TM-sl9100_flow_computer
TM-sl9100_flow_computer
SL9100
Multifunction High Performance
Flow Computer and Batcher
The information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the expressed purpose of assisting operating and maintenance personnel in the efficient use of the
instrument described herein. Publication of this information does not convey any rights to use or reproduce it or to use
for any purpose other than in connection with the installation, operation and maintenance of the equipment described
herein.
Copyright 1999
Printed in USA. All Rights Reserved.
WARNING!
This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling*
procedures must be observed during the removal, installation, or handling of internal circuit boards or devices.
*Handling Procedure
2. Personnel must be grounded, via wrist strap or other safe, suitable means, before any printed circuit board or other
internal device is installed, removed or adjusted.
3. Printed circuit boards must be transported in a conductive bag or other conductive container. Boards must not be
removed from protective enclosure until the immediate time of installation. Removed boards must be placed immediately
in protective container for transport, storage, or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern elec-
tronic designs contain components that utilize metal oxide technology (NMOS, CMOS, etc.). Experience has proven
that even small amounts of static electricity can damage or destroy these devices. Damaged components, even though
they appear to function properly, may exhibit early failure.
SAFETY INSTRUCTIONS
The following instructions must be observed.
• This instrument was designed and is checked in accordance with regulations in force EN 60950 (“Safety of
information technology equipment, including electrical business equipment”).
A hazardous situation may occur if this instrument is not used for its intended purpose or is used incorrectly.
Please note operating instructions provided in this manual.
• The instrument must be installed, operated and maintained by personnel who have been properly trained.
Personnel must read and understand this manual prior to installation and operation of the instrument.
• This instrument is internally fused. Replace the internal fuse with the following specified type and rating only:
Input Power Recommended Fuse
115 VAC 160 mA slow blow fuse
230 VAC 80 mA slow blow fuse
12-24 VDC 800 mA slow blow fuse
• The manufacturer assumes no liability for damage caused by incorrect use of the instrument or for modifications
or changes made to the instrument.
Technical Improvements
• The manufacturer reserves the right to modify technical data without prior notice.
SL9100 Flow Computer
CONTENTS
1. DESCRIPTION
1.1 Unit Description.......................................................................................................1
1.2 Unit Features...........................................................................................................1
1.3 Specifications..........................................................................................................2
2. INSTALLATION
2.1 General Mounting Hints..........................................................................................6
2.2 Mounting Diagrams.................................................................................................6
3. APPLICATIONS
3.1 Liquid Volume..........................................................................................................7
3.2 Corrected Liquid Volume.........................................................................................8
3.3 Liquid Mass.............................................................................................................9
3.4 Batching................................................................................................................10
4. WIRING
4.1 Typical Batcher Wiring..........................................................................................11
4.2 Typical Rate/Total Wiring.......................................................................................11
4.3 Wiring In Hazardous Areas...................................................................................12
5. UNIT OPERATION
5.1 Front Panel Operation Concept for Run Mode......................................................13
5.2 General Operation.................................................................................................14
5.3 Ratemeter/Totalizer Operation..............................................................................14
5.3.1 Password Protection for Rate/Total mode.............................................14
5.3.2 Relay Operation in Rate/Total mode......................................................14
5.3.3 Pulse Output in Rate/Total mode...........................................................14
5.3.4 Analog Output in Rate/Total mode.........................................................14
5.3.5 RS-232 Serial Port Operation in Rate/Total mode................................15
5.3.6 RS-485 Serial Port Operation in Rate/Total mode................................15
5.4 Batcher Operation.................................................................................................16
5.4.1 Batcher Configuration............................................................................16
5.4.2 Password Protection for Batcher mode.................................................17
5.4.3 Relay Operation in Batcher mode..........................................................17
5.4.4 Pulse Output in Batcher mode..............................................................17
5.4.5 Analog Output in Batcher mode.............................................................17
5.4.6 RS-232 Serial Port Operation in Batcher mode....................................18
5.4.7 RS-485 Serial Port Operation in Batcher mode....................................18
6. PROGRAMMING
6.1 Front Panel Operation Concept for Program Mode..............................................19
6.2 Setup Menus.........................................................................................................20
6.3 Setup Sub-Menus.................................................................................................21
6.3.1 INSTRUMENT TYPE.............................................................................21
6.3.2 SELECT FLOW EQUATION .................................................................22
6.3.3 SETUP INDICATORS (Total).................................................................22
6.3.4 SETUP INDICATORS (Density).............................................................22
6.3.5 SETUP INDICATORS (Rate).................................................................23
6.3.6 SETUP INDICATORS (Temperature)....................................................23
6.3.7 SETUP FLOW INPUT . .........................................................................24
6.3.8 SETUP AUX1 INPUT ............................................................................26
6.3.9 SET FLUID PROPERTIES ...................................................................27
6.3.10 SETUP PULSE OUTPUT....................................................................29
6.3.11 SETUP ANALOG OUTPUT..................................................................29
6.3.12 SETUP RELAYS .................................................................................30
6.3.13 SETUP CONTROL INPUTS(RATE/TOTAL)....................................... 32
6.3.14 SETUP CONTROL INPUTS(BATCH)..................................................32
6.3.15 SETUP REALTIME CLOCK(Time)......................................................33
6.3.16 SETUP REALTIME CLOCK(Date).......................................................33
6.3.17 SERIAL USAGE (RS-232/485)............................................................34
6.3.18 SERIAL USAGE (Modem Options)......................................................34
6.3.19 SETUP DATALOG/PRINT(Configure).................................................35
6.3.20 SETUP DATALOG/PRINT (Select_list)................................................36
6.3.21 ADMINISTRATIVE SETUP .................................................................36
6.3.22 SETUP NETWORK CARD .................................................................37
70
i
SL9100 Flow Computer
CONTENTS
7. PRINCIPLE OF OPERATION
7.1 General.................................................................................................................38
7.2 Flow Equations......................................................................................................38
7.3 Calculating the Expansion Factor.........................................................................41
7.4 Computation of Andrade Coefficient A and B........................................................42
7.5 Linearization Table................................................................................................43
7.5.1 Linearization Table General Information................................................43
7.5.2 Linearization Table for Pulse Inputs.......................................................43
7.5.3 Linearization Table Interpolation............................................................43
7.6 Universal Viscosity Curve (UVC)..........................................................................43
7.7 Strouhal Roshko Curve (StRo)..............................................................................43
8. TEST, SERVICE and MAINTENANCE
8.1 Test Menus............................................................................................................44
8.2 Test Sub-Menus....................................................................................................45
8.2.1 Audit Trail...............................................................................................45
8.2.2 Error History...........................................................................................45
8.2.3 Print System Setup................................................................................45
8.2.4 Keypad test............................................................................................46
8.2.5 Display test............................................................................................46
8.2.6 Calibrate Aux1 0mA...............................................................................47
8.2.7 Calibrate Aux1 20mA.............................................................................47
8.2.8 Calibrate Aux1 0V..................................................................................48
8.2.9 Calibrate Aux1 10V................................................................................48
8.2.10 Calibrate 100 ohm RTD.......................................................................48
8.2.11 Calibrate 0mA Out................................................................................49
8.2.12 Calibrate 20mA Out ............................................................................49
8.2.13 Analog In Test......................................................................................49
8.2.14 Pulse input test....................................................................................50
8.2.15 Analog out test.....................................................................................50
8.2.16 Excitation out test................................................................................50
8.2.17 Pulse out test.......................................................................................51
8.2.18 Relay test.............................................................................................51
8.2.19 Control input test..................................................................................51
8.2.20 Battery Voltage test..............................................................................52
8.2.21 Data logger utility ................................................................................52
8.3 Internal Fuse Replacement...................................................................................53
9. RS-232 SERIAL PORT
9.1 RS-232 Serial Port Description.............................................................................54
9.2 Instrument Setup by PC Over Serial Port ............................................................54
9.3 Operation of Serial Communication Port with Printers..........................................54
9.4 SL9100 RS-232 Port Pinout..................................................................................54
10. RS-485 SERIAL PORT
10.1 RS-485 Serial Port Description...........................................................................55
10.2 General ..............................................................................................................55
10.3 Operation of Serial Communication Port with PC...............................................55
10.4 SL9100 RS-485 Port Pinout................................................................................55
11. FLOW COMPUTER SETUP SOFTWARE
11.1 System Requirements.........................................................................................56
11.2 Cable and Wiring Requirements.........................................................................56
11.3 Installation ..........................................................................................................56
11.4 Using the Flow Computer Setup Software..........................................................57
11.5 File Tab................................................................................................................57
11.6 Setup Tab............................................................................................................57
11.7 View Tab..............................................................................................................58
11.8 Misc. Tab.............................................................................................................58
12. GLOSSARY OF TERMS
12 Glossary Of Terms.................................................................................................59
13. Diagnosis and Troubleshooting
13.1 Response of SL9100 on Error or Alarm:.............................................................63
13.2 Diagnosis Flow Chart and Troubleshooting........................................................64
13.3 Error & Warning Messages:................................................................................65
13.3.1 Sensor/Process Alarms.......................................................................65
13.3.2 Self Test Alarms...................................................................................66
71
ii
SL9100 Flow Computer
The isolated analog output can be chosen to follow volume flow, corrected
volume flow, mass flow, temperature or density by means of a menu
selection. Most hardware features are assignable by this method.
