lecture_notes_4th_sem_em&i_1679445846 (1)
lecture_notes_4th_sem_em&i_1679445846 (1)
ON
(Th. 3)
Name of the course: Diploma in Electrical Engineering.
(4th Semester)
ENGINEERING, ROURKELA.
2|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
CHAPTER 1
MEASURING INSTRUMENTS
MEASURING INSTRUMENTS
INTRODUCTION
Measurement is the act, or the result, of a quantitative comparison between a given quantity
and a quantity of the same kind chosen as a unit. The result of the measurement is expressed
by a pointer deflection over a predefined scale or a number representing the ratio between
the unknown quantity and the standard.
A standard is defined as the physical personification of the unit of measurement or its sub-
multiple or multiple values.
The device or instrument used for comparing the unknown quantity with the unit of
measurement or a standard quantity is called a measuring instrument.
The value of the unknown quantity can be measured by direct or indirect methods.
In direct measurement methods, the unknown quantity is measured directly instead of
comparing it with a standard. Examples of direct measurement are current by ammeter,
voltage by voltmeter, resistance by ohmmeter, power by wattmeter, etc.
In indirect measurement methods, the value of the unknown quantity is determined by
measuring the functionally related quantity and calculating the desired quantity rather than
measuring it directly. Suppose the resistance as (R) of a conductor can be measured by
measuring the voltage drop across the conductor and dividing the voltage (V) by the current
Accuracy:
Accuracy is the closeness with which the instrument reading approaches the true value of the variable
under measurement. Accuracy is determined as the maximum amount by which the result differs
from the true value. It is almost impossible to determine experimentally the true value. The true value
is not indicated by any measurement system due to the loading effect, lags and mechanical problems
(e.g., wear, hysteresis, noise, etc.).
Accuracy of the measured signal depends upon the following factors:
Intrinsic accuracy of the instrument itself;
Accuracy of the observer;
Variation of the signal to be measured; and
Whether or not the quantity is being truly impressed upon the instrument.
Precision:
Precision is a measure of the reproducibility of the measurements, i.e., precision is a measure of the
degree to which successive measurements differ from one another. Precision is indicated from the
number of significant figures in which it is expressed. Significant figures actually convey the
information regarding the magnitude and the measurement precision of a quantity. More significant
figures imply greater precision of the measurement
3|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Resolution:
If the input is slowly increased from some arbitrary value it will be noticed that the output does not
change at all until the increment exceeds a certain value called the resolution or discrimination of the
instrument. Thus, the resolution or discrimination of any instrument is the smallest change in the
input signal (quantity under measurement) which can be detected by the instrument. It may be
expressed as an accrual value or as a fraction or percentage of the full scale value. Resolution is
sometimes referred as sensitivity. The largest change of input quantity for which there is no output
of the instrument is called the dead zone of that instrument.
The sensitivity gives the relation between the input signal to an instrument or a part of the instrument
system and the output. Thus, the sensitivity is defined as the ratio of output signal or response of the
instrument to a change of input signal or the quantity under measurement.
Speed of Response
The quickness of an instrument to read the measurand variable is called the speed of response.
Alternately, speed of response is defined as the time elapsed between the start of the measurement
to the reading taken. This time depends upon the mechanical moving system, friction, etc.
MEASUREMENT OF ERRORS
In practice, it is impossible to measure the exact value of the measurand. There is always some
difference between the measured value and the absolute or true value of the unknown quantity
(measurand), which may be very small or may be large. The difference between the true or exact
value and the measured value of the unknown quantity is known as the absolute error of the
measurement.
If δA be the absolute error of the measurement, Am and A be the measured and absolute value of the
unknown quantity then δA may be expressed as
When the absolute error ε0 (=δA) is negligible, i.e., when the difference between the true value A and
the measured value Am of the unknown quantity is very small or negligible then the relative error may
be expressed as,
The relative error is generally expressed as a fraction, i.e., 5 parts in 1000 or in percentage value,
4|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
CLASSIFICATION OF INSTRUMENTS
Absolute Instruments
The instruments of this type give the value of the measurand in terms of instrument constant and its
deflection. Such instruments do not require comparison with any other standard. The example of this
type of instrument is tangent galvanometer, which gives the value of the current to be measured in
terms of tangent of the angle of deflection produced, the horizontal component of the earth’s
magnetic field, the radius and the number of turns of the wire used. Rayleigh current balance and
absolute electrometer are other examples of absolute instruments. Absolute instruments are mostly
used in standard laboratories and in similar institutions as standardizing.
Secondary Instruments
These instruments are so constructed that the deflection of such instruments gives the magnitude of
the electrical quantity to be measured directly. These instruments are required to be calibrated by
comparison with either an absolute instrument or with another secondary instrument, which has
already been calibrated before the use. These instruments are generally used in practice.
Secondary instruments are further classified as
Indicating instruments
Integrating instruments
Recording instruments
5|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
2. Digital Instruments
Signals varying in discrete steps and taking on a finite number of different values in a given range are
digital signals and the corresponding instruments are of digital type. Digital instruments have some
advantages over analog meters, in that they have high accuracy and high speed of operation. It
eliminates the human operational errors. Digital instruments can store the result for future purposes.
A digital multimeter is the example of a digital instrument.
2. Electrical Instruments
When the instrument pointer deflection is caused by the action of some electrical methods then it is
called an electrical instrument. The time of operation of an electrical instrument is more rapid than
that of a mechanical instrument. Unfortunately, an electrical system normally depends upon a
mechanical measurement as an indicating device. This mechanical movement has some inertia due
to which the frequency response of these instruments is poor.
3. Electronic Instruments
Electronic instruments use semiconductor devices. Most of the scientific and industrial
instrumentations require very fast responses. Such requirements cannot be met with by mechanical
and electrical instruments. In electronic devices, since the only movement involved is that of
electrons, the response time is extremely small owing to very small inertia of the electrons. With the
use of electronic devices, a very weak signal can be detected by using pre-amplifiers and amplifiers.
Advantages of Electrical/Electronic Instruments
Non-contact measurements are possible
These instruments consume less power
Compact in size and more reliable in operation
Greater flexibility
Good frequency and transient response
Remote indication and recording possible
Amplification produced greater than that produced in mechanical instruments
OPERATING TORQUES
Three types of torques are needed for satisfactory operation of any indicating instrument. These are
1. Deflecting torque
2. Controlling torque
3. Damping torque
Deflecting Torque/Force
Any instrument’s deflection is found by the total effect of the deflecting torque/force, control torque/
force and damping torque/force. The deflecting torque’s value is dependent upon the electrical signal
to be measured; this torque/force helps in rotating the instrument movement from its zero position.
The system producing the deflecting torque is called the deflecting system.
Controlling Torque/Force
The act of this torque/force is opposite to the deflecting torque/force. When the deflecting and
controlling torques are equal in magnitude then the movement will be in definite position or in
equilibrium. Spiral springs or gravity is usually given to produce the controlling torque. The system
which produces the controlling torque is called the controlling system.
The functions of the controlling system are
1. To produce a torque equal and opposite to the deflecting torque at the final steady position
of the pointer in order to make the deflection of the pointer definite for a particular
magnitude of current
2. To bring the moving system back to its zero position when the force causing the
instrument moving system to deflect is removed
The controlling torque in indicating instruments is almost always obtained by a spring, much less
commonly, by gravity.
8|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Damping Torque/Force
A damping force generally works in an opposite direction to the movement of the moving system.
This opposite movement of the damping force, without any oscillation or very small oscillation brings
the moving system to rest at the final deflected position quickly. Air friction, fluid friction and eddy
currents provide the damping torque/force to act. It must also be noted that not all damping force
affects the steady-state deflection caused by a given deflecting force or torque. With the angular
velocity of the moving system, the intensity of the damping force rises; therefore, its effect is greatest
when it rotates rapidly and zero when the system rotation is zero. In the description of various types
of instruments, detailed mathematical expressions for the damping torques are taken into
consideration.
When the deflecting torque is much greater than the controlling torque, the system is called
underdamped. If the deflecting torque is equal to the controlling torque, it is called critically damped.
When deflecting torque is much less than the controlling torque, the system is under overdamped
condition.
Controlling System
Spring Control
A hair-spring, usually of phosphor-bronze
attached to the moving system, is used in
indicating instruments for control purpose.
To give a controlling torque which is directly
proportional to the angle of deflection of the
moving system, the number of turns on the spring
should be fairly large, so that the deflection per
unit length is small. The stress in the spring must
be limited to such a value that there is no
permanent set.
Suppose that a spiral spring is made up of a total
length L m of strip whose cross-section is
rectangular, the radial thickness being t m and the
9|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
depth b m. Let E be Young’s modulus (N/m2) for the material of the spring. Then, if θ radians be the
deflection of the moving system to which one end of the spring is being attached, the expression for
the controlling torque is
Gravity Control
Gravity control has the following advantages when compared with spring control:
It is cheaper
Independent of temperature
Does not deteriorate with time
Consider an instrument in which the deflecting torque TD is directly proportional to the current (say)
to be measured.
Thus, if I is the current,
If the instrument is spring-controlled, the controlling torque being TC, when the deflection is θ,
TC = ksθ ( ks is spring constant)
Also, TC=TD
or ksθ = kI
10|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Thus, a gravity-controlled instrument would have a scale which is ‘cramped’ at its lower end instead
of being uniformly divided, though the deflecting torque is directly proportional to the quantity to be
measured.
Damping System
There are three systems of damping generally used. These are as follows:
Air-friction damping
Fluid-friction damping
Eddy-current damping
Air-Friction Damping
In this method, a light aluminium piston is attached to the moving system and moves in an air chamber
closed at one end. The cross-section of this chamber may be either circular or rectangular. The
clearance between the piston and the sides of the chamber should be small and uniform. If the piston
is moving rapidly into the chamber, the air in the closed space is compressed and the pressure
opposes the motion of the piston (and, therefore, of the whole moving system). If the piston is moving
out of the chamber rapidly, the pressure in the closed space falls, and the pressure on the open side
of the piston is greater than that on the opposite side. Motion is thus again opposed. Sometimes
instead of a piston, a vane, mounted on the spindle of the moving system, moves in a closed-sector-
shaped box.
Fluid-Friction Damping
In this type of damping, a light vane, attached to the spindle of the moving system, dips into a pot of
damping oil and should be completely submerged by the oil. The frictional drag in the disc is always
in the direction opposing motion. There is no friction force when the disc is stationary. In the second
system, increased damping is obtained by the use of vanes.
