ML3B Hardware Manual_MS2_040707_e_print
ML3B Hardware Manual_MS2_040707_e_print
Operation, Calibration,
Maintenance
Dear User
To help us provide the product and service excellence you expect now and in the
future, CORTEX Biophysik welcomes your comments and suggestions.
Sincerely
Nonnenstrasse 39
04229 Leipzig
Germany
Phone +49 (0)341 48 74 9-0
Fax +49 (0)341 48 74 9-50
email: [email protected]
http://www.cortex-medical.de
Table of Contents
Chapter Page
I. Before you start 5-7
§ Introduction 5
§ Intended use 5
§ Safety regulations 6
§ Interference with other devices 7
P re p a r e p at ie n t /t e st sub je ct f o r t he t e st
®
§ Fix Polar heart rate transmitter belt 19
§ Select mask and fix head cap 20
§ Connect volume transducer/sample line to mask 20
§ Fix and adjust mask 21
§ Ready for the test 22
Table of Contents
Chapter Page
VII. Maintenance & Cleaning 50-64
§ Cleaning & Maintenance Routines Overview 50
§ Cleaning & Maintenance Instructions 52
- Replace O2 Analyzer 52
- Clean & disinfect face mask 57
- Clean head cap assembly 58
- Clean housing 58
- Clean & disinfect DVT volume transducer 59
®
- Clean & disinfect Triple V volume transducer 61
- Replace sample line 63
- Visual inspection of system & parts prior to next use 63
§ Maintenance of Calibration Equipment 64
- Replace reference gas bottles 64
- Clean & replace tube set 64
- Clean gas saver 64
Please read the following instructions and safety regulations carefully before first use to obtain the
®
best performance from your MetaLyzer 3B. Keep these instructions in a safe place for future
reference.
Intended use
®
The MetaLyzer 3B is a stationary metabolic stress test system (or CPX system) for pulmonary gas
exchange measurements during ergometer or treadmill exercise. Direct measures of gas exchange
include O2 and CO2 concentration of the expired / inspired air, heart rate, ventilation, ambient
temperature and pressure. In addition, other valuable physiological variables can be obtained (e.g.
AT, VO2max), allowing the physician to completely assess a patient’s or an athlete’s
cardiopulmonary status. This device may be used with adults and children over the age of 14
years.
®
During a CPX test with MetaLyzer 3B, the patient / test subject is wearing a small face mask,
breathing out through a volume transducer fixed to the face mask. The gas samples are collected
while the patient / test subject exercises on an ergometer or treadmill. The optional 3-channel, 5-
lead ECG enables the physician to record the ECG of his/her patient / test subject and to monitor
his / her heart function throughout the test. Selected ECG sequences can be saved for online
display and / or a detailed analysis after the test (Note: The 3-channel ECG option is supported by
MS 2 version from 2.4). In addition, other external diagnostic devices (e.g. ergometers) can be
®
connected and directly controlled via the MetaSoft software.
®
As a breath-by-breath system the MetaLyzer 3B measures volume continuously and
simultaneously determines expired CO2 and O2 concentration. The CO2 output and O2 uptake
during each breath are calculated, and data sampled is transferred breath-by-breath to a PC for
immediate display. Breath-by-Breath systems allow very accurate measurements under non-
steady-state and changing ambient air conditions. They are considered appropriate for extended
CPX testing requirements aimed to accurately follow up the physiological responses during
exercise tests in which work rates change more frequently.
®
The MetaLyzer 3B is operated via the mains supply. It is directly connected to a PC for real-time
®
display of the data recorded during measurement, using MetaSoft software. The data recorded is
®
saved in the internal database of MetaSoft for a precise performance analysis after the test.
Neither CORTEX Biophysik nor its sales partners assume any responsibility for the final use of its
equipment.
For further information on metabolic stress testing please contact your local CORTEX sales
partner.
Note: Federal law restricts this device to sale by or on the order of a physician.
Safety regulations
®
For safety reasons, the MetaLyzer 3B must not be operated or used with accessories,
disposables, optional modules or analyzers other than offered and / or authorized by the
®
manufacturer. Ergometers and / or treadmills which are to be controlled via MetaSoft must meet
the legal requirements for medical equipment and may only be used with the appropriate
® ®
MetaLyzer 3B option. Computers which are connected to the MetaLyzer 3B, but were not
supplied by CORTEX Biophysik, must comply with the technical standards for electronic
equipment. They must meet the minimum requirements recommended by CORTEX Biophysik for
® ®
use with MetaSoft CPX testing software to ensure a proper function of the MetaLyzer 3B.
