Estroboscopio Manual
Estroboscopio Manual
Endoscopy
Repair Manual
Machine Translated by Google
Warranty
Stryker Endoscopy warrants the X-7000 Light Source shall Stryker Endoscopy be liable for any breach of war-
against defects in both materials and workmanship to the ranty in any amount exceeding the purchase price of the
registered owner at the time of purchase. All components product.
are covered by the warranty for a period of one year from
the date of purchase. No agent, employee or representative of Stryker Endoscopy
has the authority to bind the Company to any other war-
This warranty does not apply to any unit which has been ranty, affirmation, or representation concerning this
the subject of misuse, abuse, neglect, improper installation instrument.
or operation or that which has been altered, adjusted, or
tampered with by any person other than Stryker Endos- This warranty is valid only to the original purchaser of
copy authorized service personnel. Stryker Endoscopy products directly from Stryker Endos-
copy or from a Stryker Endoscopy authorized agent. The
The customer is responsible for returning the defective warranty cannot be transferred or assigned by the original
equipment to the factory at his or her own expense. purchaser.
Stryker Endoscopy or its representative will service the unit,
repair or replace any defective parts thereof, and return the unit. The X-7000 Light Source warranty is void if any WARN-
INGS, CAUTIONS, or NOTES are disregarded.
If, upon examination, it is determined that the fault has All Stryker products are warranted against defects in
been caused by misuse or abnormal conditions of materials and workmanship.
operation, the repairs will be billed to the customer in the
same manner as out-of-warranty repairs.
Products repaired by Stryker Endoscopy will be issued a © Stryker and Stryker Endoscopy are registered trade-
30 day repair warranty against defects in both materials marks of Stryker Corporation.
and workmanship, provided the original warranty period
has expired. This warranty applies only to products that
have been repaired by Stryker. Instruments submitted
due to defects in materials and workmanship during the
warranty period will be repaired at no charge to the customer.
Table of Contents
1.0 INTRODUCTION
1.1 Scope.................................................................................................................................................................................................1
1.2 Responsibility..........................................................................................................................................................................1
1.3 Upgrades.................................................................................................................................................................................1
7.0 APPENDIX
8.0 WARRANTY
...................................................................................................................................................................................................................24
9.0 SERVICE
...................................................................................................................................................................................................................24
1.0 Introduction
1
Machine Translated by Google
If service is needed either during or after the warranty Each diagnostic and repair procedure described in this
manual requires a technician qualified by training or experience
period:
in the following areas:
Package carefully in the original shipping con-tainer, if If a repair procedure necessitates replacement parts,
possible.
please refer to the Appendix (page 22).
2
Machine Translated by Google
2.1 GENERAL RECOMMENDATIONS NOTE: A proper electrical installation procedure will ensure that
all associated signal cabling is cor-rectly installed
The Stryker Endoscopy Model X-7000 Light Source is a prior to applying AC power to the system.
If some part of a procedure is omitted or adequate associated devices does not exceed the available cable
equipment is not used, the safety and performance of the device length.
recommended that if any element in these procedures is beyond • Ventilation is not obstructed around the unit. Avoid
the scope of the technician's training, refer to section 1.7 locations with excessive heat or temperature fluctuation
for information on obtaining fully qualified professional tions, such as direct sunlight.
precautions are not taken, results may include Light Source Operating and Maintenance Manual (Stryker P/N
3
Machine Translated by Google
4
Machine Translated by Google
Other Observations:
Staff Recommendation:
Machine Translated by Google
REPAIR CODES
NLB NCB NLL NMO NPB BBR IRR NSW NFU NFG NSH NDB NJW EBF CHR NC1 BUR BAR RCR DFR NSP NHL NFP
Integrating
Assembly
Assembly
Software
Replace
Upgrade
Ethernet
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Eprom /
Replace
Replace
Replace
Replace
Chassis
Control
Display
Shutter
Ribbon
house)
Handle
Engine
Ballast
Ballast
Switch
Power
Mirror
Board
Board
Board
Board
Board
Cover
Cable
Panel
Front
Knob
Slide
Fuse
Bulb
Bulb
Bulb
Inlet
Feet
Rod
Jaw
Jaw
Fan
Fan
Hot
(in-
Pot
AC
Bulb, Bulb
The following instructions apply to the removal of Bulb Fan Board
components from the X-7000. To remove a component
for replacement or repair will require the removal of certain Shutter Hot Engine Engine
Mirror Mount
other components. Components must be removed in the
Assembly
order listed in the following procedures.
Integrating
Rod
The table in section 4.1.1 shows the order in which
components must be removed according to the basic
physical assembly of the X-7000. To remove a Slide Front Power
component for replacement, or repair, begin with section Button, Panel Switch
4.2 and proceed accordingly to the appropriate section. Jaw Knob
Display
Board
Jaw
Assembly
7
Machine Translated by Google
8
Machine Translated by Google
Tools Required:
Basic Tool Kit
9
Machine Translated by Google
Tools Required:
Basic Tool Kit
10
Machine Translated by Google
Tools Required:
Basic Tool Kit
11
Machine Translated by Google
• Remove the fan wire harness assembly. • Remove the bulb module.
• Note the fan orientation. • Detach the wire harness from the control board.
• Remove the 2 nuts attaching the fan to the Chassis and • Remove the 2 nuts attaching the bulb fan housing assembly
remove the sub-assembly. to the chassis.
