Manual DR 5000
Manual DR 5000
Manuals / Brands / Hach Manuals / Laboratory Equipment / DR 5000 / User manual / PDF
Quick Links
Stored Programs
Lamp Replacement
Section 8 Troubleshooting
Table of Contents
Table of Contents
Safety information
Use of hazard information
Precautionary labels
Class 1 LASER
Chemical and Biological Safety
Source Lamp Safety
Overview of product
Unpack the instrument
Operating environment
Connection
Cell compartment, Multi Cell Holder, 100 mm Cell Adapter (optional)
mm adapter (optional)
Installation Multi Cell Holder
Beam path
Power the instrument on and off
Language selection
Self-Check
Overview
Tips for the use of the touch screen
Use of the alphanumeric keypad
Main Menu
Instrument Setup mode
Operator ID
Sample ID
Date and time
Display and sound preferences
Lamp Control
VIS-Lamp and UV-Lamp
Effect caused by the lamp settings on the Measuring Mode
PC and printer
Printer setup
Print data
HACH Data Trans
Password
Password deactivation
Store, recall, send and delete data
The Data Log
Auto/manual data storage
Recall stored data from the data log
Send data from the data log
Delete stored data from the data log
Store, recall, send and delete data from wavelength scan and time
course
Data storage from wavelength scan or time course
Recall stored data from wavelength scan or time course
Send data from wavelength scan or time course
Delete stored data from wavelength scan or time course
Stored Programs
Select a saved test/method; enter user-specific basic data
Stored program options
Use of program timers
Set the reading mode
Take single wavelength measurements (single reading)
Take single wavelength measurements (continuous readings)
Set the dilution factor
Run a standard adjust
Set the chemical form
Change of the default setting of the chemical form
Run a reagent blank
Analysis of samples
Add stored programs to the favorite programs list
Barcode Programs
Complete a barcode 13 mm test/vial
Select the measuring range
Select the chemical evaluation form
Change of the default setting of the chemical form
Basic test-specific and sample-specific data settings
Sample blank
Update/edit barcode tests
Manual update of a barcode test
Update an existing barcode test
Program a new test
Upgrade of the instrument software
User Programs
Program a user method
Single wavelength settings
Multi wavelength settings
Calibration settings for single and multi wavelength mode
Store a user program
Additional user-defined parameters and functions
Free programming program type
Measurement process
Enter a new element of a measuring sequence
Enter the calibration formula (evaluation formula)
Enter variables
Save a free programming user program
Select a user program
Add, edit and delete user programs from the favorites list
Add to Favorites
Edit
Delete
Favorite Programs
Recall a favorite program
Delete a favorite program
Standard Addition - monitoring/checking results
Complete a standard addition
Single Wavelength (absorbance, concentration and transmittance
measurements)
Set up single wavelength mode
Take single wavelength measurements (single reading)
Take single wavelength measurements (continuous readings)
Multi-Wavelength mode - measurements at more than one
wavelength
Set the reading mode at different wavelengths
Complete a measurement in the multi wavelength mode
Wavelength Scan mode - recording of absorbance and
transmittance spectrums
Set up the wavelength scan
Wavelength scan reading
Navigation of the wavelength scan graph or a wavelength scan analysis
Work with reference scans
Time course of absorbance/transmittance
Time course setup parameters
Time course scan reading
Analysis of time course data
Navigation of a time scan or a time scan analysis
System checks
Instrument information
Upgrade of the instrument software
Optical checks
Wavelength check
Noise check
Stray light check
Absorbance check
Drift check
Verification kit
Output checks
Lamp history
Factory service
Service time
Instrument Backup
Cleaning requirements
Spectrophotometer
Display
Cuvettes/sample cells
Lamp replacement
Tungsten lamp replacement
Deuterium lamp (UV) replacement
Fuse replacement
Filter pad maintenance
Filter pad replacement
DOC022.53.00654
DR 5000
USER MANUAL
3
Downloaded from www.Manualslib.com manuals search engine
Table of Contents
4
Downloaded from www.Manualslib.com manuals search engine
Table of Contents
5
Downloaded from www.Manualslib.com manuals search engine
Table of Contents
6
Downloaded from www.Manualslib.com manuals search engine
Section 1 Specifications
Specifications are subject to change without notice.
Performance specifications
Spectral bandwidth 2 nm
User programs 50
Operating requirements 10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
Storage requirements –25–60 °C (–13–140 °F) max. 80% relative humidity (non-condensing)
7
Downloaded from www.Manualslib.com manuals search engine
Specifications
8
Downloaded from www.Manualslib.com manuals search engine
Section 2 General Information
DANGER
Indicates a potentially or imminently hazardous situation
which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation
which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in
minor or moderate injury.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol indicates that the instrument contains a Class 1 LASER device.
LASER CLASS 1 Data: 0.3 mW; λ = 650 nm
9
Downloaded from www.Manualslib.com manuals search engine
General Information
DANGER
Potential Chemical/ Biological Exposure Hazards. Handling
chemical samples, standards and reagents can be dangerous.
Users of this product are advised to familiarize themselves
with safety procedures and the correct use of chemicals, and
to carefully read all relevant Material Safety Data Sheets.
CAUTION
The UV lamp generates UV light.
Do not look directly at an operating lamp without wearing UV
protective eye glasses.
10
Downloaded from www.Manualslib.com manuals search engine
General Information
11
Downloaded from www.Manualslib.com manuals search engine
General Information
12
Downloaded from www.Manualslib.com manuals search engine
Section 3 Installation
WARNING
Electrical and Fire Hazards. Use only the provided power
supply. Only qualified personnel should conduct the tasks
described in this section of the manual.
13
Downloaded from www.Manualslib.com manuals search engine
Installation
3.3 Connection
The DR 5000 has three USB interfaces as a standard feature,
located on the back of the instrument (Figure 1).
The USB Type A interfaces are used for communications with a
printer, USB memory stick or keyboard. A USB memory stick is
used to update instrument software.