The user can assign the standard RS-232 Serial Port for data recording,
transaction printing, or for connection to a computer or modem.
A Service or Test mode is provided to assist the user during start-up system
check out by monitoring inputs and exercising outputs and printing system
setup.
Unit Features
1.2 Unit Features:
1
SL9100 Flow Computer
1.3 Specifications:
Specifications: Auxilliary/Compensation Inputs
Environmental The compensation inputs are menu selectable for
Indoor Use temperature, density or not used.
Altitude up to 2000m
Operating Temperature: 0°C to +50°C Operation: Ratiometric
(-20°C to 55°C optional) Accuracy: 0.01% FS
Storage Temperature: -40°C to +85 C Thermal Drift: Less than 100 ppm/C
Maximum Relative Humidity : 80% for Basic Measurement Resolution: 16 bit
temperatures Update Rate: 1 update/sec minimum
up to 31°C decreasing linearly to 50% RH at Automatic Fault detection:
40°C Signal Over-range/under-range
Mains supply voltage fluctuations not to exceed Current Loop Broken
±10% of the nominal voltage RTD short
Tr a n s i e n t o v e r v o l t a g e a c c o r d i n g t o RTD open
INSTALLATION Fault mode to user defined default settings
CATEGORY II (see UL 3101-1 Annex J) Transient Protection: 500 V (Capacitive Clamp)
POLLUTION DEGREE 2 in accordance with Reverse Polarity: No ill effects
IEC 664 (see 3.7.3) Over-Voltage Limit (Voltage Input): 50 VDC
Materials: UL, CSA, VDE approved
Available Input Ranges
Listing: CE Light Industrial, Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
UL File #: E192404 Current: 4-20 mA, 0-20 mA
Resistance: 100 Ohms DIN RTD or
Display 100 Ohm DIN RTD (DIN 42-760, BS 1904):
Type: 2 lines of 20 characters, LCD Three Wire Lead Compensation
Character Size: 0.3" nominal Internal RTD linearization learns ice point
User selectable label descriptors and units of resistance
measure 1 mA Excitation current with reverse polarity
protection
Keypad Temperature Resolution: 0.01 C
Keypad Type: Membrane Keypad
Keypad Rating: Sealed to Nema 4 Control Inputs
Number of keys: 16 Switch Inputs are menu selectable for Start, Stop,
Reset, Reset/Start, Stop/Reset, Lock, Inhibit, Alarm
Enclosure Acknowledge, Print or Not Used.
Size: See Dimensions Control Input Specifications
Depth behind panel: 6.5" including mating Input Scan Rate: 10 scans per second
connector Logic 1: 4 - 30 VDC
Type: DIN Logic 0: 0 - 0.8 VDC
Materials: Plastic, UL94V-0, Flame retardant Transient Suppression: 500 V fast transient
Bezel: Textured per matt finish (Capacitive Clamp)
Equipment Labels: Model, safety, and user wiring Input Impedance: 100 KΩ
Control Activation: Positive Edge or Pos. Level
Power Input based on product definition
The factory equipped power option is internally Excitation Voltage
fused. An internal line to line filter capacitor is 110/220 VAC Powered Units
provided for added transient suppression. Menu Selectable: 5, 12 or 24 VDC @ 100mA
Order Option 1: 110VAC: 85 to 127 Vrms, 50/60 24 VDC Powered Units
Hz Menu Selectable: 5 or 12 VDC @ 100mA
Order Option 2: 220VAC: 170 to 276 Vrms, 50/60 12 VDC Powered Units
Hz 5 VDC @ 100mA
Order Option 3: 12VDC: 10.5 to 14 VDC
Order Option 4: 24VDC: 18 to 24 VDC Fuse
Internal self resetting fuse supplied
Flow Inputs:
Pulse Inputs: Relay Outputs
Number of Flow Inputs: one The relay outputs are menu assignable to
Configurations supported: single input or (Individually for each relay) Low Rate Alarm, Hi Rate
quadrature (menu selectable) Alarm, Prewarn Alarm, Preset Alarm, Temperature,
Input Impedance: 10 KΩ nominal Density, or General purpose warning (security).
Pullup Resistance: 10 KΩ to 5 VDC (menu
selectable) Number of relays: 2 (4 optional)
Pull Down Resistance: 10 KΩ to common Contact Style: Form C contacts
Trigger Level: (menu selectable) Contact Ratings: 250 VAC @ 5 amps
High Level Input 30 VDC @ 5 amps
Logic On: 3 to 30 VDC Fast Transient Threshold: 1000 V
Logic Off: 0 to 1 VDC
Low Level Input (mag pickup)
Selectable sensitivity: 10 mV & 100 mV
Minimum Count Speed: Selectable: 1 - 99 secs
Maximum Count Speed: Selectable: 40, 3k,
20kHz
Overvoltage Protection: 50 VDC
Fast Transient: Protected to 500 VDC
(Capacitive Clamp)
2
SL9100 Flow Computer
3
SL9100 Flow Computer
The system also provides a minimum implementation A supervisor password is required to gain access to
of an “audit trail” which tracks significant setup this specialized mode of operation. Normally quality,
changes to the unit. This feature is increasingly calibration, and maintenance personnel will find this
being found of benefit to users or simply required mode of operation very useful. It is also useful for
by Weights and Measurement Officials in systems factory testing.
used in commerce, trade, or “custody transfer”
applications. Many of these tests may be used during start-up
of a new system. Inputs signals may be read,
A software program is available which runs on a and output signals may be exercised to verify the
PC using a RS-232 serial port for connection to electrical interconnects before the entire system is
the Flow Computer. Illustrative examples may be put on line.
downloaded in this manner.
