Optimize_Issue_5_January_2017
Optimize_Issue_5_January_2017
SPEXCertiPrep.com/ECS
A publication by
Environmental Consulting & Supplies
a Division of
Optimize
January 2017
For all discussions, we will assume you are running QQ It’s a very good idea to purchase a hand-held
Volatiles by GC/MS (524, 624, 8260) and/or Semivolatiles Helium leak detector. It’s a lot to spend I
by GC/MS (525, 625, 8270). Other non-Mass Spec know but it’s an invaluable tool when trying
to find small leaks.
applications are not addressed here. Since it is rare for
these units to “break”, the majority of problems with QQ Never try to detect leaks by squirting a fitting
with soapy water to see if it bubbles. You can
them ends up being with conditions being set incorrectly get some of the soap into the lines and that
or for the system to develop leaks and/or active sites. would be disastrous.
Detecting and Resolving Leaks: If a leak does occur in your inlet, check the following:
Here are some good ideas to follow in order to QQ Carefully inspect the glass liner. If it is
chipped, that would cause a leak.
detect and fix leaks:
QQ Check the Viton O-ring that seals the liner. It
QQ The MSD can detect leaks where air is being
should feel elastic. If in doubt, change it.
sucked into the ion source (by virtue of the
presence of large amounts of ions 28 and QQ Replace the septum if it has been punctured
32), but it has no ability whatsoever to detect many times or has been crushed out of shape.
leaks where Helium and analytes are being Be careful not to over tighten the septum
lost OUT of the system. To detect these leaks nut as that can cause the septum to become
you need to monitor flows and/or use a leak crushed. 1/4 turn past finger tight should do
detector. it on the septum nut.
QQ Once per month, when the system is running QQ Replace the inlet seal if the face seems worn
optimally, measure the split vent flow and out or warped. This sometimes happens if it
record it in a log book. Monitoring changes has been used and reused a lot or has had the
in split vent flow is the best way to detect face cleaned to remove debris.
leaks. This applies to 5890 GCs. 6890 GCs QQ Re-seal the fitting from the column to the
will automatically add more flow if a leak inlet. Be careful not to over tighten the fitting,
exists, so in this case if you notice reduced as this can cause the ferrule to get cracked.
areas of your analytes, you will need to check
the inlet with a Helium leak detector. Coping with Active Sites
QQ Whenever you dismantle the injector to do Another common problem with inlets is the
maintenance, ALWAYS measure the split vent
flow immediately before and immediately development of active sites. An active site is
afterwards. If you have done it correctly, they any buildup of contamination that would cause
will be identical. If not, you need to check a compound to react with or adsorb to, thereby
everything you did prior to heating up the compromising the integrity of the analysis.
inlet and running the system. As above, this
applies to 5890 GCs. For 6890 GCs you will For Volatiles, active sites in the inlet generally only
need to check the inlet with a Helium leak occur if a sample foams over. Otherwise, you would
detector. typically only do your injector maintenance once a
QQ Monitor the areas of your first Internal year on average.
Standard. If the areas drop off, yet the For Semivolatile analysis, fighting active sites is
sensitivity of the MSD (as measured by the
raw abundance of ion 69 in Manual Tune) is nearly an every day battle because of the beating the
constant, check for leaks. For Semivolatile inlet takes by having numerous Methylene Chloride
analysis the leak is likely to be in the inlet. extracts injected into it. Active sites in the inlet can
For Volatiles, it’s more likely to be in the cause many compounds (especially Benzidine and the
purge and trap than the GC so you should Phenolics) to show drastically reduced response.
leak check the purge and trap.