Manual Rosemount 1056 en 68922
Manual Rosemount 1056 en 68922
00809-0100-3156, Rev AF
November 2023
Rosemount™ 1056
Intelligent Four-Wire Transmitter
Safety information
Your instrument purchase from Emerson is one of the finest available for your particular application. Emerson has
designed and tested these instruments to meet many national and international standards. Experience indicates
that its performance is directly related to the quality of the installation and knowledge of the user in operating and
maintaining the instrument.
WARNING
Precautionary signs
Failure to observe precautionary signs may result in injury or death to personnel or cause damage to equipment.
Follow all warnings, cautions, and instructions marked on or supplied with the product.
WARNING
Follow installation guidelines.
Failure to follow these installation guidelines could result in death or serious injury. If this equipment is used in a
manner not specified by the manufacturer, then the protection it provides against hazards may be impaired.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Install equipment as specified in this document.
Ensure that only qualified personnel perform the installation, operation, and maintenance of the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Follow appropriate local and national codes.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
WARNING
Electrical shock
Installation and servicing of this product may expose personnel to dangerous voltages.
Electrical installation must be in accordance with the National Electrical Code (ANSI/National Fire Protection
Agency-70) and/or any other national or local codes.
The equipment is protected throughout by double insulation.
Disconnect main power, wired to separate power source, before servicing.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being
performed by qualified personnel.
Signal wiring must be rated at least 240 V.
Non-metallic cable strain reliefs do not provide grounding between conduit connections. Use grounding type
bushings and jumper wires.
Make sure that the instrument is connected and properly grounded through a three-wire power source.
Proper use and configuration is the operator's responsibility.
Install all protective equipment covers and safety ground leads after installation.
Operate only with front panel fastened and in place.
Disconnect and lock out power before connecting the power supply.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’
equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict
physical access by unauthorized personnel to protect end users' assets. This is true for all systems used within the
facility.
2
NOTICE
Use only factory documented components for repair. Tampering or unauthorized substitution of parts and
procedures can affect performance and cause unsafe operation of your process.
NOTICE
This product generates, uses, and can radiate radio frequency energy, and this can cause radio communication
interference.
As temporarily permitted by regulation, this device has been tested for compliance within the limits of Class A
computing devices, pursuant to subpart J of part 15 of Federal Communication Commission (FCC) rules, which are
designed to take whatever measures may be required to correct the interference.
3
4
Reference Manual Contents
00809-0100-3156 November 2023
Contents
Chapter 1 Installation.................................................................................................................................7
1.1 Unpack and inspect.....................................................................................................................7
1.2 General installation information............................................................................................... 7
1.3 Mounting diagrams.....................................................................................................................8
Chapter 2 Wiring....................................................................................................................................... 15
2.1 General wiring information......................................................................................................15
2.2 Preparing conduit openings.................................................................................................... 16
2.3 Preparing sensor cable.............................................................................................................17
2.4 Power, output, and sensor connections.................................................................................17
Chapter 3 Navigating the display........................................................................................................... 29
3.1 User interface............................................................................................................................ 29
3.2 Keypad........................................................................................................................................ 29
3.3 Main display............................................................................................................................... 30
3.4 Menu system..............................................................................................................................33
Chapter 4 Start up the transmitter.........................................................................................................35
Chapter 5 Programming the transmitter.............................................................................................. 39
5.1 General programming information........................................................................................ 39
5.2 Changing start-up settings...................................................................................................... 39
5.3 Choose temperature units and automatic/manual temperature compensation............. 41
5.4 Configuring and ranging the current outputs.......................................................................41
5.5 Set security code........................................................................................................................44
5.6 Security access...........................................................................................................................45
5.7 Using hold.................................................................................................................................. 45
5.8 Resetting factory default settings........................................................................................... 46
5.9 Program alarm relays............................................................................................................... 47
Chapter 6 Programming measurements............................................................................................... 55
6.1 Introduction to programming measurements......................................................................55
6.2 Program pH measurement...................................................................................................... 55
6.3 Program oxidation reduction potential (ORP) measurement............................................. 59
6.4 Program contacting conductivity measurement.................................................................. 61
6.5 Program toroidal conductivity measurement....................................................................... 69
6.6 Program chlorine measurement.............................................................................................74
6.7 Program oxygen measurement.............................................................................................. 84
6.8 Program ozone measurement................................................................................................ 90
6.9 Program turbidity measurement............................................................................................ 92
6.10 Program flow measurement................................................................................................. 96
6.11 Program current input measurement..................................................................................98
Chapter 7 Calibration............................................................................................................................. 105
7.1 Calibrate pH sensors...............................................................................................................105
7.2 Calibrate oxidation reduction potential (ORP) sensors...................................................... 109
Rosemount 1056 5
Contents Reference Manual
November 2023 00809-0100-3156
6 Emerson.com
Reference Manual Installation
00809-0100-3156 November 2023
1 Installation
1.1 Unpack and inspect
Procedure
Inspect the shipping container.
• If it is damaged, contact the shipper immediately for further instructions.
• If there is no apparent damage, unpack the container. Ensure all items shown on the
packing list are present. If items are missing, notify Emerson immediately.
1. Install the transmitter with a sun shield or out of direct sunlight and areas with
extreme temperatures.
