Operation manual of RePure-A
CONTENTS
CONTENTS ...................................................................................................................... - 3 -
Chapter 1 Safety Guidelines ......................................................................................... - 4 -
Chapter 2 Installation .....................................................................................................- 5 -
2.1 Package Contents Verification ............................................................................- 5 -
2.2 Normal Operating Conditions ............................................................................. - 5 -
2.3 Transportation and Storage Requirements ........................................................ - 5 -
2.4 Power Requirements .......................................................................................... - 5 -
Chapter 3 Instrument Characteristics ............................................................................- 6 -
3.1 Instrument Construction ......................................................................................- 6 -
3.2 Specialty ..............................................................................................................- 8 -
Chapter 4 Operation Instruction .................................................................................. - 10 -
4.1 Power on ........................................................................................................... - 10 -
4.2 Menu Structure ..................................................................................................- 11 -
4.3 File Instructions ................................................................................................. - 11 -
4.3.1 Create or Edit File ...................................................................................- 12 -
4.3.2 Delete File ...............................................................................................- 15 -
4.3.3 Copy File .................................................................................................- 15 -
4.3.4 Rename File ........................................................................................... - 15 -
4.3.5 Run File .................................................................................................. - 15 -
4.4 Setting ............................................................................................................... - 17 -
4.4.1 Running Setting ...................................................................................... - 17 -
4.4.2 System Setting ....................................................................................... - 18 -
4.4.3 E-mail Setting ......................................................................................... - 18 -
4.4.4 GLP Report .............................................................................................- 19 -
4.4.5 Device Information ................................................................................. - 19 -
4.4.6 Software Update .....................................................................................- 20 -
4.4.7 Tm Calculator ......................................................................................... - 20 -
4.4.8 Calibration Log ....................................................................................... - 20 -
4.5 Incubate ............................................................................................................ - 20 -
4.6 Log In ................................................................................................................ - 21 -
4.6.1 New User ................................................................................................ - 22 -
4.6.2 Change Password and Delete User ....................................................- 22 -
Chapter 5 Maintenance and Troubleshooting .............................................................- 23 -
5.1 Machine Maintenance .......................................................................................- 23 -
5.2 Troubleshooting ................................................................................................ - 24 -
5.3 Notes ................................................................................................................. - 24 -
5.4 Error Messages and Correspond Solutions ..................................................... - 25 -
5.5 Cause of the Abnormal Phenomenon and Correspond Solution .................... - 25 -
Chapter 6 After-sales Service ......................................................................................- 26 -
-3-
Operation manual of RePure-A
Icon Meaning
High temperature warning:
To avoid burning, please don’t touch directly all areas labeled with safety warning and
the hot areas described in manual.
Electric shocks Warning:
To avoid electric shock accidents, please operate strictly according to the request of
electric shocks warning.
Note:
Please note that reminder contains important information and be sure to read them
carefully. Otherwise it will cause the instrument to not work properly or even damage
the instrument.
2. Safe Use
Please read the following information before using the instrument and be sure to comply with the
following basic security measures. Failure to follow the measures or the other warnings in the
manual will affect the normal work of the instrument, even damage the equipment and hurt people.
1) Prohibit using the instrument in humid, dusty, high temperature, magnetic environment.
2) Prohibit opening the cover of instrument or touching the inner device of instrument.
3) Prohibit blocking the air flow vents and be aware of gloves or rags sucked into air flow vent in the
bottom of instrument.
4) Please keep instrument clean, and maintain it regularly.
Note: Please disconnect the power immediately and contact the supplier or consult certified
mechanic for assistance when any of the following occurs.
1) Instrument be moistened by rain or water, or any other liquid.
2) Instrument can’t work properly, especially any abnormal sound or odor appears.
3) Instrument’s function changes obviously.
-4-
Operation manual of RePure-A
Chapter 2 Installation
2.1 Package Contents Verification
Please open box and check the following list when you receive the RePure-A PCR instrument.
Name Quantity
Instrument 1
Power line 1
In the event of discrepancy, please keep the original box and contact with us immediately.