The USB Type B interface is used for communications with a PC.
The optional Hach Data Trans software (see section 5.2.6.3 on
page 31) must be installed on the PC for this use.
A USB hub may be used to connect several accessories at a time.
Note: USB cables must not be longer than 3 meters (10 feet).
Figure 1 Interfaces
1 USB type A 4 Fuse
2 USB type B 5 Plug in power supply
3 On/Off switch
14
Downloaded from www.Manualslib.com manuals search engine
Installation
3.4 Cell compartment, Multi Cell Holder, 100 mm Cell Adapter (optional)
15
Downloaded from www.Manualslib.com manuals search engine
Installation
16
Downloaded from www.Manualslib.com manuals search engine
Installation
2. Identify the correct opening for the selected cell type in the
Multi Cell Holder.
17
Downloaded from www.Manualslib.com manuals search engine
Installation
18
Downloaded from www.Manualslib.com manuals search engine
Section 4 Start Up
Note: Do not turn the instrument off and on in rapid succession. Always
wait about 20 seconds before turning the instrument on again, otherwise
the electronic and mechanical systems will be damaged.
1. While turning the instrument on, touch the screen at any point
until the list for selecting a language appears (about
30 seconds).
4.3 Self-Check
19
Downloaded from www.Manualslib.com manuals search engine
Start Up
20
Downloaded from www.Manualslib.com manuals search engine
Section 5 Standard Operations
5.1 Overview
5.1.1 Tips for the use of the touch screen
The entire screen is touch-activated. To make a selection, press the
screen with a fingernail, fingertip, pencil eraser or a stylus. Do not
press the screen with a sharp object, such as the tip of a ball
point pen.
• Do not place anything on top of the screen, to prevent damage
or scratching on the screen.
• Press keys, words or icons to select them.
• Use scroll bars to move up and down long lists very quickly.
Press and hold the scroll bar, then move up or down to move
through the list.
• Highlight an item from a list by pressing it once. When the item
has been successfully selected, it will be displayed as reversed
text (light text on a dark background).
21
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Stored programs are pre-programmed methods that make use of HACH reagents and LANGE
Stored Programs (HACH cuvette tests and pipette tests.
LANGE Programs) The DR 5000 Procedures Manual contains illustrated, step-by-step procedures for analyses
using HACH programs. The working procedures for LANGE tests are included in the test packs.
User programs make "made to measure analysis" possible:
–Users can program methods they have developed themselves
User Programs
–Existing HACH and LANGE methods can be stored as user programs. The LANGE tests can
then be modified to suit the user's requirements.
Favorite Programs List of methods/tests created by the user to suit his own requirements.
Single wavelength measurements are:
Absorbance measurements: The light absorbed by the sample is measured in absorbance
units.
Single Wavelength Transmittance measurements (%): The percentage of the light that passes through the sample
and reaches the detector is measured.
Concentration measurements: A concentration factor can be entered to enable the measured
absorbance values to be converted into concentration values.
In the multi-wavelength mode, absorbance (Abs) or percentage transmittance (%T) is measured
Multi Wavelength at up to four wavelengths and absorbance differences and absorbance relationships are
calculated. Simple conversions into concentrations can also be carried out.
A wavelength scan shows how the light from a sample is absorbed over a defined wavelength
spectrum.This function can be used to determine the wavelength at which the maximum
Wavelength Scan
absorbance value can be measured.The absorbance behavior is displayed graphically during the
scan.
Time Course The time scan records the absorbance or % transmittance at a wavelength over a defined time.
The system checks menu offers a number of options, including optical checks, output checks,
System Checks
lamp history, instrument update, service time and instrument backup.
Recall Data Stored data can be recalled, filtered, sent and deleted.
In this mode, user-specific or method-specific settings can be entered: Operator-ID, Sample-ID,
Instrument Setup
Date & Time, Display & Sound, Lamp Control, PC & Printer, Password, and Module Options.
22
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
5.2.1 Operator ID
Use this option to enter up to 30 sets of operator initials (up to
5 characters each) into the instrument. This feature helps record
which operator measured each sample.
4. Press OK to confirm.
23
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
5.2.2 Sample ID
Use this option to enter up to 30 Sample Identification tags (up to
13 characters each) into the instrument. Sample IDs can be used to
specify the sample location or other sample specific information.
4. Press OK to confirm.
24
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Option 1
In the result display, press Options>More>Instrument Setup. If a
Sample ID is already assigned, select the Sample ID symbol in the
result display.
Option 2
With an USB keyboard (with US keyboard layout) or a scanner a
Sample ID can be directly entered in the result display.
25
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
4. Press the field below Save to select the lamp burning time.
26
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: After this period of time the lamp automatically turns off, after no
measurement has been made outside the corresponding range (UV/VIS).
6. Touch OK to confirm.
Note: The selected program that requires a lamp has highest priority. If the
lamp in the Lamp Control was turned off, the lamp will be turned on
automatically if it is needed for the stored program.
27
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
When the lamp is warmed up and ready, the UV icon stops flashing.
Note: If both lamps are On, the icon UV-VIS is displayed in the selected
Measurement Mode.
Note: Alternatively you can touch the UV- or VIS-Lamp icon on the
measurement screen to change the actual setting.
28
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
2. Press Printer.
29
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Printer Setup:
• Resolution: Print quality
• Paper: Paper size
Note: If an optional Thermal Printer is connected, the function "Auto Send"
on/off is available.
Note: The option Auto-Send is not available for any other printer (e.g. ink
jet printer).
Select between
• 100 dpi
• 150 dpi and
• 300 dpi
6. Press OK to confirm.
30
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Select between
• Letter
• Legal
• Executive
• A4
8. Press OK to confirm.
2. Select the data source, where the data to be printed are stored.
4. Press the PC & Printer icon to send the data (table, curve)
immediately to the printer.
2. Select PC.
31
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: The remote function is only for monitoring the data transfer.
