FIP flow sensor
FIP flow sensor
00
PADDLEWHEEL
FLOW SENSOR
SAFETY INSTRUCTIONS
General Statements
• Do not install and service the product without following the Instruction
Manual.
• This item is designed to be connected to other instruments which can be
hazardous if used improperly. Read and follow all associated instrument
manuals before using with it.
• Product installation and wiring connections should only be performed by
qualified staff.
• Do not modify product construction.
Installation and Commissioning Statements
• Remove power to the instrument before wiring input and output connections.
• Depressurize and vent the system before installing or removing the sensor.
• Check and confirm the chemical compatibility of the materials in contact
with the liquid.
• Do not exceed maximum specifications using the instrument.
• To clean the unit, use only chemical compatible products.
1
PACKING LIST
Please verify that the product is complete and without any damage.
The following items must be included:
• F3.00 Paddlewheel Flow Sensor
• Instruction Manual for F3.00 Paddlewheel Flow Sensor
DESCRIPTION
The simple and reliable paddlewheel flow sensor type F3.00 is designed for
use with every kind of solid-free liquids. The sensor can measure flow from
0.15 m/s (0.5 ft/s) producing a frequency output signal highly repeatable.
A rugged construction and a proven technology guarantee exceptional
performances with little or no maintenance required. A dedicated electronic,
with a push-pull output, is available for a safe connection to any
kind of PLC/Instrument digital input. A specially designed family of fittings
ensures an easy and quick installation into all pipe materials in sizes from
DN15 to DN600 (0.5” to 24”).
TECHNICAL DATA
General
• Pipe Size Range: DN15 to DN600 (0.5” to 24”) Please refer to Installation
Fittings section for more details
• Flow Rate Range: 0.15 to 8 m/s (0.5 to 25 ft./s)
• Linearity: ± 0.75 % of full scale
• Repeatability: ± 0.5 % of full scale
• Minimum Reynolds Number Required: 4500
• Enclosure: IP68 or IP65
• Wetted Materials:
- sensor Body: CPVC, PVDF, 316L SS
- o-rings: EPDM or FPM
- rotor: ECTFE (Halar®)
- shaft: Ceramic (Al2O3)/316L SS (only for metal sensors)
- bearings: Ceramic (Al2O3)
Specific for F3.00.H
• Supply voltage: 5 to 24 VDC ± 10% regulated
• Supply current: < 30 mA @ 24 VDC
• Output signal:
- square wave
- frequency: 45 Hz per m/s nominal
(13.7 Hz per ft/s nominal)
- type: transistor NPN open collector
- output current: 10 mA max
• Cable length: 8 m (26.4 ft) standard, 300 m (990 ft)maximum
2
Specific for F3.00.C
• Supply voltage: 3 to 5 VDC regulated or
3.6 Volt Lithium battery
• Supply current: < 10 µA max
• Output signal:
- square wave
- frequency: 45 Hz per m/s nominal
(13.7 Hz per ft/s nominal)
- min. input impedance: 100 KΩ
• Cable length: 8 m (26.4 ft) standard, 16 m (52.8 ft) maximum
Specific for F3.00.P
• Supply voltage: 12 to 24 VDC ± 10% regulated
• Supply current: < 30 mA @ 24 VDC
• Output signal:
- square wave
- frequency: 45 Hz per m/s nominal
(13.7 Hz per ft/s nominal)
- type: Push-Pull (for connection to NPN and PNP inputs)
- output current: 20 mA max
• Cable length: 8 m (26.4 ft) standard, 300 m (990 ft) maximum
Standards & Approvals
• Manufactured under ISO 9001
• Manufactured under ISO 14001
• CE
• RoHS Compliant
• GOST R
3
Maximum Operating Pressure / Temperature (25 years lifetime)
F3.00.H or F3.00.P Sensor F3.00.C Sensor
• CPVC body: • CPVC body:
- 10 bar (145 psi) @ 25°C (77°F) - 10 bar (145 psi) @ 25°C (77°F)
- 1,5 bar (22 psi) @ 80° C (176°F) - 1,5 bar (22 psi) @ 80° C (176°F)
• PVDF body: • PVDF body:
- 10 bar (145 psi) @ 25°C (77°F) - 10 bar (145 psi) @ 25°C (77°F)
- 2,5 bar (36 psi) @ 100°C (212°F) - 2,5 bar (36 psi) @ 100°C (212°F)
• SS body: • SS body:
- 25 bar (363 psi) @ 120°C (248°F) - 25 bar (363 psi) @ 100°C (212°F)
4
DIMENSIONS
5
8 Ceramic bearings
INSTALLATION
Pipe Location
• The six most common installation configurations shown in fig. 1 help in
selecting the best location in the pipeline for paddlewheel flow sensor as well
for magmeter flow sensor.
