Walk T1 -User Manual
Walk T1 -User Manual
BPL Medical
11th KM, Bannerghatta Road, Arakere, Bengaluru, INDIA 560076.
This manual is for the WALK T1, Ambulatory Blood Pressure Monitor (ABPM) System.
CAUTION: US Federal law restricts this device to sale by or on the order of a licensed practitioner
Copyright Information
All content in this manual is the proprietary information of BPL Medical and is provided solely for purposes
of operation, maintenance or service of the WALK T1 ABPM System. This manual and the WALK T1 ABPM
System described in it are protected under copyright law under which they may not be copied, in whole or in
part, without written consent of BPL Medical. BPL Medical and WALK T1 are registered trademarks of BPL
Medical, Inc. All other trademark names are the trademarks of their respective holders. The information in
this manual is furnished for guidance only, is subject to change without notice, and should not be construed
as a commitment by BPL Medical. BPL Medical assumes no liability for errors or inaccuracies that may
appear in this manual. © 2019 BPL Medical. All rights reserved.
Manufacturer Information
SunTech Medical, Inc. EMERGO Europe SunTech Medical (Shenzhen) Co., Ltd.
507 Airport Blvd, #117 Prinsessegracht 20 105 HuanGuan South Road, Suite 15
Morrisville, NC 27560-8200 2514 AP The Hague 2~3/F
Phone: 1-919-654-2300 The Netherlands DaHe Community Guanlan,
1-800-421-8626 LongHua District, Shenzhen
Fax: 1-919-654-2301 GuangDong PRC 518110
Tel: + 86-755-29588810
+ 86-755-29588986 (Sales)
+ 86-755-29588665 (Service)
Fax: + 86-755-29588829
The data obtained from ambulatory blood pressure monitors is accurate and useful for managing a wide
variety of hypertensive situations including:
• White-coat hypertension
• Resistant hypertension
• Masked hypertension
• Childhood hypertension
• Efficacy of anti-hypertensive drug therapy on a 24-hour basis
• Nocturnal hypertension
• Episodic hypertension and/or anxiety disorders
• Hypotensive symptoms
• Changes in diet and daily routine designed to reduce hypertension
USB Industry
This product and its shipping container should be kept dry ISO 7000-0626
For additional information on how these accessories fit together consult the WALK T1 Quick Start Guide (82-
0166-00).
The following safety and effectiveness issues are to be considered prior to the usage of the WALK T1
monitor.
NOTE: This device is defibrillator protected. No precautions specific to the WALK T1 are required during defibril-
lation, and defibrillation discharge has no effect on the WALK T1.
• The monitor is intended for use following consultation and instruction by a physician.
• The reliability of the device is dependent upon conformance with the operation and service
instructions, as detailed in this manual.
• This device has been designed for use on patients with normal sinus rhythms.
• The interpretation of blood pressure measurements should only be made by a physician. The
accuracy of any blood pressure recording may be affected by the position of the subject, his or her
physical condition, and use outside the operating instructions detailed in this manual.
• Safety and effectiveness of central blood pressure measurements on children under the age of 18
years of age, pregnant women and neonates have not been tested.
Disposal
This symbol indicates that the monitor contains materials which may be hazardous to human health. This
product complies with the WEEE Directive. Please return the WALK T1 monitor to BPL Medical for proper
disposal. Please dispose of other materials according to local regulations.
Allergic exanthema (symptomatic eruption) in the area of the cuff may result, including the formation of
urticaria (allergic reaction including raised edematous patches of skin or mucous membranes and intense
itching) caused by the fabric material of the cuff.
Petechia (a minute reddish or purplish spot containing blood that appears in the skin) formation or Rumple-
Leede phenomenon (multiple petechia) on the forearm following the application of the cuff, which may lead
to idiopathic thrombocytopenia (spontaneous persistent decrease in the number of platelets associated with
hemorrhagic conditions) or phlebitis (inflammation of a vein) may be observed.
This monitor is designed to perform in conformity with the description thereof contained in this operation
manual when operated, maintained and repaired in accordance with the instructions provided. The monitor
should not be modified in any way. Ensure pressure compatibility for all patients. If any abnormality occurs
in the monitor, suspend the operation immediately and disconnect it from the patient. If the monitor has
Warnings
The general warning sign indicates a potentially hazardous situation which could result in serious injury.
WARNING: Do not use in the presence of flammable anesthetics; this could cause an explosion. This device
is not suitable for use in an oxygen enriched environment.
WARNING: Do not immerse the monitor in any fluid, place fluids on top, or attempt to clean the monitor
with any liquid detergents, cleaning agents, or solvents. This may cause an electrical hazard. Do not use
the monitor if accidental wetting occurs; please return to BPL Medical (see Limited Warranty). Refer to
Maintaining and Cleaning the WALK T1 ABP System, for care instructions.