Warning: To avoid electric shock accident, instrument must be fitted with a reliable grounding.
-5-
Operation manual of RePure-A
Status LED
Lock column
Reaction block
LCD display
Air vents
USB A port
-6-
Operation manual of RePure-A
Cooling fins
Air vents
Main board fan
-7-
Operation manual of RePure-A
3.2 Specialty
1. Long service life Peltier heating units.
2. Reinforced aluminum module with anodizing technology can keep rapid heating-conducting
property and have enough corrosion resistance.
3. High heating and cooling rate, max. ramping rate 5 ℃/s, can save your precious time.
4. Self-adapting pressure hot lid makes closing lid and tightening lid in one step.
5. Air channel is in front and back and it allows machine placed side by side.
6.It has Android operation system and 10.1 inch capacitive touch screen. It has graphical menu
navigation interface and operation is very simple.
7. Built-in multiple standard program file template, can quickly edit the required files.
8. The running program and left time can be displayed in real time , allow to edit file when
program is running.
9. One-click quick incubation function can meet experiment’s needs such as denaturation,
enzyme cutting/enzyme-link and ELISA.
10. Hot lid temperature and hot lid work mode can be set to meet different experiment’s need.
11. Automatic restart after power failure. When power is restored it can continue to run
unfinished program.
12. Support USB to store and copy PCR data, user can control PCR by USB mouse.
13. Update software by USB flash.
14. Wi-Fi module built-in, one unit can control multiple PCR machine through computer or cell
phone with internet connection.
15. Email notification after experiment is over.
-8-
Operation manual of RePure-A
3.3 Performance
Model RePure-A
Capacity 96×0.2ml
Reaction Volume 5-100μl
Temperature Range 0-105℃
Max. Heating Rate 5℃/s
Uniformity ≤±0.2℃
Accuracy ≤±0.1℃
Display Resolution 0.1℃
Temperature Control Block/Tube
Ramping Rate
Adjustable 0.01-5℃/s
Gradient Temp.
Range 30-105℃
Gradient Spread Horizontal: 0.1-42℃
Hot Lid Temperature 30-115℃
Number of Programs 200,000+(USB FLASH)
Max. No. of Step 30(Multiple nested PCR experiments)
Max. No. of Cycle 200(Up to 100,000 in nested cycles)
Time Increment/
Decrement 1 Sec-600 Sec, Long PCR is available
Temp. Increment/
Decrement 0.1-10.0℃, Touchdown PCR is available
Pause Function Yes
Auto Data Protection Yes
Hold at 4℃ Forever
Record hot lid and module temperatures in real time,
Temperature Curve
with a 60-minute recording time on a single page and a sliding view.
Record the running information of each step of the program,
GLP Report
and can export CSV, PDF format report.
File Encryption
Individual program files can be encrypted
Function
User Management
Three levels of access (administrator, user, guest)
Privilege
Email Notification Email notification after experiment is over
Lock Screen Prevents accidental termination of the program being experimented
Function on.
PC-LINK Function Simultaneous on-line monitoring of more than 250 instruments
Mobile Phone APP Simultaneous on-line monitoring of more than 250 instruments
Communication
USB2.0, LAN, WIFI
Interface
LCD 10.1inch, 1280×800 Pixels,
Dimensions 385mm× 270mm× 255mm (L×W×H)
Weight 10kg
Power Supply 100-240VAC, 50/60Hz, 600W
-9-
Operation manual of RePure-A
4.1 Power on
Connect power cord to the back of instrument then plug into power source and switch to "-". .
When the instrument is on, the buzzer should be heard and the LCD screen lights up to show
the Boot screen (Figure 3). After that, the operating system starts up and then goes to the
Self-Test screen (Figure 4).During this phase, the LCD screen displays the product name,
company LOGO. When it’s over, the system performs the Home screen (Figure 5) as shown
below.