5.2.7 Password
The Password menu contains a variety of security settings to
control access to various functions. For example, prevent
unauthorized changes to stored programs or instrument
configurations.
32
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
33
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: Use this function to delete the former Password or to enter a new
one.
34
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
35
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
• Sample ID
• Operator ID
• Start Date
• Parameter
2. Press Recall Data from the Main Menu. Press Options and
then the PC & Printer icon
3. Select the data to send to the memory stick and press OK.
36
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
37
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
5.3.2 Store, recall, send and delete data from wavelength scan and time course
The instrument can store 20 Wavelength Scans and 20 Time
Course Data sets.The data can be stored manually at the user's
discretion after viewing the data.
38
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Option 1:
2. Press Options and then the PC & Printer icon to send the
data to a USB memory stick, to a printer or to a PC with Hach
Data Trans.
39
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: The advice ”Data already exist. Overwrite?“ appears when the files
were already stored. Press OK to overwrite the stored data.
Option 2:
40
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
1. Press Recall Data from the Main Menu and then Wavelength
Scan or Time Course or Options>More>Recall Data.
41
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: Select the program number by name or use the arrow keys to
scroll through the list quickly and highlight the program or press
Select by number to search for a specific program number. Use the
alphanumeric keypad to enter the test number and press OK.
42
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
43
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
3. To start the next timed activity for the Stored Program, press
the Timer icon and OK.
Note: Press Close to view the measurement screen while the timer is
running. The time will be shown left side bottom instead of the date.
Note: Press Cancel to stop the timer midway through the countdown.
Note: The Timer will beep when the end of the running time is reached.
2. Select the required mode, then press OK, then Return to return
to the result display.
44
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: The Read key is only active after the zero measurement has
been carried out.
Note: In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.
3. Press Options and then the Store icon to store the displayed
data in the Data Log.
2. Press OK to confirm.
45
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: When a dilution is in effect, the dilution icon will appear on the
display.
Note: If using undiluted samples, set the dilution factor off again.
46
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: The adjustment must be within certain limits, which vary with
each program. The allowable percentage is shown after "Adjustment“.
Note: The selected Chemical Form will appear on the display. Test
results will be calculated and displayed in this chemical form.
47
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
7. Press OK.
Note: The results calculated using the reagent blank value must lie within
the limits of the measuring range of the test/method.
Note: The reagent blank icon is shown in the result display (see arrow)
when the function is active.
48
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
3. Press Zero.
49
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
50
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Some tests can be used for different measuring ranges. After the
sample cuvette/vial has been inserted, a list of the different
measuring ranges is displayed.
Select the required measuring range by highlighting the appropriate
line.
Press Start Permanent if this measuring range is to apply to all
subsequent measurements.
51
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: The selected chemical form is displayed, but does not become
the default. To change the default, see section 5.5.3.1.
52
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Save as User Program To store the selected parameters as a User Program, see section 6.1 on page 61.
Call up saved measurement data, wavelength scans or time courses, see section 5.3 on
Recall Data
page 35.
Instrument Setup Basic data of the instrument, see section 5.2 on page 23.
Turbidity and color in the sample matrix can falsify the results of a
photometric analysis. The interference factors come from the
sample or are created by reactions with the reagents.
The influence of turbidity and/or color can be eliminated or reduced
by taking a sample blank reading.
In the barcode mode, a special cuvette/vial (TNT919) containing
the sample blank is placed in cell compartment #1 after the sample
reading has been taken and is automatically measured. The
sample reading is then corrected by adding or subtracting the blank
value. The final result is displayed, with the message "After blank
value corr.".
Some barcode tests do not require a sample blank value to be
determined, as turbidity and color are dealt with during the test
53
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Version An abbreviation assigned by the user or the version number, etc. is entered here
Exact definition of the test: the number of wavelengths at which measurements are carried out,
Measurement process the number of absorbance measurements needed, the keys to be used, any waiting periods
between measurements, etc.
Formula Definition of the evaluation formulas with which the test result is calculated
The number of variables shown in the display depends on the definition of the measurement
Variables process and the formulas. Input of the numerical values of the wavelengths, factors, constants,
etc.
This function can be used to enter abbreviations and defined times for up to four timers. Highlight
the appropriate line and press Edit.
Timer 1, Timer 2,
The timers are activated or deactivated with the control boxes on the left of the display. In the
Timer 3, Timer 4
next column, a selection can be made from a list of names that describe the corresponding work
step. In the third column, the times for each active timer are entered.
1. Highlight the line containing Name and press Edit. Use the
alphanumeric keypad to enter the name indicated in the
working procedure.
2. Press OK to confirm.
54
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
1. Highlight the line containing Version and press Edit. Use the
alphanumeric keypad to enter the name indicated in the
working procedure.
2. Press OK to confirm.
1. Highlight the Formula line in the data overview and press Edit.
55
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
λ1: Wavelength 1
U1: Conversion Factor 1 for the first chemical form
U2: Conversion Factor 2 for additional chemical forms etc.
56
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
3. Press Options>More>Edit.
57
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
1. Go to http://www.hach.com.
58
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.
59
Downloaded from www.Manualslib.com manuals search engine
Standard Operations
60
Downloaded from www.Manualslib.com manuals search engine
Section 6 Advanced Operations
61
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Program Number:
Specific test number, with which the program can subsequently be
called up from the selection list in the User Programs menu or the
Favorites menu.
3. Press OK.
Program Name:
Program Type:
62
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Units:
Select the required unit from the list and press Next.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then New.
63
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Chemical form:
Enter the chemical formula used in the display to represent the
analysis parameter.
Use the alphanumeric keypad to enter the chemical form and press
Next to enter Calibration settings.
Units:
Select the required unit from the list and press Next.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then New.
64
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Wavelength λx:
Press a λx key and use the alphanumeric keypad to enter a
wavelength. Press another λx key and enter the next wavelength. If
necessary, repeat until all the wavelengths for the formula have
been entered. The wavelengths must be in the range from
190–1100 nm. Press OK.
