Chapter 6 Summary AHL1
Chapter 6 Summary AHL1
AHL1 2025
● X-rays have a shorter wavelength, higher frequency and higher energy than visible
light. The higher energy is what makes x-rays dangerous.
● In an X-ray machine, there's a part called the cathode (negative side) that
releases electrons, which come from a component similar to a light bulb
filament. These electrons are then pushed towards another part called the anode,
which is the positive side. When the electrons hit the anode, they interact with it
and produce the radiation that is required for X-ray images.
● The anode heel effect refers to the unequal distribution of x-ray beam intensity
along the cathode-anode axis.
➢ Radiographic density
➢ Radiographic contrast
➢ Radiographic detail
➢ Distortion
○ The object being radiographed must be parallel to the recording surface and the
OID kept as short as possible so as to prevent foreshortening and increased
penumbra.
○ Distortion can also occur when the x-ray beam is not perpendicular to the
recording surface.
○ A small amount of padding beneath the patient brings the vertebral column
parallel to the recording surface, but care must be taken not to use too much
padding. This can elevate the spinal column, which produces a false narrowing of
the intervertebral spaces.
➢ Scatter radiation
○ When an x-ray photon strikes an object, it can do one of three things: pass
through the object, be absorbed by the object or produce secondary radiation.
○ Scatter radiation not only decreases the contrast of the image, but it's also a
serious safety hazard for both personnel and patients.
○ Beam-limiting devices (cones, diaphragms, collimators, and filters) are commonly
used to decrease scatter radiation by confining the primary beam to the area
being examined.
➢ Grids
○ As the thickness of the area being imaged increases, the amount of kVp required
also increases, which means more scatter radiation. To minimize this, grids are
necessary when radiographing areas 10cm or more in thickness.
○ The grids also absorb a portion of the usable x-rays, so to compensate for this
loss , the number of x-rays generated must be increased by increasing the mAs.
○ Focused grids have lead strips placed at progressively increasing angles to
match the divergence of the x-ray beam. By angling the lead strips, the cut off of
the primary beam is eliminated and radiographic density is uniform.
➢ mAs
➢ kVp
➢ Object-film distance
○ OID is the distance from the object being imaged to the recording surface.
○ Distance should be as short as possible to reduce penumbra effect and
magnification that occurs with a long OID.
➢ Radiographic film
○ X-ray film consists of three layers: a thin protective layer, an emulsion containing
silver halide crystals and a polyester film base. The emulsion coats both sides,
which gives the film greater sensitivity, increasing the speed, density and
contrast. Increasing the speed also decreases the required exposure for the
patient and personnel.
○ Film is sensitive to all types of electromagnetic radiation including gamma
radiation, particulate (alpha or beta) radiation x-rays, heat, and light.
○ Unexposed film should be stored in a cool, dry place away from chemical fumes.
It is also pressure sensitive and should be stored on an end and not laid flat on
its side.
➢ Intensifying screens
○ There are three basic types of x-ray equipment from which to choose: portable,
stationary or mobile units.
○ Basic equipment requirements for a stationary unit include the x-ray generator
system . collamiter, grid, table, tube stand and positioning aides. These
accessories allow individuals to be positioned farther from the primary beam.
○ Three main types of digital systems: computed radiology (CR), digital radiography
(DR), and charged coupled devices (CCD) technologies.
○ Digital radiographs can be enhanced and viewed on computer software that
enables contrast, brightness, zoom, and pan adjustments as well as
measurement of various anatomical structures.
○ Digital x-rays may reduce the number of repeat x-rays needed and eliminate the
need for film, chemicals and screens.
➢ Radiation safety
○ Tissues that are most sensitive to ionizing radiation are those with rapidly
growing or reproducing cells.
○ Blood-forming cells (especially lymphocytic blood cells) are relatively sensitive to
radiation. Damage to these cells reduces the resistance to infection and causes
clotting disorders.
○ Radiodermatitis (reddened, dry skin) can result from excessive, chronic, low-level
radiation exposure.
➢ Terminology
○ REM is used to express the dose equivalent to ionizing radiation. REM takes into
account the quality of radiation, so doses of different kinds of x-rays can be
compared. (1 SV= 100 REM) (MPD= maximum personal dose.
○ The recommended dose for occupationally exposed persons should not exceed
5 REM per year. Your badge should not exceed 0.05 REM per month.
○ Good radiation control programs consist of safe x-ray equipment, low-exposure
techniques , use of positioning aides, proper patient measurement, proper
positioning methods, shielding, and monitoring personal radiation exposure.
➢ Darkroom setup
○ The room must be just large enough to provide a dry bench area away from the
wet bench area to prevent chemical splashes. Cleanliness is also important.
○ The most important feature of a dark room is that it is light tight.
○ The darkroom should have adequate ventilation to prevent volatile chemical
fumes from accumulating in the room, which would cause fogged film, damage
to the electrical equipment, and health problems for personnel.
