are not designed for this instrument.
Use only those parts recommended by Humphrey Instruments to
achieve optimum performance and safety.
1. Plug the instrument into a different outlet so that the instrument and
the receiving device are on different branch circuits.
4. Move the receiving device and the instrument away from each other.
Indicates fuse.
Chapter 1 — Introduction
Contains an overview of the instrument and its functions, instrument
setup and safety precautions.
Chapter 7 — Printing
Describes how to print data on the color printer, other color printers,
and color slides.
Chapter 15 — Specifications
Lists the ATLAS instrument specifications.
Appendix
Contains a Glossary of terms as well as an Index for reference.
Patient Data must be entered to use the Full Exam feature. All input is
optional except patient name and right eye/left eye designation. The
patient and exam information is automatically saved in the database on
the hard drive.
QuickVue
QuickVue is a screening tool used for high-volume or routine
diagnostics. It captures a single image that may be saved and/or printed
after the examination. QuickVue does not require any data input prior
to the exam, and the exam may be printed without saving on the hard
drive.
2. Make sure the patient’s head is aligned properly at a 45° angle in the
ATLAS chinrest with the chin centered in the proper position. The
chin should be pressed firmly down in the chinrest.
4. Ask the patient to look at the red fixation light while you align the
projection head.
How to Get
the Best Images
1. Ask the patient to blink just before you capture the image. This
causes a tear to be distributed evenly over the cornea.
2. If the patient has ptosis, have an assistant stand to the side of the
instrument and hold the patient’s eyelid up, being careful not to block
the optical path of the instrument or press on the patients eye.
4. Ask the patient to open both eyes wide just before you capture the
image.
2. Position the projection head by moving the base until the live image
of the eye appears on the screen. Adjust the height of the instrument
by rotating the joystick handle until the rings appear.
3. Center the crosshairs anywhere inside the smallest ring of the live
image.
4. Move the joystick forward and back to focus the clearest image of the
smallest ring.
2. Quickly make any minor corrections to the focus. Then press the
button on the top of the joystick to capture the image.
CONFIGURING REMARKS Flexibility when entering patient data, groups and adding
comments is the purpose of the "Configure Remarks" feature
which is user programmable according to personal preferences.
This button is located in the Utilities menu.
With this feature remarks can be added, deleted, and edited and
Once a list is edited, the user can move the order of the remarks boxes
around by holding down the control key on the keyboard, and
dragging with the glidepad button. The size of the box can be
increased or decreased, and completely new boxes can be added
to the current selection of remarks boxes as well.
The Full Exam selection, view, and save function has been designed
to increase office efficiency and speed when capturing multiple images.
Full Exam allows you to capture multiple images of either OD or OS
within the same full exam sequence. It also allows remarks, groups,
and operator data to be entered in within the same screen as the
processed image. ATLAS users will have to identify which eye was
imaged and will be prompted by the software, while ATLAS Eclipse
users will not due to the automatic eye detection features in the chin
rest sensors of the Eclipse.
BUILDING REMARKS Entering pre-existing, edited, or new remarks that have been added
from the configure remarks button can be done from the "Build
Remarks" screen simply by highlighting the appropriate boxes and
clicking 'OK". These remarks automatically become attached to the
patient's record, and can be used as a way of identifying particular
exams. This is especially helpful when viewing older exams, or when
someone other that the original user is viewing the exam at a later
date.
Any exams that aren't saved and do not display a map will be
discarded after clicking on "Views". This button will then take you to
the default single or multivue screen.
DISPLAYING THE If multiple OD and OS images have been processed, the software will
PROCESSED IMAGES choose the two most recent images and display these two images in the
OD/OS Compare Display. If multiple OD or OS images have been
selected, then the Trend with Time Display will be shown with the
exams in chronological order.
2. Press the button on the joystick to capture the image. A single image
is captured, and the default view is displayed.
MASTERVUE CONTROL
The MasterVue Control lets you select the patient and exam to be
reviewed. You may also change to an alternate view.
Opening the
MasterVue Control
There are two ways to open the MasterVue Control.
1. Scroll down the patient list by placing the arrow on the button to the
right of the patient list and holding it down until the name you want
appears.
2. Move the pointer to the name you want and click to select. That
name now appears in the selection box at the top of the list.
3. To speed the search, you can type the first several letters of the last
name in the selection box at the top of the patient list. This will call
up the patient you wish to select, after enough letters have been
entered.
Select Exam
After you have selected a patient, the exams for that patient are
displayed in a list. Select the exam you want to review by clicking on it.
Select View
When the selected patient exam and view are correct, click the OK
button. The selected view appears.
Exam Selection
From the MasterVue Control, you can also set preferences for:
The Set Preferences button on the MasterVue Control is used to change the Map
display. For complete instruction on using this button, see “Axial Map” in Chapter
4.
• AutoSize
• Custom
Red, orange, and yellow are “hot” colors and show the steeper areas of
the cornea. Green is an “intermediate” color. Light and dark blue are
“cool” colors and show the flattest areas.
Note: The scale can be changed from diopters to millimeters in the Change
Preferences screen according to user preferences.
When you change the scale, the display changes to show the new
diopter values.
Standard
The Standard scale default ranges from 38.5 to 50.0 in 0.50 diopter
increments. This scale can be changed to a user defined default in the
Utilities Menu from the Change Preferences screen under change
Standard Scale. The Standard Scale can be set to user defined defaults,
or it can be reset to the default scale, simply by clicking on this button
and manually setting the Standard Scale to the user preferences.
MULTIVUE DISPLAYS
MultiVue Displays let you compare more than one exam or more than
one view of the same exam as well as to set preferences for the single
display. The current MultiVue displays are:
• Single Display
• Difference Display
• Exam Overview
• Custom Display
Open the MultiVue
Displays Screen
1. Use the MasterVue Control to select the patient, exam, and view you
want. This becomes the currently active exam.
The Trend with Time Display shows the currently active exam and up
to three previous exams for the same eye. The exams are displayed in
chronological order from top left to bottom right (oldest exam in upper
left corner, and current exam in bottom right).
If there are not enough previous exams to fill all four screen positions,
“No Exam Found” will appear in one or more positions in the display.
Difference Display
The Difference Display shows the previous exam for the same eye in
the first position, the currently active exam in the second position and
the diopter power difference between the two. The display also shows a
Standard diopter scale of ± 2 diopters and an AutoScale diopter scale
adjusted to accommodate the widest range of the two exams.
If you want to compare an earlier exam, click the GlidePoint control
button on the exam image you want to replace. The MasterVue Control
appears. Select the exam from the list, and click the OK button. The
new exam will be displayed and the Difference Map will be recalculated.
For more information about the Custom Display, see page 3-8.
The Exam Overview Display shows four separate views of the same
exam. You may select which views appear in this display by setting
MultiVue preferences.
The Custom Display shows up to four different exams. You may select
any combination of patients, exams, and views. This display is useful for
slides and other visual presentations.
There are two ways to open a Custom Display. First, you can select
Custom Display from the MultiVue Display screen. Second, you can
turn any MultiVue Display described previously into a Custom Display.
Click on any view to open the MasterVue Control. Then select an
alternate view and click the OK button.
4. Use the Installed Views list box to set your view preferences on this
screen. Click on an installed view in the list, “drag” the view into the
desired box, and then release the GlidePoint button.
5. The view that appears in this box will appear when Single Display is
selected on the MultiVue Displays screen.
6. Difference Map View — only axial map and tangential map views can
be loaded in this selection box.
7. These views will appear in the upper and lower halves, respectively, of
the OD/OS Compare Display.
Exam Overview
8. The views listed here will comprise the Exam Overview Display.
Note that the views will be arranged in the same order as they appear
in these boxes.
The MasterVue Control lets you select an alternate view of the data
captured by the instrument. Several alternate views are available, and
they are described in the following sections of this chapter.
OPEN MASTERVUE
CONTROL AND SELECT
VIEW
For complete instructions on how to select a data view, see Chapter 3.
1. There are two ways to open MasterVue Control:
• First, you can click the Review button on the Main Menu.
Description
The Axial Map displays the curvature of the cornea as a topographical
map, using 24 colors to represent dioptric power.
Colors
Distance
Location
You can move the pointer arrow to any part of the map and determine:
• diopter power
The large crosshair on the map denotes the point at which you are
located on the view. This is usually the vertex or “geographic center” of
the map or view. The smaller crosshair enclosed by a circle denotes the
center of the pupil, or “optical center” of the eye. The triangle identifies
the apex of the cornea as defined a best fit ellipsoid to the corneal
surface.
Diopter Scale Options
The data in the Power Map can be displayed in four different
diopter/millimeter scales:
Note: The uniform scale can only be accessed when viewing a MultiVue Display.
The CIM value is located on the display under the numerical values for
power, vertex, and semi-meridian.
If there are not enough points available to reliably fit the corneal shape,
if the fit is judged poor, or the shape factor is too negative a number to
be judged accurately, the message “Insufficient Data” will appear instead
of the shape factor value.
The Shape Factor is located on the Map display below the CIM value.
Simulated “K”
Readings
MultiVue Displays
Map data can be displayed as a:
• Single Display
• Difference Display
• Exam Overview
• Custom Display
This option is used to change only the Axial Map, Refractive Power
Map, or Tangential Map display. The settings in Set Preferences allow
the Map to be displayed with or without the pupil demarcation, the
rectangular grid, the Apex marker, the extrapolation options, and the
CIM and Shape Factor Values.
