SpectrAA Release Notes
SpectrAA Release Notes
Agilent SpectrAA Base, Pro and CFR Version 5.6 can also be used with older IEEE controlled SpectrAA instrument models, including the 50,
110/220, 600/640, 800/880 and 220/600/640/800/880 Zeeman spectrometers. However, these older instruments have not been
evaluated with the SpectrAA Version 5.6 software. Error free operation is therefore not guaranteed with these instrument models.
Agilent SpectrAA Base, Pro and CFR Version 5.6 software has been evaluated using instrument firmware version 2.12. If you are upgrading
from SpectrAA version 5.1 or earlier, it will be necessary to upgrade your firmware. Contact your local Agilent office for more details.
Registration requirements when upgrading can be found in the ‘Upgrade registration required’ column.
Upgrade
FW Release SDA Database
Name SW Version registration OS version SCM version Instrument supported The following new features were added:
Version date version version
required
SpectrAA 5.6.8 2.12 May-22 N Microsoft 2.4.1.1322 2.3.0.36 Microsoft Agilent 55B AA - Added support for Microsoft Server 2019 and Microsoft SQL
Base and Pro Windows 10 SQL Server Agilent 240 AA Express 2016 SP3. (SPAA-320)
v5.6 Professional or 2016 SP3 Agilent 240/280 FS AA - Updated software to remove Adobe Flash requirements.
SpectrAA Enterprise Agilent 240/280 Zeeman AA (SPAA-319)
CFR v5.6 Agilent SPS 4 autosampler - The Sampling Probe Height no longer resets to 150 mm after
Microsoft Server performing an instrument zero for a method from the Optimize
2019 window. (SPAA-376)
- Updated to support ADLINK USB GPIB adapter version
Microsoft Server 21.10.0. (SPAA-352)
2016 - Added the option to include or remove the Element option
when creating a report. (SPAA-310)
- Exporting results during a run to PRN and Fixed Width
outputs now works as intended. (SPAA-294)
- Lamp part number 5610107090 is now recognized by
SpectrAA. (SPAA-292)
- Sample weights, volumes and dilution in labels page are no
longer editable by using the "Auto copy" function for finished
samples. (SPAA-291)
- The edit sample labels, weights, volumes, and dilution actions
are now associated with the "Create New Worksheet" privilege.
(SPAA-290)
- The SpectrAA software no longer crashes when reducing the
sample numbers in the Develop tab. (SPAA-327)
- Renaming worksheets with invalid characters now produces
the correct error message. (SPA-329)
- X-axis values now display correctly when printing a PDF of a
signal graph from a worksheet. (SPAA-324)
-The 0.1 nm slit width lamp option is now available in the Fast
Sequential Wizard on the Optical tab. (SPAA-323)
- CCV solution is now correctly measured and reported during
the continuous calibration function. (SPAA-335)
Agilent SpectrAA Software Release Notes
Upgrade
FW Release SDA Database
Name SW Version registration OS version SCM version Instrument supported The following new features were added:
Version date version version
required
- The Audit trail now displays text indicating that changes were
made to the sample labels/weights/volumes/dilution factors
without having to save the method. (SPAA-363)
- Invalid error messages no longer appear after every sample
read when running SpectrAA on Chinese operating systems.
(SPAA-337)
- SpectrAA software Chinese (Simplified) updated to resolve
mistranslations and truncations. (SPAA-303, SPAA-62, SPAA-
399, SPAA-410, SPAA-411, SPAA-420, SPAA-421)
- SpectrAA software Chinese (Traditional) updated to resolve
mistranslations and misaligned columns. (SPAA-52)
- SpectrAA Help updated. (SPAA-340, SPAA-326, SPAA-341,
SPAA-365, SPAA-384, SPAA-200)
- SpectrAA software French mis-translations resolved. (SPAA-
343)
SpectrAA 5.5.13 2.12 Feb-20 N Microsoft 2.2.1.1276 2.2.5.35 Microsoft Agilent 55B AA - Updated to support ADLINK USB GPIB adapter version
Base v5.5 Windows 10 Server 2014 Agilent 240 AA 20.01.1. (SPAA-297)
SpectrAA 5.5 Professional or R2 64-bit Agilent 240/280 FS AA
CFR v5.5 Enterprise 64-bit Agilent 240/280 Zeeman AA
Agilent SPS 4 autosampler
SpectrAA 5.5.13 2.12 Aug-19 N Microsoft 2.2.1.1276 2.2.5.35 Microsoft Agilent 55B AA - Updated the Validate application.