®
The operation of the MetaLyzer 3B in combination with other devices (PC, ergometer) must
comply with the requirements of DIN EN 60601-1-1 / DIN EN 60601-1-1/A1.
CORTEX SmartCart trolleys are designed for exclusive use with CORTEX systems. They can only
be equipped with the appropriate CORTEX devices. Do not add trays or equipment other than
offered and / or authorized by CORTEX for the respective SmartCart model. Please note that
®
Cortex devices (CardioLyzer, MetaLyzer ) are connected to a patient / test subject. For patient
safety, PCs, printers and other devices which do not comply with EN 60601-1, must be connected
to an isolating transformer if they are operated within a patient environment, i.e. at a distance of
less than 150 cm from a patient / test subject.
Always ensure the power outlet is secured. Devices connected must meet the legal requirements
according to DIN EN 60601-1-1 / DIN EN 60601-1-1/A1.
Do not use or connect any defective, broken or non-genuine parts and / or accessories.
Defective parts and / or accessories must be replaced prior to use. If in doubt, always
contact an authorized CORTEX sales partner or CORTEX Biophysik to check the system
before use.
The system may only be opened to exchange the O2 analyzer. Prior to opening the system,
®
make sure your MetaLyzer 3B is switched off and disconnected from the mains supply.
For repairs return the system to your local CORTEX sales partner or to CORTEX Biophysik. The
®
MetaLyzer 3B and / or its parts may only be repaired by an authorized CORTEX sales partner or
by CORTEX technical support staff. Repairs undertaken by non-authorized personnel may
cause severe risk for the user and patient / test subject. The CORTEX warranty expires in
case of non-authorized repair.
® ®
Do not use a MetaLyzer 3B with an open or defective housing. The MetaLyzer 3B must not
be used with anaesthesia gas in combination with oxygen or nitrous oxide (it is not a device of
class AP or APG). Always keep the device in a dry and protected position. Never operate the
devices in or close to water (DIN VDE 0100, Part 702).
Neither CORTEX Biophysik nor its sales partners assume any liability for defects, damages or
®
any injuries caused due to improper use of the MetaLyzer 3B or due to a violence of its safety
regulations and maintenance instructions.
Gas under pressure involves certain risks for user and equipment. The calibration procedures
and routines described in this manual should therefore be performed by properly trained or
experienced staff only.
®
During intended use of the MetaLyzer 3B, no waste is produced or compounds are generated
which need to be disposed. At the end of its life cycle, the device and its accessories should be
disposed at a service center specialized in electronic scrap disposal.
For cleaning and maintenance instructions please see chapter “Maintenance & Cleaning” to
®
ensure a proper and long-time performance of your MetaLyzer 3B.
Important
§ Always use original parts offered or sold by the manufacturer or
specified as equivalent by the manufacturer.
§ Never use defective or broken parts. Visually check each part of the
equipment prior to use. Replace defective parts, accessories and/or
power supplies immediately. If in doubt, always contact an authorized
CORTEX Biophysik sales partner or CORTEX Biophysik.
