Pv-Isotest, Pv-255s, Pvchecks-pro Upgrading Procedure
Pv-Isotest, Pv-255s, Pvchecks-pro Upgrading Procedure
Tipo documento:
IOP
IOP_16220FWUpgradingProcedure_1-01.doc
Object: 16220
Oggetto:
This document is intended for internal and/or confidential use. Its copying or distribution must be
preliminarly authorized by HT-Italia. It is NOT permitted to distribuite it to any customer or its using
as market information.
Page 2 of 3
IOP: - 16220 - Firmware upgrading procedure
Scopo (Aim)
La presente IOP descrive come aggiornare il firmware degli apparati serie 16220 (PV-
ISOTEST, PV-255s)
(this document describes how to upgrade the firmware of the 16220 series (PV-ISOTEST, PV-255s)
IOP_16220FWUpgradingProcedure_1-01.doc
This document is intended for internal and/or confidential use. Its copying or distribution must be
preliminarly authorized by HT-Italia. It is NOT permitted to distribuite it to any customer or its using
as market information.
Page 3 of 3
IOP: - 16220 - Firmware upgrading procedure
2.3. Lanciare il software “Upload162x0.exe” e, usando il comando Serial, selezionare
il modello dello strumento, la COMMxx corrispondente alla voce “USB Serial Port”
associata al cavo C2006
(Launch the “Upload162x0.exe” program, set the Instrument Type and select the COMMxx
correspondent to the “USB Serial Port” item associated to C2006 cable by using the Serial
command)
2.4. Impostare “115200” come velocità di comunicazione
(set 115200 as baud rate)
2.5. Nella cartella Upload → Firmware del software selezionare l’opzione “All
firmware by Opto” e col tasto Browse selezionare il file “16220_vY-XX.bin” dal
PC dove “Y-xx = versione del FW”.
(In the Upload → Firmware tab of the software select the “All firmware by Opto” option and with
the Browse key select the “16220_vY-xx.bin” from the PC where “Y-xx = version of FW”)
2.6. Cliccare il tasto Prog per attivare l’operazione e attendere il completamento della
stessa
Click the Prog key to start the operation and wait until the end of it
2.7. Al termine dell’aggiornamento, premere il tasto ON/OFF per riaccendere lo
strumento e verificare la versione fw.
(At the end of the operation press the ON/OFF key to switch on the device)
2.8. Se la procedura di aggiornamento dovesse fallire, riprovare ad eseguire tutti i
passi impostando però su strumento e software di aggiornamento la velocità
57600.
(If the process fails, re-attempt again the complete procedure by setting 57600 as baud rate)
IOP_16220FWUpgradingProcedure_1-01.doc
This document is intended for internal and/or confidential use. Its copying or distribution must be
preliminarly authorized by HT-Italia. It is NOT permitted to distribuite it to any customer or its using
as market information.