The following action items may be performed in the
The setup mode has numerous subgrouping of Maintenance Mode:
parameters needed for flow calculations. There is
a well conceived hierarchy to the setup parameter Print Calibration/Maintenance Report
list. Selections made at the beginning of the setup Examine Audit Trail
affect offerings further down in the lists. Perform Keypad Checkout
Perform Display Checkout
In the setup mode, the flow computer activates the Perform Pulse Input Checkout
correct setup variables based on the instrument Perform Pulse Output Checkout
configuration, the flow equation, and the hardware Perform Control Input Checkout
selections made for the compensation transmitter Perform Relay Output Checkout
type, the flow transmitter type, and meter Perform Analog Input Checkout
enhancements (linearization) options selected. All Perform Analog Output Checkout
required setup parameters are enabled. All setup Calibrate Analog Inputs using the Learn Feature
parameters not required are suppressed. Calibrate Analog Output using the Learn Feature
Battery Check
A help line prompt is provided for each entry. In Datalog Printing and Clearing
addition a help message is available which may be
accessed by depressing the “HELP” key. Note that a calibration of the analog input/output will
advance the audit trail counters since it effects the
Also note that in the setup mode are parameter accuracy of the system.
selections which have preassigned industry
standard values. The unit will assume these values RS-232 Serial Port
unless they are modified by the user. The SL9100 has a general purpose RS-232 serial
port which may be used for any one of the following
Some of the process input variables have available purposes:
a “default” or emergency value which must be
entered. These are the values that the unit assumes Transaction Printing
when a malfunction is determined to have occurred Periodic Printing of Datalog
on the corresponding input. Print Internal Datalog
Remote Metering by Modem (optional)
It is possible to enter in a nominal constant value Computer Communication Link
for temperature or density by placing the desired Configuration by Computer
nominal value into both the lo and hi values. This Print System Setup
is also a convenience when performing bench top Print Calibration/Malfunction History
tests without simulators. Remote Control
4
SL9100 Flow Computer
Operation of Serial Communication Port with Operation of Serial Port with Modems (optional)
Printers The SL9100 RS-232 channel supports a number of
SL9100’s RS-232 channel supports a number operating modes. One of these modes is intended
of operating modes. One of these modes is to support operation with a modem in remote
intended to support operation with a printer in metering applications.
metering applications requiring transaction printing,
data logging and/or printing of calibration and An external modem is intentionally being used
maintenance reports. with the SL9100. This permits use with the variety
of modem standards worldwide while avoiding
For transaction printing, the user defines the items the specialized approvals required for equipment
to be included in the printed document. The user that is deemed to fall under the category of
can also select what initiates the transaction print telecommunication equipment.
generated as part of the setup of the instrument.
The transaction document may be initiated via In the modem mode, the SL9100 is assumed to be
a front panel key depression, a remote contact operating in a remote metering role. The SL9100
closure, or upon completion of a batch. will support key items in the Hayes Compatible “AT”
Command Set. In this role, the SL9100 will have the
In data logging, the user defines the items to be following special abilities:
included in each data log as a print list. The user 0. Monitor the modem status as a task of the
can also select when or how often he wishes a data system
log to be made. This is done during the setup of 1. Instruct the modem to answer an incoming call
the instrument as either a time of day or as a time ATA
interval between logging. 2. Respond to the calling modem at the programmed
baud rate and protocol
The system setup and maintenance report lists all 3. Terminate the telephone connection in event the
the instrument setup parameters and usage for connection is lost.
the current instrument configuration. In addition,
the Audit trail information is presented along with In addition, the SL9100 will be capable of initiating a
a status report listing any observed malfunctions call to a designated telephone number in the event
which have not been corrected. of a metering malfunction.
5
SL9100 Flow Computer
2. Installation
2. Installation
Termination Connectors:
Minimum Wire Gauge: 22 AWG
Maximum Wire Gauge: 14 AWG
Voltage/current limits are limited by unit specifications.
Ensure that the switch or circuit breaker chosen is suitable for the
power requirements of the unit.
SL9100
SL9100
Bezel Adaptor
Gasket
Mounting Bracket
Mounting Bracket
Dimensions
6.15
5.67 (144) (156) 5.43
0.28 (7.2) 0.5 (138)
SL9100
(13)
0.4 (10)
6.18
Dotted Line Shows Optional Bezel Kit
6
SL9100 Flow Computer
3. Applications
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• Volume flow is calculated using the flowmeter frequency output and the
K-Factor determined by the desired method.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total
• Analog Output
Rate or Total
• Pulse Output
Total
• Relay Outputs
Rate or Total Alarms
Applications:
The Flow Computer can monitor actual volume flow and total of any liquid.
Flow alarms are provided via relays and datalogging is available via analog
(4-20mA) and serial outputs.
Liquid Volume
Illustration SL9100
STOP
STOP GRAND
GRAND SCROLL
SCROLL PRE
PRE 2 2 PROC F2 METER F3 HELP HELP
ENTER ENTER
66 77 8 8 9 9 0 0– –• •
Flowmeter
7
SL9100 Flow Computer
Calculations:
• Corrected Volume at a base or reference condition is calculated using the
flow and temperature inputs as well as the thermal expansion coefficient
stored in the flow computer. Use the "SET FLUID PROPERTIES"
submenu to define reference temperature and density values for
standard conditions.
Output Results:
• Display Results
Corrected Flow Rate, Resettable Total, Non-Resettable Total,
Temperature, Density
• Analog Output
Corrected Rate, Total, Temperature or Density
• Pulse Output
Corrected Total
• Relay Outputs
Corrected Rate , Total or Temperature Alarms
Applications:
Monitoring corrected volume flow and total of any liquid. Flow alarms are
provided via relays and datalogging is available via analog (4-20mA) and
serial outputs.
Corrected
Liquid Volume
Illustration SL9100
8
SL9100 Flow Computer
Measurements:
Actual volume is measured by the flow element. Temperature is measured
by the temperature transmitter. A density transmitter can be used for direct
density measurements.
Calculations:
• The density and mass flow are measured directly or calculated using
the reference density and the thermal expansion coefficient of the liquid
(see "SET FLUID PROPERTIES" submenu)
Output Results:
• Display Results
Mass Flow Rate, Resettable Total, Non-Resettable Total,
Temperature or Density
• Analog Output
Mass Rate, Volume Rate, Mass Total, Temperature or Density
• Pulse Output
Mass Total or Volume Total
• Relay Outputs
Mass Flow Rate, Total, Temperature or Density Alarms
Applications:
Monitoring mass flow and total of any liquid. Flow alarms are provided
via relays and datalogging is available via analog (4-20mA) and serial
outputs.
Liquid Mass
Illustration
SL9100
STOP GRANDGRAND
SCROLL
SCROLL PRE PRE
2 2 DENS
PROC
TIME
METER HELP
HELP
STOP ENTER
6 7 8 8 9 9 00 •• ENTER
6 7 ––
Flowmeter Temp./Dens.
Transmitter
Mass Flow
9
SL9100 Flow Computer
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
Volume or Corrected Volume or Mass Calculations).
Calculations:
• Volume flow is calculated using the flowmeter frequency output and the
selected K-Factor method.
• Mass or Corrected Volume is calculated using the flow and temperature
inputs as well as the thermal expansion coefficient stored in the flow
computer.
• Comparison of the amount delivered with desired batch amount.
Output Results:
• Display Results
Flow Rate, Batch Total, Non-Resettable Total, Temperature or
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total (volume or corrected volume/mass)
• Relay Outputs
Batch Total, Rate, or Temperature Alarms
Applications:
Batching and monitoring flow and total of any liquid. Batching is
accomplished via relays, datalogging is available, analog (4-20mA) and
serial outputs.
Batching Illustration
SL9100
TOTAL
TOTAL RATE
RATE PRE
PRE11 F1
FLUID PRINT
PRINT
CLEAR
CLEAR MENU
MENU
START
START 1 1 2 2 33 44 55
GRAND SCROLL
STOPSTOP GRAND SCROLL PRE
PRE 22 PROC
F2 METER
F3 HELP
HELP
ENTER
ENTER
6 6 7 7 88 99 0
0 – ••
–
10
SL9100 Flow Computer
4 WIRING
{
- 17 NC 25 NC
Alarm Relay 1 18 COM RLY1 26 COM RLY3
19 NO 27 NO
{
20 NC 28 NC
Alarm Relay 2 21 COM RLY2 29 COM RLY4
22 NO 30 NO
11
SL9100 Flow Computer
4 Common
1 2
4 3
Transmitter T Diode
+
12
SL9100 Flow Computer
5. UNIT OPERATION
13
SL9100 Flow Computer
14
SL9100 Flow Computer
RS-232 Serial Port 5.3.5 RS-232 Serial Port Operation in Rate/Total mode
(Rate/Total mode)
The RS-232 serial port can be used for programming (using the Setup Disk) or
for communicating to printers, modems and computers in the Operating Mode
(Run Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Data Logging
In data logging, the user defines the items to be included in each data log (see
section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also select
when (time of day) or how often (print interval) the data log is to be made (see
section 6.3.19 SET PRINTER OUTPUT, Configure). Data logs can also be
initiated using the print key or control input when assigned to print function.