WARNING: Too frequent measurements can cause injury to the patient due to blood flow interference.
WARNING: The cuff should not be applied over a wound as this can cause further injury.
WARNING: The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double
mastectomy use the side of the least dominant arm.
WARNING: Pressurization of the cuff can temporarily cause loss of function of simultaneously used
monitoring equipment on the same limb.
WARNING: Do not use if device is dropped and/or is damaged. Have a qualified service representative check
the monitor before using again.
WARNING: Do not attach the cuff to a limb being used for IV infusions or any other intravascular access,
therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow, potentially
causing harm to the patient.
WARNING: Use only with the cuffs supplied by BPL Medical. Different cuffs have not been validated with
WALK T1 and measurements with non-validated components may not be accurate
WARNING: Use of an ACCESSORY, transducer or cable with ME EQUIPMENT and ME SYSTEMS other than
those specified may result in increased EMISSIONS or decreased IMMUNITY of ME EQUIPMENT or ME
SYSTEM.
WARNING: The WALK T1 may be interfered with by other equipment even if the other equipment complies
with CISPR Emission Requirements.
WARNING: Do not use the monitor during magnetic resonance imaging (MRI) or in an MRI environment.
The caution symbol indicates a potentially hazardous situation which may result in minor or moderate injury.
It may also be used to alert against unsafe practices.
CAUTION: When downloading data from the monitor’s communications USB port, the device should not be
in use with a patient.
CAUTION: Do not remove monitor covers, except to replace batteries. The monitor does not contain any user
serviceable components. Return monitor if service is required.
CAUTION: Do not use on neonates, pediatric patients less than 3 years old, or patients known to be readily
susceptible to bruising.
CAUTION: Do not use the monitor if it has failed its diagnostic self test, or if it displays a greater than zero
pressure with no cuff attached. The values displayed by such a monitor may be inaccurate.
CAUTION: Substitution of a component different from that supplied may result in measurement error.
Repairs should be undertaken only by personnel trained or authorized by BPL Medical.
CAUTION: The WALK T1 does not contain any user serviceable internal parts and should only be repaired by
an authorized BPL Medical service representative. Do not service the product while in use.
CAUTION: If cuff fails to deflate within two and a half minutes, instruct patient on manual removal of cuff.
CAUTION: Check that operation of the monitor does not result in prolonged impairment of the circulation of
the patient.
CAUTION: Remove batteries when device is not in use for long periods of time to prevent possible battery
leakage and product damage.
CAUTION: A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood
flow interference and potentially harmful injury to the patient.
CAUTION: Using an incorrect cuff size could result in erroneous and misleading blood pressure
measurement results.
Contraindications
The WALK T1 ABPM system should be used in conjunction with all other available medical histories and
diagnostic test information about the patient. The following are reasons to withhold use of the WALK T1
ABPM system from a patient:
CONTRAINDICATION: Do not use system on patients who have had a double mastectomy
CONTRAINDICATION: Do not use on the same arm of patients with a peripherally inserted central catheter
(PICC) line, Intraveneous (IV) or arterial line.
3 4
1. Start/Stop Button
To Power On:
Press the Start/Stop button.
To Power Off:
When not taking a reading, press and HOLD the Start/Stop button until you hear 8 beeps (approx. 5 seconds),
then release.
To Abort a Measurement:
Press the Start/Stop button any time during a reading.
To Start a Programmed ABP Study:
When time is flashing, press the Start/Stop button to take 1st reading.
To Start a Manual BP Reading:
When clock is displayed, press the Start/Stop button.
Display
Setting up the WALK T1 System involves powering the WALK T1, installing WALK PRO on a personal
computer (PC), and connecting the WALK T1 to the PC.
The WALK T1 ABPM system is packaged with everything you need to start. See Product and Accessories for
complete contents.
1. Incrementing dashes
2. Software and safety version of the monitor
3. Battery voltage followed by three quick audible beeps
4. The number of BP readings in memory followed by one long audible beep
5. Time flashing
NOTE: Ensure batteries are inserted with the correct polarity. Improper installation will prevent the monitor from
functioning. Batteries are required for WALK T1 use at all times including programming, scheduling and retriev-
ing.
CAUTION: Remove batteries when device is not in use for long periods of time to prevent possible battery
leakage and product damage.
NOTE: Install WALK PRO before connecting the USB cable to the computer.
1. Connect the USB cable to the USB connector on the ABP monitor (Figure 1; part A).
2. Connect the USB end of the cable to the USB port on your PC (Figure 1; part B).
A B
NOTE: The cable can be left connected when the PC is off and the monitor is not connected to a patient.
