TOC - EPA Method 415.3
TOC - EPA Method 415.3
Revision 1.0
June, 2003
B. B. Potter, USEPA, Office of Research and Development, National Exposure Research Laboratory
J. C. Wimsatt, The National Council On The Aging, Senior Environmental Employment Program
415.3 - 1
METHOD 415.3
DETERMINATION OF TOTAL ORGANIC CARBON AND SPECIFIC UV ABSORBANCE
AT 254 nm IN SOURCE WATER AND DRINKING WATER
1.1 This method provides procedures for the determination of total organic carbon (TOC),
dissolved organic carbon (DOC), and UV absorption at 254 nm (UVA) in source
waters and drinking waters. The DOC and UVA determinations are used in the
calculation of the Specific UV Absorbance (SUVA). For TOC and DOC analysis, the
sample is acidified and the inorganic carbon (IC) is removed prior to analysis for
organic carbon (OC) content using a TOC instrument system. The measurements of
TOC and DOC are based on calibration with potassium hydrogen phthalate (KHP)
standards. This method is not intended for use in the analysis of treated or untreated
industrial wastewater discharges as those wastewater samples may damage or
contaminate the instrument system(s).
1.2 The three (3) day, pooled organic carbon detection limit (OCDL) is based on the
detection limit (DL) calculation.1 It is a statistical determination of precision, and
may be below the level of quantitation. The determination of OCDL is dependent on
the analytical instrument system’s precision, the purity of laboratory reagent water
(LRW), and the skill of the analyst. Different TOC instrument systems have
produced significantly different OCDLs that range between 0.02 and 0.12 mg/L OC
for both TOC and DOC measurements. Examples of these data can be seen in
Section 17, Table 17.1. It should be noted that background levels of OC
contamination are problematic. The minimum reporting level (MRL) for TOC and
DOC will depend on the laboratory’s ability to control background levels (Sect. 4).
2.1 In both TOC and DOC determinations, organic carbon in the water sample is oxidized
to produce carbon dioxide (CO2), which is then measured by a detection system.
There are two different approaches for the oxidation of organic carbon in water
samples to carbon dioxide gas: (a) combustion in an oxidizing gas and (b) UV
promoted or heat catalyzed chemical oxidation with a persulfate solution. Carbon
dioxide, which is released from the oxidized sample, is detected by a conductivity
detector or by a nondispersive infrared (NDIR) detector. Instruments using any
combination of the above technologies may be used in this method.
2.2 Settleable solids and floating matter may cause plugging of valves, tubing, and the
injection needle and/or injection port. The TOC procedure allows the removal of
settleable solids and floating matter. The suspended matter is considered part of the
415.3 - 2
sample. The resulting water sample is then considered a close approximation of the
original whole water sample for the purpose of TOC measurement.
2.3 The DOC procedure requires that the sample be passed through a 0.45-µm filter prior
to analysis to remove particulate OC from the sample.
2.4 The TOC and DOC procedures require that all IC be removed from the sample before
the sample is analyzed for organic carbon content. If the IC is not completely
removed, significant error will occur. The sample, which is then free from IC
interference, is injected into a TOC instrument system. The organic carbon is
oxidized to CO2, which is released from the sample, detected, and reported as mg/L or
ppm TOC or DOC.
2.5 The UVA procedure requires that the sample be passed through a 0.45-µm filter and
transferred to a quartz cell. It is then placed in a spectrophotometer to measure the
UV absorbance at 254 nm and reported in cm-1.
2.6 The SUVA calculation requires both the DOC and UVA measurement. The SUVA is
calculated by dividing the UV absorbance of the sample (in cm-1) by the DOC of the
sample (in mg/L) and then multiplying by 100 cm/M. SUVA is reported in units of
L/mg-M. The formula for the SUVA may be found in Section 12.2.
3.1 ANALYSIS BATCH - A set of samples prepared and analyzed on the same
instrument during a 24-hour period. For a TOC/DOC analysis batch, the set may
contain: calibration standards, laboratory reagent blank and/or filter blanks, field
blank, field samples, laboratory fortified matrix sample, field duplicate sample, and
continuing calibration check standards. For a UVA analysis batch, the set may
contain: filter blanks, field samples, field blank, field duplicate sample, and
spectrophotometer check solutions with associated blank. An analysis batch is
limited to 20 field samples. QC samples are not counted towards the 20 sample limit.
QC requirements are summarized in Table 17.6.
3.2 BLANKS - Prepared from a volume of LRW (Sect. 3.9) and used as needed to fulfill
quality assurance requirements and to monitor the analytical system.
415.3 - 3
calibration standards over time by analyzing it as a sample and comparing the
results with initial analysis of the CB.
3.2.3 FILTER BLANK (FB) - The FB is an aliquot of LRW that is filtered and
analyzed using the same procedures as field samples undergoing DOC and
UVA determinations. For DOC and UVA analyses, the FB serves as the LRB.
The FB will give an indication of overall contribution of organic carbon
contamination from laboratory sources such as the LRW itself, labware
cleaning procedures, reagents, the filter apparatus, filter, and instrument
system(s).
415.3 - 4
calibration check standards (CCC), and laboratory fortified matrix samples
(LFM).
3.5 FIELD DUPLICATES (FD1 and FD2) - Two separate samples collected at the same
time and place under identical circumstances, and treated exactly the same throughout
field and laboratory procedures. Analyses of FD1 and FD2 give a measure of the
precision associated with sample collection, preservation, and storage, as well as
laboratory procedures.
3.6 INORGANIC CARBON (IC) - Carbon in water samples from non organic sources,
composed mainly from dissolved mineral carbonates and carbon dioxide. IC can
interfere with the determination of TOC and DOC if it is not removed.
3.9 LABORATORY REAGENT WATER (LRW) - The LRW may be distilled and/or
deionized (DI) water, or high pressure liquid chromatography (HPLC) reagent grade
415.3 - 5
or equivalent water which is low in TOC concentration, meeting the requirements as
stated in Section 7.2.
3.16 SOURCE WATER - Surface water or ground water that is used by a drinking water
utility to produce potable water for public consumption.
3.18 TOTAL CARBON (TC) - A measure of the OC and IC contained in a water sample.
In this method, IC is removed from the sample. Therefore, the TC reported by a TOC
instrument system will be equal to the TOC or DOC measurement.
415.3 - 6
3.19 TOTAL ORGANIC CARBON (TOC) - The gross amount of organic matter (carbon
not removed by the IC removal step) found in natural water. Suspended particulate,
colloidal, and dissolved organic matter are a part of the TOC measurement. For this
method, the TOC definition excludes the contribution of floating vegetative or animal
matter, and volatile organic matter found in source water. Settleable solids consisting
of inorganic sediments and some organic particulate are not transferred from the
sample by the laboratory analyst and are not a part of the TOC measurement.
Acronym Term
CB calibration blank
CCC continuing calibration check
commercial spectrophotometer background
COMM-BKS
solution
commercial spectrophotometer check
COMM-SCS
solution
DOC dissolved organic carbon
FB filter blank
FD field duplicate
FRB field reagent blank
IC inorganic carbon
IDC initial demonstration of capability
KHP potassium hydrogen phthalate
LFB laboratory fortified blank
LFM laboratory fortified matrix
LRB laboratory reagent blank
LRW laboratory reagent water
MRL minimum reporting level
MSDS material safety data sheet
OC-CAL organic carbon calibration standard
415.3 - 7
Acronym Term
OC-PDS organic carbon primary dilution standard
OCDL organic carbon detection limit
QCS quality control sample
SCS spectrophotometer check solution
SDWA Safe Drinking Water Act
SOP standard operating procedure
SUVA specific UV absorbance
TC total carbon
TOC total organic carbon
UVA UV absorbance
4.2 All glassware must be meticulously cleaned. Wash glassware with detergent and tap
water, rinse with tap water followed by reagent water. Non-volumetric glassware may
then be heated in a muffle furnace at 425 oC for 2 hours to eliminate interferences.