➢ Film identification
○ Permanent labeling is required for all images. This can be done during the
exposure or afterwards, but it must be done before the film is processed
○ The label should include the clinic name, date, owner’s name, address, patient
name and patient signalment.
○ Developer
■The developer’s main function is to convert the sensitized silver halide
crystals into black metallic silver.
■The developer contains 5 ingredients: a solvent, reducing agents,
restrainer, activator, and preservatives.
■Chemicals are manufactured in two forms: liquid and powder. The powder
form should never be mixed inside of the darkroom.
○ Fixer
■It removes the unchanged silver halide crystals from the film emulsion,
leaving the black metallic silver. It also hardens the film emulsion, which
decreases the susceptibility to scratches,
■The fixer contains five ingredients: a solvent, a fixing agent, an acidifier, a
hardener, and a preservative.
■It comes in two forms: powder and liquid.
○ Manually developing the x-ray film is a chemical process that depends on the
duration of the immersion in the chemicals and the temperature of the
chemicals.
○ Automatic processors can develop film more quickly and consistently provide
high quality radiographs.
○ The manual processing fixer solution, silver recovery systems, and old
radiographs can be sold to companies to reclaim the silver.
➢ Radiographic artifacts
○ They contain elements with a high atomic number (absorb more x-rays) including
water-soluble organic iodides or barium sulfate.
○ A commonly used oral form of water-soluble organic iodide is a solution of
meglumine and sodium diatrizoate. They will be absorbed into the bloodstream
and excreted by the kidneys.
○ Negative contrast agents include air, oxygen, and carbon dioxide, which will
appear radiolucent on the finished radiograph.
○ Care must be taken not to over-inflate the organs such as the bladder.
➢ Diagnostic ultrasound
○ A transducer sends low intensity, high-frequency sound waves into the soft
tissues where they interact with tissue interfaces. Some of the sound waves are
reflected back to the transducer and some transmitted into deeper tissues. The
ones reflected back to the transducer (echos) are analyzed by the computer to
produce a grey-scale image.
○ Radiographs demonstrate the size, shape and position of the organs. Ultrasound
displays the findings found on the radiograph as well as the soft tissue textures
and the dynamics of some organs, like motility of the bowel.
○ Transducers
■Sector scanners are useful when imaging areas limited by ribs, gas-filled
bowels, or lungs. Linear-array scanners are useful in areas with
unrestricted window size.
■The frequency of the transducer determines the amount of detail or
resolution of an image. As frequency increases, the wavelength gets
shorter and the shorter the wavelength is, the better the resolution of the
image will be.
○ Display modes
○ Patient preparation
■To produce the best quality images, the transducer head must be placed
in close contact with the skin, so the animal’s hair must be clipped or
shaved before the study.
■Fasting of small animals before abdominal ultrasound examination is
recommended.
○ Instrument controls
■Ultrasound equipment has many controls for adjusting the quality of the
image. Improper adjustment of any of these can greatly decrease the
quality of the image.
■Controls include brightness and contrast, depth, gain and power, and
Time Gain Compensation (TGC).
○ Artifacts
➢ Endoscopy
○ This is the best method of evaluating the digestive system because it provides
the opportunity to examine and obtain tissue samples without the invasiveness
of surgery.
○ Responsibilities of the vet techs assistitng with endoscopy include selection,
care and maintenance of the endoscopes and the care and positioning of the
patient.
○ Rigid
○ Flexible
➢ Endoscopy room
○ The room should never be so dark that proper anesthetic monitoring is inhibited.
○ All necessary equipment should be located near the endoscopy unit so that it can
be reached quickly if needed during the endoscopy. The items should be
meticulously organized.
○ Ensure the room is well-stocked before the procedure (checklists and procedure
cards will help this). All anesthetic equipment should be ready before the
procedure begins.
➢ Accessory instruments
○ THe most basic instruments include biopsy forceps, foreign body removal
forceps and a cytology brush.
○ Before each endoscopy exam, be sure each piece of ancillary equipment is
functioning and anticipate the need for these pieces of equipment.
○ They should be cleaned as soon as possible after use or the jaws of the forceps
may become locked in the closed position.
○ Endoscopes must be handled carefully during cleaning and never placed where
they could fall or be bumped.
○ Gather the following supplies: latex gloves, cleaning solution , two large basins
(for cleaning solution and distilled water), distilled water, Methyl alcohol, lint-free
gauze pads, cotton-tipped applicators, and channel-cleaning brush.
○ Biopsy instruments should be immersed in soapy water, brushed carefully with a
cleaning brush, and then rinsed.
➢ Storing an endoscope
➢ Gastrointestinal endoscopy
➢ Colonoscopy
➢ Bronchoscopy
➢ Computed tomography
➢ Nuclear medicine