• The three configurations in fig. 2 ensure that the pipe is always full: for a
correct measurement the sensor can NOT be exposed to air bubbles at any
time.
• The three installations in Fig. 3 should be avoided unless you are absolutely
sure the sensor is not exposed to air bubbles.
• In gravity-flow systems the connection to the tank must be designed so the
level does not drop below the outlet: this to avoid pipe to draw air in from the
tank causing a inaccurate measurement of sensor (see Fig. 4).
• For more information, please refer to EN ISO 5167-1.
• Always maximize distance between flow sensors and pumps.
Fig.1
6
Fig.2
Fig.3
Fig.4
Mounting position
Measuring part of sensor (rotor for paddlewheel and pins for magmeter)
should be positioned at 12% of ID where, basing on insertion theory, average
velocity can be measured.
The reading accuracy of insertion flow sensors can be affected by:
• air bubbles;
• sediments;
• friction between shaft and bearings (only for paddlewheel).
In a horizontal pipe runs, the mounting position to get the best performances is
at a 45° angle (Fig. 3) to avoid air bubbles as well sediments. Vertical position
(Fig. 2) can be chosen in case air bubbles are not present. Do not mount the
sensor on the bottom of the pipe (Fig. 1) if sediments are likely. Do not mount
paddlewheel at 90° otherwise friction can affect measurement.
Installation in a vertical pipe runs can be done fixing any orientation.
Upward flow is preferred to ensure full pipe.
7
Process connection
1. Lubricate the sensor O-rings
with a silicone lubricant. Do
not use any petroleum based
lubricant that may damage the
O-rings.
2. Lower the sensor into the fitting
making sure the alignment tab is
seated in the fitting notch.
3. Hand tighten the sensor cap.
Do not use any tool otherwise
cap and/or fitting threads may be
damaged.
WIRING
General recommendation
• Always ensure the power supply is switched off before working on the sensor.
• Always use a high quality (regulated) DC voltage supply.
WIRING DIAGRAMS
M9.07
M9.20
M9.02 M9.50 M9.08 M9.03 (FLOW SENSOR 1) M9.10
M9.00
M9.03 (FLOW SENSOR 2)
GND 5 30 16 30 7 37
FREQ. IN 6 28 14 28 8 36
V+ 7 27 13 27 9 35
DIR 6B - 15 29 - -
8
F3.00.H IP65 Sensor Connection to FLS Instruments (except M9.20)
9
F3.00.P IP68 Sensor Connection to PLC with NPN Input
10
F3.00.P IP65 Sensor Connection to PLC with PNP Input
11
F3.00.C IP68 Sensor Connection to Other Brand Instruments
12
INSTALLATION FITTINGS
Type Description
13
K-FACTOR TABLES
K-Factor is the number of pulses which a sensor produces for one liter of
measured fluid. Here below all K-Factors for water at room temperature are
listed.
K-Factor values can depend on the installation conditions. K-Factor has to
divide the frequency generated by F3.00 in order to achieve the flow rate (l/s).
Please contact your dealer for K-Factor values not included in the table.
