Data_Cubis_W--2025-e
Data_Cubis_W--2025-e
Laboratory balances
MODULAR
DESIGN
General Specifications
Power supply 100–240 V~, –15 %/+10 %, 50–60 Hz, 1.0 A
Input voltage 15 VDC, ± 5 %
Power consumption 7W (max.)
Ambient temperature Operation +5°C to +40°C
Highest relative humidity 80 % for temperatures up to 31°C, decreasing linearly to 50 % relative humidity for 40°C
Safety of electrical equipment According to EN 61010-1:2001: Safety requirements for electrical equipment for measurement,
control, and laboratory use – Part 1: General requirements
Electromagnetic compatibility According to EN 61326-1:2006: Electrical equipment for measurement, control, and laboratory use –
EMC requirements – Part 1: General requirements
Defined immunity to interference Suitable for use in industrial areas
Interference emission Class B (suitable for use in residential areas and areas that are connected to a low voltage network
that also supplies residential buildings)
Cubis® Display and Control Units
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
2
Microbalances 0.001 mg
Model 6.6S 6.6S (with DF filter 3.6P
draft shield)
Readability mg 0.001 0.001 0.001 | 0.002 | 0.005
Weighing capacity g 6.1 6.1 1.1 | 2.1 | 3.1
Tare range (subtractive) g – 6.1 – 6.1 – 3.1
Repeatability <±mg 0.001 0.001 0.003 | 0.004 | 0.005
Linearity <±mg 0.004 0.004 0.004
Corner load (test load [g]) mg 0.004 (2) 0.004 (2)** 0.005 (1)
Optimal starting point of the mg 0.82 – 0.82
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s <5 <5 <5
Typical measurement time s <8 <8 <8
External standard calibration value g 5 (E2) 5 (E2) 3 (E2)
(min. accuracy class)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size d mm 30 50 30
Weighing chamber height mm 70 15 70
Protection Protected against dust and water
Semi-microbalances 0.01 mg
Model 225S 225P 125P
Readability mg 0.01 0.01 | 0.02 | 0.05 0.01 | 0.1
Weighing capacity g 220 60 | 120 | 220 60 | 120
Tare range (subtractive) g – 220 – 220 – 120
Repeatability <±mg 0…60 g: 0.015 0…60 g: 0.015 0…60 g: 0.015
60…220 g: 0.025 60…220 g: 0.04 60…120 g: 0.06
Linearity <±mg 0.1 0.15 0.15
Corner load (test load [g]) mg 0.15 (100) 0.2 (100) 0.15 (50)
Optimal starting point of the mg 8.2 8.2 8.2
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s <2 <2 <2
Typical measurement time s <6 <6 <6
External standard calibration value g 200 (E2) 200 (E2) 100 (E2)
(min. accuracy class)
Display result s 0.2 – 0.4 0.2 – 0.4 0.2 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 85 + 85 85 + 85 85 + 85
Weighing chamber height (draft shield DU) mm 261 261 261
Protection Protected against dust and water
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
** = Measured using a standard pan
3
Analytical Balances 0.1 mg
Model 524S 524P 324S 324P 224S 124S
Readability mg 0.1 0.1 | 0.2 | 0.5 0.1 0.1 | 0.2 | 0.5 0.1 0.1
Weighing capacity g 520 120 | 240 | 520 320 80 | 160 | 320 220 120
Tare range (subtractive) g – 520 – 520 – 320 – 320 – 220 – 120
Repeatability <±mg 0.1 0.15 | 0.2 | 0.4 0.1 0.1 | 0.2 | 0.4 0.07 0.1
Linearity <±mg 0.4 0.5 0.3 0.5 0.2 0.2
Corner load (test load [g]) mg 0.3 (200) 0.4 (200) 0.3 (200) 0.4 (200) 0.2 (100) 0.2 (50)
Optimal starting point of the mg 82 82 82 82 82 82
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1 1 1 1
Typical stabilization time s <1 <1 <1 <1 <1 <1
Typical measurement time s <3 <3 <3 <3 <3 <3
External standard calibration value g 500 500 200+100 200+100 200 (E2) 100 (E2)