15
SL9100 Flow Computer
The Batcher mode is used primarily to control filling operations. The main
difference between the Batch mode and Rate/Total mode is the relay operation.
The Batch mode allows the operator to "START" the unit via the front panel or
remote input. Once started, the relays (RLY1 & RLY2) will energize and send a
contact to a flow control device (i.e. solenoid valve or pump). The flow sensor
will send a signal to the unit and total accumulation will begin. When the Prewarn
value (PRE 2) is reached, Relay 2 will drop out (this is ideal for flow slow down).
When the Batch amount (PRE 1) is reached, Relay 1 will drop out and the Batch
is complete.
Several messages will be displayed during normal batch operation (i.e. Batch
Fill, Batch Stopped). The keypad is disabled for the duration of these timed
messages (approx. 2 sec).
When the unit is programmed for batch mode, several batch operation choices
are available. These choices include: Up or Down Counting, Maximum Batch
Preset, Setpoint Checking, Batch Overrun Compensation, Auto Batch Restart,
Time Delay, Flow Signal Timeout, Maximum Drain Time, Slow Start Quantity, No
Total for Batch Delay, Start or Reset/Start, and Stop or Stop/Reset.
Batch Overrun
The Batch Overrun is used for batch applications that have slow responding
valves and a consistent batching flowrate. When the Batch Overrun is set, the
unit will compensate for batch overruns by computing an averaged overrun
value from the last four batches. This average is used to internally adjust the
batch setpoint to minimize overrun. A non zero maximum drain time must also
be specified.
Time Delay
The Time Delay for Auto Batch Restart functions as follows: When a batch
is completed, the next batch will automatically start after the amount of time
entered here.
16
SL9100 Flow Computer
17
SL9100 Flow Computer
RS-232 Serial Port 5.4.6 RS-232 Serial Port Operation in Batcher mode
(Batch mode)
The RS-232 serial port can be used for programming (using the Setup Disk) or
for communicating to printers, modems and computers in the Operating Mode
(Run Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Data Logging
In data logging, the user defines the items to be included in each data log
(see section 6.4.22 SET DATALOG/Print, Select_list). The user can also
select if unit will datalog at end of batch or when (time of day) or how often
(print interval) the data log is to be made (see section see section 6.4.21 SET
DATALOG/Print, Configure).
18
SL9100 Flow Computer
6. PROGRAMMING
The SL9100 is fully programmable through the front panel. Please review
the following usage summary before attempting to use the instrument.
SL9100
Setup Mode:
19
SL9100 Flow Computer
6.2.2
Submenu SELECT FLOW EQUATION Refer to Pages 22 for Details.
Groups
Sub-menus Notes
6.3.1 INSTRUMENT TYPE Press ENTER to enter Instrument Type sub-
INSTRUMENT TYPE menus.
Batch
INSTRUMENT TYPE Press ENTER when Batch is flashing to
Rate/Tot Batch configure the instrument as a Batcher.
Select UP to Reset to 0 and count up to preset.
BATCH COUNT MODE Select DOWN to reset to Preset and count
Up Down down to 0.
Enter the maximum allowable Batch Preset.
MAXIMUM BATCH PRESET The operator will not be able to enter a batch
1000.0 gal preset larger than this value.
Select ON to set the unit to operate using a
BATCH OVERRUN COMP Batch Overrun Compensation routine.
Off On Select OFF to inhibit Batch Overrun
Compensation routine. (See Section 5.4)
AUTO BATCH RESTART Select On to enable the Auto Batch Restart. This
Off On will automatically restart the unit at the end of
each batch run. Select Off if this is not desirable.
TIME DELAY (1-99sec) Enter Time Delay for Auto Batch Restart if selected.
10 When a batch is completed, the next batch will start
after the amount of time entered here.
NOTE:
Non zero "Max Drain
MAX DRAIN TIME (sec) Enter time (0-99 sec.) for Max. Drain Time. After batch
Time" value required 10 quantity is reached, “Batch Done” is declared when
the flow rate is “0” or the Maximum Drain Time has
for Batch Overrun expired.
Comp.
SLOW START QUANTITY
0 gal Enter a quantity for a Slow Start up. RLY 1 (slow
fill) will energize for Slow Start and RLY 2 (fast
fill) will energize after the Slow Start Quantity has
been delivered.
NO TOT FOR BATCH DLY
No Yes Select "Yes" to inhibit totalization during the auto
batch restart delay.
Advance To
SELECT FLOW EQUATION
21
SL9100 Flow Computer
6.3.2 SELECT FLOW EQUATION Press ENTER to enter Select Flow Equation
SELECT submenus.
FLOW EQUATION
Advance To
SETUP INDICATORS (To-
tal)
6.3.3 Press ENTER to begin setup of the Indicators
SETUP
SETUP INDICATORS
INDICATORS
(Total)
TOT DEC PLACES (0-3) Select the desired Total Decimal Place.
0 0-3 decimal places allowed.
Advance To
SETUP INDICATORS
(Density)
6.3.4 Press ENTER when Dens is flashing to
SETUP
SETUP INDICATORS
Total Dens Rate Temp configure the Density Indicators.
INDICATORS
(Density)
Advance To
SETUP INDICATORS
(Rate)
22
SL9100 Flow Computer
Advance To
SETUP FLOW INPUT
23
SL9100 Flow Computer
6.3.7 SETUP FLOW INPUT Press ENTER to begin setup of Flow Input.
SETUP
FLOW INPUT
NOTE: PULSE INPUT TYPE Enter the desired Pulse type. See side note.
ChA = Single Pulse ChA A=B Qx1 Up/Down
A=B = Pulse Security
Qx1 = Quadrature
Up/Down = Up/Down
PULSE TRIGGER LEVEL Select the desired Input Pulse Trigger Level.
Control
Line
10mV 100mV 2.5V
LINEAR TABLE KB
KB--01:####### P/gal
LOW FLOW RATE ALARM Enter the desired volumetric Low Rate Alarm.
####### gal/s This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
HIGH FLOW RATE ALARM Enter the desired volumetric High Rate Alarm.
####### gal/s This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
Meter Expan [xe - 6] Enter the expansion coefficient for the meter
0 ppm/f housing.
Advance To
SETUP AUX1 INPUT
25
SL9100 Flow Computer
6.3.8 SETUP AUX1 INPUT Press ENTER to begin setup of the Auxiliary 1
SETUP AUX1 INPUT Input.
AUX1 INPUT TYPE Select desired use for the Auxiliary 1 Input
None Density Temp (compensation input).
AUX LOW ALARM Enter the Low setpoint for the Temperature
###### F Alarm.
AUX HIGH ALARM Enter the High setpoint for the Temperature
###### F Alarm.
Advance To
SET fluid properties
26
SL9100 Flow Computer
6.3.9 SET FLUID PROPERTIES Press ENTER at this prompt to Set Fluid
SET FLUID Properties.
PROPERTIES
NOTE: FLUID NUMBER (0-9) Up to 10 Fluid types may be stored in the unit.