Base v5.5 Windows 10 Server 2014 Agilent 240 AA
SpectrAA 5.5 Professional or R2 64-bit Agilent 240/280 FS AA
CFR v5.5 Enterprise 64-bit Agilent 240/280 Zeeman AA
Agilent SPS 4 autosampler
SpectrAA 5.5.13 2.12 Nov-18 Y Microsoft 2.2.1.1276 2.2.5.35 Microsoft Agilent 55B AA - Added support for ADLINK USB GPIB adapter. (SPAA-115)
Base and Pro Windows 10 Server 2014 Agilent 240 AA - Added audit trail to reports. (SPAA-110)
v5.5 Professional or R2 64-bit Agilent 240/280 FS AA - Added signal graphics after every sample in reports. (SPAA-
SpectrAA 5.5 Enterprise 64-bit Agilent 240/280 Zeeman AA 111)
CFR v5.5 Agilent SPS 4 autosampler - Added ability to print calibration parameters without sample
measurement. (SPAA-112)
- Updated National Instruments driver version (16.0.0.171).
(SPAA-115)
- Audit trails are created for:
Addition of methods to worksheets (SPAA-88)
Editing of sample names, weights, volumes and dilutions
(SPAA-89)
Insertion and deletion of samples (SPAA-90)
Changes to column properties (SPAA-94)
Instrument zero operations (SPAA-95, SPAA-179)
Changes to QCP limit options (SPAA-100)
Report generation (SPAA-132)
Analysis completion (SPAA-185)
System generated analysis stop and pause actions (SPAA-
194)
Continue action for paused analysis (SPAA-197)
- Additional detail added to audit trails created for:
Changes to Read Reagent Blank with Calibration Reslope
(SPAA-97)
Changes to QCP Rate Mode (SPAA-99)
- Removed ability for unprivileged users to:
Delete methods from worksheets (SPAA-85)
2
Agilent SpectrAA Software Release Notes
Upgrade
FW Release SDA Database
Name SW Version registration OS version SCM version Instrument supported The following new features were added:
Version date version version
required
Change solution type (SPAA-101)
Run Fast Sequential and SRM wizards (SPAA-102)
Add methods to library (SPAA-146)
Load library methods to worksheet (SPAA-147)
Add reference calibrations (SPAA-150)
Create new worksheets using New from (SPAA-182)
- Removed ability to delete method which has data associated
with it from worksheets within CFR environment. (SPAA-61)
- Removed ability for Fast Sequential and SRM wizards to be
run on fully approved (read only) worksheets. (SPAA-104)
- Removed atomic symbol prefix from sample name in reports.
(SPAA-118)
- Corrected multiple issues with SRM wizard operations.
(SPAA-113, SPAA-114, SPAA-173, SPAA-214)
- Corrected multiple issues with furnace methods. (SPAA-123,
SPAA-129, SPAA-195)
- Corrected multiple issues allowing users to avoid entering a
reason for change within CFR environments. (SPAA-131,
SPAA-133, SPAA-152)
- Corrected issue with SP100 program folder access under
Program Data. (SPAA-71)
- Corrected issue with SCM Administrator password reset
using program installation files. (SPAA-72)
- Corrected issue with previewing and printing reports on
Microsoft Windows 10 operating system. (SPAA-75)
- Corrected issue with display of Residuals and Calculated
Concentration in Calibration dat.a (SPAA-79)
- Corrected issue with manually entered solution dilution
factors missing in exports from label page. (SPAA-81)
- Corrected issue with untracked changes to QC solution
definitions. (SPAA-98)
- Corrected issue with tracking changes to column properties.