Volumetric glassware should not be heated above 120 oC. Alternate cleaning
procedures, such as acid rinsing and heating at lower temperatures, may be employed,
providing that these procedures are documented in a laboratory SOP and LRBs are
monitored as per Section 9.9.
4.3 Laboratory water systems have been known to contaminate samples due to bacterial
breakthrough from resin beds, activated carbon, and filters. Laboratory water systems
should be maintained and monitored frequently for carbon background and bacterial
growth. It is recommended that the LRW be filtered through a 0.22-µm filter
membrane to prevent bacterial contamination of TOC instrument systems, reagents,
415.3 - 8
and samples. The LRW, sample transfer (pipet), glassware, and sample bottles are the
principle source for organic background in the analytical system. However, it is not
possible to control all sources of organic carbon contamination. Therefore, this
method allows for instrument background correction or adjusting the zero reference
point of the instrument for organic carbon background that is found in the analytical
system. 2 There are many ways to correct for organic carbon background. Consult the
instrument manufacturer’s operation manual for the instrument background correction
procedure. Subtraction of LRB or FB measurements from TOC, DOC, or UVA
sample results is not allowed.
4.4 High concentrations of OC, both man-made and naturally occurring, can cause gross
contamination of the instrument system, changes in calibration, and damage to valves,
pumps, tubing, and other components. It is recommended that analysis of a sample
known to have a concentration of OC > 10 mg/L OC be followed by the analysis of an
LRB. It is highly recommended that known samples containing OC concentrations
> 50 mg/L OC be diluted or not run on instruments used to analyze low-level drinking
water samples.
4.5 Source waters containing ionic iron, nitrates, nitrites, and bromide have been reported
to interfere with measurements of UVA absorbance at 254 nm. 3 The concentration of
the interferences and their effect on the UVA cannot be determined as each unique
sample matrix may produce a different UVA response for the same concentration of
interference or combination of interferences. This method does not treat or remove
these interferences. Therefore, suspected or known interferences may affect results
and must be flagged in the SUVA result as “suspected UVA interferences.”
4.6 Chloride exceeding 250 mg/L may interfere with persulfate oxidation methods.4, 5
Some instrument systems may require increased persulfate concentration and
extended oxidation times. Consult with your instrument manufacturer’s
representative or instrument operation manual for instrument settings and reagent
strengths when analyzing samples containing high levels of chloride.
4.7 Inorganic carbon (IC) interferes with TOC and DOC measurements. TOC instrument
bias due to incomplete IC removal has been reported.6, 7 If inorganic carbon is not
completely removed from the water sample, it will result in a positive or negative bias
depending on the way the instrument system calculates TOC (e.g., TOC =TC - IC,
TC = TOC + IC, or TOC = TC). When inorganic carbon (IC) is removed from the
sample prior to the TOC assay, as required in this method, TOC = TC and the method
bias is minimized.
5.0 SAFETY
5.1 Fast-moving source water, steep inclines, water conduits, and electrical hazards may
present special safety considerations for the sample collector. The sample collector
415.3 - 9
should be aware of any potential safety hazards and take necessary precautions while
collecting samples.
5.2 Each chemical reagent used in this method should be regarded as a potential health
hazard. Exposure to these compounds should be minimized and/or avoided by active
participation in safety planning and good laboratory practices.8 Each laboratory is
responsible for maintaining a current awareness file of OSHA regulations9 regarding
the safe handling of the chemicals specified in this method. Material Safety Data
Sheets (MSDS) containing information on chemical and physical hazards associated
with each chemical should be made available to all personnel involved in the
chemical analysis.
5.3 Potassium persulfate is a strong oxidizing and corrosive reagent. The analyst should
avoid eye and skin contact by wearing eye/face protection, powderless gloves and
laboratory clothing. If body tissue comes in contact with this reagent, apply large
quantities of water for at least 15 minutes (see MSDS) while removing contaminated
clothing. This reagent may cause delayed burns. Seek immediate medical attention if
the area becomes irritated or burned. This reagent can also cause a fire or explosion if
it is allowed to come in contact with combustible materials.
5.4 Protect your hands by wearing laboratory disposable gloves during the preparation
and disposal of corrosive (acids and oxidants) laboratory reagents. Do not reuse
laboratory gloves that have been discarded or are suspected of being contaminated.
415.3 - 10
capability requirements as stated in Section 9.3.2, and that FB requirements are met
(Sect. 9.9). It is recommended that the analyst review the AWWA journal article
“Selecting filter membranes for measuring DOC and UV254”, Karanfil, et. al.10, prior
to the selection of an alternative filter membrane, apparatus, and wash procedure.
Karanfil tested 11 filter membranes (0.45-µm pore size and 47-mm disc size)
representing four different manufacturers and seven different types of filter materials
for both desorption and adsorption. Hydrophilic polyethersulfone (PES) filters
available from two manufacturers (Osmonics Micro-PES and Gelman Supor 450,
both 0.45 micron absolute pore size and 47-mm disc size) and a hydrophilic
polypropylene filter (Gelman GH Polypro, 0.45 micron absolute pore size and 47-mm
disc size) were found to be the best options among those tested in the study.
6.2 INJECTION VIALS - Specially cleaned 40-mL glass vials, with cap and
polytetrafluoroethylene (PTFE)/silicone septa. Eagle-Picher TOC Certified, Cat. No.
40C-TOC/LL, Eagle-Picher Technologies®. These vials are specially cleaned by the
manufacturing process and certified to contain < 10 µg TOC. Vials may be reused if
cleaned as per Section 4.2. The PTFE/silicone septa once pierced by the sample
injector must be discarded.
6.3 INSTRUMENT SYSTEMS - The TOC and UVA procedures allow for the use of
several different types or combinations of TOC instrumental system technologies.
Examples of typical TOC instrument systems, as well as a UV spectrophotometer, are
described below. Data from these instruments may be found in Section 17. Only one
TOC instrument is required to perform this method.
415.3 - 11
combustion in air or oxygen reagent gas to produce CO2, which is then
detected by an NDIR detector.
6.4 LABORATORY REAGENT WATER TREATMENT SYSTEM - The LRW used for
the development of this method was generated using a Millipore®, Milli-Q Plus
Ultra-Pure Water Treatment System with a 0.22-µm sterile pack filter capable of
producing organic carbon free (< 0.010 mg/L OC), ultrapure deionized water.11 The
maximum amount of OC allowed in the LRW for this method is 0.35 mg/L. When
purchasing a treatment system for general laboratory use, it is recommended that a
system be purchased capable of producing LRW of the above stated quality in order
to be of use in other laboratory analyses.
6.7 PIPET, DISPOSABLE TRANSFER - Large volume bulb (15mL), non-sterile, with
flexible long stem polyethylene transfer pipet. “Sedi-Pet ™”, Fisher Scientific® Cat.
13-711-36. Pipets are used for sample transfer from the middle of a sample bottle
containing floating material (scum).
415.3 - 12
Glass bottles may be reused after cleaning (see Sect. 4.2 for glassware cleaning
instructions) or discarded.
6.10 VACUUM SOURCE - Aspirator, air flow or water flow, hand-operated or low
pressure electric vacuum pump, providing a vacuum of 15 inches of mercury (Hg) or
better. If an alternative choice is made, see note in Section 6.1.
