Process CalibratoR Mt100
Process CalibratoR Mt100
Calibrator
Applications
● Multifunctional Sensor Calibrator: Calibrate
temperature, pressure, voltage, current, resistance, and
frequency. Since it both measures and sources, you
can troubleshoot and calibrate all with one rugged tool
● Multifunctional Sensor Simulator: Simulate multiple
sensors for troubleshooting and debugging systems.
Using this calibrator, engineers can identify the fault in
the sensor, controller or wiring by simulating sensor
output signals and running carrying out test procedures.
● Hand-Held Operation. Small enough to fit easily into a
tool bag and to use in tight spaces. Runs an entire shift on a rechargeable Li-ion battery
pack. Great for testing on the field.
● 2 operating modes: Measure & Source (generate) —enable technicians to troubleshoot,
calibrate, or maintain instrumentation with just one tool.
Page 1
MicroTech MT100 Multifuntion
Process Calibrator
● The signal calibrator can output a user-defined signal
when a user enters into signal value through digit buttons.
● Simulation and measurement of thermocouple and temperature sensors like PT100,
PT1000, CU50.. etc.
● Debugging of industrial field PLCs
● Process Instrumentation: Signal Generators, Signal measurement, Valve Adjustment,
Inverter control
● PLC & Control Panel debugging by bypassing sensors and sensor calibrations
● LED & Equipment Testing or Control
Engineers can power Sensor Transmitters under test using the inbuilt 24V loop supply with
simultaneous mA measurement so you can test a transmitter out of the circuit and better isolate
problems.
Source mA with simultaneous pressure measurement to conduct valve and I/P (Input
current/Output pressure) tests in a single device. You can also test flow meters with frequency and
CPM (counts per minute) functions and perform fast linearity tests with auto-step and auto-ramp
features so you can quickly verify the mA signal at different output levels.
It can output and measure 4 to 20mA Current, Voltage, Frequency Pulses, Thermocouple and
RTD Sensor. It is widely used as an mV mA calibrator, temperature calibrator, frequency signal
calibrator, 4-20mA loop calibrator, multimeter, etc. It supports full function operating mode and
user-defined output setting.
Features
● High accuracy of 0.025% +1/2 digits; widely used as frequency calibrator, mA mV loop
calibrator, temperature calibrator, etc.
● Have a full-function operating mode, the user can enter any value through digits buttons,
and the calibrator outputs a signal directly
● Signal output simulation and measurement: 0-10KHz frequency, 4-20mA source or passive
signal, 0-11V, 0-110mV, 20-400Ω, standard R, E, S, K, J, B, T, N type thermocouple, PT100,
CU50, 24V outputs, etc.
● Offers easy and convenient handheld operation. 2200mAH rechargeable battery inside, has
battery status indicator, very suitable and convenient to use indoors or outdoors in industrial
sites.
● High-resolution colour LCD display, has backlight for different lighting conditions, large
screen clear display
● Can be used as signal generator, frequency calibrator, and multifunction process calibrator.
● Have description in Hz, V, mV, mA, RTD, TC, Step, and Ω keys, it enables the user to
operate it easily
● Internal polarity Reverse connection and Over current protection; Protection voltage 30V
Page 2
MicroTech MT100 Multifuntion
Process Calibrator
Technical Specifications
Output Parameters
FUNC OUT UNIT RANGE SET STEP ACCURACY (SP: setting point)
DC. V V 0~11.00V 0.01/0.1/1 ±0.025%SP ±0.004V
DC. mV mV 0~110.00mV 0.1/1/10 ±0.025%SP ±0.02mV
Sink/Source/Prog
DC. mA 0~24.00mA 0.01/0.1/1/4 ±0.025%SP ±0.004mA
ram
0~400Ω 1/10/0100 ±0.025%SP ±0.3Ω
Ohm Ω
400~2200Ω 1/10/0100 ±0.05%SP ±1.5Ω
Frequency KHz 0-10 0.01/0.1/1 ±0.05%SP ±2Hz
R 0~1700℃
S 0~1600℃
B 500~1800℃ ±0.1%SP ±3℃
Page 3
MicroTech MT100 Multifuntion
Process Calibrator
Measurement Parameters
R 0~1700℃
B 500~1800℃
K -200~1370℃
TC 0.1℃
E -200~1000℃ ±0.025%RD ±0.5℃ (K)
±0.025%RD ±0.3℃ (E)
J -200~1200℃ ±0.025%RD ±0.4℃ (J)
±0.05%RD ±0.5℃(T)
T -200~400℃ ±0.025%RD ±0.8℃ (N)
N -200~1300℃
0~9990 engineering
unit (for current
Range switch mA measurement and (related to engine
measure the current
of 24V output)
Page 4
MicroTech MT100 Multifuntion
Process Calibrator
Page 5