Dso3d12 Digital Oscilloscope Manual
Dso3d12 Digital Oscilloscope Manual
Contents
1 supereyes DSO3D12 Digital
Oscilloscope
2 OVERVIEW
3 Buttons Function
4 Safety Precautions
5 Quick start instructions
6 Probes
6.1 Use tips
7 Vertical system
8 Horizontal system
9 Trigger system
10 Operation Guide
11 Menu introduction
12 Parameters
13 Q & A
14 Documents / Resources
1. Green is the running state; Flashing cyan for waiting to trigger; Red is stopped.
2. Channel priority:Y ellow 1 indicates that CH1 has high priority, and the CH1 waveform can be moved up or
down.
3. Yellow: Turn on the signal generator output; Gray: Turns off the output.
4. Shows the position of the current time base window in memory.
5. Auto indicates that the trigger mode is automatic. Normal indicates that the trigger mode is normal.
6. Trigger edge: Rising edge trigger or falling edge trigger.
7. Battery level indication.
8. Trigger level position.
9. X1/X10 indicates probe attenuation, which should be set to match the probe. The goal is to tell the oscilloscope
what attenuated probe you used.
10. DC/AC coupling.
11. Vertical sensitivity: Indicates the voltage of a grid.
12. Timebase: The time of a grid.
13. Zero arrow: The baseline position, where the vertical voltage is zero.
Note: (CH2 icon is gray to indicate that it is off, if you do not need to test two signals at the same time, please
close CH2 to obtain higher sampling rate and reduce power consumption).
Buttons Function
Shift⬆ When this button is pressed, the functions on the case will be performed.
Probe attenuation, which tells the oscilloscope what probe range you
x1x10
are using. (CH1)
Note:
CH1 cannot be turned off, CH2 can be turned on or off. When two channels are not required, CH2 should be
turned off for higher sample rates and reduced power consumption.
Safety Precautions
High voltage. The x10 range of the probe measures up to 220V. An x100 probe is required for overreach.
Probe attenuation. Before measuring voltages greater than 40V, switch the probe to the X10 position.
Charge. The charging voltage is 5V, use a computer or mobile phone charger.
When measuring high-voltage or non-isolated circuits, it cannot be used while charging (powered by a built-in
battery).
Do not use a multimeter while charging.
ever insert the red probe into the current hole to measure the voltage, it will blow the fuse or even damage the
machine. After measuring the current, remember to change the red probe to the voltage hole, remember!!
When the multimeter measurement current is greater than 6A, the duration should not exceed 10 seconds to
avoid line damage caused by heating.
When measuring high voltages, do not use the oscilloscope and multimeter at the same time.
Power
Press and hold the power button on/off. When the battery is dead, use the USB cable to connect to the 5V
adapter for charging, and the charging current is about 700mA. It is best to use a computer or mobile phone for
charging. The LED lights up red when charging, and turns green when fully charged.
Auto
After the probe is connected to the signal, press the [Auto] button, and the oscilloscope will automatically detect
the signal and adjust the range. If there is no waveform, the probe may not be connected correctly, or there may
be no signal at the connection (you can use a multimeter to determine if there is a voltage).
Use of Multimeters
Check the red pen first in the voltage hole. Press the [Select] key to open the multimeter function table, select the
function and press the [OK] key. If you want to measure the current, insert the red pen into the current hole, and
pay attention to the “series” of the meter pen and the power line. When charging: Multimeter is not supported, and
the oscilloscope cannot measure high voltage or non-isolated circuits.
Use dual channels
Press the [Shift] key, and then press the [CH1/CH2] key to open/close channel 2. The default up and down keys
move the channel 1 waveform, if you need to move the channel 2 waveform, you can press [CH1/CH2] to set the
high priority of channel 2 (pay attention to the change of the 1/2 icon in the upper left corner). When not using dual
channel measurements, Channel 2 should be turned off for higher sample rates and reduced power consumption.
Probes
Probe calibration
The probes delivered with this machine have been compensated. If you use another new probe before the first
use, it must be calibrated before it can be used normally, otherwise the measurement will be inaccurate (only for
the 10X/100X). The calibration method is as follows:
Probe bandwidth
Because the 1X probe has a large input capacitance, the bandwidth of the 1X is usually within 5MHz; while the
bandwidth of the 10X is the standard bandwidth of the probe.
Use tips
High bandwidth
Because the 1X probe has a large input capacitance, the bandwidth of the probe 1X is usually within 5MHz, so
please use the 10X position when the measured signal bandwidth is greater than 5M.
Vertical system
Vertical sensitivity
Oscilloscope vertical sensitivity refers to the voltage represented by a grid in the vertical direction of the screen.
There are 8 divisions in the vertical direction of the oscilloscope. You can estimate the amplitude of the signal by
observing how many divisions the waveform occupies on the screen. For example, if the vertical sensitivity is
1v/div, and the signal occupies 3 grids on the screen, you can know that the amplitude of the signal is 3V.
