Introduction to IR spectroscopy
Introduction to IR spectroscopy
Spectroscopy
Infrared (IR) Region
E = hcW
c = v
c = v/W
What Is An Infrared Spectrum?
• A plot of measured infrared light intensity vs a property
of light
• The y-axis can be plotted in absorbance/%transmittance
units
• The x-axis is plotted in wavenumbers
In Absorbance units (quantitative)
A = log(I0/I)
A = absorbance (amount of absorption)
I0 = Intensity in the background spectrum
I = Intensity in the sample spectrum
Advantages Disadvantages
3. Stray light causes spurious readings. 3. Stray radiations do not affect the
detector.
4. To improve resolution only a small 4. A much larger beam may be used at all
amount of IR beam is allowed to pass. time. Data collection is easier.
An interferogram:
• a plot of light intensity vs OPD
• a fundamental measurement obtained by FTIR
spectrophotometer
• A superposition of sine and cosine waves
• electrical signals coming out of detector
Interferometers and Interferograms
Af = final amplitude
A1 = amplitude of fixed mirror
A2 = amplitude of mobile mirror Aligns a beam of waves in a
specific direction
Phase of a wave
• A measure of how far through the cycle a
wave is at a particular point in space and time
OR
• Position of a point within a wave cycle
For constructive interferences: ẟ = nλ, n = 0, 1, 2….(an integer)
--> Af > A1 or A2
At ZPD,
ẟ=0
n=1
Phase difference = λ
Moving the mirror in a small
distance affects light intensity of
exiting beam
When ẟ ≠ 0; ẟ = 2∆
∆ = mirror displacement
n = 1/2
Phase difference ½ λ
• Interferogram of a single • Interferograms of λ light, 3λ
wavelength of light (laser) light and their sum
passing through the
interferometer
Frequency of light (IR source) vs frequency of electrical
signal (Fourier frequency in interferogram)
F = (2Vv)/c
F = Fourier frequency of interferogram in Hz
V = frequency of light wave in Hz
v = velocity of moving mirror in cm/s
c = speed of light in cm/s
F = 2vW
F = Fourier frequency of interferogram in Hz
v = velocity of moving mirror in cm/s
W = wavenumber in cm-1
• An IR spectrum consists of two important info: light
intensity and wavenumber.
• Fourier frequencies (F) in an interferogram tell us
what wavenumber of light are present in a given light
beam
• Amplitude of F ∝ amount of light reaches detector at
that wavelength
• When light of all wavenumbers from IR source passes
through interferometer, each wavelength exists an
interferogram with the corresponding F.
• Detector measures the sum of individual F.
All wavelengths of light are in-phase and interfere constructively
SNR ∝ N1/2
Noise level
N =1
Resolution ∝ 1/δ
δ = maximum optical path difference for a scan
Sample Suggested OPD Mirror
Typical Instrumental Setting for Different Sample Types
resolution (δ/cm) displacement
(cm-1) (Δ/cm)
• Parameters:
SNR,
Resolution (Res),
Analysis time (T),
Number of scans (N)
Trading rule 1: SNR ∝ N1/2
Trading rule 3: T ∝ N
2. Interefometers
• F = 2vW, hence mirror moves freely and at constant
velocity to ensure accurate F and OPD
• Friction is avoided by having air bearing or mechanical
bearing interferometers
3. Beamsplitters
• Thin film of Ge coated with two IR transparent
windows (KBr)
• Ge has the right thickness to transmit and reflect
some IR radiation
4. IR Detectors
• Detector element converts light intensity to electrical
signal (voltage)
• Type and material of the detector element determine
sensitivity and wavelength range, speed of response
and noise (N) level
• Area of detector element ∝ noise level (< 1mm2)
• Thermal vs Quantum Detectors
Thermal Detectors
based the conversion of a temperature change,
resulting from high-intensity radiation falling on a
suitable material, into a measurable signal (e.g:
deuterated triglysine sulfate, DTGS)
Advantages Disadvantages
Advantages Disadvantages
• Fast response • Responsitivity is
• high detectivity wavelength
dependent
• Requires cooling.
The Computer
• All measured signals are digitized and sent to the computer
where the Fourier transformation takes place.
• To obtain a relative scale for the absorption intensity, a
background spectrum must also be recorded. This is usually a
measurement w/o a sample in the beam.
• A single background measurement can be used for many sample
measurements because this spectrum is characteristic of the
instrument itself.
• This can be compared to the spectrum of a sample in the beam to
determine the “percent transmittance.”
• This technique results in a spectrum which has all of the
instrumental characteristics removed. All spectral features which
are present are strictly due to the sample.
Proper Use of Spectral Processing
Rules of Spectral Processing
1. Examine original spectrum to decide if it
needs processing. Do not process spectra
automatically
2. Have a clear idea of how processing
logarithm works, the problems it may cause,
and how to avoid them
3. Retain original data
4. Record how, when, and why processing is
performed
Spectral subtraction
• Used to simplify mixture spectra by removing unwanted
contributions such as water vapor, solvents, CO2, interferents,
impurities, and unknown compounds.
• Two spectra must be plotted with y-axis units that are linearly
proportional to concentration (Abs)
Asample – Aref = Result Absorbance
• Reference features in the spectra being subtracted must be perfect
overlays (have the same width, height, shape and wavenumber
position)
Examples of a spectral subtraction
Baseline Corrections
offset
Slope baseline
• Sample too thick
• Light scattering due to large
particles
Smoothing
• Savitsky-Golay method
• Boxcar smoothing
Deconvulation