GS12D08G02-00E
GS12D08G02-00E
General Conductivity
Specifications Detectors/Sensors
GS 12D08G02-E
F07.EPS
F07-1.EPS
SC8SG
SC210G
Conductivity
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 3
■ GENERAL SPECIFICATIONS Weight:
Sensors:
1. SC4AJ: Adapter mounting type
Cable with pin terminals (applicable to SC100, FLXA21, (SC4AJ-S-AD-09-002-05): approx.450 g
SC202G and SC202SJ) Adapter mounting type
Cable with M3 ring terminals (applicable to SC450G, (SC4AJ-S-AD-15-002-05): approx.520 g
SC202u/TB) Welding socket type
Cable with M4 ring terminals (applicable to FLXA21) (SC4AJ-S-SA-NN-002-05): approx.670 g
Object of measurement: 1 or 1.5 inch welding clamp type
Conductivity of solutions (SC4AJ-S-SB-NN-002-05): approx.550 g
Measuring principle : Two-electrode system 2 inch welding clamp type
Cell constant : 0.02cm-1, 0.1cm-1 (SC4AJ-S-SC-NN-002-05): approx.670 g
Measuring range : (Note) There are weight differences among SC4AJ
For a cell constant: 0.02cm-1 sensors. In order to know the more accurate weight
In case of the combination with the SC450G, of each type of sensors, please calculate it from
FLXA21,SC202G or SC202SJ: following information. The cable weighs 75 g/m. The
0 - 0.5 µS/cm to 0 - 200 µS/cm SC4AJ with 0.02cm-1 cell constant is 15 gram
In case of the combination with the SC100: heavier than the SC4AJ with 0.1cm-1 cell constant.
0 - 100 µS/cm to 0 - 1 mS/cm SUS314L electrode is 40 g heavier than Titanium
electrode.
(Material: Titanium only, SC100 can not use
Adapters:
with SC4AJ sensor made of SUS which cell
3/4NPT stainless steel adapter (/PS)
constant is 0.02 cm-1.)
:approx. 110 g
For a cell constant: 0.1cm-1
R3/4 stainless steel adapter (/RS):approx. 110 g
In case of the combination with the SC450G
3/4NPT PVDF adapter (/PF) : approx. 35 g
, FLXA21,SC202G or SC202SJ:
R3/4 PVDF adapter (/RF) : approx. 35 g
0 - 5 µS/cm to 1 mS/cm
Straight welding socket (/SA1) : approx. 300 g
In case of the combination with the SC100:
Angle welding socket 15 (/SA2) : approx. 320 g
0 - 500 µS/cm to 0 - 5 mS/cm
Welding clamp 1 inch (/SB1) : approx. 330 g
(Material: Titanium)
Welding clamp 1.5 inch (/SB2) : approx. 305 g
In case of the combination with the SC100:
Welding clamp 2 inch (/SC1) : approx. 350 g
0 - 500 µS/cm to 1 mS/cm (Material: SUS) (note) Do not submerge the sensor itself in process water,
Temperature Range: For electrode, 0 to 110°C as the seams between the mold and the metal of
For holder, see Figure 1 the sensor are not waterproof.
Sterilization for electrode:
135°C (275°F), within 30 minutes in 2. SC8SG:
Steam Sterilization Cable with pin terminals (applicable to FLXA21,SC202G
Pressure range : For electrode, 0 to 1 MPa and SC202SJ)
For holder, see Figure 1 Cable with M3 ring terminals (applicable to SC450G,)
MPa Cable with fork terminals (applicable to SC202G and
SC202u/TB)
1.2
Cable with M4 ring terminals (applicable to FLXA21)
1.0 Object of measurement:
0.8 Conductivity of liquids
0.6 SUS316L Measuring Principle: 2-electrode system or 4-electrode
0.4 PVDF system
0.2 Cell Constants : 0.01 cm-1 or 10 cm-1
(for two-electrode system)
0 10 20 30 40 50 60 70 80 100 120 140 10 cm-1 (for four-electrode
˚C system)
F10.EPS Measuring Ranges : 0 - 0.5 µS/cm to 0 - 100 µS/cm for
Figure 1 The range of tolerance of holders (option: /PS, a cell constant of 0.01 cm-1
/PF, /RS, /RF, /SA1, /SA2, /SB1, /SB2, /SC1) for 0 - 1 mS/cm to 0 - 1000 mS/cm
temperature and pressure for a cell constant of 10 cm-1
Temperature Range: 0˚ to 100˚C (130˚C maximum only
Sample solution condition: for 0.01 cm-1 cell constant
Although flow rate is not limited in detectors, excluding those with
measurement, air bubbles should not polypropylene chambers)
be mixed in the sample solutions to Pressure : 1000 kPa max. (500 kPa maxi-
obtain correct measured values. mum for detectors with polypropy-
Temperature sensor: Pt1000 lene chambers)
Materials Flow rate of Sample Solution:
Body & Electrode : SUS316L (for all Fitting-type) or No particular limitation applies,
Titanium (only for adapter mounting although a value of less than 20 l/
type-AD), Viton O-ring min. is recommended for flow-
Isolator : PEEK through detectors.