11|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Eddy-Current Damping
When a sheet of conducting material moves in a magnetic field so as to cut through lines of force,
eddy currents are set up in it and a force exists between these currents and the magnetic field, which
is always in the direction opposing the motion. The force is magnitude of the current is proportional
to the velocity of movement of the conductor, and thus, if the magnetic field is constant, the damping
force is proportional to the velocity of the moving system and is zero when there is no movement of
the system.
Eddy-Current Damping Torque of Metal Former shows a metallic former moving in the field of a
permanent magnet.
12|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
CHAPTER-2
ANALOG AMMETERS AND VOLTMETERS
INTRODUCTION
An analog device is one in which the output or display is a continuous function of time and bears a
constant relation to its input. Measuring instruments are classified according to both the quantity
measured by the instrument and the principle of operation. Three general principles of operation are
available: (i) electromagnetic, which utilizes the magnetic effects of electric currents; (ii) electrostatic,
which utilizes the forces between electrically charged conductors; (iii) electro-thermal, which utilizes
the heating effect.
Electric measuring instruments and meters are used to indicate directly the value of current, voltage,
power or energy. In this chapter, we will consider an electromechanical meter (input is as an electrical
signal which results in mechanical force or torque as an output) that can be connected with additional
suitable components in order to act as an ammeter and a voltmeter. The most common analog
instrument or meter is the permanent magnet moving coil instrument and it is used for measuring a
dc current or voltage of an electric circuit. On the other hand, the indications of alternating current
ammeters and voltmeters must represent the rms values of the current, or voltage, respectively,
applied to the instrument.
Absolute instruments give the value of the electrical quantity to be measured in terms of the
constants of the instruments and to its deflection, no comparison with another instrument being
required. For example, the tangent galvanometer gives the value of the current to be measured in
terms of the tangent of the angle of deflection produced by the current, the radius and the number
of turns of galvanometer coil, and the horizontal component of the earth’s magnetic field. No
calibration of the instrument is thus necessary.
Secondary instruments are so constructed that the value of current, voltage or other quantity to be
measured can be determined from the deflection of the instruments, only if the latter has been
calibrated by comparison with either an absolute instrument or one which has already been
calibrated. The deflection obtained is meaningless until such a calibration has been made.
This class of instruments is in most general use, absolute instrument being seldom used except in
standard laboratories and similar institutions.
Recording instruments give a continuous record of the quantity being measured over a specified
period. The variation of the quantity being measured are recorded by a pen (attached to the moving
system of the instrument; the moving system is operated by the quantity being measured) on a sheet
of paper that moves perpendicular to the movement of the pen.
Integrating instruments record totalized events over a specified period of time. The summation,
which they give, is the product of time and an electrical quantity. Ampere hour and watt hour (energy)
meters are examples of this category.
PRINCIPLE OF OPERATION
Analog instruments may be classified according to the principle of operation they utilise. The effects
they utilise are
Magnetic effect
Heating effect
Electrostatic effect
Electromagnetic effect
Hall effect
The majority of analog instruments including moving coil, moving iron and electrodynamic utilise the
magnetic effect. The effect of the heat produced by a current in a conductor is used in thermocouple
and hotwire instruments. Electrostatic effect is used in electrostatic voltmeters. The electromagnetic
induction effect is used in induction wattmeters and induction energy meters.
Principle of Operation
The principle on which a Permanent Magnet Moving Coil (PMMC) instrument operates is that a torque
is exerted on a current-carrying coil placed in the field of a permanent magnet. The coil C has a number
of turns of thin insulated wires wound on a rectangular aluminium former F. The frame is carried on
a spindle S mounted in jewel bearings J1, J2. A pointer PR is attached to the spindle so that it moves
over a calibrated scale. The whole of the moving system is made as light in weight as possible to keep
the friction at the bearing to a minimum.
The coil is free to rotate in air gaps formed
between the shaped soft-iron pole piece (pp) of a
permanent magnet PM and a fixed soft-iron
cylindrical core IC. The core serves two purposes; (a)
it intensifies the magnetic field by reducing the
length of the air gap, and (b) it makes the field radial
and uniform in the air gap.
Thus, the coil always moves at right angles to the
magnetic field. Modern permanent magnets are
made of steel alloys which are difficult to machine.
Soft-iron pole pieces (pp) are attached to the
permanent magnet PM for easy machining in order
to adjust the length of the air gap.
14|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
A soft-iron yoke (Y ) is used to complete the flux path and to provide shielding from stray external
fields.
Deflecting Torque Equation of PMMC Instrument
Let, B = flux density in the air gap (wb/m2)
For a permanent magnet, B is constant. Also, for a given coil l, b and n are constants and thus the
product (Blnb) is also a constant, say k1.
Control Torque The control on the movement of the pointer over the scale is provided by two spirally
wound, phosphor-bronze springs S1 and S2, one at each end of the spindle S. Sometimes these springs
also conduct the current into and out of the coil.
The control torque of the springs is proportional to the angle θ turned through by the coil.
So, angular deflection of the pointer is directly proportional to the current. Thus the scale of the
instrument is linear or uniformly divided.
Damping Torque When the aluminium former (F) moves with the coil in the field of the permanent
magnet, a voltage is induced, causing eddy current to flow in it. These current exerts a force on the
former. By Lenz’s law, this force opposes the motion producing it. Thus, a damping torque is obtained.
Such a damping is called eddy-current damping.
The resistance of the shunt can be calculated using conventional circuit analysis.
Rsh = shunt resistance (Ω)
Rm = coil resistance (Ω)
Im = Ifs = full-scale deflection current (A)
Ish = shunt current (A)
I = current to be measured (A)
The voltage drop across the shunt and the meter must be same as they are connected in parallel.
The ratio of the total current to the current in the meter is called multiplying power of shunt.
Multiplying power,
Voltmeter Multipliers
For measuring higher voltages, a high resistance is connected in series with the instrument to limit
the current in the coil to a safe value. This value of current should never exceed the current required
to produce the full scale deflection. The high resistance connected in series with the instrument is
called a multiplier.
The value of multiplier required to extend the voltage range, is calculated as under:
Rsc = multiplier resistance (Ω)
Rm = meter resistance (Ω)
17|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
V = ImRm
V = Im(Rm + Rsc)
Sensitivity The moving-coil instrument is a very sensitive instrument. It is, therefore, widely used for
measuring current and voltage. The coil of the instrument may require a small amount of current (in
the range of µA) for full-scale deflection. The sensitivity is sometimes expressed in ohm/volt. The
sensitivity of a voltmeter is given by
Where, Ifs is the full-scale deflecting current. Thus, the sensitivity depends upon on the current to
give full-scale deflection.
MOVING-IRON INSTRUMENTS
Moving-Iron or MI instruments can be classified as
Attraction-type moving-iron instruments
Repulsion-type moving-iron instruments
The current to be measured, in general, is passed through a coil of wire in the moving-iron
instruments. In case of voltage measurement, the current which is proportional to the voltage is
measured. The number of turns of the coil depends upon the current to be passed through it. For
operation of the instrument, a certain number of ampere turns is required. These ampere turns can
be produced by the product of few turns and large current or reverse.
18|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The control torque is provided by two hair springs S1 and S2 in the same way as for a PMMC
instrument; but in such instruments springs are not used to carry any current. Gravity control can also
be used for vertically mounted panel type MI meters. The damping torque is provided by the
movement of a thin vane V in a closed sector-shaped box B, or simply by a vane attached to the
moving system. Eddy current damping can not be used in MI instruments owing to the fact that any
permanent magnet that will be required to produce Eddy current damping can distort the otherwise
weak operating magnetic field produced by the coil.
If the current in the fixed coil is reversed, the field produced by it also reverses. So the polarity induced
on the vane reverses. Thus whatever be the direction of the current in the coil the vane is always be
magnetized in such a way that it is attracted into the coil. Hence such instrument can be used for both
direct current as well as alternating current.
Two different designs for moving iron instruments commonly used are as follows:
Radial Vane Type In this type, the vanes are radial strips of iron. The fixed vane is attached to the coil
and the movable one to the spindle of the instrument. The instrument pointer is attached to the
moving vane spindle.
As current flows through the coil, the generated magnetic field induces identical polarities on both
the fixed and moving vane. Thus, even when the current through the coil is alternating (for AC
measurement), there is always a repulsion force acting between the like poles of fixed and moving
vane. Hence deflection of the pointer is always in the same direction irrespective of the polarity of
current in the coil. The amount of deflection depends on the repulsion force between the vanes which
19|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
in turn depends on the amount of current passing through the coil. The scale can thus be calibrated
to read the current or voltage directly.
Co-axial Vane Type I In these type of instruments, the fixed and moving vanes are sections of coaxial
cylinders. Current in the coil magnetizes both the vanes with similar polarity. Thus the movable vane
rotates along the spindle axis due to this repulsive force. Coaxial vane type instruments are
moderately sensitive as compared to radial vane type instruments that are more sensitive.
Moving iron instruments have their deflection is proportional to the square of the current flowing
through the coil. These instruments are thus said to follow a square law response and have non-
uniform scale marking. Deflection being proportional to square of the current, whatever be the
polarity of current in the coil, deflection of a moving iron instrument is in the same direction. Hence,
moving iron instruments can be used for both DC and AC measurements.
Advantages of MI Instruments
Robust construction and relatively cheap
Suitable for measuring both dc and ac
Can withstand overload momentarily
Disadvantages of MI Instruments
As the deflection is proportional to I 2, hence the scale of the instrument is not uniform. It is
cramped in the lower end and expanded in the upper portion.
It is affected by stray magnetic fields.
There is hysteresis error in the instrument. The hysteresis error may be minimized by using
the vanes of nickel-iron alloy.
When used for measuring ac the reading may be affected by variation of frequency due to the
change in reactance of the coil, which has some inductance. With the increase in frequency
iron loses and coil impedance increases.
Since large amount of power is consumed to supply I2R loss in the coil and magnetic losses in
the vanes, it is not a very sensitive instrument.
20|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
ELECTRODYNAMOMETER-TYPE INSTRUMENTS
The electrodynamometer is a transfer-type instrument. A transfer-type instrument is one that may
be calibrated with a dc source and then used without modification to measure ac. This requires the
transfer type instruments to have same accuracy for both dc and ac.
Controlling Torque The controlling torque is provided by two control springs. These springs act as
leads to the moving coil.
Operation with ac Let, i1 and i2 be the instantaneous values of current carried by the coils. Therefore,
the instantaneous deflecting torque is:
If the two coils are connected in series for measurement of current, the two instantaneous currents
i1 and i2 are equal.