Parameters MetaSoft®
The following parameters are measured and / or calculated during a measurement with
® ®
MetaLyzer 3B using MetaSoft CPX Testing software:
S = Standard
Parameter O = Option
Circulation
- HR (Heart Rate) S
- HRmax (Maximum Heart Rate) S
- Blood Pressure (systolic, diastolic, mmHg) O
- Workload:
Level (rest, load, recovery), work rate (watts), grade (watts/ min., S
grade/min.), speed (m/sec., km/h, m/h), elevation
- 3-channel ECG (5, 10,20 mm/mV, 15, 25, 75 mm/sec., Einthoven I-
III, Wilson V, Nehb, Goldberger aVR, aVF, aVL, ST segment O
elevation)
- 12-channel ECG (Einthoven, Wilson, Goldberger) O
Gas Exchange
- FIO2, FEO2 (O2 Concentration in inspired / expired gas, Vol. %) S
- FICO2, FECO2 (CO2 Concentration in inspired / expired gas, Vol. %) S
- VO2 BTPS (Oxygen Uptake BTPS, l/min) S
- VCO2 BTPS (Carbon Dioxide Output BTPS, l/min) S
- VO2 STPD (Absolute O2 Uptake STPD, l/min) S
- VCO2 STPD (Absolute CO2 Output STPD, l/min) S
- VO2/kg (Relative O2 Uptake, l/min/kg) S
- VCO2/kg (Relative CO2 Output, l/min/kg) S
- VO2max (Maximum O2 Uptake, l/min/kg) S
- METS (Metabolic Equivalent) S
- RQ (Respiratory Exchange Ratio VCO2/VO2) S
- AT (Anaerobic Threshold) S
- VO2/HR (Oxygen Pulse) S
- VO2/HRmax (Maximum Oxygen Pulse) S
- VO2 / WR (Aerobic Capacity, l/min./W) S
- O2 kinetics (Oxygen Deficit/redemption, l/min.) S
Ventilation
- RR (Respiratory Frequency or Rate, Breath/min.) S
- VT (Tidal Volume, l/Breath) S
- VE ATPS (Minute Ventilation, l/min.) S
- VE / VO2 (Ventilatory Equivalent for O2) S
- VE / VCO2 (Ventilatory Equivalent for CO2) S
- PETO2 (End-tidal O2 Partial Pressure, mmHg) S
- PETCO2 (End-tidal CO2 Partial Pressure, mmHg) S
- PACO2 (Arterial CO2 Partial Pressure, mmHg) S
Technical Specifications
Analyzers
Volume transducer Type: DVT turbine, digital
Range: 0.1 – 12 l/s
Resolution: 7ml
Accuracy: 2%
®
Type: Triple V turbine, digital
Range: 0.05 – 20 l/s
Resolution: 7 ml
Accuracy: 2%
Resistance: ≤ 0.1 kPA/l/s at 16 l/s
Safety
Safety Standards EN 60601-1 (1996)
EN 60601-1-1 (1994)
EN 60601-1-1/A1 (1996)
Class Class II, Type BF
Protection Class IP 30
EMC EN 60601-1-2 (1994)
Conformity Class II a
(according to 93/42/EWG; Appendix IX)
Certified Quality
®
The MetaLyzer 3B complies with the Medical Device Directives, MDD
0124 93/42/EEC, and has been approved to carry the CE Mark as shown.
Cable/Tube Connectors:
Socket Function
gas Socket to connect sample line
volume Socket to connect volume transducer cable
ecg Socket to connect ECG patient cable
®
hr Socket to connect Polar heart rate receiver cable
digital Digital input/output to connect external devices
(e.g. CBP 2000 Blood Pressure Module)
on / off On/off switch
Sockets
Socket Function
12-15V / 1.5A DC To connect external medical power supply.
IV. Installation
Connect device to
PC/Notebook
®
The MetaLyzer 3B is operated using
®
MetaSoft software.
®
To connect the MetaLyzer 3B base system
to your PC/notebook, use the PC connection
cable supplied.
Serial plug-in to be
inserted into PC/notebook
PC connection cable
Black cable connector to be inserted
®
into MetaLyzer 3B base system
DVT
volume transducer with
cable
®
Triple V
volume transducer with
cable
Sample line
®
Fix the POLAR receiver to the
ergometer/treadmill.
Connect bicycle
ergometer or treadmill
®
The MetaSoft exercise device control
allows an automatic and/or manual control of
a broad range of bicycle ergometers/
treadmills.
Power system
®
1. The MetaLyzer 3B is operated
exclusively through the medical power
supply.
®
2. Switch on your MetaLyzer 3B by
pressing the on / off button on the front
panel of the base system.
Prepare System
Power system for warm-up
®
Switch on your MetaLyzer 3B by pressing the on / off
button on the front panel of the base system.
Prepare patient/test
subject for the test
Fix Polar® HR Transmitter
Belt
®
Fix the POLAR transmitter belt to the
thorax of your patient/test subject, using the
elastic strap included in your basic supply
package.
®
Fig: Polar transmitter belt fixed to patient’s
thorax.
Head cap
Select mask
and fix head cap
Mask
®
Fig. Triple V volume transducer and sample line
connected to face mask.
