5Mud Balance
5Mud Balance
check
GPS-IMS-CAL-F-15
REV-02 Calibration Certificate
Customer: Bohai Drilling Engineering Company {BHDC-112}
Date of receiving : 03-04-2024 Date of Issuing : 03-04-2024
Calibration Results:
The instrument under calibration was calibrated up to full scale according to the manufacturer's recommendations
,calibration method. The following gives the under calibration readings and the corresponding reference mean
readings:
Reading on Unit Under Calibration Measurement Uncertainty
Reference Mud in 20 ˚C
gm/cm3 ± gm/cm3
Water (0.998 gm/cm3) 0.999 0.01
Glycerin (1.260 gm/cm3) 1.261 0.01
Results: The above readings indicate that the Mud Balance Error is within the permissible limit,
At the time of calibration.
• This certificate will remain valid on the same date of the next calibration specified in the
certificate, unless any breakage or damage occurs to the item being calibrated, or it is
loaded with a load more than rated for it.
• This calibration certificate refers only to the particular item submitted for calibration.
• This certificate shall not be reproduced, unless written permission has been obtained
from the lab
• This certificate is valid only with signature and stamp.
• This calibration was carried out in compliance with ISO/IEC 17025.
GPS-IMS-CAL-F-15
REV-02 Calibration Certificate
Customer: Bohai Drilling Engineering Company {BHDC-112}
Date of receiving : 03-04-2024 Date of Issuing : 03-04-2024
Calibration Results:
The instrument under calibration was calibrated up to full scale according to the manufacturer's recommendations
,calibration method. The following gives the under calibration readings and the corresponding reference mean
readings:
Reading on Unit Under Calibration Measurement Uncertainty
Reference Mud in 20 ˚C
gm/cm3 ± gm/cm3
Water (0.998 gm/cm3) 0.999 0.01
Glycerin (1.260 gm/cm3) 1.261 0.01
Results: The above readings indicate that the Mud Balance Error is within the permissible limit,
At the time of calibration.
• This certificate will remain valid on the same date of the next calibration specified in the
certificate, unless any breakage or damage occurs to the item being calibrated, or it is
loaded with a load more than rated for it.
• This calibration certificate refers only to the particular item submitted for calibration.
• This certificate shall not be reproduced, unless written permission has been obtained
from the lab
• This certificate is valid only with signature and stamp.
• This calibration was carried out in compliance with ISO/IEC 17025.