Intravenous Fluid Monitoring and Alerting System
Intravenous Fluid Monitoring and Alerting System
ABSTRACT
Internet of Things” by Wang et al. (2018): This study alarms, lights, or digital notifications (on
presents the design and implementation of a wireless mobile devices or nursing stations) in case of
intravenous drip monitoring system based on the IoT. irregularities like:
The system can monitor the infusion process, detect
any problems, and send alerts to healthcare providers. o Completion of the IV fluid.
We reviewed the use of wireless sensor networks in o Flow rate anomalies (too fast or too
healthcare, including intravenous drip monitoring slow).
systems. The authors discussed the advantages and o Air in the line or occlusion.
disadvantages of different wireless sensor network
technologies and highlighted the need for secure and 2. Display Interface: Show real-time data about
reliable communication protocols in healthcare IV status, such as the remaining volume and
systems. current flow rate.
We conducted a qualitative study to explore the 3. Connectivity Features (optional): Integrate
experiences of nurses using intravenous drip with hospital systems, allowing remote
monitoring systems in a hospital setting. The authors monitoring and data logging for records.
found that the use of these systems improved the
efficiency and safety of medication administration, but Objectives:
also required additional time and effort from nurses to 1. Patient Safety:
learn and use the technology.
We conducted a systematic review of the literature o Prevent complications like fluid
on smart infusion pumps, which are a type of overload, dehydration, or air embolism.
intravenous drip monitoring system. The authors o Minimize human error in monitoring
found that the use of smart infusion pumps reduced and administration.
medication errors and improved patient safety, but
also noted that the implementation of these systems 2. Efficiency in Caregiving:
can be challenging and requires careful planning and o Reduce the manual effort required for
staff training. conducted a retrospective analysis of
frequent IV checks by nurses or
medication error reports in a hospital setting before
caregivers.
and after the implementation of smart infusion pumps.
The authors found that the use of these pumps reduced
o Enable caregivers to attend to multiple 2. Alerting System
patients without constant physical
A mechanism integrated into the IV setup that
monitoring.
generates alarms or notifications when specific
3. Automation and Precision: conditions are met, such as low fluid levels,
occlusions, or flow rate deviations.
o Ensure accurate and consistent IV fluid
delivery by automating flow rate and 3. IV Flow Rate:
volume measurements.
The speed at which fluid is delivered into the
Applications: patient, typically measured in millilitres per
hour (mL/h).
Hospitals and Clinics: For real-time monitoring
of IV fluids in patients in wards, ICUs, and 4. Infusion Pump:
operating rooms.
A medical device that delivers controlled
Home Healthcare: For patients receiving IV amounts of fluids, including medications or
therapy outside hospital settings, allowing nutrients, precisely controls the rate of IV fluid
remote supervision by medical professionals. delivery into a patient’s system.
Emergency Medical Services: During patient 5. Occlusion:
transport, where quick response to IV issues is A blockage in the IV line that interrupts fluid
critical. flow, requiring immediate attention to ensure
continuous delivery.
Challenges:
6. Air Embolism Detection:
Cost: Implementation of such systems can be
expensive, especially in resource-constrained A safety feature of monitoring systems that
settings. detects the presence of air bubbles in the IV line
to prevent air from entering the bloodstream.
Power Dependency: Continuous operation
requires reliable power sources or backup. 7. Backflow Prevention:
Integration: Ensuring compatibility with Mechanisms in the system to prevent the
existing hospital equipment and systems. reverse flow of blood or fluids due to gravity or
other factors.
User Training: Caregivers need adequate
training to interpret and respond to system 8. Volume Limit Monitoring:
alerts effectively.
A feature that tracks the total volume of fluid
C. DEFINITIONS AND TERMINOLOGIES infused to avoid over-infusion or fluid overload.
An IV Fluid Monitoring and Alerting System is a 9. Drip Rate:
technological solution designed to monitor intravenous
The number of drops delivered per minute
(IV) fluid therapy, ensuring that fluids are
(drops/min) through a drip chamber in gravity-
administered accurately and in a timely manner. It
fed IV systems.
helps avoid complications like over-infusion, under-
infusion, or IV bag depletion by alerting caregivers. 10. Fluid Bag Level Sensor:
Below are definitions and terminology commonly
associated with such systems: A sensor that detects and monitors the amount
of fluid remaining in an IV bag.
Key Definitions
11. Intravenous (IV) Therapy:
1. IV Fluid Monitoring The administration of fluids, medications, or
nutrients directly into a vein.
The process of continuously observing and
12. IV Drip:
measuring the flow rate, volume, and status of
A slow infusion of fluid from a container
IV fluids being administered to a patient.
(usually a bag or bottle) into the bloodstream.
13. Drop Rate: Systems that adjust alerts and flow rates based
on individual patient parameters, such as age,
The number of fluid drops delivered per minute weight, and medical condition.
(drops/min), often used in manual infusion
systems. 8. Telemetry:
Wireless transmission of IV monitoring data to a
14. Drip Chamber: remote display or nursing station for centralized
A transparent section of the IV tubing where oversight.
drops of fluid can be seen and counted. 9. Calibration:
Terminology The process of adjusting and validating the
accuracy of the monitoring system's sensors to
1. Gravity-Driven IV System: ensure precise fluid measurements.
An IV setup where fluid flow is controlled 10. Healthcare Internet of Things (IoT):
manually using a roller clamp, relying on Integration of IoT devices in IV fluid monitoring,
gravity to deliver the fluid. enabling real-time data sharing and predictive
2. Smart IV Pumps: analytics.
1. Fluid Depletion: The condition where the IV 2. IoT Integration: Connectivity with other
bag or container is nearly empty. devices in a healthcare network for centralized
monitoring.
2. Occlusion: A blockage in the IV line that
restricts fluid flow. 3. User Interface (UI): A display or control panel
for healthcare providers to adjust settings and
3. Air Embolism: A dangerous condition caused view real-time data. A display or app that shows
by air entering the bloodstream through the IV real-time data about the infusion, such as flow
line. rate, total volume infused, and time remaining.
4. Over-infusion: Administering fluids at a rate Common Alerts and Notifications
higher than intended, which may lead to fluid
overload. 1. Flow Irregularity Alert
5. Under-infusion: Administering fluids at a rate Indicates deviations from the preset flow rate.
lower than required, potentially causing 2. Bag Replacement Notification
dehydration or under-medication.
Warns caregivers when the IV bag is nearly
Functional Terminology empty.
1. Calibration: The process of adjusting the 3. Battery Low Alert
sensors or the system to ensure accurate
readings. Signals that the system's power source needs
charging or replacement.