BSP Female Threaded PVC Tee Fittings BS Solvent Welding PVC Tee Fittings
for BS PN12 pipes for BS PN12 pipes
(parallel threaded female ends) (female ends for solvent welding)
Part No. DN R K-Factor Part No. DN d K-Factor
TFFV20B 15 1/2" 235,45 TFLV20B 15 1/2" 235,45
TFFV25B 20 3/4" 142,46 TFLV25B 20 3/4" 142,46
TFFV32B 25 1" 91,53 TFLV32B 25 1" 91,53
TFFV40B 32 1" 1/4 51,57 TFLV40B 32 1" 1/4 51,57
TFFV50B 40 1" 1/2 42,89 TFLV50B 40 1" 1/2 42,89
TFFV20D 15 1/2" 235,45 TFLV20D 15 1/2" 235,45
TFFV25D 20 3/4" 142,46 TFLV25D 20 3/4" 142,46
TFFV32D 25 1" 91,53 TFLV32D 25 1" 91,53
TFFV40D 32 1" 1/4 51,57 TFLV40D 32 1" 1/4 51,57
TFFV50D 40 1" 1/2 42,89 TFLV50D 40 1" 1/2 42,89
14
NPT Female Threaded PVC Tee Fittings ASTM SCH. 80 PVC Tee Fittings
for ASTM SCH. 80 pipes for ASTM SCH. 80 pipes
(NPT threaded female ends) (female ends for solvent welding)
Part No. SIZE R K-Factor Part No. SIZE d K-Factor
TFNV20B 0.50" 1/2" 235,45 TFAV20B 0.50" 1/2" 235,45
TFNV25B 0.75" 3/4" 142,46 TFAV25B 0.75" 3/4" 142,46
TFNV32B 1.00" 1" 91,53 TFAV32B 1.00" 1" 91,53
TFNV40B 1.25" 1" 1/4 51,57 TFAV40B 1.25" 1" 1/4 51,57
TFNV50B 1.50" 1" 1/2 42,89 TFAV50B 1.50" 1" 1/2 42,89
TFNV20D 0.50" 1/2" 235,45 TFAV20D 0.50" 1/2" 235,45
TFNV25D 0.75" 3/4" 142,46 TFAV25D 0.75" 3/4" 142,46
TFNV32D 1.00" 1" 91,53 TFAV32D 1.00" 1" 91,53
TFNV40D 1.25" 1" 1/4 51,57 TFAV40D 1.25" 1" 1/4 51,57
TFNV50D 1.50" 1" 1/2 42,89 TFAV50D 1.50" 1" 1/2 42,89
15
Installation on C-PVC pipes
ISO metric PVDF Tee Fittings for ISO SDR 21
ISO Clamp Saddles for ISO SDR 21 pipes
C-PVC pipes (female ends for solvent welding)
Part No. DN d K-Factor Part No. DN d K-Factor
TFIF20BC 15 20 225,06 SVIC063BVC 50 63 21,69
TFIF25BC 20 25 139,38 SVIC075BVC 65 75 14,98
TFIF32BC 25 32 94,66 SVIC090BVC 80 90 9,88
TFIF40BC 32 40 51,37 SVIC110BVC 100 110 6,06
TFIF50BC 40 50 43,07 SVIC125BVC 110 125 4,59
TFIF20DC 15 20 225,06 SVIC140BVC 125 140 3,59
TFIF25DC 20 25 139,38 SVIC160BVC 150 160 2,69
TFIF32DC 25 32 94,66 SVIC200BVC 180 200 1,65
TFIF40DC 32 40 51,37 SVIC225BVC 200 225 1,28
TFIF50DC 40 50 43,07 SVIC063DVC 50 63 21,69
SVIC075DVC 65 75 14,98
SVIC090DVC 80 90 9,88
SVIC110DVC 100 110 6,06
SVIC125DVC 110 125 4,59
SVIC140DVC 125 140 3,59
SVIC160DVC 150 160 2,69
SVIC200DVC 180 200 1,65
SVIC225DVC 200 225 1,28
SMIC250IVC 225 250 1,01
SMIC280IVC 250 280 0,79
SMIC315IVC 280 315 0,61
Installation on PP pipes
ISO Metric PP Tee Fittings for ISO SDR 11 pipes BSP Female Threaded PP Tee Fittings
(female ends for socket welding) for BS pipes (parallel threaded female ends)
Part No. DN d K-Factor Part No. DN R K-Factor
TFIM20B 15 20 212,17 TFFM20B 15 1/2" 212,17
TFIM25B 20 25 135,32 TFFM25B 20 3/4" 135,32
TFIM32B 25 32 89,36 TFFM32B 25 1" 89,36
TFIM40B 32 40 48,94 TFFM40B 32 1" 1/4 48,94
TFIM50B 40 50 42,10 TFFM50B 40 1" 1/2 42,10
TFIM20D 15 20 212,17 TFFM20D 15 1/2" 212,17
TFIM25D 20 25 135,32 TFFM25D 20 3/4" 135,32
TFIM32D 25 32 89,36 TFFM32D 25 1" 89,36
TFIM40D 32 40 48,94 TFFM40D 32 1" 1/4 48,94
TFIM50D 40 50 42,10 TFFM50D 40 1" 1/2 42,10
16
NPT Female Threaded PP Tee Fittings
ISO Clamp Saddles for ISO SDR 21 pipes for ASTM SCH.80 pipes