2. Install the system in an area where vibrations and electromagnetic and radio
frequency interference are minimized or absent.
3. Keep the transmitter and sensor wiring at least 1 ft (0.30 m) from high voltage
conductors. Ensure there is easy access to the transmitter and sample conditioning
system.
4. The transmitter is suitable for panel, pipe, or surface mounting.
Rosemount 1056 7
Installation Reference Manual
November 2023 00809-0100-3156
8 Emerson.com
Reference Manual Installation
00809-0100-3156 November 2023
Rosemount 1056 9
Installation Reference Manual
November 2023 00809-0100-3156
10 Emerson.com
Reference Manual Installation
00809-0100-3156 November 2023
A. Maximum radius
Note
Panel mounting seal integrity (4/4X) for outdoor applications is the end user's
responsibility.
Rosemount 1056 11
Installation Reference Manual
November 2023 00809-0100-3156
12 Emerson.com
Reference Manual Installation
00809-0100-3156 November 2023
Rosemount 1056 13
Installation Reference Manual
November 2023 00809-0100-3156
A. Front panel
B. Panel and pipe mount enclosure
C. Conduit openings
D. 2-in. (51 mm) pipe mount bracket
14 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
2 Wiring
2.1 General wiring information
The transmitter includes removable connectors and slide-out signal input boards for
sensors.The front panel is hinged at the bottom. The panel swings down for easy access to
the wiring locations.
Procedure
1. Wire the sensor leads to the measurement board following the lead locations
marked on the board.
2. Carefully slide the wired board fully into the enclosure slot and take up the excess
sensor cable through the cable gland.
3. Tighten the cable gland nut to secure the cable and ensure a sealed enclosure.
Rosemount 1056 15
Wiring Reference Manual
November 2023 00809-0100-3156
activate when the conductivity is within a user-selectable percentage of the limit. USP*
alarming is available only when a contacting conductivity measurement board is installed.
A. Front panel/keypad
B. Power leads
C. Alarm relay leads
D. Sensor 1 cable
E. 4-20 mA/HART®/Profibus® leads
F. Sensor 2 cable
G. Spare opening
NOTICE
Always use proper cable gland fittings and plugs for wire and cable installations.
Conduit openings accept 0.5 in (13 mm) conduit fittings or PG13.5 cable glands. To keep
the case watertight, block unused openings with Type 4X or IP66 conduit plugs.
NOTICE
Use watertight fittings and hubs that comply with your requirements. Connect the conduit
hub to the conduit before attaching the fitting to the transmitter.
16 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
NOTICE
The AC power switch is shipped in the 230 Vac position. Adjust the switch upwards to the
115 Vac position for 110 Vac to 120 Vac operation.
Rosemount 1056 17
Wiring Reference Manual
November 2023 00809-0100-3156
This power supply automatically detects DC power and accepts 20 Vdc to 30 Vdc inputs.
Four programmable alarm relays are included.
This power supply automatically detects AC line conditions and switches to the proper line
voltage and line frequency.
Four programmable relays are included.
NOTICE
Twisted pairs are required to minimize noise pickup in the flow and current sensor inputs.
For high electromagnetic interference/radio frequency interference (EMI/RFI)
environments, use shielded sensor wire. Emerson also recommends using shielded sensor
wire in other installations.
18 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
Rosemount 1056 19
Wiring Reference Manual
November 2023 00809-0100-3156
Figure 2-6: Alarm relay wiring for Rosemount 1056 switching power supply (03
ordering code)
NO1
COM1 Relay 1
NC1
NO2
COM2 Relay 2
NC2
NO3
COM3 Relay 3
NC3
NO4
COM4 Relay 4
NC4
20 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
WARNING
Electrical shock
Installation and servicing of this product may expose personnel to dangerous
voltages.
Electrical installation must be in accordance with the National Electrical Code
(ANSI/National Fire Protection Agency-70) and/or any other national or local
codes.
The equipment is protected throughout by double insulation.
Disconnect main power, wired to separate power source, before servicing.
Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualified personnel.
Signal wiring must be rated at least 240 V.
Non-metallic cable strain reliefs do not provide grounding between conduit
connections. Use grounding type bushings and jumper wires.
Make sure that the instrument is connected and properly grounded through a
three-wire power source.
Proper use and configuration is the operator's responsibility.
Install all protective equipment covers and safety ground leads after installation.
Operate only with front panel fastened and in place.
Disconnect and lock out power before connecting the power supply.
Keep sensor and output signal wiring separate from loop power wiring. Do not run
sensor and power wiring in the same conduit or close together in a cable tray.
NOTICE
Twisted pairs are required to minimize noise pickup in the flow and current sensor
inputs. For high electromagnetic interference/radio frequency interference (EMI/RFI)
environments, use shielded sensor wire. Emerson also recommends using shielded
sensor wire in other installations.