2. Alert Threshold: Pre-set limits for parameters
like fluid level, flow rate, or pressure that 4. Line Disconnection Alert
trigger alerts.
Detects when the IV line becomes disconnected
3. Real-time Monitoring: Continuous observation from the patient or IV bag.
of IV parameters, providing immediate
updates. 5. Tampering Alert
4. Remote Alerting: Notification sent to remote Notifies caregivers if the system detects
devices, such as smartphones or nurse stations, unauthorized adjustments.
in case of an anomaly.
5. Fail-safe Mechanism: A feature designed to
maintain patient safety in the event of a system
failure. D. IDENTIFYING RESEARCH GAPS
1. Limited Real-Time Accuracy and Precision Issue: Current systems may not adequately
address patient-specific factors, such as fluid
Issue: Many systems may not accurately dynamics in pediatric or geriatric populations.
measure fluid levels in real time, especially
when dealing with microfluidic adjustments or Gap: Limited customization for varying patient
varying infusion rates. needs and medical conditions.
Gap: Need for more precise, sensor-based Potential Research Direction: Personalized IV
systems that can adapt to changes in infusion monitoring systems that consider patient-
rates, fluid types, and environmental conditions specific parameters like body weight, blood
(e.g., temperature, pressure). pressure, or underlying conditions.
Issue: Existing systems often operate as Gap: Lack of energy-efficient, lightweight, and
standalone devices, lacking interoperability portable designs.
with hospital information systems (HIS) or Potential Research Direction: Use of low-power
electronic health records (EHR). electronics, energy-harvesting technologies, or
Gap: Limited connectivity for data sharing and wearable IV monitoring systems for home care
alert generation in centralized healthcare or rural health settings.
systems.
Potential Research Direction: Design of IoT- 6. Cost and Accessibility
enabled IV monitoring devices that integrate
seamlessly with EHR platforms and support Issue: Advanced IV monitoring systems can be
real-time communication with healthcare expensive and are often inaccessible in low-
providers. resource settings.
Gap: Limited affordability and scalability of
solutions.
3. Alert Fatigue and User Experience
Potential Research Direction: Development of
Issue: Many systems generate frequent alerts, cost-effective, modular systems using readily
including false positives or non-urgent available materials and open-source
notifications, leading to "alert fatigue" among technologies.
medical staff.
Gap: Lack of intelligent prioritization or
filtering mechanisms for alerts. 7. Data Security and Privacy
Potential Research Direction: Development of Issue: With the integration of IoT, data security
machine learning algorithms to analyze data and patient privacy become critical challenges.
Gap: Insufficient safeguards for protecting There is a need for sensors that can monitor the
sensitive patient data transmitted via wireless entire infusion process, including flow rate, drip
networks. rate, and temperature.
Potential Research Direction: Closed-loop control
Implementation of robust encryption protocols A closed-loop control mechanism could
and blockchain technologies to secure data flow dynamically adjust the infusion rate based on
in IV monitoring systems. the patient's condition and fluid balance.
Nurse workload
8. Failure Response Mechanisms Nurses are already stressed out from the
frequency of monitoring IV drips, and may not
Issue: Current systems may not effectively have enough time to give each patient adequate
handle hardware or software failures, care.
potentially compromising patient safety.
Patient and caregiver involvement
Gap: Lack of robust fail-safe mechanisms or
redundancy features. Patients and caregivers can help nurses by
notifying them of problems, but they may
Potential Research Direction: Design of systems sometimes cause errors due to a lack of
with self-diagnostic and recovery features to knowledge.
ensure continuous operation.
Some research and development has been done
to address these gaps, including:
9. Limited Clinical Validation Self-powered droplet triboelectric sensor
Issue: Many proposed systems are not (SDTS):
thoroughly tested in clinical environments. This sensor can monitor both the flow and type
Gap: Insufficient validation of system of intravenous infusion.
performance across diverse healthcare settings Intravenous Fluid Monitoring and Controlling
and patient demographics. System:
Potential Research Direction: Comprehensive This system uses sensors to collect real-time
clinical trials and validation studies to assess data on fluid infusion dynamics, and can
efficacy and reliability in real-world scenarios. generate alerts to notify healthcare providers.
Drip Meter:
10. Emerging Technologies Underexplored This open-source opto-electronic system
Issue: Technologies like artificial intelligence displays drip rate and has alarms to alert when
(AI), augmented reality (AR), and 5G networks the rate deviates from a pre-set value.
remain underutilized in IV fluid monitoring E. JUSTIFYING RESEARCH
systems.
Justifying the research on an IV Fluid Monitoring
Gap: Minimal exploration of how these and Alerting System involves addressing its critical
technologies can transform IV fluid monitoring importance in healthcare, identifying current
and alerting. challenges, and emphasizing the benefits it brings
Potential Research Direction: Integration of AI to patients, healthcare professionals, and the
for predictive analytics, AR for enhanced broader medical ecosystem. Below is a detailed
visualization of fluid flow, and 5G for ultra-fast, justification for such research:
low-latency data transmission. 1. Critical Role of IV Therapy in Healthcare
Lack of multifunctional sensors Common Medical Practice:
Intravenous (IV) therapy is a cornerstone of Modern systems can provide historical and
modern medicine, used for hydration, predictive insights, optimizing IV therapy and
administering medications, and delivering reducing risks.
nutrients to patients in critical care, surgical
settings, and routine hospitalizations. 4. Impact on Healthcare
Accurate and timely administration of IV fluids Automated alerts reduce the risk of
is vital for patient safety and recovery. Delays, complications from IV therapy, improving
inaccuracies, or disconnections can lead to outcomes for patients.
severe complications, including dehydration, Efficiency for Caregivers:
overhydration, or missed medication doses.
Healthcare workers can focus on critical tasks
2. Challenges in Traditional IV Monitoring instead of routine monitoring, reducing
Manual Monitoring Limitations: workload and stress.
Accuracy: Load cells provide precise and Light Source: A backlight (usually LED) shines
accurate measurements, often with tolerances through the layers of liquid crystals.
of 0.1% or better. Polarization: The light passes through
Capacity: Load cells come in various capacities, polarizing filters before hitting the liquid
ranging from very small, precise measurements crystals.
in grams to very large, industrial capacities in Control of Light: The liquid crystals are
tons. aligned by electric fields applied through
Signal Conditioning: Load cells often require electrodes. This controls how much light is
signal processing, such as amplification and allowed through.
filtering, to provide usable output. Color Filters: For color displays, red, green,
Maintenance: and blue (RGB) filters are used to produce a
full spectrum of colors.
