The document contains calibration certificates for a Universal Testing Machine (UTM) used by ITC Life Science & Tech Centre, detailing the calibration process, results, and conditions. Each certificate includes information such as the machine's specifications, calibration dates, and results of the applied loads with corresponding readings and uncertainties. The certificates emphasize the importance of maintaining calibration standards and conditions for accurate measurements and include disclaimers regarding the validity of the results.

Uploaded by

vinod tulaskar
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
15 views

ITC 1

The document contains calibration certificates for a Universal Testing Machine (UTM) used by ITC Life Science & Tech Centre, detailing the calibration process, results, and conditions. Each certificate includes information such as the machine's specifications, calibration dates, and results of the applied loads with corresponding readings and uncertainties. The certificates emphasize the importance of maintaining calibration standards and conditions for accurate measurements and include disclaimers regarding the validity of the results.

Uploaded by

vinod tulaskar
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

CC- 2920

HEMSHELL
TRANSCENDING BEYOND TESTING

CALIBRATION CERTIFICATE
Certiicate No. JHS/BRD/CAL/24-25/196A ULR NO. CC292024000000451TT Date of Calibration 19-12-2024
Mechanical/UTM, Tension, Creepand
Date of issue 21-01-2025 Discipline / GroupTorsion Testing Machine Recommended Due Date 19-12-2025

Customer Name &Address Description &ldentification of Machine Calibration Conditions


Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make / Model LLOYD/IS1 Work Instruction No. HS/CP/07
ITC Life Science & Tech centre, Machine Capacity 1KN Condition of Machine Satisfactory
Machine Serial No, 202336 Environmental Temperature 23°C
1st main, 1st phase,
Peenya Industrial Area,
Load Cell Capacity |5ON |Calibrated at Site (Customer's Premises)
Load Cell Serial No. 5ON1621 Calibration Mode Compression Mode
Bangalore-S60058 Customer ldentification No. NA
Location of the Machine Packaging Lab |Calibrated By |Mr. Arjun PN
Resolution of Machine lo.001 N
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard 1 Reference Standard 2
Name LOAD CELL WITH INDICATOR
Make / Capacity/ Class |ADIARTECH/50N/1
Load Cell Sr. No. 1973886
Indicator Sr. No. 10970
Calibration Certifiate No./CC Na. ULR-CC239023000008571F
Calibration Certificate Due On 13-09-2025
t Reported Uncertainty in % 0.15

CALIBRATION RESULTS
Applied Load Standard Reading Observed Readings in Counts % Relative
in N in Counts Mean Indication % Relative
+% Expanded
120° 240 ° Mean Error 'g' Repeatability Error
'b' Uncertainty "U

Calibrated Range: 1N TOSON


1001 1003 1003 1002 1002 -0.13 0.15 0.62
5003 5005 5010 5011 5009 -0.11 0.12 0.62
10 10008 10013 10019 10024 10019 -0.11 0.11 0.62
20 20031 20038 20047 20052 20046 -0.07 0.07 0.62
30 30048 30064 30067 30077 30069 -0.07 0.04 0.62
40 40066 40078 40090 40093 40087 -0.05 0.04 0.62
50 50083 50095 50102 50108 50102 0.04 0.03 0.62
Residual Load 0.004 0.002 0.001 N
dassifiation of Machine CLASS 1
Estimated Mean Relative Error, E =q tU 0.75 %
Notes:
1. The calibration results reported are valid at the time of and under the conditions of measurement and results relate only to the items calibrated.
2. This report shall not be reproduced, except in full without approval of Hemshell Services LLP.
3. Any correction in this certificate, invalidates the certificate.
4.The classification of a testing machine calibrated and certified to meet a specific class does not ensure that the accuracy incuding uncertainty will be less than a
specific value.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/nstrument/Equipment covered by this report due to damage causedto it after issuance of this report.
6. All the reference standards used are traceable to National Standards.
|7. Hemshell Services LLP uses Temperature compensated load cells so temperature correction is not required.
8.The machine shall in any case be Recalibrated, if it is moved to a new location necessitating dismantiing or if it is subject to measure repairs or adjustments.
9.% Expanded Measurement Uncertainty at 95% confidence level with coveragefrcor,k
ER
VADODARA
Checked By Authortsed Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar
Technical Manager
FF No. PF-HS-25 Amend. No.: 01 Amend. Date: 25-11-2024 Page no. lof 1