The prepro- 0 Select the number of the desired fluid to edit.
grammed fluid
temperature versus
viscosity and den- FLUID NAME Shows name and number of fluid selected.
sity values are typi- Generic #0 Enter the desired name using the up/down
cal. For most ac- arrow keys.
curate calculations,
the temperature COMPUTE FLUID PROPS? Select "Yes" to allow the unit to compute the
versus viscosity No Yes fluid properties. Some of the following menus
and density tables will only appear if YES is selected.
should be edited to
just cover the tem- viscosity method Select the desired viscosity method.
perature range an-
ticipated and with
Manual Andrade ASTM Some of the following Menu Items only appear
if Andrade or ASTM selected.
the actual values
for the fluid in use.
Temperature and
TEMPERATURE point V1 Enter the Temperature 1 value corresponding
density values can
###### F to Viscosity 1.
change from brand
to brand and lot to Enter the Viscosity 1 value corresponding to
lot of products.
viscosity point V1
###### CSTKS Temperature1.
Advance to
Setup pulse output
27
SL9100 Flow Computer
6.3.9.2
Density method Select the desired density method.
Density Method Some of the following Menu Items only appear
LINEAR Case Manual Linear API
if Linear selected.
Advance To
SETUP Pulse OUTPUT
Advance To
SETUP Pulse OUTPUT
28
SL9100 Flow Computer
6.3.10 SETUP PULSE OUTPUT Press ENTER at this prompt to setup the
SETUP Pulse Output.
PULSE OUTPUT
PULSE WIDTH Select the desired Pulse Width for the Pulse
10mS 100mS Output.
PULSE VALUE Enter the desired Pulse Value for the Pulse
####### gal/P Output (Units per Pulse).
Advance To
SETUP ANALOG OUTPUT
ANALOG OUT FLOW TYPE Only if Rate selected & Flow EQ. = Mass, Cor/Vol
Vol CVol/Mass Select the desired Analog Output Flow.
ANALOG OUTPUT RANGE Select the desired current range for the Analog
4-20mA 0-20mA Output.
FS ANALOG OUT 20mA Enter desired Analog Output Full Scale Value.
####### gal/m
Advance To
SETUP RELAYS
29
SL9100 Flow Computer
NOTE:
RELAY 1 USAGE If Relay 1 or Relay 2 Selected,
In Batch mode,
Relay 1 is reserved RATE TOTAL NA Select Rate, Total or NA.
for Preset,
Relay 2 is reserved
for Prewarn.
RELAY 1 DELAY sec If Rate selected, enter desired relay activation
0 delay value.
Advance To
SETUP RELAYS 3, 4
30
SL9100 Flow Computer
Advance To
SETUP CONTROL INPUTS
1. Relay activation is based on the instantaneously computed values not the averaged displayed
value. Therefore the display damping factor will not affect the relay response time. The RELAY
DELAY feature allows the user to enter a time delay for relay activation. This feature is very useful
in applications where short over/under range conditions are not considered alarm conditions.
2. When INSTRUMENT TYPE is set to batcher, Relay 1 is reserved for PRESET and Relay 2 is re-
served for PREWARN.
3. Setting the relays to NA (Not Assigned), will allow the relay activation to be controlled via the RS-
232 Serial and/or RS-485 Modbus ports.
4. Relay 3 and Relay 4 settings may be used to trigger display alarm conditions even if the relays are
not supplied.
31
SL9100 Flow Computer
6.3.13 SETUP CONTROL INPUTS Press Enter to begin setup of the Control
SETUP CONTROL Inputs.
INPUTS
(if RATE/TOTAL)
SETUP CONTROL INPUTS Select the desired Control Input for setup.
Input1 Input2 Input3
32
SL9100 Flow Computer
Advance To
SETUP REALTIME CLOCK
(Date)
DATE: MONTH,DAY,YEAR Enter the date. (Month, Day, Last two digits of
##/##/#### Year)
Advance To
SERIAL USAGE
33
SL9100 Flow Computer
##
Select the desired Baud Rate.
BAUD RATE
300 600 1200 <more>
(If <more> selected)
BAUD RATE
2400 4800 9600 19200
Select the desired Parity.
PARITY
None Odd Even
Set the Handshake.
HANDSHAKING
None Softwre Hardwre
Choose end of line termination. Only choose
<CR> if your external device automatically
DEVICE LINE FEED assigns a line feed for every <CR> carriage
<CR> <CR+LF> return.
1
Enter the time of day to perform Call Out
transmission. (HH:MM:SS)
CALL OUT TIME
##:##:##
34
SL9100 Flow Computer
CALL OUT PHONE # Call Out Phone Number to be dialed for "Call
0 Out Time" or "Print On Error/Alarm". (Up to 20
digits with "." used to pause between digits)
Advance To
SETUP DATALOG/PRINT
PRINT TIME HH:MM:SS Enter Print Time, printer will print at this time
00:00:00 every day. Enter 00:00:00 to inhibit print time.
Advance To
SETUP DATALOG/PRINT
(Select_list)
35
SL9100 Flow Computer
PRINT LIST ITEMS Use Up and Down arrow keys to view list
List Items: FLUID YES status.
FLUID Press the Print or function key for the items
TIME
PRINT LIST ITEMS that you wish to add or remove from the list.
RATE
TOTAL TIME YES Items marked with Yes will be added to the
GRAND list,
TEMP
PRINT LIST ITEMS items marked with No will be removed from
DENS
PRE1 RATE YES the list.
PRE2
VISC The Select Print List Information display
FREQ1
FREQ2 PRINT LIST ITEMS shows the current possible Datalog size.
KA-F DataLog size =001489
KB-F
Advance To
ADMINISTRATIVE SETUP
TAG NUMBER Use the up and down arrow keys to define the
FT XXXX tag number.
PRODUCT ORDER CODE This display is used to show the product order
SL91xxxxxxxxx code (model number).
UNIT SERIAL NUMBER This display is used to show the unit's serial
00000 number.
DISPLAY NEW ERR ONLY If yes is selected, an error message will only
No Yes appear once until acknoledged by user.
Advance To
INSTRUMENT TYPE
37
SL9100 Flow Computer
7. Principle Of Operation
General Case
Tf = [% input span • (temp FS - Temp low scale)] + temp low scale
RTD Case
Tf = f(measured input resistance)
Density Transmitter
density = [% input span • (density FS - density low scale)] + density low scale
Fluid Properties:
Where reference density and therm. exp. coef. are determined from two known
temperature/density points and desired reference temperature.
38
SL9100 Flow Computer
centistokes = (A exp B
(Deg F + 459.67))
Where parameter A and parameter B are derived from user entered Temperature/
Viscosity Cstk pairs.
ASTM Case
Please refer to the following documentation related to ASTM viscosity
calculations.
ASTM Designation: D 341 - 93 Standard
Viscosity-Temperature Charts for Liquid Petroleum Products
Temperature Transmitter
Standard Volume Flow = volume flow • (1 - Therm.Exp.Coef. • (Tf-Tref))
Density Transmitter
density
Standard Volume Flow = volume flow •
reference density
39
SL9100 Flow Computer
Flow Equations
The above information was obtained from "Flow Measurement Engineering Handbook, 3rd
Edition" by Richard W Miller.
40
SL9100 Flow Computer
Calculating Fluid The liquid density is a function of the flowing temperature for many fluids. This unit
Expansion Factor solves an equation which represents this physical property of the fluid based on two
known temperature/density pairs.
From this known information. the unit computes the following variables: Reference
Temperature, Reference Density, Fluid Expansion Factor.
The required fluid temperature/density pair information is available for many fluids
in one or more of the following forms:
Fluid Specific Gravity vs. Temp. Table
Specific Gravity vs. Temp. Graph
Fluid Density vs. Temp. Table
Fluid Density vs. Temp. Graph
Begin by obtaining one of the fluid properties for the fluid you are using from available
manufacturers information or Engineering Handbooks. In some cases this information
is listed on the Material Safety Data Sheet for the fluid.