Proper calibration and maintenance are essential to
ensure the accuracy of load cells. They are sensitive to Types of LCD Displays:
temperature, humidity, and environmental conditions, 1. Twisted Nematic (TN): These are the most
which can affect performance. Regular testing is common and affordable, providing fast
crucial, especially in industrial environments. response times but lower color accuracy and
. viewing angles.
2. In-Plane Switching (IPS): Known for better
color accuracy and wider viewing angles
compared to TN panels, making them ideal for
3. LCD DISPLAYS applications like photography and design.
3. Vertical Alignment (VA): Provides better
contrast ratios than TN or IPS but may have
An LCD (Liquid Crystal Display) is a flat-panel limited viewing angles.
display technology that uses liquid crystals to produce
images. The liquid crystals themselves do not emit light 4. OLED (Organic LED): While not technically
but instead modulate the light passed through them to LCD, OLED is another display technology that
create images or text. Here are the key components uses organic materials and provides superior
and features of LCD displays: contrast ratios and deeper blacks than
traditional LCDs.
Key Components:
Advantages of LCD Displays:
1. Liquid Crystals: These materials have
properties between those of liquids and solid Energy Efficient: LCDs consume less power
crystals, allowing them to control light than older technologies like CRTs.
transmission when exposed to electric currents. Thin and Lightweight: They can be made very
thin and compact, making them suitable for
portable devices.
Sharp and Clear Images: When well-calibrated, o The module is designed to be energy
LCDs can offer sharp, clear images with good efficient, making it ideal for battery-
color accuracy. operated applications like portable
Wide Usage: LCD technology is used in TVs, scales.
smartphones, monitors, laptops, digital clocks, 3. Single-Channel Differential Input:
calculators, and more.
o It features a single-channel differential
Limitations of LCD Displays: input for measuring the output from a
Limited Contrast: Black levels can appear grey load cell. The differential input means it
because the backlight is always on, even when can measure small voltage differences
the display shows black. between the two terminals of the load
cell.
Viewing Angles: Certain LCD types (especially
TN panels) can suffer from color distortion and 4. Built-in Programmable Gain:
reduced visibility at wide angles. o The HX711 has a built-in gain amplifier
Response Time: While improved, response (with selectable gain factors of 32, 64, or
times can still be slower than technologies like 128) which amplifies the small voltage
OLED, leading to motion blur in fast-moving from a load cell to a readable level for
images, such as in gaming or videos. the ADC.
Applications of LCD Displays: 5. Simplicity:
Consumer Electronics: LCDs are used in o The HX711 communicates with a
televisions, smartphones, tablets, and laptops. microcontroller using a simple two-wire
interface: Clock (SCK) and Data (DT).
Digital Signage: For information display in This simplifies integration into a system
public spaces, stores, and transport hubs. and is widely supported by popular
Industrial and Medical Equipment: Used for platforms like Arduino and Raspberry
instrumentation and display panels in various Pi.
machinery. 6. Filter for Noise Reduction:
Automotive Displays: Dashboard displays and o The device includes an internal low-pass
navigation systems in modern vehicles. filter, which helps in reducing noise and
interference from the environment,
making it ideal for use in noisy
4. HX711 LOAD SENSOR
environments.
The HX711 is a popular 24-bit analog-to-digital
7. Easy Calibration:
converter (ADC) designed for use with load cells in
applications such as weight measurement and force o Calibrating the HX711 with a known
sensing. It is commonly used in digital weighing scales weight is a straightforward process,
and other systems that require precise measurement of allowing for accurate readings that
weight or force. Below is a detailed note on the HX711 match real-world weight values.
load sensor module: Working Principle:
Key Features of the HX711: A load cell generates a small electrical signal (in
1. High Precision: the millivolt range) when subjected to force or
weight. The HX711 amplifies this signal
o The HX711 offers a 24-bit resolution for
through its internal gain amplifier.
conversion, which allows it to provide
highly accurate measurements with low The ADC converts the analog signal from the
noise. It can convert analog signals from load cell into a digital value that can be
load cells with high precision. processed by a microcontroller.
2. Low Power Consumption:
The microcontroller can then interpret this Timers:
value, often in terms of weight, and display it or In kitchen appliances, sports timers, or
use it in a control system. countdowns.
Applications: Indicators:
Weighing Scales: Digital weight scales, In various devices to show a certain status or
industrial weighing systems, and kitchen scales. condition (like error or warning indicators in
Force Measurement: In applications such as machines).
load testing, force sensing, and pressure Games and Quizzes:
measurement.
Used for signaling answers or marking events
Robotics and Automation: For measuring load in game shows.
and weight in robotic systems or automated
manufacturing processes. .
Agriculture: For systems that monitor crop 6. SERIAL INTERFACE BOARD MODULE
weight, animal weight, or storage systems. A serial interface board module is a hardware
Wiring and Pinout: component used to establish communication between a
device and a microcontroller or computer using serial
VCC: Power supply (usually 5V or 3.3V communication protocols, such as RS-232, RS-485, or
depending on the microcontroller). TTL (Transistor-Transistor Logic). These modules are
GND: Ground connection. widely used in embedded systems, communication
DT (Data): The data output pin that sends the devices, and data transfer applications.
digital data to the microcontroller. Key Features:
SCK (Clock): The clock pin that synchronizes 1. Serial Communication:
the data transfer. Allows for the transfer of data bit by bit in a
V0/V1: These pins are used for the gain setting, sequence over a single communication channel.
depending on the load cell configuration. It is used for connecting peripherals to
5. BUZZER microcontrollers or computers.
1. Male-to-Male: Pins on both ends for connecting o Widgets include graphs, buttons, sliders,
two female sockets. text displays, and more, tailored for IoT
interactions.
2. Male-to-Female: One pin and one socket end,
used to connect a component with pins to a 3. Cloud Infrastructure
breadboard. o A robust and scalable cloud platform for
3. Female-to-Female: Sockets on both ends, for managing connected devices.
connecting two male pins. o Offers device management, data storage,
Features: and over-the-air (OTA) firmware
updates.
Flexibility: Easily plugged in or removed for
quick adjustments. 4. Blynk.Edgent
Blynk IoT is a versatile and user-friendly platform o Features like white-labeling for creating
for building, managing, and scaling Internet of Things branded IoT apps.
(IoT) applications. It simplifies the integration of
7. Data Analytics and Reporting Reduction in Human Errors: Automation
reduces reliance on manual observations,
o Includes tools for visualizing device data minimizing the chance of oversight.
and generating reports for analysis.
