OPEN AMPEROMETRIC CELLS - EMEC

The ECL Series open amperometric cells are designed for measuring free chlorine in both organic and inorganic forms, featuring a constant water flow requirement and various models for different water types. Each model includes specifications for measuring range, connection, and maintenance, with a focus on stable and reliable performance. The document also outlines the construction materials and operational parameters for optimal use.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
40 views9 pages

OPEN AMPEROMETRIC CELLS - EMEC

The ECL Series open amperometric cells are designed for measuring free chlorine in both organic and inorganic forms, featuring a constant water flow requirement and various models for different water types. Each model includes specifications for measuring range, connection, and maintenance, with a focus on stable and reliable performance. The document also outlines the construction materials and operational parameters for optimal use.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

ECL Data Sheet

Open amperometric cells

ECL Series is designed for measuring free chlorine (both organic and
inorganic).
Open amperometric cells comprise an Off-line probe holders, a sensing
electrode and a flow electrode.
Probe holders can contain up to three probes (temperature, pH and ORP).

The flow of water within this cell must remain constant and within 40 l/h.
A pressure stabilizer is available for areas subject to sudden pressure
changes.
ECL6
It is recommended to instal a filter before the probe holder. ECL7
ECL12

• Chlorine probes (hypochlorous acid)


• Stable and reliable measurement even with low chlorine concentrations
values
• Acrylic body
• Continuous sampling measurement
• Proximity flow sensor controlled (ECL6; ECL6/E; ECL7; ECL12; ECL12/E)
ECL20
ECL21

R21221

CHLORINE in water can be present in different combination:


FREE CHLORINE ACTIVE: HOCl (hypochlorous acid)
COMBINED CHLORINE: Monochloramine, dichloramine, trichloramine (DPD4-DPD1 analysis system)
FREE ORGANIC CHLORINE: Free chlorine with isocyanide acid (DPD1 analysis system)
FREE INORGANIC CHLORINE: Free chlorine. (DPD1 analysis system)
TOTAL CHLORINE: Free chlorine and combined chlorine. (DPD4 analysis system)

MODELS
• ECL20 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR FRESH WATER
• ECL21 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
• ECL6 FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL7 FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL6/E FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL12 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
• ECL12/E FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL6
ECL6/E
ECL6 ECL6/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) / BROMINE
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, ORP and temperature temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL7
ECL7
Parameter FREE CHLORINE (ORGANIC AND INORGANIC)
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, Redox (PG13,5) e temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL12
ECL12/E
ECL12 ECL12/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/silver)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, ORP and temperature temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/silver
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL20
ECL20
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR FRESH WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL21
ECL21
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/silver
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.

Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
6 5 4 3 2 1
DIMENSIONS
ECL12
D D
223.20

100.00
C C

9.00

9.00
106.00

B B
60.00

6 5 4 3 2 1
201.70
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo:° Peso Netto: A
Massimo_F Massimo_F 22/07/2015 ±0.05
Materiale:
D
ECL12/E D
ECL12 CELLA AMPEROMETRICA Edizione Foglio
ECL12 00 1/1
6 5 4 3 2 1
100.00
177.30

C C
9.00

9.00
81.00

B B
60.00

171.70

A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 22/07/2015 ±0.05
Materiale:
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008 ECL12E CELLA AMPEROMETRICA
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15 Edizione Foglio
ISO 14001:2004 ECL12E 00 1/1
6OHSAS 18001:2007 5 4 3 2 1
ECL Data Sheet
6
Open amperometric cells
5 4 3 2 1
DIMENSIONS
ECL6 / ECL7
D D
223.20

100.00
C C

9.00

9.00
106.00

B B
60.00

6 5
201.70 4 3 2 1
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo:° Peso Netto: A
Massimo_F Massimo_F 21/07/2015 ±0.05
Materiale:
ECL6/E
ECL6 CELLA AMPEROMETRICA Edizione Foglio
D ECL6 00 1/1 D
6 5 4 3 2 1
100.00
177.30

C C
9.00

9.00
81.00

B B
60.00

171.70

A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 21/07/2015 ±0.05
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
Materiale:
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004 ECL6E CELLA AMPEROMETRICA Edizione Foglio
OHSAS 18001:2007 ECL6E 00 1/1
6 5 4 3 2 1
ECL Data Sheet
Open amperometric cells
DIMENSIONI
6 5 4 3 2 1
ECL21

110.00

D D

123.90
C C
10.00

10.00
B B
60.00
55.00

D
A

A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 15/07/2015 ±0.05
Materiale:

DIMENSIONALE ECL5N Edizione Foglio


ECL5N 00 1/1
6 5 4 3 2 1
6

6
ECL20

55.00 10.00
10.00
5

5
60.00

105.00
4

123.90