62faf8ad9cb0ad44e95aca9e AnyCheck 2017.10 Ver.2

The document provides instructions for using AnyCheck, an implant stability meter that measures the stiffness of the bone-implant interface through tapping. It highlights the importance of maintaining a contact angle of 0 to 30 degrees and outlines precautions to avoid measurement errors. The device indicates osseointegration levels with a range of values, helping to determine the appropriate timing for connecting prosthetics.

Uploaded by

prothesesalah62
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
10 views2 pages

62faf8ad9cb0ad44e95aca9e AnyCheck 2017.10 Ver.2

The document provides instructions for using AnyCheck, an implant stability meter that measures the stiffness of the bone-implant interface through tapping. It highlights the importance of maintaining a contact angle of 0 to 30 degrees and outlines precautions to avoid measurement errors. The device indicates osseointegration levels with a range of values, helping to determine the appropriate timing for connecting prosthetics.

Uploaded by

prothesesalah62
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Find Neobiotech online

Clinical Case Precautions

1 4
If the contact angle is off the
range (0~30 degree downward
based on the ground level),
the device will make a beeping 2017. 10 E-Ver.02
sound and button will not func-
tion. This is to reduce the error
occurred while measuring from Dental implant stability meter

2 5
various tapping angles. When it
makes a beeping sound, tilt the (tapping-motion) for assessing
tapping rod downward so that osseointegration without
it meets the range. disconnecting the healing
The tip of the tapping rod is abutment.
designed to contact the healing
abutment for the convenience

3 6 of users. Applying too much


pressure onto the healing abut-
ment or holding the device
too far away from the healing
abutment during measurement
may result in an error.

www.neobiotech.co.kr
Neobiotech Co., Ltd. E-space #1001, 36, Digital-ro 27-gil, Guro-gu, Seoul, Korea, 08381
Tel. +82-2-582-2885 Fax. +82-2-582-2883
AnyCheck Instructions
1. Turn on AnyCheck by pressing
LCD the switch button until it turns on.
AnyCheck is an implant stability meter that measures the stiffness of the alveolar bone-
implant interface through a tapping-motion. Screen 2. Slightly contact the tip of the tap-
ping rod to the upper part of the
AnyCheck reduces the risk of implant failure by indicating the most appropriate time for
healing abutment. Be cautious
connecting prosthetics. not to push the healing abutment.

0~30˚
3. Maintain the contact angle
between 0 to 30 degrees.

Characteristics ON/OFF and 4. Gently press the START button


while holding the device stable.
1. AnyCheck is easy to use due to its easy con- START Switch
trol system and small and light body design. 5. Check the measured value
displayed on the LCD screen.
2. Since AnyCheck is used without removing
the healing abutment, implant stability can 6. Range of the measured value is
between 30-85. The smaller the
be safely measured anytime right after the
measured value, the weaker the
implant placement. degree of osseointegration.
Tapping Rod
3. Measuring time is reduced to less than 3 Front Guide
seconds. The most benefit comes when
Measured value is similar to ISQ scale: above 60 – bridge
measuring multiple implants at a time. (In recommended; above 70 – single case loading possible.
case of weak osseointegration, tapping-
motion automatically stops after tapping Displayed in red – 30~59; orange – 60~64; green – 65~85.
twice. Otherwise, taps up to six times.) Measured values (IST scale) differ depending on the height of the
healing abutment. As the standard height of the healing abutment is
4. Light tapping method allows safe measure- 4mm, calculate the correct IST value according to the following chart:
ment of implant stability (30% reduced
tapping strength compared to product ‘P’)
Standard +3 7mm height Healing Abutment Displayed IST value +6
5. The degree of osseointegration is indicated Standard +2 6mm height Healing Abutment Displayed IST value +4
as IST(Initial Stability Test) value between 30 Tapping AnyCheck
Standard +1 5mm height Healing Abutment Displayed IST value +2
to 85. (Value used in product ‘O’)
Rod Body
Standard 4mm height Healing Abutment Displayed IST value
6. Failing implants may be detected in earlier Standard - 1 3mm height Healing Abutment Displayed IST value - 2
stages.
Standard - 2 2mm height Healing Abutment Displayed IST value - 4
Standard - 3 1mm height Healing Abutment Displayed IST value - 6