WARNING
This equipment generates, uses and can radiate radio frequency energy, and if not
installed and used in accordance with the instruction manual, may cause interference
to radio communications. As temporarily permitted by regulation it has not been tested
for compliance with limits for Class A computing devices pursuant to Subpart J of
Part 15 of FCC Rules, which are designed to provide reasonable protection against such
interference. Operation of this equipment in a residential area is likely to cause
interference, in which case the user, at his own expense, will be required to make
whatever measures may be required to correct the interference.
WARRANTY
ORTEC* warrants that the items will be delivered free from defects in material or workmanship. ORTEC makes
no other warranties, express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE.
ORTEC’s exclusive liability is limited to repairing or replacing at ORTEC’s option, items found by ORTEC to be
defective in workmanship or materials within one year from the date of delivery. ORTEC’s liability on any claim
of any kind, including negligence, loss, or damages arising out of, connected with, or from the performance or
breach thereof, or from the manufacture, sale, delivery, resale, repair, or use of any item or services covered by
this agreement or purchase order, shall in no case exceed the price allocable to the item or service furnished or
any part thereof that gives rise to the claim. In the event ORTEC fails to manufacture or deliver items called for
in this agreement or purchase order, ORTEC’s exclusive liability and buyer’s exclusive remedy shall be release
of the buyer from the obligation to pay the purchase price. In no event shall ORTEC be liable for special or
consequential damages.
Quality Control
Before being approved for shipment, each ORTEC instrument must pass a stringent set of quality control tests
designed to expose any flaws in materials or workmanship. Permanent records of these tests are maintained for
use in warranty repair and as a source of statistical information for design improvements.
Repair Service
If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted
in advance of its return so that a Return Authorization Number can be assigned to the unit. Also, ORTEC must
be informed, either in writing, by telephone [(865) 482-4411] or by facsimile transmission [(865) 483-2133], of the
nature of the fault of the instrument being returned and of the model, serial, and revision ("Rev" on rear panel)
numbers. Failure to do so may cause unnecessary delays in getting the unit repaired. The ORTEC standard
procedure requires that instruments returned for repair pass the same quality control tests that are used for
new-production instruments. Instruments that are returned should be packed so that they will withstand normal
transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated
ORTEC repair center. The address label and the package should include the Return Authorization Number
assigned. Instruments being returned that are damaged in transit due to inadequate packing will be repaired at
the sender's expense, and it will be the sender's responsibility to make claim with the shipper. Instruments not
in warranty should follow the same procedure and ORTEC will provide a quotation.
Damage in Transit
Shipments should be examined immediately upon receipt for evidence of external or concealed damage. The
carrier making delivery should be notified immediately of any such damage, since the carrier is normally liable
for damage in shipment. Packing materials, waybills, and other such documentation should be preserved in order
to establish claims. After such notification to the carrier, please notify ORTEC of the circumstances so that
assistance can be provided in making damage claims and in providing replacement equipment, if necessary.
*ORTEC® is a registered trademark of Advanced Measurement Technology, Inc. All other trademarks used herein are
the property of their respective owners.
iii
CONTENTS
1. DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1. PERFORMANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2. CONTROLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3. INPUTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.4. OUTPUTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.5. INDICATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.6. ELECTRICAL AND MECHANICAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.7. ACCESSORIES AVAILABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.8. RELATED EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3. INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.1. GENERAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.2. CONNECTION TO POWER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.3. CONNECTING INTO A SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.4. CONNECTING AN EXTERNAL REFERENCE INPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4. CIRCUIT DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
This manual contains up to three levels of safety instructions that must be observed in order to avoid personal
injury and/or damage to equipment or other property. These are:
DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is
not observed.
WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed.
CAUTION Indicates a hazard that could result in property damage if the safety instruction is not
observed.
Please read all safety instructions carefully and make sure you understand them fully before attempting to
use this product. In addition, the following symbol may appear on the product:
Please read all safety instructions carefully and make sure you understand them fully before attempting to
use this product.
WARNING Using this instrument in a manner not specified by the manufacturer may
impair the protection provided by the instrument.
Cleaning Instructions
! Allow the instrument to dry completely before reconnecting it to the power source.
v
vi
1
1. DESCRIPTION
The ORTEC Model 556 High Voltage Power Supply The Model 556 features a front-panel LCD meter
is a standard double-width NIM module that which can be programmed via a front-panel toggle
provides either polarity of output voltage from 50 V switch to display either output voltage or load
to 3000 V, 0 to 10 mA. The adjusted output voltage current. The output voltage can be programmed by
of the selected polarity is available simultaneously an external control via a rear-panel BNC connector.
through two SHV rear-panel connectors. A rear-
panel slide switch permits operation on either 115 V The Model 556 provides the extremely stable, low-
or 230 V ac input power, furnished through a noise, high voltage that is required for proper bias of
removable power line cord and connector. photomultiplier tubes, ionization chambers, and
semiconductor detectors.
DANGER
THIS INSTRUMENT PRODUCES EXTREMELY HAZARDOUS VOLTAGES AT A
POTENTIALLY LETHAL CURRENT LEVEL. NEVER CONNECT OR DISCONNECT
THE HIGH VOLTAGE OUTPUT CONNECTOR WITH THE POWER SWITCH ON.
NEVER CHANGE THE OUTPUT POLARITY SWITCH WITH THE POWER SWITCH ON.