Toolbar Buttons
PDF: Saves the report for the open ABP study in PDF format
To create a template:
1. In the open Programming Monitor window, enter the desired parameter settings in the form. See
Parameter settings.
2. Click the Save Template button at the bottom of the Programming Monitor window.
3. In the Assign Template Name dialog box, type a name for the template and click Save.
To open a template:
1. From the open Programming Monitor window, click the Open Template button at the bottom of the
Programming Monitor window.
2. From the Available Templates dialog box, select the template name, and then click Open.
3. The template information will populate the corresponding information in the Programming Monitor
window.
To delete a template:
1. From the Programming Monitor window, click the Open Template button at the bottom of the
Programming Monitor window.
2. From the Available Templates dialog box, select the template name, and then click Delete.
3. At the prompt, click Yes to confirm deletion.
To email a template:
An email server must be specified before a Programming Template can be emailed. See Email Settings on
page 45 for setup instructions.
Standard Tab
UI Item Description
Patient ID Patient ID for reporting and referencing data.
Patient name Enter patient name (first, middle, last).
Select/Create Click to open the Patient Information window. Use this window to use
Patient existing patient information or to create a new patient entry.
Measurement Specifies when and how often the monitor takes readings. For Awake time
Schedule: and Sleep time, select from the Hour pull-down menu to establish the start
time for these periods. From the BP Interval pull-down menu, select the
desired interval between readings (none, 5, 10, 15, 20, 30, 45, 60, 90 or 120
minutes).
Start study in 5 Slider denotes that the study will start automatically after programming;
minutes: unchecked denotes that the study will be started with the first press of the
Start/Stop button when the monitor is powered ON.
Quality Control Shows the Quality Control tab that allows setting requirements for minimum
readings in a study to pass Quality Control
Advanced Tab
UI Item Description
Max Pressure Establishes the maximum inflation pressure for the monitor (options
between 160 and 280 mmHg). Suggested setting is 30 mmHg above the
highest expected systolic BP.
NOTE: The ABP monitor will not inflate to Max Pressure with each reading;
instead it inflates to 30 mmHg above the previous systolic reading.
Intervals Set interval type. Select Fixed to set the intervals to exact times. Select
Standard for +/- 5 minutes around the selected intervals.
Display results When on, allows the patient to view the results immediately after a
measurement.
NOTE: Display Results is always on for the first 30 minutes of study.
Quality Control Tab (if quality control is selected in the programming monitor window)
UI Item Description
QC Review Period: Select start and end times for a quality control review period for the study
program.
Minimum Reading Enter the requirements for a minimum percentage of scheduled readings
Requirements: captured during the defined QC period.
The minimum number of readings captured per hour during the defined QC
period (between 1 and 12).
The minimum number of hours to contain readings for the entire study.
NOTE: To access the Quality Control tab you must first turn on the “Quality
Control” dial at the bottom of the Program Monitor window.
To determine the correct cuff size for your patient, wrap the cuff around the patient’s upper arm without
sliding the arm through the sleeve. Use the color- coded RANGE indicator on the inside of the cuff and the
bold INDEX marker to check that the arm circumference falls within the cuff range. If the arm is within range,
this cuff size is correct for your patient. If the marker is outside the RANGE indicator, select a new cuff size as
indicated by color.
To apply the ABPM cuff, simply slide the sleeve up the patient’s arm, ensuring the color size indicator is at the
top of the cuff. The cuff should be midway between the elbow and shoulder. Be sure the ARTERY indicator
is over the patient’s brachial artery, between the bicep and tricep muscles. Wrap the cuff snugly around the
patient’s upper arm. There should be space for approximately 1 finger underneath the bottom of the cuff.
Connect the BP hose to the WALK T1 monitor and to the cuff by pushing the fittings together until it is locked
into place. The hose can be released by pulling the female fitting away from the male connection. Drape the
hose over the patient’s shoulder, behind the neck and across to the opposite side of the body.
4. Attach to patient
Insert the WALK T1 into its pouch with the display showing through the window. Attach the pouch to the
patient using the belt (Figure 7).
• When the pressure in the cuff increases, the patient must avoid excess movement during
measurements. Let the cuffed arm hang loosely, slightly away from the body with the middle of
the cuff at heart level. Avoid flexing the muscles or moving the hand and fingers of the cuffed
arm.
• The patient can stop a measurement in progress by pressing the Start/Stop button
momentarily.
• If the Manual Readings setting is on, the patient can start a measurement at any time by
pressing the Start/Stop button momentarily.
• The cuff should not be removed between BP measurements.
• Before sleeping, the patient should make sure that the hose is not kinked and will not become
kinked.
• The batteries can be replaced during a study without the data being lost or interrupting the
monitor’s program. Alternatively, the monitor can be turned off without losing its data.
• Instruct the patient on how and when to fill out the patient diary.