Say, ii = i 2 = i
Electrodynamic Wattmeter The electrodynamic wattmeter consist of two fixed coils ‘a’ and ‘b’ placed
symmetrical to each other and producing a uniform magnetic field. They are connected in series with
the load and are called the Current Coils (CC). The two fixed coils can be connected in series or parallel
to give two different current ratings. The current coils carry the full-load current or a fraction of full
load current. Thus the current in the current coils is proportional to the load current. The moving coil
‘c’, in series with a high non inductive resistance Rv is connected across the supply. Thus the current
flowing in the moving coil is proportional to, and practically in phase with the supply voltage. The
moving coil is also called the voltage coil or Pressure Coil (PC). The voltage coil is carried on a pivoted
spindle which carries the pointer, the pointer moved over a calibrated scale.
Two hair springs are used for providing the controlling torque and for leading current into and out of
the moving coil. Damping is provided by air friction.
Torque Equation
Thus, the instantaneous value of the deflecting torque is proportional to the instantaneous power.
Owing to the inertia of the moving system, the pointer reads the average power. In dc circuits, the
power is given by the product of voltage and current, and hence the torque is directly proportional to
the power. Thus, the instrument indicates the power.
22|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
RECTIFIER-TYPE INSTRUMENTS
The basic arrangement of a rectifier type of instrument using a full-wave rectifier circuit. If this
instrument is used for measuring ac quantity then first the ac signal is converted to dc with the help
of the rectifier. Then this dc signal is measured by the PMMC meter. The multiplier resistance Rs, is
used to limit the value of the current in order that it does not exceed the current rating of the PMMC
meter.
These types of instruments are used for light current work where the voltage is low and resistances
high.
Hence, the sensitivity of a half-wave rectifier instrument with ac is 0.45 times its sensitivity with dc
and the deflection is 0.45 times that produces with dc of equal magnitude V.
It shows a full-wave rectifier circuit along with the input and output waveform. Average value of
voltage/current for full-wave rectifier,
So the deflection is 0.9 times in a full-wave rectifier instrument with an ac than that produced with dc
of equal magnitude V.
Sensitivity of a full-wave rectifier instrument with an ac is 0.9 times its sensitivity with dc.
Limitations
Rectifier instruments are only accurate on the waveforms on which they are calibrated. Since
calibration assumes pure sine waves, the presence of harmonics gives erroneous readings.
The rectifier is temperature sensitive, and therefore, the instrument readings are affected by
large variations of temperature.
Applications
The rectifier instrument is very suitable for measuring alternating voltages in the range of 50–
250 V.
The rectifier instrument may be used as a micrometer or low milli-ammeter (up to 10–15 mA).
It is not suitable for measuring large currents because for larger currents the rectifier becomes
too bulky and providing shunts is impracticable due to rectifier characteristics.
Rectifier instruments find their principal application in measurement in high-impedance circuits at
low and audio frequencies. They are commonly used in communications circuits because of their high
sensitivity and low power consumption.
CHAPTER-3
WATTMETER AND MEASUREMENT OF POWER
POWER MEASUREMENT
In alternating current circuits, the instantaneous power varies continuously as the
voltage and current varies while going through a cycle. In such a case, the power at
any instant is given by
where, p(t), v(t), and i(t) are values of instantaneous power, voltage, and current
respectively.
Thus, if both voltage and current can be assumed to be sinusoidal, with the current
lagging the voltage by phase-angle φ, then
where, Vm and Im are peak values of voltage and current respectively, and w is the
angular frequency.
The instantaneous power pis therefore given by
where, V and I are rms values of voltage and current respectively and cos j is power
factor of the load.
Involvement of the power-factor term cos j in the expression for power in ac circuit
indicates that ac power cannot be measured simply by connecting a pair of ammeter
and voltmeter. A wattmeter, with in-built facility for taking in to account the power
factor, can only be used for measurement of power in ac circuits.
26|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
ELECTRODYNAMOMETER-TYPE WATTMETER
Construction of Electrodynamometer-
type Wattmeter
The moving coil that is connected across the load carries a current proportional to
the voltage. Since the moving coil carries a current proportional to the voltage, it is
called the voltage coil or the pressure coil or simply PC of the wattmeter. The moving
coil is entirely embraced by the pair of fixed coils. A high value non-inductive
resistance is connected in series with the voltage coil to restrict the current through
it to a small value, and also to ensure that voltage coil current remains as far as
possible in phase with the load voltage.
The moving coil, made of fine wires, is wound either as a self-sustaining air-cored coil,
or else wound on a nonmetallic former. A metallic former, otherwise would induce
Eddy-currents in them under influence of the alternating field.
27|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The moving, or voltage coil along with the pointer is mounted on an aluminum
spindle in case jewel bearings are used to support the spindle. For higher
sensitivity requirements, the moving coil may be suspended from a torsion head
by a metallic suspension which serves as a lead to the coil. In other constructions,
the coil may be suspended by a silk fibre together with a spiral spring which gives
the required torsion. The phosphor-bronze springs are also used to lead current
into and out of the moving coil. In any case, the torsion head with suspension, or
the spring, also serves the purpose of providing the restoring torque to bring the
pointer back to its initial position once measurement is over.
The moving, or voltage coil current must be limited to much low values keeping
in mind the design requirements of the movement system. Current is lead to and
out of the moving coil through two spiral springs. Current value in the moving
coil is thus to be limited to values that can be safely carried by the springs without
appreciable heating being caused.
3. Damping System
4. Shielding System
With a spring constant K, the controlling torque provided by the spring for a final
steady-state deflection of θ is given by
TC=Kθ
Under steady-state condition, the average deflecting torque will be balanced by the
controlling torque provided by the spring. Thus, at balanced condition TC = Td
INDUCTION-TYPE WATTMETER
Induction-type wattmeters, however, following the very basic principles of mutual
induction, can only be used for measurement of ac power, in contrast to
electrodynamometer type wattmeters that can be used for power measurements in
both ac and dc circuits. Induction type wattmeters, in contradiction to
electrodynamometer-type wattmeters, can be used only with circuits having relatively
steady values of frequency and voltage.
V = voltage to be
measured I = current
to be measured
The instantaneous torque acting on the aluminum disc is proportional to (φsh · ise -
φse · ish).
Let, instantaneous value the applied voltage is
v = Vm sin w t
i = Im sin (w t - j)
where k¢ is a constant and the minus (-) sign indicating the fact the flux φsh lags
behind the voltage by 90°.
31|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The eddy emf induced in the disc due to the shunt magnet flux is
Similarly, the eddy emf induced in the disc due to the series magnet flux is
The instantaneous deflecting torque (T) acting on the disc can now be calculated as
where, V and I are rms values of voltage and current. Average torque on the
instrument is thus found to be proportional to the power in the circuit.
32|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
CHAPTER-4
ENERGYMETERS AND MEASUREMENT OF ENERGY
Energy is the total power consumed over a time interval, that is Energy = Power × Time. Generally, the
process of measurement of energy is same as that for measurement of power except for the fact that
the instrument used should not merely measure power or rate of consumption of energy, but must
also take into account the time interval during which the power is being supplied.
The unit of energy can be expressed in terms of Joule or Watt-second or Watt-hour as per convenience.
A larger unit that is most commonly used is kilowatt-hour (kWh), which is defined as the energy
consumed when power is delivered at an average rate of 1 kilowatt for one hour. In commercial
metering, this amount of 1 kilowatt-hour (kWh) energy is specified as 1 unit of energy.
Energy meters used for measurement of energy have moving systems that revolve continuously, unlike
in indicating instruments where it deflects only through a fraction of a revolution. In energy meters,
the speed of revolution is proportional to the power consumed. Thus, total number of revolutions
made by the meter moving system over a given interval of time is proportional to the energy
consumed. In this context, a term called meter constant, defined as the number of revolutions made
per kWh, is used. Value of the meter constant is usually marked on the meter enclosure.
with flux φ1 to produce a force F2 and hence a torque Td2 on the disc. Total torque is resultant of the
torques Td1 and Td2.
Let φ1 and φ2 are the instantaneous values of two fluxes having a phase difference of α between them.
Therefore, we can write
where, φ1m and φ2m are peak values of fluxes φ1 and φ2 respectively.
The flux φ1 will produce an alternating emf is the disc, given by
Similarly, the alternating emf produced in the disc due to the flux φ2 is given by
If, is considered to the impedance of the aluminum disc with power factor β then eddy current
induced in the disc due to the emf e1 can be expressed as
Similarly, eddy current induced in the disc due to the emf e2 is given by
Similarly, instantaneous torque Td2 produced due to interaction of the current i2 and flux φ1 is given by
Volatage coil–many turns of fine wire encased in plastic, connected in parallel with load.
Display dials
1. Operating System
The operating system consists of two electromagnets. The cores of these electromagnets are made of
silicon steel laminations. The coils of one of these electromagnets (series magnet) are connected in
series with the load, and is called the current coil. The other electromagnet (shunt magnet) is wound
with a coil that is connected across the supply, called the pressure coil. The pressure coil, thus, carries
a current that is proportional to supply voltage.
Shading bands made of copper are provided on the central limb of the shunt magnet. Shading bands,
as will be described later, are used to bring the flux i — Bearing produced by a shunt magnet exactly
in quadrature Pivot with the applied voltage.
2. Moving System
The moving system consists of a light aluminum disc mounted on a spindle. The disc is placed in the
space between the series and shunt magnets. The disc is so positioned that it intersects the flux
produced by both the magnets. The deflecting torque on the disc is produced by interaction between
these fluxes and the eddy current they induce in the disc. In energy meters, there is no control spring
as such, so that there is continuous rotation of the disc.
3. Braking System
The braking system consists of a braking device which is usually a permanent magnet positioned near
the edge of the aluminum disc.
The emf induced in the aluminum disc due to relative motion between the rotating disc and the fixed
permanent magnet (brake magnet) induces eddy current in the disc. This eddy current, while
interacting with the brake magnet flux, produces a retarding or braking torque. This braking torque is
proportional to speed of the rotating disc. When the braking torque becomes equal to the operating
torque, the disc rotates at a steady speed. The position of the permanent magnet with respect to the
rotating disc is adjustable. Therefore, braking torque can be adjusted by shifting the permanent
magnet to different radial positions with respect to the disc.
4. Registering System
36|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The current IP produces a flux φpt that is in same phase as IP. This flux is made to divide itself in two
parts, φg and φp. The major portion of total pressure coil flux, i.e., φg passes through the side gaps, as
reluctance of these paths are low due very small air gaps. Remaining portion of the flux, i.e., φp passes
through the disc and is responsible for production of the driving torque. Due to larger reluctance of
the path, this flux φp is relatively weaker.