6.12 VOLUMETRIC FLASK AND PIPETS - All volumetric glassware used in this
method are required to be “Class A”.
415.3 - 13
7.0 REAGENTS AND STANDARDS
NOTE: The chemicals required for this method must be at least reagent grade.
Unless otherwise indicated, it is intended that all reagents shall conform to the
specifications of the Committee on Analytical Reagents of the American Chemical
Society (ACS) and/or ACS certified, when available. Some instrument manufacturers
provide reagents specifically prepared for the optimum performance of their TOC
instruments and provide calibration services and/or calibration standards. The
analyst is allowed to use these services or prepare reagents and/or standards
according to the instrument manufacturer’s operation manual.
7.1 COMPRESSED GASES – Carbon dioxide free Ultra High Purity (UHP) grade
nitrogen gas or an optional Ultra-low level TOC gas delivery system. For combustion
based TOC systems, zero grade air and UHP grade oxygen may be needed. The use
of lesser grades of compressed gases will result in high background noise in the TOC
instrument systems. The TOC Instrument 1 described in Section 6.3.1. does not
require compressed gasses for operation.
7.2 LABORATORY REAGENT WATER (LRW) - Water that has a TOC reading of
< 0.35 mg/L and < 0.01 cm-1 UVA. Although the LRW TOC and UVA limits in this
method are 0.35 mg/L and 0.01 cm-1, respectively, the system specified in Section 6.4
is capable of producing better quality organic carbon free, ultrapure deionized water.
For optimum performance, it is recommended that LRW with < 0.05 mg/L TOC and
< 0.0045 cm-1 UVA be used for this method. Alternatively, LRW may be purchased
(ACS HPLC grade or equivalent).
7.4 O-PHOSPHORIC ACID (85%), [H3PO4, CAS# 7664-38-2] - ACS grade or better.
415.3 - 14
7.7.1 PERSULFATE REAGENT - Prepare this solution according to the instrument
manufacturer’s instructions or purchase the solution from the instrument
manufacturer. If the laboratory prepares the solution, transfer the solution to
the instrument reagent bottle and cap. It is recommended that this solution be
sparged gently with carbon dioxide free UHP grade nitrogen gas for
approximately 1 hour. If the instrument system provides continuous sparge, it
is recommended that the reagent bottles be allowed to sparge for 10 minutes to
1 hour before operating the instrument. Self contained reagent packs or other
types of reagent systems may not require reagent sparging. Discard the
solution as per expiration time/date listed in the manufacturer’s operation
manual.
415.3 - 15
7.8.1.3 CALCIUM NITRATE TETRAHYDRATE, [Ca(NO3)2 C 4H2O, CAS#
13477-34-4] - ACS grade or better.
415.3 - 16
FLASK SALT WEIGHT
(1 L) (g)
A magnesium sulfate heptahydrate, MgSO4 C 7H2O 2.565
B ammonium chloride, NH4Cl 0.594
calcium chloride dihydrate, CaCl2 C 2H2O 2.050
calcium nitrate tetrahydrate, Ca(NO3)2 C 4H2O 0.248
potassium chloride, KCl 0.283
sodium chloride, NaCl 0.281
C sodium bicarbonate, NaHCO3 2.806
sodium phosphate dibasic heptahydrate, Na2HPO4 C 7H2O 0.705
D sodium-meta silicate nonahydrate, Na2SiO3 C 9H2O 1.862
415.3 - 17
prepared using LRW preserved to pH # 2 with concentrated acid (see note,
Sect. 7.8). Filtration of the CAL standards for DOC analysis is unnecessary,
since interferences from the filtration unit are monitored via the FB.
Therefore, the OC-CAL may be applied to TOC or DOC determinations. The
OC-CAL standards must be sparged, or otherwise treated for IC removal, like
a sample following the procedure in Section 11.5.
CB – 0 1000 –
1 500 1.0 1000 0.5
2 500 2.0 1000 1.0
3 500 4.0 1000 2.0
4 500 10.0 1000 5.0
5 500 20.0 1000 10.0
415.3 - 18
analyzed. The IN-SPEC™ optical standard and background solution for a 254 nm
spectrophotometric check is NIST traceable, and is available from GFS Chemicals,
PO Box 245, Powell, Ohio 43065.
7.10.2 KHP-SCS - Prepare the KHP-SCS that will provide an absorbance similar to
the absorbance of the samples analyzed. Prepare a 1-L volumetric flask
containing approximately 500 mL of LRW. Transfer and dissolve 4.08 g
anhydrous KH2PO4 and 2.84 g anhydrous Na2 HPO4 into the 500 mL of LRW.
From the example calculation, or table located below (Sect. 7.10.2.1), transfer
the amount of OC-PDS (in mL) needed to produce the representative
absorbance of the sample into the buffered KHP-SCS and dilute with LRW to
the 1 L mark.
415.3 - 19
KHP-OC conc. = (UV254 ! 0.0018) ' 0.0144
The 5.431 mg/L is the same as 5.431 mg/L KHP. It follows that to
produce a 1-L KHP-SCS solution having a UVA absorbance of 0.08
cm-1, you will need 10.9 mL of OC-PDS as calculated below:
KHP-SCS Preparation
OC-PDS (mL
UVA@254nm ORGANIC
added per liter of
(cm -1 ) CARBON (mg/L)
LRW)
0.0738 5 10
0.1458 10 20
0.2898 20 40
0.4338 30 60
7.10.3 Verify that the KHP-SCS-BLANK and the KHP-SCS buffered solutions are at
pH 7. Check the pH by placing a drop from the SCS bottle onto pH test paper.
Do not put the pH paper into the SCS bottle. Placing the pH paper in the
bottle will contaminate the sample with organic carbon. If this happens, the
spectrophotometer check solution must be discarded and a new solution
prepared in a clean bottle. If the buffered KHP-SCSs are not at a pH of 7, the
415.3 - 20
solution must be discarded and a new solution made. Store these solutions at
approximately < 6 °C. These solutions are not preserved. In a sterile
environment these solutions may be stable for a month. However, the shelf life
of these solutions may be shortened as a result of microbial growth. Therefore,
it is recommended that the above solutions be made fresh weekly and/or be
replaced if any significant change in absorbance is noted.
8.1.1 The DOC sample must be filtered in the field or in the laboratory within 48
hours of sample collection according to the procedure detailed in Section 11.4
prior to acidification and analysis. After filtration, the DOC sample is acidified
with 1 mL of concentrated acid per 1 L of sample or the sample is preserved by
drop wise adjustment to a pH < 2 (Sect. 8.3). The DOC bottle is capped and
inverted several times to mix the acid and is stored at < 6 °C. The sample must
be analyzed within 28 days from time of collection.
415.3 - 21
8.1.2 The UVA sample must be filtered in the field or in the laboratory according to
the procedure detailed in Section 11.4. The sample used for the UVA
determination is not acidified. The UVA bottle is capped and stored at < 6 °C
for up to 48 hours from the time of collection. The UVA sample must be
analyzed within 48 hours from the time of collection.
8.2 TOC SAMPLE COLLECTION - The typical sample volume collected may vary from
40 mL to 1 L of sample. It is recommended that the sample collector coordinate the
size of collection volume with the needs of the analytical laboratory. If the TOC
sample is collected in a 40-mL injection vial, it is acidified to pH < 2 by adding 2
drops of concentrated acid. If the TOC sample is collected in a 1-L bottle, 1 mL of
concentrated acid is added or the sample is drop wise adjusted to a pH < 2 (Sect. 8.3).