• If the Day/Night button is on, instruct the patient on how to set day and night modes.
• If Event Marking is on, instruct the patient on how and when to mark events.
• Ensure the patient knows how to care for the monitor. Keep the monitor dry and do not drop it.
• If the monitor or cuff causes extreme pain, or pain not normally associated with blood pressure
measurement, the patient should remove the cuff and turn off the monitor.
• The patient should not talk during BP measurements. The patient should be seated, standing
or lying down. If seated, the patient should have legs uncrossed, feet flat on the floor with back
and arms supported.
Replace the batteries (2 AA alkaline batteries) for every study with new batteries. Failure to do so may result
in incomplete 24-hour studies.
To record an event:
Press the Event button
The monitor sounds a long audible beep for confirmation, and "rcd : 01" appears on the display (Subsequent
recordings use 02, 03, and so on). The monitor can record up to 30 events. If the patient tries to record more
than 30 events, the monitor beeps four times, and "No rcd" appears on the display.
The monitor beeps and displays either the sun or moon icon depending on the mode that it was switched to.
If switching to Night mode, the monitor will display a moon icon.
If switching to Day mode, the monitor will display a sun icon.
When the patient returns, take the cuff, monitor, and belt off and download the captured data to WALK PRO
for review.
NOTE: A green checkmark in the results indicates that the study meets that requirement. A red x indicates that
the study failed to meet that requirement.
The BP file will now open automatically and the patient’s BP data is now saved to the WALK PRO patient
CAUTION: If you do not retrieve data from the monitor, this data will be lost when you program it for the next
study.
1. From the File menu, select Open. Alternatively, from the toolbar, click the BP Data button.
2. The Open ABP Study - Select Patient dialog box appears, allowing you to select the patient file to
open. Select the patient whose BP data you wish to review by clicking on the patient’s name.
3. Click OK.
4. If the patient has more than one ABP study, a second dialog box, Choose Date, appears. Select a
study date and click OK. The patient’s data appears in the display area. You can now review the BP
study or print a report.
Each study in a patient file contains data that is displayed in the views accessed by the tabs at the bottom
of the application. The study information bar located toward the top of the view identifies the patient name,
patient ID, patient age (at the time of the study) and the study date of the displayed file.
Tabs and their contents.
• ABP Data: ABP measurement data from the monitor and relevant graphs.
• Patient Info: Patient name, ID, contact information, and physical description.
• Physician Info: Patient history, clinical information, and interpretation.
• Statistics: Statistical analysis of the ABP study.
• Time-Slice Statistics: Statistical analysis of specific user-defined time windows.
• Hourly Averages: Analysis of ABP study data including study comparison.
• Summary: Interpretative summary settings and results for current study.
Overview Graph
The scale of the vertical axis represents blood pressure (mmHg) and heart rate (bpm). The horizontal axis
displays the time in clock hours. Clicking on any point, measurement or event, in the graph highlights the
corresponding row in the table.
There are three tabs to the right of the graph: Threshold, Time-Slice, and Legend. The Threshold tab allows
the user to change the Threshold settings on the graph. The Time-Slice tab allows users to define specific
time windows in which to perform a statistical analysis. The Legend tab provides details on the configuration
of colors for the Overview Graph.
Shading on the graph indicates various periods of the study. The default colors (see Display Settings) and
definitions are:
• Blue shading: Indicates the asleep period of the study.
• Yellow shading: Indicates the white coat period, which is the first hour of the study (appears
only if enabled).
• Light orange shading: Indicates the periods used to calculate Morning Surge (appears only if
enabled).
AASI Graph
The AASI Graph tab is optional, and can be enabled by selecting Ambulatory Arterial Stiffness Index on
the Report>Configuration>Data (see Customizing and Configuring WALK PRO). The graph on this tab
plots the diastolic on systolic BP values for each measurement captured in the study. Additionally, this
tab lists the calculated AASI value, diastolic as a function of systolic, systolic as a function of diastolic,
the correlation coefficient of the linear regression equation, and the coordinate of the average systolic
and diastolic. The calculated AASI value will also be added to the Overall results on the Statistics tab.
The monitor tags data with an asterisk(*) based on criteria used to determine the validity of the data. The
exclamation point (!) is a permanent tag and is used when an accurate reading cannot be determined by
the ABP monitor. This tag cannot be changed and associated data will not be used in data analysis. The
asterisk is a tag that you can edit. Tags numbered r01 to r30 indicate events entered by the patient, and
the comments can be changed. Data can be tagged or un-tagged based on a number of factors, including
patient history, patient diary information, or other factors.
There are two methods to remove or insert an asterisk (*) tag on a measurement.
NOTE: All changes made to the ABP study are saved immediately.