Figure 2-7: Contacting conductivity signal board and sensor cable leads
Rosemount 1056 21
Wiring Reference Manual
November 2023 00809-0100-3156
Figure 2-8: Toroidal conductivity signal board and sensor cable leads
22 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
Figure 2-10: Chlorine, dissolved oxygen, and ozone signal boards and sensor cable
leads
Rosemount 1056 23
Wiring Reference Manual
November 2023 00809-0100-3156
Figure 2-12: Flow/current input signal board and sensor cable leads
24 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
Figure 2-13: Power wiring for the Rosemount 1056 115/230 Vac power supply (01 order
code)
A. To main board
B. Earth ground
C. Neutral
D. Line
Rosemount 1056 25
Wiring Reference Manual
November 2023 00809-0100-3156
Figure 2-14: Power wiring for the Rosemount 1056 85-265 Vac power supply (03
ordering code)
A. To main board
B. Earth ground
C. Neutral
D. Line
26 Emerson.com
Reference Manual Wiring
00809-0100-3156 November 2023
Rosemount 1056 27
Wiring Reference Manual
November 2023 00809-0100-3156
Figure 2-16: Power wiring for Rosemount 1056 24 Vdc power supply (02 ordering code)
A. To main board
28 Emerson.com
Reference Manual Navigating the display
00809-0100-3156 November 2023
3.2 Keypad
Press MENU to access menus for programming and calibrating the transmitter. There are
four function keys and four selection keys on the keypad.
Function keys
Four top-level menu items appear when you press MENU.
Calibrate Calibrate the attached sensor(s) and analog output(s).
Hold Suspend analog output(s).
Program Program outputs, measurement, temperature, security, and reset.
Display Program display format, language, warnings, and contrast.
Press MENU to display the main menu screen. Press MENU followed by EXIT to display the
main display.
Press DIAG to display active Faults and Warnings and detailed transmitter information and
sensor diagnostics, including:
• Faults
• Warnings
Rosemount 1056 29
Navigating the display Reference Manual
November 2023 00809-0100-3156
Selection keys
Surrounding the ENTER key, four selection keys (Up, Down, Right, and Left) move the
cursor to all areas of the screen while using the menus.
Use selection keys to:
• Select items on the menu screens.
• Scroll up and down the menu lists.
• Enter or edit numeric values.
• Move the cursor to the right or left.
• Select measurement units during operation.
30 Emerson.com
Reference Manual Navigating the display
00809-0100-3156 November 2023
Process measurements
Two process variables are displayed if two signal boards are installed. One process variable
and process temperature are displayed if one signal board is installed with one sensor.
The upper display area shows the Sensor 1 process reading. The center display area shows
the Sensor 2 process reading. For dual conductivity, you can assign the upper and center
display areas to different process variables as follows:
Process variables for upper display Process variables for center display
Measure 1 Measure 1
% Reject Measure 2
% Pass % Reject
Ratio % Pass
Ratio
Blank
For single input configurations, the upper display area shows the live process variable, and
you can assign the center display area to Temperature or blank.
Secondary values
Up to four secondary values are shown in display quadrants at the bottom half of the
screen. You can program all four secondary value positions to any displayable parameter
available.
Possible secondary values include:
• Slope 1
• Reference off 1
• GI impedance 1
• Reference impedance 1
• Raw
• mV input
• Temperature 1
• Manual temperature 1
• Manual temperature 2
• Output 1 mA
• Output 2 mA
• Output 1%
• Output 2%
• Measure 1
• Relay 1(1)
• Relay 2(1)
• Relay 3(1)
• Relay 4(1)
Rosemount 1056 31
Navigating the display Reference Manual
November 2023 00809-0100-3156
• Blank
4. The Sensor 1 process value is highlighted in reverse video. Press the selection keys
to navigate down to the screen sections that you wish to program. Press ENTER.
5. Choose the desired display parameter or diagnostic for each of the four display
sections in the lower screen.
6. Continue to navigate and program all desired screen selections. Press MENU and
EXIT. The screen returns to the main display.
For single sensor configurations, the default display shows the live process measurement
in the upper display area and temperature in the center display area. You can elect to
disable the display of temperature in the center display area using the Main Format
function. For dual sensor configurations, the default display shows Sensor 1 live process
measurement in the upper display area and Sensor 2 live process measurement in the
center display area.
32 Emerson.com
Reference Manual Navigating the display
00809-0100-3156 November 2023
Rosemount 1056 33
Navigating the display Reference Manual
November 2023 00809-0100-3156
34 Emerson.com
Reference Manual Start up the transmitter
00809-0100-3156 November 2023
WARNING
Electrical shock
Installation and servicing of this product may expose personnel to dangerous
voltages.
Electrical installation must be in accordance with the National Electrical Code
(ANSI/National Fire Protection Agency-70) and/or any other national or local
codes.
The equipment is protected throughout by double insulation.
Disconnect main power, wired to separate power source, before servicing.
Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualified personnel.
Signal wiring must be rated at least 240 V.
Non-metallic cable strain reliefs do not provide grounding between conduit
connections. Use grounding type bushings and jumper wires.
Make sure that the instrument is connected and properly grounded through a
three-wire power source.
Proper use and configuration is the operator's responsibility.
Install all protective equipment covers and safety ground leads after installation.
Operate only with front panel fastened and in place.
Disconnect and lock out power before connecting the power supply.
When the transmitter is powered up for the first time, Quick Start screens appear.
Quick Start operating tips are as follows:
a. A back-lit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or right of the ENTER
key. To scroll up or down or to increase or decrease the value of a digit, use
the keys above and below the ENTER key. Use the Left or Right keys to move
the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes.
Press EXIT during Quick Start to return the display to the initial startup screen
(Select language).