The flux φp is proportional to the current IP and is in the same phase. The flux φp is thus proportional
to the supply voltage V and lags it by an angle δ which is only a few degrees less than 90°. The flux φp
being alternating in nature, induces and eddy emf Eep in the disc, which in turn produces eddy current
I. Depending on the impedance angle β of the aluminum disc, eddy current I will lag behind the eddy
emf Eep by an angle β.
The load current I flows through the series magnet current coil and produces a flux φs. This flux is
proportional to the load current I and is in phase with it. This flux, in the same way, induces and eddy
emf Ees in the disc, which in turn produces eddy current Ies. The eddy current Ies lags behind the eddy
emf Ees by the same angle β.
Now, the eddy current Ies interacts with flux φp to produce a torque and the eddy current Iep interacts
with flux φs to produce another torque. These two torques are in opposite direction, and the resultant
torque is the difference of these two.
The resultant deflecting torque on the disc due to combined action of two fluxes φp and φs is given as
38|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Where, Z is the impedance of the aluminum disc and ω is the angular frequency of supply voltage.
Thus, in order that the speed of rotation can be made to be proportionate to the power consumed,
the angle difference δ between the supply voltage V and the pressure coil flux φp must be made 90°.
Total number of revolutions within a time interval dt is
Phase-angle Error
The meter will indicate true energy only if the phase angle between the pressure coil flux φp and the
supply voltage V is 90°. This requires that the pressure coil winding should be designed as highly
inductive and its resistance and iron losses should be made minimum. But, even then the phase angle
is not exactly 90°, rather a few degrees less than 90°. Suitable adjustments can be implemented such
39|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
that the shunt magnet flux linking with the disc can be made to lag the supply voltage by an angle
exactly equal to 90°.
by the bands is reduced. This changes the phase angle difference between φp and φpt, as can be
observed. Thus, careful adjustments of the copper shading bands position can make the phase
difference between the supply voltage V and resultant shunt magnet flux φp to be exactly 90°.
Creeping Error
In some meters, a slow but continuous rotation of the disc can be observed even when there is no
current flowing through the current coil, and
only pressure coil is energised. This is called
creeping. The primary reason for creeping is
due to over-compensation for friction.
Though the main driving torque is absent at
no-load, the additional torque provided by the
friction compensating vane will make the disc
continue to rotate. Other causes of creeping
may be excessive voltage across the potential
coil resulting in production of excessive
torque by the friction compensating device, or
vibrations, and stray magnetic fields.
41|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Creeping can be avoided by drilling two holes on the aluminum disc placed on diametrically opposite
locations. Drilling such holes will distort the eddy current paths along the disc and the disc will tend to
stop with the holes coming underneath the
shunt magnet poles. The disc can thus creep
only till a maximum of half the rotation till
one of the holes comes below the shunt
magnet pole. This effect is however, too
insignificant to hamper disc movement
during normal running operations under
load.
Creeping can also be avoided by attaching a tiny piece of iron to the edge of the disc. The brake magnet
in such a case can lock the iron piece to itself and prevent creeping of the disc. Once again, this action
is too insignificant to hamper disc movement during normal running operations under load.
On the other hand, as the disc rotates continuously in the field of the series magnet, an emf is induced
dynamically in the disc due to its linkage with the series magnet flux φs. This emf induces eddy currents
in the disc that interact with the series magnet flux to create a retarding or braking torque that opposes
motion of the disc. This self braking torque is proportional to the square of the series magnet flux or is
proportional to the
square of the load current; i.e., T φ2 I2.
b s
42|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
At higher loads, thus the braking torque overpowers the deflecting torque and the meter tends to
rotate at slower speed, and consequently reads lower than actual.
To avoid such errors, and to minimise the self-braking action, the full load speed of the disc is set at
lower values. The current coil flux φs is made smaller as compared to the pressure coil flux φp. Thus,
the dynamically induced emf that causes the braking torque is restricted as compared to the driving
torque. Magnetic shunts are also sometimes used with series magnets to compensate for overload
errors at high current values.
Creep test With pressure coil supplied with 110% of rated voltage and current coil open circuited, the
meter disc should not rotate by more than one revolution, i.e., it should not creep.
Starting test At 0.5% of rated current and full rated voltage, the meter disc should start rotating.
Phantom Loading
When the current rating of the meter under test is high, a test with actual loading arrangements would
involve considerable wastage of energy and also it is difficult to arrange for such large loads under
laboratory test conditions. In such cases, to avoid this, ‘phantom’ or ‘fictitious’ loading arrangements
are done for testing of energy meters.
43|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Phantom loading consists of supplying the shunt magnet pressure coil circuit from a rated voltage
source. The series magnet current coil is supplied from a separate low voltage supply source. It is
possible to circulate rated current through the current coil circuit with the low voltage source since
impedance of this circuit is very low. The energy indicated by the meter under phantom loading
condition is the same as the energy indication as would have been with a real load. With this
arrangement, the total energy consumed for the test is comparatively smaller. The total energy
required for the test is that due to the small pressure coil current at rated voltage and small current
coil voltage at rated current.
CHAPTER 5
MEASUREMENT OF SPEED, FREQUENCY AND
POWER FACTOR
Electrical Tachometer
Definition: The tachometer use for measuring the rotational speed or angular velocity of the machine
which is coupled to it. It works on the principle of relative motion between the magnetic field and
shaft of the coupled device. The relative motion induces the EMF in the coil which is placed between
the constant magnetic field of the permanent magnet. The developed EMF is directly proportional to
the speed of the shaft.
44|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Mechanical and electrical are the two types of the tachometer. The mechanical tachometer measures
the speed of shaft regarding revolution per minutes.
The electrical tachometer converts the angular velocity into an electrical voltage. The electrical
tachometer has more advantages over the mechanical tachometer. Thus it is mostly used for measuring
the rotational speed of the shaft.
Depends on the natures of the induced voltage the electrical tachometer is categorized into two types.
1. DC Tachometer Generator
2. AC Tachometer Generator
DC Tachometer Generator
Permanent magnet, armature, commutator, brushes, variable resistor, and the moving coil voltmeter
are the main parts of the DC tachometer generator. The machine whose speed is to be measured is
coupled with the shaft of the DC tachometer generator.
The DC tachometer works on the principle that when the closed conductor moves in the magnetic
field, EMF induces in the conductor. The magnitude of the induces emf depends on the flux link with
the conductor and the speed of the shaft.
The armature of the DC generator revolves between the constant field of the permanent magnet. The
rotation induces the emf in the coil. The magnitude of the induced emf is proportional to the shaft
speed.
The commutator converts the alternating current of the armature coil to the direct current with the help
of the brushes. The moving coil voltmeter measures the induced emf. The polarity of the induces
voltage determines the direction of motion of the shaft. The resistance is connected in series with
the voltmeter for controlling the heavy current of the armature.
45|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Now,
Advantages:
The following are the advantages of the DC Tachometer.
The polarity of the induced voltages indicates the direction of rotation of the shaft.
The conventional DC type voltmeter is used for measuring the induced voltage.
Disadvantages:
The commutator and brushes require the periodic maintenance.
The output resistance of the DC tachometer is kept high as compared to the input resistance. If
the large current is induced in the armature conductor, the constant field of the permanent magnet
will be distorted.
AC Tachometer Generator
The DC tachometer generator uses the commutator and brushes which have many disadvantages. The
AC tachometer generator designs for reducing the problems. The AC tachometer has stationary
46|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
armature and rotating magnetic field. Thus, the commutator and brushes are absent in AC tachometer
generator.
The rotating magnetic field induces the EMF in the stationary coil of the stator. The amplitude and
frequency of the induced emf are equivalent to the speed of the shaft. Thus, either amplitude or
frequency is used for measuring the angular velocity.
The below mention circuit is used for measuring the speed of the rotor by considering the amplitude of
the induced voltage. The induces voltages are rectified and then passes to the capacitor filter for
smoothening the ripples of rectified voltages.
Advantages:
The drag cup Tachogenerator generates the ripple free output voltage.
The cost of the generator is also very less.
Disadvantage:
The nonlinear relationship obtains between the output voltage and input speed when the rotor
rotates at high speed.
47|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
This varying laminated core holds the fixed coil at its one end. Then we connect this fixed coil across
the supply mains.
The electrical resonance type frequency meter measures the frequency of these supply mains. Now
there is a moving coil which is so pivoted at its top end that it can move along the extended core of the
fixed coil like a pendulum. The pointer of the instrument is so attached at the top end of the moving
coil that its tip moves along the semicircular dial. Now, we connect on a capacitor across the two leads
of the moving coil.
48|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Torque Equation
Let us consider I1 is the supply current of the fixed coil and I2 is the induced current of the moving
coil. Now, we have already mentioned that the phase angle between the supply current I1 (current in
the fixed coil) and the emf induced in the moving coil is 90°. Again there is a phase difference
between the induced emf and the induced current I2 (current in the moving coil).
Let us consider the angle of this phase difference is α. So, the actual phase difference between I1 and
I2 will be (90°-α). Therefore, we can write the expression of the torque (T) as
From the above expression of the torque, we can see that the torque will be zero when α is zero. That
means there must not be any phase difference between the induced current and the induced emf in the
moving coil.
Resonance: That can only be possible when inductive reactance of the moving coil becomes equal to
its capacitive reactance.
Again the inductive reactance (2πfL) depends upon the angular position of the moving coil on
the extended core of the fixed coil.
49|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
So, when we just switch on the supply, the fixed coil starts attracting the moving coil towards it.
This attraction due to the torque acting on the moving system.
Therefore, the moving coil starts rotating along with the pointer attached to it. As a result, the
inductive reactance of the moving coils changes. Then after certain angular rotation of the
moving coil the inductive reactance of this coil exactly becomes equal to the
capacitive reactance of the coil. At that point of time, there will be no torque acting on the
moving system of the electrical resonance type frequency meter. Therefore the pointer of the
instrument becomes stationary at that point.
If somehow the supply frequency changes, the value of inductive reactance of the instrument
also changes. Therefore the resonance of the circuit gets disturbed. Therefore again the deflecting
torque appears on the moving system and tries to rotate it further. Hence, again the inductive
reactance of the moving coils changes. And after a certain rotation again resonance occurs. So,
here again, the torque becomes zero. Therefore the pointer rests on a new position.
So, we have seen how the position of the pointer on the dial of electrical resonance type
frequency meter changes with changing the supply frequency.
Calibration
First, we supply an electrical signal of exactly 50Hz to the moving coil. Then we find the exact
position of the pointer tip on the dial and make it as 50Hz.