Hemshell Services LLP


903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujarat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.comn, www.hemshellservices. com
***nd of the Rer
CC- 2920
HEMSHELL
TRANSCENDING BEYOND TESTING

CALIBRATION CERTIFICATE
Certificate No. HS/BRD/ CAL/24-25/196B ULR NO. CC292024000000452F Date of Calibration19-12-2024
Date of Issue 21-01-2025 Mechanical /UTM, Tension, Creep and
Discipline / GroupTorsion Testing Machine Recommended Due Date 19-12-2025
Customer Name & Address
Description &ldentification of Machine Calibration Conditions
Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make /Model |LLOYD/LS1 Work Instruction No. HS/CP/07
ITC Life Science & Tech centre, Machine Capacity 1KN Condition of Machine Satisfactory
1st main, 1st phase, Machine Serial No. 202336 Environmental Temperature |23°C
Load Cell Capacity 5ON Calibrated at ISite (Customer's Premises)
Peenya Industrial Area, Load Cell Serial No. ITension Mode
Bangalore-S60058 5ON1621 |CalibrationMode
|Customer Identification No. NA
Location of the Machine
Resolution of Machine
Packaging
Lab |Calibrated By Mr. Arjun PN
|0.001
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard: Reference Standard 2
Name |LOAD CELL WITH INDICATOR
Make/ Capacity/ Class ADI ARTECH/5ON /1
Load Cell Sr. No. |1973886
Indicator Sr. No. 10970
Calibration Certificate No. / CC No. ULR-CC239023000008571F --
Calibration Certificate Due On 13-09-2025
Reported Uncertainty in % |0.13
CALIBRATION RESULTS
Applied Load Standard Reaing Observed Readings în Counts % Relative
in N in Counts Mean Indication
% Relative
t% Expanded
120 240° Error'q' Repeatability Error
Mean Uncertainty 'U
Calibrated Range: 1N TO 50N
1000 1002 1002 1001 1002 -0.15 0.10 0.50
5 4998 5002 5003 5007 5004 -0.12 0.10 0.50
10 10006 10010 10015 10019 10015 -0.09 0.09 0.50
20 20030 20034 20040 20049 20041 -0.05 0.07 0.50
30 30045 30051 30057 30063 30057 -0.04 0.04 0.50
40 40064 40079 40085 40094 40086 -0.05 0.04 0.50
50 50076 50082 50091 50100 50091 -0.03 0.04 0.50
Residual Load 0.004 0.003 0.002 N

9
Classification of Machine CLASS1
Estimated Mean Relative Error, E = q tU 0.65 %
Notes:
1. The calibration results reported are valid at the time of and under the conditions of measurement and results relate only to the items calibrated
2. This report shall not be reproduced, except in full without approval of Hemshell Services LLP.
3. Any correction in this certificate, invalidates the certificate.
4. The classification of atesting machine calibrated and certifiedto meet aspecific class does not ensure that the acuracy inchudíng uncertainty will be less thana
specific value.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/Instrument/Equipment covered by this report due to damage caused to it afterissuance of this report
6. All the reference standards used are traceable to National Standards.
|7. Hemshell Services LLP uses Temperature compensated load cells so temperature correction is not required.
8. The machine shall in any case be Recalibrated, if it is moved to anew location necessitating dismantling or if it is subject to measure repairs or adjustm ents.
9.% Expanded Measurement Uncertainty at 95% confidence level with coverage facor

Checked By
VADODARA
Authorised Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar

IFF No. FF-HS-25 Amend. No.: 01


Technical Manager
Amend. Date: 25-11-2024 Page no. lof 1

Hemshell Services LLP


903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujorat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.com, www.hemshellservices.com
***End of the Report**
HEMSHELL
CC - 2920 TRANSCENDING BEYONDTESTING

