0% found this document useful (0 votes)
30 views15 pages

Techniques for Time Domain Measurements - Application Note 5991-0420EN c20140723 [15]

This application note from Keysight Technologies discusses time domain and distance-to-fault measurement techniques using FieldFox handheld analyzers to identify discontinuities in RF and microwave systems. It explains the relationship between frequency and time domain measurements, the use of vector network analyzers for both types of measurements, and introduces the concept of gating to isolate specific reflections. Measurement examples illustrate the effectiveness of these techniques in troubleshooting and characterizing various components in the field.

Uploaded by

Alex Yang
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
30 views15 pages

Techniques for Time Domain Measurements - Application Note 5991-0420EN c20140723 [15]

This application note from Keysight Technologies discusses time domain and distance-to-fault measurement techniques using FieldFox handheld analyzers to identify discontinuities in RF and microwave systems. It explains the relationship between frequency and time domain measurements, the use of vector network analyzers for both types of measurements, and introduces the concept of gating to isolate specific reflections. Measurement examples illustrate the effectiveness of these techniques in troubleshooting and characterizing various components in the field.

Uploaded by

Alex Yang
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

Keysight Technologies

Techniques for Time Domain Measurements


Using FieldFox Handheld Analyzers

Application Note

Abstract
This application note will introduce time domain and distance-to-fault (DTF)
measurement techniques for identifying the location and relative amplitudes of
discontinuities while operating in the field. This application note will describe
the relationship between frequency domain measurements and time domain
transforms and their relationships to time resolution and range. Also shown
will be VNA configurations for characterizing band-limited devices such as
couplers, filters, antennas and waveguide components, and broadband devices
such as cables and connectors. This note will also discuss time domain
“gating,” a powerful feature that effectively isolates discontinuities in the time
domain just as a filter would isolate signal energy in the frequency domain.
Measurement examples will be provided using the Keysight Technologies, Inc.
FieldFox vector network analyzer.
Introduction

Testing and qualifying components and systems that function as part of a communications or radar
system often requires that the electrical performance of these devices achieve a certain level of
specified performance across the operating frequency range. Specifications include Voltage Standing
Wave Ratio (VSWR), return loss and insertion loss to name a few. These specifications provide a
clear distinction when the device under test (DUT) has passed or failed its performance requirements
as a function of frequency. Figure 1a shows the measured VSWR of a system where the VSWR has
exceeded the specification at several places across the measured frequency range of 8.5 to 12 GHz.
The Keysight FieldFox vector network analyzer (VNA) used in this example was configured with limit
lines to help the operator quickly identify whether the DUT has passed or failed the test.

While a frequency measurement provides useful information into the proper functioning of a system,
having only swept frequency may not provide enough information to determine the root cause of the
problem. When a system fails to meet specification, troubleshooting is often difficult as components
would then need to be swapped in and out of the system until performance once again meets the
specified requirements. Fortunately, there is another measurement technique that provides details into
the location and magnitude of any such problems. This technique relies on measurements in the time
domain and a vector network analyzer, such as the Keysight FieldFox, has the capability of displaying
the time domain characteristics of one and two port components and systems. Figure 1b shows the
time domain transformation of the VSWR measurement with the location of individual discontinuities
being displayed as a function of time. In this example, it can be assumed that the largest peak in the
time domain is associated with the component causing the out-of-spec condition for this system.
Knowing the propagation velocity of the signal in the transmission system, the physical location of the
fault can be determined and the system can be quickly repaired.

This application note will introduce time domain and DTF measurement techniques for identifying the
location and relative amplitudes of discontinuities while operating in the field.

