The SK-3502 Digital Insulation Tester is designed for insulation resistance and low resistance measurements with various test voltages ranging from 50V to 1000V. It features an auto-ranging display, multiple measurement functions, and safety precautions to prevent electrical shock hazards. The device includes various keys for measurement control, data memory, and display management, ensuring accurate and safe operation.

Uploaded by

George Gamal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
11 views

e_instruction_sk3500_3502_2

The SK-3502 Digital Insulation Tester is designed for insulation resistance and low resistance measurements with various test voltages ranging from 50V to 1000V. It features an auto-ranging display, multiple measurement functions, and safety precautions to prevent electrical shock hazards. The device includes various keys for measurement control, data memory, and display management, ensuring accurate and safe operation.

Uploaded by

George Gamal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

DIGITAL 1-2.

Display Range
Rated Test Voltage Display Range (Auto-ranging) Resolution
4-2. FUNCTION Switch
Turns the power on and selects measurement functions. After measurement, turn it to "OFF".
INSULATION TESTER
SK-3502 DIGITAL INSULATION TESTER

50V / 125V / 250V 4.000M 0.001M


High
Voltage

Alligator 4-3. MEASURE Key


500V / 1000V 40.00M 0.01M Clip Insulation Resistance and Low Resistance Measurements : Hold down this key
SK-3500 / SK-3502

Kaise
50V / 125V 200.0M 0.1M to start measurement and release it to stop measurement. Displayed values are

CE
TIMER ON 1000V
1000V PRESS
2000M
2000M 1000V & 500V
LIGHT
500V
Test Lead held automatically for 5 seconds ("DH" lights up).
400.0M 0.1M
2000M
MEMO
250V

Instruction Manual
SELECT

250V / 500V / 1000V


HIGH 2000M
OFF LOW
125V

Voltage Measurement : Press this key while measurement to hold LCD indications
MEASURE
200M

2000M 1M
ON
COMP
AC/DC V
600V MAX

DH
OFF
OFF

fig. 1 ("DH" lights up). To cancel display hold, press this key again.
5 SEC
LOW
Low 0 ADJ

1-3. Rated Measurement Range and Accuracy


FOR SAFETY MEASUREMENTS!! In case of live-line measurement, strictly observe the warnings below (refer to fig 2.) 4-4. TIMER Key : Timer Function (Continuous Measurement)
Rated Test Voltage Name of Measurement Range Measurement Range Accuracy Press this key for 3-minutes continuous measurement. Usable in Insulation
Do not hold Insulation Tester in your hands. Keep safety distance from power source
To prevent an electrical shock hazard to the operator and/or damage to the instruments, 1st effective range 0.200 to 10M 3rdg4dgt or circuit to be measured not to touch the dangerous voltage. Resistance Measurement and Low Resistance Measurement.
read this instruction manual carefully before using the Insulation Tester. WARNINGS with 50V / 125V 0 to 0.199M Black test lead : Attach black alligator clip and connect to (earth) side of the circuit. To cancel it : Press MEASURE Key during continuous measurement.
the symbol on the Insulation Tester and this instruction manual are highly important. 2nd effective range 5rdg5dgt
10.01 to 200.0M Red test lead : Hold with one hand and connect to + (positive) side of the circuit. 4-5. SELECT Key :
1st effective range 0.200 to 100M 3rdg4dgt
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution 250V / 500V Press this key to switch year, date and time display while viewing memorized data.
0 to 0.199M
(refer to instruction manual)". / 1000V 2nd effective range 5rdg5dgt Red Test Lead
4-6. Safety Lock Key : Sefety Lock Function
100.1 to 2000M
SK-3502 DIGITAL INSULATION TESTER
The key to protect the operator from high voltage accidents. To select 500V or
The symbol in this manual advises the user of an electrical 1-4. Measurement Terminal Voltage Characteristic 1000V range, turn FUNCTION Switch holding down this key.
High
Voltage

WARNING
shock hazard that could result in serious injury or even death. Minimum Resistance Alligator
Rated Torelance of Measurement Measurement Values to
Rated 4-7. 0 ADJ Key : Zero-ohm Adjustment

Kaise
Test Voltage Circuit Terminal Voltage Test Current TIMER ON 1000V
1000V PRESS
Clip
Maintain Rated Test Voltage

CE
1000V & 500V

The symbol in this manual advises the user of an electrical Usable in Low Resistance Measurement. LCD indication is adjustable into zero
LIGHT 2000M
2000M
500V
2000M
MEMO