Rosemount 1056 35
Start up the transmitter Reference Manual
November 2023 00809-0100-3156
After the last step, the main display appears. The outputs are assigned to default
values.
36 Emerson.com
Reference Manual Start up the transmitter
00809-0100-3156 November 2023
Rosemount 1056 37
Start up the transmitter Reference Manual
November 2023 00809-0100-3156
38 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Rosemount 1056 39
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
40 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Procedure
1. From the main menu, go to Program → Temperature.
2. Select automatic or manual temperature compensation
3. Set the manual reference temperature
4. Program temperature units to °F or °C.
See Figure 5-1.
Rosemount 1056 41
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
NOTICE
Always configure the outputs before ranging them.
42 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Procedure
1. Go to Program → Output → Configure.
Rosemount 1056 43
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
Enter a value for 4 mA and 20 mA (or 0 mA and 20 mA) for each output.
Procedure
1. Press MENU.
The Main menu screen appears.
2. Select Program.
3. Select Security.
The Security screen appears.
4. Enter a three digit security code for each of the desired security levels.
The security code takes effect two minutes after the last key stroke.
5. Record the security codes for future access and communication to operators or
technicians as needed.
The display returns to the Security menu screen.
Figure 5-6 displays the security code screens.
Temperature Security
---------------------------------------- Calibration/Hold: 000
Security All: 000
Diagnostic Setup
Ambient AC Power:Unk
Reset Analyzer
Postrequisites
Press EXIT to return to the previous screen. To return to the main display, press MENU and
then EXIT.
44 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Procedure
Enter the three-digit security code for the appropriate security level.
If the entry is correct, the appropriate menu screen appears. If the entry is incorrect, the
Invalid Code screen appears. The Security Code screen reappears after two seconds.
NOTICE
Once on hold, all current outputs remain on hold indefinitely.
Rosemount 1056 45
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
Procedure
1. Press MENU.
The Main Menu screen appears.
2. Select Hold.
The Hold Outputs and Alarms? screen appears.
3. Select Yes to place the transmitter on hold. Select No to take the transmitter out of
hold.
NOTICE
There are no alarm relays with this configuration. Current outputs are included with
all configurations.
NOTICE
Hold will remain on indefinitely until someone disables it.
46 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Program the deadband 0.00 µS/cm Program the change in process value
after the relay deactivates.
Set relay on-time 10 minutes Enter the time in seconds that the relay
is activated.
Rosemount 1056 47
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
Set recovery time 60 seconds Enter time after the relay deactivation
for process recovery.
Procedure
1. Go to Program → Alarms.
Figure 5-10 appears to allow you to configure the alarm relays.
48 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Factory defaults are displayed as they would appear for an installed contacting
conductivity board. USP safety appears only if alarm logic is set to USP. Interval time,
On Time, Recover time, and Hold while active only appear if the alarm is configured
as an interval timer.
Enter the desired value for the process measurement or temperature at which to activate
an alarm event.
Rosemount 1056 49
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
Procedure
Select an alarm assignment.
Select the desired relay logic to activate alarms at a high reading or a low reading. USP only
appears if a contacting conductivity board is installed.
50 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Procedure
1. Hold down the EXIT key for six seconds while in the main display mode to enter the
Expert menus.
The screen prompt Enable Expert Menu appears.
2. Select Yes.
3. Under the Alarm Settings menu, select Normal State.
4. Select the alarm condition that you prefer each time the transmitter powers up.
Rosemount 1056 51
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
Enter the time in seconds that you want the relay to be activated for.
Select whether or not to hold the current outputs for Sensor 1, Sensor 2, or both sensors
while the relay is activated.
52 Emerson.com
Reference Manual Programming the transmitter
00809-0100-3156 November 2023
Rosemount 1056 53
Programming the transmitter Reference Manual
November 2023 00809-0100-3156
54 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
6 Programming measurements
6.1 Introduction to programming measurements
The Rosemount 1056 automatically recognizes each installed measurement board upon
first power-up and each time the transmitter is turned on.
Completing Quick Start screens upon first power-up enables measurements, but you may
have to take additional steps to program the transmitter for the desired measurement
application. This section covers the following programming and configuration functions:
1. Select measurement type or sensor type.
2. Identify the preamp location (Program pH measurement).
3. Enable manual temperature correction and enter a reference temperature.
4. Enable sample temperature correction and enter temperature correction slope.
5. Define measurement display resolution (pH and amperometric).
6. Define measurement display units.
7. Adjust the input filter to control display and output reading variability or noise.
8. Select a measurement range (Program contacting conductivity measurement and
Program toroidal conductivity measurement).
9. Enter a cell constant for a contacting or toroidal sensor (Program contacting
conductivity measurement and Program toroidal conductivity measurement).
10. Enter a temperature element/RTD offset or temperature slope (Program contacting
conductivity measurement and Program toroidal conductivity measurement).
11. Create an application-specific concentration curve.
12. Enable automatic pH correction for free chlorine measurement (Program chlorine
measurement.
To fully configure the transmitter for each installed measurement board, you may use the
following:
1. Reset Analyzer function to reset factory defaults and configure the measurement
board to the desired measurement. Follow the Reset Analyzer menu to reconfigure
the transmitter to display new measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items.
Rosemount 1056 55
Programming measurements Reference Manual
November 2023 00809-0100-3156
Figure 6-1: Flow diagram for pH and oxidation reduction potential (ORP)
programming
To change any setting, scroll to the desired item and press ENTER.