Concentration Factor Kx
Multiplication factor for converting absorbance values into
concentration values.
Press a factor key and use the alphanumeric keypad to enter a
factor. If the formula includes more than one factor, press another
factor key and enter another factor and repeat until all the factors
have been entered. Press OK. When all the relevant data have
been entered, press Next.
Note: Up to five digits can be entered, including a maximum of 4 decimal
places to the right of the decimal.
65
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Chemical form:
Enter the chemical formula used in the display to represent the
analysis parameter.
Use the alphanumeric keypad to enter the chemical form and press
Next to enter Calibration settings.
66
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: The correlation coefficient (r2) is shown on the left below the
axes.
6. Press Force 0 to change the setting from Off to On. The curve
now passes through the origin of the coordinate system.
8. When the table has been completed and the curve type has
been chosen, press Done when the graph is displayed or Exit
when the table is displayed. Go to section 6.1.1.4 on page 70.
67
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
3. Press the "+" symbol again (see arrow) and enter the next
standard concentration. Repeat this sequence until all standard
conentrations (maximum of 24 solutions) have been entered.
5. Insert the zero solution into the cell compartment. Press Zero.
Repeat this sequence until all the standard solutions have been
measured (maximum of 24 solutions).
7. When the data have all been entered and the measurements
have all been completed, press Graph.
68
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
9. Press Force 0 to change the setting from Off to On. The curve
then passes through the origin of the coordinate system.
11. When the table has been completed and the curve type has
been chosen, press Done when the graph is displayed or Exit
when the table is displayed. Go to section 6.1.1.4 on page 70.
Note: The coefficients can have 5 digits and can have a positive or a
negative sign.
69
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
2. Set On and press the 0.000 key to enter the measuring range
limit. Confirm the entry by pressing OK.
70
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
3. Press the left key to enter another chemical form with the
alphanumeric keypad and press OK to confirm. Press the right
key to enter the conversion factor to calculate the concentration
of the additional chemical form from the concentration of
Chemical Form 1 and press OK to confirm.
71
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
72
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
73
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
{ } key
Elements of the measurement sequence that are to be repeated
are placed in brackets.
The left bracket "{"marks the start of the sequence that is to be
repeated and the right bracket "}" marks the end.
Note: The key showing the right bracket remains inactive until a left bracket
is entered.
1. Press {.
74
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
1. Highlight the Formula line in the data overview and press Edit.
3. Select C1: Off again and press Edit. The display changes to
C1: On.
75
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: The arrow key deletes the most recently entered element of the
formula.
76
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
6. Select C2: Off again and press Edit. The display switches to
C2: On.
77
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: When the first Cn formula has been defined, the Formula list is
automatically extended by Cn+1.
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
λ1: Wavelength 1
U1: Conversion Factor 1 for the first chemical form
U2: Conversion Factor 2 for the further chemical form etc.
1. Press Store to save the entered data. The data can be stored
under any data point (Measurement sequence, Formula, Timer,
etc.).
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
78
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: Use the scroll bar to scroll through the list quickly.
6.1.4 Add, edit and delete user programs from the favorites list
The most frequently used tests/methods in the User Program menu
can also be added to the list of favorites to simplify their selection.
Note: Use the scroll bar to scroll through the list quickly. Use the
alphanumeric keypad to enter the test number (program number) and
press OK to confirm.
79
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
6.1.4.2 Edit
6.1.4.3 Delete
80
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: Use the scroll bar to scroll through the list quickly.
Note: Press Select by Number to search for the program by number. Use
the alphanumeric keypad to enter the test number (program number) and
confirm the input by pressing OK.
3. Press Start.
Note: Use the scroll bar to scroll through the list quickly.
Note: Press Select by Number to search for the program by number. Use
the alphanumeric keypad to enter the test number (program number) and
confirm the input by pressing OK.
81
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Method:
A defined amount (concentration) of a standard solution of the test
substance is added to the sample. The detection rate should be
close to 100%.
Checklist:
82
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Sample volume:
A defined volume of a standard solution of known concentration is
added to the volume of sample specified in the procedure and the
sample is measured after each addition. In most cases three
different standard solutions are prepared and the procedure is
repeated for each of them.
Note: The units and chemical forms used for the sample are used for the
standard solutions. Be careful to ensure that to use the correct units for
subsequent entries.
83
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
2. Press Start.
84
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
6. Press the key of the value to change it. Use the alphanumeric
keypad to change the value. Press OK to confirm.
7. Press the keys to enter the standard addition volumes. Use the
alphanumeric keypad to enter the new data and press OK.
10. Use the arrow keys to select the first standard addition volume
in the table and insert the cuvette/cell with the corresponding
volume of added standard into the cell compartment.
Press Read.
Repeat the procedure from point 8 with all the other standard
addition solutions.
85
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
11. After all the standard addition solutions have been measured,
press Graph.
Note: In the curve menu, the name on the Curve key switches to
Table. Press Table to display all the data in the table again.
12. Press Ideal line to display the relationship between the added
standard solutions and the ideal line (detection rate 100%).
86
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Concentration
To select the position of the decimal point in the calculated concentration readings.
Resolution
87
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Concentration factor:
3. Press OK to confirm.
Concentration resolution:
88
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Reading mode:
2. Select the required mode, then press OK, then Return to return
to the result display.
Note: The Read key is only active after the zero measurement has
been completed.
Note: In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.
3. Press Options and then the Store icon to store the displayed
data in the Data Log.
89
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Concentration
To select the position of the decimal point in the calculated concentration readings.
Resolution
Save as User Program To store the selected parameters as a User Program, see section 6.1 on page 61
90
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
λ / Absorbance formula:
K1A1+K2A2
K1A1+K2A2+K3A3
K1A1+K2A2+K3A3+K4A4
K1A1/K2A2
(K 1 A 1 + K 2 A 2 ) / K 3 A 3
(K 1 A 1 + K 2 A 2 ) / (K 3 A 3 + K 4 A 4 )
91
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Concentration factor:
3. Press OK to confirm.
Concentration resolution:
92
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: The Read key does not become active until the zero
measurement has been completed.