Two temperature-density pairs will be required to compute the temperature
coefficient.
The reference temperature is simply chosen by the user. Common reference
temperatures are 60° F or 15° C.
However, for cryogenic fluids, the normal boiling point may also be used. In some
cases the fluid data may list properties at 100° F. This temperature may also be
used as the reference temperature.
The reference temperature should be chosen so that it is in the application
temperature range. (i.e. The application temperature range -10 to 120° F, reference
temperature of 60° F chosen.)
Enter the reference temperature you have chosen at this point.
The reference density corresponds to the reference temperature computed by the
unit.
Expansion Factor
Equations EQ1.
Spec.Grav. = Density of Fluid / Density of Water
NOTE:
1. Reference Appendix Given the reference temperature, reference specific gravity, a second temp. and a
A for several common
liquids.
second Spec.Grav., the Expansion Factor (C Factor) can alternatively be computed
as follows:
2. The Compute Fluid
Properties utility
[ ] x 1,000,000
will perform these
EQ2. Used for Liquid Mass and Corrected Volume Equations
C = 1 - (Spec.Grav.2 / Ref.Spec.Grav.)
calculations for you.
Temp.2 - Ref.Temp
[ ] x 1,000,000
EQ3. Used for Liquid Mass and Corrected Volume Equations
1- (Dens.2 / Dens.1)
C= Temp.2 - Temp.1
EQ4.
Ref.Density = (1 - c x 1-6 (T1 - Tref)) x D1
41
SL9100 Flow Computer
Computation of The flow computer solves an equation which computes the viscosity in
Andrade Cstk as a function of temperature. Two sets of known temperature/Cstk
Coef. A & B data points are required for the calculation to be performed. These are the
setup parameters in the Fluid Group.
NOTE: The Viscosity Coef. A and B are derived from two known temperature/
1. Reference Appendix
A for several common viscosity pairs. Convert these known points to units of Degrees F and
liquids. centistokes (Cstk)
2. The Compute Fluid The information is now in a suitable form to compute the Viscosity Coef. A
Properties utility
will perform these and Viscosity Coef. B using the following equation based on the fluid state.
calculations for you.
For a liquid, A and B are computed as follows:
Cstk1
A= ———————————
exp [ B / ( T1 + 459.67) ]
42
SL9100 Flow Computer
Notes:
1) A minimum of three points must be set up.
2) If "0" is entered for the frequency of any point other than point 1, the Flow
Computer assumes there are no more points above the points that preceded
them. (The display will then advance to the next setup menu.)
3) If the input frequency is above the highest or below the lowest frequency
programmed, the unit will use the last known point for the K factor in computing
the resulting actual flow.
Parameters:
Determine closest point above input signal
signal = X, K factor (correction factor) = KA
KAKA
K factor
Universal 7.6 Universal Viscosity Curve (UVC) using Hz/Cstk and K-Factor Data
Viscosity Curve A Universal Viscosity Curve is a presentation of the calibration of a turbine
flowmeter's K-Factor as a function of the Hz/cstks. It is used to represent the
combined effects of flowrate and viscosity on the calibration of the flowmeter. It
is entered as a table of point pairs in ascending order of Hz/cstks.
Strouhal Roshko 7.7 Strouhal Roshko Curve (StRo) using ST-RO Data
Curve A Strouhal Roshko Curve is a presentation of the calibration of a turbine flow-
meter's calibration as a table or curve of Strouhal number as a function of
Roshko number. It is used to represent the combined effects of flowrate, flow-
ing temperature and viscosity on the calibration of the turbine flowmeter. It is
entered as a table of point pairs in ascending order of Roshko numbers.
43
SL9100 Flow Computer
44
SL9100 Flow Computer
Print System Setup Press enter to begin printing the system setup.
Press ENTER to print
This message will display as the data
Print System Setup transmission takes place.
—— Printing ——-
Press Menu to get back to print system setup
Print System Setup top-level menu.
45
SL9100 Flow Computer
46
SL9100 Flow Computer
Calibration
Calibrate Aux1: 0mA To Calibrate: Connect Current Source (+)
Submenu Group
Iin=TB1-8 GND=TB1-4 TB1-8, (-) TB1-4. Input 0mA and press Enter.
47
SL9100 Flow Computer
Advance to
Cal RTD 100ohm
48
SL9100 Flow Computer
8.2.11
Calibrate 0mA Aout
Calibrate 0mA Aout Connect ammeter to (+) TB1-15, (-) TB1-16.
Submenu Group + TB1-15 - TB1-16 Press enter.
To trim 0mA analog output: Press CLEAR
Calibrate 0mA Aout to enable editing and enter a small negative
Enter mA: 0.00000 number (i.e. -0.100) to force a display reading,
then clear and enter small quantity measured
on your meter.
Calibrate 0mA Aout The display will return to Calibrate 0mA out.
+ TB1-15 - TB1-16 Press the down arrow key to advance to the
20mA analog out or repeat above if necessary.
8.2.12
Calibrate 20mA Aout
Calibrate 20mA Aout Connect ammeter to (+) TB1-15, (-) TB1-16.
Submenu Group + TB1-15 - TB1-16 Press enter.
8.2.13
Analog In Test Analog In Test Press enter to test the analog inputs.
Submenu Group
49
SL9100 Flow Computer
8.2.14 Pulse input test Press Enter key to test the pulse input.
Pulse input test
Submenu Group
2.5V
10mV Pulse input test Use the Up/Down arrow keys to select the
100mV Trigger level 2.5V appropriate trigger level.
40Hz
Pulse input test Use the Up/Down arrow keys to select the
3KHz
20kHz count speed 3kHz appropriate frequency range.
Pulse input test Press Menu key to return to Pulse input test
top-level menu.
Analog out test Press Menu key to return to Analog out test
top-level menu.
8.2.16 Excitation out test Press Enter to test the excitation output.
Excitation out test
Submenu Group To test the excitation output: Connect a
voltmeter to (+) TB1-1, (-) TB1-4. Press the
Excitation out test key under the desired setting to move the
*5v 12v 24v asterisk (*). The unit should output the selected
voltage.
Excitation out test Press Menu key to return to Excitation out test
top-level menu.
50
SL9100 Flow Computer
8.2.17 Pulse out test Press Enter key to test the pulse output.
Pulse out test
Submenu Group
To simulate a frequency on the pulse output:
Connect a frequency counter to (+)TB1-13,
Pulse out test (-)TB1-14. Press the key under the desired
*0Hz 1Hz 10Hz 20Hz setting to move the asterisk (*). The unit
should output the selected frequency.
Pulse out test Press Menu key to return to Pulse out test
top-level menu.
Control inputs test Press Menu key to return to control input test
top-level menu.
51
SL9100 Flow Computer
8.2.20 Battery Voltage Test Press Enter key to view the battery voltage.
Battery Voltage
test
Submenu Group
Battery Voltage Test The display will show the battery voltage.
3.312 Volts Replace battery at 2.5 VDC or below.
Battery Voltage Test Press Menu key to return to battery voltage test
top-level menu.
8.2.21 Data logger utility Press Enter to use data logger utility.
Data logger utility
Submenu Group
Data logger utility The displays shows the number of Data Logs.
Log 10 958 Max Press the Down arrow key to advance to PRT
(print) or CLR (clear).
Data logger utility Press PRINT key to output data logger logs
Log 00001 PRT CLR to printer, Press CLEAR key to clear the data
logger contents.
Data logger utility Press Menu key to return to Data logger utility
top-level menu.
52
SL9100 Flow Computer
Instructions:
1. Make sure you follow proper E.S.D. Precautions. All persons performing this
replacement must follow proper grounding procedures.
2. Turn the power to the unit off.
3. Disconnect the two piece connector rear terminal block, leaving all connections in place.
4. Remove the unit from the panel.
5. Remove the four machine screws (see fig. 1) which hold the two sections of the case
together.