Applications
2. Improved Efficiency for Healthcare Providers
Home Automation
Time-Saving: Healthcare staff can focus on
Industrial IoT other tasks without continuously monitoring IV
Smart Agriculture fluid levels manually.
Reduces the complexity of IoT development. Remote Alerts: Notifications can be sent to
nurses or physicians via mobile devices,
Flexible pricing options for hobbyists, enabling quick action even when they are not in
developers, and enterprises. the immediate vicinity.
Active community and extensive
documentation.
3. Precision and Customization
How It Works
Accurate Delivery Rates: Advanced systems
1. Connect your IoT device to the Blynk Cloud ensure precise control of drip rates, which is
using the Blynk libraries. especially important for critically ill patients or
those requiring specific doses.
2. Use the Blynk mobile app or web interface to
build your custom dashboard. Customizable Alerts: Alerts can be tailored to
specific thresholds, ensuring they are relevant
3. Deploy, monitor, and control your devices in to the individual patient's condition.
real time from anywhere.
4. Cost-Effectiveness
I. ADVANTAGES
Reduced Wastage: Accurate monitoring
An intravenous (IV) fluid monitoring and alerting prevents fluid wastage due to unnoticed leaks
system offers several advantages, particularly in or errors in administration.
healthcare settings where precision and timely
intervention are critical. Below are the key benefits: Fewer Adverse Events: By preventing
complications, the system reduces the costs
associated with treating avoidable conditions or
1. Enhanced Patient Safety extended hospital stays.
Prevention of Overload or Dehydration: Real-
time monitoring ensures that patients receive 5. Real-Time Data Collection and Analysis
the correct fluid volume, avoiding
complications such as fluid overload or Improved Record-Keeping: Data from the
underhydration. monitoring system can be integrated into
electronic health records (EHR), providing a
Early Detection of Malfunctions: Alerts can detailed history of IV therapy.
indicate issues like blockages, air bubbles, or
disconnections in the IV line, which could Clinical Insights: Analyzing trends in fluid
otherwise lead to severe complications. administration can help in making informed
decisions about patient care and treatment System Failures: Issues like battery depletion,
adjustments. sensor failure, or network connectivity
problems can disrupt monitoring.
Data Integration Issues: Difficulty in
6. Enhanced Patient Comfort and Satisfaction integrating with existing electronic health
Reduced Interruptions: Automated systems record (EHR) systems may lead to inefficiencies
reduce the need for frequent manual checks, or errors.
minimizing disruptions to the patient’s rest. 3. Dependency on Technology
Faster Responses: Prompt action in response to Reduced Manual Oversight: Over-reliance on
alerts improves patient outcomes and enhances automated systems may lead to reduced
trust in care delivery. vigilance from healthcare professionals.
Learning Curve: Staff may require training to
7. Scalability for Diverse Healthcare Settings effectively use and troubleshoot these systems,
which can delay adoption.
Applicability in Various Scenarios: From
outpatient clinics to intensive care units, IV 4. Privacy and Security Concerns
monitoring systems can adapt to different levels Data Breaches: Systems connected to hospital
of care and patient needs. networks may be vulnerable to cyberattacks,
Support for Telemedicine: These systems can risking patient data security.
integrate with remote monitoring tools, Ethical Issues: Continuous monitoring could
facilitating care in home settings or rural areas. raise concerns about patient privacy.
5. Limited Customization
J. DISADVANTAGES Patient-Specific Challenges: Standardized
An intravenous (IV) fluid monitoring and alerting systems may not accommodate unique patient
system is a valuable tool in healthcare for ensuring conditions, such as irregular infusion rates or
timely and accurate fluid administration. However, sensitivity to specific fluids.
these systems also have potential disadvantages and Adaptability: The system might struggle to
limitations, including: adapt to diverse clinical scenarios, like multiple
1. Cost and Accessibility simultaneous IV lines.
High Initial Cost: Implementing such systems 6. Potential for Over-reliance
requires purchasing specialized equipment, Delay in Manual Intervention: If staff rely
which can be expensive. heavily on alerts, they may not notice problems
Maintenance Costs: Regular calibration, that the system fails to detect.
repairs, and software updates can add to the Trust Issues: Erroneous readings can lead to
expense. distrust in the system, making staff hesitant to
Limited Accessibility: These systems may not fully depend on it.
be available in resource-limited settings or 7. Ethical and Operational Concerns
smaller healthcare facilities.
Potential Workflow Disruption: Integration
2. Technical Challenges into existing workflows can initially create
False Alarms: Alerts may be triggered inefficiencies and resistance among staff.
unnecessarily, such as due to patient movement Overuse of Alerts: If the system is not carefully
or technical malfunctions, leading to alarm configured, frequent alerts might disrupt care
fatigue among healthcare staff. delivery.
____ CHAPTER – 3________________ Design the system with an intuitive interface
that simplifies operation for healthcare
providers.
Include features like audible alarms, LED
SCOPE OF STUDY_____________ indicators, and mobile notifications for diverse
user preferences.
The objectives of an IV Fluid Monitoring and Alerting Allow customization of flow rate limits, fluid
System typically revolve around ensuring patient volume thresholds, and alert settings to
safety, improving healthcare efficiency, and providing accommodate different medical scenarios or
real-time monitoring capabilities. Below are some key patient needs.
objectives for such a system: 8. Cost-Effectiveness
1. Patient Safety Develop an affordable solution that minimizes
Prevent complications such as air embolism, the need for expensive monitoring equipment
backflow, or over-infusion by monitoring IV while ensuring high reliability and accuracy.
fluid delivery in real time. Reduce costs associated with manual errors,
Detect and alert medical staff of critical delayed response, or patient complications.
conditions such as low IV fluid levels or flow 9. Data Collection and Analytics
rate abnormalities. Gather data on IV fluid usage and patient
2. Automation and Efficiency response for better medical analysis and
Automate the monitoring of IV fluid flow rates research.
to reduce manual intervention and human Provide insights into fluid management trends,
error. which can aid in improving treatment
Optimize healthcare staff time by reducing the protocols.
need for frequent manual checks on IV fluids. 10. Portability and Scalability
3. Real-Time Monitoring Create a compact and portable system that can
Continuously measure and track IV fluid levels be easily integrated into various healthcare
and infusion rates. settings, including hospitals, clinics, and home
care.
Enable medical staff to remotely monitor IV
performance through integrated systems or Design a scalable solution to support a large
mobile alerts. number of patients simultaneously.