Note: Zero and Read are disabled until the cell compartment is
closed.
93
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Integral: On/Off The integral gives the area and the derivative of the integral gives the original function
94
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
λ Setting wavelength
2. Press the upper left key to open the numeric keypad and select
the minimum wavelength. Press OK to confirm.
3. Press the upper right key to open the numeric keypad and
select the maximum wavelength. Press OK to confirm.
Note: Do not select the same wavelength for minimum and maximum.
Note: Scan recordings of high resolution data take a longer time than
recordings of low resolution data.Selecting a larger step allows the
instrument to scan faster, but decreases the resolution of the collected
data.
95
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Cursor mode
2. The selection for this menu item determines what data are
displayed in the table. Highlight Track or Peak/Valley.
3. Press OK to confirm.
Integral
The Integral applies to the whole wavelength range of the scan.
3. Press OK to confirm.
96
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: For the next scan measurement the setting for the Integral will be
On.
Note: If manual scaling is selected, use the alphanumeric keypad to set the
limits ymin. and ymax.. The graph is adjusted to display only the values in the
selected range. If automatic scaling is selected, the instrument sets the
limits automatically so that the total range can be displayed.
4. Press OK.
97
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Curve Icon Choice of Cursor Mode Peak/Valley (cursor moves between minimum/maximum absorbance
(Choice of Cursor Mode) values) or Cursor Mode Tracking (cursor moves over each data point of the scan).
The arrow keys are used (right/left) to move the cursor (depending on the selected mode) to the
next data point. The data of the data point (wavelength/absorbance or transmittance value) are
Arrow keys highlighted on the x and y axes.
Note: Press any point on the curve to display the associated data.
This function is used to magnify the section of the curve in the vicinity of the cursor. The original
Zoom Icon
curve size can be restored by pressing the zoom icon again.
98
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Option 1:
Note: After selecting a reference scan the Reference: Off key in the
Options menu turns into Reference: On.
Note: Only scans that have the same wavelength range and step can
be displayed using the overlay option. A reference curve can only be
displayed with another active curve. Two previously stored scans
cannot be overlaid.
Note: A black and an gray box are shown in the left corner of the
display. The gray box relates to the reference scan and the black one
relates to the current wavelength scan.
4. Press the black or gray small box in the left upper corner on the
screen to switch between the actual wavelength scan and
reference scan.
99
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Option 2:
Note: After selecting a reference scan the Reference: Off key in the
Options menu turns into Reference: On.
Note: Only scans that have the same wavelength range and step can
be displayed using the overlay option. A reference curve can only be
displayed with another active curve. Two previously stored scans
cannot be overlaid.
5. Press the black or gray small box in the left upper corner on the
screen to switch between the actual wavelength scan and
reference scan.
100
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
101
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
2. Input the total time and the reading time and press OK to
confirm.
Note: In total 500 measuring steps are possible. To select a total time and
a time interval that would cause this number of measurements to be
exceeded, the time interval is defined automatically and the OK key is
inactivated.
Note: If manual scaling is selected, use the alphanumeric keypad to set the
limits ymin. and ymax.. The graph is adjusted to display only the values in the
selected range. If automatic scaling is selected, the instrument sets the
limits automatically so that the total range can be displayed.
4. Press OK to confirm.
Note: During the measurement the Zero and Read keys change to
Mark and Stop.
102
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
• Select Mark to mark the next data point collected. This mark is
not used by the instrument, but is available for the user and
may indicate a significant event, such as the addition of a
sample or other reagent. The mark is also shown in the table.
• Select Stop to stop taking sample readings.
103
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
104
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
105
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
1. Go to http://www.hach.com.
Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.
106
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
4. Press Start.
Note: During the warm-up phase of the UV lamp, the message ”Lamp
Warmup...“ is displayed.
107
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
4. Press Zero.
108
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
6. Press Start.
109
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
4. Press Zero.
110
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
111
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
4. Press Start.
112
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
This test takes a reading every minute for one hour. Every
15 minutes, linear regression is used to calculate the slope
(rate of change) for the previous 15 minute interval.
6. The last reading and slope for each 15 minute interval are
displayed. At the end of the hour, the overall values are
calculated and displayed. The test proceeds to completion.
113
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
3. Press Edit.
• Stray Light
• Photometrical accuracy
• Wavelength accuracy
5. Press Verification.
114
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
8. Insert the different filter in the given order one after the other
with the filter glass facing the front of the instrument. Press
Next after inserting a filter and closing the cell compartment
After a lamp is replaced and the Lamp History is reset, the display
of the total operating time is reset to 0.
115
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
2. Select On and then Last Service to enter the date of the last
inspection.
3. Press OK to confirm.
5. Press OK to confirm.
116
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: If the USB stick is not connected, the message "Please insert USB
Memory" is displayed. Connect a USB stick, in order to store the data.
Press OK to confirm and press Store again.
117
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: If the Backup was already stored before, the message "Data already
exists. Overwrite?" is displayed. Press OK to overwrite the data.
118
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
Note: If the USB stick is not connected, the message "Please insert USB
Memory" is displayed. Connect a USB stick, in order to store the data.
Press OK to confirm and press Restore again.
119
Downloaded from www.Manualslib.com manuals search engine
Advanced Operations
120
Downloaded from www.Manualslib.com manuals search engine
Section 7 Maintenance
CAUTION
Potential Chemical, Biological Eye and Skin Hazards.
Only qualified personnel should conduct the tasks described
in this section of the manual.
7.1.1 Spectrophotometer
• Clean the enclosure, cuvette/sample cell compartments and all
accessories with a soft damp cloth. A mild soap solution can
also be used. Do not get excess water in the cuvette/sample
cell compartments. Do not insert a brush or sharp object into
Cell Compartment #1 to avoid damaging the mechanical
components.
• Dry the cleaned parts carefully with a soft cotton cloth.