1. To change the Map setup, click the Set Preferences button on the
MasterVue Control. The View Preferences screen appears.
3. When you have finished setting preferences for the Map, click the OK
button. The MasterVue Control appears. Note: The preferences you
set will apply to all subsequent Map displays until you change the
preferences again.
TANGENTIAL MAP
Description
The Tangential Map displays the cornea as a topographical map, but
bases the calculation on the local curvature of a given meridian. It
should be used along with the Axial Map, or sagittal display, to gain
more insight into the corneal shape.
Colors
Red, orange, and yellow are “hot” colors and show the steeper areas of
the cornea. Green is an “intermediate” color. Light and dark blue are
“cool” colors and show the flattest areas.
Distance
Location
You can move the pointer arrow to any part of the map and determine:
• diopter power
The large crosshair denotes the point at which you are located on the
view. This is usually the vertex or “center” of the map or view. The
smaller crosshair enclosed by a circle denotes the center of the pupil, or
visual axis of the eye. The triangle indicates the apex of the cornea as
measured from the topographical map.
Note: The uniform map can only be accessed when viewing a MultiVue Display.
MultiVue Displays
Tangential Map data can be displayed as a:
• Single Display
• Difference Display
• Exam Overview
• Custom Display
GENERAL PREFERENCES
X Extrapolate
MAP PREFERENCES
X Show Pupil
X Show Grid
X Calculate CIM and Shape Factor
Preferences
Set Preferences
This option is used to change only the Axial Map, Refractive Power
Map, or Tangential Map display. The settings in Set Preferences allow
the Map to be displayed with or without the pupil demarcation, the
rectangular grid, the Apex marker, the extrapolation options, and the
CIM and Shape Factor Values.
3. When you have finished setting preferences for the Map, click the OK
button. The MasterVue Control appears. Note: The preferences you
set will apply to all subsequent Map displays until you change the
preferences again.
NUMERICAL VIEW
Description
Numerical View displays the diopter values for up to 5 circular areas or
zones on the cornea from 0 mm to 10 mm. The color of the numbers
represents the flatness or steepness of the cornea and corresponds to
the scale displayed on the left.
MultiVue Displays
Numerical View data can be displayed as the following:
• Single Display
• Exam Overview
• Custom Display
Description
The Keratometry View provides information simulating the data
obtained using a Keratometer. This view also provides readings for each
semi-meridian in three zones:
The data is listed by semi-meridian and zone on the left of the screen.
The Simulated Keratometer values at the top left of the display are
computed in a similar way to the computations performed by
Keratometers commonly used to measure the central 3 mm.
• Single Display
• Exam Overview
• Custom Display
MultiVue Displays
The Photokeratoscope image can be displayed as the following:
• Single Display
• Exam Overview
• Custom Display
Description
In the top portion of the screen, the Profile View plots the steepest and
flattest meridians. The inset boxes show the location of each meridian.
To plot other meridians, click on the colored circle in one of the inset
boxes and drag the circle to the location you want.
The central gray box marks the pupil’s relationship to the profile view.
The pupil indicator tracks the pupil from the Axial Map and displays it
on the Profile View to indicate the actual visual center of the eye.
The bottom part of the screen plots the difference in diopters between
the steepest and flattest meridians shown in the top part of the screen,
and the amount of corneal coverage is displayed in millimeters below.
MultiVue Displays
The Profile View image can be displayed as the following:
• Single Display
• Exam Overview
• Custom Display
DATABASE MANAGEMENT
DATABASE DESCRIPTION
The ATLAS maintains a database containing patient and exam
information. The database is stored on the instrument’s hard disk drive.
This database is essential to the MasterVue Control and MultiVue
Display capabilities.
OPENING THE
UTILITIES SCREEN
Database maintenance functions are available from the Utilities screen.
BACKUP THE
HARD DISK
An important part of maintaining your records is backing up, or
copying, all the data on the hard drive to disk. This gives you a backup
copy of all your records. If the data on your hard disk is damaged in
any way, you can use the backup copy to restore the records to your
hard disk.
Back up your data using the Backup Hard Disk option in the Utilities.
It is recommended that you backup the hard disk every day, if possible.
If you conduct patient exams on particular days of the week, you
should back up at the end of those days. You may want to run the
backup utility overnight to minimize the down time during office hours.
We recommend that you keep two versions of your backup media. You
can then alternate the media, using the oldest version to do your current
backup.
Before you attempt to back up your hard disk drive, be sure you have a
formatted media.
1. To back up all the data on your hard drive to media, click the
Complete button on the Backup screen. A message box appears.
Before you attempt to backup your hard disk drive, be sure you have a
formatted media.
1. To backup only the data that has changed since your last backup, click
the Standard button on the Backup screen. A message box appears.
FORMAT MEDIA
Before you can use a new media to backup your system, you must
format it if the media has not been preformatted by the manufacturer.
This allows the computer to read the media. You only need to format a
media the first time you use it. If you later want to reuse a formatted
media (and write over the data stored on it), you do not need to format
it again.
1. To format media for backing up the data on your hard drive, click the
Format button on the Backup screen. A message box appears.
ARCHIVING AND
RESTORING
EXAMINATIONS
When the hard disk is nearing capacity, the instrument will prompt you
to archive examinations on media or external drive using the
Archive/Restore option in the Utilities function. The instrument
prompts you to archive exams when the hard drive is nearing capacity.
You may archive at any time however.
To archive patient examinations, you may let the instrument select the
oldest exams on the hard disk. Or you can manually select:
• specific exams.
The selected exams are removed from the hard disk and stored on
external hard drive or media.
To restore patient exams, you select the exams from the external drive
or media, and create a list. The exams on the list are restored to the
hard drive.
Note: Exams which have been archived cannot be read on other instruments. To
look at exams from other instruments, use the export/import function.
FILEVUE™ ARCHIVING
Archive Exams
1. To archive patient exams, click the Archive button on the
Archive/Restore screen.
2. Click on the button next to the drive you want to use. Then insert a
disk in the selected drive or select the external disk. Click OK. The
Select Archive Mode screen appears.
This screen lets you choose how you want to select exams for archiving.
Archive Oldest Exams
Without Selection
3. To let the instrument archive the oldest exams on the hard disk
without any additional manual selection, click on the Begin button.
Then go to “Archive Exams to Media/External Drive” on page 5-9.
Select by Groups
Use the Remove All button to exclude all groups currently selected.
Use the Add Selected and Add All buttons to move groups from the
excluded list back to the archive list.
When you are finished selecting groups to archive, click on the Archive
button. Then go to “Archive Exams to Media/External Drive.”
Select by Doctor
Use the Remove All button to exclude all doctors currently selected.
Use the Add Selected and Add All buttons to move doctors names
from the excluded list back to the archive list.
Select by Patients
8. To select exams by patient for archiving, click on the Patients button
on the Select Archive Mode screen. The Archive by Patients screen
appears.
Use the Remove All button to exclude all patients currently selected.
Use the Add Selected and Add All buttons to move patient names
from the excluded list back to the archive list.
When you are finished selecting patients’ exams to archive, click on the
11. The box on the left shows patient names. The middle box shows the
exams associated with each patient. To select an exam, first click on
the patient’s name. Then click on the specific exam you want to
archive. The exam will move from the middle box to the right box,
Archive Exams.
Use the Move All button to select all exams for a given patient. To
“unselect” an exam, click on the exam in the Archive Exams box.
When you are finished selecting exams to archive, click on the Archive
button. Then continue with “Archive Exams to Media/External Drive.”
13. The instrument archives the exams and displays a message telling
you how many exams were archived. Click OK.
14. If you are finished archiving exams, click Cancel. Then click
Return to return to the Utilities Menu.
RESTORE EXAMS
Restoring exams is required only when you want to view exam data that
was previously archived. To restore a patient exam, you must know on
which media the examination was archived. There are two ways to locate
the correct media.
• Or, you may insert any media in the appropriate drive, click on
utilities then Archive/Restore. After you have selected the
exams to be restored, a message box will display the labels of
the required disks.
1. Insert the media containing the exam(s) you want to restore in the
appropriate drive.
Click No if there isn’t enough room. You’ll then have to archive other
exams to make room for the exam(s) you want to restore.
Use the Move All button to select all exams for a given patient. To
“unselect” an exam, click on the exam in the Restore Exams box.
7. If you are finished restoring exams, click Cancel. Then click Return
to return to the Utilities Menu.
EXPORTING AND
IMPORTING EXAMS
The Export/Import option in the Utilities function lets you export, or
copy, examinations from the hard drive to media. You can also import,
or add, exams to the instrument from media.
Unlike archiving, exporting does not remove the exam data from the
hard drive. Exporting and importing are used to transfer exams from
one computer system to another.
2. Click on the button next to the drive you want to use. If necessary,
insert a media in the selected drive. Click OK.
A message box appears telling you how much room is available on the
target media. Click Yes to continue if there is enough room to export
the exam(s).
4. The box on the left shows patient names. The middle box shows the
exams associated with each patient. To select an exam, first click on
the patient’s name. Then click on the specific exam you want to
archive. The exam will move from the middle box to the Export
Exams box located on the right.