- 10 -
Operation manual of RePure-A
File Library
Home
New Directory
Directory Option
File Option
Edit
Incubate
Settings
Running Options
GLP Report
System Log
Look Screen
TM Calculator
Calibration Log
Time and Time Zone
Sound Settings
WIFI Connection
Software Updates
Firmware Updates
Switch Language
Factory Mode
Email Alerts
Remote Access
Device Information
Modify Background
User
Home
Logout
Login
Registered User
Change
Password
Delete User
- 11 -
Operation manual of RePure-A
Touch “New File” button in the Home screen to create new file (Fig. 1).
Or in the Home screen, touch “File Library” button to enter File Library screen (Fig 2),then
Select a file in the File Library screen, touch “Edit” button to enter Edit File screen (Fig 3).
- 12 -
Operation manual of RePure-A
2) Run mode
There are two run modes: standard mode and fast mode. Users can choose the mode
according to their actual needs. The fast mode takes less experimental time than the standard
mode.
3) Advanced
Select a desired step, touch “Advanced” button to enter Advanced screen (Figure 4).
Parameters can be edited by selecting among Gradient, Ramp Rate, Temperature Increment or
Time Increment. After entering Gradient values the gradient distribution table will be shown at
bottom of the screen. Touch “OK” to save changes and return Edit File screen (Figure 5).
- 13 -
Operation manual of RePure-A
Figure 4 Advanced
4) Save file
In the Edit File screen , select the file name edit box, input volume of reagent , lid temperature,
as shown in Fig. 6, input the "Save as" button after completion, select the save path as shown in
Fig. 7, or click the "Run" button to run the file directly.
- 14 -
Operation manual of RePure-A
In the File Library screen, touch File Option button to select the file to delete, and the query
dialog window will pop up when touching “Delete file” button, and delete this file by touching
OK(Fig. 8). Only one delete per time is allowed to prevent any files being mistakenly deleted.
In the File Library screen, touch File Option button to select the file to copy and touch Copy file
button. Then enter the save path and touch Paste file button to finish copy.
4.3.4 Rename File
In the File Library screen, touch File Option button and select the file to rename. Then touch
Rename file button and enter new file name in the dialog window to rename files.
4.3.5 Run File
In the File Library screen, select the file needed and touch Run button or touch the "Run" button
on the new file/edit file screen to pop up the prompt dialog box. Then press the "OK" button to
- 15 -
Operation manual of RePure-A
enter the running file screen,Click the button on the right to switch to the next page (Fig. 9).
Touch Home button to back the Home screen, or touch Circle in the middle of the screen to
enter the running file screen (Fig. 10).
- 16 -
Operation manual of RePure-A
4.4 Setting
In the Home screen, touch Setting button to enter the Setting screen (Figure 11).
In the Setting screen, select the Running options to enter the Running setting screen (Fig. 12).
1) Hot lid temperature
Hot lid temperature setting range:30-115℃,The system default:105℃.
2) Hot lid work mode setting
In total there are 2 working modes and default mode is that the hot lid will be on when file
running.
3)Temperature control mode
The temperature control mode contains block control mode and tube control mode. The block
control mode is suitable for performing normal PCR and the tube control mode is for
experiments that require higher environment conditions. The system default is tube control
mode.
- 17 -
Operation manual of RePure-A
In the Setting screen, can set time and date, language, sound.
1) Time and date
Set the system time and date(Fig. 13).
2) Sound
● Keyboard sound- sound when touch button.
● Alarm sound- sound when system error occurs.
● File end sound- sound when file running is complete.
● Temp to reach sound- sound when the target temperature reaches.
In the Setting screen, select E-mail alerts to enter the E-mail Setting screen.
1) Status of email alert
After program is over it will automatically send email if chose Enable.
2) Test email
Test email function is to test whether email can send or not.
3) Save
Touch “OK” to save information and email setting record.
- 18 -
Operation manual of RePure-A
Record of every file name, time of start run this file, block number, source files path (Fig. 15).
● GLP Report- Record the time and temperature for each step.
Show the machine name(can modify), the SN., the control version and UI version (Figure 16).
