C5000 operation guide
C5000 operation guide
2024
Revised on July. 3. 2024
Measurement principle
Chemiluminescence resonance energy transfer (CRET) and neighbor-joining techniques are
widely used for bioanalytical sensors. Under alkaline hydrogen peroxide conditions, acridine
ester (AE) is oxidized to the substrate state, releasing an intense chemiluminescence signal
upon return to the ground state. Graphene oxide (GO) with excellent quenching ability can
CRET with AE-labeled DNA3, making the chemiluminescence almost non-existent and
achieving a "signal off" signal transition. The addition of the target detector induces an
adjacent contact reaction, forming an adjacent complex that detaches the AE-labeled DNA3
from GO and inhibits the CRET process. In addition, in situ cycling using nucleic acid
endonucleases can further amplify the signal and achieve highly sensitive
chemiluminescence detection.
The core detector device of this instrument is a photomultiplier tube (PMT), detected by a
single photon and transmitted to an amplifier, and add high voltage current amplification,
the amplifier will be converted from analog current to digital current, digital current will
transmit the luminescent signal to the main board and be calculated to obtain the relative
light unit (RLU), and then calculated by the standard curve to obtain the concentration of
Rear view
Side view
Revised on July. 3. 2024
Indicator status
Software basis
Menu structure
Revised on July. 3. 2024
Status bar
INSTALLATION
POWER SUPPLY
OPERATIONAL ENVIRONMENT
C5000 should be placed on a dry, clean, flat and horizontal surface in an in-door setting
away from direct sunlight, vibration and any source of electromagnetic radiation.
Temperature:10 ~ 30 ℃,
INSTALLATION PRECEDURE
switch button
Revised on July. 3. 2024
System login
Username: User
Passcode: 1
adapters
Revised on July. 3. 2024
the consumable
Go to Apply>Liquid cleaning
Revised on July. 3. 2024
draining procedure
Test procedure
1 Calibration
test kit
2 Assay protocol
bottom
reagent cuvette
order.
reagent cuvette.
Revised on July. 3. 2024
Go to Apply>User setting.
print function.
Maintenance
Please shut down the system daily to keep the analyzer in good condition.
Remove all the reagent cuvettes when the system is shut down.
Be sure to check on the remaining of the substrate and buffer before placing a test order.
Be sure to check on the waste container, empty the liquid inside if necessary.
After collection of the whole blood, allow the blood to clot by leaving it undisturbed at
Sample with visually inspected hemolysis, lipemia, or icterus should be considered for its
Sample Stability: Sample retrieved from the refrigerator should be left undisturbed till it
*Ni J, Zhu W, Wang Y, Wei X, Li J, Peng L, Zhang K, Bai B. A Reference chart for clinical biochemical tests of hemolyzed serum samples. J Clin
Lab Anal. 2021 Jan;35(1):e23561.
Revised on July. 3. 2024
FQA
What shall be done when the system fails to identify the assay item when the reagent
Make sure the calibration is added. Remove the reagent cuvette, switch to the other side QR
Do I need to scan the calibration every time when placing a new test order?
No. Scan the calibration only once when a new lot test kit is opened.
This error occurs when little chemiluminescence signals are detected. Check on the
substrate remaining, replace a new bottle of substrate if necessary. Also, makes sure all the
Switch off the analyzer and rotate the reaction tray manually. Clean the remaining liquid
HUMISS TRADING S. L.
Calle Luis Buñuel 12, MADRID 28018,Spain