Use the Move All button to select all exams for a given patient. To
“unselect” an exam, click on the exam in the Export Exams box.
When you are finished selecting exams to export, click on the Export
Now button.
5. The system exports the exams. A message box appears telling you
when the export has been completed. Click OK.
2. Click on the drive you want to import from. If necessary, insert the
media in the selected drive. Click OK.
3. A message box appears telling you how much room is available on the
media you will import to, and asking if you wish to continue. Click
Yes if there is enough room to import the exam(s).
Click No if there isn’t enough room. You’ll then have to archive other
exams to make room for the exam(s) you want to import.
5. The box on the left shows patient names. The middle box shows the
exams associated with each patient. To select an exam, first click on
the patient’s name. Then click on the specific exam you want to
import. The exam will move from the middle box to the right box,
Import Exams.
7. If you are finished importing exams, click Cancel. Then click Return
to return to the Utilities Menu.
MAINTAINING
THE DATABASE
To change information about patients, exams, doctors, groups, and
operators in your database, click on the Data button on the Utilities
The Database Maintenance screen lets you add, edit, delete, and save
information about each of these database categories.
Edit Patient Records
Edit patient records by clicking on the Patients button. The Edit
Patient Record screen appears.
This screen lets you select a patient, edit patient information, add a new
patient, or delete a patient’s record.
Select Patient
All patient names are listed in alphabetical order. Scroll up or down to
find the patient’s name by clicking on the arrow to the right of the
patient list, or “pulling down” the list by using the GlidePointer. Select
the patient you want by clicking on the patient’s name.
At this point, you may edit the patient information or delete the patient
record completely.
Edit Patient
Information
1. Enter or change the patient’s identification number by clicking in the
Patient # field. Then type the identification number.
2. If you want to change the patient’s name, click in the First Name or
Last Name field. Then type the new name.
4. Enter or change the patient’s birth date by clicking in the Birth field.
Then type the digits of the patient’s birth date in month-day-year
format.
You cannot enter a doctor’s name directly into the Docto r or Referred by field. If you
need to add a Doctor to the database, see “Edit Doctor Information” on page 5-18.
5. If you want to select or change the name in the Doctor field, click on
the drop-down arrow next to the doctor’s name. Then select the
doctor’s name from the drop-down menu.
7. After editing the patient information, you may select another patient
record to edit.
8. When you have finished editing patient records, click the Done
button. A message box appears.
2. Click the Delete button at the bottom of the screen. A message box
appears.
3. If you are sure you want to delete the patient record, click the Yes
button. The patient’s record and all examinations are removed from
the database.
The Edit Exam Record lets you select an exam, and change the Eye,
Group, and Exam Remarks (including Supplemental Remarks).
Select Patient
All patient names are listed in alphabetical order. Scroll up or down to
find the patient’s name by clicking on the arrow to the right of the
patient list, or by “pulling down” the list with the GlidePointer. Then
click on the patient’s name.
Select Exam
After you have selected a patient, the exams for that patient are
displayed in the Select Exam box, beginning with the most recent.
Select an exam to edit by clicking on the exam.
The exam number, exam date, and any information previously entered,
such as doctor and operator, are displayed.
Edit Exam Information
1. To change the Eye, click on the circle next to OD or OS.
2. If you want to change the Group from the current setting, click on
the drop-down arrow next to the group name. Then select the Group
from the drop-down menu.
You cannot enter a Group directly into this field. If you need to add a Group to the
database, see “Edit Group Information” on page 5-19.
Edit Doctor
Information
Edit doctor information by clicking on the Doctor button on the
Database Maintenance screen. The Edit Doctor Information screen
appears.
The Edit Doctor information screen lets you add, change, or delete a
doctor’s name. You may also designate the doctor’s association as
“Clinic” or “Referred.”
The Doctors who practice at the facility using this equipment should be
identified as “Clinic.” The doctors who have referred patients and sent
copies of exam files for import into this instrument should be identified
as “Referred.”
5. Click the Save button. This will associate the new doctor name with
all of the patients who were associated with the previous doctor’s
name.
Delete Doctor Name
1. Select the name of the doctor you want to delete by clicking on the
name in the doctor name list. Then click the Delete button. A
message box appears.
2. Click the Yes button. The doctor’s name is removed from the doctor
list and from any patient records and exams.
Caution: Delete permanently removes the doctor information from the database.
When you have finished editing doctor names, click the Done button to
return to the Database Maintenance screen.
Edit Group Information
2. Click in the Name field. Then type any changes to the group name.
4. Click the Save button. This will replace the old group with the new
group in all exams saved with the old group association.
Delete Group Name
1. Select the group you want to delete by clicking on the group name in
the Select Group field.
3. Click the Yes button. The group is removed from the group list and
from any exam records.
Caution: Delete permanently removes the group information from the database.
When you have finished editing groups, click on the Done button to
return to the Database Maintenance screen.
Edit Operator
Information
The Edit Operator Information screen lets you add, or change operator
names. These names are used to select the operator for each exam.
Add Operator
1. Enter the operator by clicking in the Name field and clicking on the
Add New button. Then type the operator’s name.
5. Click the Save button. The edited operator name replaces the original
operator name for all exams associated with the old operator name.
The operator names which have been labeled inactive will still appear in
database maintenance, but will not display when taking new exams.
CHANGE PREFERENCES
SCREEN
You can change the settings for general displays, and selection lists by
using the Change Preferences screen in the Utilities function.
OPERATING ENVIRONMENT
on the Main Menu. The Utilities screen appears.
1. If you want to change the Clinic Name, press and hold the left
GlidePoint or press the left control button and drag the cursor across
the name. Then press the Delete key on the keyboard. This clears the
name field.
2. Enter the new clinic name by typing the name in the Clinic Name
field.
Default Scale
Select the default scale you want to use in data displays. You may choose
either Standard, AutoSize or Custom by clicking on the circle next to
the scale you want.
To add or delete a name from these lists, use the Data Maintenance
option.
3. Follow the directions on the screen and in the software release notes
and instructions that come with the new software to complete the
OPERATING ENVIRONMENT
installation.
After you have set the date and time, click OK.
Note: It is possible to set up Windows so that the time and date may be entered and
displayed in European or “military” format, using a 24-hour clock. Please contact
your distributor or Humphrey Technical Support for assistance in changing the date
and time format.
CHANGING DISPLAYS
It is possible to view the ATLAS software in both 640 x 480 (ATLAS
LCD Screen) and 1024 x 768 resolution (external SVGA monitor). To
view the software in 1024 x 768 resolution, connect an external SVGA
monitor into the port marked VGA located on the side of the ATLAS.
The ATLAS lets you print a color copy of a patient exam in any view
using the color printer attached to the unit. You can also send an exam
to:
PRINTING
TO THE COLOR PRINTER
1. Before you print an exam, be sure the printer is turned on and the
“On Line” light is lit.
4. • If you are printing the Axial Map, you can print the curvature
values for a specific area of the display by moving the
GlidePointer to that area and pressing the space bar on the
keyboard once. The values and coordinates for the GlidePointer
position are “frozen” in the display. The Options screen then
appears.
PRINTING
Options
Cancel Cancel
8. In the Print Quality field, select either Draft for faster printing at
lower resolution, or Presentation for slower printing at higher
resolution. Presentation mode may not be available in some printers.
If this is the case, the presentation mode option will be grayed out on
the screen.
landscape orientation.
Note: The Portrait setting produces an 8 1/2” x 11” print. The image is smaller,
but the printing time is faster. The Landscape setting produces an 11” x 8 1/2”
print. The image is larger, but the printing time takes several minutes longer than the
Portrait setting.
PRINTING
TO A SLIDE FILE
To prepare color slides of an image, you will need to “save” the image
to a file on media such as a floppy disk. You can then send this file to a
photography service for processing into a color slide.
1. Display the patient exam and view you want to send to a slide file.
2. • You can “freeze” the curvature values for a specific area of the
display by moving the GlidePointer to that area and pressing the
space bar on the keyboard once. The values and coordinates for
the GlidePointer position will be set in the display. The Options
screen then appears.
Options
Cancel Cancel
PRINTING
3. Click the Print button. The Printer Setup screen appears.
7. In the Output File Type field, select the graphic file format you
want to use. The following graphic file formats can be produced:
9. A screen appears asking you to enter the drive and file name of the
slide file.
10. Insert a blank disk into drive a. Then click OK. The instrument
prints the slide file on the disk in drive a.
1. Display the patient exam and view you want to print to an image file.
Use the MasterVue Control and MultiVue Display options.
2. • If you are printing the Corneal Power Map, you can print the
curvature values for a specific area of the display by moving the
cursor to that area and pressing the space bar on the keyboard
once. The values and coordinates for the cursor position are
“frozen” in the display. The Options screen then appears.
Options
PRINTING
Print Print Screen
Cancel Cancel
7. In the Output File Type field, select the graphic file format you
want to use. Several options are available, and you should choose the
format that is compatible with the graphics software you will be
using.
9. A screen appears asking you to enter the drive and file name of the
image file.
Type the drive and file name in the following format:
a:\filename
9. a:\ designates the 3 1/2” disk drive on the instrument. The file name
can be up to eight characters long, with no spaces. The instrument
will enter the three-character extension corresponding to the file
format (i.e., BMP, TIF, PCX, or GIF). For example, exam004.bmp is
a valid file name.