CAUTION
250V

50V 0.05M
SELECT HIGH 2000M
OFF LOW
125V

shock hazard that could cause injury or material damages. by pressing this key with test leads shorted. ("0 ADJ " lights up)
MEASURE
200M
ON
COMP
AC/DC V
600V MAX

125V 1 to 1.2 times 0.125M


DH
OFF
OFF
5 SEC
LOW
Low 0 ADJ

250V of Rated 0.25M


fig. 2 4-8. LIGHT Key : Backlight
1 to 1.2mA
WARNING WARNING 5. Correct Selection of FUNCTION Switch Turns on and off the LCD backlight. The light automatically turned off after 30 seconds.
500V Test Voltage 0.5M
Measurement on High Power Line is very dangerous. It sometimes includes High Surge 1000V 1M Always confirm that FUNCTION Switch is set to the correct position. Do not 4-9. COMP Key : Comparator Function
Voltage that could cause dangerous arcs of explosive short in the instrument and could result measure voltage except for AC/DC V position. Activates Comparator Function in Insulation Resistance Measurement and Low
Short Circuit Current : 2mA or less
in serious injury to the operator. For dangerous voltage measurement on High Power Line or WARNING 6. Maximum Input Observance Resistance Measurement. For details, refer to 6-1. Comparator Function .
High Voltage Circuit, always keep the instrument away from your body without holding it in SK-3500/3502 Measurement Terminal Voltage Characteristic Do not measure voltage or current that exceed the specified maximum input values.
4-10. MEMO Key : Data Memory Function
your hands. Do not touch the insulation Tester, its Test Leads, and any part of the circuit. (V)
WARNING 7. Test Lead Detachment Press this key to memorize measuring data in Insulation Resistance Measurement and Low Resistance
1200
1100 1000V Range Detach test leads from measuring circuit before switching measurement functions Measurement or to view memorized data. For details, refer to 6-2. Data Memory Function .
or removing battery cover for battery and fuse replacement.
1. UNPACKING AND INSPECTIONS 900
4-11. HIGH / LOW Key
4-2. GENERAL WARNINGS AND CAUTIONS Usable in Comparator Function to set buzzer timing.
Inspect the instrument and acessories for transport damage. If there is any damage or missing items, WARNING 1. Children and the persons who do not have enough knowledge
ask your local dealer for replacement. Confirm that the following items are contained in the package. 700
about electric measurements must not use this instrument. 4-12. UP-DOWN Keys
Measurement Usable for various settings such as to set threshold in Comparator Function or to
1. Digital Insulation Tester 1 pce. 4. Carrying Case (1016) 1 pce. 500V Range WARNING 2. Do not measure the electricity naked or barefooted to protect
Terminal 500
2. Test Lead (100-61) 1 set 5. Batteries (1.5V R6P) 8 pcs. yourself from electrical shock hazard. set memory number in Data Memory Function.
Voltage
3. Alligator Clip (black) 1 pce. 6. Instruction Manual 1 pce. 300
250V Range WARNING 3. Be careful not to get hurt with the sharp test lead pins. 4-13. High Voltage Warning Lamp
125V
CAUTION 1. Do not polish the case or attempt to clean it with any cleansing Lights up during insulation resistance and Low Resistance measurements to
100
2. SPECIFICATIONS 0
50V fluid like gasoline or benzine. If necessary, use silicon oil or antistatic fluid. indicate high voltage generation.
0.01 0.05 0.1
0.25 0.5
1 10 100 1000 CAUTION 2. Avoid the Insulation Tester from hard mechanical shock or 4-14. Input Terminals Test Leads
2-1. GENERAL SPECIFICATIONS 0.125
vibration, high temperature and strong magnetic field.
1. DISPLAY (LCD) a. Numerical Display : 4000 count, Maximum reading 4000, with bargraph Insulation Resistance (M ) Set up EARTH Test Lead (black) as shown in fig. 4. Plug black EARTH test lead into
CAUTION 3. Remove the batteries when the instrument is out of use for a EARTH terminal, and red LINE test lead into LINE terminal.
b. Units and Symbols : , V, , M, BAT, DH, 0ADJ, MEMO, COMP, long time. The exhausted batteries might leak electrolyte and corrode the inside.
, , , bargraph and decimal point. 2. Low Resistance Measurement
2. SAMPLING RATE : 2 times / second 2-1. Display Range and Accuracy 5. MEASUREMENT PROCEDURES
3. RESPONSE TIME : Insulation resistance ( 0, center value), Display Range Accuracy Measurement 4. NAME ILLUSTRATION
Resolution 5-1. PREPARATION FOR USE
Low resistance : within 5 seconds, Voltage : within 2 seconds (Auto-ranging) (After 0 Adjustment) Current
INPUT Terminals 1. INSTRUCTION MANUAL
4. OPERATABLE TEMPERATURE & HUMIDITY : 0 to 40 , 90% RH or lower in non-condensing 20.00 0.01 200mA or more
3rdg5dgt Read INSTRUCTION MANUAL carefully to understand the specification and functions
5. STORAGE TEMPERATURE & HUMIDITY : -20 to 50 , 75 RH or lower in non-condensing 400.0 0.1 Approx. 10mA SK-3502 DIGITAL INSULATION TESTER