7.1.2 Display
• Take care not to scratch the display. Do not touch the screen
with ball pens, pencils or similar pointed objects.
• Clean the display with a soft, lint-free and oil-free cotton cloth.
Diluted window cleaner liquid can also be used.
CAUTION
Potential Chemical/ Biological Exposure Hazards.
Use proper laboratory practices whenever there is a risk of
chemical exposure.
121
Downloaded from www.Manualslib.com manuals search engine
Maintenance
The Lamp compartment is on the left side behind the display and is
provided with ventilation on the back side. The tungsten and
deuterium (UV) lamp are installed in the lamp compartment. On the
back side a fan is installed for cooling of electric components.
The ventilation system operates automatically.
Refer to section 7.2.1 on page 122, section 7.2.2 on page 123 and
Figure 8 on page 124 for instructions on replacing the tungsten
lamp and deuterium lamp (UV).
DANGER
The instrument must remain switched off until the change of
lamp has been completed.
WARNING
Burn Hazard. Wait until the lamp cools down. Contact with the
hot lamp can cause burns.
4. Place the cover and the attached fan carefully beside the
instrument (take special care with the fan cable).
Important Note: The lamp should only be held by the lamp socket.
Avoid touching the glass.
8. Insert the tungsten lamp into the lamp compartment. Push the
spring down.
9. Check that the spring and the lamp socket are positioned
correctly.
10. Use a screwdriver to screw the back cover onto the instrument
again.
122
Downloaded from www.Manualslib.com manuals search engine
Maintenance
13. Reset the Lamp History, see section 6.8.5 on page 115.
DANGER
The instrument must remain switched off until the change of
lamp has been completed.
WARNING
Burn Hazard. Wait until the lamp cools down. Contact with the
hot lamp can cause burns.
4. Place the cover and the attached fan carefully beside the
instrument (take special care with the fan cable).
7. Holding the lamp socket, lift the deuterium lamp up out of the
lamp compartment (remove the complete unit, including the
cable).
Important Note: Do not touch the glass envelope on the new lamp.
If it is touched, clean with alcohol.
10. Insert the deuterium lamp in the socket so that the safety
contact clicks into place.
11. Use a screwdriver to screw the back cover onto the instrument
again.
14. Reset the Lamp History, see section 6.8.5 on page 115.
123
Downloaded from www.Manualslib.com manuals search engine
Maintenance
DANGER
Failed fuses are generally an indication of a problem with the
equipment. If fuse failure continues, contact a service
representative for instructions on how to return the equipment
for repair. Do not attempt to repair the equipment yourself.
124
Downloaded from www.Manualslib.com manuals search engine
Maintenance
3. Pull out the plastic base over the power cable socket.
4. Lift the instrument and check the color of the filter mat. Replace
the filter mat if is dark gray or black.
8. Remove the old air filter pad and replace it with a new air filter
pad (Figure 10, item 2).
125
Downloaded from www.Manualslib.com manuals search engine
Maintenance
126
Downloaded from www.Manualslib.com manuals search engine
Section 8 Troubleshooting
Error The cuvette is soiled or there are Clean the cuvette; allow the particles to
Clean Cuvette undissolved particles in the cuvette settle
Negative result! The calculated result is negative Check the concentration of the sample
Check the programming
Error in the test database / user
No evaluation! Contact the manufacturer or a sales
database
representative
The measured absorbance is above the Dilute the sample and repeat the
Over measuring range
calibration range of the test measurement
If possible, select a test with a lower
The measured absorbance is below the
Under measuring range measurement range or use a cuvette
calibration range of the test
with a longer path length
127
Downloaded from www.Manualslib.com manuals search engine
Troubleshooting
128
Downloaded from www.Manualslib.com manuals search engine
Section 9 Replacement Parts
Description Cat.-No.
Sipper Module DR 5000 LZV485
Peltier temperature controlled Module DR 5000 LZV513
Pour-Thru Module DR 5000 LZV479
Multi-Cell Holder DR 5000 A23618
Carousel Holder (Sample changer) A23620
USB-Keyboard (keyboard layout: US) LZV582
USB-Barcode Scanner (handset scanner for entering sample IDs) LZV566
Hach Data Trans (PC software for data transfer) LZY274
Brewery Analysis Software LZV659
Sipper Tubing, Inlet and Waste A23800
Sipper Tubing, PHARMED quality LZV580
Sipper Tubing, comes standard with Sipper Module LZV780
Pour Thru Tube Kit LZV569
USB-Memory Stick 2946900
USB-Interface Cable, (1 m) LZV567
Tungsten (VIS) Lamp A23778
Deuterium (UV) Lamp A23792
Fuse A23772
Filter Pad A23766
Certified test filter set for self-checks (Verification Kit)
LZV537
(6 precision glass filters with target values)
Semi-Micro Cell, 50 mm, Glass LZP269
Macro Cell, 20 mm, Glass LZP331
Sample Cell, 1 cm Square Glass, (10 mm path length) matched pair 2095100
Sample Cell, 1 cm Square Quartz, (10 mm path length) matched pair 4822800
Sample Cell, 50 mm Rectangular Glass 2629250
Sample Cell, 50 mm Rectangular Quartz 2624450
Sample Cell, 100 mm Rectangular Glass 2629201
Sample Cell, 100 mm Rectangular Quartz 2624401
Sample Cell, 1" Round Plastic, (1 cm path length) 10 mL, w/cap 4864302
Sample Cell, 1" Square Glass, 10 mL, matched sets 2495402
Sample Cell, 1" Square Glass, 10&25 mL, matched sets 2665902
Sample Cell, 1" Square Plastic 2410212
Pour Thru Cell, 1x1 cm (Quartz/160 µL/ Glass center height 10 mm) A24209
Pour Thru Cell, 1x1 cm (Quartz/450 µL/ Glass center height 10 mm) LZV510
Pour Thru Cell, 5x1 cm (Quartz/370 µL/ Glass center height 10 mm) LZV649
Pour Thru Cell, 1" Square (1" path length) 5913700
Adapter for 100 mm Sample Cell LZY421
Dust Cover LZV728
129
Downloaded from www.Manualslib.com manuals search engine
Replacement Parts
130
Downloaded from www.Manualslib.com manuals search engine
Section 10 Contact Information
HACH Company Repair Service in the Repair Service in Canada: Repair Service in
World Headquarters United States: Hach Sales & Service Latin America, the
P.O. Box 389 HACH Company Canada Ltd. Caribbean, the Far East,
Loveland, Colorado Ames Service 1313 Border Street, Unit 34 Indian Subcontinent, Africa,
80539-0389 U.S.A. 100 Dayton Avenue Winnipeg, Manitoba Europe, or the Middle East:
Tel (800) 227-HACH Ames, Iowa 50010 R3H 0X4 Hach Company World
(800) -227-4224 Tel (800) 227-4224 Tel (800) 665-7635 Headquarters,
(U.S.A. only) (U.S.A. only) (Canada only) P.O. Box 389
Fax (970) 669-2932 Fax (515) 232-3835 Tel (204) 632-5598 Loveland, Colorado,
[email protected] Fax (204) 694-5134 80539-0389 U.S.A.