Important
®
Perform your test using MetaSoft .
For further instructions refer to your
®
MetaSoft Operator’s Manual.
Important!
VI. Calibration
General
®
To maintain the highest possible accuracy, periodic calibration of the MetaLyzer 3B analyzers
with certified calibration equipment and in accordance with the instructions described in this
manual is strongly recommended.
Depending on the prevailing altitude level of your area or the operating conditions in which you
®
intend to use your MetaLyzer 3B upon installation, it may be necessary to recalibrate its gas
®
analyzers prior to first use. It should be noted that the MetaLyzer 3B is calibrated based on a
barometric pressure level of approximately 1000 mbar when delivered. Should the mean level of
barometric pressure in your area be lower than 900 mbar, a gas and volume calibration should
®
be performed prior to first use of the MetaLyzer 3B system.
Calibration Procedures
The following calibration procedures are available:
Important
®
To ensure the highest possible accuracy, the MetaLyzer 3B analyzers
should be calibrated under conditions (temperature, pressure, humidity)
similar to the environmental conditions in which the system is operated.
Example:
®
If the MetaLyzer 3B is usually operated at normal barometric pressure, it
is strongly recommended to recalibrate its analyzers under the new
conditions if a test is to be performed at high altitude.
Calibration Practices
®
If the MetaLyzer 3B is operated in stable environmental conditions, i.e. if it is used in the same
environmental conditions or if environmental conditions in which it is used, do not change significantly
between the measurements, the following calibration practices are recommended to verify and maintain
accuracy of its analyzers:
Gas analyzers
A two-point gas calibration should be executed prior to a series of coherent tests, after oxygen
sensor exchange or at least every 30 days. Please note that an ambient air measurement must
be performed prior to each measurement/test. For instructions on how to perform this please refer
to the chapter „Perform ambient air measurement“ of this user manual.
Volume transducer
The volume transducer should be calibrated at least once a day before the first test.
Pressure analyzer
It is not necessary to recalibrate the pressure analyzer unless barometric pressure changes
significantly. It is recommended, however, to check the barometric pressure measured by
MetaLyzer from time to time (e.g. every six months) against a reference barometer. If the values
measured differ more than 10 mbar, the pressure analyzer should be recalibrated.
Important
Parts List
The calibration procedures and routines described in this manual are based on CORTEX calibration
equipment which is optionally available and can be purchased from your CORTEX Biophysik sales partner
or CORTEX Biophysik. Always use original parts offered or sold by CORTEX Biophysik or parts specified
as equivalent by CORTEX Biophysik.
Digital Barometer
Check connections
Make sure the
· volume transducer cable
· sample line
are plugged in firmly prior to test start.
Volume Calibration
Equipment required:
§ Manual Calibration Syringe (included in CORTEX Calibration Kit or equivalent syringe) or certified
automatic syringe
Calibration Steps:
®
4. Start MetaSoft .
Note:
The volume (flow) sensor should be
calibrated at the beginning of each
day.
7.
a. Select the appropriate calibration
syringe from Syringe selection
box.
Note:
The volume of the calibration
syringe used has to be within the
upper ventilation range expected
to be measured during tests with
®
MetaLyzer 3B.
8.
a. Select the appropriate volume
(flow) sensor from Volume Sensor
selection box.
Note:
This setting allows you to handle
several volume (flow) sensors. If
you have only one volume (flow)
sensor, select Standard from the
Volume Sensor drop-down
selection box.
9. Press the
Note:
Make sure that the sample line outlet
of your volume (flow) sensor is
closed with the black plug-in and the
volume (flow) sensor is connected
both to your syringe and device.
Important!
Equipment required:
§ Gas 1 (clean ambient air or span gas with 0% CO2, approx. 21% O2 , bal. in N2)
§ Calibration/Reference gas
(Reference/calibration gas from CORTEX Calibration Kit or equivalent span gas with
4-6% CO2, 14-16% O2, bal. in N2)
§ Gas Saver with power supply or 9V battery (from CORTEX Calibration Kit)
§ Set of various connection tubes to gas bottle and sample line
(CORTEX Calibration Kit or equivalent tubes with connections)
Calibration Steps:
Note:
The sample line is not yet connected to gas saver.