Entering Comments
Use the Comments column, the right-most column in the table, to keep track of patient activity during a BP
reading. While activity is only one of the many factors that can affect blood pressure, it can be helpful in
understanding a BP reading within the context of the study.
To assist the patient with tracking their activities, a patient diary template is available for printing from the
Page 27 of 68 80-0096-00-RevA 30-Apr-20
Download Library on the BPL Medical website.
You can add or delete entries appearing in the Comments pull-down list by going to the Diary Comments
section of the Configuration window. (For instructions on configuring Diary Comments, see Customizing and
Configuring WALK PRO.)
The selected symbol replaces the checkmark, and the comment appears in the table.
To define thresholds:
1. Open a study in the ABP Data tab. Within the Overview Graph tab, navigate to the Threshold tab
Thresholds
Thresholds can be set for Awake and Asleep systolic and diastolic BP. The default setting is a global
threshold set to 140/90 mmHg for Awake periods and 120/80 mmHg for Asleep periods. If desired, you can
set thresholds to match a published standard using the corresponding buttons: JNC7/AHA or ESH.
• JNC79 recommends 135/85 mmHg for Awake periods and 120/75 mmHg for Asleep periods.
• The American Heart Association (AHA)10 recommends a 24-hour average BP of 130/80 mmHg.
• The ESH11 recommends 135/85 mmHg for Awake periods and 120/70 mmHg for Asleep periods.
NOTE: The pediatric threshold is applied as a custom BP threshold, which applies to a single patient’s file.
To create a time-slice:
1. Open a study in the ABP Data tab. Within the Overview Graph tab, navigate to the Time-Slice tab
located to the right of the graph view. (Figure 16).
2. Click the New button. On the graph left-click and hold the mouse to select the starting time of the
time-slice period then drag the mouse to the end time. Releasing the mouse button determines the
ending time of the time slice. The selected time-slice will be highlighted on the Overview graph.
3. Once the time-slice is selected, a Time-Slice Settings box (Figure 17) will appear and allow the user to
name the time-slice and edit the start and end times, if necessary.
4. Click Save to create the new time-slice.
To edit a time-slice:
1. Open an ABP study and click the Time-Slice tab (Figure 16).
2. Choose the time-slice from the drop-down menu. The time-slice period will be highlighted on the
Overview graph.
3. Click Edit. The Time-Slice Settings box (Figure 17) will appear.
4. Edit the name or the start and end times.
5. Click Save to save the settings.
To delete a time-slice:
1. Display the ABP study. Choose the Time-Slice tab.
2. Choose a time-slice from the drop-down menu. The time-slice period will be highlighted on the ABP
data graph.
3. Choose Delete. Click Yes to delete the time-slice (Fig. 18).
Users are able to enter and edit Patient History, Reason for Test, Current Medications, and Physician
Interpretation. As an alternative to typing in the free text field for Current Medications, users may add
Medication, Dosage, and Frequency to dictionaries that are stored within WALK PRO.
Abbreviation Meaning
ac Before meals
bid Twice daily
pc After meals
prn As needed
qam Every day before noon
qd Every day
qh Every hour
qhs Every night at bedtime
Viewing Statistics
To view the statistical analysis for the displayed ABP study, click on the Statistics tab. This window initially
shows the following tables:
The Statistics tab can also show additional time periods or BP load charts.
NOTE: All data in these tables is included in exports, except white coat analysis values and BP load charts.
In the Statistics tab, each table includes the mean, standard deviation, coefficient of variation (CV), minimum
value, and maximum value for the following parameters: systolic, diastolic, MAP, pulse pressure.
The Statistics tab may include any of the following additional calculations if they are enabled under
Report>Configuration>Data>:
• Ambulatory Arterial Stiffness Index (AASI): AASI is calculated as one minus the regression slope of
the diastolic on systolic blood pressure. (The regression slope is calculated by plotting the diastolic
against the systolic values.) This index has been used to measure arterial stiffness and has been
shown to help predict target organ damage, cardiovascular mortality, and stroke.17
• Morning Surge: Morning Surge is calculated as the difference between the mean systolic BP during
the two hours after waking, minus the systolic BP during the Asleep hour containing the lowest BP
value measured. This calculation can help to identify cases in which the morning BP increase is
unusually high and presents a potential cardiovascular risk or a risk of stroke.14
• Coefficient of Variation: The Coefficient of Variation (CV) can be used as a measure of BP variability,
which can be a determinant of end-organ damage.18 CV is calculated as the standard deviation
divided by the mean blood pressure of a time period.
• Smoothness Index: Smoothness Index (SI) is the ratio between the average of the blood pressure
changes computed for each hour of the recording and its standard deviation. SI is generally used
to evaluate the effectiveness of antihypertensive treatment as it shows the averaged trend of blood
pressure reduction between two studies.19
• White Coat Analysis15: White coat syndrome refers to abnormally elevated blood pressure when the
patient is in a medical setting. If selected, this option does the following:
• Adds a white coat period to the statistical analysis.