TOC samples must be acidified at the time of collection. Cap the bottle or injection
vial and invert several times to mix the acid. The sample is delivered as soon as
possible to the laboratory and should arrive packed in ice or frozen gel packs. If there
is no visible ice or the gel packs are completely thawed, the laboratory should report
the conditions to the data user. Samples shipped that are improperly preserved, and/or
do not arrive at the laboratory within 48 hrs, cannot be used for compliance monitoring
under the SDWA. The sample is stored at < 6 °C, until analysis. Stored and preserved
samples must be analyzed within 28 days from time of collection.
8.3 SAMPLE pH CHECK - The pH of the preserved sample (DOC, TOC only) or filtrate
should be checked to ensure adequate acidification for the preservation. This should
only be performed by an adequately trained sample collector. Check the pH by placing
a drop from the sample onto pH test paper. Do not put the pH paper into the sample
bottle. Placing the pH paper in the sample bottle will contaminate the sample with
organic carbon. If this happens, the sample or filtrate must be discarded and a new
sample collected.
9.1 Each laboratory using this method is required to operate a formal quality control (QC)
program. QC requirements for TOC include: the initial demonstration of laboratory
capability (IDC) followed by regular analyses of continuing calibration checks (CCC),
independent quality control samples (QCS), laboratory reagent blanks (LRB), field
duplicates (FD), and laboratory fortified matrix samples (LFM). For this method, a
TOC laboratory fortified blank (LFB) is the same as a CCC (Sect. 10.3) and no LFB is
required. QC requirements for DOC include: the IDC followed by regular analyses of
CCCs, QCSs, filter blanks (FB), LFB, FDs, and LFMs.
For laboratories analyzing both TOC and DOC samples, only the DOC IDC
determination is required, as it is similar to, yet more rigorous than, the TOC IDC.
The IDC must be performed the first time a new instrument is used and/or when a new
analyst is trained.
415.3 - 22
QC requirements for UVA analysis include: the performance of the IDC followed by
the regular analysis of spectrophotometer check solutions (SCS), FBs, and FDs. For
UVA analysis, no LFB or DL determination is required.
The control of instrument background is crucial prior to the performance of the IDC.
It is required that a critical evaluation be made of the instrument background 2
associated with an instrument system before proceeding with the IDC. Once an
acceptable instrument background is established, it is safe to proceed with the IDC.
415.3 - 23
DOC quantitation, an IDC of the IC removal is performed. Please note: any
manipulation of the sample may inadvertently introduce organic carbon from
the apparatus.
The LRB during the above study was < 0.05 mg/L, thus a 20-minute sparge
time ensured that no measurable organic carbon remained in the sample.
The above sparge efficiency table should be used only as a guide. The analyst
may find that a higher flow rate may reduce the time necessary to remove the
inorganic carbon to a level at or near the TOC measurements found in the LRB.
The IC-TEST solution is also used to test alternate IC removal apparatus that
remove IC by internal chemical treatment, alternate sparging procedures,
and/or membrane IC removal. Any alternative procedure or IC removal
apparatus must be tested using the IC-TEST solution and meet the LRB
requirements as stated in Section 9.9.
415.3 - 24
value. If ±20% of the true value is exceeded, identify and correct the problem
and repeat Sections 9.2.5 and 9.2.6.
where:
t(n-1,1-alpha = 0.99) = Student's t value for the 99% confidence level with n-1 degrees
of freedom (t = 3.14 for 7 replicates)
n = number of replicates, and
S = standard deviation of replicate analyses.
If the initial OCDL exceeds 0.35 mg/L or the mean recovery of the LFB used in
the OCDL determination exceeds + 50% of the true value, then the OCDL
determination must be repeated.
415.3 - 25
the samples, or by adsorption (uptake) of DOC and UV-absorbing materials
from the samples onto the filters. Filter membranes selected for DOC and
UVA measurements must not desorb nor adsorb significant DOC and UV-
absorbing materials. Desorption is minimized by pre-washing selected filters
as described in Section 9.3.2.2. Adsorption is minimized by filtering a portion
of the sample to waste before sample collection as described in Section 9.3.2.3.
Because the filtration of relatively turbid samples may cause filters to clog, pre-
filtration may be necessary and pre-filter preparation is described in Section
9.3.2.1. Due to the possibility of lot-to-lot variations in the levels of
contamination or adsorption, it is recommended that for each filter lot, the user
determine the amount of LRW needed to wash the filters and the amount of
sample that needs to be filtered and discarded prior to collection of filtrate
(filter-to-waste volume). A minimum of three filters (from each new lot)
should be cleaned and checked for desorption/adsorption prior to using the
filters for actual samples. This evaluation must be repeated when filters are
purchased from another manufacturer or when the type of filter being used is
changed.
415.3 - 26
in Section 9.9. The volume of LRW required to obtain acceptable filter
blanks is then used to clean filters for analyses of all samples (Sect.
11.4). Filters that cannot be cleaned to meet the referenced criteria
must not be used in the preparation of DOC and UVA samples.
415.3 - 27
filter-to-waste volume is the sum of the volumes of that filtrate and any
previous filtrates in the series. For example, if the unfiltered sample
has a TOC concentration of 3.5 mg/L and the filtrate series (each filtrate
= 30 mL) have concentrations of 2.3, 3.2, and 3.4 mg/L, then a
minimum of 60 mL of sample should be filtered-to-waste prior to
collecting filtrate for DOC analyses. It is recommended that the filter-
to-waste volume be determined by performing this test on at least three
filters from each lot and averaging the results. Filters that require
large volumes of filter-to-waste should be avoided, because they will
be more subject to clogging prior to the collection of the necessary
volume of filtrate for analysis. The filter-to-waste volume that is
determined in this section must be used in the filtration procedure
described in Section 11.4.4.
9.3.3 Perform Sections 9.2.5 through 9.2.7 using filtered LFBs. The LFBs must be
prepared using the same procedure used to prepare samples (Sect. 11.4).
9.4.3 Conduct the filter membrane suitability study described in Section 9.3.2 for
UVA.
9.5 CONTINUING CALIBRATION CHECK (CCC) - With each analysis batch, analyze a
Low-CCC at or below the MRL (Sect. 9.10) prior to TOC or DOC sample analysis.
Subsequent CCCs are analyzed after every ten samples and after the last sample. The
concentrations should be rotated to cover the instrument calibration range. A Mid-
CCC is required during every analysis batch. Acceptance criteria are as follows: Low-
415.3 - 28
CCC, + 50% of true value; Mid-CCC, + 20% of true value; High-CCC, + 15% of true
value, see Section 10.3 for concentrations.
9.6 FIELD DUPLICATE (FD) - Within each analysis batch, a minimum of one set of field
duplicates must be analyzed (FD1 and FD2). Sample homogeneity and the chemical
nature of the sample matrix can affect analyte recovery and the quality of the data.
Duplicate sample analyses serve as a check on sampling and laboratory precision.
Two samples are collected at the field site and are treated exactly alike.
9.6.1 Calculate the relative percent difference (RPD) for duplicate measurements
(FD1 and FD2) using the equation:
9.6.2 Relative percent difference for field duplicates having an average concentration
of > 2 mg/L OC should fall in the range of < 20% RPD. If field duplicates in
this concentration range exhibit an RPD greater than 20%, results should be
flagged and the cause for the greater difference (e.g. incomplete IC removal or
matrix interference), investigated. UVA readings should be < 10% RPD.
NOTE: Greater variability may be observed for samples with OC approaching
the OCDL.
9.7 LABORATORY FORTIFIED BLANK (LFB) - Within each DOC analysis batch,
analyze an aliquot of reagent water or other blank matrix which has been fortified with
KHP at a concentration of 1-5 mg/L OC. Recovery for the LFB must be within ±20%
of the true value. One LFB is required with each DOC analysis batch. For the DOC
analysis, an LFB is subjected to the same preparation and analysis as a sample,
including filtration (Sect. 11.4). The LFB is not determined for the TOC or UVA
measurements.