Note: The only difference between an image file and a slide file is that the image file
has a white background by default, while the slide file has a black background by
default.
PRINTING
General
Troubleshooting
If you have problems with your Atlas instrument, you should call the
Humphrey Technical Support Line. However, before you call, you
should check the following items.
Symptoms/Solutions
1. When you are entering names or other information on the
keyboard, numbers appear instead of letters.
The Num Lock key on the keyboard may have been accidentally
pressed. When this key is pressed, some of the keyboard keys are
converted to numeric entry, rather than alphabetical characters. To
return to alphabetical entry, press the Num Lock key, which is located at
the top of the keypad on the right side of the keyboard. When Num
Lock is turned off the light below the words “Num Lock” will be unlit.
2. No video image appears on the screen.
Check cable connectors, call Technical Support if problem persists.
Troubleshooting
Zip Drive
10. Green power light does not come on, or the Zip disk cannot be
inserted.
Make sure you are using an iomega Zip power supply and that it is
connected to the Zip drive as shown in the installation guide. Make sure
the power supply is plugged into a working power outlet.
11. Instrument does not start up properly after the Zip drive is
installed.
Turn off the instrument and disconnect power from the Zip drive. Turn
on the instrument first, then immediately connect the Zip power (or use
a power strip to turn on both at the same time).
MAINTENANCE
Calibration Check
The calibration check tests the instrument to make sure it is providing
accurate measurements. A calibration bar with three mounted spheres is
placed in the chin rest assembly. Using the Calibration option in the
Utilities function, capture a series of sphere images that are analyzed to
see if the measurements are within the defined diopter ranges. The
calibration test is performed on 35, 42.25, and 55 diopter spheres.
We recommend that you run the calibration option at least once a month.
Mount Calibration Bar
Place the calibration bar in the area provided on the chin rest. Be sure
the spheres are facing the projection head.
Run Calibration
1. To run the calibration option, click Utilities on the Main Menu. Then
select the Test Calibration option by clicking the Test button.
5. When the image is aligned and focused, press the button on the
joystick to capture the image. A message screen appears listing the
results as one of the following:
• UNDER —
Image capture was under the acceptable diopter value.
• OVER —
Image capture was over the acceptable diopter value.
• PASS —
Image capture was within the acceptable diopter value.
8. When you have finished analyzing images for all of the spheres, a
message screen appears showing the final results of the calibration
test. You may print the results at this time. If the system fails the
calibration test, first try cleaning the calibration spheres with a soft
cloth. If your system fails again, call the Humphrey Technical
Support number for assistance.
Cleaning
The exterior of the ATLAS should be kept free of dust. Clean the
exterior with a dry cloth. If necessary, you can use a mild cleaning
solution, but be sure you do not allow moisture inside the instrument,
keyboard, printer, or projection head. In addition, DO NOT use a
cleaning solution on the faceplate or cone.
INTRODUCTION
MasterFit incorporates contact lens fitting capabilities and features
that are commonly used worldwide, as well as the inclusion of several
new features that should benefit all users of the software. Aspherical
fitting designs, using corneal eccentricity to fit the lens, and displaying
the final lens position on the cornea are just of few of the options
that this flexible and feature rich software package has to offer. The
software can be used to streamline the fitting process, decrease the
number of lens refits, and reduce the rejection rate of ordered lenses.
Also, it is so easy to use that it is not necessary for the doctor to inter-
act with the software after preprogramming the lens designs, nomo-
grams, lens style, and laboratories into the computer. The user is then
able to select the doctors choice of lens, simulate the fluorescein pat-
tern on the computer, and print the results, allowing the initial and
intermediary fitting steps to be assigned to the technician using the
software. After the doctor observes the fit of the trial lens chosen by
the software and performs the overrefraction, the final lens prescrip-
tion can be ordered from the lab by e-mail, greatly improving office
ACCESSING THE
MASTERFIT CONTACT LENS From the MasterVue Control Screen: (the Control Screen appears
MODULE after selecting REVIEW from the Main Menu):
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-1
REV. C PN 48113-1
FROM ANY MAP OR VIEW 1. Click on OPTIONS.
2. Click on MultiVue Displays.
3. Select MasterFit Contact Lens Module listed under
Optional Modules.
4. Click OK. This brings up the Fluorescein View
Screen.
FLUORESCEIN VIEW The Fluorescein View is the first screen in the MasterFit Contact Lens
SCREEN Module, and the main fitting screen.
Show
Fluorescein Click here to show the fluorescein pattern for this lens.
Show Details Click here to show the details of the fitting results.
Markers have also been added to the edge of the fluorescein pattern to
denote the default flat and steep axis as calculated by the software. The
9-2 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
FLUORESCEIN VIEW (FIG 1.)
Fitting Method Keratometric- uses information calculated from the “K” values
obtained in the 3mm zone combined with the manufacturers fitting
guide. If you have entered nomograms to alter the fit of the lens
based on corneal astigmatism, eccentricity, or flat or average K’s,
these will also be used in the final calculation of the lens fit.
Base Curve This box highlights the calculated base curve. Use the up/down arrows to
increase/decrease the base curve. The base curve and power are intricate-
ly related. Changing one parameter also makes a change to the other.
Each time you change the base curve you will need to recalculate the fluo-
rescein pattern by clicking on the “Show Fluorescein” button to observe
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-3
REV. C PN 48113-1
the new fit.
This box highlights the doctor preferred lens diameter for the selected
Lens Diameter lens design. Click on the up/down arrows to increase/decrease the
diameter. The diameter and optical zone are also related using the stan-
dard fitting relationship between these two parameters.
Increasing/decreasing the diameter also increases/decreases the optical
zone size. You will need to recalculate the fluorescein pattern after
changing these parameters to observe the new fit.
Select Exam Click here to select another patient or exam from the MasterVue
Control List.
Click here to access the Worksheet, which can be viewed as the contact
Worksheet
lens order form, showing the final fitting calculations. This screen can
be e-mailed to a contact lens lab for fitting instructions or printed for
filing in the chart.
Options Click here to access Preferences, Print Screen, Cancel, or Return to Main
Menu.
By clicking on the SHOW DETAILS button, (FIG. 2), you can see
the fitting sequence and methodology that the software chose to fit the
lens, and any pertinent comments about the parameters used to fit the
lens. You may also determine the success of the fit based on the nomo-
gram used.
9-4 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
SHOW DETAILS/FLUORESCEIN SCREEN (FIG. 2)
This screen will also show the position of the lens as it was fit on the
eye by the software. The software will automatically find the best posi-
tion for the lens to fit on the cornea, and display this position in a text
message located directly beneath the fluorescein pattern. This
lens/cornea relationship is determined from elevation data and the
The Axial Map (FIG. 3) can also be displayed along with the
Fluorescein View by clicking on the Axial button. This allows you to
see curvature changes on the map as they relate to the elevation values
on the tear film clearance plot or simulated fluorescein pattern.
Click here to access the Axial Map View. This screen combines the rec-
Axial Map
ommended contact lens fitting specifications, fluorescein view, clearance
plot and axial map on one screen.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-5
REV. C PN 48113-1
FLUORESCEIN/AXIAL MAP SCREEN (FIG. 3)
LENS DESIGNS/LENS The Lens Designs/Lens Types Screen (FIG. 4 ) allows you to pro-
TYPES SCREEN gram the software to fit the types of lenses that are common to a spe-
cific manufacturer or company. Most commonly used lenses can be
found in any of the popular publications of currently available lenses.
Other lesser known proprietary or protected lens designs may have to
be obtained from the companies themselves. You can access this screen
by clicking on the preferences screen from the Fluorescein View, and
then clicking on Lens Types. This screen allows you to program any type
of currently manufactured lens (spherical, aspherical, front toric, back
toric) into the software for future use. You will need to know the para-
meters of the lens and the restrictions that the manufacturer has put on
the lens before utilizing this function,( i.e., base curve, powers, diame-
ters, peripheral and secondary curves and their available incremental
steps).
9-6 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
LENS DESIGNS/TYPES SCREEN (FIG. 4)
MasterFit does require some initial data input in order to use the pre-
determined lens fitting types. So that the desired lenses may be input,
the user must know the manufacturer’s suggested recommendations and
the proper fitting methodology or be capable of making their own
selection for a specific type of lens. This only has to be done once in
order to complete the selection of lenses that you want to specify. It
This feature allows for specific types of lenses to be entered into the
Lens Types software based on the restrictions that have been placed on it by the
manufacturer of the material or lens. It is activated by clicking on the
Lens Types button. Front toric, back toric, bi-toric, and spherical
designs can be selected from this screen by clicking on the drop down
list box. This feature is very useful when programming lens types into
the software that are not currently offered in the default list.
The software allows for fitting aspheric lenses by the eccentricity value
that has been assigned to the lens. Once the aspheric box is
highlighted, the eccentricity can be entered into the form field. The
eccentricity can be adjusted by clicking on the up or down arrows to
increase or decrease the value. [The lens can then be categorized as a
front toric, back toric, or bi-toric according to the manufacturers
design.]The design can also be saved in millimeters or diopters
according to the operators choice by highlighting the box next to the
desired design.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-7
REV. C PN 48113-1
The diameter of the lens and the increments in mm in which it is manu-
factured can be set by clicking on the diameter drop down list box. If
the diameter that you wish to choose for a specific lens is not in the list
box, you can create a new diameter by clicking on the “New” button.