• Displays the first hour of the study, or white coat period, with a light blue background in the
graphs.
• Evaluates for and reports on the presence of white coat syndrome on the interpretive summary,
when present.
• Mean Arterial Pressure (MAP) - This is the average pressure in an artery over the period of one heartbeat.
In the brachial artery, it is calculated by adding the diastolic to one-third of the difference between the
systolic and diastolic readings.
• Pulse Pressure (PP) - This is calculated by subtracting the diastolic from the systolic reading. It is another
hemodynamic parameter that may serve as an indicator for cardiovascular risk.
• Pressure Rate Product (PRP) - This is the product of the average systolic reading multiplied by the
average heart rate. PRP strongly correlates to a patient’s activity level and may be a key indicator of
cardiovascular risk. As typically reported in research and clinical applications, PRP data is divided by
1000.
Use the check boxes at the top of each column to select the statistics you would like to see displayed in the
graph at the bottom of the page. You can resize the table and zoom in on the graph with your mouse pointer.
(For instructions on resizing and zooming, see Viewing an Ambulatory Blood Pressure Study.)
To generate an interpretive summary for the displayed ABP study, click the Summary tab on the bottom of
the screen and select one of the following options for interpretation:
• Joint National Committee Seventh Report (JNC)7 and American Heart Association (AHA)8 guidelines.
• European Society for Hypertension (ESH)6.
• Pediatric – AHA12
• Pediatric – Soergel Tables11
Based on the calculated dip percentage, WALK PRO determines whether the patient is one of the following:
• Dipper (normal): A patient shows a decrease of 10% or more in systolic or diastolic blood pressure
values during sleep compared to readings taken when awake.
• Non-dipper (abnormal): A patient shows no decrease or less than a 10% decrease in systolic or
diastolic blood pressure values during sleep compared to readings taken when awake.
The statistics generated during the previous study are now incorporated into the Hourly Averages window.
The table expands to include the hourly average values for the reference study, and the differences between
the displayed and the reference studies.
On the graph at the bottom of the window, both studies’ data are displayed. The dotted lines represent data
from the reference study, and solid lines represent data from the displayed study. Select Difference and the
graph will display one line representing the difference between the two studies. The same navigation tools for
the Hourly Average and ABP Data windows also apply to the comparison window.
NOTE: All report pages include the BPL Medical logo, patient demographics (all information from the Patient
Info tab), and test date. The footer of each page identifies the type of monitor used and the page number.
You can choose from the following pages to customize your report (listed in the order the pages will print,
see Figure 26):
• Summary: The Interpretive Summary appears above the data graph and summary statistics.
Summary BP statistics for the overall, awake, asleep, and white coat (if chosen) periods include: mean
systolic and diastolic, mean heart rate, BP load, and asleep dip.
• Patient Information: Includes patient history, reason for test, patient medications, and the physician
interpretation.
• Statistics: Displays detailed statistics for BP parameters according to the time periods included in the
study. Overall, awake and asleep time period results will be displayed first, followed by the white coat
period (if enabled), and finally any additional time periods programmed by the user.
• Time-Slice Statistics: Displays statistics for BP parameters for each user defined time-slice period.
• Reviewed Data: Shows all the BP readings captured during the study in table format as it is shown
NOTE: You must have a printer installed. If you do not have a printer installed, see the online Help to install a
driver to print to a file.
Symbol Explanation
+ Signifies the reading was initiated manually by the patient pressing the Start/Stop key
on the monitor.
- Indicates that the pressure to which the cuff inflated in dynamic inflation mode may
have been too low to obtain an accurate systolic reading. Readings marked with “-” are
automatically omitted from the study. Compare the tagged reading with a “-” to the BP
readings above and below it. If the omitted reading seems to be acceptable, un-tag it
from the ABP Data tab; if it does not seem acceptable, leave it “tagged”.
r Indicates that this reading was a retry reading. Retries are readings that are
automatically obtained 4 minutes after a measurement attempt fails or a questionable
reading is obtained by the monitor.
! The exclamation point is a permanent tag and is used when an accurate reading
cannot be determined by the WALK T1.
> This symbol will appear to the left of the measurement if the reading results were
above the set threshold.
r01 to These numbered tags identify a series of patient-marked events which occurred during
r30 the study.
1. From the toolbar, click the BP Data button or from the File menu, select Open. The Open ABP Study –
Select Patient dialog box appears listing all available patient files.
2. Select the patient whose BP data you wish to retrieve by clicking on the patient’s name or use the
search button to locate the patient by any information relevant to the patient’s file. Click OK.