6. The rear section of the case should detach from the rest of the case. It may be
necessary two cut the wiring label along the joint where the two sections connect. With
the rear section of the case removed the fuse will be exposed (located near the rear
terminal, AC connection).
7. Locate the Fuse F1 (see fig. 2) and unplug the fuse from its socket.
8. Insert the new fuse into the socket. Insure that the pins are fully inserted and straight.
9. Reassemble the case and install the four machine screws which join the two sections of
the case.
10. Reinstall the unit into the panel.
11. Reconnect the rear terminal block.
12. Turn the unit back on.
Fuse Specifications:
110 VAC Power: 160mA/250V, TD Wickman 19372-030-k or equivalent
220 VAC Power: 80mA/250V, TD Wickman 19372-026-k or equivalent
12/24 VDC Power: 800mA/250V, TD Wickman 19374-046-k or equivalent
fig. 1 fig. 2
Screws
(4 places) fuse
FUSE
53
SL9100 Flow Computer
Transaction Printing
Data Logging
Remote Metering by Modem (optional)
Computer Communication Link
Configuration by Computer
Print System Setup
Print Calibration/Malfunction History
For transaction printing, the user defines the items to be included in the printed document. The user
can also select what initiates the transaction print generated as part of the setup of the instrument. The
transaction document may be initiated via a front panel key depression, a remote contact closure, or
upon completion of a batch.
In data logging, the user defines the items to be included in each data log as a print list. The user can
also select when or how often he wishes a data log to be made. This is done during the setup of the
instrument as either a time of day or as a time interval between logging.
The system setup and maintenance report list all the instrument setup parameters and usage for the
current instrument configuration. In addition, the Audit trail information is presented as well as a status
report listing any observed malfunctions which have not been corrected.
The user initiates the printing of this report at a designated point in the menu by pressing the print key
on the front panel.
5 4 3 2 1
9 8 7 6 Phone Line for
Internal Modem
Option
RS-232 RS-485 1 2
1 Handshake Line
5 4 3 2 1 5 4 3 2 1
9 8 7 6 9 8 7 6
2 Transmit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
3 Receive
4 Do Not Use
5 Ground
6 Do Not Use
7 Do Not Use
8 Do Not Use
9 Do Not Use
54
SL9100 Flow Computer
10.2 General
The optional RS-485 card utilizes Modbus RTU protocol to access a variety of pro-
cess parameters and totalizers. In addition, action routines can be executed. For
further information, contact factory and request RS-485 Protocol manual for SL9100.
1 Ground
Phone Line for
2 Ground
5 4 3 2 1
Internal Modem
Option
9 8 7 6
5 4 3 2 1 5 4 3 2 1
9 8 7 6 9 8 7 6
4 TX/RX (+)
5 TX/RX (-) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
6 Do Not Use
7 Terminating Resistor (180 Ω)
8 TX/RX (+)
9 TX/RX (-)
55
SL9100 Flow Computer
The SL9100 setup program provides for configuring, monitoring and controlling a
SL9100 unit over the RS-232 link.
Sample applications are stored in disk files. The setup program calls these
Templates. You can store the setup from the program’s memory to either the
SL9100 (Downloading the file) or to a disk file (Saving the file) for later usage.
Similarly you can load the setup in program memory from either a disk file (Opening
a file) or from the SL9100 unit (Uploading a file).
The program can monitor outputs from the unit while it is running.
The program can reset alarms and totalizers.
For assistance there are mini-helps at the bottom of each screen in the program.
There is also context sensitive help available for each screen accessible by pressing
the F1 key.
11.3 Installation
For Windows 95™ or higher Click the Start button, select Run and proceed as
follows:
Type the floppy drive letter followed by a colon (:) and a backslash (\), and the word
setup. For Example:
a:\setup
Follow the instructions on your screen.
56
SL9100 Flow Computer
into groups containing various configuration and/or monitoring functions. To view the
tab windows, simply click on the tab. The previous tab window will be hidden as the
new tab window is brought to the foreground.
57
SL9100 Flow Computer
The setup software assumes the current setup has been uploaded from the flow
computer into the PC. It is important that the setup program and the SL9100 unit
are using the same setup information at all times or the data will be inconsistent. It is
best to upload or download the setup before using this feature.
To start the viewer, first check the boxes of items to view and then click the start
button. The data will appear in the appropriate sections and will be continuously
updated. The refresh rate is dependent on the number of items that are being
viewed and the baud rate of the connection. Data in the List of Values section can
be collapsed by clicking on the ‘minus’ sign in front of the group title. The data can
be expanded by clicking on the ‘plus’ sign in front of the group title. If a group is
collapsed and data in the group changes on refresh, the group will automatically
expand. Changing the view items requires stopping the current viewing, checking
the new selections and then restarting the viewer.
If communication errors occur while reading data from the SL9100 device, the word
‘Error’ will appear in place of the actual value. If the connection to the SL9100 is
lost, the viewer will time out with a message saying the device is not responding.
The viewer will attempt to communicate with the SL9100 device matching the
device ID set in the communications screen. If you are having trouble establishing
communication, compare settings for the PC and the flow computer. Also verify the
connections between the PC and flow computer.
58
SL9100 Flow Computer
Analog Output
The analog signal (4-20mA) that is generated by the SL9100. It can correspond to the Rate, Total, Temperature or
Density. This output is used primarily for transmission of process information to remote systems.
Audit Trail
The audit trail is used to track the number of changes made to the units setup program.
Batch Overrun
The SL9100 offers a batch overrun compensation routine. If batch overrun occurs due to slow valve response
time, the unit will compensate for the overrun amount on the next batch. This feature can be disabled if desired.
NOTE: A non zero "Maximum Drain Time" value is required for normal operation with this function.
Batcher
An instrument which controls the dispensing of desired batch amounts. Liquid batching systems are usually
comprised of a batch controller (batcher), flowmeter and control valve. The batcher opens and closes the valve
through the use of relays and measures the amounts of liquid being dispensed via the flowmeter.
Baud Rate
The speed of serial communication transmissions, expressed in bits per second.
Calibration Temperature
The temperature at which a flow sensor was calibrated on a test fluid.
Custody Transfer
Weights and Measure metering codes often specify several requirements for instruments and mechanisms to
prevent and track changes in the setup of an instrument which may be used in the commercial sale of goods. The
SL9100 tracks changes via the Audit Trail.
Data Logger
The capturing of information for later use and the mechanism for specifying the conditions where a capture should
be made.
Default Value
The value to be used by the instrument if a sensor failure or out of ranch signal is detected.
Expansion Factor
See C-Factor
Flow Alarm
A visual indication that the volumetric flowrate is above or below the flow alarm setpoint specified by the user.
59
SL9100 Flow Computer
Flow Equation
A recognized relationship between the process parameters for flow, temperature and density used in flow
measurements.
Follow, Alarm
Alarm relays which are non latching and whose output state is based solely on the comparison of the current
process value and the alarm setpoint (trip point).
Function Key
A key on a push-button panel or keyboard (whose function is described by the key label) used to perform an
instrument function or special routine.
Handshake
A means of controlling the information flow between two pieces of equipment to prevent the sending device from
transmitting information at a rate faster than what can be accepted by the receiver.
Hysteresis
The relay hysteresis is a "dead band" setting which allows the relay to remain energized for a given amount below
the setpoint. This is used to prevent relay chatter when the process value is near the setpoint value.
Example: If the Preset is set at 100, and the hysteresis is set at 10, the relay will energize when the rate, temp or
dens. reaches 100, the relay will remain energized until the reading falls below 90.
Input Termination
Input signal lines on digital inputs often require pullup or pulldown resistor configurations to operate properly
with different sensor configurations. The SL9100 contains such resistors and may be enabled via the setup menu.