Clicking on the “Copy” button will copy the current lens diameter and
its associated nomogram, allow you to make any adjustments that are
necessary, and then rename the Lens Design. The “Clear” button will
clear all of the diameter input boxes and allow you to begin again.
“Delete” will delete the displayed lens diameters, “Save” will save the
current lens diameter, and “Cancel” will cancel the entered diameter.
These boxes are all located next to the diameter box itself and relate only
to changes made to the diameter of the lens.
The sphere and cylinder in diopters and the incremental steps can also
be input using the up/down arrow keys. This will set the dioptric sphere
and cylinder powers as well as the incremental steps that this lens is
manufactured. The optical zone diameter of the lens can be set in the
same way described above as can the flat and steep base curves in the
case of a toric fit.
The secondary and peripheral curves as well as their width can be pro-
grammed in this screen as well. The secondary and peripheral curves
may be absolute or relative. If these curves are absolute, then they may be
entered manually without having any relationship to the base curve or
lens diameter. Relative curves imply that the radius of curvature of the
secondary and peripheral curves are based on additions to the base
curve of the lens, and the width of these curves are a percentage of the
lens diameter. The mathematical calculation that takes place from these
input boxes will determine the final radius and width of the peripheral
and secondary curves. Once the lens type input is completed, save the
lens type by entering the name of the lens and clicking on “Save”. This
new lens will now become a permanent selection in the MasterFit
Module.
9-8 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
nomogram to fit to the flat or average “K”, according to the eccentricity
entered in the “From” and “To” boxes. This allows a nomogram to be
adjusted to fit the lens from the eccentricity of the cornea. If the
nomogram calls for adjustments to be made according to the astigma-
tism on the cornea, (see Lens Nomograms below), then both of these
nomograms will take affect. If only one type of fitting adjustment is
entered the nomogram will use only that adjustment. To access this
screen click on the Shape Adjust button from the Lens Designs screen.
LENS DESIGNS/LENS The Lens Designs/Lens Nomograms Screen (Fig. 6) permits the
NOMOGRAMS SCREEN user to adjust the manufacturers lookup table (fitting guide) to fit a lens
precisely according to their own preferences. The adjustments to the
base curve can be to the average or flat “K” relative to the corneal astig-
matism that is read by the topography software’s simulated K readings.
These lookup tables are already pre-programmed into the software for
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-9
REV. C PN 48113-1
LENS DESIGNS/NOMOGRAMS SCREEN (FIG. 6)
NOTE: When entering adjustments to the base curve in diopters, the sign of the
adjustment will be the opposite in mm. In other words, fitting steeper than “K” by
.25 diopter would cause the adjustment to be +.25, while fitting flatter than “K” by
.25 diopter would cause the adjustment to be -.25D. Conversely, fitting steeper
than “K” by 0.1mm would cause the adjustment to be -.01, while fitting flatter
than “K” by 0.1mm would cause the adjustment to be +0.1.
9-10 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
The secondary and peripheral curves and their associated widths can
also be input into the boxes in the same way. There are five rows of
data that can be input to adjust the base curve and secondary and
peripheral curves. You do not have to fill in all of the boxes, but if there
is astigmatism on the cornea that is not accounted for in the nomogram,
or no peripheral curve information is entered in the boxes, the software
will not be able to adjust the nomogram for this possibility. This new
fitting table will allow you to design your own nomograms based on
doctor preferences. When you are done entering the data, save the new
lens nomogram by entering a new lens design name in the entry box
labeled Lens Designs.
The following buttons are used to input new or edit existing lens designs:
New
Click here to enter a new lens design.
Delete This gives you the option to delete a lens design and associated fitting parame-
ters.
Save
Click here to save the lens design and associated fitting parameters.
The following are used to input new or edit existing lens diameters:
New
Click here to enter a new lens diameter.
Delete This gives you the option to delete a lens diameter and associated fitting para-
meters.
Save Click here to save the lens diameter and associated fitting parameters.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-11
REV. C PN 48113-1
5. Add one or more rows of lens fitting parameters. You
may use the keyboard to enter this information OR the
up/down arrows in each box. When using the key-
board, use the TAB key to move the cursor to the next
box.
LAB PREFERENCES Contact Lens Laboratory Preferences ( FIG. 7) can also be entered to
include the name, address, telephone, fax, and e-mail address of the
doctors favorite lab, contact lens distributor or other source where the
lenses are obtained, or the supplier of a particular brand of lens materi-
al. Once this information is entered, the technician can send the final
fitting information as well as the topographical map to the lab with a
few simple keystrokes from the Worksheet via e-mail using the
MasterNet Communications Software. The lab can then design the
lens from the e-mail message, or consult with the doctor, in the case of
the more difficult cases, to obtain the best fit for the patient using the
actual topography data.
9-12 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
LAB PREFERENCES SCREEN (FIG. 7)
You will need to become familiar with the Lab Preference Screen in
order to add information about the contact lens labs you deal with. To
view this screen from the Fluorescein View Screen, click on PREFER-
ENCES, then click on LABS, or from Doctor Preferences or Lens
Designs Screens, click on Lab Preferences.
DOCTOR PREFERENCES The Doctor Preferences Screen (FIG. 8) uses the Lens Designs,
SCREEN Lens Types, Lens Nomograms, and Lab Preferences programmed
into the system by the operator. Once you have customized the screens
mentioned above to reflect the lenses you prefer, you can choose one
lens design as the default selection for each doctor or fitter in your
office. Similarly, you can add the names of lens laboratories or distribu-
tors that you deal with and choose one as the default setting for each
doctor. You can also program other important settings that allow the
doctor’s preferences to be automatically selected for each fit.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-13
REV. C PN 48113-1
DOCTOR PREFERENCES SCREEN (FIG. 8)
9-14 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
6. Set the eye image overlay. The fluorescein
pattern can be “painted” over the image of the
eye with two settings: the translucent setting
will allow the eye image to be seen through the fluo-
rescein pattern, while the opaque setting will
not, however the surrounding eye anatomy can still be
seen in the captured image from both settings.
Turning the image overlay to the off position will
show the opaque fluorescein pattern only with no eye
image.
Once the default parameters have been set, save the changes by clicking
on the “Save” button. Click on “Done” to return to the Fluorescein
View. Clicking on the “Cancel” button will cancel all entries that you
have made to the Doctor Preferences.
TO ADD A NEW LENS
1. Click on NEW and enter the name of the new
LABORATORY:
laboratory.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-15
REV. C PN 48113-1
Sending the data via e-mail using the MasterNet Communications
Software is done from the WORKSHEET, (FIG.9).
The Worksheet shows the final lens fitting calculations that are obtained
from the fitting, including any trial lens information or remarks that the
lab may require. The Worksheet also allows you to e-mail a lens order
or consultation, including the topographical map, to your lab of choice.
Entering laboratory information in the form fields in the Lab
Preferences screen allows you to instantly default to the lab of choice,
or select another lab for transmission of the data.
9-16 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
GENERAL INSTRUCTIONS SCREEN (FIG. 10)
QUICK GUIDE TO USING 1) Select the patient and exam you wish to view from the
THIS MODULE: MasterVue Control Screen, and click on MasterFit
Contact Lens Module from the list box, or,
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-17
REV. C PN 48113-1
4) Select the type of nomogram that you wish to fit with,
Topographic or Keratometric. (If your selection is
Keratometric, you must also select the lens type and
design from the preprogrammed information you have
entered.)
9-18 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
blue ticks show the default position of the flat axis while the red ticks show the steep axis. This allows the
doctor to judge the fit of the chosen lens based on the simulated fluorescein pattern in any
meridian (axis) that is desired.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ MASTERFIT CONTACT LENS MODULE / 9-19
REV. C PN 48113-1
9-20 / MASTERFIT CONTACT LENS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
10 MASTERNET™ COMMUNICATIONS SOFTWARE
OVERVIEW
MasterNet™ allows you to send and receive patient files and e-mail with
other MasterVue™ software users anywhere in the world. If you have
any questions or problems contact the Technical Support Department at
Humphrey Systems at (800) 341-6968.
ACTIVATING
MASTERNET™
To activate the MasterNet software, you must first have a user ID and a
password. Choose a name for your user ID that fits the doctor or office
name, i.e.,
[email protected].
The ID must appear exactly in the format shown in lower case letters
and may consist of a maximum of eight letters. Be careful not to
confuse “o’s” and zeros as they are read differently by the software.
Once you have selected your user ID, notify Humphrey Technical
Support Department by mail or telephone.
Once Technical Support has been notified, they will supply you with a
password. This password is private to your office and should not be
shared with others. This will ensure that correspondence via e-mail and
patient exams are kept confidential.
Configuring e-mail
MASTERNET™
Note: If a line has to be selected first before hearing a dial tone, such as a multiple
phone line, MasterNet will not work from that line. MasterNet needs a phone line
like one used for a FAX machine.
PREPARING AND
SENDING PATIENT
EXAMS AND E-MAIL
Selecting exams
to send
5. Click on the exam(s) you wish to send. They will appear in the queue
box.
To move all exams associated with a patient into the queue box, click
Move All.
6. Click Queue E-Mail. The system now prepares the exams for
sending.