ALWAYS SWITCH POWER OFF AND WAIT AT LEAST 30 SECONDS BEFORE
CONNECTING OR DISCONNECTING CABLES AND BEFORE CHANGING THE
OUTPUT POLARITY.
2. SPECIFICATIONS
! lnt Selects the internal reference source; the DIMENSIONS Standard double-width NIM module,
front-panel controls select the output amplitude. per DOE/ER-0457T.
3. INSTALLATION
3.1. GENERAL 4. Check the polarity switch on the rear panel and
set it for either positive or negative output
The Model 556 is normally used in conjunction with
polarity as required for the application.
other modular electronics and may be installed in a
standard NIM bin such as an ORTEC Model 4001A.
5. Connect a high-voltage cable from either output
Since the bin may be rack mounted, any high
connector on the 556 to the instrument to be
temperature equipment that may be installed in the
powered. Use the other output connector if a
same rack as the bin must be sufficiently cooled by
second instrument is to be operated at the same
circulating air to prevent exceeding the +50EC
output voltage.
(120EF) maximum operating temperature of the 556.
The ORTEC M127/N NIM Fan is available for
6. Set the front-panel selector switches and
forced-air cooling of a rack of equipment.
potentiometer for the desired voltage level. This
is normally specified for the instruments to
3.2. CONNECTION TO POWER which the voltage is to be applied. The adjusted
The Model 556 requires a grounded ac-power output voltage will be the sum of the settings of
source of nominal 115 V ac or 230 V ac. A rear all three controls.
panel international-standard IEC connector allows
the connection of many different types of line cords 7. Turn on the power with the switch on the front
between the ac outlet and the 556. A rear-panel panel. The indicator lamp next to the switch will
slide switch allows the choice of 115 V ac or light to show that input power is being applied.
230 V ac input voltages. The 556 is shipped with a The indicating meter at the top of the front panel
choice of two fuse holder caps to accommodate will also indicate the polarity and amplitude of
either 3AG or 5×20 mm size fuses. On 115 V ac either the 556 output voltage or load current,
operation a 1.5 A (FAST) (250V) size 3AG fuse depending on the setting of the meter switch.
should be used; on 230 V ac operation, a 0.75A(F)
(250V) size 5×20 mm fuse should be used. 3.4. CONNECTING AN EXTERNAL
REFERENCE INPUT
This power supply may be operated entirely
removed from the 4001A bin if desired, since it is The Model 556 output voltage level can be
totally self-contained and requires no dc-operating controlled by an external reference level that is
power levels from the NIM bin. However, furnished through the rear-panel BNC connector
precautions should be taken to ensure that when the Control locking switch is set at Ext. The
personnel is aware of the shock hazard at the rear range of input voltage is 0 to 6.9 V to provide an
connectors, and that air space should be provided at output from 0 to 3000 V. The front- panel voltage
the top and bottom of the instrument. level controls are ineffective for external reference
operation.
4. CIRCUIT DESCRIPTION
Figure 2 is a simplified block diagram of the Model the input of the high frequency dc-to-dc converter.
556 circuits. The converter is made up of a free-running oscillator
(Q6-9) driving chopper MOSFETs Q10 and Q11,
The 556 requires an ac power input, regardless of which alternately switch on to transfer energy
the use of an internal or external reference level. through high-voltage transformer T2. The converter
The selected reference level is applied to a precision output consists of the rectified, filtered, and doubled
regulator and controls the low voltage input level to high voltage at C17. R53 and C18 further filter the
an internal 24-kHz oscillator. The oscillator output output voltage.
voltage is stepped up for the high voltage output
through a converter transformer, and this signal is A preregulator circuit is necessary to limit the power
rectified and filtered to produce the output to each dissipation by Q5. Therefore its collector and emitter
output connector. voltages are compared at U3(5 and 6). The resulting
output represents a request for more Q5 collector
The front panel meter (M1) is connected to the voltage.
output connectors through R65, R85, and R66 when
monitoring output voltage. R66 is used to calibrate U3(1) is synchronized with the power line frequency.
the front-panel display meter. In the load current This signal clocks JK flip-flop U4 and when U3(7) is
display mode, the output ground return current is high, SCR Q12 is fired in synchronization with the
sensed by R69. The sensed voltage is inverted by power line through Q2, Q3, and D8. When SCR Q12
S7 and fed to meter M1. fires, current flows through R3 and C1, boosting the
collector voltage at Q5.
Two feedback loops, one for preregulation and one
for output regulation, operate simultaneously. To Current limiting transistor Q4 monitors the current
maintain the output voltage at a constant voltage, through R4 and prevents damage during output
U1 compares the sampled output voltage against overload by conducting current away from the base
the reference voltage appearing at its input of Q5. Also, if a high voltage output should become
terminals. The resulting error signal is amplified by shorted to ground, transistors Q14 and Q15 will
U2, Q1, and Q5. The output (emitter) of Q5 feeds conduct current away from the base of Q5. Current
pulses in the ground return path during a short the switch is in the NEG position. Regulators U5 and
circuit condition produce voltage pulses across R69 U6 provide the necessary internal ±12 V power to
which are of sufficient amplitude to turn either Q14 operate the control circuitry, the reference circuit,
or Q15 On, depending upon the setting of the and the oscillator.
POLARITY switch (R.P.). Q15 will become active if
the switch is set for POS and Q14 will be active if
[Intentionally blank]