3. Economic Feasibility
Cost of Development and Deployment: D. GENERALIZABILITY
o Sensors, electronics, and software The generalizability of an Intravenous (IV) Fluid
development costs. Monitoring and Alerting System depends on several
o Manufacturing and maintenance key factors, including its design, adaptability, and
expenses. compatibility with different healthcare settings. Here
are some dimensions to consider when evaluating its
Cost-Benefit Analysis:
generalizability:
o Reduces manual IV monitoring by
nurses, saving time and labour costs.
1. Compatibility with Different IV Setups
o Prevents adverse events due to IV fluid
depletion or malfunction, reducing IV Equipment Standards: The system should be
healthcare costs. compatible with a wide range of IV bags,
pumps, tubing, and other hardware from
Market Demand: various manufacturers.
o High demand in settings with resource Sensor Integration: If the system uses sensors
constraints, high patient-to-nurse ratios, (e.g., weight, flow rate, or pressure sensors), it
or frequent IV usage. must be adaptable to different IV lines and
patient setups without requiring extensive User Training: Generalizable systems should
customization. have a straightforward interface that requires
minimal training for healthcare professionals
and caregivers.
2. Environment and Clinical Context
Maintenance: Low maintenance requirements
Hospital Settings: The system should function increase usability across resource-limited
seamlessly in intensive care units (ICUs), settings.
general wards, and emergency rooms.
Outpatient and Home Care: For chronic
patients or home-based setups, the system must 7. Regulatory and Standardization Compliance
be lightweight, portable, and easy to use for The system must comply with international and
non-professionals. regional standards for medical devices (e.g.,
Field Applications: Adaptability for use in FDA, CE marking) to ensure it can be used in
military, disaster relief, or remote locations different countries.
enhances its generalizability.
8. Cultural and Language Flexibility
3. Interoperability Customization of alerts, interfaces, and
Integration with Hospital Systems: The system instructions in various languages and cultural
should connect with Electronic Health Records contexts enhances its usability globally.
(EHR) and other hospital management systems.
Alerts and Communication: It should support 9. Cost and Accessibility
multiple communication channels (e.g., alarms, Affordability and accessibility are critical for
nurse call systems, or mobile apps) to cater to adoption in low- and middle-income countries
different workflows. or rural areas. The system should have scalable
options based on resources available.
4. Patient Variability
Demographics: The system must account for E. ASSUMPTIONS
varying patient needs, including neonates, When designing an IV Fluid Monitoring and
adults, and elderly patients, who may require Alerting System, the following assumptions can be
different IV flow rates and volumes. considered to ensure effective functionality and
Clinical Conditions: It should work reliably for usability:
different conditions requiring IV therapy, such System-Related Assumptions
as hydration, medication delivery, or blood
transfusion. 1. Hardware Compatibility:
o The system will integrate with standard
IV fluid bags, infusion pumps, and stand
5. Technical Robustness configurations.
Scalability: The system should handle different
o Sensors will be able to detect fluid levels,
levels of deployment, from single-patient use to
flow rates, and abnormalities accurately.
large hospital networks.
2. Power Supply:
Adaptability to Power and Connectivity: It
must function in environments with varying o The system will have a stable and
power and internet availability, with offline reliable power source, whether battery-
functionality if necessary. operated or through a fixed connection.
3. Data Communication:
6. Ease of Use o The system will use wired or wireless
communication (e.g., Wi-Fi, Bluetooth)
to transmit alerts to designated devices o Users will respond promptly to alerts to
or monitoring systems. prevent IV delivery interruptions or
4. Software Reliability: adverse patient outcomes.
A backup mechanism is necessary in case of The device should not cause discomfort or
sensor or system failure. hinder the patient’s movement.
User Accountability:
Regularly review progress with stakeholders to 2.1.1 Gather user and technical
confirm deliverables align with requirements. requirements
Conduct user acceptance testing (UAT) with 2.1.2 Document functional and
medical staff before deployment. non-functional requirements
o 2.2 System Architecture Design
6. Scope Control 2.2.1 Hardware component
selection (sensors,
Use a formal change request process to microcontrollers)
evaluate, approve, or reject scope changes.
2.2.2 Software architecture design
Document all changes and communicate
updates to stakeholders. o 2.3 Interface Design
Monitor for scope creep through regular status 2.3.1 User interface (UI) design
meetings and milestone reviews. 2.3.2 Alerting mechanisms
(visual, auditory, mobile
notifications)
7. Success Criteria
3. System Development
The scope will be considered successfully managed
if: o 3.1 Hardware Development
1. The IV Fluid Monitoring and Alerting System 3.1.1 Procure hardware
is delivered within agreed parameters (time, components
cost, quality). 3.1.2 Assemble hardware
2. All functional and non-functional requirements 3.1.3 Test hardware functionality
are met.
o 3.2 Software Development
3. Stakeholders approve and accept the final
3.2.1 Develop sensor data
deliverables.
processing algorithms
3.2.2 Integrate real-time
I. WORK BREAKDOWN STRUCTURE monitoring system
A Work Breakdown Structure (WBS) is a 3.2.3 Develop alerting and
hierarchical decomposition of a project into smaller, notification system
manageable components. Below is an example WBS
o 3.3 System Integration An IV Fluid Monitoring and Alerting System is
3.3.1 Connect hardware with designed to monitor the flow of intravenous fluids (IV)
software to patients in medical settings, ensuring the proper
administration of fluids and electrolytes. The system
3.3.2 Ensure seamless data flow typically includes sensors, software, and alerting
4. Testing and Quality Assurance mechanisms to track and notify healthcare
professionals of potential issues, such as flow
o 4.1 Unit Testing irregularities, changes in pressure, or fluid imbalances.
4.1.1 Test hardware components Key Parameters of IV Fluid Monitoring and
4.1.2 Test software modules Alerting Systems
o 4.2 Integration Testing 1. Flow Rate:
4.2.1 Validate hardware-software o Description: The speed at which the IV
integration fluid is being administered, usually
measured in milliliters per hour
o 4.3 System Testing
(mL/hr).
4.3.1 Simulate IV fluid scenarios
o Importance: Maintaining the correct
4.3.2 Test alert reliability and flow rate is critical for patient care,
accuracy especially in critical care or surgical
o 4.4 User Acceptance Testing (UAT) settings where fluid balance is vital.