(a) Frequency response (b) Time response

Figure 1a. Measured VSWR as a function of frequency (1a), and the time domain response showing several discontinuities as a function of time (1b).
Time domain measurement basics

Time domain analysis is very useful


to observe the effects of mismatches
along a transmission line system.
When an RF or microwave signal
propagates along a transmission
line, a portion of the signal will be
reflected back from any disconti-
nuities encountered along the path.
Using time domain analysis, the loca-
tion of each discontinuity is displayed
as a function of time along the x-axis
and the amplitude of the reflected
signal, or S11, is plotted along the
y-axis. Knowing the propagation
velocity along the transmission line
allows the time measurement to be
scaled to physical distance. It is also
possible to examine the time domain
response of a transmitted signal, or
S21, but this measurement requires
connection at two ports of a system.
As field measurements often limit
access to only one port of a system,
this application note will primarily Figure 2. Configuration and measurement of a time domain response showing three
focus on the time domain response discontinuities along a coaxial transmission line
of reflection, or S11, measurements.

Figure 2 shows a simple example The measurement from any one-port be simultaneously displayed in time
of a time domain response for two or two-port device can be represented and frequency creates a powerful
short sections of coaxial line con- in the time domain and/or the fre- analysis and problem solving tool.
nected together with an adapter quency domain. If a measurement is Fortunately, modern test instrumenta-
and terminated in a 50-ohm coaxial made in one domain, then the other tion, such as a VNA, includes this
load. If the input to the line is excited domain can be calculated using a mathematical transformation as part
with a short impulse waveform, well-known mathematical technique of the firmware allowing the user to
then it is possible to observe the called the Fourier Transform (FT). This display either time domain data or
reflected impulse response from transform provides a universal problem frequency domain data, or both. For
each discontinuity as a function of solving method that allows one to example, Figures 1a and 1b show the
time. In this case, there are reflected examine a particular measurement frequency measurement and time
signals from the input connector (1), from an entirely different viewpoint domain transform recorded using the
the adapter (2) and the termination [2]. If a measurement is recorded Keysight FieldFox VNA.
(3). Excessively large peaks would be using a time domain method, then
representative of problems in one or a FT calculation will result in a
more of these components. It is also frequency domain representation of
possible to excite the line with a the data. Alternatively, if frequency
step waveform and details of this domain data is initially captured, then
technique are shown in the Keysight an Inverse FT (IFT) will result in a
Time Domain Application Note time domain representation of the
1287-12 [1]. data. The fact that the same data can

3
Instruments with time domain capability

There are two basic instruments A vector network analyzer, such as


capable of displaying the time domain the Keysight FieldFox N9918A VNA is
response of individual discontinuities primarily a frequency domain instru-
along a transmission system; the ment with capability of measuring the
time domain reflectometer (TDR) and reflected and transmission charac-
the vector network analyzer (VNA). teristics of one and two-port devices.
A time domain reflectometer uses a Using error-corrected data measured
traditional method of launching an in the frequency domain, the
impulse or a step waveform into the response of a network to an impulse
test device and directly measuring the or step function can be calculated
response as a function of time [3]. using the IFT and displayed as a
Using a step generator and broadband function of time. As the VNA uses
oscilloscope, such as the Keysight narrowband measurement receivers,
86100D DCA with 54754A TDR the dynamic range will be typically
Module, a fast edge is launched into higher than oscilloscope-based TDR
the transmission line. The incident systems. Also a VNA includes time
and reflected voltage waves are moni- domain capability for measuring
tored by the broadband oscilloscope band-limited devices, called bandpass
and the position of each discontinuity mode, which will be discussed later
can be displayed as a function of time. in this application note. Lastly, there
is a configuration of a FieldFox VNA,
known as a cable and antenna (CAT)
analyzer, which performs the same
frequency-to-time domain transforma-
tion but scales the time measurements
to an equivalent physical distance to
aid the operator in quickly locating
faults in RF and microwave
transmission lines in the field.