9.8 LABORATORY FORTIFIED MATRIX (LFM) - Within each TOC or DOC analysis
batch, an aliquot of one field sample is fortified with an aliquot of the OC-PDS (Sect.
7.8.3). The spike concentration used should result in an increase in the LFM
concentration of 50 to 200% of its measured or expected concentration. Over time,
samples from all routine sample sources should be fortified. For DOC analysis, the
LFM is filtered prior to acidification and analysis.
9.8.1 Calculate the percent spike recovery (%REC) using the equation:
415.3 - 29
where
A = measured concentration in the fortified sample
B = measured concentration in the unfortified sample,
and
C = fortification concentration.
9.8.2 Recoveries may exhibit a matrix dependence. If the LFM recovery falls outside
of 70 to 130% for any fortified concentration, the analyst should suspect that
inorganic carbon was not properly removed (Sect. 11.5) from the sample or that
contamination or matrix interference exists (Sect. 4) and can not be removed.
If the source of the poor recovery can not be identified, the analyst should label
the sample report “suspect/contamination or matrix interference” to inform the
data user that the sample data quality is questionable but should not be rejected.
Failure to meet the recovery criteria after repeated sampling may suggest that
the sample matrix may need further study.
9.9 LABORATORY REAGENT BLANK (LRB) AND FILTER BLANK (FB) - Within
each analysis batch, a minimum of one LRB must be analyzed. For DOC and UVA
analysis, the FB serves as the LRB. If more than one lot of filters is used in a DOC or
UVA analytical batch, a FB must be analyzed for each lot. The analyst should be
aware that additional filter blanks, up to one for each sample, are required by some
regulations (e.g., 40 CFR 141.131(d)(4)(i)).
The LRB or FB is used to assess contamination from the laboratory environment and
background contamination from the reagents used in sample processing and is treated
exactly the same as a sample. The volume of the FB must be the same as the sample
volume. If UVA is to be determined, the FB (UVA-FB) must have an absorbance of
< 0.01 cm-1 UVA. The LRB and/or the FB (DOC-FB) must be < 0.35 mg/L OC. If
0.35 mg/L OC or 0.01 cm-1 UVA is exceeded, background carbon or reagent
contamination should be suspected. The cause for significant changes in the LRB or
FB value must be identified and any determined source of contamination must be
eliminated. For the FB, this may mean redetermination of filter membrane suitability
(Sect. 9.3.2). The cause of the contamination and the corrective action used to remedy
the problem is then recorded in the instrument log for future reference.
9.10 MINIMUM REPORTING LEVEL (MRL) - The OCDL should not be used as the
MRL. For TOC analysis, it is recommended that an MRL be established no lower than
the mean LRB measurement plus 3F, or two times the mean LRB measurement,
whichever is greater. For DOC analysis, the FB is substituted for the LRB. This value
should be calculated over a period of time, to reflect variability in the blank
measurements. Although the lowest calibration standard for OC may be below
the MRL, the MRL for OC must never be established at a concentration lower
than the lowest OC calibration standard.
415.3 - 30
9.11 QUALITY CONTROL SAMPLE (QCS) - During the analysis of the IDC (Sects. 9.2,
9.3), each time new OC-PDS solutions are prepared (Sect. 7.8.3), or at least quarterly,
analyze a QCS from a source different from the source of the calibration standards.
The QCS is used to provide an independent verification of the method and the TOC
instrument system. To verify the stock or calibration solutions by comparison with the
QCS, dilute the calibration solution and QCS to a concentration in the mid range of the
calibration curve (approx. 1 - 5 mg/L TOC) in the same manner that the OC-CAL
standards are made (Sect. 7.8.4). Acceptable verification of the calibration is made
when the means of 3 analyses for both the calibration solution and QCS, having a
concentration range between 1 to 5 mg/L OC, agree to within ±20% of the true value.
If the measured QCS concentration is not within ±20% of the true value, the
calibration solution must be remade and/or the source of the problem must be
determined and corrected. Analysis of the QCS only applies to TOC and DOC
determination.
415.3 - 31
10.2 CALIBRATION CURVE - A new calibration curve is generated when fresh standards
are made (Sect. 7.8.4) or when CCCs fall out of QC limits (Sect. 10.3). Use a CB and
at least four OC-CAL standards that span the concentration range of the samples to be
analyzed. For example, if the samples to be analyzed are low in concentration (a range
falling between 0.5 to 5 mg/L OC), prepare a calibration blank and a minimum of four
TOC calibration standards (CB, OC-CAL 1 - 4, see Sect. 7.8.4). The lowest
concentration calibration standard must be at or below the MRL, which may depend on
system sensitivity. Add an additional 40 µL of H3PO4, HCl, or H2SO4, depending upon
instrument requirements (Sect. 8.0), to the 40-mL injection vial(s). Sparge the
calibration standards using the IC removal procedure in Section 11.5 prior to
calibrating the instrument. Inject the standards from low to high concentration and
calibrate the instrument. Be careful not to extend the calibration range over too wide
of a concentration range as flow injection memory may cause analytical error (Sect.
9.2.3). The optional OC-CAL 6 - 7 may be used when operating the instrument in a
higher concentration range.
NOTE: For instruments that have an internal calibration setting, the calibration is
checked by comparing the five point calibration curve with the internal calibration
point. If the five point calibration curve does not agree with the internal calibration
using the CCC criteria in Section 10.3, the internal calibration of TOC instrument
must be reset by the manufacturer or adjusted by the analyst, following the
manufacturer’s operation manual.
10.2.1 With the instrument in the ready mode, initiate the automated instrument
calibration routine as per the instrument manufacturer’s operation manual.
The computer generated calibration curve must have r2 $ 0.993 before
proceeding with analyses. Ideally the instrument calibration should be
r2 $ 0.9995 for best results. After the instrument system has been calibrated,
verify the calibration using the Continuing Calibration Check (CCC, Sect.10.3)
and QCS (Sect. 9.11).
10.2.2 Save the data from the initial calibration curve and record it in the laboratory
notebook or instrument log. The initial calibration curve serves as a historical
reference so that future calibrations curves can be compared to determine if the
slope or sensitivity of calibration has changed. If the slope or sensitivity of the
instrument changes such that QC requirements cannot be met, consult the
instrument manual or lab SOP for corrective action, which may include
instrument maintenance and recalibration.
415.3 - 32
Mid-CCC. Subsequent CCCs should alternate between low, medium, and high
concentrations, and must end the analysis batch. In summary, at least one Low-CCC
and one Mid-CCC is analyzed with each analysis batch in order to verify the
calibration curve. It is recommended that low, mid, and high CCCs be used to verify
the calibration curve over time.
10.3.1 Low-CCC - the concentration range may vary from as low as 2 times the
OCDL up to 0.7 mg/L OC. The Low-CCC is used to verify the low end of the
calibration and must be at or below the MRL, which may depend on system
sensitivity. The recovery for the Low-CCC must be within + 50% of the true
value.
10.3.2 Mid-CCC - the concentration is varied between 1.0 mg/L to 5.0 mg/L OC.
The purpose of this CCC is to verify the precision and accuracy at the
calibration range where critical source water treatment decisions are made.
The Mid-CCC concentration may be varied to meet changing regulatory
requirements. The Mid-range CCC must be within ±20% of the true value. If
it is not, the TOC instrument system must be re-calibrated.
