SOP for Out of Specification
SOP for Out of Specification
1.0 OBJECTIVE:
The purpose of this SOP to lay down the procedure for dealing with out of specification
(OOS) results.
2.0 SCOPE
This standard operating procedure is applicable to situations arising in the event of a product
failing to meet the laid down specification. Scope of implementation of the standard operating
procedure is limited to validated products, methods, processes and sample received from plant,
tested accordance with existing Pharmacopoeia or In-house methods and having laid down
specifications.
3.0 RESPONSIBILITY:
3.1 Executive-Quality Assurance - Prepare the SOP and follow-up the SOP accordingly.
3.2 Asst. Manager-Quality Assurance/Production/Quality Control-Provide the support to the
implementation of SOP and maintained the records.
3.3 Head–QA shall be responsible for insure the compliance of the procedure and implementation
of this SOP.
5.0 DISTRIBUTION
5.1 Quality Assurance
5.2 Quality Control
5.3 Production
5.4 Ware house
5.5 Engineering
6.2 Abbreviations
6.2.1 SOP : Standard Operating Procedure
6.2.2 QA : Quality Assurance
6.2.3 OOS : Out of Specification
6.2.4 cGMP : Current good Manufacturing Practices
6.2.5 HR : Human resource
6.2.6 QC : Quality Control
6.2.7 CAPA : Corrective action and Preventive action
6.2.8 USFDA : united States food and drug administration
6.2.9 CDER : Center for drug evaluation and research
7.0 PROCEDURE
7.1 QC Investigations of "Out of Specification (OOS)" shall be done in cases of:
7.1.1 Batch release testing and testing of starting materials.
7.1.2 In- Process Control testing.
7.1.3 Stability studies on marketed batches of finished products and or active pharmaceutical
ingredients, ongoing / follow up stability (no stress tests).
7.1.4 Previous released batch used as reference sample in an OOS investigation showing OOS or
suspect results.
Note: The OOS process is not applicable for In-process testing while trying to achieve a
manufacturing process end-point i.e. adjustment of the manufacturing process. (e.g. pH,
viscosity), and for studies conducted at variable parameters to check the impact of drift (e.g.
process validation at variable parameters).
7.2 Chemist finding out of specification result shall immediately stop the analysis and inform to
Head QC immediately. The Analyst should retain all the sample/standard solutions.
7.3 Head QC shall carryout the primary investigation (Phase I). The investigation shall include the
following but not limited to.
7.3.1 Phase – I: Initial Assessment
7.3.1.1 System Failure (Phase –I a Investigation)
7.3.1.1.1 Power failure
7.3.1.1.2 Calibration check failure
7.3.1.1.3 Standard/ sample spoiled by spillage/ contamination
7.3.1.1.4 Unsuitable instrument used