4
Measurement example using a horn antenna

Figure 3 shows a test configuration In practice, it could be possible that


that will be used to examine the metallic structures placed near an
frequency domain and time domain antenna system could have a nega-
responses of an over-the-air measure- tive effect on the system performance
ment using an X-band waveguide as reflections from these additional
horn antenna and a separate metal structures could create ripples large
plate placed near the antenna. The enough to cause an out-of-spec
high-gain horn antenna is connected condition. As mentioned, examining
to a short length of WR-90 waveguide only the frequency response does
transmission line and the transmis- not provide sufficient insight into the
sion line is attached to a waveguide- root cause of the problem(s). The
to-coaxial adapter for connection to time domain response under these
the VNA. The VNA is calibrated for measurement conditions will be
S11 at the plane where the adapter examined next.
is connected to the instrument port.
The measurements will include the
effects of all the components shown Coaxial Horn
adapter Waveguide antenna
in Figure 3 and also include effects
from other items in the surrounding
environment. The metal plate is a Metal
0.3 meter square aluminum plate plate
mounted to a tripod and positioned
in front of the antenna. The distance
between the metal plate and the
antenna is varied in order to examine
changes in the frequency and time Figure 3. Configuration for measuring the frequency domain and time domain responses
responses. of an over-the-air measurement using an X-band horn antenna

Figure 4 shows the measured


frequency response of the system
under three test conditions; two with
the metal plate near the antenna
and one without. The yellow curve
shows the measurement without the
metal plate. There is a small ripple
in the measured S11 response due
to interaction between the various
transmission lines and reflections
from items in the surrounding
environment. The blue trace shows
the frequency response with the
metal plate positioned at 1.7 meters
from the horn antenna. In this case,
there is a larger amount of ripple in
S11 when compared to the case with
no plate. The orange curve shows
the results when the plate is moved
closer at a 1.2 meter separation. Here Figure 4. Comparison of the measured frequency response of the X-band antenna system
the ripple is even more pronounced under three test conditions; two with the metal plate near the antenna and one without
as the reflected signal from the plate the metal plate
has larger received amplitude.

5
Measurement example using a horn antenna (continued)

Figure 5 shows the time domain


response of the antenna system
using a Keysight FieldFox VNA with
the time domain option. For the case
without the metal plate, shown as the
yellow trace, the peaks associated
with signal reflections from the sur-
rounding environment are relatively
low in comparison to those from the
adapter and antenna. This type of
measurement would provide a good
baseline for a new system installation
and could be used as a comparison as
more antenna components and struc-
tures are added around this system.
Figure 5 also shows the time domain
response when the metal plate is Figure 5. Time domain response of the antenna system under three test conditions;
positioned near the antenna. For the two with the metal plate near the antenna and one without the metal plate
measurement at 1.7 meters, shown
as the blue trace, there is a large As shown earlier in Figure 4, the
peak representing the reflection from environmental reflections introduce
the metal plate and a smaller peak to a ripple in the frequency domain
the right representing the reflection response. It would be useful to
from the legs of the tripod. When the confirm the system performance
plate and tripod are moved closer without the effects from the environ-
at 1.2 meters, shown as the orange ment. Under this test requirement,
trace, the measured time domain the VNA has a gating function that is
response shows an equivalent time configured in the time domain to filter
shift in these peaks. From these out specific reflections and allow the
measurements, you can see that the frequency domain response to be
peaks associated with the adapter viewed without the effects from these
and antenna are static as these reflections. The next section will
components were not changed during introduce the basic concepts of time
the testing. There is also a noticeable domain filtering also known as gating.
increase in the peak amplitude asso-
ciated with the metal plate when the
plate is moved closer to the antenna.
This increase in peak amplitude is
due to the reduction in space loss
as the signal now propagates over a
shorter distance.