10.3.3 High-CCC - the concentration range is varied between 5 to 50 mg/L OC. The
selection of the High-CCC should be near the concentration of the highest OC-
CAL standard used. The purpose of this CCC is to bracket the concentration
the samples that are typically analyzed and to verify the upper range of the
calibration curve. High-CCC must be within ±15% of the true value. If it is
not, the TOC instrument system must be re-calibrated.
10.4.1 Using a transfer pipet fill the spectrophotometer cell with the COMM-BKS or
KHP-SCS-BLANK (Sects. 7.9.1, 7.10.1). Use this solution to zero the
spectrophotometer.
10.4.2 After the spectrophotometer is zeroed, empty the cell, clean with LRW, rinse
with methanol, dry with N2 or reagent grade air, and fill it with the KHP-SCS
or COMM-SCS.
10.4.3 Read the UVA of the KHP-SCS or COMM-SCS. The reading must be within
10% of the expected absorbance value. Record the absorbance of the KHP-
SCS or COMM-SCS in the spectrophotometer instrument logbook. Empty the
415.3 - 33
cell, clean with LRW, rinse with methanol, and dry with N2 or reagent grade
air.
10.4.4 If the SCS absorbance criteria stated above cannot be met, discard the COMM-
SCS or the KHP-SCS and purchase new COMM-SCS or remake the KHP-
SCS. Repeat Section 10.4.
11.0 PROCEDURE
11.3 TOC SAMPLE PREPARATION - Remove the TOC sample from cold storage and
allow the sample to come to room temperature. Determine if the sample has been
preserved by acidification to a pH # 2 by placing some drops on pH paper or by
pouring some of the sample into a small beaker and checking it with a glass or solid-
state pH electrode. Do Not put the pH paper or electrode into the sample bottle. If the
pH is greater than 2, discard the sample.
415.3 - 34
11.4 SUVA SAMPLE PREPARATION - If SUVA is not being determined, proceed to
Section 11.5. The SUVA determination consists of paired sample analyses composed
of a DOC sample and a UVA sample. DOC and UVA samples may be taken from the
same bottle, or may be taken from separate field duplicate bottles. Remove the DOC
and UVA sample(s) from cold storage and allow them to come to room temperature.
The laboratory is required to document any use of alternative filters, apparatus (see
note, Sect. 6.1), or changes in the SUVA sample preparation procedure. All QC
requirements (Sect. 9) must be met.
11.4.1 Samples for DOC and UVA analysis are NOT acidified in the field. The DOC
sample is acidified after filtration as described below and the UVA sample is
not acidified at all. Determine if the sample(s) was accidentally preserved by
placing a few drops from the sample on pH paper or by pouring some of the
sample into a small beaker and checking it with a glass or solid-state pH
electrode. Do Not put the pH paper or electrode into the sample bottle.
Placing the pH paper or electrode into the sample bottle will contaminate the
sample solution with organic carbon. If this happens, the sample must be
discarded. If the UVA sample pH is # 2, check to make sure that the sample is
actually for the UVA determination. It is possible that this sample is a TOC or
filtered DOC sample and was mislabeled as a UVA sample. If the sample set
was not mislabeled or switched but accidentally preserved, the sample must be
discarded. The analyst must check the date and time of collection to ensure
that the sample holding times listed in Section 8.1 have been met.
11.4.2 Filter Cleaning - Cleaning the filter apparatus, including the filter, removes
trace organic compounds that may have been left behind in the manufacturing
process. This cleaning must be done immediately prior to sample filtration.
Rinse the filter with LRW, using the cleaning procedure used to determine
filter membrane suitability (Sect. 9.3.2.2), including the cleaning of the pre-
filter if a pre-filter is necessary.
11.4.3 Filter Blank (FB) - Use a clean filter apparatus (prepared in Sect. 11.4.2) and
filter an aliquot of LRW into an injection vial for the DOC analysis and another
aliquot of LRW into a vial for UVA analysis (Figure 1). FB volume must be
the same as the sample volume collected in Section 11.4.4. During the
development of this method, approximately 250 mL of LRW was filtered and
aliquots were poured into two 40-mL injection vials and labeled as the DOC
and UVA FBs. If the DOC and UVA analyses are coming from two separate
bottles, a filter apparatus will be needed for each bottle and an FB should be
prepared from each apparatus. Add 40 µL of H3PO4, HCl, or H2SO4 (as
required by the various instrument types, Sect. 8.0) to the 40-mL DOC-FB
injection vial. Do not acidify the UVA-FB injection vial. These vials are
paired with the respective SUVA sample and retained for DOC-FB and UVA-
FB analyses.
415.3 - 35
11.4.4 Sample Preparation - Reassemble the filter apparatus. Pour enough sample
onto the filter to saturate any adsorption sites, as determined according to the
filter-to-waste procedure in Section 9.3.2.3. Apply vacuum until no visible
water remains on the filter. Remove the vacuum, swirl the apparatus with
sample filtrate, disassemble, and discard the sample filtrate rinse. Reassemble
the filter apparatus and pour an additional aliquot of sample into the top of the
filter apparatus. Attach the vacuum and retain the filtrate. Pour one aliquot
into a 40-mL injection vial and label it to identify it as the DOC sample. Pour
a second aliquot into a 40-mL injection vial and label it to identify it as the
UVA sample. Add 40 µL of H3PO4, HCl, or H2SO4 to the 40-mL DOC
injection vial. Do not acidify the UVA injection vial. As with the DOC and
UV FBs (Sect. 11.4.3), separate filter apparatus may be used for the DOC and
UVA samples, in which case the filtrate need not be split into two aliquots.
For a sample that is difficult to filter, an additional filter apparatus or the
optional pre-filter insert apparatus may be used. The use of additional filters
may require the collection of additional FBs, collected as specified in Section
11.4.3. The resulting additional DOC-FB, UVA-FB sample filtrates are
collected, their volumes composited and then placed into their respective
injection vials.
11.5 INORGANIC CARBON REMOVAL - All OC-CALs, TOC and DOC samples,
DOC-FBs, and LRBs must be treated to remove IC prior to OC analysis. UVA
samples and UVA-FBs are not sparged with nitrogen gas or otherwise treated to
remove IC prior to analysis (See Figure 2). The laboratory is required to document
any use of alternative IC removal apparatus (Sects. 6.9, 11.5.2) or changes in the IC
removal procedure. All quality control requirements (Sect. 9.2.4) must be met.
NOTE: If a sparging apparatus is used, it should be isolated from the organic
laboratory and be free of organic contaminants.
415.3 - 36
carbon removal apparatus that may produce an efficient means for the removal
of IC. The laboratory must demonstrate sparging efficiency by the performance
of the IC removal sparging efficiency test (Sect. 9.2.4 ) and meeting the LRB
requirements as stated in Section 9.9.
12.1 TOC DIRECT READING: The TOC concentration is calculated by the automated
instrument system’s software. Follow the instrument manufacturer’s operation manual
when making instrument response adjustments for instrument system blank
corrections. The TOC calculation assumes that the sample has been properly
preserved, that only a trace amount of IC remains following the IC removal procedure,
and that any remaining IC will not contribute to the TOC measurement and result in a
calculation error. Some instrument systems calculate TOC from the difference of the
total carbon (TC) minus the IC. The analyst is reminded that the IC in the sample is
removed prior to sample analysis. Therefore, the reported TC is equal to, and the same
415.3 - 37
as, the TOC value (TOC =TC) and is read directly from the instrument’s computer or
printout.
where:
12.3 Calculations should utilize all available digits of precision, but final reported
concentrations should be rounded to two significant figures (one digit of uncertainty).