Then we slowly increase the supply frequency step by step and see the position at each of the
steps. And we mark these positions of the pointer on the dial with the corresponding supply
frequencies.
Then we reduce the supply frequency step by step and mark the corresponding positions of the
pointer on the dial with corresponding frequencies.
Measurement of Frequency
When we connect the leads of the fixed coil of an electrical resonance type frequency meter with
supply mains the position of the pointer on the dial indicates the actual frequency of the supply signal.
Conceptually, the principal of this frequency meter is quite simple. The instrument consists of a
number of metal reeds with slightly different natural frequencies fitted side by side.
Also, the instrument consists of an electromagnet placed nearer to these reeds.
After that, we connect the electromagnet across the supply main. Consequently, there will be
an alternating current flowing through the coil of the electromagnet. Therefore, the electromagnet
produces an alternating magnetic flux with the same frequency of supply current.
The natural frequency of any of the reeds matches to the double of the frequency of the alternating
flux. Therefore the resonance occurs at that particular reed. As a result, this particular reed will
vibrate with more amplitude than the others. We can see this vibration of that particular reed with
our naked eyes.
Then we can asses that the frequency of the electrical signal coming from the supply main is the
same as the natural frequency that vibrating reed.
The dimension and weight of all the reeds are not the same. Instead, they differ slightly in each read
from others. This is because for differing the natural frequency of one reed from others.
Practically this difference of natural frequency from one reed to its adjacent reed is either 1 or 0.5Hz.
Suppose we have designed a mechanical resonance frequency meter to measuring a frequency from
48 Hz to 51 Hz. If the difference of natural frequency of any reed from its adjacent reed is 1Hz, then
there will be a total of 9 reeds in the instrument fitted in
a row side by side. This is because for the above-
mentioned frequency range there will be reeds of natural
frequencies from 96 Hz (48X2) to 102 Hz (51X2)
But there will be a particular reed whose natural frequency is double of the supply frequency. Therefore
mechanical resonance occurs on that reed. Therefore the reed will vibrate most with the highest
amplitude. As a result, the vibration of this reed will be easily visible. Hence, the supply frequency will
be half of the natural frequency of that reed.
Although due to the alternating flux of the electromagnet all other reed will vibrate. But their vibration
is so slight it cannot be visible easily.
In some occasions there may be two adjacent reeds vibrate strongly with equal amplitude. In that case,
the supply frequency will be half of the average of the natural frequencies of these two adjacent reeds.
This method involves mathematical calculations. Sometimes it is required to measure the power
factor of the circuit instantaneously when the power factor of the load is varying continuously.
52|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
This purpose is served by connecting a dynamometer type power factor meter in the circuit. It
indicates the power factor of the circuit directly on the scale by the deflection of a pointer.
The current coils FF are wound with thick wire whereas pressure coils A and B are wound with fine
wire. The pressure coils fixed on the same spindle, to which a pointer is attached, constitutes the
moving system.
The pressure coil A is connected across the supply through a non-inductive resistant and pressure
coil B is connected across the supply through highly inductive choke coil of inductance L. The value of
53|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
resistance R and inductance L are so chosen that for the main supply frequency, the current in the two
pressure coils A and B is the same.
Thus the fields produced by the two coils are of the same strength. The field produced by the
coil B lags behind the field produced by the coil A slightly less than 90° because of resistance of the
coil. Accordingly while fixing the coil B the plane of this coil is displaced from the plane of the coil A
by the electrical angle which is slightly less than 90°.
However while discussing the action (working) of the instrument it will be assumed that the phase
difference between the two currents flowing through the coils A and B is 90° and same is the angle
between the planes of the coils.
Though power factor meter is an indicating instrument but no controlling torque is provided in
this instrument. The currents are being led into the moving coils A and B by fine ligaments which
exert no control.
When the instrument is connected to the load circuit, current flows through the fixed coils FF and
Moving coils A and B, flux is set by the fixed coils and moving coils.
By the alignment of two fields, torque develops i.e. the resultant field produced by the moving coils
tries to come in line with the field produced by the fixed coils and torque develops till both of them
come in line with each other.
There are three extreme conditions in which this instrument is connected in the circuit.
1. When power factor of the circuit is unity: In this case, current is in phase with circuit voltage. The
current flowing through potential coil A is in phase with the voltage which is also in phase with the
current flowing through current coil FF.
At the same time, the current flowing through potential coil B lags behind voltage as well as the
current flowing through current coil FF by 90o. Thus pressure coil A will experience a turning moment
so its plane will come in position a parallel to the plane of the current coil FF. The torque acting on the
pressure coil B is zero. Thus, the pointer indicates unity power factor on the scale.
2. When power factor of the circuit is zero lagging: In this case, current lags behind the circuit voltage
by 90°. Therefore, the current flowing through pressure coil B will be in phase with the current in current
coils FF, both being lagging behind the circuit voltage by 90°.
The current flowing through pressure coil A will lead the current in current coil FF by 90°. Thus a
turning moment acts on the pressure coil B and brings its plane parallel to the plane of current
coil FF and pointer indicates zero power factor lagging.
54|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
3. When power factor of the circuit is zero leading: In this case current leads the circuit voltage by 90o.
Therefore, the current flowing through pressure coil A lags the current in current coil FF by 90° and the
current flowing through pressure coil B lags the current in current coil FF by 180°.
Thus field produced by the moving system is just reversed to that in the case (2). Thus an opposite
turning moment acts on the pressure coil B and brings its plane parallel to the plane of current
coil FF and pointer indicates zero power factor leading.
For intermediate power factors the moving system of power factor meter takes up intermediate
positions and the pointer indicates the power factor accordingly.
The two identical moving coils A and B are fixed with their planes 120° apart and are connected across
the two remaining phases respectively through high resistances as shown in the figure.
In this case, there is no necessity for phase splitting by artificial means, since the required phase
displacement between the currents in the moving coils can be obtained from the supply itself.
When the three phase power factor meter is connected in the circuit, under balanced load conditions,
the angle through which the pointer is deflected from the unity power factor position is equal to the
phase angle of the circuit, because the two moving coils are fixed 120° apart.
The deflections in three phase power factor meter are independent of frequency and waveform
since the currents in the two moving coils are affected in the same way by any change of frequency.
1. Losses are less because of minimum use of iron parts and also give less error over a small range
of frequency as compared to moving iron type instruments.
2. They high torque is to weight ratio.
CHAPTER 6
MEASUREMENT OF RESISTANCE, INDUCTANCE&
CAPACITANCE
CLASSIFIATION OF RESISTANCE
The classification of resistances, from the point of view of measurement is as follows:
(1)Low resistances: All resistances of the order of 1Ω hand under may be classified as low resistances.
(2)Medium resistances: All resistances from 1Ω to about 0.1 MΩ come under medium resistances.
(3)High resistances: All resistances of the order above 0.1 MΩ come under high resistances.
(iii)Potentiometer method
The DC potentiometer method of measurement of resistance is used for measuring the unknown
resistance of low value. This can be done by comparing the unknown resistance with the standard
resistance. The voltage drop across the known and unknown resistance is measured and by comparison
the value of known resistance is determined.
Let understand this with the help of the circuit diagram. The R is the unknown resistance whose value
is needed to be measured. The S is the standard resistance from which the value of unknown resistance
is compared. The rheostat is used for controlling the magnitude of current into the circuit.
The double pole double throw switch is used in the circuit. The switch, when moves to position 1,
1’ the unknown resistance connects to the circuit, and when it moves to position 2, 2’ the standard
resistance connects to the circuit.
Consider that when the switch is in position 1,1’ the voltage drop across the unknown resistance is Vr
The magnitude of the current is adjusted in such a way that the voltage drop across the resistance is equal
to 1 volt.
(ii)Voltmeter–ammeter method,
(iii)Substitution method,
The Wheatstone bridge is the most commonly used circuit for measurement of medium-range
resistances. The Wheatstone bridge consists of four resistance arms, together with a battery (voltage
59|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
In the bridge circuit, R3 and R4 are two fixed known resistances, R2 is a known variable resistance and
RX is the unknown resistance to be measured. Under operating conditions, current ID through the
galvanometer will depend on the difference in potential between nodes B and C. A bridge balance
condition is achieved by varying the resistance R2 and checking whether the galvanometer pointer is
resting at its zero position. At balance, no current flows through the galvanometer. This means that at
balance, potentials at nodes B and C are equal. In other words, at balance the following conditions are
satisfied:
The detector current is zero, i.e., 1D = 0 and thus It = I3 and I2 = I4
Potentials at node B and C are same, i.e., VB = VC, or in other words, voltage drop in the arm AB equals
the voltage drop across the arm AC, i.e., VAB = VAC and voltage drop in the arm BD equals the voltage
drop across the arm CD, i.e., VBD = VCD
Thus, measurement of the unknown resistance is made in terms of three known resistances. The arms
BD and CD containing the fixed resistances R3 and R4 are called the ratio arms. The arm AC containing
the known variable resistance R2 is called the standard arm. The range of the resistance value that can
be measured by the bridge can be increased simply by increasing the ratio R3/R4.
In the loss of charge method unknown resistance is connected in parallel with the capacitor and
electrostatic voltmeter. The capacitor is initially charged to some suitable voltage by means of a battery
of voltage V and then allowed to discharge through the resistance. The terminal voltage is observed
OR
From above equation, it follows that if V, v, C, and t are known the value of R can be computed.
If the resistance R is very large the time for an appreciable fall in voltage is very large and thus this
process may become time-consuming. Also the voltage-time curve will thus be very flat and unless
great care is taken in measuring voltages at the beginning and at the end of time t, a serious error may
be made in the ratio V/v causing the considerable corresponding error in the measured value of R.
more accurate results may be obtained by change in the voltage V-v directly and calling this change as
However, from the experimental point of view, it may be advisable to determine the time t from the
discharge curve of the capacitor by plotting the curve of log v against time t. this curve is linear as shown
in the second figure and thus the determination of time t from this curve for the voltage to fall from V to
v yields more accurate results.
Loss of charge method is applicable to some high resistances, but it requires a capacitor of very high
leakage resistance as high as resistance being measured. The method is very attractive if the resistance
62|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
being measured is the leakage resistance of a capacitor as in this case auxiliary R and C units are not
required.
Actually, in this method, we do not measure the true value of resistance since we assume here that the
value of resistance of electrostatic voltmeter and the leakage resistance of the capacitor have infinite
value. But in practice corrections must be applied to take into consideration the above two resistances.
Let R1 be the leakage resistance of the capacitor. Also R’ be the equivalent resistance of the parallel
resistances R and R1.