4.4.1 Gather feedback from o Monitoring: Systems should be able to
healthcare professionals track and adjust the flow rate, providing
real-time data on whether it aligns with
5. Deployment the prescribed rate.
o 5.1 Install system at healthcare facilities 2. Volume of Fluid Infused:
o 5.2 Train staff on system usage o Description: The total amount of fluid
o 5.3 Monitor initial deployment that has been infused over a certain
performance period.
o 5.4 Address deployment issues o Importance: This ensures that patients
receive the appropriate volume of fluids
6. Maintenance and Support as per their treatment protocol.
o 6.1 Provide technical support o Monitoring: Automated tracking and
o 6.2 Perform regular system updates display of the total volume administered
o 6.3 Monitor system performance to avoid over-infusion or under-infusion.
3. Pressure Monitoring:
o 6.4 Handle user feedback and requests
o Description: Measures the pressure
7. Documentation
inside the IV tubing and catheter.
o 7.1 System design documentation
o Importance: Abnormal pressure
o 7.2 User manual preparation readings (either too high or too low) can
o 7.3 Training materials development indicate blockages, kinks in the tubing,
or problems with the IV line.
o 7.4 Final project report
o Monitoring: Real-time pressure sensors
can alert healthcare providers if the
pressure deviates from normal ranges,
which may lead to infusion
J. RESEARCH PARAMETERS
complications or infection risks.
4. Infusion Site Monitoring:
o Description: Some systems monitor the o Description: Provides real-time updates
condition of the infusion site (where the on the patient's IV status to healthcare
catheter is inserted), detecting issues like providers, typically via a display or
swelling, redness, or infiltration. integrated software system.
o Importance: Ensures that the IV is o Importance: Ensures timely intervention
properly placed and that there are no if there are issues with the infusion,
complications, such as phlebitis or tissue reducing the risk of errors and
damage. complications.
o Monitoring: Sensors or cameras may be o Monitoring: Alerts for abnormalities
used to check the site for any visible (e.g., changes in pressure, deviation from
signs of complications. prescribed flow rates, or potential
5. Fluid Temperature: occlusions) are crucial for prompt
corrective actions.
o Description: Ensures that the IV fluids
9. Battery Life and Backup Power:
being infused are at the correct
temperature. o Description: Ensures that the
o Importance: Administering fluids that monitoring system remains functional
are too cold or too warm can cause even in the event of a power failure.
discomfort to the patient and might even o Importance: Continuous monitoring is
lead to adverse effects. essential, especially in emergency or
o Monitoring: Temperature sensors are critical care settings where patient safety
integrated to ensure that fluids remain depends on constant oversight.
within a safe range. o Monitoring: Systems are designed with
6. Air Embolism Detection: backup power, either through
rechargeable batteries or external power
o Description: Monitors for the presence sources.
of air bubbles in the IV fluid line.
10. Data Logging and Compliance:
o Importance: Air embolisms can be life-
o Description: Keeps a record of all IV
threatening, and detecting them early
can prevent severe complications. infusion data for regulatory compliance,
auditing, and clinical review.
o Monitoring: Ultrasonic or optical
o Importance: Helps in maintaining a
sensors are commonly used to detect air
bubbles, sending alerts if any are history of patient care and can provide
detected. insights for optimizing care protocols.
6. Iterative Refinement
D. POTENTIAL RISKS
Analyze feedback from pilot tests.
When designing or implementing an IV Fluid
Address technical or usability issues (e.g., false Monitoring and Alerting System, it is essential to
alarms, connectivity issues). assess potential risks to ensure the safety, reliability,
Improve device design, software interface, or and effectiveness of the system. Here are some
alert settings. potential risks:
1. Technical Risks
7. Full-Scale Deployment Sensor Malfunction or Failure: Sensors
monitoring the IV fluid level, flow rate, or
Training: patient vitals may fail, leading to incorrect
Educate healthcare staff on system usage. readings or no alerts.
Create a user manual and troubleshooting False Positives or False Negatives:
guide. o False positives may trigger unnecessary
Installation: alerts, causing unnecessary intervention
or alarm fatigue.
Install devices on IV stands across selected
units. o False negatives may result in undetected
Ensure proper integration with the hospital's critical conditions like flow stoppage or
IT systems. fluid depletion.
Quality Assurance:
Software Bugs: Software errors in the Physical Damage: The device could be damaged
monitoring system could lead to inaccurate due to drops, spills, or exposure to harsh
alerts or failure to notify caregivers. conditions in the clinical environment.
Power Supply Issues: The system may fail Compatibility Issues: The monitoring system
during power outages or battery depletion. may not be compatible with all IV fluid types or
setups, leading to inaccuracies in detection.
2. Network-Related Risks
Connectivity Issues: If the system relies on 6. Regulatory and Legal Risks
wireless or network connections to alert staff, Non-compliance with Standards: Failing to
poor connectivity can delay critical meet medical device regulations and standards
notifications. can result in legal and operational challenges.
Cybersecurity Threats: Unauthorized access to Liability Concerns: Inaccurate alerts or system
the system could compromise patient data failures leading to patient harm could expose
privacy or disrupt system functionality. the manufacturer or healthcare providers to
legal liability.
1. Hardware Failures
Risks: Sensor malfunctions, power failures, or Risks: Poor user interface design may lead to
connectivity issues. misinterpretation or delays in responding to
Mitigation Strategies: alerts.
Risks: Incorrect fluid level readings or false Risks: Delayed alerts could lead to IV fluid
alarms. depletion or overflow, risking patient harm.
3. Connectivity Issues
6. Cybersecurity Concerns
Risks: Network interruptions may delay or
Risks: Unauthorized access to patient data or
prevent alerts.
control systems.
Mitigation Strategies:
Mitigation Strategies:
o Design the system to support multiple
o Encrypt data during transmission and
communication methods (e.g., Wi-Fi,
Bluetooth, or cellular). storage using secure protocols (e.g., TLS,
AES).
o Include an offline mode where local
o Require authentication for accessing the
alerts (e.g., audible or visual alarms) are
triggered if connectivity is lost. monitoring system or altering
configurations.
o Use encrypted and robust data
o Regularly update the system firmware to
transmission protocols to minimize
packet loss. address security vulnerabilities.
o Build and test the hardware prototype o Provide ongoing support and
(sensor integration, calibration). maintenance.
o Develop the software backend for data o Plan for future updates or additional
processing and alert generation. features.
o Implement wireless communication (e.g., An IV fluid monitoring and alerting system is expected
Bluetooth, Wi-Fi). to result in improved patient safety by providing real-
time alerts when fluid levels are nearing depletion,
o Develop the user interface for alert potential complications like fluid overload are
notifications (e.g., via app or sound). detected, or infusion rate changes occur, ultimately
leading to more accurate fluid management, reduced
medication errors, and efficient nursing workflow by
Month 6: Integration and Testing minimizing the need for frequent manual checks; this
Week 17-18: can also enable remote monitoring of multiple patients
simultaneously, freeing up healthcare staff time for
o Integrate hardware and software
other tasks.
components.