www.hach.com [email protected] Tel +001 (970) 669-3050
Fax +001 (970) 669-2932
[email protected]
HACH LANGE GMBH HACH LANGE LTD HACH LANGE LTD HACH LANGE GMBH
Willstätterstraße 11 Pacific Way Unit 1, Chestnut Road Hütteldorferstr. 299/Top 6
D-40549 Düsseldorf Salford Western Industrial Estate A-1140 Wien
Tel. +49 (0)2 11 52 88-320 GB-Manchester, M50 1DL IRL-Dublin 12 Tel. +43 (0)1 9 12 16 92
Fax +49 (0)2 11 52 88-210 Tel. +44 (0)161 872 14 87 Tel. +353(0)1 46 02 5 22 Fax +43 (0)1 9 12 16 92-99
[email protected] Fax +44 (0)161 848 73 24 Fax +353(0)1 4 50 93 37 [email protected]
www.hach-lange.de [email protected] [email protected] www.hach-lange.at
www.hach-lange.co.uk www.hach-lange.ie
DR. BRUNO LANGE AG HACH LANGE FRANCE HACH LANGE SA DR. LANGE NEDERLAND
Juchstrasse 1 S.A.S. Motstraat 54 B.V.
CH-8604 Hegnau 33, Rue du Ballon B-2800 Mechelen Laan van Westroijen 2a
Tel. +41(0)44 9 45 66 10 F-93165 Noisy Le Grand Tél. +32 (0)15 42 35 00 NL-4003 AZ Tiel
Fax +41(0)44 9 45 66 76 Tél. +33 (0)1 48 15 68 70 Fax +32 (0)15 41 61 20 Tel. +31(0)344 63 11 30
[email protected] Fax +33 (0)1 48 15 80 00 [email protected] Fax +31(0)344 63 11 50
www.hach-lange.ch [email protected] www.hach-lange.be [email protected]
www.hach-lange.fr www.hach-lange.nl
HACH LANGE APS HACH LANGE AB HACH LANGE S.R.L. HACH LANGE S.L.U.
Åkandevej 21 Vinthundsvägen 159A Via Riccione, 14 Edif. Arteaga Centrum
DK-2700 Brønshøj SE-128 62 Sköndal I-20156 Milano C/Larrauri, 1C- 2ª Pl.
Tel. +45 36 77 29 11 Tel. +46 (0)8 7 98 05 00 Tel. +39 02 39 23 14-1 E-48160 Derio/Vizcaya
Fax +45 36 77 49 11 Fax +46 (0)8 7 98 05 30 Fax +39 02 39 23 14-39 Tel. +34 94 657 33 88
[email protected] [email protected] [email protected] Fax +34 94 657 33 97
www.hach-lange.dk www.hach-lange.se www.hach-lange.it [email protected]
www.hach-lange.es
HACH LANGE LDA HACH LANGE SP.ZO.O. HACH LANGE S.R.O. HACH LANGE S.R.O.
Av. do Forte nº8 ul. Opolska 143 a Lešanská 2a/1176 Roľnícka 21
Fracção M PL-52-013 Wrocław CZ-141 00 Praha 4 SK-831 07 Bratislava –
P-2790-072 Carnaxide Tel. +48 (0)71 342 10-83 Tel. +420 272 12 45 45 Vajnory
Tel. +351 214 253 420 Fax +48 (0)71 342 10-79 Fax +420 272 12 45 46 Tel. +421 (0)2 4820 9091
Fax +351 214 253 429 [email protected] [email protected] Fax +421 (0)2 4820 9093
[email protected] www.hach-lange.pl www.hach-lange.cz [email protected]
www.hach-lange.pt www.hach-lange.sk
HACH LANGE KFT. HACH LANGE S.R.L. HACH LANGE HACH LANGE SU
Hegyalja út 7-13. Str. Leonida, nr. 13 8, Kr. Sarafov str. ANALİZ SİSTEMLERİ
H-1016 Budapest Sector 2 BG-1164 Sofia LTD.ŞTİ.
Tel. +36 (06)1 225 7783 RO-020555 Bucuresti Tel. +359 (0)2 963 44 54 Hilal Mah. 75. Sokak
Fax +36 (06)1 225 7784 Tel. +40 (0) 21 201 92 43 Fax +359 (0)2 866 04 47 Arman Plaza No: 9/A
[email protected] Fax +40 (0) 21 201 92 43 [email protected] TR-06550 Çankaya/ANKARA
www.hach-lange.hu [email protected] www.hach-lange.bg Tel. +90 (0)312 440 98 98
www.hach-lange.ro Fax +90 (0)312 442 11 01
[email protected]
www.hach-lange.com.tr
131
Downloaded from www.Manualslib.com manuals search engine
Contact Information
132
Downloaded from www.Manualslib.com manuals search engine
Section 11 Limited Warranty
Hach Company warrants its products to the original purchaser against any defects that are due to faulty
material or workmanship for a period of one year from date of shipment unless otherwise noted in the
product manual.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its
option, it will repair or replace the defective product or refund the purchase price excluding original
shipping and handling charges. Any product repaired or replaced under this warranty will be warranted
only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable
components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned
without authorization from Hach Company.