6
Gating and frequency response

The basic concept of gating is to To continue with the antenna shown with gated response, there
introduce a time filter to remove example discussed in the previous is a reduction in the ripple across
unwanted reflections from the time section, Figure 7a shows the time the measured frequency span as the
domain response. This time filter domain response with and without reflections from the environment
works in the same way as a fre- a gating function. In this example, have been filtered out from the
quency filter would in the frequency a bandpass gate is configured to measurement. Gating is a very power-
domain. Once the undesired respons- remove the effects of the metal plate ful tool to examine the reflection
es are removed from the time domain, and the environment. The yellow and transmission properties of a
the frequency domain response of the trace is the original measurement component or system by selectively
remaining devices can be observed. including the peak from the metal removing specific responses from the
Figure 6 is a simplified example of an plate reflection. The blue trace shows measurements in any domain.
ideal time domain response having the time domain response with the
three peaks. For this example, the bandpass gate configured to filter any It should be noted that when a
time filter or “gate” is configured to response to the right of the antenna. discontinuity along a transmission
remove the two outside peaks. By With the bandpass gate applied, line reflects energy, that portion of
adjusting the gate’s time center and the time response only includes the the energy will not reach subsequent
span, an equivalent bandpass filter effects of the coaxial adapter and the discontinuities further down the line.
can be created in the time domain. antenna. It is also possible to remove This effect can hide, or “mask,” the
Once the gate is activated, the upper a single peak, or group of adjacent true response of the later discontinu-
and lower peaks are removed from peaks, using a notch gate. ity. This condition also occurs when
the measurements in the time and the transmission path has insertion
frequency domains. Along with the Figure 7b shows a comparison loss along the length of the line. The
bandpass gate, the FieldFox VNA has between the original frequency gating function does not compensate
another type of gate that can notch response including the plate and for the masking effect and masking
out a single peak, or grouping of environmental reflections, shown in will be discussed in the next section
peaks, leaving behind the remaining yellow, and the response when the of this application note.
peaks above and below this filter gate is applied, shown in blue. As
response. In either case, once the
gate is properly configured in the time
domain, the user can return the VNA
to the frequency domain to examine
the frequency response of the system
with the effects of the filtered peaks.

(a) Time response

(b) Frequency response

Figure 6. Simplified example of an ideal Figure 7. (7a) time and (7b) frequency domain responses of the X-band antenna system
time domain response having three peaks with and without a gate function applied to the measured data
and the results when a time “gate” is
applied to the original response

7
Masking in coaxial lines

Figure 8 shows the time domain the reflected signals could be below It should be noted that the measured
response of a short length of RG58 the dynamic range of the VNA and amplitude on any time domain display
coaxial cable with the end of the would not be observed on a time is actually an average value taken
cable terminated in an open circuit. domain display. This would also occur across the entire measured frequency
The time response shows a single if the transmission path contained a range. Therefore, if the transmission
large peak representing the reflec- very large discontinuity that reflected system includes couplers, filters and
tion from the end of the cable. It is most of the energy and prevented any other items with varying insertion
expected that an ideal open circuit later reflections from being observed loss, then applying a single factor
would have an S11 of 0 dB resulting on the time domain display. to compensate for the insertion loss
from all of the energy being reflected as a function of frequency may not
back to the source. As shown in If the insertion loss of the cable is provide accurate amplitude results.
Figure 8, the measured peak for this known, the Keysight FieldFox VNA This effect will be shown in the next
test configuration has a lower value and CAT analyzers allow the cable section of this note.
at –4.2 dB. The measured peak is loss to be entered into the instru-
lower by twice the average cable loss ment. Compensating for the cable
resulting from the combination of the loss will improve the amplitude
signal loss in the forward direction measurements in the time domain.
and then the loss passing through It should be noted that coaxial
the cable a second time after being cable insertion loss is a function of
reflected from the open. The insertion frequency with the loss increasing
loss of the cable is “masking” the at higher frequencies. The cable
true response from the open circuit. insertion loss factor can be entered
Gating around the peak response as a single value representing the
from the open would not correct the average loss or, when using the CAT
amplitude measurement as gating is mode DTF, can be entered into a table
a post-process that occurs after the of frequency/loss pairs. The FieldFox
measurement has been acquired. CAT mode also includes a set of
It is important to note that if the built-in cable specifications for quick
transmission system has very high selection based on the cable types
insertion loss, then the amplitude of installed in the system.