Mounting adapter : Plyvinylidene difluoride (for /PF and /
RF) or SUS316, SUS316L
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Nov.21, 2012-00
<<Contents>> <<Index>> 4
(Note) No limitation applies to flow rate (flow velocity) as far Diameter: 9.2 mm
as measurement is concerned. Take care, however, Material : Thermoplastic PVC
when using flow-through detectors. Electrodes or
the inner walls of a liquid chamber may wear put 4. SC210G:
drastically at higher flow speeds if a measured
solution contains slurry. Air bubbles should not be Cable with ring terminals (applicable to FLXA21,
mixed in the sample solutions to obtain correct SC202G/TB and SC202SJ/TB)
measured values. Cable with M3 ring terminals (applicable to SC450G,
RTD for Temperature Compensation: SC202u/TB)
Pt1000 (built into the sensor) Cable with pin terminals (applicable to FLXA21,SC202G
Construction : Rainproof encapsulation (compat- and SC202SJ)
ible with the JIS C0920 Japanese Cable with M4 ring terminals (applicable to FLXA21)
Industrial Standard) Object of measurement:
Weight : Conductivity of solutions
Screw-in type Measuring principle : Two-electrode system
approximately 1.3 kg (excluding the cable) Cell constant : 0.05 cm-1, 5 cm-1
Flow-through type (SCS14 chamber) Measuring range : 0 - 0.5 mS/cm to 0 - 200 mS/cm
approximately 3.1 kg (excluding the cable) (Cell constant: 0.05 cm-1)
Flow-through type (SCS14 chamber, flanged) 0 - 200 mS/cm to 0 - 20 mS/cm
approximately 4.5 kg (excluding the cable) (Cell constant: 5 cm-1)
Flow-through type (polypropylene chamber) Temperature Range: 0 to 105°C
approximately 2.7 kg (excluding the cable) (chamber material: SCS14)
Flow-through type (polypropylene chamber, flanged) 0 to 100°C
approximately 3.2 kg (excluding the cable) (chamber material:
Cable Polypropylene)
0.3 kg for 5.5-m length; 0.5 kg for 10-m Pressure range : 0 to 1 MPa
length; 0.9 kg for 20-m length. (chamber material: SCS14)
Process Connection: Screw-in or flow-through 0 to 500 kPa
Construction of Wetted Part: (chamber material: Polypropylene)
• Sensor-holding base: Measuring solution condition:
SUS316 and fluoro-rubber when using screw- Although flow rate is not limited in
in type holder or the chamber made of measurement, less than 20 l/min
stainless steel. PP and fluoro-rubber when is recommended for flow-through
using the chamber made of PP. type. If slurry is included in
• 0.01 cm-1 cell constant, two-electrode sensor: sample solutions in flow-through
SUS316 and ethylene chloride trifluoride type detectors, the electrode part
• 10 cm-1 cell constant, two-electrode sensor: and the inside of solution chamber
reinforced epoxy resin and graphite may be worn significantly.
• 10 cm-1 cell constant, four-electrode sensor: Air bubbles should not be mixed
polyvinylidene difluoride, glass and platinum in the sample solutions to obtain
• Stem (flow-through type): correct measured values.
SCS14 or polypropylene resin Temperature sensor: Thermistor (PB36NTC)
Installation : Wet part Materials
• Screw-in type—held by the process piping SC210G-A : For sensor, SUS 316 stainless
• Flow-through type (polypropylene chamber) steel, Viton (O-ring) and
—mounted on a pipe (nominal Polytrifluorochloroethylene
diameter of 50 mm ±2 in.) For body, SUS316 stainless steel,
• Flow-through type (SCS14 chamber) polypropylene and Viton (O-ring)
—held by the process piping SC210G-B : For sensor, Platinum, glass and
Viton (O-ring)
3. WU41: Dedicated cable for the SC8SG For body, SUS316 stainless steel,
polypropylene and Viton (O-ring)
Cable : Six multicore wire
Construction : JIS C0920 watertight (equal to
Applicable transmitter/converter with various detectors NEMA 4)
Converter:
Yes N.A. N.A. Yes N.A. N.A. Yes N.A. N.A.