The scope’s vertical sensitivity range is 10mV / div ~ 10V / div (1X probe). In the oscilloscope button panel, the
keys for adjusting the vertical sensitivity are ‘mV’ and ‘V’ button. Pressing the ‘mV’ button decreases the vertical
sensitivity, which is used to measure signals with smaller voltage amplitude. Press the ‘V’ button to increase the
vertical sensitivity for measuring signals with larger voltage amplitudes.
Vertical position
How to move the waveform up and down: Press the up and down buttons directly to move the waveform up and
down. The number in the upper left corner is yellow 1 for movable channel 1, click the “CH1/CH2” key to switch to
green 2 movable channel 2.
Horizontal system
Time base
The time base refers to the time represented by each grid in the horizontal direction on the oscilloscope screen.
There are 12 grids in the horizontal direction of the oscilloscope. The sampling time that can be observed on the
screen = ” time base” * 12.The user can estimate the signal period by observing how many grids a waveform
period occupies. For example, the current time base is 500us, and one cycle of the signal occupies 2 grids,so the
signal period is 1ms (1KHz).
The oscilloscope time base range is 5ns ~ 10s. The buttons for adjusting the time base in the oscilloscope button
panel are ‘ns’ and ‘s’ button. When observing high-frequency signals, you should press the ‘ns’ button to decrease
the time base. When observing low frequency signals, you should press the ‘s’ button Increase the base.The time
base in the figure is 200us/div.
What you need to know is that the sampling process of the oscilloscope is: “Sampling”>>”Processing” >> “Display”
When adjusted to a large time base, the screen waveform refresh will be slow due to the longer time taken by the
oscilloscope to sample the waveform.
In the stop mode, the user can move the waveform left and right.
Trigger system
About trigger
The oscilloscope will capture a frame of waveform only after the waveform meets the pre-set conditions. This
action of capturing the waveform according to the conditions is the trigger.
Trigger level
In the edge trigger mode, the trigger condition appears only when the waveform crosses up or down the trigger
level. The trigger level is automatically adjusted by default. Here’s how to adjust manually:
Trigger Sources
The small arrow on the right, yellow indicates that the trigger source is CH1, and other color indicates that the
trigger source is CH2. Which channel is used as a trigger can stabilize the waveform of which channel, and two
channels cannot be triggered at the same time.
Trigger Mode
Trigger mode: automatic trigger, normal trigger, single trigger. Automatic triggering is more commonly used. The
normal trigger will not refresh the screen when there is no waveform, suitable for measuring some pulse signals
that come occasionally, etc., and refresh the waveform once in a pulse. The single-trigger procedure is described
in 7.2 below.
Operation Guide
50%
The role of 50% is to quickly return to the center, including zero point offset, trigger position, trigger level will
quickly return to the middle position with one button, operation method: directly click “50%”.
Single trigger
Suitable for sporadic single pulse capture. A single trigger is when the single button is pressed, and the
oscilloscope samples a frame of waveform and then stops. That is, only one frame of waveform is taken per
press. It is important to note that this sampling must be triggered, if it is not triggered, the upper left corner of the
oscilloscope flashes “||”, and the sampling is not completed until there is a waveform triggeTr.o cancel a single
trigger, click “Stop” to stop.
Procedure:
Screenshot
Click the “Save” button to capture the current screen.
View images
After pressing the Shift key, than press the Save key to enter the Image View window. In this window, the user can
delete or browse historical screenshots. Click the “x1x10” button to delete the current screenshots. Click the “ok”
button to zoom in on the waveform. After zooming in, click “ok” to return to the homepage. Click the ” Save ”
button to return to the oscilloscope window.
Signal generator
Click the ” Gen ” button to call up the window of the signal generator, The signal amplitude is 2.5V, the maximum
frequency of the sine wave is 5M, and the other waves are 1M. Click the ” Stop” button to turn on/off the signal
generator output. Click the “Single” button to reset the number to: 1.00Khz, 50% Click the “s” and “ns” keys to
switch the output waveform: sine wave, square wave, triangle wave, etc. Click “Ok” or “Gen” to confirm and
exit.Square wave can adjust the duty cycle, other waveforms do not work.
Note: When the set frequency is greater than 1MHz, the square wave signal may show small horizontal jitter.
Cursor
Press the “Shift” +”s” key to turn the horizontal cursor on/off.
Press the “Shift” +”ns”key to turn the vertical cursor on/off.
Cursor
Press the “Shift” +”s” key to turn the horizontal cursor on/off.
Press the “Shift” +”ns”key to turn the vertical cursor on/off.
Press the left and right arrow keys directly to move the horizontal cursor 1. After pressing the Shift key, press
the horizontal arrow key again to move the horizontal cursor 2. The same goes for vertical cursors.