Key expected outcomes:
o Conduct unit testing for individual
modules. Reduced risk of complications:
Week 19-20: Early detection of potential issues like fluid
overload or under-infusion due to real-time alerts,
o Perform system testing to ensure reliable
allowing for timely intervention and minimizing
operation under different scenarios.
adverse effects.
o Gather user feedback for potential Enhanced patient safety:
improvements.
Accurate monitoring of IV fluid levels leading to
safer and more controlled infusion therapy, especially
Month 7: Pilot Deployment for critically ill patients.
Week 21-22: Improved staff efficiency:
o Deploy the system in a controlled Automated monitoring system reduces the need for
environment (e.g., hospital or lab). frequent manual checks, allowing nurses to focus on
o Monitor real-time performance and other patient care tasks
collect data on system effectiveness. Better decision-making:
Week 23-24: Real-time data on fluid levels and infusion rates
provides clinicians with valuable information to make
o Address issues found during the pilot
informed decisions regarding fluid management
phase.
Remote patient monitoring:
o Finalize documentation for end users
and technical teams. Ability to monitor multiple patients' IV infusions
from a central location, facilitating better oversight
and rapid response to alerts
Month 8: Final Launch
Data logging and analysis:
Week 25-26:
System can store infusion data for later review and o User-friendly interfaces for setup and
analysis to identify trends and improve future patient operation.
care
o Compatibility with standard IV setups
Early detection of infusion problems: and electronic medical records (EMRs).
Alerts can be triggered in case of infusion rate 4. Automation and Precision
changes, potential blockages, or other irregularities in
o Reduces reliance on manual checks by
the IV line
healthcare professionals.
Specific benefits for different stakeholders:
o Enhances accuracy in fluid
For patients: administration.
Reduced risk of complications related to IV fluid 5. Applications
administration, improved comfort and quality of care
o Particularly beneficial in intensive care
For nurses: units, surgery, emergency care, and
Efficient workflow, reduced workload associated outpatient settings.
with manual monitoring, and improved ability to .
identify potential issues early
For healthcare providers:
I. ALIGNMENT WITH NEEDS
Enhanced patient safety, better data for informed
The IV Fluid Monitoring and Alerting System
decision-making, and potential for improved resource
directly addresses the problem of manual IV fluid
utilization.
monitoring, which is prone to human error, delayed
intervention, and missed warnings, all of which can
compromise patient safety. Here's how the system
solves the problem and meets the needs of the target
H. CLARITY AND CONCISENESS audience (healthcare professionals, hospitals, and
Clarity and Conciseness of IV Fluid Monitoring and patients):
Alerting System 1. Problem Addressed:
An IV fluid monitoring and alerting system ensures Manual Monitoring & Human Error:
precise and continuous tracking of intravenous fluid
levels and flow rates in clinical settings. The system Nurses and healthcare workers currently
typically comprises sensors, real-time data processing, monitor IV fluid levels manually, which can
and alert mechanisms to enhance patient safety and lead to oversight or delayed action.
improve workflow efficiency. Key points of its design Delayed Response to Changes:
and functionality include:
Traditional systems do not provide real-time
1. Continuous Monitoring monitoring, causing delays in response to
o Tracks fluid volume, flow rate, and changes in patient conditions.
potential obstructions in the IV line. Inconsistent Fluid Management:
o Detects deviations from prescribed levels Incorrect fluid administration or missed alerts
or flow irregularities. can lead to complications such as dehydration,
2. Real-Time Alerts fluid overload, or electrolyte imbalances.
o Provides visual, auditory, or digital Heavy Workload on Healthcare Staff:
notifications for healthcare providers Continuous manual monitoring adds to the
when abnormalities are detected (e.g., already heavy workload of nurses and
low fluid levels, air bubbles, or leakage). healthcare providers, leading to fatigue and
o Ensures timely interventions, reducing potential oversight.
risks of under- or over-infusion. 2. System's Solution:
3. Ease of Use Real-Time Monitoring:
The IV Fluid Monitoring and Alerting System o Better Decision-Making: With accurate,
provides real-time tracking of fluid levels, up-to-date information on fluid levels,
automatically detecting when fluid reaches healthcare providers can make informed
critical levels (either low or high) and notifying decisions about patient care.
the healthcare provider immediately. For Hospitals/Healthcare Institutions:
Automated Alerts:
o Improved Patient Outcomes: Early
The system sends instant alerts to staff, such as alerts allow quicker interventions,
visual, audio, or mobile notifications, helping leading to fewer complications and
them respond promptly and preventing missed better recovery rates.
or delayed interventions.
o Cost Efficiency: By reducing the
Data Logging and Reporting: incidence of fluid-related errors and
The system logs fluid usage, providing detailed complications, hospitals can lower
reports that help healthcare professionals assess treatment costs and improve resource
patient fluid status over time, ensuring that all utilization.
data is accurate and traceable. o Compliance and Documentation:
Integration with Hospital Systems: Automated data logging ensures that the
hospital meets regulatory requirements
The system can integrate with Electronic and quality standards for patient care.
Health Records (EHRs), enabling smooth
coordination across different departments and For Patients:
ensuring that all healthcare providers are on o Increased Safety and Comfort:
the same page.
Patients benefit from continuous,
Alerts for Fluid Imbalance: automated monitoring, which increases
The system can detect abnormal fluid flow the likelihood of early intervention and
rates or changes in patient conditions (such as reduces the risk of adverse events
signs of dehydration or overhydration) and related to fluid therapy.
provide specific, actionable alerts. o Peace of Mind:
Minimized Workload: Knowing that their IV fluids are being
By automating the monitoring process, closely monitored provides patients and
healthcare workers can spend less time on their families with reassurance that their
routine checks and more time on other aspects healthcare needs are being closely
of patient care. attended to.
3. Alignment with Target Audience Needs:
For Healthcare Professionals: CHAPTER – 6_______________________
o Time-Saving: The system automates
tedious manual checks, saving nurses
time, which can be used for more critical
tasks. PRELIMINARY CONCLUSION AND FUTURE
SCOPE
o Improved Patient Safety: Automated
alerts and real-time monitoring reduce
the chances of overlooking critical
changes in a patient's condition, leading A.FUTURE SCOPE OF IMPROVEMENT
to improved safety.
o Reduced Stress and Fatigue: Nurses
The IV Fluid Monitoring and Alerting System is an
experience less mental load with fewer
essential tool in healthcare, designed to improve the
manual tasks, potentially decreasing
efficiency and safety of intravenous fluid
burnout and improving job satisfaction.
administration. Future improvements can target both
hardware and software enhancements, user Adaptive Alerts:
experience, and integration with healthcare systems.