(SPAA-134)
- Corrected issue with missing GMT time and date for selected
audit trail entries. (SPAA-141)
- Corrected issue with loss of audit trail on re-saved
worksheets. (SPAA-137)
- Corrected issue with password protection security action on
application start-up. (SPAA-158)
- Corrected issue with saving worksheets created using New
from within CFR environment. (SPAA-183)
- Corrected issue with saving of statistical limits. (SPAA-191)
- Corrected issue with application of idle time lock. (SPAA-206)
- Included additional multi-element coded lamps. (SPAA-157)
- Updated CFR privileges documentation. (SPAA-117)
3
Agilent SpectrAA Software Release Notes
SpectrAA 5.4.06 2.11 Sep-16 Y Microsoft 1.10.303.1272 2.2.4.33 Microsoft Agilent 55B AA - SpectrAA Base, Aglilent SpectrAA Pro and Agilent SpectrAA
Base and Pro Windows 7 Windows Agilent 240 AA CFR are supported on Windows 10 Pro 64-bit and Windows 10
v5.4 Professional or 2008 Native Agilent 240/280 FS AA Enterprise 64-bit.
SpectrAA 5.4 * Enterprise 64-bit Directory Agilent 240/280 Zeeman AA - Added Windows 10 Logitech Camera driver support. (SPAA-
CFR v5.4 SP1 Microsoft Agilent SPS 4 autosampler 54)
Microsoft Server 2014 - Added Windows 10 Software Registration support. (SPAA-1)
Windows 10 R2 64-bit - Added Windows 10 SPS 4 driver support. (SPAA-2)
Professional or - Agilent SpectrAA CFR is supported on Microsoft Windows
Enterprise 64-bit Server 2008 R2 and Microsoft Windows Server 2012 R2.
- Updated National Instruments driver to Windows 10
compatible version. (15.5.0.49152)
* The Agilent - Added Windows 10 compatible Logitech Camera driver
SpectrAA CFR version. (13.80.853.0)
software will not - Updated Windows 7 Logitech Camera driver version.
work under the (13.51.823.0)
following: - Updated dual.exe to work on Windows 10. (SPAA-21)
- Microsoft - Updated Optimize → Optimize Signal to ensure the controls
Windows 7 for the autosampler are active, and the autosampler probe can
Professional or continue to be moved to a solution to optimize the signal.
Enterprise 32-bit (SPAA-3)
SP1 - Multiple HC lamps can now be optimized (SPAA-4).
- Microsoft - Enabled the signal graphics to be reliably saved with
Windows 10 worksheet. (SPAA-5)
Professional or - Optimized the drop in signal response rate when going in to
Enterprise 32-bit the optimize dialog. (SPAA-6)
- Operating - When the optimize page is opened and closed, the probe
systems other height now remains at the set probe height. (SPAA-7)
than English - When using QCP testing, Linear calibration type and a failed
correlation coefficient result occurs it is now correctly flagged
with a '+'. (SPAA-9)
- When using QCP testing, Linear Origin calibration type and a
failed correlation coefficient result occurs it is now correctly
flagged with a '+'. (SPAA-9)
- The correct parameters for the Validate.exe are used for
testing GTA120. (SPAA-10)
Flow of 0.3 LA/min
Hot Inject checkbox is not selected as default
- The "Fatal Error: Time moved error" no longer appears when
an HCL lamp has peaked and the Optimize Lamp graphic
becomes active on the Optimize page. (SPAA-11)
- The default current for the Se UltrAA lamp is now correctly
set to 10 mA. (SPAA-12)
- Removed incorrect text “If you are using an ULTRAA lamp,
you should use a lamp current of 15mA. (Note - the default
lamp current is 10mA).” from the Se Cookbook. (SPAA-13)
- QC Standard test failure in SpectrAA Base is correctly flagged
with "Q" in the analysis grid. (SPAA-14)
- Traditional Chinese native language is no longer supported.
- Installation of SpectrAA CFR is only supported under the
following scenarios:
SpectrAA, SDA and SCM all on the same PC
SpectrAA, SDA and SCM all on separate PCs (where SDA
can be installed on a Windows Server)
SpectrAA on the instrument PC, SDA and SCM on a separate
PC, or Windows Server.
SpectrAA 5.3.9 2.07 Apr-15 Y Microsoft 1.9.303.1269 2.2.4.26 Microsoft Agilent 55 AA The following items have been implemented in this version of
Base and Pro Windows 7 Windows Agilent 240 AA SpectrAA:
v5.3 2008 Native
4
Agilent SpectrAA Software Release Notes
Upgrade
FW Release SDA Database
Name SW Version registration OS version SCM version Instrument supported The following new features were added:
Version date version version
required
SpectrAA Professional 64- Directory Agilent 240/280 FS AA - Agilent SPS 4 autosampler support including the variable
CFR v5.2 5.2 (38) bit SP1 Microsoft Agilent 240/280 Zeeman AA pump speed feature.