For vertical cursors, Y1: The voltage at the dashed line above. Y2: The voltage at the dashed line below.
dY=Y1-Y2, which is the voltage difference between the two dashed lines. The vertical (voltage) cursor can only
measure one channel, and whichever channel has the highest priority, measure the voltage of which channel.
Menu introduction
Channel
The channel menu includes two options, “Coupling” and “Probe”. The coupling can be set “DC coupling” and “AC
coupling”. DC coupling passes the signal directly. AC coupling is equivalent to series connection with the
capacitor. Probes are available in ” 1X” / “10X”/ “100X”, this option corresponds to the oscilloscope probe. When
the probe is switched to the 10X position, this option should be selected as ” 10X”. The same goes for the 1X.
Measurement
There are 14 measurement options available, as shown below: Frequency, Peak-to-Peak, Duty cycle, Amplitude,
RMS, Average, Period, +Pulse width, -Pulse width, Max, Min,Top, Base, -Duty cycle.
The difference between peak-to-peak and amplitude(AMP) is shown in the figure above.
Trigger
There are four options in the trigger menu: “Trigger Mode” ,“Trigger Type” , “Trigger level”and “Trigger source”.
Trigger Mode can be selected from “Auto” and “Normal”. For “Auto” mode, if a waveform trigger is detected, the
waveform will be used for triggering. At this time, the waveform can be displayed stably. If the oscilloscope cannot
detect the trigger condition, it will automatically force the trigger to display the waveform, but because there is no
suitable trigger condition, the waveform will not be displayed steadily.For “Normal” mode, The oscilloscope will not
refresh the display until there is a waveform trigger. Therefore, in this mode, the screen waveform may not refresh
or there is no waveform. It is suitable for measuring discontinuous waveforms and requires users to adjust the
trigger level in advance. Trigger type includes rising and falling edges. Rising edge trigger means that the
waveform crosses the trigger level line upward, and a trigger condition occurs. Trigger level include “automatic”
and “manual”. “Auto” is suitable for trigger mode = auto, the oscilloscope detects the waveform in real time and
automatically adjusts the trigger level. “Manual” When the “single” is used or the trigger mode is “Normal”, the
“Manual” option should be used. Under this option, the trigger level can be manually adjusted by the user. Trigger
source indicates which channel’s waveform is currently used for triggering. In the dual-channel mode, CH1 or CH2
can be selected. If the channel is used as the trigger source, its waveform can be stable. If the waveform
frequencies of the two channels are different, the waveform of the other channel will be unstable.
Display
“Persistence” is to make the historical waveform remain on the screen for a while and then disappear. Optional
persistence options are: None, 1s, Infinite. When Infinity is selected, the waveforms that have appeared on the
screen will remain forever and will not disappear. “roll” is to display the waveform while sampling. Only when the
time base is greater than or equal to 200ms will the roll mode be entered. If roll is turned off, the waveform will be
displayed after one screen is sampled, resulting in a long time before the screen is refreshed for large time base.
“XY” mode is a way that channel 1 is used as the X axis and channel 2 is displayed on the Y axis. It can be used
to view Lissajous graphics etc. “Hi-rate” refers to the waveform refresh speed. When turned on, many frame
waveforms will be superimposed. When turned off, only one waveform is displayed at a time. Note that it is always
“on” when the time base <=200ns.
Settings When the FFT option is turned on, the CH1 waveform can be FFT calculated and displayed. FFT display
options are: logarithmic, linear, and music spectrum. Measuring FFT results is not supported.The calibration
option is used to calibrate the zero point of the oscilloscope. Self-calibration is required when the outside
temperature changes or when there is a zero offset due to aging. Remove the probes before calibration.
Furthermore:
1. In the settings, you can set whether to automatically shut down, if you choose automatic shutdown, the
oscilloscope will automatically shut down 15 minutes after no key operation.
2. You can choose whether to turn on sound or not.
3. Choice of language: Chinese or English.
4. Keep this menu open, and long press “s” to switch the background color of the measurement item.
Parameters
±40V x1
width*hight* thickness 145*86*33
Max voltae Shell size
±400V x10 mm
Multimeter parameters
Function Range precision
Pay attention 600mA hole with 1A fuse, 10A hole without fuse
600.0Ω ±(1.5%+3)
6.000kΩ/60.00kΩ/600.0kΩ ±(1%+3)
6.000MΩ ±(1.5%+5)
60.00nF/600.0nF/6.000μF ±(10%+5)
Capacitance
60.00μF/600.0μF ±(15%+5)
Voice assistants
You must be in a quiet environment to use the voice assistant.The Shift+V(ch1) keys can turn the voice assistant
on/off. Wake up first to perform voice control, and if there is no voice command 20 seconds after waking up, it will
automatically go to sleep. Note that due to the size of the voice, speech rate, accent, etc.,we cannot guarantee
the accuracy of recognition. Wake-up command: “Hello, ZEEWEII” (Pronounced “Hello,see wave”) or “Hello,mini
scope”.
Command words function
Q&A
Documents / Resources
Manuals+,