The final calculation is rounded up or down according to Standard Methods 1050B.15
13.1 Table 17.1 summarizes the 3-day organic carbon detection limit (OCDL) study for five
TOC instruments systems. The DOC determination ranged from 0.02 to 0.08 mg/L
OCDL and the TOC determination ranged from 0.04 to 0.12 mg/L OCDL. All source
415.3 - 38
water samples reported in Section 13 and the Section 17 Tables were sparged for 20
minutes to remove inorganic carbon interferences.
13.2 Table 17.2 and associated sub-tables illustrate the single instrument precision and
accuracy for each of the five TOC instrument technologies.
13.3 Tables 17.3 and 17.4 illustrate the instrument differences and performances for five
TOC instruments analyzing seven different source water matrices.
13.4 In all cases, the TOC instruments had difficulty in analyzing the Saint Leon well water.
The Saint Leon well water had a moderately high inorganic carbon content of
approximately 100 mg/L IC, and a low organic carbon content of 0.2 to 0.6 mg/L OC.
The Saint Leon well water organic carbon content was near the organic carbon
detection limit. The low OC concentration produced the greatest differences between
instrument responses. For low TOC samples with high IC, differences between
instrument responses may be more apparent due to possible IC interference.
13.5 The TOC, DOC and SUVA procedures of this method are dependent on the operation
manual for the TOC instrument system and the UV spectrophotometer as provided by
the respective instrument manufacturers. However, all performance criteria and
quality control requirements described in this method, as summarized in Tables 17.5
and 17.6, must be met.
14.1 Pollution prevention encompasses any technique that reduces or eliminates the
quantity or toxicity of waste at the point of generation. Numerous opportunities for
pollution prevention exist in laboratory operations. The EPA has established a
preferred hierarchy of environmental management techniques that places pollution
prevention as the management option of first choice. Whenever feasible, laboratory
personnel should use pollution prevention techniques to address their waste generation.
When wastes cannot be feasibly reduced at the source, the Agency recommends
recycling as the next best option.
14.2 For information about pollution prevention that may be applicable to laboratories and
research institutions, consult Less is Better: Laboratory Chemical Management for
Waste Reduction, available from the American Chemical Society's Department of
Government Relations and Science Policy, 1155 16th Street N.W., Washington D.C.
20036, (202)872-4477.
14.3 For recycle information, contact the US EPA, Pollution Prevention and WasteWise
program, http://www.epa.gov/wastewise/ .
415.3 - 39
15.0 WASTE MANAGEMENT
15.1 The U.S. Environmental Protection Agency requires that laboratory waste management
practices be conducted consistent with all applicable rules and regulations. The
Agency urges laboratories to protect the air, water, and land by minimizing and
controlling all releases from hoods and bench operations, complying with the letter and
spirit of any sewer discharge permits and regulations, and by complying with all solid
and hazardous waste regulations, particularly the hazardous waste identification rules
and land disposal restrictions. For further information on waste management, consult
The Waste Management Manual for Laboratory Personnel, available from the
American Chemical Society at the address listed in Section 14.2.
15.2 The laboratory should consult with local authorities prior to disposal of any waste to
publicly owned treatment works (POTW) and receive permission for that disposal.
16.0 REFERENCES
1. Glaser, J. A.; Foerst, D. L.; McKee G. D.; Quave, S. A.; Budde, W. L. Trace Analyses
for Wastewaters. Environ. Sci. Technol. 1981, 15 (12), 1426-1434.
2. Benner, R.; Storm, M. A Critical Evaluation of the Analytical Blank Associated with
DOC Measurements by High-Temperature Catalytic Oxidation. Mar. Chem. 1993, 41,
153-160.
3. Standard Method 5910B: Ultraviolet Absorption Method. In Standard Methods for the
Examination of Water and Wastewater, Eaton, A. D.; Clesceri, L. S.; Greenberg, A. E.,
Eds.; American Public Health Association; Washington, DC, 1995; 19th ed.
4. Aiken, G.R. Chloride Interference in the Analysis of Dissolved Organic Carbon by the
Wet Oxidation Method. Environ. Sci. Technol. 1992, 26 (12), 2435-2439.
5. Sakamoto, T.; Miyasaka, T. TOC Analysis Study Confirming the Accuracy of a Method
for Measuring TOC by Wet Oxidation. Ultrapure Water 1987, 24-31.
6. Potter, B. B.; Wimsatt, J. C. Preprints of Extended Abstracts, Vol 42 (1), 223rd National
Meeting of the American Chemical Society, Orlando, FL, April 7-11, 2002; American
Chemical Society Division of Environmental Chemistry: Cape Girardeau, MO, 2002;
Paper 60, 559-564.
415.3 - 40
8. American Chemical Society, Committee on Chemical Safety. Safety in Academic
Chemistry Laboratories, Vol. 2, Accident Prevention for Faculty and Administrators, 7th
ed.; American Chemical Society: Washington, DC, 2003.
10. Karanfil, T.; Erdogan, I.; Schlautman, M. A. Selecting Filter Membranes for Measuring
DOC and UV254. J.—Am. Water Works Assoc. 2003, 95 (3), 86-100.
11. Standard Method 1080: Reagent-Grade Water. In Standard Methods for the Examination
of Water and Wastewater, Eaton, A. D.; Clesceri, L. S.; Greenberg, A. E., Eds.; American
Public Health Association; Washington, DC, 1995; 19th ed.
12. Schaffer, R. B.; Van Hall, C. E.; McDermott, G. N.; Barth, D.; Stenger, V. A.; Sebesta,
S. J.; Griggs, S. H. Application of a Carbon Analyzer in Waste Treatment. J.—Water
Pollut. Control Fed. 1965, 37 (11), 1545-1566.
13. Van Hall, C. E.; Barth, D.; Stenger, V. A. Elimination of Carbonates from Aqueous
Solutions Prior to Organic Carbon Determination. Anal. Chem. 1965, 37 (6), 769-771.
14. Kaplan, L.A. Comparison of High-Temperature and Persulfate Oxidation Methods for
Determination of Dissolved Organic Carbon in Freshwaters. Limnol. Oceanogr. 1992, 37
(5), 1119-1125.
15. Standard Method 1050B: Significant Figures. In Standard Methods for the Examination
of Water and Wastewater, Eaton, A. D.; Clesceri, L. S.; Greenberg, A. E., Eds.; American
Public Health Association; Washington, DC, 1995; 19th ed.
16. Standard Method 5510: Aquatic Humic Substances. In Standard Methods for the
Examination of Water and Wastewater, Eaton, A. D.; Clesceri, L. S.; Greenberg, A. E.,
Eds.; American Public Health Association; Washington, DC, 1995; 19th ed.
415.3 - 41
17.0 TABLES, DIAGRAMS, FLOWCHARTS, AND VALIDATION DATA
a
Organic Carbon Detection Limits were determined by analyzing 7 replicates over 3 days.
b
LRW fortified as specified in the table.
c
%RSD = percent relative standard deviation
d
%REC = percent recovery
INSTRUMENT:
1: UV/Persulfate/Wet Oxidation with Permeation/Conductivity Detection
2: Elevated Temperature/Catalyzed/Persulfate/Wet Oxidation/Nondispersive
Infrared Detection (NDIR)
3: UV/Low Temperature/Persulfate/Wet Oxidation/NDIR
4: Catalyzed/Combustion Oxidation(680 °C)/NDIR
5: High Temperature Combustion Oxidation/NDIR
415.3 - 42
17.2 SINGLE TOC INSTRUMENT PRECISION AND ACCURACY
Unfortified Sample
Fortified Sample Conc.
Source Conc.
Water Mean %RSD Mean %REC
415.3 - 43
17.2 SINGLE TOC INSTRUMENT PRECISION AND ACCURACY, cont’d.
Unfortified
Fortified Sample Conc.