Figure 8. Test configuration and time domain response of a short length of RG58 coaxial cable terminated
in an open

8
Measurement example with bandpass ilter

Figure 9 is the test configuration for function is completely reflected from through the filter in the forward
a system that includes two coaxial a single point on the transmission direction and then passes through
cables interconnected with an inline line. As shown on this measurement, the filter a second time in the reverse
bandpass filter. The longer cable is the time response from the open direction after being reflected from
terminated in either a 50-ohm load or has a rather complex shape. The the open. For this example, the band-
an open ended cable. Being able to reflection from the open is modified pass filter is masking the reflected
swap terminations at the end of the by the transmission response of the response from the open.
transmission system is very useful bandpass filter as the signal passes
when troubleshooting a system and
attempting to identify the end of a
transmission system. An open or
short is typically used to provide a
large reflection for identifying the
end of the cable in the time domain
response.

This simple example will show the


effects of masking as the reflected Figure 9. Test configuration for a system that includes two coaxial cables interconnected
with a bandpass filter and terminated in either an open or a 50-ohm load
signal passes through the inline band-
pass filter. The measured frequency
domain response of S11 is shown in
Figure 10a. For this measurement,
the cable is terminated in a 50-ohm
load to simulate the effects of a
system with the proper termination.
The time domain mode is activated
and the measured time response is
shown in Figure 10b. In this case,
the cable is terminated with an open
in order to see a large reflection in
the time response at the end of the
cable. As shown in Figure 10b, the
time domain display shows a series
of peaks associated with the various
components along the transmission (a) Frequency response
path. The first peak on the left is the
reflection from the input cable. The
next grouping of peaks is associated
with the bandpass filter. A filter is
a set of resonators that interact in
the time domain to create a unique
response. Unless a reference filter is
used for comparison, it may be diffi-
cult to determine if the filter response
appears correct when examining a
time domain measurement. Keysight
provides several excellent application
notes on filter tuning using time
domain techniques [4, 5]. The last set
of peaks on the right are associated
with the open circuit. It is typically (b) Time response
expected that an open circuit would
result in a single peak as the impulse Figure 10. Measured frequency domain response of S11 (10a) and (10b) measured time
domain response showing the reflection characteristics from the input connector, filter
and open ended cable
9
Measurement example with bandpass ilter (continued)

To examine the effects of masking


through the inline bandpass filter
shown in this example, Figure 11
shows the time and frequency
response of this system with a gate
function applied around the reflected
signal from the open. Figure 11a
shows the time domain response
with and without the gate. The blue
curve shows the time response with
the gate centered over the reflection
from the open ended cable. Figure 11b
shows the frequency domain
response of gated signal from the
open. This response shows that the
system’s inline filter is masking the (a) Time response
ideal response from the open ended
cable. In this case, the frequency
domain response shows a typical
bandpass transmission response.
As mentioned, the gated response
around the open shows the results
when the signal passes twice
through the filter, once in the forward
direction and once in the reverse
direction after reflecting from the
open. This type of measurement
would be useful when verifying
that system components have been
properly installed and when the user
has access to only one end of the
system.

It should be noted that in Figure 11b,


(b) Frequency response
the edges of the gated frequency
response have increasing amplitude.
Figure 11. Time domain response of the transmission system with an inline bandpass
This increase is not a function of
filter (11a) and (11b) frequency domain response of the gated signal from the open ended
the inline bandpass filter but rather
cable having a masking effect
an artifact of the signal processing
involved in the gating function. It
may be possible to reduce the edge
effects in the frequency domain by
adjusting the characteristics of the
gate function from the front panel of
the FieldFox VNA. Details on optimiz-
ing the gate window are described
in the FieldFox User's Guide and the
Keysight Time Domain Application
Note 1287-12 [1].