Inhibit Totalizer
"Inhibit Total" is a Control Input 1 setting that is used to stop the totalization. If enabled, a voltage level on control
input 1 will inhibit the total as long as the voltage is present. This feature is useful during meter proving and in
applications that provide a sensor to signal the flow computer when fluid is present.
K-Factor
A scaling factor derived from the pulses produced by a flowmeter output, expressed in pulses per unit (i.e. pulses/
gallon)
Limit Setpoint
An alarm trip point setting which specifies the value or magnitude of a process parameter necessary to activate an
alarm indicator or control relay.
Linear Flowmeter
A flow measurement device whose output is proportional to flow.
Linearization
The mathematical correction of a nonlinear device. The SL9100 uses a linearization Table which is made up of
input/output values and makes interpolations of the table to arrive at a "linearized" measurement.
LinTbl
Abbreviation for Linearization Table.
60
SL9100 Flow Computer
Mass Flow
Mass Flow is inferred by the volumetric flow and density (or implied density) of a fluid.
Max Window
The max. window time sets the maximum sample time (1 to 99 sec) for the ratemeter.
Parity
A method for detecting errors in transmissions of serial communications data.
Preset
A set point used to trigger the relay outputs of the SL9100.
Print Interval
The print interval allows the SL9100 to transmit information to the serial port at selectable time intervals.
Private Code
An operator password code which authorizes changes to the setup of the instrument but blocks access to the
Service/Calibration/Test mode. The private code also blocks the clearing of the Grand Total.
Process Parameters
Any sensor information which has been scaled to engineering units including Flow, Temperature and Density.
Pulse Output
The pulse output of the SL9100 is available for remote accumulation of the total or sent to peripheral devices,
such as a PLC. The output can be scaled using the Pulse Output Scaling Constant.
Quad
Abbreviation for Quadrature. Quadrature signals are used for direction control. Two flowmeter signals are output
with a 90° phase shift. The counter counts UP when channel A precedes channel B, and counts DOWN when
Channel A lags Channel B.
Quick Update %
This feature is used to disable the rate averaging filter when a significant change in the flow rate occurs. The user
can enter the percent of change needed to be detected to disable the averaging feature. This is especially useful
during start-up and shutdown of flow.
61
SL9100 Flow Computer
Ratemeter
Any device used to display the speed of a process. The ratemeter in the SL9100 displays flow rate.
Ref. Dens.
Abbreviation for Reference Density. This is the fluid density at reference conditions of temperature.
Ref. Temp.
Abbreviation for Reference Temperature. This represents the base or reference condition to which corrected flow
will be computed.
Roshko
A parameter defined as: Ro = f • temperature correction
cstk
Single_Pulse
The Single_Pulse setting is used for flowmeters with single pulse outputs.
Strouhal
A calibration parameter defined as temperature corrected K-factor for a turbine flowmeter.
Time Constant
A damping factor for an averaging filter for the analog output. (see also Rate Averaging Filter)
Totalizer
Any device which accumulates and displays a total count.
UVC
Abbreviation for Universal Viscosity Curve. A presentation of the combined flowrate/viscosity calibration for a
turbine flowmeter.
Visc Coef
Abbreviation for Viscosity Coefficient. One or more coefficients in an equation used to describe the viscosity as a
function of temperature for a fluid.
Volume Flow
The measurement of volumetric flow.
62
SL9100 Flow Computer
Some alarms are self clearing. Other alarms require the user to
acknowledge and clear the alarm. Press the ENTER button to
acknowledged and clear alarms. Alarms may reassert themselves if
the alarm condition is still present.
63
SL9100 Flow Computer
No
No system or process
errors present.
64
SL9100 Flow Computer
AUX INPUT TOO LOW! 4-20 mA Input current at aux • Check wiring
input smaller than 3.5 mA: • Check function of
• Faulty Wiring sensor
• Transmitter not set to "4-20
mA"
• Transmitter defective
MODEM NOT PRESENT The setup expects modem • Check setup for
usage and a modem is not proper baud rate,
responding. parity, etc.
• Check modem
connection and
cycle power to the
unit.
• Replace modem
EXTENDED PFI LOCKUP Unit was operated with an • Check data in unit.
input power level lower than Totalizer may have
safe operating range for an inaccuracies
extended period of time. • Investigate
brownout cause.
65
SL9100 Flow Computer
66
SL9100 Flow Computer
Appendix A
Water
Temp Deg F Centistoke #/gal
194 0.33 8.060
40 1.55 8.350
Mil C-7024
Temp Deg F Centistoke #/gal
100 1.02 6.315
60 1.374 6.370
JP-8/Jet A
Temp Deg F Centistoke #/gal
185 0.83 6.340
-40 12.00 7.068
JP-4
Temp Deg F Centistoke #/gal
185 0.49 5.862
-40 2.75 6.659
Dexron III
Temp Deg F Centistoke #/gal
212 7.72 6.788
32 270.00 7.344
AvGas
Temp Deg F Centistoke #/gal
185 0.34 5.483
-40 1.48 6.267
PAO
Temp Deg F Centistoke #/gal
185 2.10 6.571
-40 269 6.973
MIL-PRF-5606
Temp Deg F Centistoke #/gal
250 3.8 6.578
80 18.0 7.107
MIL-PRF-83282
Temp Deg F Centistoke #/gal
185 4.55 6.669
40 85.00 7.110
MIL-PRF-23699
Temp Deg F Centistoke #/gal
250 3.55 7.518
80 44 8.070
NOTE: The tabulated values are typical. For most accurate calculations, the temperature versus viscos-
ity and density tables should be edited to just cover the temperature range anticipated and with the
actual values for the fluid in use. Temperature and density values can change from brand to brand
and lot to lot of products.
67
SETUP MENUS
instrument Instrument batch count max. batch batch over- auto batch Time delay flow signal max. drain slow start no tot for
type mode preset run comp. restart timeout time quantity batch delay
type
setup indica- total de- volume units total dec density de- mass units density dec density rate time rate de- rate dec rate average quick temp de-
setup indicators tors scriptor places scriptor places default base scriptor places filter update % scriptor
excitation pulse input pulse trig- low pass input max window k-factor Average Average change change Low flow high flow meter ex-
setup FLOW INPUT voltage type ger type filter termination type KA-Factor KB-Factor table A table b rate alarm rate alarm pansion
aux1 input aux1 signal current aux low aux full aux low aux high calibration h20 density
setup aux1 INPUT type type range scale scale alarm alarm temperature at 4degc
set fluid properties fluid number fluid name Compute fluid Viscosity temperature viscosity temperature viscosity density fluid class reference temperature density temperature
properties method point 1 point 1 point 2 point 2 method temperature point 1 point 1 point 2
setup pulse output pulse ouput pulse width pulse value density reference expansion viscosity viscosity
68
usage point 2 density factor coef a coef b
setup analog output analog ou- analog out analog out analog out analog out analog out
put usage flow type range low scale full scale damping
setup
RELAYs relay usage RELAY relay relay mode relay set- relay hyster-
setup RELAYS 1, 2, 3, 4 delay duration point esis
setup con-
setup control inputs control in- control in- control in-
trol inputs
put 1 usage put 2 usage put 3 usage
1, 2, 3
These functions will only appear with appropriate
setup real time clock setup real clock type select time of day enter date settings in other functions.
time clock clock am/pm
serial hard- DEVICE ID BAUD RATE PARITY HANDSHAKE device line modem op- modem init MODEM Auto CALL OUT day CALL OUT CALL ON ER- CALL OUT NO NUMBER OF
serial usage ware feed tions master answer of week time ROR/alarm REDIALS
setup setup data- output page length top margin datalog print time print interval enable print print end of clear total print list HANGUP IF
log/print format only key batch if print items 2min. INACTIVE
datalog/print
administrative setup tag number operator supervisor software product unit serial sensor serial display new
password password version order code number number error only
69