OPEN AMPEROMETRIC CELLS - EMEC
OPEN AMPEROMETRIC CELLS - EMEC
ECL Series is designed for measuring free chlorine (both organic and
inorganic).
Open amperometric cells comprise an Off-line probe holders, a sensing
electrode and a flow electrode.
Probe holders can contain up to three probes (temperature, pH and ORP).
The flow of water within this cell must remain constant and within 40 l/h.
A pressure stabilizer is available for areas subject to sudden pressure
changes.
ECL6
It is recommended to instal a filter before the probe holder. ECL7
ECL12
R21221
MODELS
• ECL20 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR FRESH WATER
• ECL21 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
• ECL6 FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL7 FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL6/E FOR FREE CHLORINE (ORGANIC AND INORGANIC)
• ECL12 FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
• ECL12/E FOR FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL6
ECL6/E
ECL6 ECL6/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) / BROMINE
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, ORP and temperature temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL7
ECL7
Parameter FREE CHLORINE (ORGANIC AND INORGANIC)
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/copper; on request gold/copper)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, Redox (PG13,5) e temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL12
ECL12/E
ECL12 ECL12/E
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes (platinum/silver)
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Suitable as probe
pH, ORP and temperature temperature
holder for
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/silver
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL20
ECL20
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR FRESH WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/copper
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
ECL21
ECL21
Parameter FREE CHLORINE (ORGANIC AND INORGANIC) FOR SALT WATER
Measuring range 0-10 mg/l (0-10 ppm) resolution: ± 0.05
Connection 2 wires (+red; -black)
Measuring system amperometric - 2 electrodes
Ph working range 6-8 pH
First polarization: 2 h about
Run-in-time
Next polarizations: 50 min. about
Response time T90: 2 min. about
Zero point
See Operating manual: “Probe alignment”
adjustment
Slope calibration See Operating manual: “Probe alignment”- DPD1 method
Alcalinity min 100 ppm
Working
5-40° C (41-104°F)
temperature
Pressure 0.4 - 5 bar (5.8 - 72.5 PSI)
Cable (standard) 2 m (6.6 ft); 1 m if assembled on panel
Working flow 40 l/h
Fittings for
connection to the 6x8
sample pipeline
Electrode: platinum/silver
Material
Measurement cell: metacrylate (PMMA)
Mounting On flat vertical surface (panel, support, etc.).
Storage Frost and dry protected (5-40° C)
Regular control of the signal
Maintenance SHORTEN THE MAINTENANCE INTERVALS APPROPRIATELY DEPENDING
ON WATER QUALITY.
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
OHSAS 18001:2007
ECL Data Sheet
Open amperometric cells
6 5 4 3 2 1
DIMENSIONS
ECL12
D D
223.20
100.00
C C
9.00
9.00
106.00
B B
60.00
6 5 4 3 2 1
201.70
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo:° Peso Netto: A
Massimo_F Massimo_F 22/07/2015 ±0.05
Materiale:
D
ECL12/E D
ECL12 CELLA AMPEROMETRICA Edizione Foglio
ECL12 00 1/1
6 5 4 3 2 1
100.00
177.30
C C
9.00
9.00
81.00
B B
60.00
171.70
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 22/07/2015 ±0.05
Materiale:
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008 ECL12E CELLA AMPEROMETRICA
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15 Edizione Foglio
ISO 14001:2004 ECL12E 00 1/1
6OHSAS 18001:2007 5 4 3 2 1
ECL Data Sheet
6
Open amperometric cells
5 4 3 2 1
DIMENSIONS
ECL6 / ECL7
D D
223.20
100.00
C C
9.00
9.00
106.00
B B
60.00
6 5
201.70 4 3 2 1
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo:° Peso Netto: A
Massimo_F Massimo_F 21/07/2015 ±0.05
Materiale:
ECL6/E
ECL6 CELLA AMPEROMETRICA Edizione Foglio
D ECL6 00 1/1 D
6 5 4 3 2 1
100.00
177.30
C C
9.00
9.00
81.00
B B
60.00
171.70
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 21/07/2015 ±0.05
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
Materiale:
ISO 9001:2008
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004 ECL6E CELLA AMPEROMETRICA Edizione Foglio
OHSAS 18001:2007 ECL6E 00 1/1
6 5 4 3 2 1
ECL Data Sheet
Open amperometric cells
DIMENSIONI
6 5 4 3 2 1
ECL21
110.00
D D
123.90
C C
10.00
10.00
B B
60.00
55.00
D
A
A Progettato da: Controllato da: Data: Toll. Gen. Peso Lordo: Peso Netto: A
Massimo_F Massimo_F 15/07/2015 ±0.05
Materiale:
6
ECL20
55.00 10.00
10.00
5
5
60.00
105.00
4
123.90
Via Donatori di sangue, 1 - 02100 Rieti - Italy Specifications subject to change without notice.
ISO 9001:2008
Progettato da:
Tel. +39 0746 2284 1 - Fax +39 0746 2284 2 - http://www.emec.it EN R1-05-15
ISO 14001:2004
Massimo_F
OHSAS 18001:2007
3
3
DIM