3. If the patient has more than one ABP study, a second dialog box, Choose Date, will be displayed.
Select a date and click OK.
4. The patient’s data appears in the Display Area. You can now review the BP data or print a report.
The exported files will be exported to the selected file location as either a *.ASC, *.GDT, *.XML or *.AWP4, file
depending on the format chosen.
If you are editing a configuration profile, a window opens to display a number of configuration options which
are described in detail on the following pages.
NOTE: On each tab, clicking a Default button returns the settings of that tab to the default factory settings
Hardware Settings
The Hardware Settings window (Figure 29) allows you to configure your computer to communicate with your
ABP monitor via a specified port. You may also test the settings that you have selected.
If you receive an error message, “Cannot communicate with ABP device,” one of the following applies:
• You have not selected the appropriate port or device.
• The device is not connected properly.
Data Settings
The Data Settings tab (Figure 30) allows you to configure the following settings:
Additional calculations: Select which additional parameters will be displayed when viewing a patient file.
Options include:
• AASI
• Smoothness Index
• Morning Surge
• White Coat Analysis
• Coefficient of Variation
Selecting any of these options will make them visible on corresponding tabs and graphs.
Display Settings
This feature allows you to customize the color scheme of the graphs and charts (Figure 31). Customizable
palettes allow you to fine-tune the color scheme for optimum reproduction by your printer. Shading
on the graph indicates the period of the study. The default colors (which can be changed using the
Configuration>Display tab) are the following:
Report Settings
This feature allows you to format your reports and to choose what information is included in them.
To format a report:
1. From the Configure menu, select Preferences. Select a profile and click Edit.
2. Click the Reports tab and then the Formats tab.
3. Select which items will be included in your One Page report, Full Report, and Standard Report. Select
or de-select the checkboxes on the right to make these changes (Figure 26).
Export Settings
This feature allows you to export patient studies as either ASCII, XML or GDT output (Figure 34), which is
useful when the data will be merged with a database. GDT is a format commonly used to share computer
medical records.
To change export settings, navigate to report>configure, then click on the Export tab within the Configuration
window.
AWP4 export:
AWP4 file output can be configured to use a standard encryption key (less secure) or custom encryption key
(more secure). Standard encryption files can be opened and imported by any WALK PRO installation without
a password. Using a custom encryption key requires the receiving party to enter the encryption key on study
import.
NOTE: if a custom encryption key is lost or forgotten, data in AWP4 files are not recoverable. The encryption key
should be kept in a secure location.
NOTE: Codes do not automatically indicate that a reading is invalid; they serve only as guides to help you review
the data.
1. Administrative: Full access to entire functionality of WALK PRO, including editing Admin Tool settings.
2. User: Full access to entire functionality of WALK PRO, except for editing Admin Tool settings.
When login security is enabled, you can also set the software to automatically logoff users after a specified
period of inactivity in WALK PRO.
Further, the user of the device bears sole responsibility for any malfunction that results from improper use,
faulty maintenance, improper repair, damage or alteration by anyone other than BPL Medical or authorized
service personnel.
While usage will have an impact, it is expected the monitor will be in service for 7 years. Typically, an
electromechanical pump determines the lifetime of the monitor. Service and support, including relevant
accessories, will be available up to 7 years following the last date this product is manufactured by BPL
Medical. For a list of service centers, see Section "19. Technical Assistance" on page 61.
The bladder needs to be inserted back into the cuff sleeve so the pneumatic hose portion of the bladder goes
Page 59 of 68 80-0096-00-RevA 30-Apr-20
through the tube opening on the cuff. Please note that the pneumatic hose connection should face upward
when using the Orbit ABPM cuff on either the right or the left arm.
The WALK T1 does not contain any user serviceable internal parts and should only be opened by
an authorized service representative. To return for service, please send to your nearest BPL Medical
office, listed above, care of Support and Service. Alternatively, please visit our website, https://www.
bplmedicaltechnologies.com/, to request more information.
The monitor’s calibration can now be checked against a calibrated mercury column.
1. Place a t-tube (part #98-0030-00) between the hoses connecting the monitor and the cuff.
2. Wrap the cuff around a suitably sized can or bottle. This acts as the reservoir for the unit.
3. Attach the third end of the “T” tube into a calibrated mercury column, which gives you access to the
bulb and a reference.
4. Using the bulb of the calibrated mercury column, inflate the cuff to 250 mmHg.
5. Once the pressure has stabilized at this level, the LCD should match the mercury column by ± 2.0
mmHg.
6. Check the unit against the column every 50 mmHg from 250 to 50 mmHg. The monitor should
be within ± 2.0 mmHg. If it is not, the monitor needs to be returned to the service department for
recalibration or repair.
NOTE: To return the WALK T1 to its normal operating mode, remove and re-insert one of the batteries.