®
4. Start MetaSoft and select Manual
Gas Sensor Calibration Method.
Note:
A two-point reference gas calibration
should be executed prior to a test
series or at least every 30 days. It
must be executed if the oxygen
analyzer has been exchanged.
7.
a. Select the appropriate gas
concentrations for Gas 1
(ambient air or span gas) and
Gas 2 (calibration/reference
gas from CORTEX Calibration
Kit or see manufacturer label
of calibration gas bottle) from
the “Gas 1” and “Gas 2”
selection boxes in the gas
sensor calibration screen.
8. Press the
Note:
The sample line still remains
unconnected to gas saver.
Note:
Please close the reduction valve of the
calibration/reference gas bottle and
disconnect the sample line from gas
saver.
Equipment required:
§ Gas 1 (clean ambient air or span gas with 0% CO2, approx. 21% O2 , bal. in N2)
§ Calibration/Reference gas
(Reference/calibration gas from CORTEX Calibration Kit or equivalent span gas with
4-6% CO2, 14-16% O2, bal. in N2)
§ CORTEX Automatic Gas Calibrator with power supply (from CORTEX Calibration Kit)
§ Set of various connection tubes to gas bottle(s) and sample line (CORTEX Calibration Kit)
§ Connection Cable PC - Automatic Gas Calibrator (CORTEX Calibration Kit)
Calibration Steps:
Note: The gas calibrator can also be operated on either regular or rechargeable batteries. Use of
alkaline batteries (AA / LR6 size) or rechargeable batteries (AA size) with a capacity of at least
1,500 mAh is recommended. The batteries compartment is located on the bottom side of the unit.
Nearly exhausted batteries are indicated by a flashing On / LowBat LED. In this case you will be
able to complete the calibration process before exchanging the batteries.
Notes:
The Automatic Gas Calibrator allows you to perform the calibration procedure without needing to
disconnect or unplug any connections even when using two gas bottles with different calibration
gases.
If you are calibrating using ambient air and one calibration gas (e.g. from CORTEX calibration kit),
the calibration gas bottle has to be connected to the GAS 2 inlet. The GAS 1 inlet is unused in that
case.
If you are using a second calibration gas instead of ambient air, please connect this bottle to the
GAS 1 inlet.
The following steps refer to a calibration process using ambient air and one calibration gas (from
CORTEX calibration kit).
4. Check if the bottle valve (black knob) on the pressure reduction valve is closed.
To set the proper output pressure turn the pressure knob towards higher pressure (clockwise in '+'
direction) until you reach the mechanical stop. Now screw it back out again ('-' direction) by 2.5 turns.
Remove the protective cap from the gas bottle and screw the reduction valve in.
Note: If your reduction valve is equipped with a manometer please adjust the pressure to a value of
14 - 29 PSI (1 - 2 Bar). Avoid pressure settings higher than 29 PSI (2 Bar).
5. Connect the GAS 2 inlet at the automatic gas calibrator and the gas bottle using the appropriate tube
from your calibration kit.
Plug the end of the sample line into the short tube which is already connected to the calibrator
(Sample Line socket).
Fig.
Calibration equipment and test setup
for performing an automatic gas
calibration.
®
8. Start MetaSoft and select General
from the Configuration Folder, then
click on the General Settings icon.
Note:
A two-point reference gas calibration
should be executed prior to a test
series or at least every 30 days. It
must be executed if the oxygen
analyzer has been exchanged.
Notes:
Please close the gas bottle(s) after finishing the calibration process since a complete pressure
tightness of the calibrator's internal pneumatic system cannot be guaranteed.
If you won't be using the automatic gas calibrator for a long time, please remove the batteries.
Pressure Calibration
Equipment required:
§ Reference barometer
Calibration Steps:
Attention:
The barometric pressure must be
entered in mbar. Change the
barometric pressure measured in
mmHg to mbar using the following
formula:
P[mbar] = 1.33322 * P[mmHg]
5. Press the
Important
button.
The maintenance and cleaning instructions described in this manual are intended to provide
®
guidance on routine maintenance of the MetaLyzer 3B and its accessories to satisfy your
internal quality assurance and as a matter of good hygienic practice.