CALIBRATION CERTIFICATE
Certificate No. |HS/BRD/CAL/24-25/196C ULR NO. CC292024000000453F Date of Calibration 19-12-2024
Date of Issue |Mechanical /UTM, Tension, Creep and
|21-01-2025
Discipline/ GroupTorsion Testing Machíne Recommended Due Date19-12-2025
Customer Name & Address Description & ldentification of Machine Calibration Conditions
Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make/ Model LLOYD/LS1 Work Instruction No. HS/CP/07
Machine Capacity
ITC Life Science & Tech centre, 1st Machine Serial No.
1KN
202336
Condition of Machíne Satisfactory
23°C
Environmental Temperature
maln, 1st phase, Peenya Industrial Load Cell Capacity 100N Calibrated at Site (Customer's Premises)
Area, Bangalore-560058 Load Cell Serial No. 100N1154 Calibration Mode Compression Mode
Customer Identification No. NA
Location of the Machine Packaging Lab |Calibrated By Mr. Arjun PN
Resoution of Machine 0.001
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard1 Reference Standard 2
Name ILOAD CELL WITH INDICATOR |LOAD CELL WITH INDICATOR
Make/Capacity/ Class |ADIARTECH/500N/1 ADI ARTECH/50N/1
Load Cell Sr. No. 1991218 1973886
Indicator Sr. No. HS/F/LCI/039/2023 10970
Calibration Certificate No./ CCNo. ULR-CC239023000008570F (CC- 2390 ULR-CC239023000008571F
Calibration Certificate Due On |13-09-2025 13-09-2025
+Reported Uncertainty in % 0.11 ).15

CALIBRATION RESULTS
Applied Load Standard Reading Observed Readings in Counts % Relative
in N in Counts Mean indication % Relative
±% Expanded
120 240 Mean Error'q' Repeatability Error
Uncertainty 'U"
Calibrated Range: 1N to 100N
1001 1003 1002 1002 1002 -0.13 0.10 0.62
5 5003 5008 5011 5012 5010 -0.15 0.08 0.62
10 10008 10013 10017 10019 10016 -0.08 0.06 0.62
20 20031 20040 20047 20052 20046 -0.08 0.06 0.62
50 50083 50088 50093 50097 50093 -0.02 0.02 0.62
50 10005 10011 10020 10022 10018 -0.13 0.11 0.62
60 12006
16006
12016
16019
12019
16021
12023
16024
12019
16021
olela-0.10
-0.11 0.06
0.03
0.62
0.62
100 20008 20023 20027 20029 20026 -0.09 0.03 0.62
Residual Load 0.001 0.003 0.002 N
Classification of Machine |CLASS 1
Estimated Mean Relative Error., E =atU 0.77 %
Notes:
1. The calibration results reported are valid at the time of and under the conditions of measuremnent and results relate only to the items calibrated.
2. This report shall not be reproduced, except in full without approval of Hemshell Services LLP.
3. Any correction in this certificate, invalidates the certificate.
4. The classification of a testing machine calibrated and certified to meet a specific class does not ensure that
the accuracy including uncertainty will be less than a
specific value.
S. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/Instrument/Equipment covered by this report due to damage caused to it after issuance of this report.
6. AIl the reference standards used are traceable to National Standards.
7.Hemshell ServDces LLP uses Temperature compensated load cells so temperature correction is not
required.
8. The machine shall in any case be Recalibrated, if it is moved to a new location necessitating
dismantling or if it is subject to measure repairsor adjustments.
9.% Expanded Measurement Uncertainty at 95% confidence level with
coveragefacior

KGES
VADODARA
Checked By
Mr. Vinay Mahadíik Authorised Signatory
Mr. Vinod Tulaskar
FF No. FF-HS-25 Amend. No.: 01 Amend. Date: 25-11-2024
Technical Manager
Page no. lof 1

Hemshell Services LLP


903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujarat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.com, www.hemshellservices.com
****End of the Renorte
NDIA

HEMSHELL
TRANSCENDING BEYOND TESTING
CC - 2920

CALIBRATION CERTIFICATE
Certificate No. HS/BRD/CAL/24-25/196D ULR NO. CC292024000000454F Date of Calibration 19-12-2024
Mechanlcal /UTM, Tension, Creep and
Date of Issue 21-01-2025 Discipline/ GroupTorsion Testing Machne Recommended Due Date 19-12-2025

Customer Namne & Address Description &ldentification ofMachine Calibration Conditions


Name UTM Referred IS standard IS: 1828 (Part 1):2022
Make /Model LLOYD/LS1 Work Instruction No. HS/CP/07
ITCLIfe Science &Tech centre, Machine Capacity 1KN Condition of Machíne
Satisfactory
1st main, 1st phase, Machine Serlal No. 202336 Environmental Temperature 23°C
Load Cell Capacity 100N Callbrated at Site (Customer's Premises)
Peenya Industrial Area,
Bangalore-560058
Load Cell Serial No. 100N1154 Calibration Mode Tension Mode
Customer Identification No. NA
Location of the Machine Packaging Lab Calibrated By Mr. Arjun PN
Resolution of Machine 0.001
DETAILS OF REFERENCE STANDARD USED FOR CALIBRATION
Description of Reference Standard Reference Standard 1 Reference Standard2
Name LOAD CELL WITH INDICATOR LOAD CELL WITHINDICATOR
Make/ Capacity/ Class ADI ARTECH/50ON /1 JADI ARTECH/5ON/1
Load Cell Sr. No. 1991218 1973886
Indicator Sr. No. HS/F/LCI/039/2023 10970
Calibration Certificate No./ CC No. ULR-CC239023000008570F CC-2390 ULRCCZ39023000008571F
Calibration Certificate Due On 13-09-2025 13-09-2025
t Reported Uncertainty in % 0.08 J0.13
CALIBRATION RESULTS