The test is then repeated with the unknown resistance R disconnected and the capacitor discharging
through R1. The value of R1 obtained from this second test and substituted into the expression,
In order to get the value of R. The leakage resistance of the voltmeter, unless very high should also be
taken into consideration.
MEGGER
63|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
One of the most popular portable type insulation resistance measuring instruments is the mega-
ohmmeter or in short, meggar. The meggar is used very commonly for measurement of insulation
resistance of electrical machines, insulators, bushings, etc.
The traditional analog deflecting-type meggar is essentially a permanent magnet crossed-coil shunt
type ohmmeter.
The instrument has a small permanent magnet dc generator developing 500 V dc (some other
models also have 100 V, 250 V, 1000 or 2500 V generators). The generators is hand driven, through
gear arrangements, and through a centrifugally controlled clutch switch which slips at a predefined
speed so that a constant voltage can be developed. Some meggars also have rectified ac as power
supply.
The moving system in such instruments consists of two coils, the control coil CC and the deflecting
coil CD. Both the coils are mounted rigidly on a shaft that carries the pointer as well. The two coils
move in the air gap of a permanent magnet. The two coils are arranged with such numbers of turns,
radii of action, and connected across the generator with such polarities that, for external magnetic
fields of uniform intensity, the torque produced by the individual coils are in opposition thus giving
an astatic combination. The deflecting coil is connected in series with the unknown resistance RX
under measurement, a fixed resistor RD and then the generator. The current coil or the
compensating coil, along with the fixed resistance RC is connected directly across the generator. For
any value of the unknown, the coils and the pointer take up a final steady position such that the
torqueses of the two coils are equal and balanced against each other. For example, when the
64|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
resistances RX under measurement is removed, i.e., the test terminals are open-circuited, no current
flows through the deflecting coil CD, but maximum current will flow through the control coil CC. The
control coil CC thus sets itself perpendicular to the magnetic axis with the pointer indicating ‘∞ Ω’
as marked in the scale. As the value of RX is brought down from open circuit condition, more and
more current flows through the deflecting coil CD, and the pointer moves away from the ‘∞ Ω’ mark
clockwise on the scale, and ultimately reaches the ‘0 Ω’ mark when the two test terminals are short
circuited.
The surface leakage problem is taken care of by the guard-wire arrangement. The guard ring and
the guard wire diverts the surface leakage current from reaching the main moving system and
interfering with its performance.
DIGITAL MULTIMETER
A digital multi-meter is an electronic instrument which can measure very precisely the dc and ac
voltage, current (dc and ac), and resistance. All quantities other than dc voltage is first converted
into an equivalent dc voltage by some device and then measured with the help of digital voltmeter.
The procedures of measurement of different quantities are described below:
For measurement of ac voltage, the input voltage, is fed through a calibrated, compensated
attenuator, to a precision full-wave rectifier circuit followed by a ripple reduction filter. The resulting
dc is fed to an Analog Digital Converter (ADC) and the subsequent display system. Many
manufacturers provide the same attenuator for both ac and dc measurements.
For current measurement, the drop across an internal calibrated shunt is measured directly by the
ADC in the ‘dc current mode’, and after ac to dc conversion in the ‘ac current mode’. This drop is
often in the range of 200 mV (corresponding to full scale).
Due to the lack of precision in the ac–dc conversions, the accuracy in the ac range is generally of the
order of 0.2 to 0.5%. In addition, the measurement range is often limited to about 50 Hz at the lower
65|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
frequency end due to the ripple in the rectified signal becoming a non-negligible percentage of the
display and hence results in fluctuation of the displayed number. At the higher frequency end,
deterioration of the performance of the ADC converter limits the accuracy. In ac measurement the
reading is often average or rms values of the unknown current. Sometimes for measurement of
current, a current-to-voltage converter may also be used.
The current under measurement is applied to the summing junction at the input of the op-amp. The
current in the feedback resistor IR is equal to the input current IIN because of very high input
impedance of the op-amp. The current IR causes a voltage drop across one of the resistors, which is
proportional to the input current IIN. Different resistors are employed for different ranges.
For resistance measurement the digital multi-meter operates by measuring the voltage across the
externally connected resistance, resulting from a current forced through it from a calibrated internal
current source. The accuracy of the resistance measurement is of the order of 0.1 to 0.5% depending
on the accuracy and stability of the internal current sources. The accuracy may be proper in the
highest range which is often about 10 to 20 MΩ. In the lowest range, the full scale may be nearly
equal to>200 Ω with a resolution of about 0.01 Ω for a 4½ digit digital multi-meter. In this range of
resistance measurement, the effect of the load resistance will have to be carefully considered.
Interface of the output with external equipment Possible to connect an external instrument with
is not the output
possible. reading.
Simple in construction. Very complicated in construction.
CHAPTER-7
SENSORS AND TRANSDUCERS
A sensor is a device that measures a physical quantity and converts it into a signal which can be read
by an observer or by an instrument.
For example, a mercury thermometer converts the measured temperature into expansion and
contraction of a liquid which can be read on a calibrated glass tube. A thermocouple converts
temperature to an output voltage which can be read by a voltmeter. For accuracy, all sensors need to
be calibrated against known standards.
In everyday life, sensors are used everywhere such as touch sensitive mobile phones, laptop’s touch
pad, touch controller light, etc. People use so many applications of sensors in their everyday lifestyle
that even they are not aware about it. Examples of such applications are in the field of medicine,
machines, cars, aerospace, robotics and manufacturing plants. The sensitivity of the sensors is the
change of sensor’s output when the measured quantity changes. For example, the output increases 1
volt when the temperature in the thermocouple junction increases 1°C. The sensitivity of the
thermocouple element is 1 volt/°C. To measure very small charges, the sensors should have very high
sensitivity.
The term transducer is commonly used in two senses; the sensor, used to detect a parameter in one
form and report it in another (usually an electrical or digital signal), and the audio loudspeaker, which
converts electrical voltage variations representing music or speech to mechanical cone vibration and
hence vibrates air molecules creating sound.
1. Linearity
Its input vs output characteristics should be linear and it should produce these characteristics in
balanced way.
2. Ruggedness
A transducer should be capable of withstanding overload and some safety arrangements must be
provided with it for overload protection.
3. Repeatability
The device should reproduce the same output signal when the same input signal is applied again and
again under unchanged environmental conditions, e.g., temperature, pressure, humidity, etc.
6. No Hysteresis
It should not give any hysteresis during measurement while input signal is varied from its low value to
high value and vice versa.
7. Residual Reformation
There should not be any deformation on removal of input signal after long period of use.
Selection of Transducer
Following are the factors which need to be considered while selecting a transducer:
High input impedance and low output impedance, to avoid loading effect.
CLASSIFICATION OF TRANSDUCERS
On the basis of application transducer may be classified as primary and secondary transducer.
When the input signal is directly sense by the transducer, the non-electrical energy is converted
into electrical energy directly then, this type of transducer is known as primary transducer. For
example thermistor, senses the temperature directly and causes the change in resistance with
the change in temperature.
When the input signal is first sensed by some sensor or detector, then its output signal is feed
to the other instrument as an input. The output of this instrument is given as the input of
transducer for converting into electrical energy. This type of transducer is in secondary
transducer classification. For example in the case of pressure measurement, we use bourdon
tube to convert pressure into displacement, then the pressure is converter into output voltage
with the help of LVDT. Here the secondary transducer is LVDT. We will discuss the bourdon tube
and LVDT in this chapter.
1. Active transducer
2. Passive transducer
Active transducer does not require any auxiliary power source to produce their output. It is also
called self generating type transducer.
Active transducer is also known as self generating type transducer. It develops their own
voltage or current from the physical phenomenon being measured. Active transducers generate
electric current or voltage directly in response to environmental stimulation. Examples of active
transducers are thermocouples and piezoelectric accelerometers. Thermocouples produce a
voltage related to a temperature of two metals and if the two junctions are at different
temperatures, electricity is generated.
Passive transducers are also known as externally powered transducers. It derives the power
required for energy conversion from an external power source. Passive transducers produce a
change in some passive electrical quantity, such as capacitance, resistance, or inductance, as
result of stimulation. These usually require additional electrical energy for excitation. A simple
example of a passive transducer is a device containing a length of wire and a moving contact
touching the wire. The position of the contact determines the effective length of the wire,
varying the resistance of the length of wire. Other examples of passive transducers are strain
gauges, resistance temperature detectors (RTDs), and thermistors.
1. Analog transducer
2. Digital transducer
1. Resistive transducer
2. Capacitive transducer
3. Inductive transducer
4. Photoelectric transducer
5. Photovoltaic transducer
1. Linear displacement
2. Rotary displacement
1. Transducer
2. Inverse transducer.
70|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Characteristics of Transducer
Known values of the measured (input) are applied to a sensor (measurement system) for the
purpose of observing the sensor (system) output. The main characteristics of transducer are
given below:
Sensitivity It can be defined as the ratio of the incremental output and the incremental input.
While defining the sensitivity, we assume that the input-output characteristic of the instrument
is approximately linear in that range.
Range The range of the sensor is the maximum and minimum values of applied parameter that
can be measured.
Resolution The smallest difference between measured values that can be discriminated. For
example, it corresponds to the last stable figure on a digital display. This specification is the
smallest detectable incremental change of input parameter that can be detected in the output
signal. Resolution can be expressed either as a proportion of the reading (or the full-scale
reading) or in absolute terms.
Accuracy The accuracy of the sensor is the maximum difference that will exist between the
actual value and the indicated value at the output of the sensor. Again, the accuracy can be
expressed either as a percentage of full scale or in absolute terms.
Linearity The linearity of the transducer is an expression of the extent to which the actual
measured curve of a sensor departs from the ideal curve. Fig. 14.2 shows a somewhat
exaggerated relationship between the ideal, or least squares fit, line and the actual measured
or calibration line. Linearity is often specified in terms of percentage of non-linearity, which is
defined as,
RESISTIVE TRANSDUCER
The transducer whose resistance varies because of the environmental effects such type of transducer
is known as the resistive transducer. The change in resistance is measured by the ac or dc measuring
devices. The resistive transducer is used for measuring the physical quantities like temperature,
displacement, vibration etc.
The measurement of the physical quantity is quite difficult. The resistive transducer converts the
physical quantities into variable resistance which is easily measured by the meters. The process of
variation in resistance is widely used in the industrial applications.
The resistive transducer can work both as the primary as well as the secondary transducer. The primary
transducer changes the physical quantities into a mechanical signal, and secondary transducer directly
transforms it into an electrical signal.
