ITC 1
ITC 1
HEMSHELL
TRANSCENDING BEYOND TESTING
CALIBRATION CERTIFICATE
Certiicate No. JHS/BRD/CAL/24-25/196A ULR NO. CC292024000000451TT Date of Calibration 19-12-2024
Mechanical/UTM, Tension, Creepand
Date of issue 21-01-2025 Discipline / GroupTorsion Testing Machine Recommended Due Date 19-12-2025
CALIBRATION RESULTS
Applied Load Standard Reading Observed Readings in Counts % Relative
in N in Counts Mean Indication % Relative
+% Expanded
120° 240 ° Mean Error 'g' Repeatability Error
'b' Uncertainty "U
CALIBRATION CERTIFICATE
Certificate No. HS/BRD/ CAL/24-25/196B ULR NO. CC292024000000452F Date of Calibration19-12-2024
Date of Issue 21-01-2025 Mechanical /UTM, Tension, Creep and
Discipline / GroupTorsion Testing Machine Recommended Due Date 19-12-2025
Customer Name & Address
Description &ldentification of Machine Calibration Conditions
Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make /Model |LLOYD/LS1 Work Instruction No. HS/CP/07
ITC Life Science & Tech centre, Machine Capacity 1KN Condition of Machine Satisfactory
1st main, 1st phase, Machine Serial No. 202336 Environmental Temperature |23°C
Load Cell Capacity 5ON Calibrated at ISite (Customer's Premises)
Peenya Industrial Area, Load Cell Serial No. ITension Mode
Bangalore-S60058 5ON1621 |CalibrationMode
|Customer Identification No. NA
Location of the Machine
Resolution of Machine
Packaging
Lab |Calibrated By Mr. Arjun PN
|0.001
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard: Reference Standard 2
Name |LOAD CELL WITH INDICATOR
Make/ Capacity/ Class ADI ARTECH/5ON /1
Load Cell Sr. No. |1973886
Indicator Sr. No. 10970
Calibration Certificate No. / CC No. ULR-CC239023000008571F --
Calibration Certificate Due On 13-09-2025
Reported Uncertainty in % |0.13
CALIBRATION RESULTS
Applied Load Standard Reaing Observed Readings în Counts % Relative
in N in Counts Mean Indication
% Relative
t% Expanded
120 240° Error'q' Repeatability Error
Mean Uncertainty 'U
Calibrated Range: 1N TO 50N
1000 1002 1002 1001 1002 -0.15 0.10 0.50
5 4998 5002 5003 5007 5004 -0.12 0.10 0.50
10 10006 10010 10015 10019 10015 -0.09 0.09 0.50
20 20030 20034 20040 20049 20041 -0.05 0.07 0.50
30 30045 30051 30057 30063 30057 -0.04 0.04 0.50
40 40064 40079 40085 40094 40086 -0.05 0.04 0.50
50 50076 50082 50091 50100 50091 -0.03 0.04 0.50
Residual Load 0.004 0.003 0.002 N
9
Classification of Machine CLASS1
Estimated Mean Relative Error, E = q tU 0.65 %
Notes:
1. The calibration results reported are valid at the time of and under the conditions of measurement and results relate only to the items calibrated
2. This report shall not be reproduced, except in full without approval of Hemshell Services LLP.
3. Any correction in this certificate, invalidates the certificate.
4. The classification of atesting machine calibrated and certifiedto meet aspecific class does not ensure that the acuracy inchudíng uncertainty will be less thana
specific value.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/Instrument/Equipment covered by this report due to damage caused to it afterissuance of this report
6. All the reference standards used are traceable to National Standards.
|7. Hemshell Services LLP uses Temperature compensated load cells so temperature correction is not required.
8. The machine shall in any case be Recalibrated, if it is moved to anew location necessitating dismantling or if it is subject to measure repairs or adjustm ents.
9.% Expanded Measurement Uncertainty at 95% confidence level with coverage facor
Checked By
VADODARA
Authorised Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar
CALIBRATION CERTIFICATE
Certificate No. |HS/BRD/CAL/24-25/196C ULR NO. CC292024000000453F Date of Calibration 19-12-2024
Date of Issue |Mechanical /UTM, Tension, Creep and
|21-01-2025
Discipline/ GroupTorsion Testing Machíne Recommended Due Date19-12-2025
Customer Name & Address Description & ldentification of Machine Calibration Conditions
Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make/ Model LLOYD/LS1 Work Instruction No. HS/CP/07
Machine Capacity
ITC Life Science & Tech centre, 1st Machine Serial No.