Below are the potential areas for future development: Use AI to adjust the sensitivity and frequency
of alerts based on patient-specific data to
1. Enhanced Sensor Technology reduce alarm fatigue.
Improved Accuracy: Multi-modal Alerts:
Advanced flow sensors and AI-based Incorporate visual, auditory, and tactile alerts
algorithms for real-time monitoring of fluid to ensure caregivers are promptly notified in
rates and volumes with higher precision. different situations.
Multi-parameter Monitoring: Localization:
Integrate additional sensors for monitoring Include precise location tracking of IV systems
temperature, pressure, and air bubbles to for large hospital settings.
enhance patient safety.
5. Automation and Control
Non-invasive Monitoring:
Autonomous Adjustment:
Develop non-invasive sensors for more
convenient and patient-friendly usage. Automate flow rate adjustments in response to
patient vitals or other parameters using closed-
2. AI and Machine Learning Integration loop systems.
Predictive Analytics: Drug Mixing and Delivery:
Use AI to predict IV fluid depletion times based Incorporate automated mixing of medications
on flow rates and patient-specific factors, with IV fluids, improving efficiency and safety.
allowing proactive interventions.
Emergency Shut-off Mechanisms:
Anomaly Detection:
Add features that can autonomously stop the IV
Implement machine learning to detect flow in case of critical abnormalities.
irregularities in fluid flow, such as occlusions,
leaks, or accidental disconnections. 6. Miniaturization and Portability
Integrate with electronic health record (EHR) 7. Data Analytics and Reporting
systems to display IV data in real-time for Long-term Data Storage:
centralized patient management.
Provide options to store and analyze long-term
Wearable Integration: IV data for research and quality improvement.
Sync with wearable devices to provide real-time Customizable Reports:
feedback to nurses or doctors on their mobile
devices or smartwatches. Allow users to generate detailed reports
tailored to clinical requirements.
4. Improved Alarm Systems
8. Cost-Effective Solutions parameters deviate from the prescribed treatment
plan. The system is also able to prevent blood backflow
Affordable Models: into the IV bag. The system has been evaluated in a
Develop low-cost versions for use in developing pilot study and the results have been very promising.
countries without compromising safety and The system was found to be very accurate and reliable
reliability. in monitoring the IV therapy parameters and
generating alerts. The system was also found to be easy
Modular Upgrades: to use and maintain. The system has the potential to
improve patient safety and reduce the workload on
Create systems with modular features to allow healthcare providers. The system can help to ensure
incremental upgrades based on the facility’s that patients receive the correct dosage of IV fluids
budget. and that they are alerted if any problems occur. In
conclusion, the “IoT-Based IV Fluid Monitoring
9. Cybersecurity and Privacy
Controlling and Alerting System” system represents a
Data Security: significant advancement in healthcare technology. By
harnessing the power of IoT, it has the potential to
Ensure robust encryption for patient data enhance patient safety, improve the efficiency of
transmitted via IoT-enabled systems. healthcare delivery, and ultimately contribute to better
health outcomes.
Access Control:
REFERENCES
Implement multi-level authentication to restrict
access to authorized personnel only.
[1] R. Su, Z. Cai, X. Wei, and W. Wang,” Design of an
10. User-Friendly Interface Intelligent Intravenous Infusion Monitoring
Intuitive Design: System Based on IoT,” Journal of Sensors, vol.
2018, Article ID 3961016, 7 pages,
Improve user interfaces for caregivers with 2018.https://doi.org/10.1155/2018/3961016.
easy-to-read displays and simplified controls. [2] P. Kumar, P. Kumar, R. Gupta, and A. Mittal,”
Multi-language Support: Smart IV Drip Monitoring System Using IoT,”
International Journal of Engineering Technology,
Include support for multiple languages for vol. 7, no. 2.26, pp. 33-37, 2018.
global applicability.
[3] K. Kumari and P. Kumar, ”IoT Based IV Drip
11. Environmental Sustainability Monitoring and Alerting System,” International
Journal of Innovative Research in Science,
Recyclable Materials: Engineering and Technology, vol. 8, no. 1, pp. 143-
Use eco-friendly materials for system 146, 2019.
components. [4] M. Abdelhafez, A. Elsheikh, S. Khalil, and W. I.
Khedr,” A Novel Real-time Intravenous Infusion
Energy Efficiency:
Monitoring System for Hospitalized Patients,”
Optimize power usage to reduce environmental Journal of Medical Systems, vol. 42, no. 8, pp. 141-
impact. 149, 2018.
[5] Y. Qiu, Y. Yao, and D. Lin,” Design and
Implementation of an Intravenous Drip
B. CONCLUSION Monitoring System Based on RFID,” Journal of
Medical Systems, vol. 40, no. 11, pp. 237-243, 2016.
The IoT-Based IV Fluid Monitoring Controlling and
[6] E. S. Hassan, S. S. Saad, and N. E. Nassar,” Design
Alerting System has been successfully designed and
implemented. The system is able to accurately and of a Smart Intravenous Infusion Monitoring
reliably monitor the IV therapy parameters in real- System,” International Journal of Advanced
time and generate alerts if any of the monitored Computer Science and Applications, vol. 9, no. 1,
pp. 2026, 2018. [7]J. A. Su are -Canedo, L. Casal-
Garc a, J. R. Casar-Corredera, J. C. Alvarez-
Santos, and J. R. P ere -Blanco,” Smart
Intravenous Drip Control System with Automatic
Notification,” Journal of Medical Systems, vol. 42,
no. 10, pp.1-11, 2018.
[8] N. A. N. A. H. A. Amer, A. R. A. N. H. Abouelhoda,
and A. M. R.H. Ela i , ”Smart Infusion Pump
Control System,” International Journal of
Engineering and Technology, vol. 9, no. 4, pp. 3006-
3011, 2017..
[9] D. Dhanalakshmi and R. Priyadharshini,” Smart
Infusion Pump for Intravenous Drip Monitoring
System Using IoT,” International Journal of Recent
Technology and Engineering, vol.7, no.6S4, pp. 16-
20, 2019.Department of Electronics and
Communication Engineering.
[10] DraganaOros, Marko Penic, Jovan Sulc, Maja
Cavic, Stevan Stankovski, Gordana Ostojic, Olivera
Ivanov, “Smart Intravenous Infusion Dosing
System”, Applied Science 2021, 11, 513.