56 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value (pH) is displayed in bold type. Refer to Figure 6-1 to select a
measurement type.
The default value (Analyzer) is displayed in bold type. Refer to Figure 6-1 to select a preamp
location.
Rosemount 1056 57
Programming measurements Reference Manual
November 2023 00809-0100-3156
The default value (Off) is displayed in bold type. Refer to Figure 6-1 to select a solution
temperature correction algorithm.
The default value (-0.032 pH/°C) is displayed in bold type. Refer to Figure 6-1 to enter a
custom coefficient.
The default value (0.01 pH) is displayed in bold type. Refer to Figure 6-1 to select a
resolution.
58 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value (02 seconds) is displayed in bold type. Refer to Figure 6-15 to enter the
input filter value.
The default value (Low) is displayed in bold type. Refer to Figure 6-1 to select the reference
impedance.
Rosemount 1056 59
Programming measurements Reference Manual
November 2023 00809-0100-3156
The default value is displayed in bold type. Refer to Figure 6-1 to select the measurement
type.
The default value is displayed in bold type. Refer to Figure 6-1 to select the preamp
location.
60 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value is displayed in bold type. Refer to Figure 6-1 to enter the input filter value.
The default value is displayed in bold type. Refer to Figure 6-1 to select the reference
impedance.
Rosemount 1056 61
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
62 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
4. To change any settings, scroll to the desired item and press ENTER.
The default value is displayed in bold type. Refer to Figure 6-15 to select the sensor type.
Rosemount 1056 63
Programming measurements Reference Manual
November 2023 00809-0100-3156
The default value is displayed in bold type. Refer to Figure 6-15 to select a measurement
type.
64 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value is displayed in bold type. Refer to Figure 6-15 to enter the cell constant.
The default value is displayed in bold type. Refer to Figure 6-15 to set the RTD offset.
The default value is displayed in bold type. Refer to Figure 6-15 to enter the RTD slope.
Rosemount 1056 65
Programming measurements Reference Manual
November 2023 00809-0100-3156
The default value is displayed in bold type. Refer to Figure 6-15 to select a temperature
compensation.
The default value is displayed in bold type. Refer to Figure 6-15 to enter the slope.
The default value is displayed in bold type. Refer to Figure 6-15 to complete this function.
66 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value (02 seconds) is displayed in bold type. Refer to Figure 6-15 to enter the
input filter value.
Procedure
Enter the custom curve data. Press ENTER.
The display confirms the determination of a custom curve fit to the entered data by
displaying Figure 6-28.
If the custom curve fit is not completed or is unsuccessful, Figure 6-29 appears. The
transmitter returns to the screen shown in Figure 6-27.
Rosemount 1056 67
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
If you selected 4-electrode for the sensor type in the Quick Start menus, enter a cell
constant and a calibration factor using the transmitter keypad.
The cell constant is needed to convert measured conductance to conductivity, as displayed
on the transmitter screen. The calibration factor entry is needed to increase the accuracy
of the live conductivity readings, especially at low conductivity readings below 20 µS/cm.
Both the cell constant and the calibration factor are printed on the tag attached to the
four-electrode sensor cable.
68 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
Rosemount 1056 69
Programming measurements Reference Manual
November 2023 00809-0100-3156
5. To change any setting, scroll to the desired item and press ENTER.
The default value is displayed in bold type. Refer to Figure 6-31 to select a sensor model.
70 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value is displayed in bold type. Refer to Figure 6-31 to select the measurement
type.
Rosemount 1056 71
Programming measurements Reference Manual
November 2023 00809-0100-3156
The default value is displayed in bold type. Refer to Figure 6-31 to enter the cell constant.
The default value is displayed in bold type. Refer to Figure 6-31 to enter the slope.
72 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
The default value is displayed in bold type. Refer to Figure 6-31 to enter the reference
temperature.
The default value is displayed in bold type. Refer to Figure 6-31 to enter the input filter
time.
Rosemount 1056 73
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
Enter the custom curve data and press ENTER.
The display confirms the determination of a custom curve fit to the entered data by
displaying Figure 6-42.
If the custom curve fit is not completed or is unsuccessful, Figure 6-43 appears and the
transmitter returns to the screen shown in Figure 6-41.
74 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Figure 6-44 is a detailed flow diagram to guide you through all basic programming and
configuration functions.
To configure the measurement board for free chlorine:
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to free chlorine. Press ENTER.
Figure 6-45 appears; factory default settings are shown.
5. To change any setting, scroll to the desired item and press ENTER.
Rosemount 1056 75
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
1. Select Free Chlorine.
2. Press ENTER.
Procedure
Refer to Figure 6-44 to select a measurement unit.
76 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
Refer to Figure 6-44 to enter an input filter value.
Select pH correction
Figure 6-49 displays the screen from which you can select live/continuous pH correction or
manual pH correction.
Procedure
Refer to Figure 6-44 to select the pH correction.
Procedure
Refer to Figure 6-44 to enter a pH value manually.
Rosemount 1056 77
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
Refer to Figure 6-44 to select a display resolution.
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER.
Figure 6-52 will appears; factory default settings are shown.
5. To change any setting, scroll to the desired item and press ENTER.
78 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
Refer to Figure 6-44 to select a measurement type.
