1- Possible causes for drying problems
1- Possible causes for drying problems
d) Device faults
Device fault Corrective
Vacuum pump inefficient or leaky Check the maximum evacuation times on the sterilization
protocol:
®
Vacuklav 23-B/ 23 B+ c 3:40 min
®
Vacuklav 31 und 31-B/31 B+ c 3 min
®
Vacuklav 24-B/24 B+ c 2:30 min
®
Vacuklav 24 /30-B/ 30 B+ c 2 min
®
Euroklav 23V-S/ 23 VS+ c 3:40 min
®
Euroklav 29V-S/ 29 VS+ c 3 min
possible pump replacement
Pressure release inlet blocked or disabled Small parts in the condensate return hose;
®
Upgrade chamber steam generator (all Vacuklav ) or
®
chamber T piece on the pressure release filter (Euroklav );
pressure release filter choked;
Device internal hoses kinked;
®
Effluent hose kinked, too long, water skin etc. (Vacuklav 24,
24-B, 24-B/L,30-B, 24 B+, 24 B/L+, 30 B+)
Page 1 of 2
Possible causes for drying problems of autoclaves
Vacuklav, Euroklav, Pro-Class, S-Class
Euroklav®23V-S/ 29V-S, S-Class up to SW 3.34 NOT FOR Euroklav® 23-S and 29-S
Service Menu Abb. Old value New value Description
Pressure P 18 500 350 Control pressure drying
Temperature T 10 145 148 Pre-heating overheat.
T 11 118 120 Control. PH Univ-prog.
T 12 118 120 Control. PH Quick-prog.
T 14 118 120 Control. PH Spec.-prog.
T 16 100 120 Control. PH basic position
T 17 138 143 Control. PH drying Univ.-prog.
T 18 138 143 Control. PH drying Quick-prog.
T 20 138 143 Control. PH drying Spec.-prog
Frequency a 11 13 10 Number drying Univ.-prog.
a 14 13 10 Number drying Spec.-prog.
a 16 5 7 Number Under.press. Univ.-prog.
a 19 5 7 Number Under.press. drying. Spec.-prog.