(min. accuracy class) (E2) (E2)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 85 + 85 85 + 85 85 + 85 85 + 85 85 + 85 85 + 85
Weighing chamber height (draft shield DU) mm 261 261 261 261 261 261
Protection IP54 in accordance with IEC 60529
Precision Balances
Model 5203S 5203P 3203S 2203S 2203P 1203S
Readability mg 1 1|2|5 1 1 1 | 10 1
Weighing capacity g 5,200 1,200 | 2,400 | 3,200 2,200 1,010 | 2,200 1,200
5,200
Tare range (subtractive) g - 5,200 - 5,200 – 3,200 – 2,200 – 2,200 – 1,200
Repeatability <±mg 1 1 1 1 1|6 0.7
Linearity <±mg 5 5 5 3 5 2
Corner load (test load [g]) mg 2 (2,000) 2 (2,000) 2 (1,000) 2 (1,000) 3 (1,000) 2 (500)
Optimal starting point of the g 0.82 0.82 0.82 0.82 0.82 0.82
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1 1 1 1.5
Typical stabilization time s <1 <1 <1 <1 <1 <1
Typical measurement time s <2 <2 <2 < 1.5 < 1.5 < 1.5
External standard calibration value g 5,000 5,000 2,000 2,000 (E2) 1,000 (E2) 1,000 (E2)
(min. accuracy class)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 140 + 140 140 + 140 140 + 140 140 + 140 140 + 140 140 + 140
Weighing chamber height (draft shield DE) mm 172 172 172 172 172 172
Protection Protected against dust and water
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
4
Precision Balances
Model 623S 623P 323S
Readability mg 1 1|2|5 1
Weighing capacity g 620 150 | 300 | 620 320
Tare range (subtractive) g – 620 – 620 – 320
Repeatability <±mg 0.7 1|2|4 0.7
Linearity <±mg 2 5 2
Corner load (test load [g]) mg 2 (200) 4 (200) 2 (200)
Optimal starting point of the g 0.82 0.82 0.82
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2
Typical stabilization time s < 0.8 < 0.8 < 0.8
Typical measurement time s <1 <1 <1
External standard calibration value g 500 (E2) 500 (F1) 200 (E2)
(min. accuracy class)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 140 + 140 140 + 140 140 + 140
Weighing chamber height (draft shield DE) mm 172 172 172
Protection Protected against dust and water
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
5
Precision Balances
Model 6202S 6202P 5202S 4202S
Readability mg 10 10 | 20 | 50 10 10
Weighing capacity g 6,200 1,500 | 3,000 | 6,200 5,200 4,200
Tare range (subtractive) g – 6,200 – 6,200 - 5,200 – 4,200
Repeatability <±mg 7 7 | 20 | 40 6 7
Linearity <±mg 20 50 10 20
Corner load (test load [g]) mg 20 (2,000) 50 (2,000) 10 (2,000) 30 (2,000)
Optimal starting point of the g 8.2 8.2 8.2 8.2
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2 2
Typical stabilization time s 1 1 0.8 0.8
Typical measurement time s < 1.5 < 1.5 <1 <1
External standard calibration value kg 5 (E2) 5 (F1) 5 2 (E2)
(min. accuracy class)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 206 + 206 206 + 206 140 + 140 206 + 206
Protection IP54 in accordance with IEC 60529
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
6
Precision High Capacity Balances
Model 70201S 36201S 36201P 20201S
Readability mg 100 100 100 | 1,000 100
Weighing capacity g 70,200 36,200 10,200 | 36,200 20,200
Tare range (subtractive) g – 70,200 – 36,200 - 36,200 – 20,200
Repeatability <±mg 100 100 100 | 500 100
Linearity <±mg 500 200 200 200
Corner load (test load [g]) mg 500 (20,000) 300 (10,000) 300 (10,000) 300 (5,000)
Optimal starting point of the g 82 82 82 82
operating range*
Sensitivity drift between +10 to +30°C ±ppm/K 3 2 2 2
Typical measurement time s 1.5 1.5 1.5 1.5
External standard calibration value kg 20 (F1) 10 (F1) 10 (F1) 10 (F1)