®
To maintain your MetaLyzer 3B at peak performance it is strongly recommended to observe
these maintenance and cleaning instructions. Non-observance of these instructions may result in
®
a failure to properly operate your MetaLyzer 3B system and/or in a severe safety risk for the
user. The warranty expires in case of non-observance of the maintenance and cleaning
instructions.
Please contact your local CORTEX Biophysik sales partner or CORTEX Biophysik for a
preventive maintenance agreement.
Replacement of O2 Analyzer
The oxygen analyzer (sensor) is an electro-chemical cell (fuel cell) which is used up after 18
®
months of use. It should therefore be replaced after 18 months. The MetaLyzer 3B base system
®
integrates an audio/visual self-test routine. A message will appear on your MetaSoft screen prior
to measurement in case the oxygen analyzer is used up and needs to be replaced.
The O2 analyzer should be replaced by your local CORTEX sales partner or properly trained or
experienced technical support staff only. To replace the analyzer, refer to the instructions
included in the next chapter of this manual.
Replacement parts and consumables can be purchased from your local CORTEX sales partner
or CORTEX Biophysik GmbH.
Preventive Maintenance
®
To maintain a long-time performance of your MetaLyzer 3B, the system should be thoroughly
checked for wear or damage of critical parts (e.g. filter, pneumatic components) every twelve
months.
Return the system to your local CORTEX sales partner or CORTEX Biophysik for a preventive
maintenance inspection, or contact your CORTEX sales partner or CORTEX Biophysik for a
maintenance agreement.
Replace O2 Analyzer
®
This section describes how to replace the O2 analyzer of your MetaLyzer 3B stationary cardiopulmonary
exercise system. Remember to replace the analyzer after 18 months of use. The below step-by-step
instructions make it easy for you to remove and replace the O2 analyzer as often as necessary. To ensure
®
utmost operator safety and to maintain a long-time performance of your MetaLyzer 3B system, however,
service procedures should be performed by properly trained personnel only.
For further assistance please contact your local CORTEX sales partner or technical support of CORTEX
Biophysik.
Equipment required:
®
§ MetaLyzer 3B O2 analyzer, gas-tight sealed (CORTEX Part No. 000-45-020)
§ Screw driver
Steps:
Turn off and unplug your device.
(Fig. 1)
Remove cover.
(Fig. 2)
O2 analyzer
(Fig.4)
(Fig.5)
(Fig.6)
(Fig.7)
Cable connection
Tube connections
(Fig.8)
(Fig.9)
(Fig.10)
(Fig.11)
(Fig.12)
(Fig.13)
Note: Protective attire is required of personnel handling contaminated devices. Manual cleaning must be
done in a manner that protects personnel handling the devices from aerosolization and splashing of
infectious material.
Cleaning
§ After the test, remove the volume transducer and the head cap from the face mask. Soil is wiped
from the device surface with a moist sponge or cloth.
§ Inspect the mask and its components for damages at all stages of handling. If damage is detected
on any of the components, it should be identified and replaced.
§ Soak or rinse the device in water of 20° – 45°C. Use a pH neutral mild detergent. If an enzyme
product is required, soak for two minutes. Remove, examine and extend the soak time for
components with dried-on matter for up to five minutes. Soaking of components over five minutes
may be detrimental, causing damage to the component’s surface.
§ After soaking, the mask must be thoroughly rinsed with clean water to remove the detergent
residuals and debris from the components.
§ Dry mask thoroughly using a soft clean cloth or disposable paper towels.
The head cap assembly should be removed and cleaned after each patient. The cleaning of reusable
devices usually begins soon after use.
§ The head cap assembly, including the clips, can be hand-washed in water of 20° – 45°C, using a non-
oily, mild detergent.
§ Rinse thoroughly in water. Machine or line dry. Do not iron the head cap.
§ Inspect the head cap for damages at all stages of handling. If damage is detected on any of its
components (e.g. the clips), it should be identified and replaced.
Clean housing
®
Before cleaning the MetaLyzer 3B base system disconnect any tubes and/or cables from the system.
Use a moist cloth to wipe the instrument clean. Liquids must not be allowed to enter the device. Any
hospital-grade cleaning agent and disinfectant may be used.
Danger
Turbine
Electronic assembly
§ After disassembly of the volume transducer check its parts for mechanical damages (turbine,
sieve, housing of electronic assembly, cable connectors).