Attaching an
e-mail message
As soon as exams have been added to the queue a cover letter screen
will appear so that an e-mail message can be composed to attach to the
exams. To compose a message.
2. Move the cursor to the left of the desired name on the address line.
A small arrow will appear. When it does, click on it. the name will
highlight indicating it is chosen.
2. Type in the real name. Press Enter. The cursor will move to the next
line.
4. Click Done. The name and e-mail address will be added to the list.
5. Move the cursor to the left of the name on the address line. A small
arrow will appear. When it does, click on it. The name will highlight
MASTERNET™
indicating it is chosen.
1. Click Done.
2. Click on the name or names desired by moving the cursor to the left
of a name until the arrow appears and click. The selected names will
be highlighted.
3. If the person’s name does not appear on the address list add to the
list using Add a Name. Then select the new name by clicking on it.
4. Click Done.
2. Move the cursor to the message box. Type in the message (i.e.:
patient history).
Sending or
queuing mail
This is a good time to make sure the phone line is connected to the
phone jack on the side of the ATLAS (see Section 1) and into a data
line phone jack in the wall.
1. Click “Send Now”. Only the selected exam will be sent. The mail
will be encoded into a format that is unreadable by others as the red
bar moves across the screen.
2. Click Yes (if you are ready to have the modem dial). A dial tone will
be heard along with other noises and beeps coming from the modem
and the transmission will begin. It takes about 30-40 seconds per
exam to transmit the data to Humphrey via Masternet. When the
modem is done transmitting the line will disconnect and will return
back to the prepare e-mail screen.
1. Click Done.
3. Click on Send in the Send menu. The send mail queue will appear.
Exams which have not been sent will be highlighted in red.
1. Select the exam you wish to view by moving the cursor to the left of
the name desired. When the small arrow appears, click it. The line will
highlighted in blue to show it is selected.
2. Click View/Edit Selected Mail. The cover letter will appear. You
may edit it at this time. You may also print it by clicking Print.
1. Select the exam or exams you wish to send by moving the cursor to
the left of the name desired. When the small arrow appears, click it.
The line will be highlighted in blue to show it is selected.
MASTERNET™
2. Click Send Selected Mail. Only the selected exam will be sent.
The mail will be encoded into a format that is unreadable by others as
the red bar moves across the screen.
Note: You cannot send mail that has been previously sent. If you select sent mail you
will be informed.
3. Click Yes (if you are ready to have the modem dial). A dial tone will
1. Click Send All Mail. All mail in the queue will be encoded into a
format that is unreadable by others as the red bar moves across the
screen.
2. Click Yes (if you are ready to have the modem dial). A dial tone will
be heard along with other noises and beeps coming from the modem
and the transmission will begin. It takes about 30-40 seconds per
exam to transmit the data to masternet.com. When the modem is
done transmitting the line will disconnect and will return back to the
Mail Queue screen. The computer will advise how many pieces of
mail have been sent.
Sending an e-mail
message without an
exam attached
To communicate via e-mail with other colleagues without attaching an
exam.
To retrieve e-mail
1. Click on MasterNet™ in the Main Menu
3. Click on Yes (if you are prepared for the modem to dial). A dial tone
will be heard along with other noises and beeps coming from the
modem and the transmission will begin. It takes about 30-40 seconds
per exam to transmit the data from masternet.com. When the
modem is done transmitting the line will disconnect and will return
back to the Received E-Mail screen. The computer will advise how
many pieces of mail have been received.
Before exams can be viewed they must be imported. Mail which has
been retrieved but not imported is highlighted in blue. You need not
import all the exams you have retrieved at once but can import selected
exams.
You may wish to view the cover letter of a retrieved piece of e-mail to MASTERNET™
see if you wish to import it. Often there is more than one patient exam
included in a piece of e-mail. By viewing the cover letter you can see
how many exams are included and which patients they are for.
1. Select the piece of mail you wish to view by moving the cursor to the
left of the name desired. When the small arrow appears, click it. The
line will be highlighted in blue to show it is selected.
4. You will find the list of exams included at the bottom of the text
portion of the cover letter.
To import
selected exams
1. Select the piece or pieces of mail you wish to import by moving the
cursor to the left of the name desired. When the small arrow appears,
click it. The line will be highlighted in blue to show it is selected.
3. Sometime not all the files associated with the patient’s exam have
been sent to you. Often the bit map (.BMP), the video image of the
eye itself, is not sent to save transmission time. If a file has not been
sent, a dialog box will inform you. If you still want to import the
exam, click Yes. If you wish to import all the selected exams
regardless if files are missing, Click Yes to All.
To import all
retrieved exams
1. Click Import All Exams. All retrieved pieces of mail will be
imported and their printing will change from blue to black to show
this.
2. Sometime not all the files associated with the patient’s exam have
been sent to you. Often the bit map (.BMP), the video image of the
eye, is not sent to save transmission time. If a file has not been sent, a
VIEWING AND
PROCESSING MAIL
1. Click on MasterNet™ in the Main Menu
To view the
cover letter of an exam
You may wish to view the cover letter of a piece of mail in the mail MASTERNET™
queue log.
1. Select the piece of mail you wish to view by moving the cursor to the
left of the desired name. When the small arrow appears, click it. The
line will be highlighted in blue to show it is selected.
2. Click View/Edit Selected Mail. The cover letter will appear. You
may also print it by clicking Print.
After you have sent queued mail the mail will still remain on the Mail
Queue log. When you wish to remove mail from the log:
1. Select the mail you wish to delete by moving the cursor to the left of
the desired name. When the small arrow appears, click it. The line will
be highlighted in blue to show it is selected. Continue until you have
selected all the pieces of mail you wish to delete.
2. Click Delete Selected Exams. The exam name will no longer appear
on the log.
To view the
cover letter of an exam
You may wish to view the cover letter of a piece of mail in the received
mail log.
1. Select the mail you wish to view by moving the cursor to the left of
the desired name. When the small arrow appears, click it. The line will
be highlighted in blue to show it is selected.
2. Click View /Edit Selected Mail. The cover letter will appear.
3. If the name of the sender is not currently in the address list a dialog
box will appear and ask you if you wish to add the name and address.
If you do, click yes.
4. You will find the list of exams included at the bottom of the text
portion of the cover letter.
To import exams
You may have chosen to retrieve mail but did not chose to import the
exams at the time. When you wish to import an exam so that you can
view it and work with it on your ATLAS system.
1. Select the mail you wish to import by moving the cursor to the left of
the desired name. When the small arrow appears, click it. The line will
be highlighted in blue to show it is selected.
3. Sometime not all the files associated with the patient’s exam have
been sent to you. Often the bitmap (.BMP), the video image of the
eye itself, is not sent to save transmission time. If a file has not been
sent, a dialog box will inform you. If you still want to import the
exam, click Yes. If you wish to import all the selected exams
regardless if files are missing, Click Yes to All.
Even after exams are imported into the ATLAS data base the record
will remain on the Received E-Mail log. From time to time you will wish
to remove names from the log.
1. Select the mail you wish to delete by moving the cursor to the left of
the desired name. When the small arrow appears, click it. The line will
be highlighted in blue to show it is selected.
Caution: If mail is deleted from the Received Mail log before the exams are
imported the exams will be lost. They cannot be retrieved again from masternet.com
because when they are first retrieved they are removed from the mail server memory.
CONNECTING TO
TECHNICAL SUPPORT
This function allows the Technical Support Department at Humphrey
Instruments to connect to your computer via the modem.
1. Make sure the phone line is connected to the phone jack on the side
of the ATLAS (see Section 1) and into a data line phone jack in the
wall.
The axial map is a simple way of describing the overall shape of the
cornea. This curvature or shape map may obscure small details but will
give a gross approximation of the corneal curvature. Axial maps
assume the cornea is a spherical shape instead of a true asphere, and
therefore incur more error in the periphery as a rule. Even though
values are given in diopters or millimeters, the axial map is actually a
calculation of curvature values instead of power values. Although these
two values are intimately related, there is not a one to one
correspondence. There may be a difference in the power values that are
shown for an axial map and a refractive power map on the same patient.
This is also true when comparing an axial map with a tangential map.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ ADVANCED REFRACTIVE DIAGNOSTICS MODULE / 11-1
REV. C PN 48113-1
in its refractive state measured in diopters. This map uses the arc step
algorithm reconstruction, along with Snell’s Law, to measure the
refractive nature of light rays as they pass through the corneal surface.
This calculation takes into account spherical abberiation and more
closely matches the true dioptric power of the cornea.
When displaying elevation maps, a best fit spherical surface that most
closely matches the corneal surface is used for comparison. The
software will choose which surface will be the best fit for a particular
eye, and this will vary from patient to patient. The computer then
attempts to “fit”, or superimpose, this sphere onto the corneal surface.
Elevation is described as the difference between the best fit sphere and
the actual corneal surface measured in microns. Elevation values can be
positive or negative numbers. The reason for this is that the corneal
surface can fall “above” or “below” the best fit reference sphere. The
best way to describe this map would be to imagine a cross-section of
the cornea, similar to a profile view, and how the corneal surface might
pass through, fall below, or rise above a reference surface if it were
plotted out on a graph.