10
Relating time measurements to physical distance

Once it has been determined that round trip travel time, tRT. The round as the dielectric will have a VF of 0.7
a system’s insertion loss is higher trip time is divided by two for an as the dielectric constant of PTFE is
than expected or the return loss and estimate of the one-way travel time 2.1. When using a VNA to measure
associated VSWR are out of spec, it to the associated discontinuity. The the location of faults, it is important
becomes necessary to find the physi- measured DTF is the one-way time to accurately enter the velocity factor
cal location of the fault(s) along the multiplied by the velocity of the sig- into the VNA. When using a FieldFox
transmission system. If the velocity at nal, vcable, in the medium. The velocity VNA to measure DTF, once the VF
which the signal travels in the cable is given in meter/second but may is correctly entered, the FieldFox
is known, then the physical location also be calculated in feet/second. It marker displays both the time and
of the damage can be calculated from is known that the signal velocity in a the corrected one-way distance to
a reflection measurement and using vacuum is equal to the speed of light, the discontinuity. Also, when using a
the following equation. c. The signal velocity in a medium FieldFox configured in CAT mode, all
( is the speed of light, c, multiplied time domain display parameters are
DTF (meters) = (t 2
RT
(vcable ) by the Velocity Factor (VF) which is
a value less than or equal to 1. The
automatically converted to distance.
The Keysight Application Note [6]
Where vcable = (VF)(c) in meters/second and VF = 1/
√ er VF is inversely related to the square provides additional information on
With a VNA such as the FieldFox, a root of the dielectric constant, er , of estimating the velocity factor for an
marker placed on a peak in the time the transmission line. For example, unknown cable.
domain response represents the a coaxial cable that uses solid PTFE

Relationship between frequency span and pulse width


When a VNA measures the As an example, Figure 12 shows the
S-parameters of a component or measured time domain response
system in the frequency domain, of two coaxial cables connected
the data is measured across a user- together with a coaxial adapter and
selectable frequency range (span). terminated in a 50-ohm load. Figure 12a
shows the time response using a
When operating in the time domain VNA configured with a 500 MHz
mode, the pulse width of the impulse span. Here the individual disconti-
is approximately related to the inverse nuities are not observable due to the
of the frequency span. wide pulse width in the time domain.
Figure 12b shows the measurement (a) Time response with 500 MHz span
1
Pulse width (seconds) ≈ when the frequency span is increased
Frequency span
to 4 GHz. In this case the individual
A wider span will result in more discontinuities are clearly observed
detail, or resolution, in the time using a narrower pulse width.
domain. This is especially useful
when attempting to identify two While having a large frequency span
closely spaced discontinuities in improves the time domain resolution,
the time domain. The equation is it has a negative effect on the total
approximate due to the fact that a time span or maximum range that
window is applied to the measured can be measured. The next section
data in order to improve the dynamic discusses the relationship between
range of the time domain response frequency span and the maximum (b) Time response with 4 GHz span
which slightly increases or decreases measurable time.
the pulse width. Additional details Figure 12. Measured time domain response
concerning the window function can of two coaxial cables connected together
be found in Keysight Time Domain with a coaxial adapter and terminated in a
Application Note 1287-12 [1]. 50-ohm load