Limitations
This warranty does not cover:
• Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared),
terrorism, civil strife or acts of any governmental jurisdiction
• Any product not used in accordance with the instructions furnished by Hach Company
This warranty contains the sole express warranty made by Hach Company in connection with its
products. All implied warranties, including without limitation, the warranties of merchantability and fitness
for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in
your state the above limitation may not apply to you. This warranty gives you specific rights, and you
may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person
is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive
remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in
no event shall Hach Company be liable for any incidental or consequential damages of any kind for
breach of warranty or negligence.
133
Downloaded from www.Manualslib.com manuals search engine
Limited Warranty
134
Downloaded from www.Manualslib.com manuals search engine
Index
A M
Absorbance Check ................................................. 111 Main Menu ............................................................... 22
Absorbance Formula ......................................... 90, 91 Maintenance .......................................................... 121
Alphanumeric keypad .............................................. 21 Mark ....................................................................... 103
Audio signals ........................................................... 25 Measurement Process .................................54, 55, 72
B Measuring sequence ................................................ 73
Barcode Programs .......................................51, 52, 53 Multi Wavelength ............... 22, 64, 66, 90, 91, 92, 93
edit ..................................................................... 53 N
Basic data ................................................................ 42 Name ....................................................................... 54
C Noise Check ........................................................... 108
Calibration formula ................................................... 75 O
Calibration Settings .................................................. 66 Operating environment ............................................ 13
Chemical Form ............................................43, 47, 53 Operator ID .............................................................. 23
Chemical form .................................................... 51, 52 creating .............................................................. 23
Cleaning delete ................................................................. 23
Display ............................................................. 121 Optical Checks ...... 106, 107, 108, 110, 111, 112, 113
Spectrophotometer .......................................... 121 Output Checks ........................................................115
Concentration Factor ............................ 87, 88, 90, 92 P
Concentration Resolution ..................... 87, 88, 90, 92 Password ................................................................. 32
Continuous Reading .......................................... 43, 45 activate .............................................................. 32
Cursor Mode ...................................................... 94, 96 PC and Printer ................................................... 29, 31
Cuvettes/cells ........................................................ 121 Print data .................................................................. 31
D Printer Setup ............................................................ 29
Data Printing data ............................................................. 31
delete ........................................................... 35, 37 Program ................................................................... 61
recall .................................................................. 35 Programm overview ................................................. 22
send ................................................................... 35 R
store .................................................................. 35 Reading Mode ....................................... 43, 44, 88, 89
Data Storage ............................................................ 38 Reagent Blank ................................................... 43, 48
Date ......................................................................... 25 Recall Data ................22, 38, 43, 53, 88, 91, 95, 101
Date and Time ......................................................... 25 Reference Scan ................................................. 94, 99
Dilution Factor ..............................................43, 45, 53 S
Display ............................................................. 21, 121 Safety Information ...................................................... 9
Display and Sound ................................................... 25 Sample blank ........................................................... 53
Drift Check ..............................................................112 Sample ID ................................................................ 24
F create ................................................................ 24
Factory Service .......................................................116 delete ................................................................. 24
Favorites ............................................... 22, 49, 80, 81 Save as User Program .......................... 43, 53, 88, 90
delete ........................................................... 80, 81 Scale & Units .................................... 95, 97, 101, 102
edit ..................................................................... 80 Security List ............................................................. 32
recall .................................................................. 81 Select View .............................................................. 96
Filter Settings ........................................................... 36 Send Data .....................................36, 43, 53, 94, 101
Formula .............................................................. 54, 55 Service Time ...........................................................116
Free programming ..........................72, 73, 75, 78, 80 Single Reading ......................................................... 43
H Single Wavelength .......22, 43, 45, 63, 66, 87, 88, 89
HACH Data Trans .............................................. 31, 36 Software ........................................................... 58, 106
Hazard Information .................................................... 9 Standard Addition .............................................. 43, 82
I Standard Adjust ................................................. 43, 46
Instrument Backup ..........................................117, 118 Stored Data .............................................................. 35
Instrument Information ........................................... 105 Stored Programs ..........22, 42, 43, 45, 46, 47, 48, 49
Instrument Setup .......22, 23, 43, 53, 88, 91, 95, 101 send Data .......................................................... 43
Integral ............................................................... 94, 96 Stray Light Check ....................................................110
Interfaces ................................................................. 29 System Check .........................................................116
Internet ............................................................. 58, 106 System Checks 22, 105, 107, 108, 110, 111, 112, 113,
L ................................................................ 115, 116, 117
Language ................................................................. 19
135
Downloaded from www.Manualslib.com manuals search engine
T
Time & Interval ............................................... 101, 102
Time Course ..........22, 38, 39, 40, 41, 101, 103, 104
Time Course Scan Reading ................................... 102
Timer .................................. 25, 43, 53, 54, 56, 87, 90
Touch Screen .......................................................... 21
U
Unpack the instrument ............................................. 13
Update .......................................................54, 58, 106
User Programs ................................................... 22, 61
V
Variables ......................................................54, 56, 78
Verification Kit .........................................................113
Version ............................................................... 54, 55
View Table ............................................................. 101
W
Warranty ................................................................ 133
Wavelength ........................................................ 87, 90
Wavelength Check ................................................. 107
Wavelength Range .................................................. 95
Wavelength Scan .........22, 38, 39, 40, 41, 94, 95, 98
Wavelength setting ................................................ 101
Wavelength spectrum .............................................. 94
136
Downloaded from www.Manualslib.com manuals search engine