High Voltage correctly. 3. SAFETY PRECAUTIONS is highly important for safety measurement.
6. OPERATABLE ENVIRONMENT : Indoors, <2000m ASL Open Circuit Voltage : 6V or more Overload Protection : by 0.5A / 600V fuse
High

LCD Warning Lamp


Voltage

7. TEMPERATURE COEFFICIENT : 0 to 18 , 28 to 40 ; add 5 rdg 5dgt to accuracy 2. BATTERY


3. AC/DC Voltage Measurement LIGHT Key
8. IP PROTECTION DEGREE : IP40
9. POWER SUPPLY : 1.5V R6P or LR6(AA) battteries x 8 3-1. Display Range and Accuracy Safety Lock Key BAT
TIMER Key
TIMER ON 1000V
1000V PRESS
2000M 1000V & 500V
LIGHT
Input
2000MΩ

SELECT Key )
500V

10. MAXIMUM RATED POWER : 4.5VA (rated power in auto power save : 1mVA MAX) Range Resolution Accuracy 3. FUSE
2000M

Resistance
MEMO
250V

FUNCTION Switch
SELECT HIGH 2000M
OFF LOW

11. OPERATABLE SUPPLY VOLTAGE : 8.8V to 14.4V ("BAT" indication at 8.8V or less)
125V

DCV/ACV fixed 0V to 600V Approx.


MEASURE
200M
ON

MEASURE Key
COMP
AC/DC V

12. CONTINUOUS OPERATING TIME (with manganese cell) : Insulation resistance 1V 3rdg5dgt 600V MAX

Fuse Testing
AUTO 10V to 600V 660k DH
5 SEC
OFF
OFF
0 ADJ Key
COMP Key
LOW
Low Ω 0 ADJ

measurement ; [approx. 14 hours in 50V, approx. 14 hours in 125V, approx. 14 Max. Input Voltage : 600V rms Overload Protection : 900V rms (for 1 minute) LOW OL
hours in 250V, approx. 12 hours in 500V, approx. 10 hours in 1000V (50V:SK-3500,
1000V:SK-3502) ], AC V : approx. 30 hours, Low resistance : approx. 13 hours.
) UP-DOWN Keys MEMO Key, HIGH/LOW Key

13. NUMBER OF MEASUREMENT : 3. SAFETY PRECAUTIONS 4. BUILT-IN CLOCK SETTINGS


Black EARTH Test Lead Attach alligator clip to Black
1) Insulation Resistance (ON for 5 seconds, OFF for 25 seconds) Correct knowledge of electric measurements is essential to avoid unexpected danger CE Kaise Earth Test Lead. To memorize date and time of measuring data in Data Memory Function, set built-in clock in
50V : manganese cell : more than 1800 times, alkaline cell : more than 5000 times (measure 0.05M ) *SK-3500 such as operator's injury or damage to the instrument. Read carefully and observe the the following procedures. (For the function details, refer to 6-2. Data Memory Function ).
125V : manganese cell : more than 1600 times, alkaline cell : more than 4500 times (measure 0.125M ) following precautions for safety measurements. Setting Procedures :
250V : manganese cell : more than 1500 times, alkaline cell : more than 4000 times (measure 0.25M )
3-1. WARNINGS Set FUNCTION Switch from "OFF" to "LOW " holding down
500V : manganese cell : more than 1400 times, alkaline cell : more than 3500 times (measure 0.5M )
CHI SHING

WARNING 1. Checks of Body and Test Lead UP-DOWN Keys together. Keep the keys holding down until setting
1000V : manganese cell : more than 900 times, alkaline cell : more than 2500 times (measure 1M ) *SK-3502 Alligator Clip

G
HIN
Before measurement, confirm the body of this instrument and handle insulators of the Test items (year, month, date, hour: minute) are indicated on LCD.

