AQUANET International Ltd. offers a Cooling Water Analyser designed for continuous monitoring of nitrite levels in cooling water, featuring an LCD display and external data output options. The device operates on a 12-15 V DC power source, has a measuring range of 0-4500 mg/l, and includes a calibration potentiometer for accuracy adjustments. It is suitable for integration with dosing management systems and ships' engine room management systems.

Uploaded by

abdediyaa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
4 views

NitriteMonitor

AQUANET International Ltd. offers a Cooling Water Analyser designed for continuous monitoring of nitrite levels in cooling water, featuring an LCD display and external data output options. The device operates on a 12-15 V DC power source, has a measuring range of 0-4500 mg/l, and includes a calibration potentiometer for accuracy adjustments. It is suitable for integration with dosing management systems and ships' engine room management systems.

Uploaded by

abdediyaa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

AQUANET International Ltd.

Specialist Instrumentation and Software Engineers


for the Automated Water Quality Control

Cooling Water Analyser


DESCRIPTION
The instrument comprises a body fitted with internal sensing components and a
temperature sensor.
6 mm stainless steel couplings are provided so that a sample of the cooling water can
pass continuously through the body for analysis of nitrite level.

Results of the analysis are displayed on the LCD display and the data is available
externally as a 4-20mA signal for use by a dosing management system such as an
Aquanet AQ652, AQ655 or for inclusion into a ships engine room management system.

Data is also available in raw digital format for those companies wishing to write their
own protocol handler for OEM software applications.

Calibration:
The calibration potentiometer is available through the front panel and following a cooling
water sample drop test the potentiometer can be used to adjust the LCD to agree with
the test results.

The LCD continuously displays the level of Nitrite and the sensor temperature.

Power Source:
The user provides an external 12-15 V dc power cube

There are no internal user serviceable parts.

SPECIFICATIONS

Voltage 12 - 15 volts DC @ 250mA 4-20 mA signal 0 –3000 mg/l (as Nitrite)

Cables (The unit has 3 wire connections) Alarms Instrument status via LCD
Red +12– 15Vdc
Green 0 Vdc
Yellow 4-20mA output
Enclosure
Rating IP 65
Connections Double Ferrule 6mm Stainless compression Dimensions 130 x 56 x 50 mm

Measuring Range 0-4500 mg/l


Environmental
Accuracy +/- 5% FSD Operational temperature 5 to 60°C
Storage temperature 0 to 35°C
Storage RH 0-95% non condensing
Temperature 0-60°C
(Requires a cooler for higher temperatures)

Controls Calibration Potentiometer

Note: All specifications may be subject to change without notice

Flow Rate 0.5—1.5 litres/minute

AQUANET International Ltd., 8 The Galloway Centre,


Hambridge Lane, Newbury, Berkshire, UK RG14
5TL
www.aquanetint.com
Tel +44 (0)1635 580680, e-mail tech@aquanetint.
Sample Applications
Automated dosing and control with AQ652 Cooling Water management system linked to other Aquanet sys-
tems AQPro Software with output to ships management system.

Simple monitor with manual dosing and output signal optionally connected to ships computer management system

AQUANET International Ltd., 8 The Galloway Centre,


Hambridge Lane, Newbury, Berkshire, UK RG14
5TL
www.aquanetint.com