11
Relationship between frequency span and time span

When a VNA measures the T  T                   
S-parameters of a component or and a velocity factor (VF) equal to 0.66
system in the frequency domain, the
Span (GHz) Max time (microseconds) Range (meters)
data is measured across a selected
frequency range (span) and stored 2.5 2 396
at discrete points across this span. 5.0 1 198
The frequency spacing between data
points, Δf, is directly proportional
to the frequency span and inversely
proportional to the total number of Under some conditions, it may be It is important to note that it is not
measurement points, as shown in possible to configure a VNA to dis- possible to simultaneously optimize
the following equation. play a time domain response beyond for narrow pulse width, discussed in
the calculated maximum time span. the previous section, and maximum
 equency span
∆f = Under this condition, it is possible time range discussed here. A narrow
(Points —1)
to observe a spurious response, also pulse width requires a wide frequency
called an alias, which appears to span but a wide frequency span
Along with the frequency span, the be real response but is actually an reduces the maximum time span.
number of points is also selectable artifact of the mathematical transfor- Therefore a tradeoff between the
by the user. The frequency spacing, mation from the frequency domain time resolution and the time span
Δf, is then automatically calculated into the time domain. The FieldFox will need to be made. Fortunately
by the VNA. The FieldFox VNA can be provides a table showing whether the for some types of devices, there is
configured with the number of points current measurement is alias-free. another time domain mode available
set from 101 to 10,001. For example, When the FieldFox flags the pos- on the FieldFox VNA that will further
using a span of 1 GHz and 101 points, sibility for an alias, there is a simple improve the time resolution without
the data is then measured every test to verify if a response is real. sacrificing time span and this mode
10 MHz = Δf. First, place a marker on the peak in will be introduced in the next section
question and note the time (distance) of this application note.
When measuring electrically long to the fault. Next, change the start or
components and systems, it is impor- stop distance on the analyzer. A true
tant that the VNA is properly config- time domain peak will not move in
ured to measure all the way to the time (distance) but an alias response
end of the transmission path. This is will appear to move. More informa-
especially important when long cable tion concerning aliases can be found
runs are installed in a system such in the FieldFox User's Guide and the
as in tower mounted antennas. We Keysight Time Domain Application
find that the maximum time span is Note 1287-12 [1].
directly proportional to the number of
measurement points in the frequency
domain and inversely proportional
to the measured frequency span as
shown in the following equation.
Here, the divide by 2 is the result that
reflection measurements are a round
trip through the transmission system.
For transmission measurements, such
as S21 and S12, the divide by 2 is not
required.
1 (Points —1)
Maximum time ≈ =
2∆f 2 (Frequency span)

12
Coniguring bandpass and lowpass time domain modes

The FieldFox VNA and CAT analyzers waveguide components are included
have two time domain mode selections as part of the transmission path.
for optimizing time measurements.
The modes are lowpass and band- When using bandpass mode, the
pass. The mode selection depends on instrument’s start and stop frequency
the operating frequency range of the settings should be set to approxi-
DUT. For example, when measuring mately the operating range of the
broadband devices, such as coaxial DUT. When the DUT is broadband, the
cables and adapters, their operating lowpass mode would be configured
frequency range extends from DC and over a much larger frequency range.
into the RF and microwave frequency The lowpass mode extrapolates the
ranges. For these devices, the VNA measured frequency range down to
is ideally operated in the lowpass DC and then mirrors the measured
mode for all time domain measure- frequency range data about DC to
ments. When the device is specified greatly improve the time resolution
to operate over a narrow frequency capability. Additional information
range then the bandpass mode on regarding these modes can be found
the VNA will provide the best results. in the FieldFox User’s Guide and the
Bandpass mode is primarily used Keysight Application Note 1287-12 [1].
when filters, couplers, antennas and

Conclusion

The time domain feature on the


FieldFox is a powerful tool for verifying
and troubleshooting complex
components and systems. Beyond
the examples discussed in this note,
there are numerous applications and
instrument configurations for per-
forming time domain measurements
that are discussed in other Keysight
application notes listed in the refer-
ence section of this note. Keysight
will gladly supply these notes through
the website or with live technical sup-
port. Keysight also provides local and
on-site training for all handheld and
benchtop VNA and CAT analyzers.