Applied Load Standard Reading Observed Readíngs in Counts % Relative


in N in Counts Mean Indication % Relattve
+% Expanded
120° 240 Mean Error 'q' Repeatabilíty Error
Uncertainty 'U
Calibated Range: 1N to 10ON
1002 1004 1002 1002 1003 -0.05 0.15 0.50
5 4998 5001 5003 5006 5003 -0.11 0.10 0.50
10 10006 10010 10015 10018 10014 -0.08 0.08
20 20030
0.50
20040 20046 20048 20045 -0.07 0.04 0.50
50 50076 50085 50090 50093 S0089 -0,03 0.0 0.50
9997 10003 10007 10009 10006 -0.09 0.06 0.50
60 12000 12014 12019 12020 12018 -0.15 0.05 0.50
80 16000 16016 16022 16024 16021 0.13 0.05 0.50
100 20001 20019 20021 20027 20022 -0.11 0.04 0.50
Residual Load 0.002 0.004 0.003 N
Classification of Machine CLASS 1
Estimated Mean Relative Error, E = q tU| 0.65 %
Notes:
1. The calíbration results reported are valid at the time of and under the conditions of measurement and results relate only to the ítems calibrated.
|2. This report shall not be reproduced, except in ful without approval of Hemshell Services LLP.
3. Any correction in thís certificate, invalídates the certificate.
4. The classification of a testing machine calíbrated and certified to meet a specific class does not ensure that the accuracy including uncertainty will be less than a
specific value.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of malfunctioning of standard
/Instrument/Equipment covered by this report due to damage caused to it after issuance of this report.
6. All the reference standards used are traceable to National Standards.
7. Hemshell Services LLP uses Temperature compensated load cells so temperature
correction is not required.
8. The machine shall in any case be Recalibrated, if it is moved to a new location
necessitating dismantling or if it is subject to measure repairs or adjustments.
9.% Expanded Measurement Uncertainty at 95% confidence level with coverage factor k2

Checked By
VADODARA J
Authorised Signatory
Mr. Vínay Mahadik Mr. Vinod Tulaskar

|FF No. FF-HS-25 Amend. No.: 01 Technical Manager


Amend. Date: 25-11-2024
Page no. lof 1

Hemshell Services LLP


903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujarat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.com, www.hemshellservices.com
***End of the Ren
HEMSHELL
TRANSCENDING BEYOND TESTING

CALIBRATION CERTIFICATE FOR EKTENSION


Certificate No. - HS/BRD/CAL/EXT/24-25/196E Page 1 of 1
Date of Issue - 21-01-2025
Customer Name & Address MACHINE DETAILS Calibration Conditions
Type UTM Calibrated On: |19-12-2024
Make: LLOYD Calibration due: (Recommended) 19-12-2025
ITC Life Science &Tech centre, Model: LS1 Mode: EXTENSION
1st main, 1st phase, Peenya 0.01 mm
Max Capacity: 1KN Resolution
Industrial Area, Bangalore
560058 SR.NO.: 202336 Room Temp.: 23°C
Condition 0f M/C: GOOD |Customer Id NA
Location: Packaging Lab Calibrated By: Mr. Arjun PN
DETAILS OF DIGIMATIC CALIPER USED FOR CALIBRATION

Make / Capacity Mitutoyo/ 200mm


Sr. No. / ID 04256342 / HS /F /DC / 012 / 2014
Calibration Certificate No. &Date V-240330-4-2 DT. 04-04-2024
Calibration Due On 04-04-2025
Calibrated By |Godrej &Boyce Mfg. Co. Ltd
|Calibrated at Temperature in °C |20+/- 3°C
Calibration Results

Standard
Set readings Observed Reading in Machine Console %Repeatability fo Accuracy
In mmn Reading In mm
Mean Reading Error Error
mm