CHI SHING
IS
CH
2)Low Resistance 1 ON for 5 seconds, OFF for 25 seconds)
Lead have no cracks or any other damages. Dust, grease and moisture must be removed. NOTE : To reset clock settings into "2008 01.01 00:00",
manganese cell : more than 700 times, alkaline cell : more than 2500 times Red LINE Test Lead
WARNING 2. Warning for High Power Line Measurements press 0 ADJ Key after .

Kaise
14. FUSE : Fast-acting 0.5A/600V (6.332mm) x 1

Ka
High Power Line (High Energy Circuits) such as Distribution Transformers, Bus Bars and Press MEMO Key. Set last 2 digits of "year" (blinking)

CE
e is
15. FUNCTION : Comparator, Data memory, Timer, Live wire warning, High voltage warning, Safety

CE
Large Motors are very dangerous. High Power Line sometimes includes High Surge by UP-DOWN Keys.
lock, Automatic discharge, Zero-ohm adjustment, Backlight, Display hold, Auto power save fig. 3 fig. 4
Voltage that could cause explosive short in the instrument and could result in shock Press MEMO Key again. Set "month" (blinking) by
16. DIELECTRIC STRENGTH : 6.88kV 50Hz sine wave for 1 minute (between circuit and chassis)
hazard. For voltage measurement of High Power Line, do not touch Insulation Tester, UP-DOWN Keys.
17. COMPLIANT STANDARD : JIS C 1302:2002 (Insulation resistance) 4-1. LCD
its Test Leads, and any part of the circuit. Repeat and and set the all items until "minute". To
18. SAFETY LEVEL : CE Marking approved (IEC-61010-1, CAT 600V CLASS and EMC Test passed.)
WARNING 3. Warning for High Voltage Measurements return to the last item, press COMP Key.
19. DIMENSIONS & WEIGHT : 120(H) x 130(W) x 70(D)mm, Approx. 600g (Including batteries)
Even for Low Energy Circuits of electric/electronic appliances, such as heating When the setting is done, press MEASURE Key. Setup
20. ACCESSORIES : 100-61 Test Lead x 1 set, 945 Alligator Clip (black) x 1, 1016 Carrying
elements, small motors, line cords and plugs, High Voltage Measurements are very date and time are displayed.
Case, 1.5V R6P(AAA) batteries x 8, Spare fuse (0.5A/600V) 1 (built-in), Instruction Manual
dangerous. Do not touch lnsulation Tester, Test Leads, and any part of the circuit. Turn the power off to finish clock settings.
21. OPTIONAL ACCESSORIES : 766B Black Test Prod (with 767B Test Pin), 940R Red Alligator Clip
Generally, shock hazard could occur then the current between the circuit, that involves V
NOTE Built-in clock is powered by internal lithium battery.
2-2. MEASUREMENT SPECIFICATIONS more than 33V rms or 46.7V DC or peak, and ground goes up to 0.5mA or more. Time settings will be incorrect when it is exhausted. For
(Accuracy guaranteed condition : 23575RH in non-condensing) replacement, ask KAISE AUTHORIZED SERVICE AGENCY or
WARNING 4. Dangerous Voltage Measurement Procedure M : Lights up in insulation resistance DH : Lights up while holding
1. Insulation Resistance Measurement (50V : SK-3500 only, 1000V : SK-3502 only) the manufacturer, KAISE CORPORATION.
For dangerous voltage measurement, strictly observe the warnings below. (refer to fig. 1) measurement LCD indications
1-1. Rated Test Voltage and Maximum Effective Reading Do not hold Insulation Tester and test leads in your hands. Keep safety distance from : Lights up in low resistance 0 ADJ : Lights up in zero-ohm adjustment 5. SYMBOL MARK
power source or circuit to be measured not to touch the dangerous voltage. measurement MEMO : Lights up in data memory function The following symbol marks shown on the instrument and instruction manual are
Rated Test Voltage (DC) Max. Effective Reading Center Reading
Black test lead : Attach black alligator clip. V : Lights up in voltage measurement COMP : Lights up in comparator function listed in IEC 61010-1 and ISO 3864.
SK-3500 50V / 125V 200M 5M Red test lead : Attach red alligator clip to test lead pin. : Lights up in AC voltage measurement : Lights up if measuring value
250V / 500V 2000M 50M Turn off the power of the circuit to be measured when connecting test leads. : Lights up or blinks at high voltage exceeds a threshold
Caution (refer to instruction manual.) ~ Alternating Current (AC)
SK-3502 125V 200M 5M Dangerous voltage is applied Double Insulation
After measurement, before detaching alligator clips (test leads), turn the circuit input and output : Lights up if measuring value