(NPT threaded female ends)
Part No. DN d K-Factor Part No. DN d K-Factor
SVIC063BME 50 63 27,50 TFNM20B 0.50" 1/2" 212,17
SVIC075BME 65 75 18,56 TFNM25B 0.75" 3/4" 135,32
SVIC090BME 80 90 12,44 TFNM32B 1.00" 1" 89,36
SVIC110BME 100 110 7,59 TFNM40B 1.25" 1" 1/4 48,94
SVIC125BME 110 125 5,77 TFNM50B 1.50" 1" 1/2 42,10
SVIC140BME 125 140 4,49 TFNM20D 0.50" 1/2" 212,17
SVIC160BME 150 160 3,38 TFNM25D 0.75" 3/4" 135,32
SVIC200BME 180 200 2,07 TFNM32D 1.00" 1" 89,36
SVIC225BME 200 225 1,60 TFNM40D 1.25" 1" 1/4 48,94
SVIC063DME 50 63 27,50 TFNM50D 1.50" 1" 1/2 42,10
SVIC075DME 65 75 18,56
SVIC090DME 80 90 12,44
SVIC110DME 100 110 7,59
SVIC125DME 110 125 5,77
SVIC140DME 125 140 4,49
SVIC160DME 150 160 3,38
SVIC200DME 180 200 2,07
SVIC225DME 200 225 1,60
SMIC250IME 225 250 1,27
SMIC280IME 250 280 0,99
SMIC315IME 280 315 0,77
17
Installation on PVDF pipes
ISO Metric PVDF Tee Fittings for ISO SDR 33
ISO Clamp Saddles for ISO SDR 33 pipes
pipes (female ends for socket welding)
Part No. DN d K-Factor Part No. DN d K-Factor
TFIF20B 15 20 225,06 SVIF063BF 50 63 20,58
TFIF25B 20 25 139,38 SVIF 075BF 65 75 14,09
TFIF32B 25 32 94,66 SVIF 090BF 80 90 9,29
TFIF40B 32 40 51,37 SVIF 110BF 100 110 5,69
TFIF50B 40 50 43,07 SVIF 125BF 110 125 4,31
TFIF20D 15 20 225,06 SVIF 140BF 125 140 3,36
TFIF25D 20 25 139,38 SVIF 160BF 150 160 2,52
TFIF32D 25 32 94,66 SVIF 200BF 180 200 1,55
TFIF40D 32 40 51,37 SVIF 225BF 200 225 1,20
TFIF50D 40 50 43,07 SVIF 063DF 50 63 20,58
SVIF 075DF 65 75 14,09
SVIF 090DF 80 90 9,29
SVIF 110DF 100 110 5,69
SVIF 125DF 110 125 4,31
SVIF 140DF 125 140 3,36
SVIF 160DF 150 160 2,52
SVIF 200DF 180 200 1,55
SVIF 225DF 200 225 1,20
Installation on PE pipes
ISO Metric PVC Tee Fittings
for PE SDR 11 pipes (PE end connectors ISO Clamp Saddles for PE SDR 11 pipes
for electrofusion or butt welding) Part No. DN d K-Factor
Part No. DN d K-Factor SVIC063BME 50 63 27,39
TFIV20BE 15 20 193,70 SVIC075BME 65 75 18,75
TFIV25BE 20 25 134,07 SVIC090BME 80 90 12,41
TFIV32BE 25 32 85,29 SVIC110BME 100 110 7,57
TFIV40BE 32 40 48,68 SVIC125BME 110 125 5,76
TFIV50BE 40 50 41,68 SVIC140BME 125 140 4,49
TFIV20DE 15 20 193,70 SVIC160BME 150 160 3,37
TFIV25DE 20 25 134,07 SVIC200BME 180 200 2,02
TFIV32DE 25 32 85,29 SVIC225BME 200 225 1,60
TFIV40DE 32 40 48,68 SVIC063DME 50 63 27,39
TFIV50DE 40 50 41,68 SVIC075DME 65 75 18,75
SVIC090DME 80 90 12,41
SCIC110DME 100 110 7,57
SVIC125DME 110 125 5,76
SVIC140DME 125 140 4,49
SVIC160DME 150 160 3,37
SVIC200DME 180 200 2,02
SVIC225DME 200 225 1,60
SMIC250IME 225 250 1,27
SMIC280IME 250 280 0,99
SMIC315IME 280 315 0,77
18
Installation on Metal pipes
316L SS Threaded Tees
(BSP Female Threads)
Part No. DN R K-Factor
TFFX20 15 1/2" -
TFFX25 20 3/4 157,06
TFFX32 25 1" 92,84
TFFX40 32 1" 1/4 51,52
19
Correction formula for K-Factor calculation according to real internal
diameter:
Example:
Nominal Pipe Size (DN) = 100 mm
New Internal Diameter = 104 mm
Forumla: K-Factor_NEW = (51.02 x 100²) / 104² = 20,52
ORDERING DATA
F3.00.H.XX Paddlewheel Flow Sensors (Remote version)
Main
Power Flow Rate Weight
Part No. Version Length wetted Enclosure
supply Range (gr.)