SC402G (Note 3)
Converter: SC450G Note2 N.A. Yes Note2 N.A. Yes Note2 N.A. Yes
Analyzer: FLXA21 Yes Yes N.A. Yes Yes N.A. Yes Yes N.A.
Note1: Applicable when option code /TB (screw terminal) specified for SC202G/SC202SJ.
Note2: Both pin and M3 ring can be used for SC450G, but M3 ring are recommended.
Note3: SC402G has been terminated. T07-1.EPS
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 5
■ MODEL AND SUFFIX CODES
1. SC4AJ
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Nov.21, 2012-00
<<Contents>> <<Index>> 6
2. SC8SG
*1 : The cell constant is 0.01cm-1 when the combiration of measuring range R31 and Electrode configuration - T is chosen.
The cell constant is 10cm-1 when the combination of measuring range R61 and Electrode configuration - T is chosen.
*2 : Electrode configuration - F cannot be chosen when R31 is chosen. For process where can give contamination to a detector, a
four-electrode detector, the combination of R61 and - F, should be used.
*3 : If a welding socket (K9208BK) needs to be ordered beforehand, either place a separate order or prepare one by referring to the
extemal view later in this brochure.
*4 : Impossible use for the SC400G.
*5 : Used for connection to FLXA21. *6 : Used for connection to SC450G, SC202u/TB.
*7 : The model is not equipped with a mounting hardware, please place an order on the SC8SG with option code /PS or /SS
when you select flow-through model.
T03.EPS
The PP chamber can have cracks or splits unless it is not supported by a mounting hardware.
3. WU41
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 7
Spare Parts for SC8SG
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 8
Spare Parts for SC210G
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 9
■ DIMENSIONS
1. SC4AJ
Sensor SC4AJ-u-AD-09 Sensor SC4AJ-u-AD-15
Unit : mm Unit : mm
\19.05
152
\19.05
90
Pin-type Ring-type
Cable length (3m, 5m, 10m, 15m, 20m)
Approx. 350 Pin-type Ring-type
Cable length (3m, 5m, 10m, 15m, 20m)
Approx. 350
F12.EPS
F13.EPS
Unit : mm
Approx. 49
3/4NPT
or R3/4 *
O-ring Unit : mm
\24.9
\33.8
\38
95 5
130.5
Approx. 185
Pin-type Ring-type
Cable length (3m, 5m, 10m, 15m, 20m)
Approx. 350
51
158
8.5
Socket 8.5
et
55 thickness Sock ess
n
(56.3
F15.EPS
thick
)
f42 f42
(43.5) F16.EPS
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 10
Sensor SC4AJ-S-SB-NN
\19.05
Unit : mm
/SB1
70 6.35
12.7
89.7
Approx. 150
Pin-type Ring-type
Cable length (3m, 5m, 10m, 15m, 20m) φ25.4
Approx. 350
/SB2
F17.EPS
12.7
φ38.1
F18.EPS
Sensor SC4AJ-S-SC-NN
Unit : mm
\19.05
\64
/SC1 Unit : mm
80 6.35
115.7
Approx. 170
Pin-type Ring-type
12.7
Approx. 350
φ50.8
F20.EPS
F19.EPS
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 11
2. SC8SG
Screw-in Model (with welding socket)
Unit : mm
Pin terminals
Unit : mm
95 125
f30
f30
Unit : mm
Pin terminals
Locknuts
(M48 x 2) Fork terminals
Approx.78 Welding socket for Screw-in type (K9208BK)
3 22 27
221
185
3
C1 (5)
125 f42
f45
f50 60.2 Material: SUS316
f30 (Note) If you make the welding socket for screw-in type,
refer to the above drawing.