250V / 500V / 1000V 2000M 50M power off and discharge the all capacitors. to this terminal. CE Marking Conformity
BAT : Low battery warning falls below a threshold
1 2 3 4
5-2. INSULATION RESISTANCE MEASUREMENT (M) Plug black EARTH test lead into EARTH terminal, and red LINE test lead into LINE terminal. Set buzzer timing. Press High / LOW Key and select either or . Replace the batteries and fuse when "BAT" lights up on LCD or fuse blows out.
Set FUNCTION Switch to "LOW ". "- - - - " and time are indicated on LCD. Use Turn the power off and detach test leads from the circuit.
WARNINGS DOWN Key ( ) to change the time indication into year and date indications.
Buzzer sounds if measuring value exceeds a threshold with flashing .
Take the carrying case off. For easy taking off, insert your fingers into test plug
If it is lower than threshold, indicates .
Do not measure live-wire insulation resistance to avoid serious damage to the instrument and electric Measurement range is selectable among "AUTO", "20.00 range", and "400.0 slots behind input terminals, and lift up the instrument.
Buzzer sounds if measuring value falls below a threshold with flashing .
shock hazard. Before starting measurement, turn off the power to the circuit to be measured without fail. range" by UP Key ( ). Default setting is "AUTO". Unscrew Battery Cover and remove exhausted batteries.
If it is higher than threshold, indicates .
Dangerous high voltage is applied to the measurement terminals when measuring insulation resistance. To avoid Insert 8 of new 1.5V R6P (AA) batteries in correct polarity.
electric shock hazard, do not touch test lead pins or the other metal parts, and circuit or object being measured. AUTO : "- - - - " or "Auto " indications Press COMP Key to finish comparator settings. Fuse : Replace a blownout fuse into a new one. (One spare fuse is included inside
To avoid electric shock hazard, do not touch measured circuit or object after measurement 20.00 range : " - -.- -" indication (Confirm by decimal point.) Press MEASURE Key or TIMER Key to start Comparator Measurement. Buzzer battery cover.)
without discharging them. For details, see "Automatic Discharge Function" mentioned below. 400.0 range : " - - -.- " indication (Confirm by decimal point.) sounds according to the setup condition. Fix Battery Cover and tighten the screw.
flashes and buzzer sounds.
Read 3. SAFETY PRECAUTIONS carefully to avoid electric shock hazard and serious damage to the instrument. Adjust ZERO in the following "Zero-ohm Adjustment Function" procedures. Setting Example : Measure NOTE : Remove the batteries when the instrument is out of use for a long time.
Connect test leads to the object to be measured. Refer to fig.8. 4M The exhausted batteries might leak electrolyte and corrode the inside.
Plug black EARTH test lead into EARTH terminal, and red LINE test lead into LINE terminal. Set a buzzer on
WARNING : If connected to live-wired object that has over 10V, and bargraph flashes to prevent Set comparator to sound a 10M or lower.
Set FUNCTION Switch to the suitable insulation measurement function. High
Voltage
Fuse 1.5V R6P (AA)
dangerous measurement. Turn off the power of the object to be measured. (Live wire warning function). buzzer at 10M or lower. batteries
SK-3500 : 50V / 125V / 250V / 500VSK-3502 : 125V / 250V / 500V / 1000V (0.5A/600V)
Measurement by Pressing MEASURE Key When measuring value is 4M, Spare Fuse
To select 500V or 1000V (High voltage Safety Lock key
Press MEASURE Key. High Voltage Warning Lamp lights up.Keep it holding down during measurement. flashes and buzzer sounds.
warning function) : Set FUNCTION 1000V PRESS

When measuring value is 40M ,


TIMER ON 1000V V
Read the measurement value on LCD. Buzzer sounds at 10 or less if Comparator Function is not used.
2000M
2000M 1000V & 500V
LIGHT

Switch holding down Safety Lock Key.


500V
2000M F0.5A

lights up.
MEMO
250V 600V

Release MEASURE Key to stop measurement. Displayed values are held fig.8
SELECT HIGH 2000M

If they are selected without pressing


LOW
MEASURE 125V 1.5V R6P(AA) 1.5V R6P(AA)
200M

automatically for 5 seconds ("DH" lights up). Measure


ON
COMP
AC/DC V

Safety Lock Key, LCD indicates "Err" to 600V MAX

OFF
OFF 40M
1.5V R6P(AA) 1.5V R6P(AA)

Detach test leads and set FUNCTION Switch to "OFF".


