Polymer supplier selection
Polymer supplier selection
INTRODUCTION
While the study of extractables and leachables (E&L) is being demanded by the regulatory
authorities, the information gathered is important for many other aspects of medical device
development and release including product design, process optimization, troubleshooting
and quality.
Whereas the regulators are dictating that this information is to be used to help ascertain the safety,
help direct the biological testing, E&L testing provides a wealth of information covering the full
product lifecycle. Figure 1. below illustrates a simplified view of a plastics medical device lifecycle.
One can consider all the E&L present within a device as either Intentionally or Non-Intentionally
Added Substances (IAS and NIAS, respectively) from the various stages of the lifecycle. If they
are intentionally added, they have a purpose, so simply removing them is not a trivial undertaking
– you need to satisfy their purpose another way. Non-intentionally added substances can be
considered as removable without an impact on the devices’ intended properties, however, the
source of these can be more
difficult to pinpoint, and sometimes they can be the product of an IAS – either a degradation
product or the result of a chemical reaction caused through the various stresses a device
undergoes during its production.
1
[ WHITE PAPER ]
300
producers for medical use is a necessity. Unfortunately, there 100 ppm
have been examples where medical device manufacturers 200 ppm
200
have switched from a high-grade to a lower grade polymer
resulting in harm to patients. While there is no accepted 400 ppm
100
definition of a medical grade polymer, material manufacturers
turn to chemical and material characterization techniques
to provide assurances to those downstream. As regulators 0
0 1 2 3
are now pointing out, characterization from the supply chain
is key in understanding the final product’s risk profile. There
Figure 3. Heatflow versus time for polyamide 6 film stabilized with different
has been a standard created by the VDI, which defines and
amounts of Irganox 1098 in air at 120OC.
describes what is to be understood by the term “Medical
Grade Plastics”.¹
MEDICAL DEVICE MANUFACTURERS healthcare greenhouse gases emissions are 2.2 GT/year,
The regulations are increasingly expanding to include more of almost 5% of the total global emissions. In the United States
the supply chain, but the burden remains on the company that alone, 14,000 metric tons of healthcare waste is generated
places the product on the market, typically the medical device per day. This not only has an environmental cost, but also
manufacturer. In designing out potentially harmful chemicals an immediate monetary cost too, recent reports stating
from the device, it is critical to understand what impact this is that despite efforts to reduce spending on waste, it’s still
going to have in the intended application. around 25% of healthcare spending in the US and healthcare
spending itself is at 18% of GDP.²
A key piece here is durability testing. Also referred to as
fatigue or stress testing, durability tests ensure that devices This being said, there are efforts to introduce more
are fit for purpose – most importantly that they do not fail sustainability to the medical device market.
unintentionally in humans and lead to severe consequences.
The primary purpose is to physically and repetitively load
SUMMARY
(with forces, torques, rotations, strains, displacements or
Developing materials for medical devices is a complex
pressures) the devices, or portions of it, in a way that is very
challenge. The safest route to satisfy regulations is to simplify
similar to the loading encountered in the body over time.
the formulation but doing this at the same time as improving
Implantable medical devices that withstand repeated loading
properties is counter intuitive. The regulations themselves
throughout their lifetimes such as spinal implants, breast
recognize this, and the increased emphasis on quality not only
implants or heart valves must be thoroughly evaluated for
of the final product, but of the product development process
performance and reliability in many aspects. For example,
itself. Therefore, the entire value chain will be working closer
the durability and fatigue resistance of the implant can be
together to satisfy the product demands and to generate the
evaluated by performing a complete device fatigue test-to-
chemical and material characterization required to ensure
success to prove adequate lifetime such as ISO 5840 for heart
product safety.
valves, or test-to-failure of breast implants as defined in ISO
14607. Insights from tests such as these are crucial for both
References
the design optimization to minimize failure risks and final
1. VDI, Medical Grade Plastics, July 2019.
validation that is required for regulation.
2. W.H. Strank, T.L. Rogstad, N. Parekh, JAMA, 2019;322(15):1501-
1509. doi:10.1001/jama.2019.13978, https://jamanetwork.com/
journals/jama/article-abstract/2752664.
END-OF-LIFE
With the focus on patient safety being of paramount
importance, the sustainability aspects of the healthcare
industry make for stark reading. Most, if not all, the steps
taken to reduce contamination, to reduce risk, such as single-
use products, incineration or landfill rather than reclaiming/
recycling adversely impacts the environment. The global
Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
Waters and The Science of What’s Possible are trademarks of Waters Corporation. All other
trademarks are the property of their respective owners.
T: 1 508 478 2000
F: 1 508 872 1990
©2021 Waters Corporation. Produced in the U.S.A. June 2021 720007295EN GJ-PDF waters.com