The resistive transducer is designed by considering the variation of the length, area and resistivity of
the metal.
1. Potentiometer – The translation and rotatory potentiometer are the examples of the resistive
transducers. The resistance of their conductor varies with the variation in their lengths which is used
for the measurement of displacement.
72|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
2. Strain gauges – The resistance of their semiconductor material changes when the strain occurs
on it. This property of metals is used for the measurement of the pressure, force-displacement etc.
3. Resistance Thermometer – The resistance of the metals changes because of changes in
temperature. This property of conductor is used for measuring the temperature.
4. Thermistor – It works on the principle that the temperature coefficient of the thermistor
material varies with the temperature. The thermistor has the negative temperature coefficient. The
Negative temperature coefficient means the temperature is inversely proportional to resistance.
1. Both the AC and DC, current or voltage is appropriate for the measurement of variable resistance.
2. The resistive transducer gives the fast response.
3. It is available in various sizes and having a high range of resistance.
POTENTIOMETER
The potentiometers or the “pots”, as it is commonly known in the electric circles, is a three terminal
variable resistor. Out of its three terminals, two of them are fixed and one is a varying (linear / rotary)
terminal.
In the circuit diagram shown below, the terminals of the potentiometer are marked 1, 2 and 3. The
voltage supply is connected across terminals 1 and 3, positive lead to terminal one while negative lead
to terminal three. The terminal 2
is connected to the wiper.
LINEAR POTENTIOMETER
Another easy method than calculating the resistance with the help of a bridge connection is to
connect a constant current source in series with the potentiometer. Thus a voltage will be
developed. This voltage can be measured and hence the resistance, R = V/I.
ROTARY POTENTIOMETER
The rotary type potentiometers are used mainly for obtaining adjustable supply voltage to a part of
electronic circuits and electrical circuits. The volume controller of a radio transistor is a popular
example of a rotary potentiometer where the rotary knob of the potentiometer controls the supply
to the amplifier.
THERMISTOR
The thermistor is a kind of resistor whose resistivity depends on surrounding temperature. It is
a temperature sensitive device. The word thermistor is derived from the word, thermally sensitive
resistor. The thermistor is made of the semiconductor material that means their resistance lies
between the conductor and the insulator. The variation in the thermistor resistance shows that
either conduction or power dissipation occurs in the thermistor. The circuit diagram of thermistor
uses the rectangular block which has a diagonal line on it.
Construction of Thermistor
The thermistor is made with the sintered mixture of metallic oxides like
manganese, cobalt, nickel, cobalt, copper, iron, uranium, etc. It is available in the
form of the bead, rod and disc. The different types of the thermistor are shown
The bead form of the thermistor is smallest in shape, and it is enclosed inside the solid glass rod to
form probes.
The disc shape is made by pressing material under high pressure with diameter range from 2.5 mm
to 25mm.
75|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The relation between the absolute temperature and the resistance of the thermistor is
mathematically expressed by the equation shown below.
The resistance temperature coefficient of the thermistor is shown in the figure below. The graph
below shows that the thermistor has a negative temperature coefficient, i.e., the temperature is
inversely proportional to the resistance. The resistance of the thermistor changes from 10 5 to 10-
2 at the temperature between -100°C to 400°C.
Types of Thermistor
The thermistor is classified into types. They are the negative temperature coefficient and the
positive temperature coefficient thermistor.
Advantages of Thermistor
4. The upper thermistor limit of the temperature depends on the physical variation of the
material, and the lower temperature depends on the resistance reaching a large value.
5. The self-heating of the thermistor is avoided by minimising the current passes through it.
6. The thermistor is installed at the distance of the measuring circuit. Thus the reading is free from
the error caused by the resistance of the lead.
RESISTANCE THERMOMETER
The resistance thermometer or resistance temperature detector (RTD) uses
the resistance of electrical conductor for measuring the temperature. The resistance of
the conductor varies with the time. This property of the conductor is used
for measuring the temperature. The main function of the RTD is to give a positive
change in resistance with temperature.
The metal has a high-temperature coefficient that means their temperature increases with the
increase in temperature. The carbon and germanium have low-temperature coefficient which
shows that their resistance is inversely proportional to temperature.
Gold and silver are not used for RTD because they have low resistivity. Tungsten has high resistivity,
but it is extremely brittles. The copper is used for making the RTD element. The copper has low
resistivity and also it is less expensive. The only disadvantage of the copper is that it has low linearity.
The maximum temperature of the copper is about 120ºC.
The RTD material is made of platinum, nickel or alloys of nickel. The nickel wires are used for a
limited temperature range, but they are quite nonlinear.
The following are the requirements of the conductor used in the RTDs.
1. The resistivity of the material is high so that the minimum volume of conductor is used for
construction.
2. The change in resistance of the material concerning temperature should be as high as possible.
3. The resistance of the material depends on the temperature.
77|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The resistance versus temperature curve is shown in the figure below. The curves are nearly linear,
and for small temperature range, it is very
evident.
The lead wire is used for connecting the resistance element with the external lead. The lead wire is
covered by the insulated tube which protects it from short circuit. The ceramic material is used as
an insulator for high-temperature material and for low-temperature fibre or glass is used.
The tip of the resistance thermometer is placed near the measurand heat source. The heat is
uniformly distributed across the resistive element. The changes in the resistance vary the
temperature of the element. The final resistance is measured. The below mention equations
measure the variation in temperature.
Linear Approximation
The linear approximation is the way of estimating the resistance versus temperature curve in the
form of the linear equation.
Quadratic Approximation
The quadratic approximation gives the accurate approximation of the resistance temperature curve.
The approximation is expressed in the form of
the quadratic equation.
The resistance thermometer is very less sensitive, and the metal used for making the resistive
element is less expensive.
Symbol
1. Define thermistor and resistive temperature detector briefly with some examples.
2. Explain construction of thermistor with its characteristics.
3. Write down the types and advantages of thermistors.
4. Explain construction of RTD with its operation characteristics.
5. Differentiate between thermistor and RTD.
STRAIN GAUGES
The strain gauge is an electrical transducer; it is used to measure mechanical surface tension. Strain
gauge can detect and convert force or small mechanical displacement into electrical signals. On the
application of force a metal conductor is stretched or compressed, its resistance changes owing to the
fact both length and diameter of conductor change. Also, there is a change on the value of resistivity
of the conductor when it is strained and this property of the metal is called piezo-resistive effect.
Therefore, resistance strain gauges are also known as piezo-resistive gauges. The strain gauges are
used for measurement of strain and associated stress in experimental stress analysis. Secondly, many
other detectors and transducers, for example the load cell, torque meter, flow meter, and
accelerometer employ strain gauge as a secondary transducer.
Let us consider a strain gauge made of circular wire. The wire has the dimensions:
80|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
length L, area A, diameter D before being strained. The material of the wire has a resistivity ρ.
Poisson’s ratio
The gauge factor is defined as the ratio of per unit change in resistance to per unit change in length.
Where, ε= = Strain
Inductive Transducers
The inductive transducers work on the principle of the magnetic induction of magnetic
material. The induction of the magnetic material depends on a number of variables like the
number of turns of the coil on the material, the size of the magnetic material, and the
permeability of the flux path. In the inductive transducers the magnetic materials are used in
the flux path and there are one or more air gaps. The change in the air gap also results in change
in the inductance of the circuit and in most of the inductive transducers it is used for the
working of the instrument. There are two types of inductive transducers, they are given below:
The amount of voltage change in either secondary winding is proportional to the amount of movement
of the core. Hence, we have an indication of amount of linear motion. By noticing whether output
voltage is increased or decreased, we can determine the direction of motion.
Advantages of LVDT
Linearity is good up to 5 mm of displacement.
Output is rather high. Therefore, immediate amplification is not necessary.
Output voltage is stepless and hence the resolution is very good.
Sensitivity is high (about 40 V/mm).
It does not load the measurand mechanically.
It consumes low power and low hysteresis loss also.
Disadvantages of LVDT
Uses of LVDTs
The following are the major applications of LVDTs.
1. It is used for measuring the displacement having a range from few mm to cm. The LVDT directly
converts the displacement into an electrical signal.
2. It is also used as the secondary transducer. The LVDT is used as a device for measuring the force,
weight and pressure. Some of the LVDT used for measuring the load and pressure.
3. The LVDT sensor works as the main transducer, and that changes dislocation to electrical signal
straight.
4. Some of these transducers are used to calculate the pressure and load.
6. Other applications like power turbines, hydraulics, automation, aircraft, and satellites.
CAPACITIVE TRANSDUCERS
The capacitive transducer is used for measuring the displacement, pressure and other physical
quantities. It is a passive transducer that means it requires external power for operation. The capacitive
transducer works on the principle of variable capacitances. The capacitance of the capacitive
transducer changes because of many reasons like overlapping of plates, change in distance between
the plates and dielectric constant. The capacitive transducer contains two parallel metal plates. These
plates are separated by the dielectric medium which is air, material, gas or liquid. In the normal
capacitor the distance between the plates are fixed, but in capacitive transducer the distance between
them are varied.
The capacitive transducer uses the electrical quantity of capacitance for converting the mechanical
movement into an electrical signal. The input quantity causes the change of the capacitance which is
directly measured by the capacitive transducer.
83|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The capacitors measure both the static and dynamic changes. The displacement is also measured
directly by connecting the measurable devices to the movable plate of the capacitor. It works on with
both the contacting and non-contacting modes.
Principle of Operation
The equations below express the capacitance between the plates of a capacitor
εr – relative permittivity
The schematic diagram of a parallel plate capacitive transducer is shown in the figure below.
of transducer is given as
Where, C – capacitance
The capacitive transducer is mainly used for measurement of linear displacement. The capacitive
transducer uses the following three effects.
2. The change in capacitance is because of the change in distances between the plates.
3. The capacitance changes because of dielectric constant.
The equation below shows that the capacitance is directly proportional to the area of the plates. The
capacitance changes correspondingly with the change in the position of the plates.
as
The sensitivity of the displacement is constant, and therefore it gives the linear relation between the
capacitance and displacement.
The capacitive transducer is used for measuring the angular displacement. It is measured by the
movable plates shown below. One of the plates of the transducer is fixed, and the other is movable.
The angular movement changes the capacitance of the transducers. The capacitance between them is
maximum when these plates overlap each other. The
maximum value of capacitance is expressed
as
θ – Angular displacement in radian. The sensitivity for the change in capacitance is given as
The 180° is the maximum value of the angular displacement of the capacitor.
The capacitance of the transducer is inversely proportional to the distance between the plates. The
one plate of the transducer is fixed, and the other is movable. The displacement which is to be
measured links to the movable plates.