- 19 -
Operation manual of RePure-A
Copy the new software to the root directory, insert the USB flash disk into the USB port, then
touch “Software update”, return to Android desktop when updating, and immediately reopen the
updated App .
4.4.7 Tm Calculator
Touch the fields to enter concentrations and the primer sequences,then touch Tm calculator (Fig.
17) to calculate the melting temperatures.
Click the "Calibration Log" button in the setting interface, as shown in Figure 18.
4.5 Incubate
In the Home screen, touch Incubate button to enter the Incubate wizard screen (Fig. 19).
- 20 -
Operation manual of RePure-A
In the Incubate wizard screen, select the desired block and enter the block temperature and the
hold time. Then touch Start button to enter Incubate Running screen (Fig. 20).
4.6 Log In
In the Home screen, touch User button to enter the User screen (Fig. 21).
NOTE: Administrator user's initial password is “123456”, you can change the password
after logged in.
- 21 -
Operation manual of RePure-A
In the User screen, touch New user button to enter the New User screen (Fig. 22). Enter user
name and password, then touch OK button to enter the User screen. In the User screen,select
the Registered user who needs to log in and press the "Login" button(Fig. 23).
Only after login can passwords be changed and users be deleted, can other users be deleted
from the login administrator account.
- 22 -
Operation manual of RePure-A
A. Clean the holes on reaction block with neutral soap solution (Do not use any solvent
containing strong alkali, strong alcohol and organic solution)
B.Spaces near the left, right air vents and under the machine should be always kept free. It is
very important to clean the dust around air vents regularly.
C. Clean the reaction block regularly and remove any residue inside cavities to prevent any
affects to the temperature control (we suggest to use soft cloth).
2) Replace fuse
The machine has two fuses. Replace it according to the following once damage occurs.
Note: Please contact your supplier for repair when encountering any problem replacing
fuses.
- 23 -
Operation manual of RePure-A
5.2 Troubleshooting
1) Dissatisfactory Results
Biological, programmatic and hardware problems may contribute to any unsatisfied results of
experiment. In order to distinguish the hardware problems from other possible problems, the
machine is equipped with built-in self-test hardware and self-diagnosis software. Below is the
detailed description. According to the experiences, most problems are related to biological and
programmatic factors.
FAQ as follows:
A. Incorrect or insufficient reactants.
B. Denaturation temperature is too high or too low. It is suggested to set the temp. within range
of 90-95℃ for 40 seconds. The duration can be adjusted in accordance with reaction volume.
C. Annealing temperature is too high or too low. It should be between 55-70℃ and 20 to 30
chains.
D. Reactant concentration is too high or too low.
E. Preparation process without special treatment.
F. Time and temperature value in progress are not appropriate.
G. The temperature of sample is slightly too low while the temp. of block is slightly too high.
H. Check whether the PCR tubes are well placed. Smearing little mineral oil on the surface of
holes will increase the thermal conductivity.
2) Machine’s self-test and self-diagnosis function
The machine will run the self-test program when booting. The software and hardware of
machine will show the results so as to inform users the potential problems, minimize the failure
of experiment and display the error message when problem occurs.
5.3 Notes
1) Power
A. No special requirement for the power supply. Any AC power source within range of 100-240V
is applicable. However, to prevent from causing any damage to the machine, it is better to apply
low voltage fluctuation power sources, otherwise consider to install power supply regulator.
B. It’s prohibited to cut off the power to terminate a running experiment. It’s very harmful to the
machine.
2) LCD screen
Avoid using UV disinfection to clean the machine to prevent any damage.
Avoid any bump or scratch to the LCD screen while using.
3) Note about cleaning
Avoid any liquid getting inside the machine while cleaning the base of machine. Due to the
possible usage of radioactive substance during experiment, please carefully handle cleaning. It
is not suitable to use the machine in moisture or hot environment.
- 24 -
Operation manual of RePure-A
Note: Please do read the notes carefully! You may damage the machine if don’t operate
according to the above requirements.
Note: If the above problems can’t be solved, please contact the supplier in time.
- 25 -
Operation manual of RePure-A