Procedure
Refer to Figure 6-44 to select measurement units.
Rosemount 1056 79
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
Refer to Figure 6-44 to enter the input filter value.
Procedure
Refer to Figure 6-44 to select the display resolution.
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER.
Figure 6-57 appears; factory default settings are shown.
5. To change any settings, scroll to the desired item and press ENTER.
80 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
Refer to Figure 6-44 to select the monochloramine measurement type.
Procedure
Refer to Figure 6-44 to select measurement units.
Rosemount 1056 81
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
Refer to Figure 6-44 to enter the input filter value.
Procedure
Refer to Figure 6-44 to select the display resolution.
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to chlorine. Press ENTER.
Figure 6-62 appears; factory default settings are shown.
82 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Rosemount 1056 83
Programming measurements Reference Manual
November 2023 00809-0100-3156
84 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to oxygen. Press ENTER.
Figure 6-68 appears; factory default settings are shown.
Rosemount 1056 85
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
1. From the Configure Oxygen screen, select Units.
Figure 6-70 appears.
Procedure
1. From the Configure Oxygen screen, select Units.
The Units screen appears.
86 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Rosemount 1056 87
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
1. From the Configure Oxygen screen, select Pressure Units.
Figure 6-74 appears.
88 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Rosemount 1056 89
Programming measurements Reference Manual
November 2023 00809-0100-3156
Figure 6-77 is a detailed flow diagram for ozone programming to guide you through all
basic programming and configuration functions.
To configure the ozone measurement board:
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to ozone. Press ENTER.
Figure 6-78 appears; factory default settings are shown.
5. To program any displayed function, scroll to the desired item and press ENTER.
Note
The transmitter automatically detects ozone measurement boards. You don't need
to select a measurement type.
90 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Rosemount 1056 91
Programming measurements Reference Manual
November 2023 00809-0100-3156
Figure 6-82 is a detailed flow diagram for turbidity programming; use it to guide you
through all basic programming and configuration functions.
To configure the turbidity measurement board:
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
92 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
Refer to Figure 6-82 to complete this function.
Procedure
Refer to Figure 6-82 to complete this function.
If you select the total suspended solids (TSS) calculation, the following screen will be
displayed.
Rosemount 1056 93
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
1. Refer to Figure 6-82 to complete this function.
Note
Based on user-entered NTU data, calculating TSS as a straight line curve could cause
TSS to go below zero. The following screen lets you know that TSS will become zero
below a certain NTU value.
Figure 6-84 shows the potential for calculated TSS to go below zero.
94 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
If you see this screen, you need to re-enter NTU and TSS data.
Procedure
Refer to Figure 6-82 to complete this function.
Rosemount 1056 95
Programming measurements Reference Manual
November 2023 00809-0100-3156
Procedure
Refer to Figure 6-82 to complete this function.
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to flow. Press ENTER.
96 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Rosemount 1056 97
Programming measurements Reference Manual
November 2023 00809-0100-3156
98 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
4. Select Sensor 1 or Sensor 2 corresponding to the current input. Press ENTER.
The factory default is Pulse Flow, not mA Input.
5. To enable the current input functionality, override the factory default and select mA
Input.
Figure 6-90 appears. The default measurement type is displayed in bold font.
Rosemount 1056 99
Programming measurements Reference Manual
November 2023 00809-0100-3156
100 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
— μg/L
— mg/L
— g/L
— NTU
— FTU
— FNU
— ft/sec
— m/sec
— none
102 Emerson.com
Reference Manual Programming measurements
00809-0100-3156 November 2023
104 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
7 Calibration
Calibration is the process of adjusting or standardizing the transmitter to a lab test or
a calibrated laboratory instrument or standardizing to some known reference, such as a
commercial buffer.
The transmitter's auto-recognition feature will enable the appropriate calibration screens
to allow calibration for any single sensor configuration or dual sensor configuration.
Completing the Quick Start when first powering up the transmitter enables live
measurement, but does not ensure accurate readings in the lab or in process. Calibrate
with each attached sensor to ensure accurate, repeatable readings.
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
Figure 7-5 through Figure 7-7 may appear if the auto calibration is unsuccessful.
106 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
108 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to ORP. Press ENTER.
110 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
the temperature correction used by the Rosemount 1056. To avoid errors, turn off
temperature correction in both the transmitter and the standard instrument.
2. Place the sensor in a solution of known conductivity and make the transmitter
reading match the conductivity of the standard solution. Use this method if you can
easily remove the sensor from the process piping and have a standard available.
Be careful using standard solutions with conductivity less than 100 μS/cm. Low
conductivity standards are highly susceptible to atmospheric contamination. Avoid
calibrating sensors with 0.01/cm cell constants against conductivity standards with
conductivity greater than 100 μS/cm. The resistance of these solutions may be too
low for an accurate measurement.
3. To calibrate a 0.01/cm sensor, check it against a standard instrument and another
0.01/cm sensor while both sensors are measuring water with conductivity between
5 and 10 μS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide,
saturate the sample with air before taking measurements.
To ensure adequate flow past the sensor during calibration, take the sample
downstream from the sensor. For best results, use a flow-through standard cell.
If the process temperature varies greatly from the ambient temperature, keep the
connecting lines short and insulate the flow cell.