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 12
Flow-through type (Chamber Material: PP) + Mounting bracket (/PS) --Screw Jointed--
Unit : mm
Pin terminals
Fork terminals
Approx. 78
Mounting hardware
Option : /PS
Rc 1/2
Ring terminals
1/2NPT
40
203 87 99 50A pipe
305 (O.D. f60.5)
170
217
47 Rc 1/2 53
100
1/2NPT
Flow-through type (Chamber Material: PP) + Mounting bracket (/PS) --Flange Jointed--
Unit : mm
Pin terminals
Fork terminals
Approx. 78
4 - f15 holes
12 Mounting hardware
Option : /PS
Ring terminals
fC
fD
215
47
98
100
Flange rating fC fD
JIS 10K 15 FF 70 95
ANSI Class150 1/2 FF 60.5 88.9
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 13
Flow-through type (Chamber Material: SCS14) --Screw Jointed--
Unit : mm
Pin terminals
Approx. 78
Fork terminals
Locknuts
Ring terminals
Rc1/2 or
1/2NPT(F)
172
Pin terminals
Fork terminals
Approx. f58 Rc1/2 or
1/2NPT(F)
Approx. 78
4 - fh holes fG
Ring terminals
fC
fD
293
t
20
55
203
70
100
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 14
3. WU41 for SC8SG
Unit : mm
100
64.5 PG9 Approx. Pin terminal
White
Brown
Green
\ 29
Yellow
Black
Red
Cable lengh: 5.5 m, 10 m, 20 m
13
11
12 Earth
WU41e.eps
4. SC210G
Screw-in Type Flange Type
Clearance for maintenance access Unit : mm
91 82
H + 50
Unit : mm
Clearance for maintenance access
91 82
H + 60
G 1/2
Approx. Cable inlet
4 - f19 hole H + 172
G 1/2 t
Cable inlet
Mounting screw fC
R1-1/2 or 1-1/2NPT fD
Gage diameter H
Approx.
H + 184
Approx.
H H + 10 f30 26 28
SC210G-B SC210G-A
Sensor length code H
- L015 162
Flange rating fC fD t
- L030 312
f30 26 28 JIS 10K 50 RF 120 155 16
- L050 512
ANSI Class150 2 RF 120.7 152.4 19.1
- L100 1012
SC210G-B SC210G-A Sensor length : H = 150, 300, 500, 1000, 1500, 2000 JPI Class150 2 RF 120.6 152 19.5
- L150 1512
Note : ANSI flange with serrations.
- L200 2012
G 1/2
Cable inlet
G 1/2
Cable inlet 100
Rc1/2 1/2NPT
40 47 170
42
172 f58
58
Rc1/2
47
1/2NPT
Rc1/2 or 1/2 NPT 99 Inlet
JIS 50A pipe (O.D. f60.5)
Inlet
Upright/horizontal position
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 15
Flow-through type + Mounting hardware (/PS) Flow-though type
(screw connection, chamber material: polypropylene) (screw connection, chamber material: SCS14)
Clearance for maintenance access Unit : mm
Unit : mm
Clearance for maintenance access
91 82
91 82 210
210
G 1/2
G 1/2 Cable inlet
t
t Cable inlet
Approx. 100
419
Mounting Approx.
Hardware 4 - fh fD fC 409 Outlet
Outlet
Option /PS fD fC
4 - fh
75 205 f58
91
215
58
Inlet
Inlet
91 JIS 50A pipe (O.D. f60.5) Fange rating fC fD t fh
Upright/horizontal position JIS 10K 15 RF 70 95 12 15
Flange rating fC fD t fh ANSI Class150 1/2 RF 60.5 88.9 11.2 15.7
JIS 10K 15 FF 70 95 12 15 JPI Class150 1/2 RF 60.3 89 10.9 16
ANSI Class150 1/2 FF 60.5 88.9 12 15
(Note) ANSI flange is serration.
G 1/2
G 1/2
Cable inlet
Cable inlet
264
410
123
Mounting screw
R1-1/4 or 1-1/4NPT Gage diameter 123
155
89
Mounting screw
R1-1/4 or 1-1/4NPT
28 26 Gage diameter
199
(option: /SS)
Unit : mm
(Chamber of Conductivity detector) Ø30
approx. 101.8 50A pipe
(provided by the user)
20
"/ANSI"Adaptor of cable inlet
55 45
26
f15
1/2NPT
203
G1/2
70
100
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 16
■ WIRING DIAGRAM
11 11
Temperature Temperature
12 12
13 13
14 14
Electrode Electrode
15 15
16 16
11
Temperature
12
13
14
Electrode
15
16
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Sep.26, 2011-00
<<Contents>> <<Index>> 17
■ TABLE OF CORROSION-RESISTANT MATERIALS
Note: This table shows corrosion resistances against each specified chemical only. If two or more kinds of chemical are
mixed in a sample, the properties may be different from those shown in this table.
