0% found this document useful (0 votes)
11 views1 page

HPLC Application 18229 Pharma ABN Test 12 Mix On Luna 5um C18 (2) 150x4.6mm 1ml-Min

The document outlines the LC conditions for testing a mix of pharmaceuticals using a Luna 5 µm C18(2) column. It details the mobile phase composition, gradient profile, flow rate, temperature, and detection method. Additionally, it includes a note about the SecurityGuard™ Guard Cartridge System to extend column lifetime.

Uploaded by

cianrango
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
11 views1 page

HPLC Application 18229 Pharma ABN Test 12 Mix On Luna 5um C18 (2) 150x4.6mm 1ml-Min

The document outlines the LC conditions for testing a mix of pharmaceuticals using a Luna 5 µm C18(2) column. It details the mobile phase composition, gradient profile, flow rate, temperature, and detection method. Additionally, it includes a note about the SecurityGuard™ Guard Cartridge System to extend column lifetime.

Uploaded by

cianrango
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

2/4/25, 9:15 AM about:blank

App ID 18229

Pharma ABN Test 12 Mix on Luna 5um C18(2) 150x4.6mm 1mL/min


Order Items Used in This
Application

00F-4252-E0
Luna 5 µm C18(2) 100 Å, LC Column 150 x 4.6
mm, Ea

LC Conditions 1. Pyridine
Column: Luna 5 µm C18(2) 100 Å, LC Column 150 x
4.6 mm, Ea 2. Acetaminophen

Brand Name: Luna 3. Quinine


Part No: 00F-4252-E0
4. Acebutolol
Phase Name: C18(2)
5. Chlorpheniramine
Elution Type: Gradient
Mobile Phase: A: water w/ 0.1% formic acid 6. Triprolidine
B: acetonitrile w/ 0.1% formic acid
7. Prednisolone
Gradient Profile: Step No. Time(min) %A %B
1 0 95 5 8. 4-Chlorobenzoic acid
2 1.67 95 5
9. 4-Chlorocinnamic acid
3 21.69 5 95
4 21.83 95 5 10. Diazepam

5 30.56 95 5 11. Diflunisal

Flow Rate: 1 mL/min 12. Hexanophenone


Temperature: 45°C
Detection: UV-Vis Abs.-Variable Wave.(UV) @ 254 nm
(25°C)
Sample Note: SecurityGuard™ Guard Cartridge System
extends column lifetime. - SecurityGuard
Cartridges, C18 4 x 3.0mm, 10/Pk Part No.:
AJ0-4287 - Holder Part No.: KJ0-4282

Have question or want more details on implementing this method? We would love to help!
Page 1 of 1 Visit www.phenomenex.com/Chat to get in touch with our technical specialist