materials
0.15 to 8 m/s
F3.00.H.01 Hall 5 - 24 VDC L0 CPVC/ EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.02 Hall 5 - 24 VDC L0 CPVC/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.03 Hall 5 - 24 VDC L1 CPVC/ EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.04 Hall 5 - 24 VDC L1 CPVC/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.05 Hall 5 - 24 VDC L0 PVDF/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.06 Hall 5 - 24 VDC L0 PVDF/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.07 Hall 5 - 24 VDC L1 PVDF/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.08 Hall 5 - 24 VDC L1 PVDF/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.09 Hall 5 - 24 VDC L0 316SS/EPDM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.10 Hall 5 - 24 VDC L0 316SS/FPM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.11 Hall 5 - 24 VDC L1 316SS/EPDM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.12 Hall 5 - 24 VDC L1 316SS/FPM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.13 Hall 5 - 24 VDC L0 CPVC/EPDM IP65 250
(0.5 to 25 ft./s.)
20
0.15 to 8 m/s
F3.00.H.14 Hall 5 - 24 VDC L0 CPVC/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.15 Hall 5 - 24 VDC L1 CPVC/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.16 Hall 5 - 24 VDC L1 CPVC/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.17 Hall 5 - 24 VDC L0 PVDF/EPDM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.18 Hall 5 - 24 VDC L0 PVDF/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.19 Hall 5 - 24 VDC L1 PVDF/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.20 Hall 5 - 24 VDC L1 PVDF/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.21 Hall 5 - 24 VDC L0 316SS/EPDM IP65 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.22 Hall 5 - 24 VDC L0 316SS/FPM IP65 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.23 Hall 5 - 24 VDC L1 316SS/EPDM IP65 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.H.24 Hall 5 - 24 VDC L1 316SS/FPM IP65
(0.5 to 25 ft./s.)
21
F3.00.C.XX Paddlewheel Flow Sensors (Remote version)
Main
Power Flow Rate Weight
Part No. Version Length wetted Enclosure
supply Range (gr.)
materials
0.15 to 8 m/s
F3.00.C.01 Coil 3 - 5 VDC L0 CPVC/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.02 Coil 3 - 5 VDC L0 CPVC/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.03 Coil 3 - 5 VDC L1 CPVC/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.04 Coil 3 - 5 VDC L1 CPVC/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.05 Coil 3 - 5 VDC L0 PVDF/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.06 Coil 3 - 5 VDC L0 PVDF/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.07 Coil 3 - 5 VDC L1 PVDF/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.08 Coil 3 - 5 VDC L1 PVDF/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.09 Coil 3 - 5 VDC L0 316SS/EPDM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.10 Coil 3 - 5 VDC L0 316SS/FPM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.11 Coil 3 - 5 VDC L1 316SS/EPDM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.12 Coil 3 - 5 VDC L1 316SS/FPM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.13 Coil 3 - 5 VDC L0 CPVC/EPDM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.14 Coil 3 - 5 VDC L0 CPVC/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.15 Coil 3 - 5 VDC L1 CPVC/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.16 Coil 3 - 5 VDC L1 CPVC/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.17 Coil 3 - 5 VDC L0 PVDF/EPDM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.18 Coil 3 - 5 VDC L0 PVDF/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.19 Coil 3 - 5 VDC L1 PVDF/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.20 Coil 3 - 5 VDC L1 PVDF/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.21 Coil 3 - 5 VDC L0 316SS/EPDM IP65 600
(0.5 to 25 ft./s.)