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to conductivity. Press ENTER.
4. To calibrate conductivity or temperature, scroll to the desired item and press ENTER.
5. Enter a cell constant only if you are installing the sensor for the first time, replacing
the probe, or troubleshooting.
This procedure sets up the transmitter for the probe type connected to it. Each type
of probe has a specific cell constant.
Prerequisites
CAUTION
Procedure
1. Zero the transmitter by electrically connecting the conductivity probe as it will
actually be used and placing the measuring portion of the probe in the air.
2. Select Zero Cal from the Conductivity Calibration screen.
112 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Prerequisites
• Clean the probe.
• If necessary, check and standardize the temperature reading using a calibrated
thermometer.
Procedure
1. Submerge the probe in a sample of known conductivity.
2. If necessary, adjust the displayed value to correspond to the conductivity value of
the sample.
3. Turn temperature correction off and use the conductivity of the standard.
If in process calibration is successful, Figure 7-14 appears and the transmitter returns to
the Conductivity Calibration screen.
Procedure
1. Submerge the conductivity probe in a bath.
2. Measure the conductivity of a grab sample of the same bath water with a separate
laboratory instrument.
3. Adjust the transmitter reading to match the conductivity reading of the lab
instrument.
114 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
5. Press ENTER.
The display shows the live value measured by the sensor.
If the meter calibration is successful, the screen returns to the Conductivity Calibration
screen.
If meter calibration is unsuccessful, Figure 7-16 appears.
6. Select Conductivity.
The following screen appears.
Postrequisites
The following sections show the initial display screen that appears for each calibration
routine. Use the live screen prompts to complete calibration.
116 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Setting up the cell constant sets up the transmitter for the probe type connected to it. Each
type of probe has a specific cell constant.
For toroidal conductivity sensors, the default value is 3.00000/cm.
Procedure
1. Electrically connect the conductivity probe as it will actually be used and place the
measuring portion of the probe in air.
2. From the Conductivity Calibration screen, select Zero Cal.
The following screen appears.
Prerequisites
Clean the probe and check and standardize the temperature reading before performing
this procedure.
Procedure
1. Submerge the probe in a sample of known conductivity.
2. Select In Process Cal from the Conductivity Calibration screen.
The following screen appears.
118 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
CAUTION
• Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least
24 hours.
The purpose of in process calibration is to establish the slope of the calibration curve.
Because stable chlorine standards do not exist, calibrate the sensor against a test run on a
grab sample of the in process liquid. Several manufacturers offer portable test kits for this
purpose. Observe the following precautions when taking and testing the grab sample:
• Take the grab sample from a point as close to the sensor as possible. Be sure that
taking the sample does not alter the flow of the sample to the sensor. It is best to install
the sample tap just downstream from the sensor.
• Chlorine solutions are unstable. Run the test immediately after taking the sample. Try
to calibrate the sensor when the chlorine concentration is at the upper end of the
normal operating range.
To calibrate free chlorine:
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to free chlorine. Press ENTER.
4. Select Free Chlorine. Press ENTER.
Related information
Zero sensor
Calibrate sensor in process
120 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to total chlorine. Press ENTER.
4. Select Total Chlorine. Press ENTER.
5. To calibrate total chlorine or temperature, scroll to the desired item and press
ENTER.
Related information
Zero sensor
Calibrate sensor in process
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to pH-independent free chlorine. Press
ENTER.
4. Select pH Ind. Free Cl. Press ENTER.
Related information
Zero sensor
Calibrate sensor in process
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
122 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Related information
Zero sensor
Calibrate sensor in process
Zero calibration is necessary because oxygen sensors, even when no oxygen is present
in the sample, generate a small current called the residual current. The transmitter
compensates for the residual current by subtracting it from the measured current before
converting the result to a dissolved oxygen value. Zero new sensors before placing them
in service and whenever you replace the electrolyte solution. The recommended zero
standard is five percent sodium sulfide in water, although you can also use oxygen-free
nitrogen. The Rosemount 499ATrDO sensor, used to determine trace (ppb) oxygen levels,
has very low residual current and does not normally require zeroing. The residual current
in the Rosemount 499ATrDO sensor is equivalent to less than 0.5 ppb oxygen.
The purpose of in process calibration is to establish the slope of the calibration curve.
Because the solubility of atmospheric oxygen in water as a function of temperature
and barometric pressure is well known, the natural choice for a full-scale standard is
air-saturated water. However, air-saturated water is difficult to prepare and use, so the
universal practice is to use air for calibration. From the point of view of the oxygen sensor,
air and air-saturated water are identical. The equivalence comes about because the sensor
really measures the chemical potential of oxygen. Chemical potential is the force that
causes oxygen molecules to diffuse from the sample into the sensor where they can be
measured. It is also the force that causes oxygen molecules in air to dissolve until the
water is saturated with oxygen. Once the water is saturated, the chemical potential of
oxygen in the two phases (air and water) is the same.
Oxygen sensors generate a current directly proportional to the rate at which oxygen
molecules diffuse through a membrane stretched over the end of the sensor. The diffusion
124 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
rate depends on the difference in chemical potential between oxygen in the sensor and
oxygen in the sample. An electrochemical reaction, which destroys any oxygen molecules
entering the sensor, keeps the concentration (and the chemical potential) of the oxygen
inside the sensor equal to zero. Therefore, the chemical potential of oxygen in the sample
alone determines the diffusion rate and the sensor current.