1. The capacitive transducer uses for measurement of both the linear and angular displacement. It
is extremely sensitive and used for the measurement of very small distance.
2. It is used for the measurement of the force and pressures. The force or pressure, which is to be
measured is first converted into a displacement, and then the displacement changes the capacitances
of the transducer.
3. It is used as a pressure transducer in some cases, where the dielectric constant of the transducer
changes with the pressure.
4. The humidity in gases is measured through the capacitive transducer.
5. The transducer uses the mechanical modifier for measuring the volume, density, weight etc.
The accuracy of the transducer depends on the variation of temperature to the high level.
87|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
Piezoelectric Transducers
The piezoelectric transducers work on the principle of piezoelectric effect. When mechanical
stress or forces are applied to some materials along certain planes, they produce electric
voltage. This electric voltage can be measured easily by the voltage measuring instruments,
which can be used to measure the stress or force.
There are certain materials that generate electric potential or voltage when mechanical strain
is applied to them or when the voltage is applied to them, they tend to change the dimensions
along certain plane. This effect is called as the piezoelectric effect. This effect was discovered
in the year 1880 by Pierre and Jacques Curie. Some of the materials that exhibit piezoelectric
effect are quartz, Rochelle salt, polarized barium titanate, ammonium dihydrogen, ordinary
sugar etc.
The physical quantities like stress and force cannot be measured directly. In such cases the
material exhibiting piezoelectric transducers can be used. The stress or the force that has to be
measured is applied along certain planes to these materials. The voltage output obtained from
these materials due to piezoelectric effect is proportional to the applied stress or force. The
output voltage can be calibrated against the applied stress or the force so that the measured
value of the output voltage directly gives the value of the applied stress or force.
The piezoelectric effect can be made to response mechanical deformations of the material in
many different modes. The modes can be thickness expansion, transverse expansion, thickness
shear and face shear. The mode of motion affected depends on the shape of the body relative
to the crystal axis and location of the electrodes. The mechanical deformation generates a
charge and this charge appears as a voltage across the electrodes.
The magnitude and polarity of the induced surface are proportional to the magnitude and
direction of the applied force F. The polarity of induced charge depends upon the direction of
applied force. The charge is given by,
Q=d×F
F = applied force
88|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
1. Thickness shear
2. Face shear
3. Thickness expansion
4. Transverse Expansion
Application
They are used for quality assurance, process control and for research and development in many
different industries.
This transducer works on the principle of Hall Effect. When a current conducting material is
placed in the transverse magnetic field then the difference of potential is produced between
the opposite edges of the conductor. This effect is known as Hall Effect. The magnitude of the
voltage depends upon the current, the strength of magnetic field and the property of the
conductor.
Hall
strip EH
2
4
3
Transve
rse magnetic field
Let the current pass through edge 1 and 2 of the conductor and the output leads is connected
to edge 3 and 4. The edge 3 and 4 are at same potential when there is no transverse magnetic
field passing through the conductor. When a transverse magnetic field passing through the
conductor, an output voltage appears across the output leads. This output voltage is
proportional to the current and the field strength. The output voltage is given by,
EH = KH .t.I.B
89|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
B = Flux density
The voltage produces may be used for measurement of the current I or the magnetic field
strength B.
The Hall Effect is used for measurement of the location or displacement of the elements. The Hall
Effect transducer is located in the gap adjacent the permanent magnet. The field strength produced
in the gap, adjacent to the permanent magnet in the gap, where the Hall Effect element is located,
is varied by changing the position of a ferromagnetic plate. The output voltage of the Hall Effect
transducer is proportional to the field strength in the gap which is a function of the position of the
ferromagnetic plate from the structure.
3. Measurement of Current
The Hall Effect transducer used to measure the current in conductor without the need for
interrupting the circuit and without making electrical connection between the conductor
and the meter. The current passes through the conductor and sets up a magnetic field
surrounding the conductor. The magnetic field is proportional to the current.
CHAPTER-8
OSCILLOSCOPE
The cathode ray oscilloscope (CRO) is a type of electrical instrument which is used for showing the
measurement and analysis of waveforms and others electronic and electrical phenomenon. It is a very
fast X-Y plotter shows the input signal versus another signal or versus time. The CROs are used to
analyze the waveforms, transient, phenomena, and other time-varying quantities from a very low-
frequency range to the radio frequencies.
The CRO is mainly operated on voltages. Thus, the other physical quantities like current, strain,
acceleration, pressure are converted into the voltage with the help of the transducer and thus
represent on a CRO. It is also used for knowing the waveforms, transient phenomenon, and other time-
varying quantity from a very low-frequency range to the radio frequencies.
The CRO has Stylus (i.e., a luminous spot) which moves over the display area in response to an input
voltage. This luminous spot is produced by a beam of electrons striking on a fluorescent screen. The
normal form of the CRO uses a horizontal input voltage which is an internally generated ramp voltage
called “time base”.
The horizontal voltage moves the luminous spot periodically in a horizontal direction from left to right
over the display area or screen. The vertical voltage is the voltage under investigation. The vertical
voltage moves the luminous spot up and down on the screen. When the input voltage moves very fast
on the screen, the display on the screen appears stationary. Thus, CRO provides a means of the
visualizing time-varying voltage.
The cathode ray tube is the vacuum tube which converts the electrical signal into the visual signal. The
cathode ray tube mainly consists the electron gun and the electrostatic deflection plates (vertical and
horizontal).The electron gun produces a focused beam of the electron which is accelerated to high
frequency.
The vertical deflection plate moves the beams up and down and the horizontal beam moved the
electrons beams left to right. These movements are independent to each other and hence the beam
may be positioned anywhere on the screen.
91|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
The electron gun emits the electrons and forms them into a beam. The electron gun mainly consists of
a heater, cathode, a grid, a pre-accelerating anode, a focusing anode and an accelerating anode. For
gaining the high emission of electrons at the moderate temperature, the layers of barium and
strontium is deposited on the end of the cathode.
After the emission of an electron from the cathode grid, it passes through the control grid. The control
grid is usually a nickel cylinder with a centrally located co-axial with the CRT axis. It controls the
intensity of the emitted electron from the cathode.
The electron while passing through the control grid is accelerated by a high positive potential which is
applied to the pre-accelerating or accelerating nodes.
The electron beam is focused on focusing electrodes and then passes through the vertical and
horizontal deflection plates and then goes on to the fluorescent lamp. The pre-accelerating and
accelerating anode are connected to 1500V, and the focusing electrode is connected to 500V. There
are two methods of focusing on the electron beam. These methods are
Electrostatic focusing
Electromagnetic focusing.
3. Deflecting Plate
The electron beam after leaving the electron gun passes through the two pairs of the deflecting plate.
The pair of plate producing the vertical deflection is called a vertical deflecting plate or Y plates, and
the pair of the plate which is used for horizontal deflection is called horizontal deflection plate or X
plates.
The front of the CRT is called the face plate. It is flat for screen sized up to about 100mm×100mm. The
screen of the CRT is slightly curved for larger displays. The face plate is formed by pressing the molten
glass into a mould and then annealing it.
The inside surface of the faceplate is coated with phosphor crystal. The phosphor converts electrical
energy into light energy. When an electronics beam strike phosphor crystal, it raises their energy level
and hence light is emitted during phosphorous crystallization. This phenomenon is called fluorescence.
5. Glass Envelope
It is a highly evacuated conical shape structure. The inner surface of the CRT between the neck and the
screen is coated with the aquadag. The aquadag is a conducting material and act as a high-voltage
electrode. The coating surface is electrically connected to the accelerating anode and hence helps the
electron to be the focus.
After moving from the accelerating anode, the beam comes under the effect of the deflecting plates.
When the deflecting plate is at zero potential, the beam produces a spot at the centre. If the voltage
is applied to the vertical deflecting plate, the electron beam focuses at the upward and when the
voltage is applied horizontally the spot of light will be deflected horizontally.
, where
For measurement of current, the current under measurement is passed through a known non
inductive resistance and the voltage drop across it is measured by CRO, as mentioned above. The
current can be determined simply by dividing the voltage drop measured by the value of non inductive
94|ELECTRICAL MEASUREMENT AND INSTRUMENTATION
resistance. When the current to be measured is of very small magnitude, the voltage drop across non-
inductive resistance (small value) is usually amplified by a calibrated amplifier.
Measurement of Frequency
It is interesting to consider the characteristics of patterns that appear on the screen of a CRO when
sinusoidal voltages are simultaneously applied to the horizontal and vertical plates. These patterns are
called Lissajous patterns.
Lissajous patterns may be used for accurate measurement of frequency. The signal, whose frequency
is to be measured, is applied to the Y-plates. An accurately calibrated standard variable frequency
source is used to supply voltage to the X-plates, with the internal sweep generator switched off. The
standard frequency is adjusted until the pattern appears as a circle or an ellipse, indicating that both
signals are of the same frequency. Where it is not possible to adjust the standard signal frequency to
the exact frequency of the unknown signal, the standard is adjusted to a multiple or sub-multiple of
the frequency of the unknown source so that the pattern appears stationary.
Let us consider an example. Suppose sine waves are applied to X and Y plates as shown in figure
below. Let the frequency of wave applied to Y plates is twice that of the voltage applied to the X plates.
This means that the CRT spot travels two complete cycles in the vertical direction against one of the
horizontal direction.
The two waves start at the same instant. A Lissajous pattern may be constructed in the usual way
and an 8 shaped pattern with two loops is obtained. If the two waves do not start at the same instant
we get different pattern for the same frequency ratio. The Lissajous pattern for the other frequency
ratios can be similarly drawn. Some of these patterns are shown in figures.
It can be shown that for all the above cases, the ratios of the two frequencies is
The above rule, however, does not hold for the Lissajous patterns with free ends as shown in figure.
The simple rule mentioned above needs the following modifications:
Two lines are drawn, one horizontal and the other vertical so that they do not pass through any
intersections of different parts of the Lissajous curve. The number of intersections of the horizontal
and the vertical lines with the Lissajous curve are individually counted. The frequency ratio is given by
The ratio of frequencies when open ended Lissajous patterns are obtained can also be found by
treating the open ends as half tangencies as shown in figure.
There are some restrictions on the frequencies which can be applied to the deflection plates. One
obviously, is that the CRO must have the bandwidth required for these frequencies. The other
restriction is that the ratio of the two frequencies should not be such as to make the pattern too
complicated otherwise determination of frequency would become difficult. As a rule, ratios of as high
as 10:1 and as low as 10:9 can be determined comfortably.