(min. accuracy class)
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 400 + 300 400 + 300 400 + 300 400 + 300
Protection IP54 in accordance with IEC 60529
* = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
7
Verified Models with EC Type Approval Certificate: Micro- and Ultramicrobalances
Model 6.6S-xCE 2.7S-xCE 3.6P-xCE
Accuracy class* mg k k k
For verified models: EC type approval certificate D09-09-015, Type: MSX
Scale interval d* mg 0.001 0.0001 0.001
Weighing capacity max* g 6.1 2.1 3.1
Calibration value e* mg 1 1 1
Min. load min* mg 0.1 0.01 0.1
Tare equalization range (subtractive) g < 100 % from max. weighing capacity
Application range according to DIR* g 0.001 – 6.1 0.001 – 2.1 0.001 – 3.1
Optimal starting point of the operating range** mg 0.82 0.082 0.82
Typical stabilization time s <5 <7 <5
Typical measurement time s <8 < 10 <8
External standard calibration value g 5 2 3
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5…+40°C | Without “isoCAL” function: +15 … +25°C
Display result s By selection of 1 of 4 optimized filter levels
(depending on the set filter level)
Weighing pan size d mm 30 20 30
Weighing chamber height mm 70 70 70
(draft shield DM)
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
8
Verified Models with EC Type Approval Certificate: Analytical Balances 0.1 mg
Model 524S-xCE 524P-xCE 324S-xCE 224S-xCE 324P-xCE 124S-xCE
Accuracy class* k k k k k k
For verified models: EC type approval certificate D09-09-015, Type: MSX
Scale interval d* mg 0.1 0.1 | 0.2 | 0.5 0.1 0.1 0.1 | 0.2 | 0.5 0.1
Weighing capacity max* g 520 120 | 240 | 520 320 220 80 | 160 | 320 120
Calibration value e* mg 1 1 1 1 1 1
Min. load min* mg 10 10 10 10 10 10
Tare equalization range (subtractive) g < 100 % from max. weighing capacity
Application range according to DIR* g 0.01–520 0.01–520 0.01–320 0.01–220 0.01–320 0.01–120
Optimal starting point of the mg 82 82 82 82 82 82
operating range**
Typical stabilization time s <1 <1 <1 <1 <1 <1
Typical measurement time s <3 <3 <3 <3 <3 <3
External standard calibration value g 500 500 200+100 200 200+100 100
(min. accuracy class) (E2) (E2) (E2) (E2)
Application range (temperature) With “isoCAL” function: +5 … +40°C | Without “isoCAL” function: +15 … +25°C
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 85 + 85 85 + 85 85 + 85 85 + 85 85 + 85 85 + 85
Weighing chamber height (draft shield DU) mm 261 261 261 261 261 261
Protection IP54 in accordance with IEC 60529
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
9
Verified Models with EC Type Approval Certificate: Precision Balances
Model 623S-xCE 623P-xCE 323S-xCE
Accuracy class* mg K K K
For verified models: EC type approval certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1|2|5 1
Weighing capacity max* g 620 150 | 300 | 620 320
Calibration value e* mg 10 10 10
Min. load min* mg 20 20 20
Tare equalization range (subtractive) < 100 % from max. weighing capacity
Application range according to DIR* g 0.02 – 620 0.02 – 620 0.02 – 320
Optimal starting point of the g 0.82 0.82 0.82
operating range**
Typical stabilization time s < 0.8 < 0.8 < 0.8
Typical measurement time s <1 <1 <1
Application range (temperature) With “isoCAL” function: +5 … +40°C | Without “isoCAL” function: +10 … +30°C
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 140 + 140 140 + 140 140 + 140
Weighing chamber height (draft shield DE) mm 172 172 172