When you calibrate the sensor, the chemical potential of oxygen in the standard
determines the sensor current. Whether the sensor is calibrated in air or air-saturated
water is immaterial. The chemical potential of oxygen is the same in either phase.
Normally, to make calculating solubility in common units (like ppm dissolved oxygen)
simpler, it is convenient to use water-saturated air for calibration. Automatic air calibration
is standard. Simply expose the sensor to water-saturated air. The transmitter measures
the sensor current. When the current is stable, the transmitter stores the current and
measures the temperature using a temperature element inside the oxygen sensor. Enter
the barometric pressure.
From the temperature, the transmitter calculates the saturation vapor pressure of water.
Next, it calculates the pressure of dry air by subtracting the vapor pressure from the
barometric pressure. Using the fact that dry air always contains at least 20.95 percent
oxygen, the transmitter calculates the partial pressure of oxygen. Once the transmitter
knows the partial pressure of oxygen, it uses the Bunsen coefficient to calculate the
equilibrium solubility of atmospheric oxygen in water at the prevailing temperature. At
77.0 °F (25 °C) and 29.92 in (760 mm) Hg, the equilibrium solubility is 8.24 ppm. Often, it is
too difficult or messy to remove the sensor from the process liquid for calibration. In this
case, you can calibrate the sensor against a measurement made with a portable laboratory
instrument. The laboratory instrument typically uses a membrane-covered amperometric
sensor that has been calibrated against water-saturated air.
To calibrate oxygen:
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to oxygen. Press ENTER.
4. Select Oxygen. Press ENTER.
5. To calibrate oxygen or temperature, scroll to the desired item and press ENTER.
Related information
Zero sensor
Calibrate sensor in air
Calibrate sensor in process
126 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Zero calibration is necessary because ozone sensors, even when no ozone is in the sample,
generate a small current called the residual or zero current. The transmitter compensates
for the residual current by subtracting it from the measured current before converting the
result to an ozone value. Zero new sensors before placing them in service and whenever
you replace the electrolyte solution. The best zero standard is deionized water.
To calibrate ozone:
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to ozone. Press ENTER.
5. To calibrate ozone or temperature, scroll to the desired item and press ENTER.
Related information
Zero sensor
Calibrate sensor in process
Prerequisites
Run the sensor in the zero solution for at least two hours before zeroing.
Procedure
From the Calibration screen, select Zero Cal.
128 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
If the in process calibration is successful, the transmitter returns to the Calibration screen.
If calibration is unsuccessful, Figure 7-26 may appear.
correction. If you turn temperature correction off, the transmitter uses the manual
temperature you enter in all temperature correction calculations.
To calibrate temperature:
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to the desired measurement. Press
ENTER.
4. Select Temperature. Press ENTER.
6. If the temperature offset is greater than 5 °C from the default value, Figure 7-27
may appear.
Related information
Enter temperature compensation
130 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to turbidity. Press ENTER.
4. Select Turbidity. Press ENTER.
Procedure
1. Immerse the sensor in filtered water having very low turbidity and measure the
sensor output.
2. Increase the turbidity of the filtered water by a known amount (typically 20 NTU) and
measure the sensor output again.
3. On the transmitter, select Slope.
The transmitter takes the two measurements, applies a linear correction (if necessary),
and calculates the sensitivity. Sensitivity is the sensor output (in mV) divided by turbidity.
A typical new sensor has a sensitivity of about 10 mv. As the sensor ages, the sensitivity
decreases.Figure 7-32 illustrates how turbidity two-point calibration works.
132 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Before beginning the calibration, the transmitter does a dark current measurement.
Dark current is the signal generated by the detector when no light is falling on it. The
transmitter subtracts the dark current from the raw scattered light signal and converts the
result to turbidity. In highly filtered samples, which scatter little light, the dark current can
be a substantial amount of the signal the sensor generates.
Figure 7-33 appears if the slope calibration is successful.
Procedure
1. On the transmitter, select Standard from the Calibration screen.
2. Put the sensor in the standard and press ENTER on the transmitter.
Before beginning the calibration, the transmitter does a dark current measurement.
Dark current is the signal generated by the sensor even when no light is falling on
it. The transmitter subtracts the dark current from the raw scattered light signal and
converts the result to turbidity. In highly filtered samples, which scatter little light,
the dark current can be a substantial amount of the signal the sensor generates.
Figure 7-36 appears if the standard calibration is successful.
134 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
Procedure
1. On the transmitter's Calibration screen, select Grab.
Procedure
1. Press MENU.
2. Select Calibrate. Press ENTER.
3. Select Sensor 1 or Sensor 2 corresponding to Flow. Press ENTER.
136 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
2. Enter the known K factor provided with the flow sensor specifications.
The transmitter returns to the Pulse Flow Calibration screen, showing the updated K
factor.
138 Emerson.com
Reference Manual Calibration
00809-0100-3156 November 2023
2. To suspend the totalizer, select Stop. To re-enable the totalizer, select Resume. To
reset the totalizer count to zero, select Reset.
The live totalizer count result is displayed at the bottom of the Totalizer Control
screen.
140 Emerson.com
Reference Manual
00809-0100-3156 November 2023