Manual_ADKS-4 Portable four-in-one gas detector
Manual_ADKS-4 Portable four-in-one gas detector
Instruction manual
ADKS-4
1
1. Product introduction
This product has four kinds of gas detection sensors, which can detect
Ex(Explosive gas),H2S,CO,O2 through natural diffusion method. It has
excellent sensitivity and perfect repeatability, easy to use and maintain, and
can be used for industrial site safety monitoring.
The equipment has high reliability, and the shell can meet the requirements
of waterproof, dustproof and explosion-proof.
The detector is widely used in petroleum, chemical, environmental protection,
metallurgy, gas transportation, biochemistry, medicine and other industries
2
1 Alarm indicator light 5 Right button
3
2.2 Working principle: electric-chemical type and catalyze burning
type
3.Parameters
s threshold threshold
EL
S m
4
Ex :Explosive gas
5
Ex alarm threshold O2 alarm threshold
4.5 Alarm
When the device detects that Ex, H2S, CO and O2 exceed the alarm
threshold, the indicator light and screen light up, the vibration motor starts,
and the buzzer sounds.
6
4.6 Status of equipment
Press the left button after entering the " Detecting " page to view the device
date, time and battery status.
4.7 Shutdown
Long press the middle button for 3 seconds on the detection page, and the
screen will display the shutdown interface. Left click to confirm shutdown,
right click to cancel.
4.8 Charging
When the battery voltage of the device is below 3.5V, the device will
automatically shut down. Charge the battery in time.
When the battery is full, the screen will display "Charge Over".
5. Menu operations
5.1 Press the middle key to enter the main menu interface
5.2 In the menu selection interface, the top button corresponds to the left
selection item of the display, the bottom button corresponds to the right
selection option of the display, and the middle button is the confirm, save or
enter sub-menu function
5.3 If you do not operate in the menu interface for more than 5 seconds after
entering the main menu, you will automatically return to the normal detection
interface.
7
Main menu structure and operations
The menu cursor can be moved by the top button and
the bottom button, and the middle button can enter the
selected menu item. If the cursor moves to "Exit", the
detection main interface can be returned.
Zero adjustment
This function is used to correct the zero drift of the
instrument by manually adjusting the compensation
value in the case of zero drift of the display value of the
instrument.
Move the cursor to the "Zero Adjust" menu item and
press the middle button to enter the correct password
in the password interface to enter the "Select Gas"
page.
Select the gas item that needs to be adjusted to Zero,
enter the "Zero Adjust" interface, press the left button to
select the digit to be adjusted, and press the right
button to adjust the value. Press the middle key to
save the current setting.
Note: The zero compensation value should not deviate
too much from the actual zero drift, otherwise the actual
display value will deviate too much.
8
Gas calibration
This function is for the internal use of the manufacturer
only.
Alarm record
The menu of Alarm record includes three sub-menus:
"Alarm View", "Alarm Clear" and "Exit".
"Alarm View":Used to view the history of alarm records,
press the left button to scroll up, press the right button
to scroll down.
"Alarm Clear":Used to clear the alarm record, press the
left button to confirm the clear alarm record, press the
right button to cancel.
"Exit":Exit the current menu.
9
Date & Time Settings
Move the cursor to the "Time set" menu item and press
the middle key to enter the "Year ","
Month","Day","Hour" and "Min" setting menu items
respectively.
Language Settings
Move the cursor to the "Language" menu item and
press the middle button to enter the language selection
interface. You can switch between Chinese and
English menu languages.
10
Factory Mode
This function is only for internal use by the
manufacturer
6.Precautions in Use
6.1 During the operation of the device, the charging speed will slow down, so
it is recommended to shut down and charge.
6.2 When explosive gas seriously exceeds the range, the equipment may not
be able to detect correctly, and even cause sensor damage, so do not directly
use explosive gas with higher concentration such as lighter for detection.
6.3 Corrosive gases (such as high concentration of chlorine gas) may
damage the sensor, so avoid contact with such gases during storage or use,
and avoid using or storing this product under long-term high temperature,
high humidity and strong sunlight.
6.4 Do not wipe the product with corrosive solvents.
6.5 In order to ensure the detection accuracy, the machine should be
11
calibrated regularly, and the verification cycle should not exceed one year.
6.6 Non-professional personnel shall not disassemble this product.
6.7 This equipment is only used for gas leakage assessment of explosive gas,
not environmental explosive gas safety assessment instrument, and it is
prohibited to be used for assessing environmental explosive gas safety.
6.8 Some silicone rubber products contain sulfides and halogen volatiles that
may cause chemical poisoning of the sensor and lead to failure, so avoid
contact with such items when stored.
6.9 Drops or violent vibrations may permanently damage the equipment.
7. Storage environment
Equipment storage environment temperature -10~55℃, relative humidity less
than 85%, and good ventilation conditions, the air should not contain
corrosive harmful gases.
8. Packing accessories
1PCS packing box,1PCS host,1PCS charger,1PCS instruction manual,1PCS
certificate & warranty card.
12
H2 0~1000ppm 0.1ppm 200 500
Cl2 0~20ppm 0.1ppm 5 10
HCl 0~20ppm 0.1ppm 5 10
SO2 0~20ppm 0.1ppm 5 10
NO 0~250ppm 0.1ppm 20 125
NO2 0~20ppm 0.1ppm 5 10
CO2 0~5000PPM 0.1%VO 0.2%VOL 0.5%VOL
L
PH3 0~20ppm 1ppm 5 10
HC 0~50ppm 1ppm 10 25
N
C2H4 0~100ppm 1ppm 20 50
O
O3 0~100ppm 0.1ppm 20 50
CH2 0~40ppm 0.1ppm 8 20
O
C6H6 0~1000ppm 1ppm 200 500
C7H8 0~1000ppm 1ppm 200 500
C8H1 0~1000ppm 1ppm 200 500
0
VOC 0~1000ppm 1ppm 200 500
C2H3 0-250ppm 1ppm 50 125
Cl
CH3 0-30ppm 1ppm 6 15
OH/
CH4
O
C4H8 0-90ppm 1ppm 18 45
C2H6 0-80ppm 1ppm 16 40
13
O
The test range of the customized instrument is subject to the actual factory