DH
5 SEC

prevent the measurement.


LOW

fig.5 lights up.


Low 0 ADJ 1.5V R6P(AA) 1.5V R6P(AA)

Supporting Functions : Comparator Function, Data Memory Function, Backlight, Setting Procedure in LOW Resistance Measurement 1.5V R6P(AA) 1.5V R6P(AA)

Selected rated voltage, "- - - - M" and time are indicated on LCD. Use DOWN fig.10
Auto Power Save (Refer to 4-8 and 6-1 to 6-3) Set FUNCTION Switch to "LOW ".
Key () to change the time indication into year and date indications. Press COMP Key. "COMP" lights up and enter comparator setting mode.
Continuous Measurement
Connect an alligator clip on black EARTH test lead to the ground side of the object to be
Press TIMER Key. High Voltage Warning Lamp lights up, and starts 3-minutes continuous Threshold and or blink in the right part of LCD. 8. CARRYING CASE USAGE
measured. As for the measurement of nongrounded object, connect it to optional part. Press UP-DOWN Keys and select suitable threshold from the following table.
measurement. Buzzer sounds at 10 or less if Comparator Function is not used. Easy storage of carrying case cover when taking measurement.
Connect red LINE test lead to the object to be measured.
Read the measurement value on LCD. FUNCTION Selectable Threshold Value Safe and stable hands-free measurement is possible by wearing shoulder strap
WARNING If connected to live-wired object that has over 10V, and bargraph flashes to prevent
Press MEASURE Key to stop continuous measurement. Displayed values are held around your neck with releasing a side snap as shown in fig.11.
dangerous measurement. Turn off the power of the object to be measured. (Live wire warning function). LOW 0.5/1/2/3/4/5/10/20/30/50/100/200
automatically for 5 seconds ("DH" lights up).
Kaise CE
CHI SHING Detach test leads and set FUNCTION Switch to "OFF". NOTE : Cannot select 30or higher if measurement range is fixed at 20.00.
Connect to ground terminal Supporting Functions : Comparator Function, Data Memory Function, Backlight, Set buzzer timing. Press High / Low Key and select either or .
Press COMP Key to finish comparator settings.
取説

Auto Power Save (Refer to 4-8 and 6-1 to 6-3)


Press MEASURE Key or TIMER Key to start Comparator Measurement. Buzzer
説明ラベル

Power supply side


SK-3502 DIGITAL INSULATION TESTER

High
Voltage
ZERO-OHM ADJUSTMENT FUNCTION sounds according to the setup condition.
Warning This function cancels resistance of test leads and fuse to display actual resistance To Cancel Comparator Function : Press COMP Key for 1 second or more ("COMP" disappears).
TIMER ON 1000V
1000V PRESS
OFF
Be sure to turn off the value of the object being measured.
breaker of the circuit
2000M
2000M 1000V & 500V
LIGHT

MEMO
500V
2000M

250V Plug red and black test leads to input terminals, and set FUNCTION Switch to "LOW ". 6-2. DATA MEMORY FUNCTION
to be measured.
SELECT HIGH 2000M
OFF LOW

Short-circuit metal pins of the test leads and press MEASURE Key.
MEASURE 125V
200M

Load side (effective for Insulation Resistance Measurement and Low Resistance Measurement)
ON
COMP
AC/DC V
600V MAX

OFF

Press 0 ADJ Key after LCD indications become stable. Confirm "0.000" is
DH OFF

fig.6
5 SEC

Up to 100 data can be memorized together with their measured date and time.
LOW
Low 0 ADJ

indicated on LCD. ("0 ADJ " lights up)