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
10
Verified Models with EC Type Approval Certificate: Precision Balances
Model 6202S-xCE 6202P-xCE 5202S-xCE 4202S-xCE
Accuracy class* K K k K
For verified models: EC type approval certificate D09-09-015. Type: MSX
Scale interval d* g 0.01 0.01 | 0.02 | 0.05 0.01 0.01
Weighing capacity max* g 6,200 1,500 | 3,000 | 6,200 5,200 4,200
Calibration value e* g 0.1 0.1 0.1 0.1
Min. load min* g 0.5 0.5 1 0.5
Tare equalization range (subtractive) < 100 % from max. weighing capacity
Application range according to DIR* g 0.5 – 6,200 0.5 – 6,200 1 – 5,200 0.5 – 4,200
Optimal starting point of the g 8.2 8.2 8.2 8.2
operating range**
Typical stabilization time s <1 <1 < 0.8 < 0.8
Typical measurement time s < 1.5 < 1.5 <1 <1
Application range (temperature):
With “isoCAL” function +5 … +40°C +5 … +40°C +5 … +40°C
Without “isoCAL” function +10 … +30°C +10 … +30°C +10 … +30°C
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 206 + 206 206 + 206 206 + 206 206 + 206
Protection IP54 in accordance with IEC 60529
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
11
Verified models with EC Type Approval Certificate: High load balances
Modelle 36201S-0CE 36201P-0CE 20201S-0CE
Accuracy class* K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 100 100 | 1.000 100
Weighing capacity max* g 36,200 10,200 | 36,200 20,200
Calibration value e* g 1 1 1
Min. load min* g 5 5 5
Tare equalization range (subtractive) < 100% from max. weighing capacity
Application range according to DIR* g 5 – 36,200 5 – 36,200 5 – 20,200
Optimal starting point of the g 82 82 82
operating range**
Typical stabilization time s < 1.5 < 1.5 < 1.5
Typical measurement time s <2 <2 <2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result s 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
(depending on the set filter level)
Weighing pan size (W + D) mm 400 + 300 400 + 300 400 + 300
IP protection IP54
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) Chapter 41, the optimal operating range is defined from 820d to maximum weighing capacity.
Depending on the installation location and environmental conditions, the value could be higher.
12
Cubis® Leveling
Ø The Cubis® shows the level indicator on the display and provides support for rapid leveling (as standard on the display and service units
MSA and MSU; on the MSE there are only symbols to support manual leveling).
1 Fully automatic, motorized Q-Level leveling at the touch of a button (available for all Cubis® weighing modules with a weighing capacity
of > 6.1 g and < 6200 g).
13
Cubis® Optional Accessories
* Not available for Precision high capacity models with a weighing capacity of > 20,200 g.
14
Pipette Calibration Hardware and Software
Pipette calibration kit (hardware) for models with 0.1 mg and 0.01 mg readability YCP04MS
Consists of moisture trap and all required adapters
Pipette calibration kit (hardware) for microbalance weighing modules 6.6S and 3.6P VF988
Consists of moisture trap and all required adapters
Pipette Tracker pipette calibration software. Software and user manual in English only. YCP04-PT
Pipette Tracker Pro pipette calibration software, for use in regulated areas, networkable and validatable, YCP04-PTPro
according to the 21 CFR Part 11 regulations. Software and user manual in English only.
Documentation basis for validation (IQ, OQ) of Pipette Tracker PRO version. YCP04-VTK
All documents are in English only.