1KN
202336
Condition of Machíne Satisfactory
23°C
Environmental Temperature
maln, 1st phase, Peenya Industrial Load Cell Capacity 100N Calibrated at Site (Customer's Premises)
Area, Bangalore-560058 Load Cell Serial No. 100N1154 Calibration Mode Compression Mode
Customer Identification No. NA
Location of the Machine Packaging Lab |Calibrated By Mr. Arjun PN
Resoution of Machine 0.001
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard1 Reference Standard 2
Name ILOAD CELL WITH INDICATOR |LOAD CELL WITH INDICATOR
Make/Capacity/ Class |ADIARTECH/500N/1 ADI ARTECH/50N/1
Load Cell Sr. No. 1991218 1973886
Indicator Sr. No. HS/F/LCI/039/2023 10970
Calibration Certificate No./ CCNo. ULR-CC239023000008570F (CC- 2390 ULR-CC239023000008571F
Calibration Certificate Due On |13-09-2025 13-09-2025
+Reported Uncertainty in % 0.11 ).15
CALIBRATION RESULTS
Applied Load Standard Reading Observed Readings in Counts % Relative
in N in Counts Mean indication % Relative
±% Expanded
120 240 Mean Error'q' Repeatability Error
Uncertainty 'U"
Calibrated Range: 1N to 100N
1001 1003 1002 1002 1002 -0.13 0.10 0.62
5 5003 5008 5011 5012 5010 -0.15 0.08 0.62
10 10008 10013 10017 10019 10016 -0.08 0.06 0.62
20 20031 20040 20047 20052 20046 -0.08 0.06 0.62
50 50083 50088 50093 50097 50093 -0.02 0.02 0.62
50 10005 10011 10020 10022 10018 -0.13 0.11 0.62
60 12006
16006
12016
16019
12019
16021
12023
16024
12019
16021
olela-0.10
-0.11 0.06
0.03
0.62
0.62
100 20008 20023 20027 20029 20026 -0.09 0.03 0.62
Residual Load 0.001 0.003 0.002 N
Classification of Machine |CLASS 1
Estimated Mean Relative Error., E =atU 0.77 %
Notes:
1. The calibration results reported are valid at the time of and under the conditions of measuremnent and results relate only to the items calibrated.
2. This report shall not be reproduced, except in full without approval of Hemshell Services LLP.
3. Any correction in this certificate, invalidates the certificate.
4. The classification of a testing machine calibrated and certified to meet a specific class does not ensure that
the accuracy including uncertainty will be less than a
specific value.
S. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/Instrument/Equipment covered by this report due to damage caused to it after issuance of this report.
6. AIl the reference standards used are traceable to National Standards.
7.Hemshell ServDces LLP uses Temperature compensated load cells so temperature correction is not
required.
8. The machine shall in any case be Recalibrated, if it is moved to a new location necessitating
dismantling or if it is subject to measure repairsor adjustments.
9.% Expanded Measurement Uncertainty at 95% confidence level with
coveragefacior
KGES
VADODARA
Checked By
Mr. Vinay Mahadíik Authorised Signatory
Mr. Vinod Tulaskar
FF No. FF-HS-25 Amend. No.: 01 Amend. Date: 25-11-2024
Technical Manager
Page no. lof 1
HEMSHELL
TRANSCENDING BEYOND TESTING
CC - 2920
CALIBRATION CERTIFICATE
Certificate No. HS/BRD/CAL/24-25/196D ULR NO. CC292024000000454F Date of Calibration 19-12-2024
Mechanlcal /UTM, Tension, Creep and
Date of Issue 21-01-2025 Discipline/ GroupTorsion Testing Machne Recommended Due Date 19-12-2025
Checked By
VADODARA J
Authorised Signatory
Mr. Vínay Mahadik Mr. Vinod Tulaskar
Standard
Set readings Observed Reading in Machine Console %Repeatability fo Accuracy
In mmn Reading In mm
Mean Reading Error Error
mm
Note: According to standard ASTM-E-2309 acceptable %Accuracy &Repeatability error is +/-1% of reading
THE REPEATED SET OF READINGS ARE DIFFER IN PLACE TAKEN EARLIER.
R
oVADODARA
Checked By Approved Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar
Technical Manager
NOTE:
1. This report refers only to the particular item calibrated at customer site.
2. The calibration results reported are valid at the time of and under the conditions of measurement.
|3. This report may not be reproduced, except in full without prior permission of Hemshell Services LLP.
4. Any correction in this certificate, invalidates the certificate.
5. HemshellServices LLP is not liable for any change in calibration data and performance specifications on account of
malfunctioning of Standard/Instrument/Equipment covered bythis report due to damage caused to it after issuance of this report.
DOC NO.HS/03 ISSUE NO.05 DT.27.07.24
NOTE:
1. This report refers only to the particular item calíbrated at customer site.
l2. The calíbration results reported are valid at the time of and under the conditions of measurement.
3.This report may not be reproduced,except in full without prior permission of Hemshell Services LLP.
4. Any correction in this certificate, invalidates the certificate.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of
malfunctioning of standard/Instrument/Equipment covered by this report due to damage caused to it after iSSuance of this report.