Measurement by Pressing MEASURE Key Data Memorize Procedure
Press MEASURE Key. High Voltage Warning Lamp and on LCD light up. Keep it NOTES Press MEMO Key during 5-seconds display hold
holding down during measurement. Zero Adjustment Function is effective when indicating value is 10 or less. If 0 after Insulation Resistance or Low Resistance Memory fig.11
Read the measurement value on LCD. ADJ Key is pressed over 10, "Err " is indicated. Number
Measurement. ("MEMO" lights up).
Release MEASURE Key to stop measurement. Displayed values are held If "OL" remains indicated under measurement after shorting test lead, fuse LCD indications and Memory Number blinks.
automatically for 5 seconds ("DH" lights up). blowout or test lead disconnection may occur. 9. MAINTENANCE
Memory Number is selectable by UP-DOWN Keys.
Discharge the measured object in the following "Automatic Discharge Function" proecdures. Measurement stops automatically after 30 minutes of continuous measurement for 20 or lower. Memory 9-1. PERIODICAL CHECK AND CALIBRATION
If the selected number is already used, the
Number 1
Detach test leads and set FUNCTION Switch to "OFF". memorized data is displayed. Periodical check and calibration is necessary to make safety measurements and to
Supporting Functions : Comparator Function, Data Memory Function, Backlight,
5-4. AC/DC VOLTAGE MEASUREMENT (V) maintain the specified accuracy. The recommended check and calibration term is once
Press MEMO Key for 1 second or more to Press UP Key
Auto Power Save (Refer to 4-8 and 6-1 to 6-3) WARNINGS memorize a data in the selected Memory Number. a year and after the repair service. This service is available at KAISE AUTHORIZED
Memorized items are "measurement value", "rated Memory SERVICE AGENCY through your local dealer.
Continuous Measurement Do not measure any voltage that might exceed maximum input value (600V AC/DC). Number 2
Press TIMER Key. High Voltage Warning Lamp and on LCD light up, and starts Always connect test leads to secondary side of a breaker. Connection to voltage" and "measured date and time". 9-2. REPAIR
3-minutes continuous measurement. primary side may cause electric shock hazard. NOTES Repair service is available at KAISE AUTHORIZED SERVICE AGENCY through your
WARNING : Dangerous high voltage is continuously applied to test lead pins. Pay To avoid electric shock hazard, pay attention not to short-circuit live wire with test lead pins. The previously memorized data is deleted if a new data is overwritten in the local dealer. Pack the instrument securely with your name, address, telephone number
careful attention not to get an electric shock. Read 3. SAFETY PRECAUTIONS carefully to avoid electric shock hazard and serious damage to the instrument. same Memory Number. and problem details, and ship prepaid to your local dealer.
Read the measurement value on LCD. In Insulation Resistance Measurement, when measuring over 2000M under Check the following items before asking repair service.
Plug black EARTH test lead into EARTH terminal, and red LINE test lead into LINE terminal.
Press MEASURE Key to stop continuous measurement. Displayed values are held 250V/500V/1000V ranges, or measuring over 200M under 50V/125V ranges, LCD
Set FUNCTION Switch to "AC/DC V". "Auto V" and time are indicated on LCD. Check the battery connection, polarity, and capacity ("BAT" lights up or not).
automatically for 5 seconds ("DH" lights up). can display up to 2500M or 250M each, but memorized data should be "OL".
Use DOWN Key () to change the time indication into year and date indications. Confirm that FUNCTION Switch is set to the correct position.
Discharge the measured object in the following "Automatic Discharge Function" procedures.
Measurement range is selectable among "AUTO", "DCV", and "ACV" by UP Key View the Memorized Data Confirm that the body of this instrument and handle insulators of the Test Leads
Detach test leads and set FUNCTION Switch to "OFF".
(). Default setting is "AUTO". Set FUNCTION Switch to Insulation Resistance Measurement function or "LOW have no cracks or any other damages.
Supporting Functions : Comparator Function, Data Memory Function, Backlight,
", and press MEMO Key. ("MEMO" lights up) Check if any noise affects the instrument. This instrument is fully shielded
Auto Power Save (Refer to 4-8 and 6-1 to 6-3) AUTO ("Auto V" indication) : 10V to 600V AC/DC is measurable. AC or DC are
Memorized data is displayed together with its Memory Number. Use UP-DOWN against noise, but possibly to be affected by very strong noise.
NOTES selected automatically. NOTE : No DC ( ) indication.
Keys to scroll the display.
Measurement value might be unstable depending on the object to be measured. DCV fixed ("V" lights up) : 0V to 600V DC is measurable.
Use SELECT Key to display measured date and time. Rated voltage in the bottom-left
WARRANTY