Special Applications
Density determination kit for solids and liquids for weighing modules with a readability of < 1 mg YDK01MS
Density determination kit for solids and liquids for weighing modules with a readability of 1 mg YDK02MS
Q-Grip, flexible holder for weigh-in containers and filters up to 120 mm diameter (replaces the original weighing pan; YFH01MS
for Cubis® models with 0.01 and 0.1 mg readability)
Q-Grid grid weighing pan for Cubis® models with 10 mg or 100 mg readability for weighing in laboratory hoods, safety weighing YWP03MS
cabinets, or workbenches (smaller areas exposed to draft on the weighing pan; replaces the standard weighing pan)
Weighing Tables
Weighing table made from synthetic stone, with vibration dampening YWT03
Wall console YWT04
Weighing table made from wood with synthetic stone for precise, reliable measurements YWT09
Weighing Accessories
Weighing scoop made from chrome nickel steel, 90 + 32 + 8 mm 641214
Aluminum weighing scoop, 4.5 mg (250 pieces) for ultramicrobalance and microbalance models 6565-250
Aluminum weighing scoop, 52 mg (50 pieces) for ultramicrobalance and microbalance models 6566-50
Support arm for 10 | 100 mg precision weighing modules for raising MSE, MSU, and MSA display and control units YDH01MS
Support arm for precision weighing modules with 100 mg | 1 g readability and weighing capacity > 20 kg YDH02MS
for raising MSE, MSU, and MSA display and control units
Hook for below-balance weighing for precision weighing modules with 100 mg | 1 g readability and weighing capacity 69EA0040
> 20 kg (not for verified models, CE mark)
The brand name and logo for Bluetooth® wireless technology are owned by Bluetooth SIG Inc. The use of this brand name and trademark by
Sartorius AG is under license. Other brand names and trademarks are the property of their respective owners.
15
Balance Dimensions
Ultramicrobalance | Microbalance Control Unit MSA | MSU with E-box
All dimensions are given in millimeters
16
Ultramicrobalance | Microbalance Control Unit MSE with E-box
All dimensions are given in millimeters
17
Ultramicrobalance | Microbalance Weighing Module with DM Draft Shield
All dimensions are given in millimeters
18
Ultramicrobalance | Microbalance Weighing Module with DF Draft Shield
All dimensions are given in millimeters
19
Semi-microbalances with Motorized Draft Shield – Control Unit MSA | MSU with E-box
All dimensions are given in millimeters
20
Semi-microbalances with Motorized Draft Shield – Control Unit MSE with E-box
All dimensions are given in millimeters
240
206
205
250
373
360
125
91
56
209 211
127 217 240
249
211
300 173
404
21
Analytical Balances with Manual DU Draft Shield – Control Unit MSA | MSU
All dimensions are given in millimeters
22
Analytical Balances with a Manual DU Draft Shield – Control Unit MSE
All dimensions are given in millimeters
23
Precision Balances with a Readability of 1 mg and Manual DE Draft Shield – Control Unit MSA | MSU
All dimensions are given in millimeters
24
Precision Balances with a Readability of 1 mg and Manual DE Draft Shield – Control Unit MSE
All dimensions are given in millimeters
25
Precision Balances with a Readability of 1 mg and Framed DR Draft Shield – Control Unit MSA | MSU
All dimensions are given in millimeters
26
Precision Balances with a Readability of 1 mg and Framed DR Draft Shield – Control Unit MSE
All dimensions are given in millimeters
206 206
154 154
121.5
94
127 217 211
209 240
364 249
105
240
265
27
Precision Balances with No Draft Shield – Control Unit MSA | MSU
All dimensions are given in millimeters
28
Precision Balances with No Draft Shield – Control Unit MSE
All dimensions are given in millimeters
29
Precision High Capacity Balances
All dimensions are given in millimeters
±1
126
365 33 265
400
400
16
16
300
±1
407
±1
455
239.5
239.5
30
Precision High Capacity Balances
All dimensions are given in millimeters
±1
126
365
ca. 390
ca. 370
ca. 350
ca. 600
ca. 595
ca. 327
394 383
31
Sartorius Lab Instruments GmbH & Co. KG
Weender Landstrasse 94-108
37075 Goettingen, Germany
Phone +49.551.308.0
Fax +49.551.308.3289
[email protected]
www.sartorius.com
Technical specifications subject to change
without notice.
Printed in the EU on paper bleached without chlorine. | W
Publication No.: W--2025-e140309
Order No.: 98649-011-60
Ver. 03 | 2014