Urea Archem Validity aV-280 Application Parameters Ver 1a

The document outlines the application settings for the Urea Kinetic test using the Validity aV-280 analyzer, detailing sample volumes, reagent volumes, calibration types, and affected tests. It specifies a 2-point linear calibration method with specific concentration points and emphasizes the importance of using high-quality materials and following laboratory validation plans. Additionally, it warns against the contamination effects of certain tests on Urea Kinetic results and vice versa.

Uploaded by

ashrafmogy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
11 views1 page

Urea Archem Validity aV-280 Application Parameters Ver 1a

The document outlines the application settings for the Urea Kinetic test using the Validity aV-280 analyzer, detailing sample volumes, reagent volumes, calibration types, and affected tests. It specifies a 2-point linear calibration method with specific concentration points and emphasizes the importance of using high-quality materials and following laboratory validation plans. Additionally, it warns against the contamination effects of certain tests on Urea Kinetic results and vice versa.

Uploaded by

ashrafmogy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Application Settings: Urea Kinetic

Ver 1a

Analzyer Name and Model: Validity aV-280


Reagent Settings
Sample Volume
Test name Urea Kinetic Sample Vol. 2 Samp. For Dilut. Dilut. Vol.
Automatic Dilution 0 0
Pri. Sec.
Wavelength 340 700 Reagent Volumes
R1 R2
Method Start End Volume 160 37
Assay Point Fixed Time 19 23 Barcode

Decimal Place 1 Auto Check Up Rerun


Samp. For Dilut. Dilut. Vol.
Unit mg/dL Substrate
Low High Depleted % 0 0
Linearity Range 3 300 Linearity % 0 0
Correlation A 1 B 0 Out of Line-
(Y=AX+B) rity Range 0 0

CALIBRATION

Calibration Type: 2 Point Linear Points 2


K Value Blank Type Water

Point 0(Blank) 1(Std) 2(Std) 3(Std) 4(Std) 5(Std) 6(Std) 7(Std)


Concentration 0 10
Sample Volume 2 2
Sample Vol. For Dilut.
Diluent Volume

Affected Tests (These tests R1-R2 Bilirubin Total


will have significant R1-R2 HDL Cholesterol
contamination effect R1
on Urea Kinetic )

Effected Tests (These tests R1


will have significant
contamination effect
from Urea Kinetic )

Note: The given settings should be only as guideline. For best performance use high quality Archem Calibr. and
QC materials, GLP rules and laboratuary VALIDATION MASTER PLAN. Archem doesn't accept misusage of tests
preparing/running procedure and wrong settings parameters.