Source Sample Conc.
Water Mean Mean %REC
415.3 - 44
17.2 SINGLE TOC INSTRUMENT PRECISION AND ACCURACY, cont’d.
Unfortified Sample
Fortified Sample Conc.
Source Conc.
Water Mean %RSD Mean %REC
415.3 - 45
17.2 SINGLE TOC INSTRUMENT PRECISION AND ACCURACY, cont’d.
Unfortified Sample
Fortified Sample Conc.
Source Conc.
Water Mean %RSD Mean %REC
415.3 - 46
17.2 SINGLE TOC INSTRUMENT PRECISION AND ACCURACY, cont’d.
Unfortified Sample
Fortified Sample Conc.
Source Conc.
Water Mean %RSD Mean %REC
415.3 - 47
17.3 PRECISION AND ACCURACY DATA FOR DOC AND SUVA MEASURED IN
SEVEN SOURCE WATERS ON FIVE INSTRUMENTSa
17.3.2 DOC Measurements for Seven Source Waters, Fortified with KHP, Three
Replicate Instrument Injections on Five Instruments
415.3 - 48
17.3 PRECISION AND ACCURACY DATA FOR DOC AND SUVA MEASURED IN
SEVEN SOURCE WATERS ON FIVE INSTRUMENTSa, cont’d.
17.3.3 Mean SUVA Calculation Based on the DOC Data in 17.3.1 for Five
Source Waters
SUVA b (L/mg-M)
UVA
Source Water Inst #1 Inst #2 Inst #3 Inst #4 Inst #5 Mean
(cm-1 )
Boulder Creek 0.4324 2.62 3.08 2.84 2.97 3.58 3.02
Shingobee R. 0.7440 2.50 2.88 2.75 2.77 3.25 2.83
Bolton Well 0.2364 1.86 2.28 2.01 1.91 2.62 2.14
Ohio R. (Fernbank) 0.7267 2.60 3.01 2.90 2.88 3.43 2.97
Muddy Creek 1.124 2.95 3.46 3.33 3.20 3.89 3.37
Great Miami R. 0.8948 2.81 3.33 3.07 3.05 3.69 3.19
St. Leon Well 0.0771 1.46 1.93 1.38 1.83 3.13 1.95
a
For instrument identification (by type) see Section 6.3.
b
SUVA calculated as described in Section 12.2.
415.3 - 49
17.4 PRECISION AND ACCURACY DATA FOR TOC MEASURED IN SEVEN
SOURCE WATERS ON FIVE INSTRUMENTSa
17.4.1 TOC Measurements for Seven Source Waters, Three Replicate Instrument
Injections on Five Instruments
Total Organic Carbon, mg/L, Unfortified Samples
Std
Source Water Inst #1 Inst #2 Inst #3 Inst #4 Inst #5 Mean %RSD
Dev
Boulder Creek 1.73 1.38 1.47 1.46 1.26 1.46 0.17 12
Shingobee R. 3.16 2.62 2.72 2.84 2.45 2.76 0.26 10
Bolton Well 1.32 1.05 1.16 1.12 0.93 1.12 0.14 13
Ohio R. (Fernbank) 3.02 2.46 2.58 2.81 2.31 2.64 0.28 11
Muddy Creek 4.24 3.41 3.18 4.04 3.34 3.64 0.47 13
Great Miami R. 3.51 2.89 2.92 3.42 2.72 3.09 0.35 11
St. Leon Well 0.66 0.39 0.45 0.28 0.32 0.42 0.15 35
17.4.2 TOC Measurements for Seven Source Waters, Fortified with KHP, from
Replicate Instrument Injections on Five Instruments
Std
Source Water Inst #1 Inst #2 Inst #3 Inst #4 Inst #5 Mean %RSD %RECb
Dev
Boulder Creek 12.1 11.3 11.2 11.4 11.0 11.4 0.43 4 99
Shingobee R. 13.0 12.7 12.6 12.5 12.1 12.6 0.32 3 98
Bolton Well 11.4 11.4 11.0 11.2 10.8 11.1 0.28 3 100
Ohio R. (Fernbank) 13.2 13.1 12.6 12.8 12.1 12.8 0.45 4 101
Muddy Creek 14.6 13.8 13.5 13.7 13.1 13.7 0.54 4 101
Great Miami R. 13.8 13.2 13.0 13.6 12.3 13.2 0.59 5 101
St. Leon Well 11.1 10.5 10.7 10.2 10.0 10.5 0.41 4 101
a
For instrument identification (by type) see Section 6.3.
b
% Recovery calculated as described in Section 9.8.
415.3 - 50
17.5 INITIAL DEMONSTRATION OF CAPABILITY (IDC) REQUIREMENTS
(SUMMARY)
415.3 - 51
Method Requirement Specification and Acceptance Criteria
Reference Frequency
Sect. 9.3.2 Initial Prior to the first use of FB < 0.35 mg/L OC and/or
Demonstration of filters and whenever a < 0.01 cm -1 UVA. Sample
Filter Membrane manufacturer or filter filtrate OC within + 15% of
Suitability type is changed. unfiltered sample OC.
Sect. 9.4.1 Initial Prior to first Test two wavelengths between
Spectrophotometer instrument use and 220 and 340 nm. Check
Check annually thereafter. manufacturer’s operation manual
for acceptance limits.
Sects. 9.4.2, Spectrophotometer Prior to analysis of UVA within 10% of expected
10.4 Performance Check samples. absorbance value.
415.3 - 52
17.6 QUALITY CONTROL REQUIREMENTS (SUMMARY)
415.3 - 53
Method Requirement Specification and Acceptance Criteria
Reference Frequency
Sect. 9.6 Field Duplicate One FD is collected and FD > 2 mg/L OC < 20%
(FD) Analyses analyzed with each analysis RPD. UVA < 10% RPD.
batch.
Sect. 9.7 Laboratory One LFB is analyzed with Concentration of 1-5 mg/L
Fortified Blank every DOC analysis batch. OC using KHP. Recovery
(LFB) analysis must be within + 20% of true
value.
Sect. 9.8 Laboratory One LFM is analyzed with Recovery outside 70-130%
Fortified Matrix every TOC or DOC warrants investigation of
(LFM) analysis batch. Spike matrix effect.
concentration should result
in an increase in the LFM
concentration of 50 to
200% of its measured or
expected concentration.
Sect. 9.11 Quality Control The QCS is analyzed The analyzed value of a 1-5
Sample (QCS) during the IDC, after each mg/L QCS must be within
new calibration curve, each ±20% of the true value.
time new calibration
solutions are prepared, or at
least quarterly.
Section Calibration Curve A new calibration curve is Calibration curve must have
10.2 generated when fresh r2 > 0.993 before proceeding
standards are made and/or with analyses.
when CCCs are out of QC
limits.
Section Spectrophotometer The day to day performance The UVA of the KHP-SCS or
10.4 performance check of the spectrophotometer is COMM-SCS reading must be
checked using the COMM- within 10% expected
SCS and/or KHP-SCS prior absorbance values. Analysis
to analyzing any UVA of LRW must result in UVA
sample(s). of < 0.01 cm-1.
415.3 - 54
FIGURE 1: FILTER BLANK PREPARATION
LRW
WASH FILTER*,
DISCARD
FILTER LRW*,
DISPENSE INTO
40-ML VIALS
DESIGNATE AS
FILTER BLANK (FB)
DOC - FB UVA - FB
ADD ACID NO ACID
SPARGE NO SPARGE
ANALYZE ANALYZE
415.3 - 55
FIGURE 2 : SAMPLE PREPARATION
WATER SAMPLE
415.3 - 56