7. Regulatory Requirements To Practice Ct - Scan
7. Regulatory Requirements To Practice Ct - Scan
Effective Date:
Issue History
Issue No. Descriptio Originator Approved By
n of Date
Change
1 Surur Kedir Name Signature
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
Contents
1. Purpose...........................................................................................................................3
2. Scope..............................................................................................................................3
3. Responsible person........................................................................................................3
4. Definitions and Abbreviations.......................................................................................3
4.1 Definitions.......................................................................................................3
4.2 Abbreviations.................................................................................................. 4
5. Work Instruction............................................................................................................4
6. Reference Document......................................................................................................7
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
1. Purpose
This procedure is intended to establish basic/mandatory requirements required for license
in Computer Tomography (CT) Scanning and to assist inspectors to decide uniformly
whether a radiological facility should be licensed, not begin or stop its services by the
power vested to the ETA by proclamation No. 571/2008. This regulatory requirement will
be available and distributed to users beforehand so as to guide and inform them about the
license process to get authorization to practice diagnostic radiological services.
2. Scope
This regulatory requirement applies to Computer Tomography (CT) scanning facilities
practicing diagnostic imaging application with x-ray machine for the purposes of
obtaining diagnostic information of patients as prescribed by appropriate medical
personnel.
3. Responsible person:
DG, DDG, R/N/T/NAC
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
5. Work Instruction
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
Film viewing screen accompanied with the The SD of the CT tomography dose
with uniform light necessary documents number over a 500 index (CTDI)
intensity and color including the service and mm2 region of The CTDI for a
should be available. operating manuals, results interest for water or single slice for each
F. Area of the of acceptance tests, and tissue equivalent available beam
exposure room calibration certificate for material at the center shaping filter and for
Area of the exposure the required machine and around the each available slice
room shall not be less parameters of the x-ray periphery of thickness should not
than 28 m2 for CT machine. phantoms < 1.5% of deviate more than ±
scan. 1. Deviation of kVp the baseline 20% from the
N.B. The exposure Accuracy ≤ ±10% baseline.
room shall have 2. Deviation of the X-ray Documentation
sufficient ventilation timer ≤ ±10% Requirements Irradiated slice
(such as a ventilation 3. Output inconsistency Full documentation thickness
window with a height ≤ 5% shall be presented The FWHM of the
above 2m from 4. The first half value with the application dose profile should
outside ground level layer in the X-ray for Authorization to not differ more than ±
and position in such a beam incident to the the Authority such 20% from baseline.
way that backscatter Patient must be as: High contrast
radiation will not fall greater than or equal Machine resolution (spatial
on windows of to: specifications (with resolution)
opposite building, a) 3.8 millimeters of manufacturer’s The FWHM of the
verandah, etc) aluminum at information, point spread function
G. Wall thickness maximum and of a pin, or the edge
120kVp; or
Wall thickness of at b) 4.2 millimeters of minimum response function of
least 17cm concrete aluminum at parameters that an edge should not
or its equivalent (with 130kVp. could be set, date of differ more than ±
no radiation leakage manufacture, etc) 20% from baseline.
to the surroundings, and shipping Low contrast
i.e. maximum of certificate as basic. resolution
1µSv/hr at all Floor layout of the Polystyrene pins of
locations occupied by facility, 0.35 cm diameter
a member of the Copies of inserted in a uniform
public and 7.5µSv/hr educational body water phantom
to the control cubicle qualifications of should be visible in
and any workers, the image.
occupationally
accessed location by Properly signed and
radiation workers) sealed completed
H. Warning light & notification &
signs application
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE
Logo Company Name: Document No. Effective Date:
6. List of Forms
Ionizing Radiation Notification APNT-01 form OF/ETA/099
Application form to use diagnostic x- ray equipment and facility APDR-01
form OF/ETA/120
7. Reference Document
ETA Management Manual
PLEASE MAKE SURE THAT THIS IS THE CORRECT ISSUE BEFORE USE