Connect EARTH terminal (+ polarity) to the ground side of the object to be measured if it is grounded. NOTE : No DC ( ) indication. SK-3500/3502 are warranted in their entirety against any defects of material or workmanship
~ corner of LCD switches into "time" "date" "year" as pressing SELECT Key.
Measurement value can be lower under this connection that is suitable for circuit measurement. Pay ACV fixed ("V" lights up) : 0V to 600V AC is measurable. under normal use and service within a period of one year from the date of purchase of the original
attention to lead wires connected to LINE terminal (- polarity) not to touch the ground or any other object. Measured Time Measured Date Measured Year purchaser. Warranty service is available at KAISE AUTHORIZED SERVICE AGENCY through your
Connect black alligator clip to () earth side and red test lead to () side of the circuit to be measured.
When measuring over 2000M under 250V/500V/1000V ranges, or measuring over 200M under local dealer. Their obligation under this warranty is limited to repairing or replacing SK-3500/3502
NOTE : Connect insulation tester IN PARALLEL to the circuit.
50V/125V ranges, LCD can display up to 2500M or 250M each, but memorized data should be "OL". returned intact or in warrantable defect with proof of purchase and transport charges prepaid.
Read the measurement value on LCD.
KAISE AUTHORIZED DEALER and the manufacturer, KAISE CORPORATION, shall not be liable
AUTOMATIC DISCHARGE FUNCTION After measurement, set FUNCTION Switch to "OFF".
fig.9 for any consequential damages, loss or otherwise. The foregoing warranty is exclusive and in lieu
After insulation resistance measurement, electricity is charged to the measured Supporting Functions : Display Hold, Backlight, Auto Power Save (Refer to 4-3, 4-8 and 6-3)
Delete All of the Memorized Data of all other warranties including any warranty of merchantability, whether expressed or implied.
object. To avoid electrical shock hazard, discharge them in the following procedures. NOTES Set FUNCTION Switch from "OFF" to "AC/DC V" holding down UP-DOWN Keys together. This warranty shall not apply to any instrument or other article of equipment which shall
Stop measurement without detaching red and black test leads from the measured object. Warning buzzer sounds when measuring 601V or more. Start to delete the data. have been repaired or altered outside of KAISE AUTHORIZED SERVICE AGENCY, nor
The function starts the discharge automatically. Wait until it is completed. "OL" is indicated when measuring 1000V or more. Buzzer sounds when all of the memorized data is deleted. which have been subject to misuse, negligence, accident, incorrect repair by users, or
and bargraph flashes during the discharge. The number of bargraph shows discharge level. Set FUNCTION Switch to "OFF" before restarting the measurement. any installation or use not in accordance with instructions provided by the manufacturer.
Detach test leads from the measured object after stop flashing and bargraph is disappered. 6. FUNCTIONS
6-3. AUTO POWER SAVE (Effective for the all measurement functions) KAISE AUTHORIZED DEALER
6-1. COMPARATOR FUNCTION
Power turns off automatically after approx. 15 minutes of last switch operation to
(effective for Insulation Resistance Measurement and Low Resistance Measurement)
conserve battery life. (Small current consumption remains. After measurement, always
Under Discharge Buzzer sounds and or flashes if measuring value exceeds or falls below the threshold.
set FUNCTION Switch to "OFF".)
discharge completed Setting Procedure in Insulation Resistance Measurement
fig.7 Set FUNCTION Switch to Insulation Resistance Measurement function.
5-3. Low Resistance Measurement () 7. BATTERY AND FUSE REPLACEMENT
Press COMP Key. "COMP" lights up and enter comparator setting mode.
WARNINGS Threshold and or blink in the right part of LCD. WARNINGS
Press UP-DOWN Keys and select suitable threshold from the following table.
Do not measure voltage in Low Resistance Measurement function to avoid To avoid electrical shock, detach instrument and test lead from circuit when to
electric shock hazard and serious damage to the instrument. FUNCTION Selectable Threshold Value replace batteries and fuse. Set FUNCTION Switch to "OFF".
Before measurement, to avoid electric shock hazard, turn off the power of the 50V / 125V 0.1/0.2/0.3/0.4/0.5/1/2/3/5/10/20/30/50/100/200M After replacing batteries and fuse, be sure to tighten the screws of battery cover.
circuit to be measured and discharge the all capacitors. 250V /500V / 1000V 0.1/0.2/0.3/0.4/0.5/1/2/3/5/10/20/30/50/100/200/500/1000/2000M Always use the specified fuse. Do not use the instrument shorting fuse holder or
Read 3. SAFETY PRECAUTIONS carefully to avoid electric shock hazard and serious damage to the instrument. without using the fuse. Fuse type : Fast-acting 0.5A/600V ( 6.3x32mm) Product specifications and appearance are subject to change without notice due to continual improvements.
* 50V : SK-3500 only, 1000V : SK-3502 only