13
References
[1] Keysight Application Note 1287-12, Time Domain Analysis Using a Network
Analyzer, Literature Number 5989-5723EN, May 2012.
[2] Brigham, E. Oran, The Fast Fourier Transform, Prentice-Hall, 1974.
[3] Keysight Application Note 1304-2, Time Domain Reflectometry Theory,
Literature Number 5966-4855E, May 2006.
[4] Keysight Application Note 1287-8, Simplified Filter Tuning Using Time
Domain, Literature Number 5968-5328E, July 2000.
[5] Keysight Application Note 1287-10, Network Analysis Solutions Advanced
Filter Tuning Using Time Domain Transforms, Literature Number
5980-2785EN, March 2001.
[6] Keysight Application Note, Techniques for Precise Cable and Antenna
Measurements in the Field, Literature Number 5991-0419EN, August 2012.

Carry precision with you.


Every piece of gear in your field kit had to prove its worth. Measuring up and
earning a spot is the driving idea behind Keysight’s FieldFox analyzers. They're
equipped to handle routine maintenance, in-depth troubleshooting and anything in
between. Better yet, FieldFox delivers Keysight-quality measurements—wherever
you need to go. Add FieldFox to your kit and carry precision with you.

Related literature Number


FieldFox Combination Analyzers, Technical Overview 5990-9780EN
FieldFox Microwave Spectrum Analyzers, Technical Overview 5990-9782EN
FieldFox Microwave Vector Network Analyzers, Technical Overview 5990-9781EN
FieldFox Handheld Analyzers, Data Sheet 5990-9783EN
FieldFox Handheld Analyzer, Coniguration Guide 5990-9836EN
FieldFox N9912A RF Analyzer, Technical Overview 5989-8618EN
FieldFox N9912A RF Analyzer, Data Sheet N9912-90006
FieldFox N9923A RF Vector Network Analyzer, Technical Overview 5990-5087EN
FieldFox N9923A RF Vector Network Analyzer, Data Sheet 5990-5363EN

Download additional application notes, watch videos, and learn more:


www.keysight.com/find/FieldFox

14
15 | Keysight | Techniques for Time Domain Measurements Using FieldFox Handheld Analyzers – Application Note

myKeysight For more information on Keysight


Technologies’ products, applications or
www.keysight.com/find/mykeysight
services, please contact your local Keysight
A personalized view into the information most relevant to you.
office. The complete list is available at:
www.axiestandard.org www.keysight.com/find/contactus
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and Americas
semiconductor test. Keysight is a founding member of the AXIe consortium. Canada (877) 894 4414
ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of Brazil 55 11 3351 7010
the PCI Industrial Computer Manufacturers Group. Mexico 001 800 254 2440
United States (800) 829 4444
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty Asia Paciic
Keysight’s commitment to superior product quality and lower total cost Australia 1 800 629 485
of ownership. The only test and measurement company with three-year China 800 810 0189
warranty standard on all instruments, worldwide. Hong Kong 800 938 693
India 1 800 112 929
Keysight Assurance Plans Japan 0120 (421) 345
www.keysight.com/find/AssurancePlans Korea 080 769 0800
Up to five years of protection and no budgetary surprises to ensure your Malaysia 1 800 888 848
instruments are operating to specification so you can rely on accurate Singapore 1 800 375 8100
measurements. Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
www.keysight.com/quality
Keysight Technologies, Inc. Europe & Middle East
DEKRA Certified ISO 9001:2008 Austria 0800 001122
Belgium 0800 58580
Quality Management System
Finland 0800 523252
Keysight Channel Partners France 0805 980333
www.keysight.com/find/channelpartners Germany 0800 6270999
Get the best of both worlds: Keysight’s measurement expertise and product Ireland 1800 832700
Israel 1 809 343051
breadth, combined with channel partner convenience.
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
ATCA, AdvancedTCA, and the ATCA logo are registered US trademarks of the PCI Industrial Russia 8800 5009286
Computer Manufacturers Group.
Spain 0800 000154
www.keysight.com/find/FieldFox Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637

For other unlisted countries:


www.keysight.com/find/contactus
(BP-07-10-14)

This information is subject to change without notice.


© Keysight Technologies, 2012 - 2014
Published in USA, August 3, 2014
5991-0420EN
www.keysight.com

You might also like