11-2 / ADVANCED REFRACTIVE DIAGNOSTICS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
ADVANCED REFRACTIVE DIAGNOSTICS MODULE
the map. This may not always be true, however, as very high
astigmatism (red color) on an axial map may show up as a negative
number on an elevation map. This is due to the fact that the corneal
surface is “falling away” or sloping at a faster rate away from the
reference surface than the rest of the cornea. Therefore it never
“crosses through” the profile of the reference surface as described
above. This can create confusion as clinicians are used to looking at the
hotter colors as “steeper” on the axial map and the cooler colors as
“flatter”. One must remember that in elevation maps the measurement
is an elevation “difference” from a best fit sphere, and the axial and
elevation maps cannot be compared point to point in this manner.
Wavefront error describes the condition of the light rays as they pass
through the corneal surface and are refracted onto the retina through
the clear media and crystalline lens in the human eye. The greater the
concentration of wavefront error, the more the light is not refracted
properly onto the retina, which can result in aberrations- often described
by patients as ghosting, halos, or glare- even though the patient may be
able to read the eye chart fairly well. This is usually described as
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ ADVANCED REFRACTIVE DIAGNOSTICS MODULE / 11-3
REV. C PN 48113-1
“quality of vision” and can cause great visual discomfort to the patient
even though they may see well using conventional vision testing
methods.
1) Insert the diskette into the proper drive on the computer, then from
the Main Menu, click on Utilities, then Install New Software.
TO INSTALL THIS MODULE 2) Follow the onscreen instructions, clicking on “yes” or “OK” to the
system prompts. The software will load after the system has rebooted.
1) From the MasterVue Control screen, click on the patient and exam
that you wish to view, then click on MultiVue, or
TO ACCESS THIS MODULE 2) From any single view display, click on options, then MultiVue Display,
then OK.
4) If you wish to change the order of the views, the exam, or patient
you have selected, simply click on the view (map) you wish to change,
and from the MasterVue Control screen select the new patient, exam, or
view that you wish to change to. After clicking OK, your new selection
will appear.
5) Only one patient, exam and eye may be viewed at a time in the
Advanced Refractive Diagnostics Module. If this module is “greyed
out” in your software, then the software has not been loaded onto the
instrument. This must be done in order to view this module. For more
information on installing the software onto the instrument, please refer
to the instructions for installing new software.
11-4 / ADVANCED REFRACTIVE DIAGNOSTICS MODULE HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
12 SIMULATED ABLATION MODULE
These known values can be input into the data boxes and calculated to
obtain a simulation of what a spherical (PRK) or astigmatic (PTK) laser
ablation might look like on the eye even before the surgery is ever done.
This tool can then be used for patient education, pre-surgical
information gathering purposes, or to compare pre-op and simulated
post-op results to determine the final result of the surgery. Another
useful function of the Simulated Ablation Module is to compare the
“perfect” ablation that the computer simulates with the actual ablation
itself to monitor the homogeneity and uniformity of the ablation
created on the eye by the excimer laser.
Once the data is input into the computer, the operator can determine
the depth of the proposed ablation in microns, or manipulate the
optical zone of the ablation for better centering over the visual axis to
account for pupil decentration. You may also increase or decrease the
size of the pupil in its dim light (dilated) state to determine interaction
with the edge of the ablation zone. Interaction of the pupil with the
ablation zone edge often accounts for patient complaints of ghost
images, and glare and halos from oncoming headlights while driving at
night.
3) The system will reboot, and the software module will then load. You
must install Revision A5 first in order to install or view this module.
TO ACCESS THIS
MODULE
1) From the MasterVue Control screen, click on the patient and exam
that you desire to view.
2) Click on Simulated Ablation from the Views list box, then click OK.
3) The Simulated Ablation display will appear showing the axial map,
and the input data boxes.
5) Enter the desired post-op refraction that you wish the patient to have.
6) Manipulate the ablation zone over the visual axis by clicking on the
Horizontal or Vertical arrow buttons to decenter the ablation to the
desired position. (the default position will be the corneal vertex)
7) Adjust the pupil size to the desired diameter by clicking on the arrow
buttons to increase or decrease the size of the pupil.
10) To view the final parameters after the ablation has been performed,
click on Parameters, and the results will be displayed on the screen with
the simulation.
12) Click on Simulated Ablation Difference, then OK, and the pre-op
and simulated “post-op” maps will appear with the difference in
power/curvature calculated below in the difference map.
13) If this module is “greyed out” in your software then it has not been
loaded onto the system software. To load this software, please see the
installation instructions.
The three statistical indices that PathFinder uses to gauge the shape,
regularity, and toricity of the corneal surface are Corneal Irregularity
Measurement (CIM), Shape Factor (SF), and Mean Toric
Keratometry(TKM). Isolating and identifying normal versus
abnormalpopulations using these three parameters by examining
hundreds oftopography exams has been done in order to achieve the
high level ofsensitivity and specificity that the software is able to
accomplish.
STATISTICAL INDICES AND Shape Factor (SF) is a measure of the asphericity of the cornea, and
NORMAL POPULATIONS aderivative of eccentricity, which is a well known calculation of
cornealshape used by contact lens fitters. Shape factor can be used
todetermine whether the corneal is more oval or elliptical shaped,
byassigning a factor, or index, to represent the shape of that surface.
Shape factors are unique and different from eccentricity in that it is
possible to calculate a negative, or oblate shape, as well as a positive,
or prolate shape. The less spherical, or more elliptical the cornea is in
the horizontal meridian, the more the cornea will resemble a prolate
shape. One could imagine this shape being like an egg turned on its
pointed end, as in this example, where the center of the cornea is
much steeper in curvature than the periphery. Highly positive or
prolate shape factors may imply that a pathology such as keratoconus
may exist. The less spherical, or more elliptical the cornea is in the
vertical meridian indicates an oblate shape. This shape would be the
reverse of the egg on it's pointed end, and would resemble an egg
lying on its side, such as this .
Normal Shape Factors usually range from .13 - .35 in the normal
population but can be in the .2 - .3 range. When the shape factor
reaches .7 and above, the cornea begins to exhibit a more conical
shape, and usually indicates some sort of pathology or abnormal
shape.
The mathematical
2
formula for shape factor is:
Ep = e = 1 - pp where Ep is a prolate shape
and
2
Eo = _-e _
2
1-e where Eo is an oblate shape.
67% of the population falls in the normal range. The infrared Models
992 & 995 have more noise, which accounts for the difference in the
ranges. A bell shaped distribution curve is not seen due to the fact
that no negative numbers can be calculated using CIM, resulting in a
skewed distribution plot.
The distribution appears as a bell shaped curve where the mean value
for TKM is 44.5 and 96% of the population falls between 41.25 and
47.25 diopters.
Definition of Conditions Isolating and identifying normal versus abnormal populations has
Identified by PathFinder been determined using the three statistical indices (CIM, SF, MTK) in
the examination of hundreds of topography exams. These exams act
as a control group for the software to identify when a cornea is
normal, just as when there is an abnormal condition.
To Access the PathFinder PathFinder can be accessed two ways within the software. From the
Corneal Analysis Main Menu screen, click on "Review", then select the patient and
exam that you wish to evaluate. Select PathFinder Corneal Analysis
from the drop down “Views” list box. Click on "OK" and the
software will run and instantly provide the analysis using the standard
Axial Map. The software can also be run from the MultiVue
Displays/Optional Modules Screen. By clicking on the entry labeled
"PathFinder Corneal Analysis" after selecting the desired patient and
exam, the software will display the findings. Scaling can be changed
from .25 diopter to .50 diopter increments simply by clicking on the
AutoSize or Standard buttons respectively. To print the analysis,
return to the Main Menu, or cancel the selection, click on Options,
and follow the onscreen prompts.
NORMAL RESULT
EXPLANATION SCREENS By placing the pointer on any of the colored bars which show the
statistical index results and clicking once, you can access the
explanation screens which show the normal population distribution
curve and a short explanation of that particular index. This
explanation also includes the normal ranges and a general statement
as to what abnormal readings for this index might indicate. This
screen will remain the same for each patient as it is designed to give a
reference to the user to show where a particular patient might fall in
the normal population distribution. This screen also serves to help
educate both the user and the patient as to the meaning of each of the
descriptors used in the PathFinder Corneal Analysis.
THE PROBABILITY DISTRIBUTION SCREEN FOR NORMAL TKM STATISTICAL INDEX (FIG. 5)
THE PROBABILITY DISTRIBUTION SCREEN FOR SHAPE FACTOR STATISTICAL INDEX (FIG. 7)
CONCLUSIONS The PathFinder Corneal Analysis software can be a very useful tool
for confirming a diagnosis of keratoconus, uncovering an "occult" or
subclinical keratoconus, distinguishing contact lens distortion from
true keratoconus, pre-screening prior to refractive surgery, and
establishing a normal baseline topographical map of the corneal
surface. Revealing these conditions at an earlier stage may allow the
practitioner to recommend another course of treatment, suggest
alternative methods of visual rehabilitation, or simply change an ill
fitting contact lens. PathFinder is simple, quick, and easy to use, and
SPECIFICATIONS
MODEL 991 MODEL 992
DIMENSIONS
WORKING DISTANCE
70mm 70mm
FIELD OF VIEW
11.4mm 12.5mm
NUMBER OF RINGS
20 22 (18 superiorly)
DIOPTRIC RANGE
9 to 108 diopters 9 to 108 diopters
OPTICS
REPEATABILITY, TEST
OBJECTS*
± .10 diopters ± .10 diopters
REPEATABILITY, NORMAL
CORNEAS
± .25 diopters ± .25 diopters
LUMINOSITY
Visible Infrared
ELECTRICAL
REQUIREMENTS
100/120/220/240/ 100/120/220/240/
volts AC ± 10% volts AC ± 10%
VOLTAGE
100-120/220-240 VAC
CURRENT
2A MAX/1A MAX
FREQUENCY
50-60 HZ
FUSE
T3.15 A, 250 V, Slow Blow
WARRANTY
For one year from the date of delivery to the original purchaser, Seller
warrants its ATLAS (“the instrument”) to be free from defects in
material or workmanship. In the event of failure, Seller’s obligation is
limited to repairing or replacing on an exchange basis parts which have
been promptly reported by the Purchaser during such one-year period as
being defective and are so found by Seller upon inspection.