§ Rinse turbine in warm water (< 50°C). Do not hold the turbine under running water. This
may damage the turbine.
§ Put turbine into a disinfecting solution (see disinfectants recommended for face mask).
§ After disinfecting, first rinse the turbine in pure water, then in distilled water to avoid
calcification after air drying. Air dry the turbine. Do not use mechanical dryers, paper or
textile for drying to keep the turbine free from dust and fibre deposits.
§ The housing of the electronic assembly may be surface-cleaned, using a fluff-free, weakly
moistened cloth. Moisture must not get into the housing.
§ To reassemble the volume transducer, follow the opposite instructions by pushing turbine back
into electronic assembly. The parts must be thoroughly dried before assembling.
®
The Triple volume transducer must be cleaned and disinfected after each patient. This means one test
can be performed while the contaminated parts are being cleaned. The standard package includes one
volume transducer; exchange units are available on order from your local CORTEX sales partner or the
manufacturer.
®
The Triple V volume transducer consists
of the following parts:
Housing
- housing
- turbine
- electronic assembly with cable
Turbine
Electronic assembly
®
Disassembly of Triple V volume
transducer:
Cleaning
1. Turbine:
§ Rinse the turbine/transducer in warm water (< 50°C). Do not hold the turbine/transducer under
running water. This may damage the turbine/transducer.
§ Put the turbine/transducer into a disinfection solution (see disinfectants recommended for face
mask).
§ After disinfection first rinse the turbine/transducer in pure water, then in distilled water to avoid
calcification after air drying. Air-dry the turbine/transducer. Do not use mechanical dryers,
paper or textile for drying to keep the turbine free from dust and fibre deposits.
2. Electronic Assembly:
The housing must be surface-cleaned, using a fluff-free, weakly moistened cloth only. Moisture must
not get into the housing.
3. Housing:
The housing can be easily cleaned using a non-oily, mild soap. Rinse thoroughly in running water.
Inspect the tube set at all stages of handling. Should you detect damages or breakages, replace the
defective tube immediately. It is recommended to replace the entire set after 12 months’ of use.
Replacement tube sets can be purchased from your local CORTEX Biophysik sales partner or CORTEX
Biophysik.
VIII. Annexes
Warning: Do not try to repair defective parts yourself or to open the system as this would
invalidate your warranty!
Operation
No power. - Power supply is not - Connect power supply
connected
Visual error messages
The sample line is blocked. - Sample line obstructed. - Remove obstruction (e.g. kink).
- Filter obstructed. - Replace sample line.
- Internal tubes obstructed. - Contact your local sales partner.
Please check calibration of - Calibration values got lost. - Recalibrate CO2 sensor.
CO2 sensor.
- If the error message still appears
after recalibrating the sensor, a
memory error may have occurred.
Contact your local sales partner or
CORTEX Biophysik to arrange for
a return of the system.
Please check calibration of O2 - Calibration values got lost. - Recalibrate O2 sensor.
sensor.
- If the error message still appears
after recalibrating the sensor, a
memory error may have occurred.
Contact your local sales partner or
CORTEX Biophysik to arrange for
a return.
Please check calibration of - Calibration values got lost. - Recalibrate volume transducer.
volume transducer.
- If the error message still appears
after recalibrating the sensor, a
memory error may have occurred.
Contact your local sales partner or
CORTEX Biophysik to arrange for
a return.
Please check calibration of - Calibration values got lost. - Recalibrate pressure sensor.
pressure sensor.
- If the error message still appears
after recalibrating the sensor, a
memory error may have occurred.
Contact your local sales partner or
CORTEX Biophysik to arrange for
a return of the system.
Via email
X. Trademarks
XI. Disclaimer
THE INFORMATION IN THIS GUIDE IS SUBJECT TO CHANGE AT ANY TIME AND WITHOUT
NOTICE.
CORTEX Biophysik GmbH reserves the right to make changes in its designs and specifications at
any time and without prior notice. The information provided by this document is believed to be
accurate and reliable. However, no responsibility is assumed by CORTEX Biophysik GmbH for its
use; nor for any infringements of patents or other rights of third parties resulting from its use.
Photos and images for representation purposes only.
©
CORTEX Biophysik GmbH 1993 – 2004. All rights reserved.
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
.....................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................
......................................................................................................................................