BPL Medical, Inc. warrants each instrument to be free from defects in material and workmanship. Liability
under this warranty covers servicing of the instruments when returned from the customer’s facility prepaid
to the prospective factory depending on location. BPL Medical will repair any component(s) or part(s) that it
finds to be defective during the period of this limited warranty. Should a defect become apparent, the original
purchaser should notify BPL Medical of the suspected defect. The instrument should be carefully packaged
and shipped prepaid to the appropriate service center.
The instrument will be repaired in the shortest possible time and returned prepaid by the same shipping
method as received by the factory.
This limited warranty is void if the instrument has been damaged by accident, misuse, negligence, or serviced
by any person not authorized by BPL Medical.
This limited warranty contains the entire obligation of BPL Medical and no other warranties expressed,
implied, or statutory are given. No representative or employee of BPL Medical is authorized to assume any
further liability or grant any further warranties except as set herein.
BPL Medical
11th KM, Bannerghatta Road, Arakere, Bengaluru, INDIA 560076.
This equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this document. This equipment has been tested and found
to comply with the limits for medical devices to IEC60601-1-2: 2014. These limits are designed to provide
reasonable protection against harmful interference in a typical medical installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee
that interference will not occur in a particular installation. If this equipment does cause harmful interference
to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
WARNING: The WALK T1 should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the WALK T1 should be observed to verify normal operation in the configuration in
which it will be used.
The WALK T1 is intended for use in a Home Healthcare Environment and is suitable for use in all
establishments, including domestic establishments and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic purposes. This equipment has been tested
and found to comply with the limits for medical devices to IEC 60601-1-2: 2014.
Immunity test Applies to Compliance level Electromagnetic Environment-Guidance for
Home Healthcare Environment
Electrostatic All device ± 2, 4, 6, 8kV Floors should be wood, concrete or ceramic tile.
discharge (ESD) input and contact If floors are covered with synthetic material, the
IEC 61000-4-2 output ± 2, 4, 8, 15kV air relative humidity should be at least 40 %.
connections discharge
and cables
Radiated RF EM All device 10V/m Radiated electromagnetic fields should be at
fields IEC 61000-4-3 input and 80 MHz to levels characteristic of a typical location in a
output 2700MHz Home Healthcare Environment
connections 80% AM at 1kHz
and cables
Radiated All device See Table A This device has been subjected to RF wireless
RF Wireless input and below communication bands from cell phones, and
communication output other communication devices
equipment IEC connections
61000-4-3 and cables
Electrical fast N/A N/A N/A
transient/burst IEC
61000-4-4
Note: In the event of an error, the device will auto-recover within 5 seconds.
E = [6/d] √P
Recommended separation distances between portable and mobile RF communications equipment and
the WALK T1
Rated maximum Separation distance according to frequency of transmitter (m)
output power of
transmitter (W) 150 kHz to 80 MHz 80 MHz to 800 MHz 800MHz to 2.5GHz
d = 1.2 √P d =1.2 √P d = 2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the monitor is used exceeds the applicable RF compliance level above, the monitor should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the monitor.
b) Over the frequency range 150kHz to 80 MHz, field strengths should be less than 3V/m
Table A – Test specifications for the device’s Signal Input Parts/Signal Output parts to RF wireless
communication equipment.
Test Band a) Service b) Modulation Maximum Distance IMMUNITY TEST LEVEL
Frequency (MHz) b) Power (W) (m) (V/m)
(MHz)
358 380 – 390 TETRA 400 Pulse 1,8 0.3 27
Modulation
b) 18Hz
450 430 – 470 GMRS 460, FM c) 5 kHz 2 0.3 28
FRS 460 deviation 1
kHz sine
710 704 - 787 LTE Band Pulse 0,2 0.3 9
745 13, 17 Modulation
780 b) 217Hz
810 800 - 960 GSM Pulse 2 0.3 28
870 800/900, Modulation
930 TETRA 800, b) 18Hz
iDEN 820,
CDMA 850,
LTE Band 5
1720 1700 - 1990 GSM 1800, Pulse 2 0.3 28
1845 CDMA Modulation
1970 1900, GSM b) 217Hz
1900, DECT,
LTE Band
1, 3, 4, 25,
UMTS
Battery Disposal
The BPL Medical WALK T1 contains a small lithium ion battery on a Printed Circuit (PC) board that contain
materials which may be hazardous to human health. The battery cannot be easily removed and therefore the
WALK T1 must be disposed of in an environmentally responsible way or returned to BPL Medical. A prepaid
return label can be obtained. Please follow local rules and regulations for the safe disposal of the WALK T1.
Cuff
Do not return used cuffs. Used blood pressure cuffs may be contaminated medical waste and should be
dealt with in accordance to your local regulations for medical waste.