40
Nitric acid 10 20 10 30 10 30 10 100 Strong acid
80 25 60 Weak acid
Sulfuric acid 3 20 5 30 5 20 5 30
3 100 5 100 10 60 5 100
Phosphoric acid 30 60 15 30 5 30 5 30
30 100 5 b 25 100 5 60
Ammonia water 15 80 10 b 10 b 10 b
15 100 28 65 28 65 28 65
Caustic potash 10 b 10 60 10 b
25 b 25 b 25 b
Alkali
Strong alkali
Caustic soda 20 80 20 30 20 60 20 30
Weak alkali
20 100 20 b 20 b 20 b
Potassium carbonate 5 b 5 b 5 b
35 b 35 b 35 b
Sodium carbonate sat. 100 25 b 25 b 25 b
Zinc chloride 20 b 20 60 20 b
Aluminum chloride 25 25 10 b
25 25 25 b
Ammonium chloride 35 40 25 b 25 20 25 b
Potassium chloride sat. 60 sat. 60 sat. 60
Chlorides
resistance against
Sodium nitrate 50 b 50 b 50 b
all salts normally
Sodium sulfite used 20 b 20 b
Hydrogen peroxide 10 30 10 30 10 30
Sodium hypochlorite 10 90 2 60 to 90 2 60 to 90 15 30
20 80
Others
Potassium bichromate 10 b 10 20 10 b
Alcohol 96 70 100 b 80 60 80 100
Acetic acid 100 70 100 70 10 60 10 100
Phenol 100 20 95 30 100 20 100 20
Aromatic solvent 100 20 100 25 100 20 100
(Note) b: Shows temperatures up to the boiling point. PVDF: Polyvinylidene difluoride
T09.eps
CAUTION
Select the material of wetted parts with careful consideration of process characteristics. Inappropriate
selection may cause leakage of process fluids, which greatly affects facilities. Considerable care must be
taken particularly in the case of strongly corrosive process fluid such as hydrochloric acid, sulfuric acid,
hydrogen sulfide, and sodium hypochlorite. If you have any questions about the wetted part construction
of the product, be sure to contact Yokogawa.
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
<<Contents>> <<Index>> 18
■ ) and write
Thank you for inquiry about YOKOGAWA Conductivity Detector/Sensor. Please check (✔) the appropriate box (■
down the relevant information in the underlined blanks.
1. General Items
Name of your company:
Person in charge : Belongs to: (Phone No.: )
Name of plant :
Measuring point :
Purpose of use : ■ Indication ■ Record ■ Alarm ■ Control
Power supply : V AC, Hz
2. Measuring Conditions
(1) Liquid temperature : to , Normal [˚C]
(2) Liquid pressure : to , Normal [kPa]
(3) Flow rate : to , Normal [l/min.]
(4) Flow speed : to , Normal [m/s]
(5) Slurry or fouling components: ■ No ■ Yes
(6) Name of measuring liquid :
(7) Component of measuring liquid :
(8) Others :
3. Installing Location
(1) Ambient temperature :
(2) Installing location : ■ Outdoors ■ Indoors
(3) Others :
4. Specification Requirements
(1) Measuring Range :
(2) Transmission output : ■ 4 to 20 mA DC ■ 0 to 20 mA DC
(3) Detector/Sensor : SC4AJ ■ 2-electrode system (0.02 cm-1) ■ 2-electrode system (0.1 cm-1)
SC8SG ■ 2-electrode system (0.01 cm-1) ■ 2-electrode system (10 cm-1)
■ 4-electrode system (10 cm-1)
SC210G ■ 2-electrode system (0.05 cm-1) ■ 2-electrode system (5 cm-1)
(4) Mounting type : SC4AJ ■ Adapter mounting ■ Welding socket ■ Welding clamp
SC8SG ■ Screw-in ■ Flow-through
SC210G ■ Screw-in ■ Flange ■ Flow-through
■ Screw-in with gate valve
(5) Cable length : SC4AJ ■3m ■5m ■ 10 m ■ 15 m ■ 20 m
SC8SG ■ 5.5 m ■ 10 m ■ 20 m
SC210G ■ 3 m ■5m ■ 10 m ■ 15 m ■ 20 m
(6) Dedicated cable for SC8SG : WU41 ■ 5.5 m ■ 10 m ■ 20 m
(7) Others :
All Rights Reserved. Copyright © 2000, Yokogawa Electric Corporation GS 12D08G02-E 8th Edition Oct.08, 2010-00
Subject to change without notice