• Remove the 4 screws from the fan mounts. • Note the fan orientation.
• Remove the malfunctioning fan(s) and replace. Ensure the correct • Remove the 4 nuts attaching the bulb fan to the fan
orientation. mounts/fan ducts.
• Reinstall the 2 fan mounts with the 4 screws. • Reinstall the new fan onto the mounts/duct with 4 bolts,
ensuring the correct orientation.
• Reinstall the 2 nuts.
• Reinstall the fan assembly onto the chassis with 2 nuts.
• Reinstall the black ribbon cable, fan wire harness assem-
bly, and console cover. • Reconnect the wire harness.
• Perform Electrical Current Leakage Test (procedure 6.3). • Replace the bulb module and cover.
12
Machine Translated by Google
13
Machine Translated by Google
14
Machine Translated by Google
15
Machine Translated by Google
• Remove the jaw handle from the actuator shaft. • Remove the set screw from the bottom of the potentiom-
another knob.
• Replace the jaw handle onto the actuator shaft. • Loosely screw the set screw into the bottom of the potenti-
tiometer knob.
• Secure the jaw handle in place using the set screw.
• With the screw hole facing down, insert the potentiometer
• Reinsert the set screw and reinstall the cover. knob onto the slide pot.
16
Machine Translated by Google
• Remove power cord and console cover (procedure 4.2). 4.5.4 Jaw Assembly Replacement
• Remove the jaw knob and potentiometer knob (procedures • Remove power cord and console cover (procedure 4.2).
4.5.1 and 4.5.2,).
• Remove the front panel (procedure 4.5.3).
CAUTION: DO NOT DAMAGE THE JAW INTERLOCK
CABLE WHEN REMOVING THE FRONT • Remove the motor mount (procedure 4.4.5).
PANEL.
• Disconnect the ribbon cable from the inside of the front • Remove the 4 Phillips screws.
panel.
• Remove the jaw assembly and replace.
• Disengage the six tabs which hold the front panel onto
the chassis. • Reinsert the 4 Phillips screws.
• Remove the jaw interlock cable from the display board. • Replace the ESST wire harness, motor mount, front
panel, and cover.
• Remove the display board (procedure 4.5.6).
17
Machine Translated by Google
• Remove power cord and console cover (procedure 4.2). • Remove power cord and console cover (procedure 4.2).
• Remove the jaw knob and potentiometer knob • Remove the front panel (procedure 4.5.3).
(procedures 4.5.1 and 4.5.2).
• Detach the flex cable from the display board (J1).
• Remove the front panel (procedure 4.5.3).
• Unclip the 6 front panel clips from the chassis.
• Cut the zip tie over the cables.
• Remove the ESST ring cable from the display board.
• Disconnect wire harness from display board.
• Remove the ribbon cable from the display board.
• Push in the two snap clips and push out the switch from
inside the console. • Remove the power switch wire harness from the display
board.
• Insert a new power switch. • Push the board onto the front panel clips.
• Reconnect the wire harness. • Reattach the power switch wire harness.
• Replace the zip tie, front panel, and console cover. • Reattach the ESST ring wire harness.
• Perform Electrical Current Leakage Test (procedure 6.3). • Reattach the front panel ribbon cable.
18
Machine Translated by Google
Tools Required:
Basic Tool Kit
• Locate the fuse holder below the power cord inlet (see
Figure 1).
19
Machine Translated by Google
20
Machine Translated by Google
6.1 REQUIRED SKILLS • Disconnect the power cord from the power source.
• Basic electronic techniques • Disconnect the video cables from the console.
21
Machine Translated by Google
• Touch the positive test lead (red probe) to the alligator clip
and the negative lead (black probe) to the ground post.
22
Machine Translated by Google
7.0 Appendix
• MAP 0337
23
Machine Translated by Google
8.0 Warranty
This Stryker Endoscopy product is warranted to the original purchaser to be free from defects in material and workmanship for the
following times:
• One year following the date of delivery and an additional 30 days following repair
This warranty extends to all purchases and is limited to the repair or replacement of the product without charge when returned in the
original shipping case to:
Stryker Endoscopy
5900 Optical Court
San Jose, CA 95138
Stryker Endoscopy cannot accept responsibility for returns or replacements which have not been authorized. This warranty does not
cover damages caused by misuse or by failure to follow the procedures outlined in this manual or demonstrated by Stryker
Endoscopy representatives.
9.0 Service
The user is advised to return a malfunctioning unit to Stryker Endoscopy for repair or replacement, where specialized equipment and
technicians are available to perform repairs while maintaining full product quality and safety.
In the event that the user decides to undertake repair procedures, Stryker Endoscopy recommends that these be carried out only by
qualified technicians with proper test equipment listed in this manual, so that the safety of operators and patients may not be
compromised.
1. Contact Stryker Endoscopy at 1-800-624-4422, or call your local Stryker Endoscopy sales representative.
2. Clean and sterilize all parts that will be returned for service. Follow the instructions provided in the user manual.
3. Package all the components carefully in the original shipping container if possible.
4. Ship the product pre-paid and insured to:
The product described in this manual is continually being reviewed, and improvements may be made without notice.
Stryker and Stryker Endoscopy are registered trademarks of Stryker Corporation.
24
Machine Translated by Google
1000-400-755 Rev. A