Warranty Procedure
The procedure for warranty claims shall be as follows:
2. This warranty covers all parts, labor, travel, and expenses for the
warranty period. At the Seller’s discretion, the repair of the ATLAS
may be made in the Seller’s Repair Depot. In this case, all shipping
costs will be paid by the Seller unless the Purchaser’s ATLAS is found
upon inspection not to be eligible for repair under the warranty, in
which case the Purchaser shall be responsible for one half the
shipping costs.
Exclusions
This warranty does not cover consumable items such as operating
supplies, paper, printer ink cartridges, disk media and will not apply if
repair or parts replacement is required because of accident, neglect,
misuse, transportation or causes other than ordinary use, or by supplies
that do not meet the proper operating specifications of the Seller. The
warranty does not apply if the configuration of the computer has been
altered, or if software not supplied by the Seller has been used. This
warranty also does not apply to any articles that have been repaired or
altered except by the Seller.
All data stored on the hard disk, external zip drive and disk media are
the Purchaser’s records, and it is his or her responsibility to preserve the
integrity of the files. Seller is not responsible for the loss of patient files
stored on the hard disk, external drive, or disk media. Customers are
responsible for restoring their own data after a hard disk has been
repaired.
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ WARRANTY AND SERVICE CONTRACT / 16-1
REV. C PN 48113-1
The entire risk as to the quality and performance of the software is with
the Purchaser. The Seller does not warrant that the software will meet
the Purchaser’s requirements, that the operation of the software will be
uninterrupted or error-free, or that all software errors will be corrected.
The Purchaser assumes the responsibility for the selection of the
programs and hardware to achieve the Purchaser’s intended results, and
for the installation, use, and results obtained from the instruments and
programs.
This warranty does not extend to any disk media that has been damaged
as a result of accident, misuse, abuse, or as a result of service or
modification by anyone other that the Seller. Should such software
prove defective following its purchase, the Purchaser (not the Seller)
assumes the entire cost of all necessary service, repair, or correction.
The Seller shall have no liability or responsibility to the original
Purchaser or to any other person or entity with respect to any claim,
loss, liability, or damage caused or alleged to be caused directly or
indirectly by any software supplied by the Seller.
Every reasonable effort has been made to ensure that the product
manual and promotional materials accurately describe the instrument’s
specifications and capabilities at the time of publication. However,
because of ongoing improvements and product updates, the Seller
cannot guarantee the accuracy of printed materials after the date of
publication and disclaims liability for changes, errors, or omissions.
Seller shall in no event be liable to the Purchaser for the loss of profit,
loss of use or consequential damages. Purchaser agrees that Seller will
not be liable for any damages caused by Purchaser’s failure to fulfill the
Purchaser’s responsibilities as to proper installation, use, management,
and supervision of the instrument. This warranty shall only apply to the
original Purchaser and shall not, in any way, be transferable or
assignable.
16-2 / WARRANTY AND SERVICE CONTRACT HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
MERCHANTABILITY OR WARRANTY OF FITNESS FOR A
PARTICULAR PURPOSE. (ALL INSTRUMENT SPECIFICATIONS
ARE SUBJECT TO CHANGE WITHOUT NOTICE.)
SERVICE CONTRACT
A Warranty Extension Agreement (Service Contract) is available after
the one-year warranty period expires. This Warranty Extension is for
one year and is subject to the Terms and Conditions applicable to the
HUMPHREY ATLAS™ AND ATLAS ECLIPSE™ WARRANTY AND SERVICE CONTRACT / 16-3
REV. C PN 48113-1
16-4 / WARRANTY AND SERVICE CONTRACT HUMPHREY ATLAS™ AND ATLAS ECLIPSE™
REV. C PN 48113-1
Appendix GLOSSARY OF TERMS
Alternate View
A data display other than the standard Corneal Power Map (e.g.,
Numerical View, Photokeratoscope View).
Archive
Process of removing exams from the hard disk and saving them on an
external drive or disk media; used for freeing hard drive space on the
computer.
Backup
Process of using high-capacity media to create a complete duplicate of
all information on the hard disk.
Button
An object displayed on the media screen that can be selected with the
GlidePoint and activated by clicking.
Calibration Bar
Frame with three mounted spheres that is placed in the chin rest
assembly in order to capture images used to calibrate the system.
Calibration Test
Process of checking the ATLAS system to make sure it is providing
accurate measurements.
Chin Rest
Assembly that cradles the patient’s chin and forehead during an
examination.
Clicking
To quickly press and release a button on the GlidePoint while the
GlidePoint is positioned over a selection on the screen.
Computer
GLOSSARY
ATLAS instrument component that contains the central processing unit,
random-access memory, and disk drive.
Cursor
Blinking bar shape on the computer screen that shows where the next
character you type will appear.
Database
A complete record of all patient and exam information in the
instrument. The database is stored on the instrument’s hard disk.
Default Settings
Settings that first appear in Selection Lists when new records are added
to the system. These settings can be changed via the Change
Preferences option on the Utilities Menu.
Difference Display
A MultiVue display that shows the currently selected exam, the previous
exam for the same eye, and the dioptric power differences between the
two.
Disk
Standard IBM-compatible format 3½” disk used to export and archive
small numbers of patient exams.
Drop-Down Menu
A list of choices that drops down from an area of the screen when
activated by the GlidePoint control buttons.
Ethernet
Allows you to network the ATLAS with different computers, printers or
office management systems to share information and exams.
Export
Process of copying exams from the hard drive to a Floptical disk or
floppy diskette; typically used to transfer exams from one instrument to
another.
Field
A portion of a screen display that contains information such as a
patient’s name or date of examination.
Format
Process of electronically preparing a disk media so that it can be used to
store data.
Full Exam
A comprehensive eye examination that captures up to four different
images of a patient’s eye and allows the operator to select the best
image.
GlidePoint
Pad used to move a pointer around the screen and make selections by
“clicking” one of the control buttons located on the two upper corners.
Group
A user-defined category attached to exams in order to select them for
retrieval at a later date.
Import
Process of copying exams from a Floptical disk or floppy diskette to the
hard drive; typically used to transfer exams from one instrument to
another.
GLOSSARY
Joystick
Device used to focus the image of a patient’s eye and trigger the image
capture routine.
Joystick Lock
Control located in the joystick locks the projection head in place while
the instrument is being moved.
Keratometric Fit
In the optional Contact Lens Module, a method of fitting lenses that uses
nomograms published by contact lens manufacturers; lens parameters are
defined by using only the central three millimeters of the cornea.
Keyboard
Typewriter-like device used to enter information such as names, birth
dates, and examination remarks.
Landscape Orientation
An image printed across the wide axis of a sheet of paper, i.e., 11” wide
by 8½” long.
Main Menu
The MasterVue screen display that first appears when the system is
turned on.
MasterVue Control
Screen that lets you select the patient and exam to be reviewed, as well
as the type of view to be displayed (e.g., Corneal Power Map or
Numerical View).
Media
Standard Media (disk, diskette, IBM Compatible) used to export, backup
and archive patient exams.
Message Box
A small box that “pops” on top of the computer screen display to tell
you important information.
MultiVue Display
Display that lets you compare more than one exam on the screen at the
same time.
Numerical View
Data display that shows the dioptric values for up to five circular areas,
or zones, on the cornea from 0 mm to 10 mm.
Photokeratoscope View
Data display that shows the actual image of the eye as captured by the
ATLAS system.
Portrait Orientation
An image printed across the narrow axis of a sheet of paper, i.e., 8½”
wide by 11” long.
Printer
Device used to produce color printouts of patient exams.
Projection Head
ATLAS instrument component that projects rings and houses the video
cameras used to capture images of a patient’s eye.
QuickVue Exam
MasterVue screening tool used for high-volume or routine diagnostics
that require only a single image capture.
Restore
Process of placing archived exams back on the hard disk drive.
Selection List
A list of choices, such as patient names or data views, that appear on
the screen; choices can be made by positioning the arrow over a
selection and clicking the GlidePoint button.
Topographic Fit
GLOSSARY
In the optional Contact lens Module, a proprietary method of fitting
lenses that uses the entire curvature of the cornea as drawn from the
ATLAS topographical map of the patient’s eye.
Topographical
A display that graphically represents three-dimensional features.
Utilities
Collection of system “housekeeping” functions such as controlling the
appearance of screen displays, and setting the date and time.