Server 2008 - Added the latest software registration version.
R2 SP1 64-bit
The following defects have been fixed in this version of
SpectrAA:
- Fixed some Japanese mistranslations and display issues.
(TT199515, TT200174, TT212468, TT213755)
- Resolved an issue where Japanese worksheets would not
open if the file name used em-sized characters. (TT222673)
- Data in a Japanese langauge report no longer wraps to the
next line when there is a negative value. (TT213757)
- The User Defined QC test now passes when all pass criteria
are met. (TT220888)
- The ETC 60 functional test in the Validate application now
correctly loads the method. (TT237947)
- The VGA 77 test specifications in the Validate application are
now updated to the correct values. (TT238019)
- The SpectrAA Help has been updated.
SpectrAA 5.2 (38) 2.07 May-13 N Microsoft 1.9.303.1269 2.2.4.26 Microsoft Agilent 50/55 AA The following defects have been fixed in this version of
Base and Pro Windows 7 Windows Agilent 110/140/220/240 AA SpectrAA:
v5.2 Professional 64- 2008 Native Agilent 240/280 FS AA - Cal Zero is no longer subtracted from non-origin calibration
SpectrAA 5.2 bit SP1 Directory Agilent 800/880 AA algorithms when a replicate is masked in Cal Zero.
CFR v5.2 Microsoft Agilent 880/220/240/280 - A worksheet run will now pause on a reslope failure where
Server 2008 Zeeman AA the reslope has been triggered by a QC failure.
R2 SP1 64-bit
The following only applies to the SpectrAA CFR software.
The Varian 21CFR11 Assistant is no longer available.
The 21 CFR Part 11 compliance is now provided using the
Spectroscopy Configuration Manager (SCM) and
Spectroscopy Database Administrator (SDA).
The SCM provides you with the tools to manage your 21 CFR
Part 11 environment. SCM provides the means to create,
configure and maintain data in relation to system security,
user management and data paths. The SCM is used to set the
privileges and profiles of each user. SCM controls which
application/functions may be run by a particular user. It also
establishes the level of authority a user may have with regard
to signatures and accessing certain parts of an application. A
Group, Project, User, User Profile and Privileges are created
and assigned to a Station (name of the computer where the
Instrument software has been installed) in SCM. For more
details on SCM please refer to the Help.
5
Agilent SpectrAA Software Release Notes
SpectrAA 5.2 (32) 2.07 Dec-11 Y Microsoft Agilent 50/55 AA The following features applies to SpectrAA Base and Pro:
Base and Pro Windows 7 Agilent 110/140/220/240 AA - Modification to Agilent SpectrAA Base and Pro Version 5.2 to
v5.2 Professional 64- Agilent 240/280 FS AA implement software registration, which must be completed
SpectrAA 5.1.27** bit SP1 Agilent 800/880 AA after installation. Please note that you have up to 60 days to
CFR v5.1 Agilent 880/220/240/280 register your software. The registration process is free and
** Agilent Zeeman AA simple. You will be prompted to register the first time you run
SpectrAA CFR the application.
software v5.1 is - Modification to Agilent SpectrAA Base and Pro Version 5.2 to
only supported implement National Instruments driver version 488.2.2.7.3 for
on Microsoft Windows 7 64-bit compliance.
Windows XP 32- - Modification to Agilent SpectrAA Base and Pro version 5.2 to
bit SP3 fix a problem with integrate repeat measurement mode.
- Modification to Agilent SpectrAA Base and Pro version 5.2 to
fic a problem with Japanese and Simplified Chinese
translations in the Cookbook.
- To enable operating with the Windows 7 operating system,
the default location for the software executables and
configuration files have changed.
The SpectrAA Base and Pro install folder and associated
executables can be found in
C:\Program Files(x86)\Agilent\SP100\Run.
The SpectrAA Base and Pro DATA folder is located in
C:\Users\Public\Documents\Agilent\sp100.
The SpectrAA Base and Pro SAMPLERS and WORKGRP.&
folders are located in
C:\ProgramData\Agilent\sp100.
Agilent SpectrAA CFR version 5.1 (27) has not been updated.