Range Calibrated: 50mmn TO 500 mm


50 50.00 50.02 50.05 50.08 50.05 0.12 0.10
100 100.00 100.04 100.09 100.11 100.08 0.07 0.08
200 200.00 200.09 200.15 200.19 200.14 0.05 0.07
300 300.00 300.11 300.17 300.20 300.16 0.03 0.05
400 400.00 400.13 400.16 400.22 400.17 0.02 0.04
500 500.00 500.15 500.18 500.24 500.19 0.02 0.04

Note: According to standard ASTM-E-2309 acceptable %Accuracy &Repeatability error is +/-1% of reading
THE REPEATED SET OF READINGS ARE DIFFER IN PLACE TAKEN EARLIER.

R
oVADODARA
Checked By Approved Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar
Technical Manager
NOTE:
1. This report refers only to the particular item calibrated at customer site.
2. The calibration results reported are valid at the time of and under the conditions of measurement.
|3. This report may not be reproduced, except in full without prior permission of Hemshell Services LLP.
4. Any correction in this certificate, invalidates the certificate.
5. HemshellServices LLP is not liable for any change in calibration data and performance specifications on account of
malfunctioning of Standard/Instrument/Equipment covered bythis report due to damage caused to it after issuance of this report.
DOC NO.HS/03 ISSUE NO.05 DT.27.07.24

Hemshell Services LLP


903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujarat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.com, www.hemshellservices.com
HEMSHELL
TRANSCENDING BEYOND TESTING

CALIBRATION CERTIFICATE FOR SPEED


Certificate No. - HS/BRD/CAL/SPD/24-25/196F Page 1 of1
Date of Issue 21-01-2025
Customner Name & Address MACHINE DETAILS Calibration Conditions
UTM Calibrated On: 19-12-2024
|Type
Make: LLOYD Calibration due: (Recommended) 19-12-2025
ITC Life Science &Tech centre, 1st Model: Mode: |SPEED
main, 1st phase, Peenya Industrial Max Capacity: 1KN Resolution 0.01 mm /min.
Area, Bangalore-560058 SR.NO.: 202336 Room Temp.: |23°C
Condition Of M/C: GOOD Customner ld NA
|Location: Packaging Lab Calibrated By: Mr. Arjun PN
DETAILS OF DIGITAL STOP WATCH USED FOR CALIBRATION
Make/ Capacity RACER / 30 MIN.
Sr. No. /ID HS / F/ DSW /032/ 2023
Calibration Certificate No. & Date |V-240330-4-1 / DT: 05-04-2024
Calibration Due On 05-04-2025
Calibrated By |Godrej &Boyce Mfg. Co. Ltd.
Calibrated at Temperature in oC 24.8 °C
Calibration Results

Set Speed Time Taken


(mm /min.) %Repeatability Remarks
(mm/sec/ms) Mean Reading Error

|Range Calibrated: (50 TO 500) mm/min.


50 01:00:08 01:00:10 01:00:09 1:00:09 0.06 Satisfactory
01:00:10
100
200
01:00:12
01:00:14 01:00:15
01:00:13
01:00:16
1:00:12 0.08
Satisfactory
1:00:15 0.06 Satisfactory
01:00:15 01:00:17
300
400 01:00:18 01:00:20
01:00:19
01:00:21
1:00:17
1:00:20
0.11
0.08
Satisfactory
Satisfactory
01:00:20
500 01:00:20 01:00:18 1:00:19 0.06 Satisfactory
Error Allowed is +/-1%of Standard Readings
Asper Class 1.
HA
SERD
Checked By VADODARA J Approved Signatory
Mr. Vinay Mahadik Mr. Vinod Tulaskar
Technical Manager

NOTE:
1. This report refers only to the particular item calíbrated at customer site.
l2. The calíbration results reported are valid at the time of and under the conditions of measurement.
3.This report may not be reproduced,except in full without prior permission of Hemshell Services LLP.
4. Any correction in this certificate, invalidates the certificate.
5. Hemshell Services LLP is not liable for any change in calibration data and performance specifications on account of
malfunctioning of standard/Instrument/Equipment covered by this report due to damage caused to it after iSSuance of this report.
DOC NO.HS/04 ISSUE NO.05 DT.27.07.24|

HemshellServices LLP
903/1, Makarpura GIDC Estate, Makarpura,
Vadodara 390010. Gujarat, India.
Tel:+91 265 3100370/71
[email protected] www.hemetek.com, www.hemshellservices. com