22
0.15 to 8 m/s
F3.00.C.22 Coil 3 - 5 VDC L0 316SS/FPM IP65 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.23 Coil 3 - 5 VDC L1 316SS/EPDM IP65 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.C.24 Coil 3 - 5 VDC L1 316SS/FPM IP65 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.02 Push-Pull 12 - 24 VDC L0 CPVC/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.03 Push-Pull 12 - 24 VDC L1 CPVC/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.04 Push-Pull 12 - 24 VDC L1 CPVC/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.05 Push-Pull 12 - 24 VD C L0 PVDF/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.06 Push-Pull 12 - 24 VDC L0 PVDF/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.07 Push-Pull 12 - 24 VDC L1 PVDF/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.08 Push-Pull 12 - 24 VDC L1 PVDF/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.09 Push-Pull 12 - 24 VDC L0 316SS/EPDM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.10 Push-Pull 12 - 24 VDC L0 316SS/FPM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.11 Push-Pull 12 - 24 VDC L1 316SS/EPDM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.12 Push-Pull 12 - 24 VDC L1 316SS/FPM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.13 Push-Pull 12 - 24 VDC L0 CPVC/EPDM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.14 Push-Pull 12 - 24 VDC L0 CPVC/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.15 Push-Pull 12 - 24 VDC L1 CPVC/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.16 Push-Pull 12 - 24 VDC L1 CPVC/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.17 Push-Pull 12 - 24 VDC L0 PVDF/EPDM IP65 250
(0.5 to 25 ft./s.)
23
0.15 to 8 m/s
F3.00.P.18 Push-Pull 12 - 24 VDC L0 PVDF/FPM IP65 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.19 Push-Pull 12 - 24 VDC L1 PVDF/EPDM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.20 Push-Pull 12 - 24 VDC L1 PVDF/FPM IP65 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.21 Push-Pull 12 - 24 VDC L0 316SS/EPDM IP65 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.22 Push-Pull 12 - 24 VDC L0 316SS/FPM IP65 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.23 Push-Pull 12 - 24 VDC L1 316SS/EPDM IP65 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.00.P.24 Push-Pull 12 - 24 VDC L1 316SS/FPM IP65 650
(0.5 to 25 ft./s.)
24
F3.01.X.XX Paddlewheel Flow Sensors (Compact version)
Power Main wetted Flow Rate Weight
Part No. Version Length Enclosure
supply materials Range (gr.)
0.15 to 8 m/s
F3.01.H.01 Hall 5 - 24 VDC L0 CPVC/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.02 Hall 5 - 24 VDC L0 CPVC/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.03 Hall 5 - 24 VDC L1 CPVC/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.04 Hall 5 - 24 VDC L1 CPVC/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.05 Hall 5 - 24 VDC L0 PVDF/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.06 Hall 5 - 24 VDC L0 PVDF/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.07 Hall 5 - 24 VDC L1 PVDF/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.08 Hall 5 - 24 VDC L1 PVDF/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.09 Hall 5 - 24 VDC L0 316SS/EPDM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.10 Hall 5 - 24 VDC L0 316SS/FPM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.11 Hall 5 - 24 VDC L1 316SS/EPDM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.H.12 Hall 5 - 24 VDC L1 316SS/FPM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.01 Coil 3 - 5 VDC L0 CPVC/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.02 Coil 3 - 5 VDC L0 CPVC/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.03 Coil 3 - 5 VDC L1 CPVC/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.04 Coil 3 - 5 VDC L1 CPVC/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.05 Coil 3 - 5 VDC L0 PVDF/EPDM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.06 Coil 3 - 5 VDC L0 PVDF/FPM IP68 250
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.07 Coil 3 - 5 VDC L1 PVDF/EPDM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.08 Coil 3 - 5 VDC L1 PVDF/FPM IP68 300
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.09 Coil 3 - 5 VDC L0 316SS/EPDM IP68 600
(0.5 to 25 ft./s.)
25
0.15 to 8 m/s
F3.01.C.10 Coil 3 - 5 VDC L0 316SS/FPM IP68 600
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.11 Coil 3 - 5 VDC L1 316SS/EPDM IP68 650
(0.5 to 25 ft./s.)
0.15 to 8 m/s
F3.01.C.12 Coil 3 - 5 VDC L1 316SS/FPM IP68 650
(0.5 to 25 ft./s.)
26
SPARE PARTS
Weight
Part No. Name Description
(gr.)
F3.SP2.6 Sensor Cap SS AISI 316 Sensor Cap, for SS Hall and Coil versions 205
F3.SP4.2 Rotor KIT ECTFE (Halar®) rotor with Machined Ceramic Shaft and Bearings 8
27
1 4-pole Cable Plug
2 Sensor Cap
3 O-Rings
4 Rotor KIT
28
NOTE
29
NOTE
30
NOTE
31
I2366 - IMF300E - REV-02 - 06/2016